Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Similar documents
Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

ML reconstruction for CT

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

DUE to beam polychromacity in CT and the energy dependence

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Non-Stationary CT Image Noise Spectrum Analysis

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Ch. 4 Physical Principles of CT

Digital Image Processing

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Introduction to Biomedical Imaging

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

USING cone-beam geometry with pinhole collimation,

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Artifacts in Spiral X-ray CT Scanners: Problems and Solutions

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Scatter Correction Methods in Dimensional CT

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Image Acquisition Systems

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Computer-Tomography II: Image reconstruction and applications

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

Noise weighting with an exponent for transmission CT

A new calibration-free beam hardening reduction method for industrial CT

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

MEDICAL IMAGING 2nd Part Computed Tomography

DUAL energy X-ray radiography [1] can be used to separate

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Spectral analysis of non-stationary CT noise

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Development of a multi-axis X-ray CT for highly accurate inspection of electronic devices

CT: Physics Principles & Equipment Design

CT vs. VolumeScope: image quality and dose comparison

Frequency split metal artifact reduction (FSMAR) in computed tomography

Quality control phantoms and protocol for a tomography system

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction

Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender

Principles of Computerized Tomographic Imaging

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Suitability of a new alignment correction method for industrial CT

RADIOLOGY AND DIAGNOSTIC IMAGING

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

Tomographic Reconstruction

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms

Streak artifacts arising from metal implants such as dental fillings, surgical clips, coils, wires, and orthopedic hardware may obscure important diag

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

BME I5000: Biomedical Imaging

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Feldkamp-type image reconstruction from equiangular data

Fundamentals of CT imaging

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

MEDICAL IMAGING 2nd Part Computed Tomography

Accelerated quantitative multi-material beam hardening correction(bhc) in cone-beam CT

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

An approximate cone beam reconstruction algorithm for gantry-tilted CT

Computer-Tomography I: Principles, History, Technology

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach

Attenuation map reconstruction from TOF PET data

GPU implementation for rapid iterative image reconstruction algorithm

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Enhanced material contrast by dual-energy microct imaging

Edge-Preserving Denoising for Segmentation in CT-Images

Central Slice Theorem

3-D Compounding of B-Scan Ultrasound Images

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS

Metal artifacts reduction in x-ray CT based on segmentation and forward-projection

Transcription:

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study B. De Man, J. Nuyts, P. Dupont, G. Marchal and P. Suetens Medical Image Computing, ESAT-PSI, K.U.Leuven, B-3000 Leuven, Belgium Department of Nuclear Medicine, K.U.Leuven, B-3000 Leuven, Belgium Department of Radiology, K.U.Leuven, B-3000 Leuven, Belgium Abstract Metal streak artifacts are an important problem in X-ray computed tomography. A high-resolution 2D fan-beam computed tomography simulator is presented. Several potential causes of metal streak artifacts are studied using phantom measurements and simulations. Beam hardening, scatter, noise and exponential edge-gradient effect are identified as important causes of metal streak artifacts. Furthermore, also aliasing effects and object motion can be responsible for certain metal streak artifacts. I. INTRODUCTION In X-ray computed tomography (CT), the presence of strongly attenuating objects - such as dental fillings - causes typical streak artifacts in the reconstructed images (figure 1). Attempts to reduce those streaks are commonly referred to as metal artifact reduction (MAR). Several researchers have tried to remove metal streak artifacts using various approaches, which - to our knowledge - are all based on the assumption that measured data affected by metal objects are useless for the reconstruction. These corrupt data are either ignored [1, 2] or replaced by synthetic data [1, 3]. validation and we analyze streaks with different origin. A. Simulator II. METHODS The simulator is based on a typical fan-beam geometry (figure 2a) and is developed in IDL (Interactive Data Language, Research Systems Inc., Boulder, Colorado). All parameters were adjusted for the Siemens Somatom Plus 4 CT-scanner. The third dimension, perpendicular to the scanning plane, is not taken into account. Spiral artifacts [17, 18, 19] and axial partial volume effects [16] are beyond the scope of this article. Figure 2: (a) Typical fan-beam CT geometry and (b) projector The projector (figure 2b) is implemented in C for fast computation: for every detector element the intersections with all rows (or columns) are calculated and the interpolated values are added. Figure 1: An example of metal streak artifacts in X-ray computed tomography In our opinion, a complete understanding of the streak generation processes is indispensable for a well-founded solution to the metal streak artifact problem. In CT literature, several causes of (more general) streak artifacts can be found: beam hardening [4, 5, 6, 7, 8, 9], scatter [10, 11], noise [4], exponential edge-gradient effect (EEGE) [12], detector under-sampling [13], view under-sampling [14], object motion [15] and axial partial volume effect [16]. To investigate which of these causes are relevant to metal streak artifacts with modern CT-scanners, we developed a CT simulator and we performed a number of measurements and simulations. In the following sections we describe the simulator and its Figure 3: Simulated beam aperture: (a) sampling scheme and (b) resulting intensity distribution For ease of computation, the X-ray source is modeled as a uniformly radiating straight line. The finite sizes of the focal spot and the detector elements are modeled by sampling at high resolution (typically 0.1mm) followed by a summation of the photon flux over focal spot width (0.6mm) and detector element width (1.2mm). The resulting beam aperture is shown in figure 3. The spectrum of the X-ray tube is modeled by summing 5 monochromatic simulations at well-chosen energies (figure 4).

