Manual Supplement. Manual Title: ESA615 Users Supplement Issue: 2 Print Date February 2012 Date: 2/13 Revision/Date: 1 Page Count: 18

Similar documents
Manual Supplement. Manual Title: ESA615 Users Supplement Issue: 1 Print Date February 2012 Date: 6/12 Revision/Date: 1 Page Count: 9

Manual Supplement. Manual Title: ESA609 Users Supplement Issue: 7 Print Date: February 2014 Date: 2/19 Revision/Date: 1 Page Count: 13

ESA615 Electrical Safety Analyzer

ESA615. Electrical Safety Analyzer. Getting Started Manual

ESA615. Users Manual. Electrical Safety Analyzer

Ansur ESA612. Users Manual. Plug-In

ESA614. Electrical Safety Analyzer. Getting Started Manual

Electrical Safety Analyzers

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Ansur ESA620. Users Manual. Plug-In

ESA601. Electrical Safety Analyzer. Getting Started Manual

LOW WEIGHT, SMALL SIZE,

The low weight, small size, one-touch solution. Portable appliance testing. The Fluke 6000 series PAT testers

Manual Supplement. Manual Title: 1760 Users Supplement Issue: 4 Print Date: June 2006 Issue Date: 5/14 Revision/Date: 2, 5/10 Page Count: 4

Operating Instructions

The low weight, small size, one-touch solution. Portable appliance testing. The Fluke 6000 series PAT testers

Operating Instructions

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Fluke 43B. Power Quality Analyzer. Getting Started. EN June Fluke Corporation, All rights reserved. Printed in The Netherlands.

Manual Supplement. Manual Title: 1760 Users Supplement Issue: 5 Print Date: June 2006 Issue Date: 2/18 Revision/Date: 2, 5/10 Page Count: 5

HYAMP. Quick Start Guide

vpad-mini Electrical Safety Analyzer Operating Manual

rigel 288 Electrical Medical Safety Analyser Key Features Applications

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER KEW6205

Portable Appliance Tester KT75 OPERATORS MANUAL

277 Plus Electrical Medical Safety Analyser

Medical Device Testing Explained

RTD. Process Calibrator. User s MANUAL. PDF created with pdffactory Pro trial version

Advanced Test Equipment Rentals ATEC (2832)

DX4085 Z-Meter Tester

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

BeneHeart R12/BeneHeart R12A Electrocardiograph. Service Manual

5 Performance Verification

Application Note. Use of Palm Switches with QuadTech Hipot Testers

Read the whole of this document before unpacking your Izon equipment or proceeding to the steps below.

Model 8010 High Power Device Test Fixture Interconnection Reference Guide

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

Rigel Medical Product Brochure Medical Test Equipment. rigelmedical.com

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

PRODUCT RANGE MEDICAL TEST EQUIPMENT FROM

ClareHAL 104 Electrical Safety Tester

ES601Plus. International Automated Electrical Safety Analyzer. Operating Manual

Rigel Instruction Manual 331A575 Issue 2.0. Hand-Held Electrical Safety Analyser. October Seaward Group Issue 2.0

Date received into service; / /.

Fluke Connect Modules Technical Data

UL TEST REPORT AND PROCEDURE

Multiple Patient Leakage and Mains on AP Leakage Measurements with the and Scanner

PRIMETEST PRIMETEST. Operating Instructions

Rigel Medical. Leaders in biomedical test solutions. Performance Analyzers. Vital Signs Simulators. Electrical Safety Analyzers

DM3058/DM3058E Digital Multimeter

GT-650/GT-900 User's Manual

AB MSAI Research Labs Pvt. Ltd., 201, IInd Floor, Aggarwal Cyber Plaza, Netaji Subhash Place, Pitampura, Delhi

SECUTEST S2 N+w Test Instrument with Automatic Test Sequences for IEC and Draft IEC 62638

Mini Digital Multimeter

Installation RA5112-C-16A-C20. Basic Rack Power Distribution Unit

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

Leak Current Measurement - Essential to Electrical Safety. Compliance with IEC :2005 Ed 3.0, JIS T :2012

Series Watt DC Power Supplies

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com

Features and Benefits

Micro-Ohmmeters. Digital 10A. Models 6240 & Measure low resistance with high accuracy. Selectable test current and resistance ranges

