OP-W Series User Manual

Similar documents
OCEAN WATERPROOF WEIGHING SCALE USER MANUAL

IDS-801 Series Compact Bench Counting Scale. User Manual

AL-Series INSTRUCTION MANUAL AL-311 AL-201S. American Weigh Scales, Inc. Analytical Balances American Weigh Scales, Inc. pcs% kgct.

Oxford E Series. Instruction Manual

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series

CONTENTS. PRECAUTIONS Introduction Installation... 7

EHC-WF. User/Technical Manual. Contents subject to change without notice

American Weigh Scales H-110. User Manual. H-110 (110lb x 0.05lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 4.

MWP SERIES. High Precision

1.0 Description. 2.0 Unpacking. 3.0 Installation

GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide

Manual Compact Scale / Balance PCE-BS 300 / 3000

SAC / SHC / SHC-CH SERIES

PANTONE Formula Scale 3

User Manual. Waterproof Weighing Scale. ELW Plus / ESW Plus ELW Plus Wipower

Service Manual. Operators Manual. Ultegra Health Scale by Fairbanks Scales, Inc All rights reserved Revision 7 04/17

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

MC Piece Counting Scale

Optima OP-900-LD Score Board Indicator (with Remote Control) User Manual

PX Series Balances. Quick Start Guide. Please download the user manual from

Table of Contents. 3-1 Specifications & Features Front Panel Display Keyboard. 3-3 Rear Panel. 3-4 Power Supply. 4.

AD SERIES SCALEIBALANCE INSTRUCTION MANUAL


Compact Balances. EK-610j EK-6100j. in an Authorized A&D Dealer

uline.com OVERVIEW OF CONTROLS. k g. Tare Menu. Exit. Print. Units PRINT

STANDARD OPERATING PROCEDURE AA SERIES AA-160 PRECISION BALANCE

User Manual. Models: BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1

EB300 Weighing Scale Operation Manual

Operating instruction Platform scale

TWP-W. Weighing Indicator Service Manual

Ultegra Baggage Scale. SCB-R9000-B Series 31 x 31 Baggage Scale. Operator Manual

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1.

BW/BWS Weighing Indicator

Precision Balance W3200 Series Operating Manual

Operator Manual. Ultegra MAX. SCB-R9050 Series 21 x 21 Bench Scale by Fairbanks Scales, Inc All rights reserved Revision /15

PLATINUM PRO6000 PROFESSIONAL SERIES DIGITAL COUNTERTOP / PORTABLE BALANCE

FCS-30/60/150 kg (3 DIFFRENT SCALES) Shipping scale

TWP series. Weighing Indicator Service Manual

Weighing Indicator M701. Handrail scales. REV: M1, April,05,2014 (v )

QC-3265 Checkweigher User s Manual

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

Instruction Manual for. Torbal AGCN Series Balances

KW serial. User s guide. Bench Scales UGKW-E0302

PS-IN202 Weighing Indicator User Manual

MPB SERIES Electronic Balances. r1 v1.09. Operation Manual

Visit for discount prices on Acculab Scales SVI SERIES PROFESSIONAL PRECISION INDUSTRIAL SCALE SVI-10A SVI-20B SVI-50C SVI-100E SVI-200F

OHAUS SCOUT BALANCE SCALE

Weighing Indicator USER S GUIDE. globalindustrial.com. globalindustrial.com. 11 Harbor Park Drive Port Washington, NY 11050

EHC-C. User / Technical Manual. Contents subject to change without notice

BALANCES.COM. Contents

NB/NHB Precision Balance

STANDARD OPERATING PROCEDURE ACB SERIES ACB-600H PRECISION BALANCE

CONTENTS 1.0 CAUTIONS AND PREPARATION 2.0 DISPLAY AND KEY: 7.0 DEALER S SETTING CAUTIONS PREPARATION DISPLAY

Professional Benchtop Conductivity meter BC3020

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance.

ES210. Animal & Vet Scale User s Manual (V1611) Anyload Transducer Co. Ltd

The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.

SCS3MT080, SCS3MT100, SCS3MT120, SHDHS1500 & SPLS1000 User Manual

YJ Carat and Gold Scales

Sartorius Research. R 160 P Electronic Semi-Microbalance

This document hosted by: Ranger 7000 Scales Instruction Manual

Discovery. Ohaus Discovery. Semi-Micro and Analytical Balances. The Professional s Choice For Analytical Applications!