Absorption coefficients of water, iron and amalgam are calculated from data available on-line at http://physics.nist.gov by averaging over the appropriate parts of the spectrum. half maximum (FWHM) of the LSF was calculated by fitting a Gaussian. 2) phantoms A number of phantoms were measured and simulated. A first phantom consisted of a water bowl and a cylindrical iron rod (7; 0 mm) positioned eccentrically in the water. The second series of phantoms consisted of a Plexiglas plate, with one, two or three cylindrical amalgam fillings. Figure 5 shows two reconstructions from artifact-free simulations (a) for the water bowl with iron rod (phantom 1) and (b) for the plexi plate with three amalgam fillings (phantom 2). Figure 4: Real (solid line) and simulated (arrows) X-ray spectrum The continuous rotation of the tube-detector unit is simulated by sampling at small angular (or temporal) intervals followed by a summation of the photon flux over the view angle (or time) interval. Simulations show that the finite integration time is highly determining for the line spread function (LSF). Measurements of the LSF indicate that this integration time is about larger than the expected. We attribute this excess integration time to the ion collection time, this is the time needed for the ions to drift towards the measuring electrodes in the Xenon detectors. This result is in good agreement with the theoretical! #"%$ ion collection time of calculated as &(' where )+* is the inter-electrode distance,, the applied voltage and -./ 10 1# 32 $ '!4 5 the ionic mobility taken from [20]. This so-called after-glow effect is modeled by adding to each view an appropriate number of samples from the next view. According to the manufacturers, all non-linearities in the detector response are perfectly compensated for. Detector cross-talk is modeled by including a detector overlap of 1/6 at both sides. The detector array is shifted over one fourth of the detector element width to remove the redundancy inherent to a standard 6 fan-beam acquisition. In accordance with [10, 11] a constant scatter level was chosen. Real scatter profiles usually contain very little high frequencies, and therefore, a constant scatter profile is expected to be a good approximation, at least for the purpose of this article. Poisson noise was added to the simulated intensities. The noise level was tuned so that the simulated intensities had the same standard deviation as in the measurements. B. Validation 1) LSF We determined the LSF of both the real CT scanner and the simulator, using an iron wire (78 90: mm) positioned at different positions perpendicular to the scanning plane. The full width at Figure 5: Artifact-free simulations: (a) phantom 1: water bowl with iron rod and (b) phantom 2: plexi plate with 3 amalgam fillings C. Reconstruction Two reconstruction algorithms have been implemented. The first is the direct fan-beam filtered backprojection algorithm from Herman [21]. The backprojector (figure 6) is implemented in C for fast computation: for every pixel, the intersection with the detector array is computed. The corresponding interpolated value is assigned to the pixel. Figure 6: Backprojector A second reconstruction algorithm consists of a rebinner followed by a parallel beam reconstruction. This was used only for validation of the direct fan-beam reconstruction. D. Streak Analysis The effect of beam hardening was investigated by comparing polychromatic and monochromatic simulations. For the latter, attenuation coefficients of the phantoms were defined so that the resulting intensities were in the same order of magnitude as in the polychromatic case. Similarly, the effects of scatter, noise and exponential