HYPOT III. Quick Start Guide

Autoranging True RMS Multimeter User Manual

Operating Instructions PRO-AB /6.15. Leakage Current Measuring Adapter

Agilent 66000A 1200 W DC Modular Power System, GPIB

UL TEST REPORT AND PROCEDURE

SECUTEST BASE(10) and PRO Tester for Measuring the Electrical Safety of Devices per VDE , IEC and IEC

Introduction. Chapter 2

A COMPACT GUIDE TO SAFETY TESTING: PROTECTION CLASSES

Shipping/Receiving ATEC (2832) x802 Order Additional Equipment ATEC (2832)

Model 2657A-LIM-3 LO Interconnect Module

PQ-Box 100, 150, 200 & 300 Power Quality Analyser Clamps & Accessories

Multi-Function Ground Resistance Tester Model 6471

UL TEST REPORT AND PROCEDURE

HypotULTRA. Quick Start Guide SAFETY CHECKLIST. Survey the test station. Make sure it is safe & orderly.

SECUTEST SII + Test Instrument for Measurements per BGV A3 and Operational Safety Regulation

Supplement. Manual Title: DSP-4000 Series User Supplement Issue: 3 Print Date: July 2001 Issue Date: 10/02 Revision/Date: Page Count: 4

Visit us at

GammaGT MI 3311 Instruction manual Ver. 2.0, Code no

DALI Professional Controller-4. Control unit Operating instructions

Rigel Electrical Safety Analyser

Modular Power System 1200 W per mainframe GPIB

2200 Series. Quick Start Guide SAFETY CHECKLIST. Safety Made Simple

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

UL TEST REPORT AND PROCEDURE

User Guide True RMS Multimeter Extech EX205T

MPM-U653-DCF-B M/N: MPM-U653-DCF-B. 650W Medical AC / DC SPECIFICATION FOR SWITCHING POWER SUPPLY. Revision History

Central Offi ce Connector Block Accessories

Pro Logger II. Designed and manufactured in Australia.

Model 7705 Control Module

LIFEPAK CR PLUS/EXPRESS Defibrillator TABLE OF CONTENTS

METREL test and measurement accessories: 3-phase AktivGT / Machine adapter A1322 Instruction manual Version 1.0, Code no.

UL TEST REPORT AND PROCEDURE

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Operating Instructions

Installation Instructions Model LFMWLT450, Class 1 Series AC-DC Power Supply

OPERATING INSTRUCTION

Wa r n i n g-wa r n i n g-wa r n i n g

PROFITEST INSTALLATION

Mini Digital Multimeter

Transcription:

Manual Title: Supplement Issue: 2 Print Date February 2012 Date: 2/13 Revision/Date: 1 Page Count: 18 This supplement contains information necessary to ensure the accuracy of the above manual. This manual is distributed as an electronic manual on the following CD-ROM: CD Title: ESA615 CD Rev. & Date: 3/2012 CD PN: 4105850 2012 2013 Fluke Corporation. All rights reserved. Printed in the U.S.A.

Manual Supplement Change #1, 62104, 62223 On page 19, prior to Mains Voltage Test, add the following Note: Note The standard set in the setup menu applies to all manual measurement modes. The Automated tests use the standard set for the selected test sequence. On page 61, below the first paragraph add the following Note: Note The product can show the last 200 test results. All results saved on the instrument can be exported to a PC. 2/13 1

On page 62, replace Table 6 with: Table 6. Factory Supplied Test Sequences Test Sequence Description [1] 60601 3rd Patient Monitor 3 rd Ed., Class I, 5 ECG 60601 3rd Defibrillator 3 rd Ed., Class I, 2 Paddles and 3 ECG 60601 3rd Infusion Device 3 rd Ed., Class II, 1 No AP 60601 3rd Ultrasound Device 3 rd Ed., Class I, 1 Probe 60601 3rd Generic Device 3 rd Ed., Class I, No AP 60601 3rd System 3 rd Ed., Class I, No AP 62353-Alt. Patient Monitor 62353-Alt. Defibrillator 62353-Alt. Infusion Device 62353-Alt. Ultrasound Device 62353-Alt. Generic Device NFPA99 Patient Monitor Class I, 5 ECG Class I, 2 Paddles and 3 ECG Class II, 1 No-AP Class I, 1 Probe Class I, No AP Class I, 5 ECG 2 2/13

