Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC. Table 1: Parts required. Part Number Description Qty

Similar documents
Agilent IO Libraries L.02.xx or higher

How do I apply the patch files?

Agilent MassHunter Workstation Software Offline Qualitative and Quantitative Analysis

KG-TOWER Software Download and Installation Instructions

Agilent E2094M IO Libraries

Lab 03. Windows Operating Systems (Cont.)

HP / HP HPIB Driver Installation and Troubleshooting ChemStation Version A.07.0x- A.08.0xon Windows NT

FREQUENTLY ASKED QUESTIONS (FAQ) FOR ICS's 488.2V3 CONTROLLER LIBRARY See AB48-43 for questions about ICS's 488.2V4 Driver Library

12 Lead ECG CardioVu Software Installation Guide

Wireless N PCI Adapter RNX-N250PC2 Quick Installation Guide. Rev:

Multi User License Installation

DOWNLOAD AND INSTALL ERDAS IMAGINE

82341C HPIB Installation and Troubleshooting Agilent ChemStations on Windows 95 and Windows NT

USB driver and Software Installation. UPS Monitoring and Management Software

MassLynx 4.2 SCN 989 Release Notes

SCN 848 Release Notes

KG-TOWER Software Download and Installation Instructions

Installing the Driver

Advance Technologies; Automate the World.

TIE1.80InstallationGuideUK

Lab: Install a NIC in Windows XP

Eaglesoft 18.1 Installation Instructions

The port replicator may have features that are not supported by your computer.

BLUETOOTH 2.0 MINI USB ADAPTER. Quick Start Guide TABEL OF CONTENTS

Batch Eligibility Long Term Care claims

DEPOSITPARTNER INSTALLATION GUIDE Desktop Application

Agilent G1732AA MSD Security ChemStation

Setting File Creation Software for North America. Installation Instructions

Workstation Setup Instructions Release 9.5

Installing the C++ Development Tools

USB-MIDI Driver installation and settings...1 Windows XP users... 1

ACS Technical Bulletin

UPGRADE INSTALLATION PROCEDURES

Bluetooth Application Note

Windows 7 Professional 64 bit Installation and Configuration for MassLynx or Empower Controlled Ethernet Instrument Communication

APPLICATION BULLETIN ICS ICS. ELECTRONICS division of Systems West Inc. FREQUENTLY ASKED QUESTIONS (FAQ) FOR ICS's 488.2V4 CONTROLLER LIBRARY AB48-43

Gigabit Installation Guide

Installation of Sangoma A101/102/104/108/200 card Drivers on Windows XP/2003. ============================================================

7.1 Virtual Surround Sound Effect Gaming Headset. User Manual TNIEXPRESS.COM 10/1/2016

Kinetica 5.1 Kinetica Installation Guide

DEPOSITPARTNER INSTALLATION GUIDE Web Application for Windows Operating Systems

MF9300. Software setup guide

User Manual. Backup Utility for MiniStation MiniStation. v 2.0

Sage Fixed Assets Single User. Installation guide

Installing Diskeeper on Your Network

This document provides additional information for the installation and use of the Keithley I/O Layer.

Sage Fixed Assets Single User

1. Login as an administrator

ZP 500 Thermal Printer

Document authored by: Native Instruments GmbH Software version: (02/2013)

1100 Liquid Chromatograph

Before you attempt to upgrade from the Live200 system to LiveLAN 3.1, make sure that you have the following:

Lab: Install a NIC

Installing Intellicus DotNet Client on Windows. Version: 16.0

Trusted Stackware series. Rev D.O.I-Net Co., Ltd. Document No:TST E

Access PayrollSE v5.50a

PQA600A and PQASW Picture Quality Analyzer Application Install

Installation of Hot Disk v 7.3 and later on a PC with Windows 7, 8 or 10

Agilent G1701DA MSD Productivity ChemStation Software

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2

Quick Install Guide. Wireless-G Network PCI Adapter. WLP54G 3a

Agilent MassHunter Workstation Software for Triple Quad LC/MS

Installing Diskeeper on Your Network

Eclipse 6.2 Conversion

For additional information, see these documents on the Breeze 2 software DVD:

USING THE VENABLE RLC WINDOWS SOFTWARE VERSION 4.5

Breeze 2 Installation and Configuration Guide

Standalone License Installation

Client Configuration Guide

Waters 2707 Autosampler ICS v1.20 Installer

Pursuit 7 for Windows

Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide

ONLY IF YOU HAVE VISTA: If you are using Windows Vista, you will need to download the driver from link on the web version of this KB article.

