Radiotherapy Plan Competition TomoTherapy Planning System. Dmytro Synchuk. Ukrainian Center of TomoTherapy, Kirovograd, Ukraine

Similar documents
Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017

Analysis of RapidArc optimization strategies using objective function values and dose-volume histograms

Transitioning from pencil beam to Monte Carlo for electron dose calculations

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research

Dose Calculation and Verification for Tomotherapy

Monaco Concepts and IMRT / VMAT Planning LTAMON0003 / 3.0

Real-time monitoring of linac performance using RT plans and logfiles

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Treatment Planning Optimization for VMAT, Tomotherapy and Cyberknife

Radiation therapy treatment plan optimization

Using a research real-time control interface to go beyond dynamic MLC tracking

ADVANCING CANCER TREATMENT

The MSKCC Approach to IMRT. Outline

Brilliance CT Big Bore.

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

IMRT site-specific procedure: Prostate (CHHiP)

TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:

Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience

Dosimetric impact of the 160 MLC on head and neck IMRT treatments

Dosimetric Analysis Report

TOMOHD. Product Specifications

High Throughput Computing and Sampling Issues for Optimization in Radiotherapy

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Raising the Bar in IMRT QA

An optimization framework for conformal radiation treatment planning

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

ATC Conformance Statement Overview

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Technical Publication. DICOM Conformance Statement. iplan RT Dose 4.1. Document Revision 1. December 4, Copyright BrainLAB AG

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

ADVANCING CANCER TREATMENT

DICOM Conformance Statement. Leksell GammaPlan & Leksell SurgiPlan 10.2

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

Op#miza#on of a Fast CT System for More Accurate Radiotherapy. Erica Cherry, Meng Wu, and Rebecca Fahrig

Machine and Physics Data Guide

A closer look at CT scanning

Conformal Radiation Therapy. with. Cobalt-60 Tomotherapy

Initial Clinical Experience with 3D Surface Image Guidance

How would, or how does, the patient position (chin extended) affect your beam arrangement?

DAILY LINAC QA BEAM QA

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

TEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( )

Basic Radiation Oncology Physics

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

Quality assurance of a helical tomotherapy machine

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

MINIMAX OPTIMIZATION FOR HANDLING RANGE AND SETUP UNCERTAINTIES IN PROTON THERAPY

ph fax Deming Way Middleton WI USA

Dose Calculations: Where and How to Calculate Dose. Allen Holder Trinity University.

Two Effective Heuristics for Beam Angle Optimization in Radiation Therapy

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

Iterative regularization in intensity-modulated radiation therapy optimization. Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006.

Computer-Tomography II: Image reconstruction and applications

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

arxiv: v3 [physics.med-ph] 6 Dec 2011

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

ViewRay System Commissioning

Computational modelling and analysis: IOP Medical Physics Group

A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

Dose Calculation and Optimization Algorithms: A Clinical Perspective

FAST, precise. qa software

ATTENTION! The whole content of the lecture with all the animations can be dowloaded from:

TRAJECTORY- BASED TREATMENT PLANNING AND DELIVERY FOR CRANIAL RADIOSURGERY

radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI

A Fully-Automated Intensity-Modulated Radiation Therapy Planning System

Interactive Treatment Planning in Cancer Radiotherapy

Four-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose Distributions Using Sinogram Modification

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

A comparative analysis for verification of IMRT and VMAT treatment plans using a 2-D and 3-D diode array

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear

A Mimetic Algorithm for Simultaneous Multileaf Collimator Aperture Shape and Dosimetric Optimization in CyberKnife Robotic Radiosurgery

CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY

A Fully-Automated Intensity-Modulated Radiation Therapy Planning System

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Good Morning! Thank you for joining us

Applied Optimization Application to Intensity-Modulated Radiation Therapy (IMRT)

Trajectory optimization for dynamic couch rotation during volumetric modulated arc radiotherapy

Performance of a hybrid Monte Carlo-Pencil Beam dose algorithm for proton therapy inverse planning

CONTOURING ACCURACY. What Have We Learned? And Where Do We Go From Here? BEN NELMS, PH.D. AUGUST 15, 2016