Manual Supplement Table 6. Factory Supplied Test Sequences (cont.) Test Sequence Description [1] NFPA99 Defibrillator NFPA99 Infusion Device NFPA99 Ultrasound Device NFPA99 Generic Device ANSI/AAMI ES-1 Patient Monitor ANSI/AAMI ES-1 Defibrillator ANSI/AAMI ES-1 Infusion Device ANSI/AAMI ES-1 Ultrasound Device Class I, 2 Paddles and 3 ECG Class II, 1 No AP Class I, 1 Probe Class I, No AP Class I, 5 ECG Class I, 2 Paddles and 3 ECG Class II, 1 No AP Class I, 1 Probe ANSI/AAMI ES-1 Generic Device Class I, No AP [1] Class designations refer to applicable electrical safety standard definitions for devices, not FDA medical device definitions. On page 64, under the second step 3, after the 3rd paragraph add the following Note: Note A USB keyboard or barcode reader can be used to type in the A.P. name. 2/13 3

On page 65, following step 5 add: Note A USB keyboard or barcode reader can be used to edit the test sequence name. On page 66, replace Table 7 with and add a note after the Table: Table 7. Test Settings for Test Sequences Test Setting Pause after power on Pause before power off Power on delay Power off delay Description If set to yes, delays the start of the test by the time set in the Power on delay parameter when power is applied to the DUT. If set to yes, delays the start of the test by the time set in the Power off delay parameter when power is removed from the DUT. The time the Product will wait before it does the subsequent step after power is applied to the DUT. The range is 0 to 9999 seconds. The time the Product will wait before it does the subsequent step after power is removed from the DUT. The range is 0 to 9999 seconds. Default Value No No 2 sec 0 sec Test Speed When set to Normal, the Product does a leakage current measurement in 5 Normal 4 2/13

Manual Supplement Test Mode seconds and an insulation resistance test in 1 minute. When set to Rapid, the Product measures leakage current as fast as possible and an insulation resistance test in 3 seconds. When set to Automatic, the Product does each test sequence step automatically. When set to Step-by-Step, you must push the Next Step softkey to move to the subsequent step in the sequence. Automatic Halt on Test Failure When set to Yes, the Product stops the test sequence when a fault is sensed. Yes Multiple PE Tests If set to Yes, the Product prompts you to repeat PE tests or continue. No Multiple Non-Earth Leakage Patient Lead Records storage [1] If set to Yes, the Product prompts you to repeat non-earth leakage tests or continue. When set to Store all, all results are put into the test results. If set to Store worst/last, only the worst measured value is put into the test results. If set to store worst/last and all measured values are the same, then the last value is put into the test results. No Store worst/last Insulation Test Voltage Sets the insulation test voltage to 250 V dc or 500 V dc. 500 V dc Reverse Polarity If set to No, reverse polarity tests will be skipped. Yes [1] Applies to leakage current tests only. 2/13 5

Note You can change the Test Speed parameter on a test sequence to decrease the time necessary to do a test sequence. On page 68, following step 6 add the following Note: Note A USB keyboard or barcode reader can be used to edit the test sequence name. On page 69, following step 3, replace the Note with: Note Push softkey F1 to toggle between an A to Z, Z to A or by date sort of the test sequences. On page 69, replace step 5 with: 5. Push the next Next softkey. On page 69, at the bottom of the page add the following Note: A USB keyboard or barcode reader can be used to type in the Device Under Test information. You can use a barcode reader to automatically move through each field if it automatically sends a carriage return. Note 6 2/13

Manual Supplement On page 70, replace the third paragraph with the following: To add comments when a test step ends, push the Add Comments softkey. When you see individual results of a Test Summary, you can push the Add Notes softkey to type in more information about the test step. On page 70, following step 6 add the following Note: Note A USB keyboard or barcode reader can be used to edit the test results name. On page 76, under Specifications, replace the Altitude and Power with: Altitude 100 V/115 V ac mains supply voltage. 5000 m 230 V ac mains supply voltage... 2000 m Power 100 V/115 V power outlet... 90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum 230 V power outlet... 180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum 2/13 7

On page 77, under Detailed Specifications, replace the Voltage section with: Voltage Ranges (Mains voltage)... 90.0 V to 132.0 V ac rms 180.0 V to 264.0 V ac rms Range (Point-to-point voltage) 5000 m... 0.0 V to 150 V ac rms 2000 m... 0.0 V to 300.0 V ac rms Accuracy... (2 % of reading + 0.2 V) 8 2/13