Eaglesoft 19 Installation Instructions

Delphi Workstation Setup Instructions. June 3, 1009

Training Bulletin TITLE: CHIP-CARD BROWSER RECOMMENDATIONS AUDIENCE: GENERAL MANAGERS, FRONT DESK STAFF DATE: DECEMBER 12. Intro

How do I apply the patch files? The instructions shown here are only for the version patch.

Agilent ICP-OES SVS 2 Productivity Package Installation and Upgrade Instructions

JPRO Fleet Service Portal User s Guide

Remote Deposit Service

Sage Fixed Assets. Network Installation Guide for Administrators. for Depreciation, Tracking, Planning, and Reporting

10ZiG Technology. Thin Desktop Quick Start Guide

Technical Bulletin DS2

Select the Akeni Pro Server installation file that matches your operating system and double-click on the file.

SMT107 User Manual User Manual (QCF42); Version 3.0, 8/11/00; Sundance Multiprocessor Technology Ltd. 1999

Thermal Transient Test

NET-DYN USB Dual Band (Mediatek) Installation Guide. This manual is divided into three parts: Windows XP, Windows 7 / 8 / 8.

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control

IMPORTANT. Installing your EASE Scan Tool Software

Installation Guide. Last Revision: Oct 03, Page 1-

Nikon Message Center

Sage 500 ERP Business Intelligence

Prolender Miracle Integration Guide 8/21/07

WinDSX New Installations

Millennium Expert/Enterprise Installation Guide Version Table of Contents

Installing the IBM ServeRAID Cluster Solution

Thuraya SG-2520 Software Upgrade Manual

INSTALLATION AND SIGN-ON

Entis XL. Software Installation. Copyright 2004 Enraf Ltd Page 1 of 29

Remote. Installing setup and. is used the. Measure. first. are using a. may differ. numbers for your. the e mail with the. on your server. Open.

Transcription:

1 OF 21 Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC Parts required Table 1: Parts required Part Number Description Qty 289000764 Kit, HPIB Card, Inter, PCI 1 Procedure The procedure to set up this card consists of four sections: 1. Removing the National Instrument Drivers, VISA and NI Explorer Software and installing the 82350B card 2. Installing the drivers 3. Configuring IO interface of the card 4. Configuring GPIB addresses of the A1100/1200 modules.

2 OF 21 Removing the National Instruments Drivers, VISA and NI Explorer Software 1. Click Start Control Panel. Click Add or Remove Programs. 2. Scroll down the list and Select the NI-VISA program for removal if it is listed 3. Click the Change Button 4. Click Remove All.

5. Click Next. 3 OF 21

4 OF 21 6. Verify that Remove is selected and click OK. 7. Click Finish.

5 OF 21 8. If NI Measurement & Automation Explorer is listed, select it and click the Remove button. If NI- VISA it is not listed, go to Step 10.

6 OF 21 9. Click Yes at the warning message. The program will be uninstalled. 10. Back at the Add/Remove screen, select the NI-488.2 program for removal. Click the Change Button.

