Olympus X_6 Installation Instruction (OEM djust) If you are not an experienced repair tech please consider the alternate (easy install) version dditional Items Included ut Not Shown: PN 07 Olympus X Illuminator ssy PN 077 Hex Key mm PN 060 Hex Key mm M VH tape PN 05 Pot-Plate-Cable ssembly PN 0 SHCS M x 0.7 x 0mm PN 07 Power Supply - XP Power 5V Fixed lade US Input/US Output PN 07 Cable ssy.5mm ID x.5mm OD R plug to US 6 foot The illuminator may be powered by plugging the cable into the power supply provided, or into a suitable US port on a computer or other device. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Power supplies are subject to substitution without notice due to availability issues and changes in regulations. Pot Knob (save original from microscope) Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step : Place the microscope upside down, unplug the existing X illuminator cable, and unscrew the existing illuminator unit from microscope port as shown, using the mm hex key provided. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step : Disassemble back cover. Note that after installation of the Nanodyne illuminator, there should be no power connections left in place to the DC power inlet (red circle). ** Safety and misc. ** The original illuminator circuitry (red arrow) is completely replaced by the Nanodyne equipment which is powered by a universal input wall plug power supply with low voltage DC output. Regardless whether the original circuitry is left in place or removed, be sure there is no possibility of electrical hazard in case an unfamiliar user connects power to the old DC input. The original circuitry may be left in place, or removed as it is no longer used. We recommend removing the power cable from the DC input connector. It should be cut short so there is no possibility it could be connected to the original circuitry, or insulated with heat shrink tubing or electrical tape. Han-Seung Yang //09 Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step : Pull out original pot cable from the existing cable ties (red circles). Unscrew using mm hex key and pull out the base unit as shown above. Step : Unscrew and pull out the base cover as shown in left. Han-Seung Yang //09 Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step 5: Lift up the iris diaphragm cover (it is taped as shown), unscrew and pull out the iris diaphragm unit, and remove another piece of taped cover from the base unit as shown in left. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 5 OF
Step 6: Open the front panel, disconnect pot cable as shown, and remove the knob from the pot. Unscrew original PC panel as shown below. Save the screws. of them will be needed to attach the Nanodyne pot plate. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 6 OF
Step 7: Remove the original PC panel and pot cable as shown above. Place Nanodyne Pot Plate Cable ssembly using of original screws, then reinstall original Pot Knob (saved from existing system) as shown below. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 7 OF
Step 8: Cut small pieces of M VH tapes, attach to the back sides of the LIM and SET buttons, then put them back in the original places. They are now just used to cover the holes. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 8 OF
Step 9: Reinstall front panel and secure the cables as shown above with the existing tie wraps, to keep the cables out of the moving parts of the microscope, and the optical path from the illuminator. Insert black rubber guide (red circle) to the right place and attach the original taped cover from the base unit as shown in left. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 9 OF
Step 0: Reinstall the iris diaphragm unit using original hex screws as shown above and attach the original taped iris diaphragm cover as shown in left. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) 0 OF
Step : Insert and secure the base unit using original hex screws as shown above. Place the bottom cover and make sure all the original circuitry cables are disconnected (red circles) as shown in left. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step : Reinstall the back cover, insert PN 0 hex screw (red circle) to the screw hole of the Nanodyne adapter as shown above, and insert Nanodyne illuminator unit to the microscope port. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
Step : Tighten the hex screw to secure Nanodyne illuminator in place, connect the pot cable, and secure the cables as shown in left (using M VH tape if needed). Plug C adapter to the Nanodyne illuminator and now it is ready to use. *Note: The detailed Pot Cable insertion procedure to the TPI Main ssy will be shown in next page. Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Tape to secure wires. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF
efore Insertion Partial Insertion Connect the socket at the end of the Pot Cable ssembly to the mating socket of the illuminator, as shown in the pictures on this page. fter partially inserting the socket to the mating socket of the illuminator with your finger, use other mechanical tools such as driver or tweezer to firmly insert the socket all the way by pushing back edge of the socket as indicated with the red arrow in the image (top right). The socket cannot be fully engaged by pushing on the wire attached as the wires would just collapse. If you want to pull the socket out, pull the wire straight out by firmly gripping the black heat shrink tubing. Full Insertion Copyright 07 Nanodyne Measurement Systems. Document authorized for installation of Nanodyne equipment only. Han-Seung Yang //09 PN 08 Olympus X_6 Illuminator System Installation (OEM djust) OF