Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 05: QUALITY ASSURANCE AND CALIBRATION METHODS

Similar documents
Validation of a Direct Analysis in Real Time Mass Spectrometry (DART-MS) Method for the Quantitation of Six Carbon Sugar in Saccharification Matrix

Error Analysis, Statistics and Graphing

Systematic errors Random errors

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

Interactive Charts: A Powerful New Tool for Understanding Chromatography Data

Chemical Polarimeter (Order Code CHEM-POL)

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

To Plot a Graph in Origin. Example: Number of Counts from a Geiger- Müller Tube as a Function of Supply Voltage

Unit 1 Area, Volume, and Density

FAQs: Quality Systems Manual (QSM) Version 5.0

Chapter 5. Calibration methods: regression and correlation

Section ADDED NEW LANGUAGE

Procedures for Group Quantitation

14.2 The Regression Equation

How do you roll? Fig. 1 - Capstone screen showing graph areas and menus

Parallel or Perpendicular? How Can You Tell? Teacher Notes Page 1 of 6

Screening Design Selection

Reading: Graphing Techniques Revised 1/12/11 GRAPHING TECHNIQUES

LIGHT: Two-slit Interference

2.4. A LIBRARY OF PARENT FUNCTIONS

Lab 3: Acceleration of Gravity

Z-TEST / Z-STATISTIC: used to test hypotheses about. µ when the population standard deviation is unknown

Graphical Analysis of Data using Microsoft Excel [2016 Version]

Data Objectives. The same process can be applied to determine how much simplification is appropriate when describing a geochemical system.

Math 182. Assignment #4: Least Squares

We have seen that as n increases, the length of our confidence interval decreases, the confidence interval will be more narrow.

Chemical Equilibrium CHAPTER 15. Chemistry: The Molecular Nature of Matter, 6 th edition By Jesperson, Brady, & Hyslop

UV-6 Series Double Beam UV/Vis Spectrophotometers

Administrivia. Next Monday is Thanksgiving holiday. Tuesday and Wednesday the lab will be open for make-up labs. Lecture as usual on Thursday.

0 Graphical Analysis Use of Excel

Present. John Phillips made a motion to approve the minutes from the September 29 meeting which was seconded by Lee Wolf. All were in favor.

Equipment Overview: Safety:

Using Excel for Graphical Analysis of Data

LED MINISPECTROMETER LMS-R

Skyline High Resolution Metabolomics (Draft)

Tutorial 7: Automated Peak Picking in Skyline

Four Types of Slope Positive Slope Negative Slope Zero Slope Undefined Slope Slope Dude will help us understand the 4 types of slope

Data processing. Filters and normalisation. Mélanie Pétéra W4M Core Team 31/05/2017 v 1.0.0

Four equations are necessary to evaluate these coefficients. Eqn

Everything you did not want to know about least squares and positional tolerance! (in one hour or less) Raymond J. Hintz, PLS, PhD University of Maine

CERTIFICATE OF ANALYSIS

Image Processing

UV / VIS SPECTROPHOTOMETER

Supporting Information

Using Adjustable Slits to Reduce Interfering Element Effects

2 Ranking Task: complete this section and submit it to your instructor before you begin the lab.

Voluntary State Curriculum Algebra II

Lesson 76. Linear Regression, Scatterplots. Review: Shormann Algebra 2, Lessons 12, 24; Shormann Algebra 1, Lesson 94

EURL proficiency test webtool guideline

Types of Edges. Why Edge Detection? Types of Edges. Edge Detection. Gradient. Edge Detection

ES-2 Lecture: Fitting models to data

How to Analyze Materials

Physics 1020 Experiment 3. Acceleration of Falling Objects

Condence Intervals about a Single Parameter:

Pattern recognition systems Lab 3 Hough Transform for line detection

Integrated Math 1. Integrated Math, Part 1

W7 DATA ANALYSIS 2. Your graph should look something like that in Figure W7-2. It shows the expected bell shape of the Gaussian distribution.

IB Chemistry IA Checklist Design (D)

Fusion AE LC Method Validation Module. S-Matrix Corporation 1594 Myrtle Avenue Eureka, CA USA Phone: URL:

packet-switched networks. For example, multimedia applications which process

Software Engineering 2 A practical course in software engineering. Ekkart Kindler

Introduction to Motion

X-Ray fluorescence and Raman spectroscopy

INTRODUCTION TO LABVIEW

Lecture 8. Divided Differences,Least-Squares Approximations. Ceng375 Numerical Computations at December 9, 2010

GPR Analyzer version 1.23 User s Manual

Integrated Math I. IM1.1.3 Understand and use the distributive, associative, and commutative properties.

