DICOM Conformance Statement

Similar documents
DICOM Conformance Statement

DICOM Conformance Statement

MediaWorkStation. MWS DICOM Push Interface. DICOM Conformance Statement. Version 2.0. English

DICOM Conformance Statement

DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number:

KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement

DICOM Conformance Statement for GALILEI. Software Version V6.0

DICOM Conformance Statement

KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR /2017/PI-E 1/32

DICOM Conformance Statement

DICOM Conformance Statement AIDA HD Connect

DICOM Conformance Statement June 24, 2011 Version 0.4 ImageWorks Internal document number:

Uscan. DICOM Conformance Statement

DICOM Conformance Statement VISIQ 1.0

Technical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG

SonoWin Picture Archiving System

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG

Lumify 1.8.x DICOM Conformance Statement

DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT STORAGE SCU FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-053A or XIDF-056A (MIIXR0001EAB)

DICOM Conformance Statement. DSI Release 4.4. Document Number June Copyright Philips Medical Systems Nederland B.V.

DICOM Conformance Statement

Topcon Medical Systems

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT

MediaWorkStation. DICOM Worklist Interface. DICOM Conformance Statement. Version 1.1. English

REGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version).

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

FluoroPRO. DICOM conformance Statement

QAngio XA 7.0. DICOM Conformance Statement

DAR-9400f DICOM Conformance Statement Revision A

Villa Sistemi Medicali DRF 4343

DICOM Conformance Statement for IMAGEnet 5

DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE)

Endo Mail. DICOM Conformance Statement KARL STORZ AIDA compact II Version 2.2. KARL STORZ GmbH & Co. KG

Endo Mail. DICOM Conformance Statement KARL STORZ AIDA Version 2.3. KARL STORZ GmbH & Co. KG

CMT MEDICAL TECHNOLOGIES

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T

DICOM Conformance Statement

DICOM Conformance Statement

Kodak Point of Care CR systems. DICOM Conformance Statement

DICOM Conformance Statement

DICOM. Conformance Statement For VINNO Medical Ultrasound System Revision 2.0. Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd.

DICOM CONFORMANCE STATEMENT

DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

Atlas-Based Autosegmentation. DICOM Conformance Statement DICOM 3.0

DAR-9400f/DAR-9500f with Gateway Option DICOM Conformance Statement. Revision D

DICOM Conformance Statement

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage

Technical Publications

DICOM Conformance Statement. report 42. Version 3.2. Reporting Application for Cardiovascular MR and CT

DICOM CONFORMANCE STATEMENT

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2

Fluoroscan InSight V5.0.6 DICOM Conformance Statement

DICOM Conformance Statement

DAR-8000 Series DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR DP10/DP20/DP30 SERIES DIGITAL ULTRASONIC DIAGNOSTIC IMAGING SYSTEM

DICOM Conformance Statement. Allura Centron 1.0

DICOM Conformance Statement. ECG PageWriter TC A.06.01

Technical Publications

DICOM CONFORMANCE STATEMENT for Keratograph ( )

This conformance statement specifies the compliance of EponaTech s Metron-DVM to the DICOM standard Version 3.0.

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003

Whole Body X-ray CT System Supria. DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR Diagnostic Ultrasound System

DICOM Conformance Statement. DPI DICOM Conformance Statement. Document Number: Rev. A (New Platform) 1/6/2006 JBC

Lumify DICOM Conformance Statement

S517-E118 Dec DICOM Conformance Statement

Video Plus Series DICOM Conformance Statement

DICOM Conformance Statement for FLEXAVISION HB/FD

DICOM 3.0 Conformance Statement for PlatinumOne Systems

Technical Publication. DICOM Conformance Statement. iplan RT Dose 4.1. Document Revision 1. December 4, Copyright BrainLAB AG

DICOM Conformance Statement

Sonovision DICOM Conformance Statement

Retinal imaging control software NM. DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR DIGITAL RADIOGRAPHY SYSTEM. MODEL Winscope Plessart TM PDR-01A

DICOM CONFORMANCE STATEMENT

DICOM Conformance Statement. Forum

Technical Publications

C-View 1.0. DICOM Conformance Statement. MAN Rev 001

DICOM Conformance Statement

DICOM Conformance Statement FORUM

CONFORMANCE STATEMENT

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM Conformance Statement

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1

Technical Publications

Artwork consists of twenty three (23) 8½ x 11 inch pages.

DAR Revision Date: July,

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

DICOM Conformance Statement

No. : S Rev. : A. DICOM Conformance Statement For Side Station Rev02.00

Technical Publications

DICOM Conformance Statement for DR-200/DRU-200

DICOM Conformance Statement

DICOM Conformance Statement FORUM

SIEMENS. DICOM Conformance Statement

Transcription:

DICOM Conformance Statement Application Annex: AneurysmFlow R1.1 On Interventional Workspot R1.5 Koninklijke Philips N.V. 2018 All rights are reserved.

