Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems

Similar documents
Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Advanced Reconstruction Techniques Applied to an On-Site CT System

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Translational Computed Tomography: A New Data Acquisition Scheme

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

Quality control phantoms and protocol for a tomography system

Simulation of Radiographic Testing for POD Assessment

AUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL

Image Acquisition Systems

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

Evaluation of Computed Tomography Techniques for Material Identification in Low Level and Intermediate Level Waste

Introduction to Biomedical Imaging

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

DUAL energy X-ray radiography [1] can be used to separate

Coordinate Measuring Machines with Computed Tomography

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Frequency-based method to optimize the number of projections for industrial computed tomography

Combining Analytical and Monte Carlo Modelling for Industrial Radiology

Process Monitoring using Three Dimensional Computed Tomography and Automatic Image Processing

CIVA Computed Tomography Modeling

The PolyCT Increasing the sample throughput using multiple rotation axes

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Industrial Computed Tomography Innovations

Radiographic Simulator artist: Version 2

Three-Dimensional Laser Scanner. Field Evaluation Specifications

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis

Reconstruction Methods for Coplanar Translational Laminography Applications

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control

Fraunhofer Entwicklungszentrum Röntgentechnik, EZRT

Process Monitoring using three dimensional Computed Tomography and Automatic Image Processing

Dr. Javier Santillan, San Carlos, CA

Digital Volume Correlation for Materials Characterization

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications

Boosting the Efficiency of Wind Power Plants. Stringent Demands for Boosting the Efficiency of Wind Power Plants

Industrial Computed Tomography

Digital Image Processing

Identification of Shielding Material Configurations Using NMIS Imaging

Optiv CT160 Accurate measurement for every detail. Computed Tomography System

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis

Suitability of a new alignment correction method for industrial CT

Recognition and Measurement of Small Defects in ICT Testing

Modelling I SPECIFIC REQUIREMENTS ON COMPUTER MODELING OF RADIOGRAPHIC INSPECTIONS OF NUCLEAR COMPONENTS

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Simulating the RF Shield for the VELO Upgrade

Pipe Inspection with Digital Radiography

Scatter Correction Methods in Dimensional CT

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Boosting the Efficiency of Wind Power Plants. Stringent Demands for Boosting the Efficiency of Wind Power Plants

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition

3D Computed Tomography (CT) Its Application to Aerospace Industry

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques

Non-Stationary CT Image Noise Spectrum Analysis

Application of 450 kv Computed Tomography to Engine Blocks with Steel Liners

Industrial Radiography Simulation by MCNPX for Pipeline Corrosion Detection

CT Systems and their standards

Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

The new generation of industrial computed tomography. The Desktop CT exact S

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

COMPUTER SIMULATION OF X-RA Y NDE PROCESS COUPLED

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography

SUPPLEMENTARY INFORMATION

3D X-ray Laminography with CMOS Image Sensor Using a Projection Method for Reconstruction of Arbitrary Cross-sectional Images

Experimental accuracy assessment of different measuring sensors on workpieces with varying properties

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Robot-based real-time ultrasonic tomography for industrial NDT applications

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

High Precision Optical Instrumentation for Large Structure Position Monitoring: The BCAM System Applied to the CMS Magnet

Ch 22 Inspection Technologies

Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies

Basics of treatment planning II

Application of MCNP Code in Shielding Design for Radioactive Sources

Enhanced material contrast by dual-energy microct imaging

HIGH PRECISION SURVEY AND ALIGNMENT OF LARGE LINEAR COLLIDERS - HORIZONTAL ALIGNMENT -

Ch. 4 Physical Principles of CT

High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology

AUTOMATED 4 AXIS ADAYfIVE SCANNING WITH THE DIGIBOTICS LASER DIGITIZER

Transcription:

More Info at Open Access Database www.ndt.net/?id=16704 Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems Michael BÖHNEL 1, Andreas GRUBER 2, Nils REIMS 1, Michael SALAMON 1 1 Application Specific Methods and Systems, Development Center for X-ray Technology EZRT, Fraunhofer Institute for Integrated Circuits IIS, 90768 Fuerth, Germany; Phone: +49 911 58061-7660; e-mail: michael.boehnel@iis.fraunhofer.de, nils.reims@iis.fraunhofer.de, michael.salamon@ iis.fraunhofer.de 2 Application Center for CT in Metrology, Fraunhofer Institute for Integrated Circuits IIS in cooperation with the Wolfgang Pfeiffer Stiftung, 94469 Deggendorf, Germany; e-mail: andreas.gruber@iis.fraunhofer.de Abstract X-ray computed tomography (CT) is an established tool for non-destructive testing (NDT) up to 450 kev. Due to limitations regarding object size and material density there is an increasing need of high energy systems in the megavoltage regime, e.g. applying linear accelerators (linac) as photon source. To achieve quantitative results for dimensional measurements or being able to specify minimum detectable defect sizes the overall contributions of components of the CT system to measurement uncertainties have to be determined. The existing methods for kev energy systems cannot always be transferred to high energy applications as the basic physical photon interaction effects are different and the actual size of the measurement volume is significantly larger compared to classical X-ray CT systems. In this contribution we present specially designed methods and phantoms for the high energy X-ray imaging facility at the Fraunhofer Development Center X-ray Technology (EZRT) in Fürth, Germany, for gaining quantitative system information. The methods are optimized for a time saving measurement parameterisation to be able to repeat the procedures for checking the system status on a regular basis. Keywords: computed tomography (CT), Radiographic Testing (RT), XXL-CT, high energy CT, linear accelerator, geometrical accuracy, measurement phantoms 1. Introduction Since March 2013 the Fraunhofer Development Center X-ray Technology (Fraunhofer EZRT) in Fuerth operates an European wide unique X-ray computed tomography system for very large objects (VLO). The so called XXL-CT system is the result of several years of intense planning and simulation work of Fraunhofer scientists to push the current limits in object size to the next scale. Conventional CT devices with X-ray tubes up to 450 kev or 600 kev maximum photon energy in the field of non-destructive testing (NDT) are restricted regarding the specimen dimensions of well below 1.5 m due to the available work space. Yet for objects in this scale the materials or material compositions which can be penetrated by this radiation are either weak absorbing - e.g. components with low atomic numbers - or the maximum specimen size has to be drastically reduced for high-z materials. There are a few CT facilities available in Europe, which use linear accelerators as photon source that are capable to offer higher penetrations lengths in massive and high-z materials. Nevertheless the overall object size is mostly limited to the above mentioned dimensions, thus the main application field is the local radiography of e.g. welds using radiograms. The focus of the Fraunhofer EZRT research was to enable CT measurements in NDT or security applications for objects with even larger dimensions. Examples hereby are fully mounted jet or rocket engines, subparts of airplane wings and fins, wind power rotor blades, complete cars and car chassis as well as sea freight containers [1] (see Figure 1). The XXL- CT provides the capability to inspect specimen with a size up to 5 m in at least one spatial direction. The huge size of the XXL-CT poses new challenges to ensure a high quality of radiographic images or CT-projections in terms of correct proportions and accurate geometry of the acquired images and reconstructed 3D data sets. In this paper we present some methods

and specially designed measurement phantoms to quantify the geometric accuracy of the XXL-CT. Particular attention is paid to characterize the used manipulation system. Figure 1. Interior view of the Fraunhofer EZRT high energy facility. The laboratory offers with its size (20m x 20m floor space and 14 m high work room) enough space for the XXL-CT system and the handling of large CTspecimen e.g. sea freight containers. 2. Material and methods 2.1 General information XXL-CT The XXL-CT system consists mainly out of three components, the X-ray source, a digital detector and the manipulation system, which are briefly introduced (more detailed description in [2]). As X-ray source a 9 MeV SILAC linear accelerator from Siemens is used. The emitted bremsstrahlung spectrum is generated by a rotating tungsten transmission target with a maximum dose rate of 12 Gy/min in 1 m distance of the focal spot. The mean photon energy of the spectrum is about 2 MeV [3] and therefore allows the penetration of e.g. 10 cm lead, 70 cm aluminum or up to 30 cm cast iron. To scan VLOs the detector system needs to cover a distance of several meters and must cope with high energy radiation. Possible options are flat panel or line detectors. At present there are hardly any flat panel detectors available, which can be stacked sufficiently close together to span large distances of active area. Furthermore at high photon energies in the MeV regime

