MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner

Similar documents
State-of-the-Art IGRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI

Image Guidance and Beam Level Imaging in Digital Linacs

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

Tumor motion during liver SBRT

Using a research real-time control interface to go beyond dynamic MLC tracking

1. Learn to incorporate QA for surface imaging

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Novel methods and tools for MR Commissioning and Quality Control

ADVANCING CANCER TREATMENT

Is deformable image registration a solved problem?

Initial Clinical Experience with 3D Surface Image Guidance

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

ADVANCING CANCER TREATMENT

8/3/2016. Image Guidance Technologies. Introduction. Outline

8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Virtual Phantoms for IGRT QA

Coverage based treatment planning to accommodate organ deformable motions and contouring uncertainties for prostate treatment. Huijun Xu, Ph.D.

The MSKCC Approach to IMRT. Outline

Ini$al Performance Test Results on an Elekta MR-Linac

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Advanced Radiotherapy

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

Overview of Proposed TG-132 Recommendations

Real-time monitoring of linac performance using RT plans and logfiles

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Feasibility of 3D Printed Patient specific Phantoms for IMRT QA and Other Dosimetric Special Procedures

Raising the Bar in IMRT QA

Volumetric Cine Imaging for On-board Target Localization in Radiation Therapy. Wendy Beth Harris. Graduate Program in Medical Physics Duke University

Use of MRI in Radiotherapy: Technical Consideration

Auto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011

Commissioning of MR-only simulation for radiotherapy planning

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

NA-MIC National Alliance for Medical Image Computing Subject Hierarchy

Thank-You Members of TG147 TG 147: QA for nonradiographic

Good Morning! Thank you for joining us

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved

Dosimetric Analysis Report

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Image Co-Registration II: TG132 Quality Assurance for Image Registration. Image Co-Registration II: TG132 Quality Assurance for Image Registration

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication

Introduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy

Using Pinnacle 16 Deformable Image registration in a re-treat scenario

Development and validation of clinically feasible 4D-MRI in radiotherapy

Real Time Tumor Motion Tracking with CyberKnife

Motion management for MRI-guided abdominal radiotherapy. Bjorn Stemkens

Fast Dynamic MRI for Radiotherapy

PCRT 3D. Scalable Architecture System. User-Friendly. Traceable. Continuos Development

iplan RT Image Advanced Contouring Workstation - Driving Physician Collaboration

Mathematical methods and simulations tools useful in medical radiation physics

ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017

IMRT site-specific procedure: Prostate (CHHiP)

Optical Guidance. Sanford L. Meeks. July 22, 2010

Interior Tomography Approach for MRI-guided Radiation Therapy

ViewRay System Commissioning

Advanced Targeting Using Image Deformation. Justin Keister, MS DABR Aurora Health Care Kenosha, WI

TG 132: Use of Image Registration and Fusion in RT

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)

Dose Calculation and Optimization Algorithms: A Clinical Perspective

Automated Quality Assurance for Image-Guided Radiation Therapy

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Integrated proton-photon treatment planning

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Computational modelling and analysis: IOP Medical Physics Group

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Estimating 3D Respiratory Motion from Orbiting Views

Use of Deformable Image Registration in Radiation Therapy. Colin Sims, M.Sc. Accuray Incorporated 1

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Oncentra Brachy. Anatomy-based treatment planning for HDR/PDR brachytherapy

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Chapter 9 Field Shaping: Scanning Beam

Michael Speiser, Ph.D.

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

Comprehensive treatment planning for brachytherapy. Advanced planning made easy

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

3DVH FAQs. What is PDP questions

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

PET/CT multimodality imaging for radiotherapy planning in lung cancer The medical physicist point of view Isabelle Gardin Rouen CHB and Quant.I.

A Geant4 based treatment plan verification tool: commissioning of the LINAC model and DICOM-RT interface

Financial disclosure. Onboard imaging modality for IGRT

Proposal Title: Dataflow computing for continuous real-time adaptive radiotherapy

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:

Transcription:

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner

About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation (ART, IGRT) Infrastructure maintainance (DICOM servers, workflows) MRI (commissioning, QA, protocol development) Research Starlit project 16M Fast MRI-acquisition MLC tracking

Topics Simultanous MRI /Why MRI-linac Image quality MRI-linac at the UMC Utrecht First clinical action of MRI-linac Research Adaptive replanning Online imaging Real-time imaging Latency Geometric fidelity MLC tracking

Why online MRI-guidance? Image guidance Solve anatomic uncertainty Reduce the margins! At the moment of radiation With diagnostic image quality T 2 Weighted MRI of rectum, Courtesy Martijn Intven

Why online MRI-guidance? esophageal cancer, courtesy: Gert Meijer

Why online MRI-guidance? 3D T 2 -TSE 2D T 2 -TSE CT Irradiating individual lymph nodes Lymph node visualization, courtesy Tristan van Heijst

Why online MRI-guidance? Solve uncertainty due to motion Resolve and visualize motion Discrimination of target and surroundings Independent of linac position Calypso Projektionen ONLINE MRI

MRI-guidance research and developments Utrecht, The Netherlands 1.5 T MRI, 7 MV linac www.elekta.com/mr-linac Edmonton, Canada 0.5 T MRI, 6 MV linac www.magnettx.com Viewray, Cleveland, USA (www.viewray.com) 0.35 T MRI, 3 Co sources or 6 MV linac (Feb 2014) Clinical since 2014 Australia MR linac project, Sydney (www.sydney.edu.au) 1.0 T MRI, 6 MV linac