Issued by: Philips Medical Systems Nederland BV, a Philips Healthcare company, P.O. Box 10.000 5680 DA Best The Netherlands Internet: www.philips.com/dicom Document Number: ICAP-PF.0034850 Date: 19-Sep-2018

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 3 of 12 Table of Contents 1. INTRODUCTION... 4 1.1. REVISION HISTORY... 4 1.2. TERMINOLOGY... 4 2. DATA SPECIFICATIONS... 5 2.1. SUPPORTED IOD S... 5 2.1.1. Acceptance Criteria... 5 2.1.2. Contents of Created IOD s... 5 2.1.2.1. List of created SOP Classes... 6 2.1.2.2. X-Ray Angiographic Image Storage SOP Class... 6 2.1.2.3. Secondary Capture Image Storage SOP class... 10

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 4 of 12 1. Introduction This DICOM Conformance Statement annex is applicable to the AneurysmFlow R1.0 for Interventional Workspot hosting platform, later referred to as AneurysmFlow AneurysmFlow is a software tool intended to provide relevant information to the intervention list during cerebral aneurysm embolization treatment, based on quantification of blood flow changes. AneurysmFlow is a software medical device and is intended to be used in combination with a Philips interventional X-ray system and 3DRA AneurysmFlow is a software product (Interventional Tool) that provides color coded and vector field representation of a digital subtraction angiography (DSA). It can quantify blood flow rates in the artery based on DSA and 3DRA. It can visualize blood flow patterns in an aneurysm based on DSA. It can also provide a side by side visual and quantitative comparison between two acquisitions. 1.1. Revision History The revision history below provides dates and differences among individual document versions. Table 1: Revision History Document Version Date of Issue Status Description 00 19-Sep-2018 Authorized Final Version for AneurysmFlow R1.1 on Interventional Workspot R1.5 1.2. Terminology DICOM Digital Imaging and Communications in Medicine IOD Information Object Definition UID Unique Identifier VR Value Representation DSA Digital Subtraction Angiography MAFA Mean Aneurysm Flow Amplitude

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 5 of 12 2. Data Specifications 2.1. Supported IOD s This section specifies each IOD accepted and / or created by AneurysmFlow. ACCEPTED CREATED The applicable IOD is accepted for storage in the repository of the hosting platform and supported for import in AneurysmFlow 1.1 for viewing and analysis. The AneurysmFlow 1.1 supports generation of derived by using the applicable IOD and is able to store this in the repository of the hosting platform. Table 2: Supported IOD s IOD Support Name UID ACCEPTED CREATED X-Ray Angiographic Image Storage SOP Class 1.2.840.10008.5.1.4.1.1.12.1 Yes Yes Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 No Yes 2.1.1. Acceptance Criteria This section specifies the acceptance criteria applied by AneurysmFlow R1.1 to which a set should adhere before it can be imported into the application. This can be criteria on the highest level (e.g. from a certain manufacturer or system model) or certain DICOM attributes mandatory to be present into the set holding a specific value. In case one or more Philips private attributes are required, then a list of supported Philips system models will be mentioned. Table 3: Accepted system models Manufacturer Modality System Model Name(s) Not applicable Not applicable Not applicable X-Ray Angiographic Image Storage SOP Class IOD Table 4: Accepted transfer syntaxes per IOD Transfer Syntax Name UID Name UID 1.2.840.10008.5.1.4.1.1.12.1 Implicit VR Little Endian Explicit VR Big Endian Explicit VR Little Endian JPEG 2000 Image Compression JPEG 2000 Image Compression (Lossless Only) JPEG Baseline (Process 1) JPEG Extended (Process 2 & 4) JPEG Lossless, Non-Hierarchical, FOP (Process 14) RLE Lossless 1.2.840.10008.1.2 1.2.840.10008.1.2.2 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.91 1.2.840.10008.1.2.4.90 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.51 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.5 Table 5: Accepted attribute values Attribute Name Attribute Number Values / Comments Not applicable Not applicable Not applicable 2.1.2. Contents of Created IOD s This section specifies in detail the attribute contents of created objects. Attributes are grouped together by its corresponding module as specified by DICOM standard. Philips private attributes are excluded for specification.