edge-gradient effect (EEGE) were investigated by comparing scatter-, noise- and EEGE-free simulations and simulations with scatter, noise or EEGE. The scatter-to-primary ratio was arbitrarily chosen to be 0.0001. The noise level was determined by choosing an unattenuated flux of < or =?>@ < photons per detector, depending on the phantom. Under these circumstances, both the standard deviations in the projections and the degree of artifact were comparable to those in the measured data. The effect of noise was also investigated by scanning a phantom multiple times under identical circumstances. Averaging of A intensity sinograms allows to reduce the noise by a factor B A. Averaged images and original images were compared. The EEGE needs some further explanation (see also [12]). Two modes of simulation were used. In the first mode, the projection sums are calculated as the logarithm of the integral over the focal spot and over the detector elements of the photon flux: C EDGFIH J#KLJ#MONQPSR#T * CVU J#W1XQY Z D U #[V\]#[V\_^`\_a b R#cdRNQP (1) Figure 7: FWHM of the LSF versus view number: measured (black) and simulated (gray) N_o determines the beam width and the distance between N]p two consecutive beams at that particular position. Distance and angle q determine the speed at which a beam sweeps by at that particular position: r 5suttwv xr gqyzm{#ge R N]p}R#~e U q [. NeP where is the un-attenuated intensity, b c and the areas of focal spot and detector, the attenuation coefficient and f(g#h i the line defined by ^ and a. This is the correct simulation procedure. In the second mode, the projection sums are calculated as the mean over the focal spot and over the detector elements of the attenuation line integrals. C b ckj K j M U j WlXmY Z U n[\]n[v\_^8\_a (2) This second mode allows to eliminate the EEGE. The influence of motion of the scanned object was simulated by calculating two sinograms for the object at two slightly different positions and taking a part of both sinograms to synthesize a new sinogram. Influence of detector and view sampling was investigated by reconstructing an image making use of data from only a subset of the detectors or of the views. A. Validation III. RESULTS & DISCUSSION A good Gaussian fit of the LSF was obtained in most projections (except where the projection of the wire coincides with the border of the patient table). Good agreement was obtained between measured and simulated FWHM of LSF as a function of view angle (figure 7). The typical cyclic behaviour has an easy intuitive explanation: The width N_o of the LSF depends mainly on three parameters: the distance N]p from the iron wire to the focal spot, the distance from the iron wire to the center of rotation and the angle q between (1) the beam through the iron wire and (2) the line connecting the iron wire and the center of rotation. Distance Figure 8: Water bowl with iron rod (left) and plexi plate with 3 amalgam fillings (right): measured (top) versus simulated (bottom) Good agreement is obtained between measured and simulated images for both phantom 1 and phantom 2 (figure 8). All ~QˆŠ reconstructions are windowed in the interval. ƒ 10 90 6 B. Beam Hardening Figure 9 shows the polychromatic simulations. For phantom 1, we observe dark streaks in the directions of highest attenuation. Also the more general cupping artifact is present, but this can not be seen with the current windowing. In the case of two or more metal objects (phantom 2) additional dark streaks connecting the metal objects can be seen. Figure 9: Polychromatic simulations: (a) phantom 1 and (b) phantom 2

