Characterizing the parallax error in multi-pinhole micro-spect reconstruction

Similar documents
USING cone-beam geometry with pinhole collimation,

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

SINGLE-PHOTON emission computed tomography

Single and Multipinhole Collimator Design Evaluation Method for Small Animal SPECT

Iterative SPECT reconstruction with 3D detector response

Performance Evaluation of the Philips Gemini PET/CT System

Design and assessment of a novel SPECT system for desktop open-gantry imaging of small animals: A simulation study

Introduction to Positron Emission Tomography

Simulations in emission tomography using GATE

Iterative CT reconstruction using shearlet-based regularization

GPU implementation for rapid iterative image reconstruction algorithm

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

Fast 3D iterative image reconstruction for SPECT with rotating slat collimators

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Medical Imaging BMEN Spring 2016

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Multi-pinhole SPECT Calibration: Influence of Data Noise and Systematic Orbit Deviations

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

Absolute Quantification for Small-Animal PET. Vincent Keereman, Roel Van Holen, Christian Vanhove, Pieter Mollet, Stefaan Vandenberghe Z X

Image Acquisition Systems

Performance Evaluation of radionuclide imaging systems

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Philips SPECT/CT Systems

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies

Spectral analysis of non-stationary CT noise

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Customizable and Advanced Software for Tomographic Reconstruction

Continuation Format Page

Fits you like no other

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Fast iterative reconstruction for helical pinhole SPECT imaging

3/19/2013. Outline. Objective. SPECT System Description. Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system

Fits you like no other

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

Mathematical methods and simulations tools useful in medical radiation physics

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT

Nuclear Instruments and Methods in Physics Research A

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

James R Halama, PhD Loyola University Medical Center

Constructing System Matrices for SPECT Simulations and Reconstructions

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY

Suitability of a new alignment correction method for industrial CT

Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

Introduction to Emission Tomography

A Projection Access Scheme for Iterative Reconstruction Based on the Golden Section

Non-Stationary CT Image Noise Spectrum Analysis

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Reconstruction from Projections

Ming-Wei Lee 1, Wei-Tso Lin 1, Yu-Ching Ni 2, Meei-Ling Jan 2, Yi-Chun Chen 1 * National Central University

A prototype table-top inverse-geometry volumetric CT system

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Optimization of scanner parameters for dual energy micro-ct

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

ACOMPTON-SCATTERING gamma camera records two

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

SPECT (single photon emission computed tomography)

Quantitative imaging for clinical dosimetry

Inveon User Notes. Inveon Scanners and Inveon Acquisition Workplace 2.0

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SPECT reconstruction

DUAL energy X-ray radiography [1] can be used to separate

CIVA CT, an advanced simulation platform for NDT

3D-OSEM Transition Matrix for High Resolution PET Imaging with Modeling of the Gamma-Event Detection

An approximate cone beam reconstruction algorithm for gantry-tilted CT

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

Few-view single photon emission computed tomography (SPECT) reconstruction based on a blurred piecewise constant object model

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005

Mode-Field Diameter and Spot Size Measurements of Lensed and Tapered Specialty Fibers

Digital Volume Correlation for Materials Characterization

Emission Computed Tomography Notes

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

Effects of the difference in tube voltage of the CT scanner on. dose calculation

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

Basics of treatment planning II

EUDET Telescope Geometry and Resolution Studies

Effect of normalization method on image uniformity and binding potential estimates on micropet

INTERNATIONAL STANDARD

A publicly accessible Monte Carlo database for validation purposes in emission tomography

Estimating 3D Respiratory Motion from Orbiting Views

Transcription:

Characterizing the parallax error in multi-pinhole micro-spect reconstruction B. Vandeghinste, Student Member, IEEE, J. De Beenhouwer, R. Van Holen, Member, IEEE, S. Vandenberghe, and S. Staelens Abstract The usage of pinholes is very important in preclinical micro-spect. Pinholes can magnify the object onto the detector, resulting in better system resolutions than the detector resolution. The loss in sensitivity is usually countered by adding more pinholes, each projecting onto a specific part of the detector. As a result, gamma rays have an oblique incidence to the detector. This causes displacement and increased uncertainty in the position of the interaction of the gamma ray in the detector, also known as parallax errors or depth-of-interaction (DOI) errors. This in turn has a large influence on image reconstruction algorithms using ray tracers as a forward projector model, as the end-point of each ray on the detector has to be accurately known. In this work, we used GATE to simulate the FLEX Triumph- I system (Gamma Medica-Ideas, Northridge, CA), a CZT-based multi-pinhole micro-spect system. This system uses 5 mm thick CZT pixels, with 1.5 mm pixel pitch. The simulated information was then used to enhance the image resolution by accurately modeling the DOI. Two hundred point sources were simulated and rebinned to use the DOI information. This data was then used in a GPU-based iterative reconstruction algorithm taking the simulated DOI into account. The average displacement was then determined for all point sources, and the FWHM was calculated in three dimensions, by fitting the point sources with 3D Gaussians. We show that the displacement is reduced by 83% on average. We also show a 15% resolution gain when only 5 DOI levels are used. I. INTRODUCTION THE usage of pinholes is very important in preclinical micro-spect. Pinholes can magnify the object onto the detector, resulting in a better system resolution than detector resolution. The loss in sensitivity is usually countered by using multiple pinholes per collimator, each pinhole projecting onto one specific part of the detector. Such imaging geometry implies oblique incidence of gamma rays to the detector. This causes displacement and increased uncertainty in the position of the interaction of the gamma ray in the detector. These Manuscript received November 10, 2011. This work was supported in part by a PhD grant to Bert Vandeghinste of the Institute for the Promotion of Innovation through Science and Technology in Flanders (IWT-Vlaanderen), a postdoctoral fellowship of the Research Foundation Flanders (FWO) to Roel Van Holen, and by the CIMI project, an IBBT-project in cooperation with: Barco nv, DCILABS, IBA Dosimetry, GE and DSC Labs. IBBT is an independent multidisciplinary research institute founded by the Flemish Government, to stimulate ICT innovation. Bert Vandeghinste, Jan De Beenhouwer, Roel Van Holen, Stefaan Vandenberghe and Steven Staelens are with the Department of Electronics and Information Systems, MEDISIP, Ghent University IBBT IBiTech, De Pintelaan 185 block B, B-9000 Ghent, Belgium (e-mail: bert.vandeghinste@ugent.be). J. De Beenhouwer is also with The Vision Lab, University of Antwerp, 2610 Wilrijk, Belgium. S. Staelens is also with the Molecular Imaging Centre Antwerp, University of Antwerp, 2650 Edegem, Belgium. errors are known as parallax errors or depth-of-interaction (DOI) errors. The DOI effect can accurately be compensated for when the systematic imperfections are compensated by measuring the system response at a grid of discrete locations in the field of view (FOV) of the camera. Unfortunately, this reponse can not always be reliably measured on different systems, due to difficulty to position a point source in such a grid of locations. Moreover, these measurements are especially tedious for non-stationary systems. A different approach to compensate for the DOI effect is to directly incorporate the physical processes leading to the effect into the reconstruction. A more accurate forward projection operator will then provide projections closer to the measured data. In a ray-driven forward projector, each ray is connected from the middle of one certain detector pixel to a pinhole, and then traced through the voxel space (Fig. 1). A simplified model would put the detector end-points at the depth corresponding with the mean free path for this specific isotope and this specific detector material. For 99m Tc, this is at approximately 2.57 mm depth in CZT detectors. The DOI essentially determines the angle of each traced ray. A wrong estimation of the DOI could thus lead to wrong sampling of the voxel space. This effect will be most apparent at the edge of the FOV for each individual pinhole. Furthermore, this effect will be larger for relatively thick detector materials compared to the detector pixel width, which is the case for CZT detectors. These detectors have high energy resolution, high intrinsic spatial resolution, good detection efficiency and a high count rate [1], making them a good choice for preclinical SPECT imaging. The DOI effect has previously been reported for CsI and HPGe detectors [3]. The aim of this work is to characterize the influence of the parallax error and investigate the possible gains of modeling this physical effect in reconstruction directly, specifically for a multi-pinhole micro-spect system with pixellated CZT detectors. It has been shown that depth information can be obtained by determining the correlation between a signal s amplitude and its rise time [2], although this is no standard practice yet. It could thus be beneficial to compensate for this effect without measuring the DOI directly, e.g. through system simulations. A. SPECT camera model II. MATERIALS AND METHODS The FLEX Triumph-I system (Gamma-Medica Ideas, Northridge, CA, USA) was used as a model for a typical multipinhole CZT-based micro-spect scanner. This preclinical