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 6 of 12 Abbreviations used in the Module table for the column "Presence of Value" are: The attribute is always present with a value EMPTY The attribute is always present without any value (attribute sent zero length) VNAP The attribute is always present and its Value is Not Always Present (attribute sent zero length if no value is present) ANAP The attribute is present under specified condition if present then it will always have a value The abbreviations used in the Module table for the column "Source" are: AUTO The attribute value is generated automatically CONFIG The attribute value source is a configurable parameter COPY The attribute value source is another SOP instance FIXED The attribute value is hard-coded in the application IMPLICIT The attribute value source is a user-implicit setting MPPS The attribute value is the same as that use for Modality Performed Procedure Step MWL The attribute value source is a Modality Worklist USER The attribute value source is explicit user input 2.1.2.1. List of created SOP Classes Table 6: List of created SOP Classes SOP Class Name SOP Class UID X-Ray Angiographic Image Storage SOP Class 1.2.840.10008.5.1.4.1.1.12.1 Secondary Capture Image Storage SOP Class 1.2.840.10008.5.1.4.1.1.7 2.1.2.2. X-Ray Angiographic Image Storage SOP Class Table 7: IOD of Created X-Ray Angiographic Image Storage Instances Information Entity Module Presence Of Module Patient Patient Module Study General Study Module Series General Series Module Equipment General Equipment Module Image General Image Module Image Pixel Module Cine Module Multi-Frame Module Display Shutter Module X-Ray Image Module X-Ray Acquisition Module X-Ray Table Module XA Positioner Module DX Detector Module VOI LUT Module SOP Common Module Table 8: Patient Module Patient's Name 0010,0010 PN VNAP Patient ID 0010,0020 LO VNAP Patient's Birth Date 0010,0030 DA VNAP Patient's Sex 0010,0040 CS VNAP

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 7 of 12 Table 9: General Study Module Study Date 0008,0020 DA VNAP Study Time 0008,0030 TM VNAP Accession Number 0008,0050 SH VNAP Referring Physician's Name 0008,0090 PN VNAP Study Instance UID 0020,000D UI Study ID 0020,0010 SH VNAP Table 10: General Series Module Series Date 0008,0021 DA ANAP Series Time 0008,0031 TM ANAP Modality 0008,0060 CS Series Description 0008,103E LO ANAP Performing Physician's Name 0008,1050 PN ANAP Related Series Sequence 0008,1250 SQ ANAP >Study Instance UID 0020,000D UI >Series Instance UID 0020,000E UI >Purpose of Reference Code Sequence 0040,A170 SQ VNAP Series Instance UID 0020,000E UI Series Number 0020,0011 IS VNAP Performed Procedure Step Start Date 0040,0244 DA ANAP Performed Procedure Step Start Time 0040,0245 TM ANAP Performed Procedure Step ID 0040,0253 SH ANAP Table 11: General Equipment Module Manufacturer 0008,0070 LO Philips FIXED Institution Name 0008,0080 LO VNAP COPY copied from source Station Name 0008,1010 SH Manufacturer's Model Name 0008,1090 LO Interventional Workspot Device Serial Number 0018,1000 CONFIG Software Versions 0018,1020 LO 1.5.x CONFIG where x is the detailed application SW version.

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 8 of 12 Table 12: General Image Module Image Type 0008,0008 CS ANAP Acquisition Date 0008,0022 DA ANAP Content Date 0008,0023 DA VNAP Acquisition Time 0008,0032 TM ANAP Content Time 0008,0033 TM VNAP Instance Number 0020,0013 IS VNAP Patient Orientation 0020,0020 CS ANAP Lossy Image Compression 0028,2110 CS ANAP Icon Image Sequence 0088,0200 SQ ANAP >Samples per Pixel 0028,0002 US >Photometric Interpretation 0028,0004 CS >Rows 0028,0010 US >Columns 0028,0011 US >Bits Allocated 0028,0100 US >Bits Stored 0028,0101 US >High Bit 0028,0102 US >Pixel Representation 0028,0103 US >Pixel Data 7FE0,0010 OW/OB ANAP Table 13: Image Pixel Module Samples per Pixel 0028,0002 US Photometric Interpretation 0028,0004 CS Rows 0028,0010 US Columns 0028,0011 US Bits Allocated 0028,0100 US Bits Stored 0028,0101 US High Bit 0028,0102 US Pixel Representation 0028,0103 US Pixel Data 7FE0,0010 OB/OW VNAP Table 14: Cine Module Cine Rate 0018,0040 IS ANAP Frame Time 0018,1063 DS Table 15: Multi-Frame Module Number of Frames 0028,0008 IS Frame Increment Pointer 0028,0009 AT