C. Scatter Figure 10 shows the simulations with added constant-level scatter. For phantom 1, we observe dark streaks in the directions of highest attenuation. Again, also cupping is present but not visible with the current windowing. For phantom 2, additional streaks connecting the metal objects can be seen. Even a very small scatter-to-primary ratio causes significant streaks. Also some bright streaks are present bordering the dark streaks. This is probably due to the negative values in the reconstruction kernel. Note that these artifacts are very similar as the polychromatic artifacts in figure 9 (see also [10] and [11]). This type of streaks consists of thin lines alternately dark and bright and are also distinguishable in figure 8. Noise artifacts are non-linear artifacts, just like beam hardening and scatter artifacts ([4]). The degree of artifact strongly depends on the magnitude of the measured intensities and thus of the total integrated attenuation. As a consequence, even the amount of scatter and beam hardening have a strong influence on the severity of the noise artifacts. F. Motion Figure 13 shows a simulation where the iron rod was moved over about 0 mm after 700 views. Streaks can be seen in two directions: from the rod to the focal spot at the two positions of inconsistency, this is between view 700 and view 701 and between view 1056 and view 1. Figure 10: Simulation with constant-level scatter: (a) phantom 1 and (b) phantom 2 D. Exponential Edge-gradient Effect Figure 11 shows simulations with the exponential edge-gradient effect. For phantom 1, no streaks are observed. For phantom 2, dark streaks can be seen connecting edges with equally-signed gradients, while white streaks can be seen connecting edges with opposite gradients. More generally, the EEGE is known to cause streaks tangent to long straight edges. For instance, using different windowing, dark streaks can be seen along the edges of the plexi plate. Additionally, a number of streaks can be seen radiating from the metals. Figure 13: Simulation of object motion G. Aliasing Figure 14a shows a simulation of phantom 1 using only half of the detectors for the reconstruction. Clearly, typical artifacts due to detector under-sampling (circular patterns tangent to strong edges) can be seen. Similarly, figure 14b shows a simulation of phantom 1 using only a quarter of the views for the reconstruction. Again typical artifacts due to view under-sampling (streaks starting at a certain distance from the center) are observed. Figure 11: Simulation with EEGE: (a) phantom 1 and (b) phantom 2 E. Noise Figure 12: Error due to noise determined by simulations with and without noise: (a) phantom 1 and (b) phantom 2 (windowing: Œ;Ž # # e šlœ ) Figure 12 shows the error due to noise determined from simulations with and without noise. Streaks can be seen in directions of highest attenuation and connecting metal objects. Figure 14: Simulations with (a) detector under-sampling and (b) view under-sampling Note that even in the artifact-free simulations (figure 5), aliasing errors (such as Moire patterns) are present, although not visible with the current windowing. However, appropriate reconstruction kernels should be able to reduce aliasing artifacts as much as wanted, possibly at the expense of the spatial resolution of the reconstruction. Moreover, neither the aliasing artifacts, nor the artifacts due to object motion are limited to the case of metal objects, but because of their high attenuation values, the presence of metal objects makes the artifacts more prominent. IV. FUTURE WORK For the scope of this article, we focussed on one specific scanner type. Results must be generalized for other scanner

