Standard Operating Procedure Inhouse Solar Simulator

Similar documents
Simulator System SUN ABET

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode)

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017

Series 370 S SoftwSystem are Qui Switch/Mult ck Start G imeter uide Quick Start Guide

Series 2600A System SourceMeter

Four-Point Probe System

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

2.2. Facilities Requirements

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

FC 1800 #1 Operation Manual University of Notre Dame

Condensed AFM operating instructions:

RTT98 VINTAGE TURNTABLE

Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement

PANalytical X-Ray Diffractometer with PW3830 Generator. Standard Operating Procedure. Revised on 02/15/2007

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Operating Procedure for Horiba Raman Microscope

K2400 Electroplating program Users Manuals - updated Aug

IMPORTANT SAFETY INSTRUCTIONS

Model 2600B-PM V Protection Module with 1 A Clamp. Description / April 2015 *PPA * 1

Automation of wafer handling

Installation and Getting Started Guide

TIF785 Photo Tachometer

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Photo Emission Tech., Inc. 760 Calle Plano, Camarillo, CA USA Tel: (805) Fax: (805)

KIP Fold 1000 Offline User Manual

Model: VA/230W UPS Backup System

Bent Pins Inside the Sample Chamber

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number:

Raman Lab - Laser 1 operation

Model 2657A-LIM-3 LO Interconnect Module

Short instructions. Olympus FV1000

IMAGING GROUP x 1300 imaging array; 20mm x 20mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee

Startup Software Quick Start Guide

HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE

TMA Standard Operating Procedure [Updated April 30, 2015]

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual

Computer. Test Fixture SPA. Key board

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1

SOLAR LIGHTLINE SERIES

Model 8020-KHV. Kelvin Keithley Triaxial Connector Card. Description / October 2014 *P * 1

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06

IMAGING GROUP x 2048 imaging array; 13.5mm x 13.5mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee

Premium Continuous Ink Supply Systems.

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

Model kw Pulse Mode Interactive

Moving-Minds.com Treadmill Desk Operation

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions

Standard Operating Procedure of Triboindenter (Hysitron TI 950)

High-Temperature Susceptometer (Kappabridge)

Int883 INSTRUCTION MANUAL. IR-Preheating System BASIC TROUBLESHOOTING GUIDE. 12 Contact the vendor.

4 Instrument Instructions for CE

PIXIS: 512 IMAGING GROUP

Model 8020-STC. Kelvin Standard Triaxial Connector Card. Description / October 2014 *P * 1

Contents. - i - Ver.:2

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

The Microscope: A Tool of the Scientist

Instruction Manual Extreme Temperature (XT) Data Logger

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System

NVR Security System Quick Start Guide

What is an SMU Instrument, and How Do You Decide Which One is Right for Your Application? Bona, Oh

MT-150. RENOGY 150A Peak High Precision Watt Meter and Power Analyzer E. Philadelphia St., Ontario CA Version: 1.

kena Instruction Manual T-1050

7 Digital Photo Frame

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se

Prepulse Inhibition (PPI) Boxes

20W Tungsten-Halogen Light Source. Operation Manual

UHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 07

1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd 2

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1

Field Update Guide. for Raven Viper Pro

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

BGS (at min. setting) BGS (at max. setting)

User Manual. Binocular Stereo Microscope. MicroscopeNet.com

Portable Document & Image Scanner. User s Manual

BGA IRDA-WELDER User Manual. Model: T870A

SUSS SB6E BONDER SOP. October 2013

Panalytical XPert Powder X-Ray Diffraction SOP

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff.

University of Minnesota NanoFabrication Center Standard Operating Procedure

Quick Start. Getting Started

PerkinElmer Series 200 HPLC: Operating Procedure

Models 707B and 708B Switching Matrix

User's Manual. New ipad Protection Case with detachable Bluetooth keyboard. Product number: SI54072

Survival Laser SL-002PB Laser Parts Bundle w/g-1 Lens Assembly & Operation Instructions

Survival Laser SL-005PB Laser Parts Bundle Assembly & Operation Instructions

Ref ID D S/N. JP10D00123 Vintage Wafer size 200 mm Software rev. B or higher. Test Head Hp E3101A 19 Para.

Sense Autonomous 2_11. All rights reserved.

Steps for LCMS-8040 Triple Quadrupole

K Service Source. Macintosh II/IIx/IIfx

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system.