Control of packed goods - net version. Scales with TMC Terminal MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS. Manual Number: KTI-07-01/05/06/A

Dear user, Thank you very much for purchasing LS533H rotary laser. Please read this instruction manual before operating it.

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED

Operating instructions METTLER TOLEDO BD balances

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

Division Services. Model: DPAS-300. Division Portable Bulk Load Accumulator Scale INDUSTRIAL PH (901) FAX (901)

PS-C60K-D. Counting Scale Service Manual

PW-II Service Manual PW-II SERVICE MANUAL

GX-200/GX-300/GX-400 GX-600/GX-800/GX-1000 GX-2000/GX-3000/GX-4000/GX-6100 GX-6000/GX-8000

Ranger 7000 Scales Instruction Manual

CONTENT COMPONENTS OF PROJECTOR... 2 SPECIFICATION... 3 INSTRUCTION MANUAL... 3 CHARGING THE PROJECTOR... 3 SET UP... 4

7 Digital photo frame Model No.: DF-704

Counting Scale. FC-500Si. FC-50Ki FC-20Ki FC-10Ki. FC-500i. FC-2000i FC-1000i WM+PD B

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

QB Series. Operation Manual. Precision Balance. To be the best by every measure

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

Kilotech. KRS SWP Series. Counter-top weighing scale. NTEP (USA) and Measurement Canada approved.

Optima OP-902 User Manual

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

ESA SERIES ELECTRONIC BALANCE

DL Series NEWDOLBO OWNER S MANUAL. Simple Weighing Scale 9002-DL

PLATINUM PRO1001V DIGITAL COUNTERTOP / PORTABLE BALANCE


Instruction Manual for. Torbal AGZN Series Balances

INTRODUCTION PARTS & ACCESSORIES STRUCTURE USING YOUR EXPLORĒ ATTACHING THE STRAP OPERATING YOUR EXPLORĒ

AELP PALLET BEAMS SCALES

Sartorius XQ 6100, XQ 6100 B.

IGB Series. Operation Manual. Platform Scale. Kg/Lb Specification IMPORTANT

BALANCES.COM. Contents

QUICK START GUIDE PACKAGING. 1 x 220 V/12 VDC 1A mains adapter. Accompanying documentation. 1 x K3 weight indicator

Analytical Balance GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD C

WUNDER SA.BI.SRL EN USER MANUAL. MULTIFUNCTIONS ELECTRONIC SCALE MOD. W100 LED. User Manual

Using the Cashmaster you will notice some changes in the way you work:

Using the Cashmaster you will notice some changes in the way you work:

ANALYTICAL / SEMI-MICRO BALANCES

Transcription:

User Manual

Warnings Avoid using the balance in any harsh environment. Only trained professionals are allowed to do repair work. Turn power off before connecting/disconnecting the power source. Avoid overloading or dropping material onto the platform as that could damage the balance. Material that has a static charge could influence the performance of the balance. Electromagnetic interference can affect the reading of the balance. Unplug the power before cleaning and use a piece of soft cloth (do not use any aggressive cleaning agent); Make sure no liquid enters the balance.

User Manual / Table of Contents Getting Started... 4 Overview... 4 Contents... 5 Preparation... 5 Scale Settings... 6 Settings Menu... 6 Units Menu... 8 Si.CAL-0 Menu... CALIB Menu... 5 FILTER Menu... 7 OIML Menu... 7 QUIT Menu... 20 RESET Menu... 20 C/C2 Menu... 2 ADC Menu... 2 RST Menu... 22 ZERO Menu... 22 GSET Menu... 23 G2SET Menu... 24 More Info... 25 Troubleshooting... 25 Technical Support... 25 3/25

User Manual / Getting Started Getting Started Overview Thank you for using Optima scales for your important application. Here is the list of models this manual covers: Model # Max Weight (grams) Accuracy (grams) OPW-P3 3,000 0.2 OPW-P6 6,000 0.5 OPW-P5 5,000 OPW-P30 30,000 2 4/25