the detector efficiency is very low due to thin sensor layers. Additionally the physical interaction of the radiation in the specimen is dominated by Compton scattering. This causes a high amount of scattered radiation which reduces the image quality of flat panel arrays. In contrast line detectors can be set up with pre collimators to absorb scattered radiation in the vertical direction. In our case we have decided to choose a 4 m wide line detector array (LDA). The indirect converting detector is equipped with a 10 mm thick Cadmium Tungstate scintillator and a pixel pitch of 0.4 mm. This results in nearly 10,000 pixels. The calculated quantum efficiency based on Monte-Carlo simulations is about 45 % for a 9 MeV spectrum. The manipulator for the XXL-CT-system consists generally of two towers and a rotation stage (see Figure 2). The two towers are 8 m high steel framework structures that are equipped with vertical guide rails on which a slide is mounted. The slides connected to spindle drives carry on the one hand the linear accelerator (linac tower) and on the other hand the line detector array (detector tower). To gain the absolute vertical position information of the X-ray components both towers are equipped with an absolute position measuring system with a resolution in the sub micrometer regime. The rotation stage consists of a steel table with 3.0 m in diameter. The rotation device is specified to a maximum load of 12.5 tons and is located directly in front of the detector tower. This overall configuration allows to scan a cylindrical volume of 3.2 m in diameter and nearly 5 m in height. Two different scan modes for CT measurements are implemented: A sinogram mode that acquires continuously single line images during the constant rotation of the specimen stage. The result is a single CT slice through the object at the given vertical scan position. The second mode creates single 2D projections (radiographs) of the object. Hereby the two towers move the linac and the LDA synchronously in the vertical direction upwards while consecutive line images are acquired. Afterwards the specimen is rotated by one angular step and the procedure is repeated to gain a set of projections that can be reconstructed. Figure 2. Schematic sketch of the XXL-CT system. 2.2 Methods and phantoms for determining the geometric accuracy of the CT- system

After the general introduction of the XXL-CT the quantitative characterization of the system will be discussed. In this case we will focus on the following topics: 1. Determination of detector gaps of the LDA 2. Lateral accuracy of the manipulation system during vertical movement 3. Geometrical precision in XXL-CT data 2.2.1 Detector gap The used 4 m wide LDA consists out of three separate detector modules of approximately 1.4 m length each. Therefore the mounted LDA possesses - due to the limits in the mechanical precision of the production of a single module and the required connection interface - between the middle detector module and its left and right neighboring module two physical gaps, where the scintillator segments of the sensors are not directly connected. The actual distance of the spacing of the pixel grids between two connected modules cannot be measured mechanically and needs to be determined in the radioscopic image. Otherwise the gained radiography does not represent the correct geometric projection as the two gaps are unaccounted for in the composed image consisting of left, middle and right detector module partial images. To gain the actual gap dimensions the following method was used. A calibrated ball bar measurement phantom (3 balls with defined and well known distances of the ball centers) is positioned directly in front of the LDA. The axis of the bar is aligned parallel to the LDA in a way that two balls are located in front of the middle module and the third ball is located in the corresponding outer module area. A single projection of the ball bar is acquired by a vertical scan. Then the radiography is analyzed by measuring the distances of the ball centers (see Figure 3). As the detector gap is not considered in the image, the distance between the outer ball and the two inner balls is disturbed and appears shortened. By using the known reference distances of the ball centers in respect to the measured distances to compensate the magnification in the image, the gap distance can be calculated. Figure 3. Analyzed radiography of a ball bar phantom to determine the detector gap. The left ball is positioned in the left detector module area and the two balls at the right hand side are positioned in the middle detector module area. Between the left and middle ball is the unknown detector gap. A comparison of the known distances of the ball centers in respect to the measured distances allows calculating the gap length between the left and middle detector module [4]. 2.2.2 Lateral accuracy of the manipulation system in respect to the vertical movement As mentioned in the previously, the XXL-CT system is able to generate huge 2D projection images by moving the radiation source and the LDA simultaneously upwards in the vertical

