Accuracy study. Elekta Synergy S. High precision radiation therapy using Elekta Synergy S. UMC, Utrecht, Netherlands. Institution: Purpose:

Similar documents
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Leksell SurgiPlan. Powerful planning for success

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

INSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

Comprehensive treatment planning for brachytherapy. Advanced planning made easy

Calibration of Video Cameras to the Coordinate System of a Radiation Therapy Treatment Machine

The MSKCC Approach to IMRT. Outline

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

Evaluation of the Elekta Synergy concept for patient positioning in image guided radiotherapy

Raising the Bar in IMRT QA

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Photon beam dose distributions in 2D

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Determination of rotations in three dimensions using two-dimensional portal image registration

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

Basic Radiation Oncology Physics

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

FAST, precise. qa software

Using a research real-time control interface to go beyond dynamic MLC tracking

Quick Reference Datasheet For All RIT113 Packages

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

Evaluation report. X-ray tomographic image guided radiotherapy systems CEP10071

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Automated Quality Assurance for Image-Guided Radiation Therapy

Clinical implementation of photon beam flatness measurements to verify beam quality

1. Learn to incorporate QA for surface imaging

Table of Contents. About AQUA

Dynalog data tool for IMRT plan verification

Volume Interaction Techniques in the Virtual Simulation of Radiotherapy Treatment Planning

Chapter 9 Field Shaping: Scanning Beam

S. Guru Prasad, Ph.D., DABR

TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:

Advanced Radiotherapy

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

PET-CT in Radiation Treatment Planning

Protocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070

enal++: a new and effective off-line correction protocol for rotational setup errors when using a robotic couch

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear

ADVANCING CANCER TREATMENT

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

Calibration and quality assurance for rounded leaf-end MLC systems

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Tumor motion during liver SBRT

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

The Dose Junction Issue Associated with Photon Beams for Large Volume Radiation Therapy and the Sensitivity to Set-up Error

Oncentra Brachy. Anatomy-based treatment planning for HDR/PDR brachytherapy

Feasibility study of performing IGRT system daily QA using a commercial QA device

ADVANCING CANCER TREATMENT

Performance and characteristics of an IR localizing system for radiation therapy

If it matters to you, it matters to us

Lucy Phantom MR Grid Evaluation

SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR

Introduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Code of practice: Create a verification plan for OCTAVIUS Detector 729 in Philips Pinnacle³

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

Image Guidance and Beam Level Imaging in Digital Linacs

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

8/3/2016. Image Guidance Technologies. Introduction. Outline

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Version 5.6. Quick Start Guide

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

TOMOHD. Product Specifications

A NEW APPROACH IN THE DETERMINATION OF PATIENT S REFERENCE POINT IN CONFORMAL EXTERNAL RADIOTHERAPY

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Estimating 3D Respiratory Motion from Orbiting Views

CARS 2008 Computer Assisted Radiology and Surgery

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

An approach for measuring the spatial orientations of a computed-tomography simulation system

Evaluation of AutoQA Lite TM Image Quality Measurement Software

LEAD LOCALIZATION. Version 1.0. Software User Guide Revision 1.1. Copyright 2018, Brainlab AG Germany. All rights reserved.

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

Quality assurance tests for the Gamma Knife Icon image guidance system

An algorithm for stereotactic localization by computed tomography or magnetic resonance imaging

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

Technical note: Using k-means clustering to determine the number and position of isocenters in MLC-based multiple target intracranial radiosurgery

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

LUP. Lund University Publications. Institutional Repository of Lund University

An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking

Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016

Transcription:

S T E R E O T A C T I C R A D I A T I O N T H E R A P Y Elekta Synergy S High precision radiation therapy using Elekta Synergy S Institution: UMC, Utrecht, Netherlands Purpose: The primary application of Elekta Synergy S at UMC, Utrecht, is frameless intracranial high precision radiation therapy, using multi or single fraction, static or arc techniques. A convergent beam irradiation (3) technique is used to treat patients on this machine, using XVI and 3D VolumeView TM imaging to verify the patient position, correct the table position and then reconfirm the new position. Accuracy study

