A HAAG-STREIT COMPANY MÖLLER-WEDEL OPTICAL GmbH, Rosengarten 10, D-22880 Wedel with Control Unit Ident.-No. 773 190 01 Valid from Hardware / Software Rev. / SN 1.0 IDENT.-NO. 217 321 09 Document Version 08.03.2013
Content 1. IMPORTANT INFORMATION... 3 Liability to Functions and Damage... 4 2. INTENDED USE... 5 2.2 Description of modules... 5 2.2.1 Control Unit... 5 2.2.2 LED-Head... 6 3. ASSEMBLING / STARTING UP... 7 3.1 Unpacking and inspection of delivery... 7 3.2 Starting up... 7 4. TECHNICAL DATA... 7 CE EG Declaration of conformity WEEE and RoHS declaration of conformity Ident.-Nr.: 217 321 09 2
1. Important information Safety Remarks This equipment must only be used following the instruction manual. Please read the instructions first. It is only allowed to use the equipment in the intended way. The equipment is only for use in clean and dry environment. Also see next paragraph. The valid accident prevention regulations of Employer s for Electrical Systems and Operating Materials are to be observed. Special accident prevention regulations may occur out of the specific use of the instrument and have to be considered. The year of manufacturing and the serial number of the equipment is documented on the identification label. Keep this manual for later use. Safety conditions for operating the Equipment The following conditions are to observe for save operating of the instrument: Only for indoor use. Use only in clean and dry environment. No conductance dirt cover or condensation. Equipment must be adapted to the operating conditions in time to avoid any condensation. Operating Temperature 5-40 C. Humidity less 80% up to 32 C linear decreasing to 50% at 40 C. Height above zero normal not more than 2000 m. Tolerance of mains voltage not above 10% from nominal value. Mains voltage 100-240V~ AC depending on country. Operating the equipment in mains, where transient over voltage peaks exceed the common values (more than allowed according to Over Voltage Class III according to EC664) is forbidden. It is not allowed to cover power supplies, to short-circuit the output or to pull them by cable. Defect or damaged supplies and cables must be disabled and substituted by original MÖLLER-WEDEL OPTICAL spare parts. Power plugs are only allowed to clean with a soft and dry cloth. Before cleaning disconnect them from the main voltage and the equipment. Do not short-circuit batteries or accumulators. Do not burn or open. Only use sealed batteries Additional restrictions in use may result of the measurement application and are listed in the technical data. There you will also find other base data. MAINTENANCE AND CARE Modifications or maintenance must only be carried out by persons explicitly authorised by MÖLLER-WEDEL OPTICAL GmbH. Only original parts of MÖLLER-WEDEL OPTICAL must be used for maintenance. After maintenance or technical modifications the equipment must be re-adjusted according to the technical instructions. In case of technical inquiries the numbers on the relevant parts must be indicated. Before cleaning or maintenance always disconnect from the mains and other equipment Ident.-Nr.: 217 321 09 3
Liability to Functions and Damage If the equipment is modified or repaired by not explicitly authorised persons, in case of improper maintenance (as far as not performed by MÖLLER-WEDEL OPTICAL) or in case of improper handling, any liability of MÖLLER-WEDEL OPTICAL is excluded. Accessories Electrically driven accessories are permitted on the equipment only if its technically safe application is documented by a notified person (relevant certificate must be available). Declaration of Conformity The declarations according to CE, environment and disposal are attached in the end of the manual. Ident.-Nr.: 217 321 09 4
2. Intended Use The is used as light source in Autocollimators and Collimators. Function description 2.1 Description of / Overview The consists of: Control Unit (Fig. 1.1), One or two LED-heads (Fig. 1.2) and Power supply (Fig. 1.3). 3 2 1 Fig. 1: Components of. 2.2 Description of modules 2.2.1 Control Unit At the control unit the power supply and one or two LED heads will be connected. This type of illumination is working besides with supplied power supply also with 4 NiCd- or NiMH-accumulators - type AA - (not included in delivery). The accumulators can be inserted in the battery case at the back of the case of Control Unit (attend to polarity!). Thereby a mains-independent operation is possible. In case of connecting up a power supply in the battery circuit is switched off. Please note that the batteries can not be charged by the control unit. An external charger is needed. Ident.-Nr.: 217 321 09 5
Use only NiCd- or NiMH-accumulators. Conventional batteries can damage the LED Head due to the higher voltage. With the lower two switches (s. Fig. 2.1) the LED Heads can separately be switched on and off. The regulation of the brightness is also carried out separately for both LED heads through the two upper switches(s. fig. 2.2). Rotating the upper switches clockwise increases the LED intensity, rotating the switches counterclockwise reduces the intensity 2 1 Fig. 2: Control Unit 2.2.2 LED-Head The LED-Head consists of a cable and a LED-head with five high performance LEDs. Four versions with the colours green, red, blue and white are available. The dominant wavelength is given in the technical data. Security advice : Due to the high brightness avoid direct eye contact. Do not look into the bright beam. 2.2.3 Power supply The power supply is for supplying the control unit with power Only use the supplied power supply! Using other power supplies can cause damage on the Control Unit and the LEDs. Ident.-Nr.: 217 321 09 6
3. Assembling / Starting up 3.1 Unpacking and inspection of delivery The is available in 8 versions: Description Article No.. - green - incl. control unit and universal power plug 217 321 - red - incl. control unit and universal power plug 217 322 - blue - incl. control unit and universal power plug 217 323 - white - incl. control unit and universal power plug 217 324 2-arms, green - incl control unit and universal power plug 217 325 2-arms, red - incl. control unit and universal power plug 217 326 2 arms, blue - incl. control unit and universal power plug 217 327 2-arms, white - incl. control unit and universal power plug 217 328 Each version contains: one Control Unit one (single arm) or two (two armed) LED-Heads and one power supply. Check the delivery immediately after receipt for completeness and damage 3.2 Starting up Connect the LED Head(s) and the Power Supply with the Control Unit. Connect the power supply with the mains. As described under 2.2.1 accumulators can alternatively be used. Install the LED-Head in the condenser mount of the Collimator or Autocollimator. Turn on the with the switches described (fig. 2.1) and adjust the brightness to the desired value 4. Technical Data Red green blue Dominant wavelength 640 nm 520 nm 470 nm Half-width of the spectral output 20 nm 40 nm 30 nm Power Supply Power Supply 100V-240VAC/5VDC optional:4 NiMH- or NiCd-accumulators Ident.-Nr.: 217 321 09 7
Ident.-Nr.: 217 321 09 8
Ident.-Nr.: 217 321 09 9