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 9 of 12 Table 16: Display Shutter Module Shutter Shape 0018,1600 CS Shutter Left Vertical Edge 0018,1602 IS Shutter Right Vertical Edge 0018,1604 IS Shutter Upper Horizontal Edge 0018,1606 IS Shutter Lower Horizontal Edge 0018,1608 IS Table 17: X-Ray Image Module Image Type 0008,0008 CS Samples per Pixel 0028,0002 US Photometric Interpretation 0028,0004 CS Frame Increment Pointer 0028,0009 AT Bits Allocated 0028,0100 US Bits Stored 0028,0101 US High Bit 0028,0102 US Pixel Representation 0028,0103 US Pixel Intensity Relationship 0028,1040 CS Lossy Image Compression 0028,2110 CS Table 18: X-Ray Acquisition Module KVP 0018,0060 DS VNAP Exposure Time 0018,1150 IS VNAP Radiation Setting 0018,1155 CS Imager Pixel Spacing 0018,1164 DS ANAP Pixel Spacing 0028,0030 DS Table 19: X-Ray Table Module Table Motion 0018,1134 DS VNAP Table Angle 0018,1138 DS ANAP Table 20: XA Positioner Module Distance Source to Patient 0018,1111 DS ANAP Distance Source to Patient 0018,1111 DS ANAP Positioner Motion 0018,1500 CS VNAP Positioner Primary Angle 0018,1510 DS VNAP Positioner Secondary Angle 0018,1511 DS VNAP Positioner Primary Angle Increment 0018,1520 DS VNAP

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 10 of 12 Positioner Secondary Angle Increment 0018,1521 DS VNAP Table 21: DX Detector Module Imager Pixel Spacing 0018,1164 DS Pixel Spacing 0028,0030 DS Table 22: VOI LUT Module Window Center 0028,1050 DS Window Width 0028,1051 DS Table 23: SOP Common Module Instance Creation Date 0008,0012 DA ANAP Instance Creation Time 0008,0013 TM ANAP SOP Class UID 0008,0016 UI 1.2.840.10008.5.1.4.1.1. 12.1 SOP Instance UID 0008,0018 UI Instance Number 0020,0013 IS ANAP 2.1.2.3. Secondary Capture Image Storage SOP class Table 24: IOD of Created Secondary Capture Image Storage SOP Class Instances Information Entity Module Presence Of Module Patient Patient Module Study General Study Module Series General Series Module Equipment General Equipment Module CONDITIONAL SC Equipment Module Image General Image Module Image Pixel Module SOP Common Module Table 25: Patient Module Patient's Name 0010,0010 PN COPY copied from source Patient ID 0010,0020 LO COPY copied from source Patient's Birth Date 0010,0030 DA COPY copied from source Patient's Sex 0010,0040 CS COPY copied from source

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 11 of 12 Table 26: General Study Module Study Date 0008,0020 DA COPY copied from source Study Time 0008,0030 TM COPY copied from source Accession Number 0008,0050 SH VNAP Referring Physician's Name 0008,0090 PN VNAP COPY copied from source Study Instance UID 0020,000D UI COPY copied from source Study ID 0020,0010 SH COPY copied from source Table 27: General Series Module Series Date 0008,0021 DA COPY copied from source Series Time 0008,0031 TM COPY copied from source Modality 0008,0060 CS COPY copied from source Series Instance UID 0020,000E UI COPY copied from source Series Number 0020,0011 IS VNAP COPY copied from source Related Series Sequence 0008,1250 SQ VNAP >Study Instance UID 0020,000D UI >Series Instance UID 0020,000E UI >Purpose of Reference Code Sequence 0040,A170 SQ EMPTY Table 28: General Equipment Module Manufacturer 0008,0070 LO Philips FIXED Institution Name 0008,0080 LO VNAP COPY copied from source Manufacturer's Model Name 0008,1090 LO Interventional Workspot Device Serial Number 0018,1000 ANAP CONFIG Software Versions 0018,1020 LO 1.5.x CONFIG where x is the detailed application SW version. Table 29 : SC Equipment Module Modality 0008,0060 CS ANAP Conversion Type 0008,0064 CS WSD

DICOM Conformance Statement application annex: AneurysmFlow R1.1 on Interventional Workspot R1.5 Page 12 of 12 Table 30: General Image Module Instance Number 0020,0013 IS Patient Orientation 0020,0020 CS VNAP Table 31: Image Pixel Module Samples per Pixel 0028,0002 US Photometric Interpretation 0028,0004 CS Planar Configuration 0028,0006 US Rows 0028,0010 US COPY copied from source Columns 0028,0011 US COPY copied from source Bits Allocated 0028,0100 US 8 Bits Stored 0028,0101 US 8 High Bit 0028,0102 US 7 Pixel Representation 0028,0103 US 0000 COPY copied from source Pixel Data 7FE0,0010 OW/OB COPY copied from source Table 32: SOP Common Module SOP Class UID 0008,0016 UI 1.2.840.10008.5.1.4.1.1.7 FIXED SOP Instance UID 0008,0018 UI COPY copied from source Instance Creation Date 0008,0012 DA ANAP COPY copied from source Instance Creation Time 0008,0013 TM ANAP COPY copied from source Instance Number 0020,0013 IS ANAP COPY copied from source