direction. A very important issue to gain geometrical correct projections of an object is the lateral precision of the manipulation system, as a lateral shift of the focus position and/or the center of the LDA during the vertical movement leads to a distorted image. The lateral accuracy is directly linked to the proper alignment of the guide rails which carry the slides onto which the components are mounted. During the installation of the XXL-CT system the alignment of the guiding rails of both towers were checked independently by using a laser tracker. This measurement tool indicates lateral deviations in the sub micrometer regime. But the approach is very time consuming (several hours of preparation and measurement time), expensive (up to 1000 Euros) and the availability for regular short-term checks is limited as external companies must be hired. Therefore the ideal solution was to use the existing imaging system in combination with a special measurement phantom to check the overall accuracy of the both towers within a short measurement time. This ensures a performance control which can be conducted in between measurement campaigns or on a regular basis to document the system status. The basic idea of the new method is that a 2D projection (using the maximum vertical scan height of about 5 m) of a preferably straight object is acquired by the XXL-CT system. The shape of this ideal straight object in the image should also be absolute straight in the case of a perfect path of both towers. Any lateral deviation in the trajectory of the linac tower and/or the detector tower leads to a distortion in the shape of the object in the image and can be quantitatively analysed. To apply this method a new wire phantom was developed and constructed. The measurement phantom consists of a 6 m long rigid plastic tube that is vertical mounted. Into the tube a 4 mm thick and nearly 5.5 m long steel wire was inserted and clamped using a screw driven device allowing for tightening the wire. Due to the applied strong tension and the vertical alignment of the wire within the plastic tube (to eliminate bending by gravitation) an ideal straight absorbing steel object was approximated. To analyse the 2D projection of the wire phantom a MATLAB routine was programed. Hereby in every line of the 2D image, which corresponds to a certain vertical height z in the object coordinate system, the center of the wire was determined. This was realized by fitting an appropriate function (linear combination of a linear function and a Gaussian) into the line profile (graphical representation of intensity and pixel coordinate) of the absorption pattern caused by the wire. The fit result of the expected value µ of the Gaussian represents the center of the wire position in the image with respect to the pixel coordinate system. The data set of all determined positions (µ, z) of the wire center µ in pixel coordinates and the corresponding vertical heights z were used to calculate a mean ideal straight path by a linear regression. For all data sets the lateral deviation D in respect to this ideal straight line was calculated. With the known measurement parameters magnification and pixel pitch the lateral deviation D in units of the pixel coordinate system can finally be transformed into a distance d in units of millimeter. 2.2.3 Geometrical precision in XXL-CT data To perform detailed analysis of CT-data concerning variance comparisons or determining dimensions and tolerances of objects the measurement uncertainty of the CT-system needs to be determined. For conventional CT-systems there are several methods available to ensure the quality in metrology, e.g. using calibrated phantoms for regular quality controls or performing a calibration of the CT machine. Unfortunately these established procedures cannot directly be transferred to the XXL-CT system. The main reason is the size and availability of measurement phantoms: They are simply much too small to cope with the huge measurement volume of nearly 38 m³. As a first step to approach this issue the quantification of the basic