1. PRODUCT INTRODUCTION Description 1-2. Feature...

Model 8101-PIV Test Fixture User s Guide. Overview. SMU Device holder (near latch) Two Leaded Part Plungers. PIV Device holder (near hinge)

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

X-3 LED. Owner s Manual ROBOHEAD

Working Copy For Review Purposes Only!

kena TM Instruction Manual T-1050

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Installation and Getting Started Guide

Transcription:

Standard Operating Procedure Inhouse Solar Simulator By Sanchit Khatavkar and Guru Burkul National Centre for Photovoltaic Research and Education (NCPRE), Indian Institute of Technology-Bombay, Powai, Mumbai- 400076 INDIA Contamination Category: Litho/Analytical Safety Precautions:

Do not look directly onto the light source. If under any circumstances you need to look in the light, please wear the UV protection goggles placed near the system. Do not bring or place any liquid especially water near the system. Please take all necessary cautions needed when working with electrical equipments. Operating Instructions: TURN ON Sequence: 1. Check whether the power supply cable is plugged in the switchboard and then turn the switchboard ON. 2. Turn the knob to 75% (Upto the mark on the power supply). 3. Only after this is done, Press the switch at the back which will power ON both the lamp and the fan. 4. Make sure that the units knob is set to Watts. 5. Wait for 5 mins, you will see the units increasing. 6. After 5 minutes, very gently, slightly rotate the knob so that the power supply unit shows a stable 148 Watts on the display before proceeding with the experiment. 7. This completes the lamp turn ON procedure. 8. Switch ON the power supply to the computer, Keithley 2400 SMU and the Vacuum pump. 9. Switch ON the Keithley 2400 SMU by pressing the Power switch located on the front panel. 10. Place your sample onto to the chuck. 11. For organic solar cells, place the 4.3cm chuck below the light source such that the sample is exactly at the center of the light beam (Dia - 4 cm). Connect the leads of the chuck to the Keithley SMU. 12. For silicon solar cells, place the vacuum chuck below the light source. i) Place the sample onto the chuck and turn ON the vacuum pump which is located below the table. ii) Ensure that the sample is at the center of the light beam. iii) Make front and back contacts using micromanipulator and connect the leads to Keithley SMU. 13. Switch ON the PC and click icon Lab Tracer 2 14. You will be getting a screen as shown below:

15. Check the settings by clicking the SETUP 2401 link before performing the IV sweep. You will get the window shown below. 16. Click on the Source tab. Enter the start and stop values of voltage according to the measurement (dark and illuminated) in the Source tab as shown in the image above. More sweep delay is useful only for devices like organic solar cells. 17. After that clicking on Advanced tab, make the four wire measurement as shown below

18. After making sure the Settings are correct, click Run Test to start the sweep. 19. By default, the I-V plot will be displayed at the end of the sweep, if not, then click on the Graph tab. You can save the graph if needed. To save the data, click on the Datasheet tab. On this tab, there is a Save data button. On clicking that, you will be asked for the filename against which you want to save the data. Data can be saved in.txt format. Please save data in a directory having your name in E:\. Please do not save data on the desktop. 20. To do another measurement, if you are using Keithley 2401 Sourcemeter, goto File -> Exit as shown below. This will exit you from Labtracer software.

21. If you are using Keithley 2400 sourcemeter, then click on the Close datacenter button. You will exit from Datacenter window only. 22. The GPIB connects the Labtracer setup to the Keithley 2400 SMU. 23. On clicking Run Test, the Keithley 2400 SMU should start the sweep. If this does not happen, refer to Annexure I for troubleshooting. 24. To Measure DARK IV, please close the shutter that blocks illumination and do the measurement. 25. During dark I-V measurement Si solar cell please keep the compliance 100 ma. TURN OFF Sequence: 26. Turn the knob slowly to zero. 27. Wait for 15mins for the lamp to cool down. 28. Turn OFF the lamp and fan switch at the back. 29. Turn off the computer and the Keithley Sourcemeter. 30. Turn OFF the switchboard. 31. For Solar cells with top and back metal contacts like Silicon Solar cells, use the probing manipulator to probe the top contact and the vacuum chuck will act as the back contact as shown below.

For Solar Cells with both contacts on top and light falling on the top, use two probing manipulators to probe both the contacts. Annexure I Troubleshooting for Keithley 2400 issue. 1. Click Menu. In the Main Menu enter Communication. 2. In Communications setup, click GPIB. 3. Make sure that the Address is seen as 24. If 24 is seen blinking, then it is OK. 4. Press ENTER and you will get GPIB Protocol. If 488.1 is blinking, then it is OK. 5. Press EXIT to come out of the Main Menu