Patient Set-ups and Tumor Localizations

Goals. Scope/exclusions. Outline: follow outline of MPPG (plus rationale & some implementation experiences)

Commissioning, beam matching and small field dosimetry experience with Elekta Versa


A Genetic Algorithm approach to Full Beam Configuration Inverse Planning in Coplanar Radiotherapy

Transcription:

Radiotherapy Plan Competition 2016 TomoTherapy Planning System Dmytro Synchuk Ukrainian Center of TomoTherapy, Kirovograd, Ukraine

Beam Geometry 6MV fan beam 3 jaw options 1.0, 2.5 and 5 cm 64 leaves binary collimator Leaf measurement at isocenter 0.65cm

Delivery regimes HELICAL Continuous gantry rotation at a constant rate Continuous couch movement in superior direction DIRECT Discrete 2-12 static directions Continuous couch movement in superior direction for each direction separately

Objective function F Plan F Targets F OARs where F ROI voxels (D prescr D calc ) 2 importance dose penalties DVH penalty ROI volume The smaller volume we can control the better

Beam blocking None (Default) Critical Structure Directional Blocking Critical Structure Complete Blocking Critical Structure Allowed Beam Direction Prohibited Beam Direction Target Target Target Directional Blocking only those beam directions that first go through the target are allowed Complete Blocking all beam directions that go through the completely blocked structure are prohibited

Troubleshootings: Time cumbersome Works only for PTV_BREAST Steel needed additional structures How I did my first Plan http://www.ncbi.nlm.nih.gov/pmc/articles/pmc4121994/

Additional Contours Dmin2 Dmin2 Dmin1 Dmin1 Dmin1 0,5 cm margin around the PTV to create steep dose gradient around the target Dmin2 2,5-3 cm margin around the PTV to avoid high doses in body

Additional Contours Breast Right Block LUNG RIGHT BLOCK HEART BLOCK LUNG LEFT BLOCK LUNG LEFT Plan Brest Right Block 3 mm outer ring completely blocked to achieve no more than 2Gy dose inside right breast HEART BLOCK part of the hard that is further than 3 cm from PTV (completely blocked) HEART Plan - part of the hard that is closer than 3 cm to PTV (here I optimize) LUNG LEFT BLOCK part of the left lung that is further than 3 cm from PTV (completely blocked) Dmin Lung Left part of the left lung that is closer than 1.5 cm to PTV LUNG LEFT Plan part of the left lung that is closer than 3 cm to PTV excluding Dmin Lung Left

Additional contours Avoidances LUNG RIGHT LUNG LEFT Top LUNG RIGHT LUNG RIGHT BLOCK LUNG LEFT Top Emergence of PTV SC leads to: LUNG LEFT Top part of the left lung that is closer than 3 cm to PTV on the slices where PTV SC is present LUNG RIGHT BLOCK part of the right lung that is 1.5 cm lover than PTV SC Avoidances some volumes where some dose problems are predicted or found

Additional contours PTV Min BLOCK Block structure in the inferior part of the body made as the prolongation of the blocked regions PTV Min small part of the PTV made to shrink the volume of minimum dose

General Parameters Calculation Grid Fine resolution equal to imported planning images (256x256) Field width 2.5 cm was set according to the PTV length Beamlet Mode is a general mode in clinical practice - optimizer uses pre-computed dose volumes and simply re-weightings them Max beamlet open time MF Average beamlet open time for used beamlets Pitch dcouch/rotation field width

General Parameters Calculation Grid Fine resolution equal to imported planning images (256x256) Field width 2.5 cm was set according to the PTV length Beamlet Mode is a general mode in clinical practice - optimizer uses pre-computed dose volumes and simply re-weightings them Max beamlet open time MF Average beamlet open time for used beamlets Pitch dcouch/rotation field width

Starting restrictions

First Results First I protect OARs

Further optimization Now I work with Targets Don t forget about OARs

Further optimization

Final restrictions

Final Histogram Full Dose Final Dose

Final Dose Distribution

Main difference from my first plan

Evaluation Old Results New Results

Thank You!