Manual Supplement On page 80, replace Factory-Supplied Test Sequences with: Factory-Supplied Test Sequences 60601 3 rd Edition Patient Monitor 60601 3 rd Edition Defibrillator 60601 3 rd Edition Infusion Pump 60601 3 rd Edition Ultrasound Device 60601 3 rd Edition Generic Device 60601 3 rd Edition System 62353-Alt. Patient Monitor 62353-Alt. Defibrillator 62353-Alt. Infusion Pump 62353-Alt. Ultrasound Device 62353-Alt Generic Device NFPA99 Patient Monitor NFPA99 Defibrillator NFPA99 Infusion Pump NFPA99 Ultrasound Device NFPA99 Generic Device ANSI/AAMI ES-1 Patient Monitor ANSI/AAMI ES-1 Defibrillator ANSI/AAMI ES-1 Infusion Pump ANSI/AAMI ES-1 Ultrasound Device ANSI/AAMI ES-1 Generic Device 2/13 9

Change #2 On page 12, replace Figure 5 with: 10 2/13

Manual Supplement Connect ESA615 to grounded mains socket. To protective earth or any exposed conductive surface on the enclosure RA LL LA RL V1 R F L N C1 Connect the DUT AC power cord to the equipment outlet on the Analyzer gtv113.eps Figure 5. DUT Connections to the Product 2/13 11

On page 14, replace Figure 7 and add the How to connect a PC to the Product: How to Connect a PC to the Product To connect the Product to a PC: Connect a USB port on your PC or laptop to the Mini B USB device port on the Product. Or Plug in an XStick wireless USB dongle to your PC USB port. Products available over wireless will be listed by serial number. Connect to a single Product. See Figure 7. 12 2/13

RA LL LA RL V1 R F L N C1 Manual Supplement RA LL LA RL V1 R F L N C1 Figure 7. Product to PC Connection gtv129.eps 2/13 13

On page 22, replace Figure 11 with: To protective earth or any exposed conductive surface on the enclosure RA LL LA RL V1 R F L N C1 Connect the DUT AC power cord to the equipment outlet on the Analyzer gtv112.eps Figure 11. Ground Wire (Protective Earth) Resistance Measurement Connections 14 2/13

Manual Supplement On page 53, replace How to Use the 1210 Adapter with: How to Use the 1-to-10 Adapter The 1-to-10 Adapter, an optional accessory, increases the number of lead or applied parts connections to the Product from 5 to 14. The adapter connects a maximum of 10 leads together into a one lead that is connected to one of the input jacks of the Product. The other four Product input jacks can also be used in conjunction with the Adapter. More leads can be added with multiple 1-to-10 Adapters. The example in Figure shows one application of the Adapter. The Defibrillator/Monitor in the example has 10 ECG leads, two pacer leads, and two defibrillator paddles which must be connected together, and in groups if single function, for current leakage per IEC62353. The example shows the ECG leads to be snap type connectors and two BJ2ECG adapters are shown connected to the Adapter. If the ECG leads do not have snap connectors, then the Universal Snap to Banana Adapter can be used to make the connections to the Adapter. The common lead from the Adapter is plugged into the RA jack (1st jack) of the Product. Using four sheathed test leads with alligator clips, connect the two defibrillator paddles into the LL and LA Product jacks and the two pacer leads into the RL and V1 jacks. Set the connection that ties all five Product jacks together. This will measure leakage current in all fourteen leads. The applied part group of 1, 2, and 2 lets you test groups of applied parts of one function. On page 54, replace Figure 31 with: 2/13 15

Defibrillator/Pacer DEFIB OFF PACER Caution: To prevent damage to the Product, do not apply more than 30 V dc to the AP (ECG). Check unkown connections using a DMM before connecting to ESA615. ESA615 Alligator Clips 1-to-10 and BJ2ECG snap adapters Use leads shorted for electrical safety testing only. Do not short leads for ECG simulation. gtv120.eps Figure 31. 1 to 10 Adapter Connections 16 2/13

Manual Supplement On page 56, replace Figure 32 with: 2/13 17

ECG Monitor ESA615 1-to-10 and BJ2ECG snap adapters Use leads shorted for electrical safety testing only. Do not short leads for ECG simulation. gtv121.eps Figure 32. ECG Lead Connection with 1 to 10 Adapter 18 2/13