11. Select Remove All and click Next. 7 OF 21

8 OF 21 12. Click Next. Files are removed and ni-488.2 is uninstalled. 13. Click Finish.

9 OF 21 14. Back at the Add/Remove Programs screen, Click the x to exit. Click the x to close Control Panel. 15. Click Start Shutdown. Select Shutdown and click OK. 16. After the system has powered down, unplug the system Power cord, and remove the side panel of the computer. 17. Remove the National Instruments GPIB adapter from its PCI slot in the workstation and install the Agilent (HP) 82350 PCI GPIB into the same PCI slot. IMPORTANT: Examine the Agilent 82350 card to determine whether it is a Revision A card (82350A) or a Revision B card (82350B). The 82350A card is on the left side in the picture below and the 82350B card is on the right side. The Revision A card (82350A) will not fit into the PCS slots of the M57 PC. Only the Revision B card (82350B) can be installed in this PC. Install the drivers and software for the Agilent (HP) 82350 PCI GPIB adapter. 1. Replace the computer side panel, plug in the power cord and power on the computer. 2. After the system has booted, logon as Administrator. 3 Wait for the Welcome to the Found New Hardware Wizard message to occur and click Cancel. This message occurs because the Agilent 82350 PCI GPIB adapter has been detected and the driver has not been installed yet.

10 OF 21 Note: The message below on the right side may also occur. Click Cancel in that case as well. 4. For the Agilent 82350B adapter, the Agilent Technologies I/O Libraries for Instrument Control should be used. If the Operating system is Windows XP (On the M57/M58 PC from Waters), this is in the Suite 15 subfolder in the C:\Drivers\Agilent directory. If the Operating system is Windows 7 (On the M58 PC/C20x PC from Waters), this is in the Suite 16 subfolder in the C:\Drivers\Agilent directory. 5. Note: All installation screenshots below are taken from those of the Suite 15 libraries. Similar screens will be displayed for the Suite 16 libraries. To install the version you need, double click setup.exe in the appropriate driver folder in Windows Explorer If you receive a dialog box informing you that other software needs to be installed prior to installing the Suite 15 (or Suite 16) libraries then install those first.

11 OF 21

12 OF 21 6. The InstallShield Wizard begins the I/O Libraries installation. Click Next at the first install screen.

13 OF 21 7. Accept the Agilent License agreement on the next screen and click Next. 8. Select Typical as the setup type and click Next.

9. Click Install at the final screen. 14 OF 21

15 OF 21 10. The Agilent SICL and VISA IO Libraries files are installed. When the file install is complete, Click the blue Finish button.

11. Verify that the drivers have been installed correctly by checking Agilent Technologies 82350B PCI GPIB is listed under GPIB Interfaces in Device Manager (Right click My Computer >Manage). 16 OF 21

17 OF 21 12. Close the Computer Management window. Configuring the IO Interface of the card 1. Click the blue and white IO icon in the task bar to run the Agilent Connection Expert 2. Close the welcome screen 3. For the 82350B card, the SICL Interface ID must be changed to hpib7. To do this click the Change Properties button, click in the SICL Interface ID and remove the gpib0 name and replace it with hpib7 so the screen appears as below. Click OK.

18 OF 21 Configuring GPIB addresses of the A1100/1200 modules 1.Attach IEEE cables from the 82350B card to the all A1100/1200 System modules. Cycle power on all the HP modules. 2. Confirm that the module that the IEEE connection is going through (detector, autosampler or pump) can be seen on the bus in the Connection Expert window. If the device cannot be seen on the bus it may indicate that the module requires powering down and then powering back up again.

19 OF 21 Note: In the screenshot above the cable from the 82350B card is going to the pump and the pump has an address of 28. This is just for illustrative purposes. Normally the connection should be made through the detector. 3. Set the IEEE addresses of the module (obtained through the game boy controller on the stack) by double clicking on the specific module listed under the HPIB7 interface. 4. Launch MassLynx. In the MassLynx Inlet Editor, select Tools - Instrument Configuration and select the appropriate Agilent 1100/1200 modules communication. 5. On the Configure HPIB page, set the IEEE addresses for the modules at the Configure HPIB communication tab, set the appropriate addresses for each module and the corresponding PC connection.

20 OF 21 Note: As stated earlier, normally the PC connection should be made through the detector using either the UV or the DAD. The IEEE addresses can be obtained through the game boy controller on the stack. 6. On this same page, click the checkbox next to Use HP-IO libraries for communications to select it (the default state is non-selected), click Next 7. Click Finish to complete inlet configuration

21 OF 21 Note: In the Select Configuration Addresses box you may see that only one of the modules has an address which may be displayed as something similar to hpib7,22. This is a bug related to displaying the addresses and can be ignored. 7. You should now have successful communications with your system