types. Extension to 3D, and in particular analysis of the axial partial volume effect is needed. The next step is to try to reduce streak artifacts. One possible approach is to prevent artifacts during the acquisition. Noise can be reduced by using high mas-settings. Beam hardening can be minimized by using pre-filtering. The partial volume effect due to z-gradients can be reduced by making thin slices. Unfortunately, all these measures are in direct conflict with other clinical concerns such as low patient dose, tube life, etc... Our aim is to apply iterative reconstruction [19] to reduce metal streak artifacts. The major advantage of this approach is the possibility to use a model of the acquisition, taking into account polychromaticity, scatter, noise, EEGE and any other imperfections. A general iterative reconstruction scheme is shown in figure 15. Figure 15: Iterative Reconstruction V. CONCLUSIONS A high-resolution 2D CT simulator has been developed and validated. We have proven that all investigated causes of streaks are effectively able to produce streaks. Beam hardening, scatter, noise and EEGE are the most important causes of metal streak artifacts. Under extreme circumstances, also object motion and aliasing effects produce streak artifacts. Artifacts due to axial gradients need further investigation. VI. ACKNOWLEDGMENTS This work is supported by the Flemish Fund for Scientific Research (FWO), grant number G.0106.98. P. Dupont is postdoctoral researcher of the FWO. We wish to thank the people from Siemens (Forchheim) for the information about their CT scanners. We also wish to thank François De Cock, Patrick Lievens and Philippe Vervloesem (Dept. of Radiology) for their professional assistance in the experiments. Finally, we wish to thank Els Tijskens for the production of the Plexiglas phantom and Anne Vander Speeten for the use of the water phantom. VII. REFERENCES [1] D.D.Robertson, J.Yuan, G.Wang and M.W.Vannier, Total Hip Prosthesis Metal-Artifact Suppression Using Iterative Deblurring Reconstruction, J. Comput. Assist. Tomogr., vol. 21, 1997 pp. 293-298 [2] G.Wang, D.L.Snyder, J.A.O Sullivan and M.W.Vannier, Iterative Deblurring for CT Metal Artifact Reduction, IEEE Trans. Med. Imaging, vol. 15, 1996 pp. 657-664 [3] W.A.Kalender, R.Hebel and J.Ebersberger, Reduction of CT Artifacts Caused by Metallic Implants, Radiol., vol. 164, 1987 pp. 576-577 [4] A.J.Duerinckx and A.Macovski, Nonlinear Polychromatic and Noise Artifacts in X-Ray Computed Tomography Images, J. Comput. Assist. Tomogr., vol. 3, 1979 pp. 519-526 [5] A.J.Duerinckx and A.Macovski, Polychromatic Streak Artifacts in Computed Tomography Images, J. Comput. Assist. Tomogr., vol. 2, 1978 pp. 481-487 [6] P.M.Joseph and R.D.Spital, A Method for Correcting Bone Induced Artifacts in Computed Tomography Scanners, J. Comput. Assist. Tomogr., vol. 2, 1978 pp. 100-108 [7] D.D.Robertson and H.K.Huang, Quantitative bone measurements using x-ray computed tomography with second-order correction, Med. Phys., vol. 13 no. 4, 1986 pp. 474-479 [8] J.M.Meagher, C.D.Mote and H.B.Skinner, CT Image Correction for Beam Hardening Using Simulated Projection Data, IEEE Trans. Nucl. Sci., vol. 37 no. 4, 1990 pp. 1520-1524 [9] P.M.Joseph and C.Ruth, A method for simultaneous correction of spectrum hardening artifacts in CT images containing both bone and iodine, Med. Phys., vol. 24 no. 10, 1997 pp. 1629-1634 [10] P.M.Joseph and R.D.Spital, The effects of scatter in x-ray computed tomography, Med. Phys., vol. 9, 1982 pp. 464-472 [11] G.H.Glover, Compton scatter effects in CT reconstructions, Med. Phys., vol. 9, 1982 pp. 860-867 [12] P.M.Joseph and R.D.Spital, The exponential edgegradient effect in x-ray computed tomography, Phys. Med. Biol., vol. 26, 1981 pp. 473-487 [13] P.M.Joseph, R.D.Spital and C.D.Stockham, The effects of sampling on CT images, Computerized Tomography, vol. 4, 1980 pp. 189-206 [14] P.M.Joseph and R.A.Schulz, View sampling requirements in fan beam computed tomography, Med. Phys., vol. 7, 1980 pp. 692-702 [15] G.H.Glover and N.J.Pelc, An algorithm for the reduction of metal clip artifacts in CT reconstructions, Med. Phys., vol. 8, 1981 pp. 799-807 [16] G.H.Glover and N.J.Pelc, Nonlinear partial volume artifacts in x-ray computed tomography, Med. Phys., vol. 7 no. 3, 1980 pp. 238-248 [17] G.Wang and M.W.Vannier, Stair-Step Artifacts in Threedimensional Helical CT: An Experimental Study, Radiol., vol. 191 no. 1, 1994 pp. 79-83

[18] J.A.Brink, Technical aspects of helical (spiral) CT, Radiologic Clinics of North America, vol. 30 no. 5, 1995 pp. 825-841 [19] J.Nuyts, B.De Man, P.Dupont, M.Defrise, P.Suetens and L.Mortelmans, Iterative reconstruction for helical CT: a simulation study, Phys. Med. Biol., vol. 43, 1998 pp. 729-737 [20] D.J.Drost and A.Fenster, A xenon ionization detector for digital radiography, Med. Phys., vol. 9 no. 2, 1982 pp. 224-230 [21] G.T.Herman, Image Reconstruction From Projections, Orlando: Academic Press, 1980, pp. 161-179