a. b. Fig. 1: Ray tracing in pixel-driven projectors. Grey pixels show the detected events. The difference between the solid and dotted lines shows the possible angles for the projection operators. (a) Little influence for relatively straight rays. (b) A larger oblique angle leads to a larger parallax error. Only the detector pixel width (1.5 mm) and height (5 mm) are to scale. tri-modal PET/SPECT/CT system consists of four 80-by-80 pixels CZT detector heads with 1.6 mm pixel width and 5 mm pixel depth. Each head is fitted with a 5-pinhole Tungsten collimator. The pinhole diameters are specified at 0.5 mm and the opening angle is 32 degrees. The focal length is 75 mm and the rotation axis-detector distance was 42 mm, giving a magnification factor of 1.79. The methods developed by Bequé et al. [4], [5] were used to accurately calibrate the SPECT geometry. The measured projections from three 99m Tc point sources embedded in an acrylic disk phantom were segmented and analyzed to find the 15 centroids per projection angle per detector head. All centroids were then used in an optimization scheme, resulting in the focal length and the location for each pinhole, the distance between the detector and the rotation axis, the offset of the detector center, the detector tilt and skew and the angle offset between the 4 detector heads. B. Monte Carlo Simulation An accurate representation of this system was built in the GEANT4-based GATE v3.1.2 [6], which had been improved with forced-detection to speed up the simulation process [7]. The CZT cameras were modeled (Fig. 2) to ensure realistic attenuation and scattering in the detector itself by using the correct material properties and thicknesses. The data was generated by placing two hundred 99m Tc point sources (6.5 MBq each) 10 mm apart spherically from each other in one volume. The number of projection views was set to 60 views over 360 degrees. All data was stored as ROOT 1 output, which allows us to post-process all acquired 1 A Data Analysis Framework, http://root.cern.ch Fig. 2: CZT detector modeled in GATE. data without having to re-simulate the photon incidence on the detector. C. Processing Two different sets of data were generated from the ROOT output. The first dataset was rebinned to simulate a detector with no DOI information, by summing all detected events according to the location of their respective deposited energy. This can be regarded as the function of CZT in the current available systems. The second dataset is generated out of the same data, but binned into 5 DOI levels, by uniformly quantizing the depth of each deposited ray. A histogram of the depth was stored simultaneously per pixel (Fig. 3). Fig. 4 shows the histogram for events inside the photopeak. This removes the scattering events inside the detector. A second simulation generated a third dataset, where a highly attenuating material replaced the CZT detector. This setup simulates a perfect absorber. All events were absorbed instantly when the detector was hit. This data serves as a reference to the other datasets, as no parallax error can occur. This allows us to investigate the best possible results in the ideal case. D. Reconstruction A reconstruction algorithm was developed for multi-pinhole micro-spect imaging. This algorithm consists of a raydriven forward projector and a voxel-driven back projection. These are combined in the iterative OSEM algorithm. Pinhole resolution recovery was based on a multi-ray approach using 7 rays [8]. These 7 rays intersect the circular opening of the pinhole in a hexagonal pattern with one of the points in the centre. This effectively subsamples the pinhole. Pinhole penetration was taken into account by calculating the effective pinhole diameter d r using the nominal diameter d, the acceptance angle of the pinhole and the attenuation of the collimator material µ as parameters [9]:

Fig. 3: Histogram of detected photons according to depth of detection. Full line: measured DOI. Dashed line: scattered photons. Dotted line: photonscatter free. Fraction 0.20 0.18 0.16 0.14 0.12 0.10 0.08 0.06 0.04 0.02 0.00-1 0 1 2 3 4 5 6 Depth (mm) Fig. 4: Histogram of photon detection depth, for photons in the photopeak window. d r = s d d + 2 µ tan. (1) 2 Three different volumes were reconstructed: 1) For the non-ideal single-layer dataset, 7 rays were traced from the center of each detector pixel at a depth equal to the mean free path trough the pinhole, calculating which voxels to sample along the ray. 2) For the 5-layer dataset, 7 rays were drawn per pixel per DOI layer. The endpoint of each ray was chosen to be in the centre of each layer. The sampled voxel values along this ray were then multiplied by the fraction of photons received by each layer, as determined by the simulations (Fig. 4). This amounts to weighted subsampling of the DOI per pixel. For the back projector, the same principle was applied, but without taking the photon fraction into account. 3) The reference dataset was reconstructed similarly to the single-layer dataset, putting the end point of each ray at the top plate of the detector. Fig. 5: Central slice of data reconstructed with 5 DOI levels. All datasets were reconstructed to a 320 320 320 voxel space with a 0.2 0.2 0.2 mm 3 voxel pitch. All voxel samples were determined through 3D texture interpolation in CUDA. This leads to blobs with only a small discretization error. The data was reconstructed using 5 iterations and 6 subsets. These parameters lead to a 2-minute reconstruction time when DOI is not corrected for, and 10 minutes reconstruction time when 5 DOI levels are used. All reconstruction software was implemented in CUDA, and ran on a Tesla M2070 GPU (NVIDIA, Santa Clara, CA, USA). E. Post-processing The reconstructed images were automatically segmented and scanned for all point sources. Three 2D Gaussians were fitted to each point source inside a 20 20 20 pixels window, resulting in a mean location of the peak µ x,y,z and a standard deviation x,y,z. The x-axis is defined as being perpendicular to the detector face, the y-axis as the tangential, and the z-axis as parallel to the rotational axis. As a first measure, the location was compared to the positions of the point sources from the ideal simulation, using the `2-norm of the difference between those locations (in mm): cost = 1 X ~p n ~µ n 2 n 2, (2) n with ~µ the estimated peak position, ~p the reference position from the simulation with the perfect absorber (Section II-C) and n the number of point sources measured. This metric thus represents the displacement caused by inaccurate DOI modeling. As a second measure, the full-width-at-half-maximum (FWHM) was compared to the FWHM of the ideal simulation results, to quantify the loss in resolution when DOI is not adequately accounted for: FWHM x,y,z =2 p 2 ln 2 x,y,z. (3) III. RESULTS Figure 5 shows data reconstructed with 5 DOI levels, showing how the outside FWHMs have broadened compared