High precision radiation therapy using Elekta Synergy S Elekta Synergy S commissioning team Marjolein Baarda Gijsbert Bol Minique Boere Arther van Bruggen Corine van Es Fred Groen Mark Harms Erick Kouwenhoven Stephen Kwa Rogier Schokker Theo van Soest Ties Timmers Gitta van Vliet Eric Westzaan Hans Welleweerd Andre Wopereis Acceptance tests were performed on the following items to ensure the suitability of Elekta Synergy S for the job in hand. Mechanical/geometrical accuracy tolerance 0.5mm Field definition: strip test/calibration Rotational accuracy of diaphragm/collimator, gantry and table Sources of errors such as mechanical sag or misalignments are addressed. Method and results Beam modulator TM Strip test (1) measurements have been performed on Beam Modulator TM every week since installation to assess the accuracy of the leaf position. So far there has been no structural deviation recorded beyond our tolerance of 0.5mm for both relative positioning error (RPE) and absolute positioning error (APE), consequently no adjustments have been required. Figure 1: (above) set-up of the strip test. (far left) absolute positioning error (in 0.1mm): top average for every leaf pair bottom average for every abutment. relative positioning error (in 0.1mm): top average for every leaf pair bottom average for every abutment.

Collimator rotation For acceptance of the diaphragm rotation a film was placed perpendicular to the beam axis on the table. A fieldsize of 0.8 x 20cm was set and 12 beams were irradiated at 15 increments. Figure 2: (far left) a diaphragm rotation film circular profile of the 12 beams (24 peaks) This was performed with the gantry at 0, 90 and 180. The film was scanned into the software at high resolution (0.2mm). The beam axes of the beam were reconstructed using a circular profile (2). The isocenter here is defined as the point with the smallest maximum distance to any beam axis. The isocenter is determined by minimizing this maximum distance with movement of the isocenter. The diaphragm rotation error is the maximum distance found: every beam-axis should hit the tolerance circle. Figure 3: collimator rotation 10MV gantry 0: dev. 0.08mm Gantry rotation To measure the accuracy of the gantry rotation, a film was placed in the trans axial position. A fieldsize of 0.8 x 20 was set and six beams were irradiated at 30 increments from 15 to 165, and a further six beams from 210 to 360. This was done with the diaphragm at 90 and rotation clockwise and, with the diaphragm at 270 and rotation counter-clockwise. Figure 4: (far left) Spoke film with gantry rotation Gantry rotation ccw. 10MV coll. 270: dev. 0.38mm

Table rotation A similar test was performed to ensure accuracy of the table position; a film was placed in the coronal position. Table rotations from 90 to +75 were used. To determine the AB error the test was repeated at gantry angles 0 and 180 and to determine the GT error the test was repeated at 90 and 270. Figure 5: (top) table rotation 6MV GT gantry 90/270 : dev. 0.22mm (bottom) table rotation 6MV AB gantry 0/180 : dev. 0.36mm

Gantry sag Gantry sag can also be a source of error and the extent of this required investigation. A film was placed on the table in the coronal plane and five strips of 0.8 x 10cm irradiated (extending to, but not over, the center), firstly with the gantry at 0 and then with the gantry at 180. The average distance of the corresponding beam axes was calculated at zero and the measured value was 0.92 and 1.05mm for 6MV and 10MV. Gantry sag was 0.5mm but this averages out with the convergent beam irradiation (CBI) technique to produce a negligible effect, see figure 6b. Figure 6: (above) 6a: gantry sag measurement with film 6b: influence of gantry sag on penumbra of an AP-PA technique Geometrical calibration of XVI A ball bearing test was used to generate the flex map. The flex map corrects for the sag and flex of the XVI panel during gantry rotation. The ball bearing is positioned at the laser intersection isocenter and the exact position of the ball bearing relative to the isocenter is determined using MV beams and iviewgt TM. The ball bearing is then placed in the exact isocenter and the position of the ball bearing is measured on the XVI at all gantry angles. The flex map can be determined from this data. Figure 7: geometrical calibration of XVI