User Manual / Getting Started Contents Scale main unit Scale pan User Manual AC adapter : 2V Preparation. Carefully remove contents from the package which may include fragile parts. 2. Carefully assemble the scale by placing the scale pan. 3. Place the scale in a firm and flat surface, and avoid location where the scale maybe subject to vibration, sunlight, airflow, contact with water, and strong electromagnetic wave. 4. Level the scale by using the bubble level indicator (if available) and adjusting the feet of the scale. 5. Locate scale in the following condition:. Temperature : 0 o C ~ 40 o C 6. Plug in the scale to a surge protected power outlet if using the AC adapter. You may use the internal rechargeable batteries also once it has charged. 7. Press the ON/OFF key in the front panel to turn the scale on and wait for it to finish initialization (i.e. when the display shows a weight of 0). 8. Let the scale warm up for minute before calibration and use. 9. Make sure any weight placed on the scale is not beyond the rated maximum. 0.If scale is not used for an extended period of time, turn off the scale and unplug the AC adapter (if it s connected). 5/25

Scale Settings Settings Menu After the scale has been turned on and in the normal weighing mode, to get to the settings menu, follow the procedures below: Press the PLU key, holding it for more than 2 seconds. Press ENTER key 6/25

Enter the password 505388 by using the up/down arrow keys to increase/ decrease the number, and the left/right keys to traverse to the digits of interest. Press Enter key, and now you are ready to change the settings of the scale. Units will be the first configuration the user can change. For each settings, press Enter key to change the value of that particular parameters by using the up/down keys to increase/decrease the parameters. Press Enter key to confirm. Use the left/right arrows to traverse among the menu settings. To exit without saving settings, press the U key (esc). To exit with saving settings, select Quit from the menu and press the Enter key as shown below: 7/25

Press Enter key after Quit (as shown above), and the indicator will go back to normal weighing mode. Units Menu Select the weighing unit, the place of the decimal, the maximum capcacity, and the first rang division and the second rang division, if it exists. Select the menu Units 8/25

Press Enter key Change the parameter to 0 (kg) or (g). Press Enter key. Select 0 to 5 to choose the number of decimals after the decimal point Press Enter key On the following screen, enter the maximum capacity of the scale. For the example below a 5kg max is set. Note the decimal places that was set in the previous step. Press Enter key. 9/25

Press Enter key. The screen above lets user enter the division of the scale. The options are 0, 02, 05, 0, 20, 50. If we follow the previous example, if we choose 0 we will obtain the division/accuracy of 0.00 kg. Press Enter key to configure the second rang or to exit from the Units menu. If the unit does not have the second rang option, the message will not show (the configuration of the second rang is the same as the configuration of the first rang). To exit with saving, select the menu Quit, and press the Enter key. To continue the rest of the scale configurations, use the left/right arrow keys to select the setting of interest and press the Enter key. 0/25

Si.CAL-0 Menu This setting enables user to make a linear calibration. Calibration can be done with 2 steps up to 0 steps maximum. Take the Earth Gravity into consideration. The gravity parameters can be programmed from the setting menu Gset and G2set (explained later in the document) For example, we will calibrate the scale in 3 steps: 0kg, 0.5kg, and kg. When the pan is empty press Enter key. The screen will the blinking while measuring zero. Si-0 is shown to indicate that the measure of zero has been taken and that the unit is ready to take another measurement. /25

Press the Enter key to program the calibration value. In the example above, the value 0.5kg has been programmed by using the arrow keys. Put the mass of 0.5kg on the pan, let it stabilize, and press Enter key. The screen will blink while taking the value of the weight. When the measurement is done, Si-02 will be shown to indicate that it s ready for a 3rd measurement. 2/25

Press Enter key to program the 3rd calibration value. The value of kg has been programmed (using the arrow keys). Put the mass of kg on the pan, let it stabilize, and press Enter key. The screen will blink while measuring the weight value. When the measurement is done, Si-03 will be shown to indicator that it s ready for a 4th measurement. 3/25

Calibration can be done up to 0 values maximum. To finish calibration and to go back to the main settings menu, press the left arrow key. To exit with saving configuration, select the menu Quit and press Enter key. To continue with more configurations, using the left and right arrows to traverse, and press Enter key to select the settings you want to change. 4/25

CALIB Menu This setting enables the normal calibration in 2 steps. Press the Enter key and the internal counts is displayed (Note: may not be exactly the same as this number) Press Enter key. When the pan is empty, press the Enter key to calibration the zero value. 5/25