What s New in Empower 3

Section 2.3: Simple Linear Regression: Predictions and Inference

LABQUEST 2 DATA LOGGER

Laboratory One Distance and Time

Section 1.5. Finding Linear Equations

v Prerequisite Tutorials Required Components Time

1

Dealing with Data in Excel 2013/2016

Notes 9 4 Finding Exponential Equations

Steps for LCMS-8040 Triple Quadrupole

Automated Submission for Analysis Processing (ASAP) Operator s Guide Dionex Corporation

v MODFLOW Stochastic Modeling, Parameter Randomization GMS 10.3 Tutorial

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

UNIT 1: NUMBER LINES, INTERVALS, AND SETS

Fundamentals of Operations Research. Prof. G. Srinivasan. Department of Management Studies. Indian Institute of Technology, Madras. Lecture No.

1.1 calculator viewing window find roots in your calculator 1.2 functions find domain and range (from a graph) may need to review interval notation

STAT 113: Lab 9. Colin Reimer Dawson. Last revised November 10, 2015

ITSx: Policy Analysis Using Interrupted Time Series

Page 1 of 21. Friday, June 30, Mr. Eric Cureton THE EI GROUP, INC. 15 E SALEM AVENUE, SE SUITE 301 ROANOKE, VA TEL: FAX: (540)

Stage terrain 3A Heissenstein. Electrical surveying

Grade 8 Middle School Math Solution Table of Contents: Virginia State Standards Alignment

MS data processing. Filtering and correcting data. W4M Core Team. 22/09/2015 v 1.0.0

The NGAL Test Application Note for Abbott Architect c8000

: Imaging Systems Laboratory II. Laboratory 2: Snell s Law, Dispersion and the Prism March 19 & 21, n 1 n 2

Curve fitting. Lab. Formulation. Truncation Error Round-off. Measurement. Good data. Not as good data. Least squares polynomials.

THIS IS NOT REPRESNTATIVE OF CURRENT CLASS MATERIAL. STOR 455 Midterm 1 September 28, 2010

Agilent G6855AA MassHunter Personal

A4.8 Fitting relative potencies and the Schild equation

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

IQC monitoring in laboratory networks

Agilent ChemStation for UV-visible Spectroscopy

Minnesota Academic Standards for Mathematics 2007

Chapters 5-6: Statistical Inference Methods

Transcription:

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 05: QUALITY ASSURANCE AND CALIBRATION METHODS

5-0. International Measurement Evaluation Program Sample: Pb in river water (blind sample) : Certified level = 62.3 ± 1.3nM Panel a : 181 Laboratories. : 18 labs. à more than 50 % above the certified level 4 labs. à more than 50 % below the certified level Panel b : 9 different National Measurement Institutes : all results were close to the certified range

Panel a: 181 Labs. using recognized analytical procedures 18 labs. à more than 50 % above the certified level 4 labs. à more than 50 % below the certified level Certified range : 62.3 ± 1.3nM

Panal b: 9 different National Measurement Institutes Certified range : 62.3 ± 1.3nM

5-0. International Measurement Evaluation Program Sample : Pb in river water (blind sample) : Certified level = 62.3 ± 1.3nM Panel a : 181 laboratories : 18 labs. à more than 50 % above the certified level 4 labs. à more than 50 % below the certified level Panel b : 9 different National Measurement Institutes : all results were close to the certified range Periodic blind check sample is required to demonstrate continuing reliability

5-2 Method validation Detection Limit ( = lower limit of detection) - The smallest quantity of analyte that is significantly different from the blank.

Fig. 5-2 Curves show distribution of measurements expected for a blank and a sample whose concentration at the detection limit.

5-2 Method validation From Fig. 5-2 - We assume that the standard deviation of the signal from samples near the detection limit is similar to the standard deviation from blanks

Why is it ~1 % at 3s? * d.f. (n=6), t= 3.7 * d.f. (n= infinite), t =2.5 Curves in Fig 5-2 are student s t distribution for 6 degrees of freedom. They are broader than the corresponding Gaussian distributions.

From Fig. 5-2 5-2 Method validation - A procedure that produces a detection limit with ~99 % chance of being greater than the blank. (= only ~1% of measurements for a blank are expected to exceed the detection limit.) -

5-2 Method validation From Fig. 5-2 If a sample contains analyte at the detection limit, there is a 50 % chance of concluding that analyte is absent because its signal is below the detection limit (false negative).

A procedure that produces a detection limit with ~99 % chance of being greater than the blank 1. After estimating the detection limit from previous experience with the method, prepare a sample whose concentration is ~ 1 to 5 times the detection limit. 2. Measure the signal from n replicate samples (n > 7). 3. Compute the standard deviation (s) of the n measurements. 4. Measure the signal from n blanks (containing no analyte) and find the mean value, y blank. 5. The minimum detectable signal, y dl, is defined as Signal detection limit: y dl = y blank + 3 S (5-3) *3s : 99.7 %

A procedure that produces a detection limit with ~99 % chance of being greater than the blank 5. The minimum detectable signal, y dl, is defined as Signal detection limit: y dl = y blank + 3 S (5-3) 6. The corrected signal, y sample - y blank = m x sample concentration (5-4) y sample m : signal observed for the sample : slope of the linear calibration curve (à next slide) 7. The detection limit is obtained by substitute y dl in 5-3 for y sample in 5-4. Because detection limit is the sample concentration which satisfies 5-3, y dl - y blank = 3 S (5-3) y sample - y blank = m x sample concentration Detection limit = 3 s / m