XVI image quality The CATPHAN phantom is used to measure the image quality of the Elekta Synergy S kv images on both contrast and detail resolution. Detail >7lp/_cm_ and low contrast value < 2%. Chain test This involved executing a full patient procedure using a phantom and film to record the dose, and aluminum bars to mimic bones. Figure 8 shows the scope of the chain test. The phantom used for the chain test was a polystyrene block 30 x 30 x 10, containing four aluminium bars and a film placed in mid-plane. Figure 8: scope of the chain test Figure 9: chain text phantom in trans axial, coronal and sagittal view The process involved taking a CT scan of the phantom, scan data was transferred to Nucletron PLATO where a CBI plan was created centred on the bars and film, and then the plan data was sent to Elekta Synergy S and the XVI workstation. The phantom was then set-up on the Elekta Synergy S table and a VolumeView TM image was performed. The phantom position was corrected following image registration on the XVI workstation and then the phantom was irradiated using the plan created in PLATO. Finally the film in the phantom was analyzed and by subtracting the dose delivered by the CT and VolumeView TM scans, the CBI positioning error could be determined. Figure 10: chain test - determining the geometrical error Y-profile, error on center = 0.24mm (right) X-profile, error on center = 0.05mm

Conclusions The chain test is repeated every month, the strip test every three weeks and XVI ball bearing test and image quality test are performed weekly. Results showed that the maximum deviation from the center, for Beam Modulator TM, was 0.15mm for the diaphragm rotation and 0.40mm for gantry rotation. Table isocentric rotation accuracy was defined as 0.35mm. The chain test demonstrated a geometrical error of less than 1mm. This enabled us to conclude that Elekta Synergy S meets the criteria for high precision radiation therapy as defined by EORTC recommendations. Our future plans are to use Elekta Synergy S to treat brain metastases, meningiomas, pituitary, lung, prostate and some ad-hoc high precision patients not only with static treatment techniques, but also with dynamic techniques. Figure 11: results of spoke films for acceptance Figure 12: chain test results Dynamic techniques are work-in-progress and not available on the current product configuration. Utrecht is an Elekta research site. References 1 Sastre-Padro M. An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments. Phys. Med. Biol. 2004 Jun 21: 49(12): 2631-43. 2 Treuer H. Hoevels M, Luyken K, Gierich A, Kocker M, Muller RP, sturm V. On isocenter adjustment and quality control in linear accelerator-based radiosurgery with circular collimators and room lasers. Phys. Med. Biol. 2000 Aug: 45(8):2331-42. 3 Pastyr O, Hartmann GH, Schlegel W, Schabbert S, Treuer H, Lorenz WJ, Sturm V. Stereotactically-guided convergent beam irradiation with a linear accelerator: localization technique. Acta Neuorochir. (Wien). 1989;99 (1-2): 61-4.

Fighting serious disease n Stereotactic Neurosurgery n Gamma Knife surgery n Functional Mapping n Precision Radiation Therapy n Image Guided Radiation Therapy n Stereotactic Radiation Therapy Corporate Head Office Stockholm, Sweden Tel +46 8 587 254 00 Fax +46 8 587 255 00 info@elekta.com Worldwide Product Support Center Tel +44 01293 654068 Fax +44 01293 654655 info.europe@elekta.com North America Atlanta, USA Tel +1 770 300 9725 Fax +1 770 448 6338 info.america@elekta.com Europe, South America, Africa & the Middle East Tel +44 1293 654068 Fax +44 1293 654655 info.europe@elekta.com Japan Kobe, Japan Tel +81 78 241 7100 Fax +81 78 271 7823 info.japan@elekta.com www.elekta.com Asia-Pacific Hong Kong, China Tel +852 2891 2208 Fax +852 2575 7133 info.asia@elekta.com Document No. 4513 371 0654 08:06 2006 Elekta Ltd. All mentioned Elekta trademarks and registered trademarks are the property of Elekta AB (publ) and/or its subsidiaries. Nucletron PLATO is a trademark of Nucletron BV. All rights reserved. No part of this document may be reproduced in any form without written permission from the copyright holder. Specifications subject to change without notice.