geometrical precision of the XXL-CT was in the focus of our research. In the following section a brief insight into this issue is presented (more details in [4]). The intention was to develop a method to determine radial and vertical deviations in at least some distinct parts of the measurement volume. To conduct a measurement in a short time a measurement phantom geometry had to be chosen, that a single horizontal CT slice of the object (sinogram mode) is sufficient to extract radial and vertical position information that can be compared to a high precision reference measurement (laser tracker). The main reason was that a single CT slice (diameter 3.2 m) with high image quality and full resolution of 333 µm voxel size takes about 12 min compared to 30 hours for a conventional CT of a volume covering the same horizontal area and 5 cm of height (e.g. to analyze an expanded object in the vertical dimension). The solution was to use two identical truncated cones (see Figure 4). The measurement of the radiuses (D1 and D2) and centers of both cones in the CT image allows to determine the radial position to a reference point in the CT slice (e.g. center of CT slice) as well as the height h h = H* [0.5 + (D2-D1 ) /(2*(D0-d0)]), which is equivalent to a height of the slice in relation to the known position of the cone surfaces. Figure 4. Alignment of two truncated cones to determine the radial and vertical position from a single CT slice [4]. In total six of these cone pairs were installed along the radius of the rotation stage (see Figure 5). In Figure 6 the corresponding analysis of the radiuses of a CT image is displayed. The values are used to calculate the height of the CT slice in relation to the top of the surface of the cone pairs and the radius of the cone pairs from the center of rotation. For every single cone a reference measurement of the exact position was conducted by a high precision laser tracker system. The comparison of the laser tracker results and the results from the CT-slice measurement leads to a statement of the geometrical precision.

Figure 5. Picture of the measurement phantom on the rotation stage covering the radius of 1.5 m (left) and detail of the alignment of the truncated cones (right). Figure 6. Example of a CT slice with the measurement of the radiuses of all cones [4]. 3. Results and discussion 3.1 Detector gap Table 1 shows the resulting dimensions of the detector gaps on the left and right side of the center detector module to its neighbours. In relation to the actual pixel pitch of 0.4 mm a correction method needs to be applied to obtain an undistorted 2D projection of an object overlapping the center detector module in the image. One possibility is to fill the gap between the subimages of the concerning detector modules by a linear interpolation method. Then the stiched image can be transferred to a conventional CT reconstruction algorithm. In respect to create a correct CT reconstruction directly out of the three subimages - without further corrections - it also would be feasible to use an Algebraic Reconstruction Technique (ART) which can handle spatially separated detector modules. Table 1. Results of the detector gap measurement Left detector gap in mm Right detector gap in mm 0.52 0.68 3.2 Lateral accuracy of the manipulation system in respect to the vertical movement Figure 7 displays the measurement result of the lateral accuracy of the manipulation system. The horizontal axis represents the displacement of an object detail which is located in the center of the rotation stage. A negative value denotes that the object appears to be shifted to the left side of the LDA (viewpoint source towards detector) and vice versa. The vertical axis in the graph describes the vertical height of the object measured from the rotation stage surface (z = 0 mm). The red data points represent the calculated lateral deviations from the method mentioned in section 2.2.2. The broad distribution of the data is mainly caused by the statistics of the measurement and the simplified fitting model to approximate the absorption

pattern of the steel wire. A detailed study of possible sources of error has not been conducted so far. For the purpose of better visibility a 5-point moving average (solid black line) was applied to the data. Most of the data points are located in the rage of +/- 0.1 mm in the examined vertical scan heights between z = 600 mm and z = 4500 mm. Therefore the mean lateral deviation is well below the effective pixel size of 333 µm (pixel pitch 400 µm and magnification of approx. 1.2). Consequently the lateral accuracy of the tower manipulation system is sufficiently accurate that a blur of object details across several pixels is no issue. Furthermore the overall shape of the curve and order of magnitude of the lateral deviations are in good agreement with laser tracker measurements during the installation of the manipulation system which proofs the reliability of the described method. The proposed method is adequate to check in regular terms the performance of the tower manipulators. Figure 7. Result of the lateral deviation of an object detail located in the center of the rotation stage due to alignment imperfections of both towers during the vertical movement. The resulting displacement of an object detail is well below the dimension of the effective pixel size of 333 µm.