The screen will blink while taking the zero value. For example, if we want to program the value of kg, use the arrow keys to change to kg. Put the calibration weight (in this case kg) on the pan and press Enter key. The screen will blink while taking the value and the setting will exit automatically. 6/25

To exit with saving the configuration, select the Quit menu, and press Enter key. To continue with other settings, use the left/right keys to traverse the menu, and press Enter key to enter the menu of interest. FILTER Menu The FILTER menu must NOT be modified. OIML Menu This setting has two values: 0: Scale does NOT follow OIML recommendations : Scale does follow OIML recommendations There are 3 sub-parameters: RS: Manual zero limit - Adjusts 0-99% from the bottom of the scale; It s the capacity for zero-ing when pressing the Zero key OR: Initial auto zero limit - Adjusts 0-99% from the bottom of the scale; It s the capacity for zero-ing when turning on the scale OT: Zero tracking limit - Adjust 0-99% of the division [e]; Adjusts automatically of zero drifting Here is an example 7/25

Press Enter key The above mode indicators the scale does not follow the OIML recommendations and the sub-parameters RS, OR and/or OT can be programmed. Press Enter key With this configuration of the RS sub-parameter, the scale is capable of absorbing the zero deviation (2% from the bottom of the scale). For example, if the bottom of the scale is 5.000 kg, the scale will absorb up to 0.00 kg b pressing the Zero key. Press Enter key. 8/25

With this configuration of the OR sub-parameter, the scale is capable of staring with an initial deviation of the load cell by 6% from the bottom of the scale. Press Enter key. With this configuration of the OT sub-parameter, the scale can absorb the zero drifting which are inferior or equal to 40% of the division value. For example, if the scale is configured for the division of 0.00 kg = g, 0.4 g zero deviations will be absorbed. Press Enter key. To exit and save the configuration, select Quit menu and press Enter key. 9/25

QUIT Menu This menu allows quitting of the configuration menu while saving all the changes made. Simply press Enter key at the menu Quit as shown below: To quit configuration menu without saving changes, press U (esc) key. RESET Menu The RESET menu resets all the configuration to the default settings. Note that this may leave the device unusable until the user re-calibrates the scale properly. 20/25

C/C2 Menu This configures the scale as an individual or double platform. 0: one platform : two platforms but only one work at a time 2: two platforms working simultaneously ADC Menu This menu is composed of two sub-menus. SR: speed of ADC (5, 0, 20): 20 is the highest speed GN: Amplification of load cell (, 2): 2 is when increase of amplification is needed To quit with save, select Quit menu and press Enter key. To quit menu without changes, press U (esc) key. 2/25

RST Menu This menu allows reseting of the scale based on the offset value of the load cell. With value of, it resets the offset value by initializing it. To quit with save, select Quit menu and press Enter key. To quit menu without changes, press U (esc) key. ZERO Menu This menu allows adjusting the number of divisions starting from which it will return to zero. There can be 6 values, with 0 meaning that the function is not acting. The values from to 5 determines the number of divisions starting from which it will return to zero. To quit with save, select Quit menu and press Enter key. To quit menu without changes, press U (esc) key. 22/25

GSET Menu This menu allows scale calibration based on earth s gravity in different location. Press Enter key. Enter the gravity acceleration value without decimal point of the location you are interested in. If this value is modified, a linear calibration (Menu Si.CAL-0) needs to be performed. Press Enter key. To quit with save, select Quit menu and press Enter key. To quit menu without changes, press U (esc) key. 23/25

G2SET Menu This menu allows the scale to be used with a different earth s gravity from where the scale was initially calibrated without the need to re-calibrate. Press Enter key. Enter the gravity acceleration value without the decimal point of the new location s earth gravity. Press Enter key. To quit with save, select Quit menu and press Enter key. To quit menu without changes, press U (esc) key. 24/25

User Manual / More Info More Info Troubleshooting Q: The scale is not showing an accurate reading. A: Make sure the scale is placed on a firm and flat surface and that there are no external interference. Please revisit the Preparation section of this manual. Also, please make sure the calibration procedure has been performed correctly. It won t hurt to run through the calibration process again when in doubt!!!! Technical Support Please email support@optimascale.com if you have any further questions. 25/25