Signal vs. Concentration

5-3. Standard Addition Matrix : everything in the unknown, other than analyte Matrix effect : a change in the analytical signal caused by anything in the sample other than analyte. See next slide (Fig. 5-4)

Fig. 5-4 Calibration curves for perchlorate (ClO4 - ) in pure water and in groundwater - The matrix have unknown constituents that you could not incorporate into standard solutions to make a standard curve. - Hence, the method of standard addition is required

5-3. Standard Addition Standard Addition : Known quantities of analyte are added to the unknown. From the increase in signal, we can deduce how much analyte was in the original unknown. Assumption for standard addition : A linear response to analyte : The matrix would have the same effect on added analyte as it has on the original analyte in the unknown. When we add a small volume of concentrated standard e.g., the same substance as analyte, we don t change the concentration of matrix much.

5-3. Standard Addition - When we add a small volume of concentrated standard to an existing unknown, we do not change the concentration of the matrix very much - The standard (S) is the same substance as the analyte (X). I k([s] + [X] ) ([S] + [X] ) S + X f f f f = = (5 7) I k[x] [X] - X i I : emission intensity S : standard X : unknown concentration (analyte) V : volume i : initial f : final V = V 0 (initial volume of unknown sample) + V S (volume of standard added) i [X] f = [X] i (V 0 /V), [S] f = [S] i (V S /V) (5-8) - From 5-7 & 5-8, we can solve [X] i because everything else in eq.5-7 is known.

5-3. Standard Addition Two methods for standard addition 1) Variable total volume - It is useful when we measure a property of the analyte without consuming solution such as an electrical measurement with Ion-Selective Electrode. 2) Constant total volume - Every flask contains the same concentration of unknown and differing concentration of standard. - It is useful when the chemical analysis consumes solution such as atomic or mass spectroscopy. So we need separate flasks.

5-3. Standard Addition 1) Variable total volume - It is useful when we measure a property of the analyte without consuming solution such as an electrical measurement with Ion-Selective Electrode. - We add a small volumes of standard and measure the signal, We repeat this procedure several more times. - Standard additon should increase the analytical signal to between 1.5 and 3 times its original value. I k([s] + [X] ) ([S] + [X] ) S + X f f f f = = (5 7) I k[x] [X] - X i [X] f = [X] i (V 0 /V), [S] f = [S] i (V S /V) (5-8) i From equation 5-7 and 5-8, (5-9)

Fig. 5-5 Data for standard addition experiment with variable total volume (Ascorbic acid in orange juice)

A Graphic Procedure for Standard Addition : 1) Variable total volume * Standard addition should increase the analytical signal to between 1.5 and 3 times its original value. (e.g., B = 0.5A to 2A ).

5-3. The uncertainty in the x-intercept Standard deviation of x - intercept = s y m 1 n + m 2 2 y å( xi - Sy : Standard deviation of y (Eq. 4-20) m : the absolute value of the slope of the least squares line (Eq. 4-16) n : the number of data points (= 9 for Fig. 5-6) y : the mean value of y for the 9 points x i : the individual values of x for the 9 points x : the mean value of x for the 9 points Uncertainty in the x-intercept : 0.09 8 mm (for Fig. 5-6 ) Confidence interval = ± t (standard deviation of x-intercept) * degree of freedom = n - 2 x) 2

5-3. Standard Addition Two methods for standard addition 2) Constant total volume - Every flask contains the same concentration of unknown and differing concentration of standard. - It is useful when the chemical analysis consumes solution such as atomic or mass spectroscopy. So we need separate flasks.

2) graphical Procedure for multiple solutions with constant total volume

2) graphical Procedure for multiple solutions with constant total volume I S ([S] + [X] ) + X f f = (5 7) I [X] - X i I S+ X = = I X I X [X] ([S] i f [S] + [X] [X] f i + f ) I X [X] i [X] f Standard deviation of x - intercept = s y m 1 n + m 2 2 y å( xi - x) 2

5-4. Internal Standards Internal Standard : a known amount of compound, different from analyte, that is added to the unknown. : signal from analyte is compared with a signal from the internal standard to find out how much analyte is present - They are specially useful for analyses in which the quantity of sample analyzed or the instrument response varies slightly from run to run for reasons that are difficult to control. - i) When the small quantity (ul) of sample solution injected is not reproducible (chromatography). ii) When sample loss occurs during sample preparation steps prior to any manipulations.

Fig. 5-7. Chromatographic separation of unknown (X) and internal stnadard (S) Area of analyte signal (A x ) / concentration of analyte [X] = F (Area of standard signal (A s ) / concentration of standard [S]) F : response factor * To calculate F, we prepare a known mixture of standard and analyte to measure the relative response of the detector to the two species.