3.3 Geometrical precision in XXL-CT data Table 2 and 3 summarizes the results of the deviations in radial and vertical direction of the CT measurement in respect to the reference measurement by the laser tracker system. In total three different vertical positions (z) were under investigation. Hereby position z = 0 mm denotes the lowest realizable position of the truncated cone phantom, which is about 200 mm above the rotation stage surface. The measurement phantom was aligned in a way that the inner two cones define the center of the rotation axis (radius R = 0 mm) and four further radial positions could be accessed (R = 379 mm up to R = 1480 mm). The deviations of the vertical directions are homogenously distributed within the different height positions z. The maximum value of 0.38 mm is still passable compared to the overall size of the CT volume. Nevertheless especially the radial deviations are showing higher values which exhibit a nearly linear behaviour. This is a clue that a systematic error occurs that leads to a radial distortion. Most likely the discrepancy can be explained by an incorrect value for the magnification factor M which was determined by measuring the FDD (focus detector distance) and FOD (focus object distance) manually. An incorrect magnification can cause a scaling of the radial component of a fan beam CT slice. By the help of the results a magnification M can be derived that should compensate the radial deviations to a lower level. Currently this issue is under investigation and further measurements are planned. Table 2. Results of the geometrical precision in radial direction height deviation of CT measurement to reference (in mm) position R = 0 mm R = 379 mm R = 740 mm R = 1100 mm R = 1480 mm z = 1800 mm 0.12-0.65-1.24-1.88-2.56 z = 900 mm 0.11-0.61-1.24-1.85-2.45 z = 0 mm 0.03-0.66-1.30-1.88-2.53 Table 3. Results of the geometrical precision in vertical direction height deviation of CT measurement to reference (in mm) position R = 0 mm R = 379 mm R = 740 mm R = 1100 mm R = 1480 mm z = 1800 mm 0.17 0.25 0.38 0.26 0.19 z = 900 mm 0.10 0.19 0.29 0.20 0.17 z = 0 mm 0.00 0.07 0.19 0.11 0.07 4. Conclusion In this paper we introduced a novel CT system for scanning very large objects (VLO) like cars or sea freight containers. The new scale in size of the XXL-CT system poses also new challenges in the dimension of detector, photon source and size of the manipulation system to generate on the one hand huge 2D radiographic images and CT data sets of VLOs at all. On the other hand one should also consider the quality of the measurement results concerning dimensional accuracy for the analysis of CT data in the post processing chain. We presented the first steps towards the determination of the dimensional accuracy of a XXL-CT system by focusing on methods to characterize the used components and their influence on the geometric accuracy. E. g. the linked three line detector modules display due to their construction two gaps with distances up to 1.7-fold of the pixel pitch. We figured out a method to quantify these gaps between detector modules and to correct them to obtain geometrical correct 2D projections of objects. A specially designed wire phantom was proposed to quantify the lateral precision of the tower system which synchronously manipulates the X-ray source and the line detector array during the vertical scan movement. The overall deviations in lateral direction

were determined to be below +/- 0.1 mm during a vertical moving range of about 5 m. In comparison to the effective pixel size of 333 µm the deviations are in a magnitude which does not enable a blur of object details across several pixels. The wire phantom allows checking for the mechanical system status in simple, fast and inexpensive manner that it can be used for regular quality controls of the alignment of the manipulation system. A further phantom was specified and built up to measure the geometrical precision of CT-slices in vertical and radial direction within a single sinogram acquisition. The analysis of the phantom measurement revealed a discrepancy of the radial component in the CT-data which causes a radial distortion of the CT volume. A first examination of the linear dependency of the increase by the radial deviation in respect to the object radius led to a correction of the magnification factor. The future goal is to extend these phantom experiments to quantify the overall dimensional accuracy of the XXL-CT system for metrology applications like variance analysis shown in Figure 8. Figure 8. Variance analysis between CT and CAD data of a car chassis rear-body. The colors indicate deviations between CAD model and the acquired CT data. References 1. M. Firsching et al., 3-D scanning of sea freight containers using MeV X-rays, IEEE Nuclear Science Symposium and Medical Imaging Conference Record, IEEE, 2013 2. T. Fuchs et al., High-energy 3-d X-ray computed tomography on very large objects, to be published. 3. M. Salamon et al., Applications and Methods with High Energy CT Systems, 5th International Symposium on NDT in Aerospace, 2013 4. A Gruber, Entwicklung eines Verfahrens zur Bestimmung der Messgenauigkeit bei Groß-CT-Anlagen, Bachelor Thesis, Technische Hochschule Deggendorf, March 2014.