TABLE I: Averaged results (mean±std.dev., in mm). FWHM x FWHM y FWHM z Displacement Perfect Absorber 0.483±0.110 0.422±0.162 0.463±0.111 0.000 No DOI 0.568±0.054 0.469±0.150 0.557±0.204 0.333 5 DOI levels 0.471±0.049 0.406±0.121 0.486±0.143 0.057 Increase 17.08% 14.33% 12.75% - Perpendicular FWHM Tangential FWHM 0.6 0.5 0.4 0.3 0.2 0.1 0 (a) (b) Fig. 6: Two color coded plots showing (a) the perpendicular FWHM and (b) the tangential FWHM. to the central FWHM. Figure 6 shows the mean location of the point sources reconstructed from 5 DOI levels. The point sources have been color coded to show FWHM x and FWHM y. All figures show the central slice. The displacement cost for all point sources combined was found to be at 0.333 mm when no DOI information was available. This improved to 0.057 mm total displacement when 5 depth levels were modeled. The perpendicular resolution gain with 5-level DOI modeling is 0.117 mm in the center to 0.054 mm on the outside, or a 19.4% to 10.4% increase in resolution. The tangential FWHM shows a 0.072 mm (12.7%) increase in resolution in the center, and a 0.030 mm (10%) increase in resolution on the outside. The axial FWHM increases by 0.007 to 0.099 mm (2.1% to 15.0%). These results have been averaged and are summarized in Table I. IV. DISCUSSION The photoelectric effect is the main contributor to the DOI histogram shown on Fig. 3. Simulations show that scattered photons have a large presence near the top and the bottom of the detector. The photons at the bottom are primarily caused by backscatter, depositing low energy photons directly after changing direction. This leads to a larger influence for the deepest possible rays. Both scatter fractions can be minimized by windowing the photopeak energy. The resulting DOI information is shown on Fig. 4. This can be solved for other isotopes by building an accurate attenuation model for the detector and quantifying the photon fractions as shown in Section II-A. The results clearly show that the parallax effect is the cause of two main problems. A first problem is the broadening of the FWHMs, leading to resolution loss in the reconstructed images. A significantly higher resolution is possible both perpendicularly as well as tangentially. The influence increases when moving radially outward, and can thus be regarded as a 3D magnification effect. A second problem is the displacement of point sources. This displacement is of major importance in SPECT/CT systems. Usually, co-registration is very simple on these systems as the SPECT and CT are located on the same gantry. With this displacement, only the central region would be perfectly coregistered, and a large mismatch would be noted for voxels on the outside of the FOV. As the CT image is used to correct for attenuation and the partial volume effect, it is very important that there is no mismatch. It was shown that this effect could be corrected for by using good DOI modeling. Although conventional detectors lead to an on average 0.333 mm displacement of the pointsources, this converges to almost 0 when 5 DOI levels are used. V. CONCLUSION We have characterized the parallax error for a CZT-based multi-pinhole micro-spect system. The DOI should be corrected for when SPECT is using CT as a correction map, as a mismatch between SPECT and CT voxels is unavoidable otherwise. Each point source was displaced 0.333 mm on average when no DOI modeling was added. This displacement was reduced to 0.057 mm with the introduction of 5 DOI levels. The axial resolution increased with 15%, while the transaxial resolution increased with an average of 71 µm, which is in the range of 13%. This characterization shows the importance of modeling the DOI in multi-pinhole micro-

SPECT systems, when using relatively thick detectors such as 5-mm CZT. In the future, we will investigate the possibilities of on-thefly correction of this effect during ray tracing. The information from this study will be used to model this effect accurately in reconstruction itself, without prior knowledge of the DOI. ACKNOWLEDGMENT The authors would like to thank Dirk Meier of Gamma- Medica Ideas for his help and comments on CZT detector modeling. REFERENCES [1] T. Funk et al., A new CdZnTe-based gamma camera for high resolution pinhole SPECT, IEEE NSS Conference Record, 2004. [2] L. Guerin, L. Verger, V. Rebuffel and O. Monnet, A New Architecture for Pixellated Solid State Gamma Camera Used in Nuclear Medicine, IEEE Trans Nuc Sci, vol. 55, no. 3, pp. 15731580, 2008. [3] A. B. Hwang, K. Iwata and B. H. Hasegawa, Simulation of depth of interaction effects for pinhole SPECT, IEEE NSS Conference Record, 2001. [4] D. Bequé, J. Nuyts, G. Bormans, P. Suetens and P. Dupont, Characterization of acquisition geometry of pinhole SPECT. IEEE Trans Med Imag, vol. 22, no. 5, pp. 599-612, 2003. [5] D. Bequé, J. Nuyts, P. Suetens and G. Bormans, Optimization of geometrical calibration in pinhole SPECT, IEEE Trans Med Imag, vol. 24, no. 2, pp. 180-190, 2005. [6] S. Jan et al. GATE: a simulation toolkit for PET and SPECT, Phys Med Biol, vol. 49, pp. 4543-4561, 2004. [7] J. De Beenhouwer, Fast GATE multi-pinhole SPECT simulations, Proc of the 2010 MIC, Knoxville, Tennessee, 2010. [8] C. Vanhove, A. Andreyev, M. Defrise, J. Nuyts and A. Bossuyt, Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study, Eur J Nucl Med Mol Imaging, vol. 34, no. 2, pp. 170-180, 2007. [9] H. O. Anger, Radioisotope cameras, Instrumentation in Nuclear Medicine, G. J. Hine, Ed. New York: Academic, vol. 1, pp. 485-552, 1967.