MRI-linac design at the UMC Utrecht Accelerator MLC beam J Overweg et al 2009 Proc. Intl. Soc. Mag. Reson. Med. 17 Bore design High field strength Diagnostic image quality Good field homogeneity Radiation transparency Coil gap Uniform gamma attenuation in cryostat Split gradient coil Active shielding 0T zone

MRI-linac design at the UMC Utrecht Slipring Cooling equipment Power supplies & electronics MLC & accelerator waveguide RF waveguides Modulator

MRI-linac at the UMC Utrecht More recent: Upgrade to CE-configuration Set-to-work Finished in July 2018 First patient in pre-ce state Lagendijk and Bakker, MRI guided radiotherapy - A MRI based linear accelerator Radiotherapy and Oncology Volume 56, Supplement 1, September 2000, 220 (ESTRO 19)

First patient on the MRL: First in Man Aim: Online workflow Demonstrate MRI technical feasibility and safety in the clinical setting Patients Pre-Treatment (N=5): CT Bone metastasis Palliative intervention Treatment: 8 Gy in single fraction 3/5 field IMRT Raaymakers et al. PMB. 2017;62(23):41-50 Deformable registration Delineations PseudoCT AutoPlanning Independent Calc PV OK

First in Man: Summary Total time 54 minutes for entire procedure 15 min patient set-up 7 min preparation MRI 4 min registration and contour validation 11 min planning 10 min delivery and beam-on MRI Bone match using EPID and pseudo CT DRR Theraview software Displacement vector (3D): 0.3 mm

1/7 Hz intra-beam 3D MRI during First-in-Man First in Man: Dose reconstruction Independent (geometry, state) 3D MR imager Time-stamped MRI data + time-stamped linac geometry Forward dose calculation First live dose reconstruction Dose validation M Glitzner et al 2015 Phys. Med. Biol. 60 8869

Research: Adaptive replanning per beam Delivery of the subsequent beam to linac emulator (Delivery time T beam ) Timepoint correlation to dynamic anatomies DYN beam MRLTP replanning Beam angles N = N -1 Beam dose reconstruction using DYN beam anatomies Mid-position volume update using mean DVFs over T beam Total delivered dose update Beam 21 Mid-position 72 96 120 Kontaxis et. al 2017 PMB 62 7233 192 168 144

Inter-beam replanning regime Delivery of the subsequent beam to linac emulator (Delivery time T beam ) Timepoint correlation to dynamic anatomies DYN beam MRLTP replanning Beam angles N = N -1 Beam dose reconstruction using DYN beam anatomies Mid-position volume update using mean DVFs over T beam Total delivered dose update 96 Kontaxis et al. PMB 2017

Baseline drift correction and margin reduction Recon_PTV - Adaptive Recon_PTV - Adaptive Kontaxis et. al 2017 PMB 62 7233

MR-linac therapy guidance: real time, MRI-guided MLC-tracking Response (tracking) quality Image fidelity geometric accuracy Target/OAR Surrogate Temporal quality real-time

asynchronous program MR-linac therapy guidance: real time, MRI-guided MLC-tracking A typical day of an real-time MRI-linac system Take an image ( ) Process the image ( ) Estimate motion in the image ( ) Process motion into a suitable adaptation ( ) (segment shift) Apply adaptation as long as it is suitable ( ) Be done fast and start all over Acquisition Preprocessing Motion estimation Plan adaptation MLC tracking

MR-linac therapy guidance: Imaging latency Object in MRI physical object ΔT Frame rate = 35 fps Latency = 0.28s Frame rate = 2 fps Latency = ~0.18s Physical motion Image Framerate Latency Apply tracking position Update rate Latency

Image source: https://www.premiumbeat.com/blog/3-tips-for-dealing-with-rolling-shutter/ MRI-latency: How long does the image acquisition take Normal sensors Integration time + readout time, i.e. CCD Special sensors, i.e. DSLR MRI sensors WHERE and WHEN is my image formation

MRI latency analysis Hypothesis: High sensitivity of k-space-based read-out to trajectory linear reverse linear high-low Acceleration using partial sampling/partial Fourier P. Borman et al., in revision

MRI latency analysis: Cartesian results linear reverse linear high-low

MRI latency analysis con d: Radial results Generally, ΔT=t acquisition /2 Golden angle sampling: Temporal weighting filter penalizes importance of old spokes Signal processing can reduce latency about 50%, T=t acquisition /4

Latency summary Acquisition Preprocessing Motion estimation Plan adaptation MLC tracking Runtime estimation Acquisition: 180ms 550ms + image age! Preprocessing + motion estimation: ~80ms Plan adaptation (simple tracking, gating): 5ms MLC tracking: ~30ms

The proof of the pudding: MLC tracking on the MRI linac 5mm ZrO 2 MR slice AGAR GEL ~200 mm EPID

Experiments: MRI-guided position feedback Estimated latency 10Hz 41.94 5Hz 127.11

MRI-guidance at UMC Utrecht Prototype phase finished Clinical proof of concept Beam & plan QA CE clearance in due time (July 2018) Strong research line Real-time imaging and MLC tracking Adaptive replanning See what you treat