Oscilloscope System Recovery

Similar documents
Read This First *P *

DSA8300 Digital Serial Analyzer Microsoft Windows 7 Ultimate Operating System Restore Instructions

How to install the software of ZNS8022

Acronis Backup & Recovery 11 Beta Advanced Editions

Image Backup and Recovery Procedures For Windows 7

CASPER SECURE DRIVE BACKUP

Reinstalling the Operating System on the Dell PowerVault 745N

How to Use imageprograf Firmware Update Tool (Version or later) (Mac OS) Canon Inc.

C A S P E R USER GUIDE V ERSION 5.0

Pro-face/Xycom Workstation Recovery Media Software Installation Instructions for Microsoft MS-DOS 6.22

GV Director Restoring USB Image / Frame and Panel Installation

Lab: Advanced Installation of Windows XP. Introduction

Instructions to update Digimax application software (a.k.a. firmware)

Acronis Disk Director 11 Home. Quick Start Guide

AWG5000 and AWG7000 Series Windows XP Operating System Restore

Lab - Install Windows 7 or Vista

Instrument Software Update Instructions

menu of icons which you can select which item you want to use to startup the computer.

CASPER SECURE SERVER EDITION 3.0 USER GUIDE

NEC Versa 500D/550D Windows 98 Upgrade Information

Format Hard Drive Using Windows 7 Recovery Disk

Appendix A: Metasys Web Access (MWA) Server Installation Guidelines

Boot Camp Installation & Setup Guide

Applied ICT Skills MS Windows

Acronis Backup & Recovery 10 Advanced Editions

Lesson 1: Preparing for Installation

Parallels Desktop 4.0 Switch to Mac Edition. Migrate your PC Tutorial.

Instructions For Formatting Hard Drive Windows 7 Command Prompt

Basic Edition A Windows Embedded Partner

C A S P E R USER GUIDE V ERSION 10

HOW TO: Perform a Hardware Independent Restore (HIR)

Flex Focus 1202 Software Installation Procedure

Integrating RDX QuikStor into Windows Backup with Full System Recovery for Windows Server OS

AOEMI Backupper V4.0

OS Update. Application. User s Guide. (ClassPad OS Version 3.04)

Veritas System Recovery Disk Help

Boot Camp Installation & Setup Guide

NEC Versa 4200 Windows 98 Upgrade Information

Fedora 12. For guidelines on the permitted uses of the Fedora trademarks, refer to fedoraproject.org/wiki/legal:trademark_guidelines.

Intel Server RAID Controller U2-1 Integration Guide For Microsoft* Windows NT* 4.0

Boot Camp Installation & Setup Guide

CONTENTS. SysReturn_Manaual Updata 06/4/24. Chapter1 Introducing SysReturn Introduction to SysReturn Features and Functions...

Fiery proserver User Guide

Instrument Software Update Instructions. X-Series Multi-touch instruments

A+ Guide to Software: Managing, Maintaining, and Troubleshooting, 5e. Chapter 7 Fixing Windows Problems

CASPER SECURE DRIVE BACKUP. for BitLocker Drive Encryption S TARTUP D ISK C REATOR G UIDE

Q-Stress 4.6 Software Installation & Upgrade

Lionstracs Groove X-R

Fiery proserver User Guide

How To Force Restore A Computer That Won Boot Up After System

Veritas System Recovery Disk Help

3 INSTALLING WINDOWS XP PROFESSIONAL

Disk Imaging with Knoppix

A+ Guide to Managing and Maintaining Your PC, 7e. Chapter 16 Fixing Windows Problems

Home. User Guide. Home. Laplink Software, Inc. User Guide MN-PCMHOME-EN-08 (REV. 1/15/2014)

IT Essentials v6.0 Windows 10 Software Labs

How to update OS image from V1421 to V1501 by Exe file

Stage 1 - The text based setup - Basic configuration (Partitioning and Formatting)

C A S P E R T ECH EDITION 5.0 USER GUIDE

NG6000 Firmware Update Process

Free. User Guide. Free. Laplink Software, Inc. User Guide MN-PCMFREE-EN-08 (REV. 2/2013)

8 MANAGING SHARED FOLDERS & DATA

Professional. User Guide. Professional. User Guide MN-PCMPRO-EN-06 (REV. 07/2010)

Full System Restore Manually Windows 7 No Disk

HDD Recovery Information for Your Toshiba Computer

CONTENTS. Features and Functions...3. Preparation...7. Proceed Installation (Hardware Version)...7. Proceed Installation (Software Version)...

Windows 2000 Safe Mode

Quick Setup & Getting Started

Parallels Desktop for Mac

User Guide. Rebit Backup.

Laplink PCmover Windows Upgrade Assistant

UPS WorldShip Install on a Workgroup Remote

User Guide. PCmover Professional for Windows XP. Laplink Software, Inc. User Guide MN-PCMPRO-XP-EN-08 (REV. 2014/07/08)

Intel Solid State Drive Firmware Update Tool

User's Guide. User Notice. Copyright

Intel Entry Storage System SS4000-E

Using iscsi with BackupAssist. User Guide

Infinova SS200T3R Embedded VMS Box

SAS Installation Instructions Windows 2003, XP, 2000, NT. Workstation Installation Guidelines

3. IMPORTANT: When prompted, accept the default Installation directory.

Full System Restore Manually Run Cmd Prompt

Hitachi Align Tool. User's Guide

Home. User Guide. Home. Laplink Software, Inc. User Guide MN-PCMHOME-EN-08 (REV. 10/2012)

LexisNexis PCLaw Installation & Upgrade Guide

Supplementary Instructions for Windows NT 4.0. Contents

FAQ No Q: What should we know before we start Windows 10 upgrade?

PGP NetShare Quick Start Guide Version 10.2

Perform Manual System Restore Xp Safe Mode Winxp

PARAGON ADAPTIVE RESTORE

Password Changer User Guide

Agilent MassHunter Workstation Software for Triple Quad LC/MS

Getting Started. HP Business PCs

OS Installer. Installation Instructions

8000 Series Instruments Windows 2000 Operating System Restore

Quick Start Guide. Laplink Software, Inc. Quick Start Guide MN-PCMVR-EN-01 (REV. 02/07) w w w. l a p l i n k. c o m / s u p p o r t

Maxxyz Compact Operating System Installation

This installation guide is intended for customers who are installing NMIS for the first time. Included are the recommended hardware specifications

NEC Versa LX (AGP Video) Important Information

InfoWatch CryptoStorage. User Guide

Firmware Update Guide

Transcription:

Instructions Oscilloscope System Recovery The oscilloscope application (firmware) runs on a Microsoft Windows platform. While the instrument is designed to operate reliably for many years, the loading or incomplete removal of additional Windows applications can cause problems in the stability of the operating system. Severe cases may require you to reload the base operating system and firmware. For this purpose, Teledyne LeCroy provides a recovery application and a backup image. The recovery procedure is easy to perform. Both the application used and the delivery method vary by oscilloscope model and year. Although the recovery procedures may seem similar, there are important differences, and using the incorrect procedure can cause system failure. Follow these guidelines to choose the correct procedure. Model Series Windows OS Delivery Method Procedure WaveSurfer 3000 CE N/A N/A HDO4000/HDO4000A HDO6000/HDO6000A 10 Partition on hard drive HDO4000/HDO6000 System Recovery HDO8000/HDO8000A MDA800/MDA800A 10 USB Key (see Note) System Recovery from USB Key DDA/SDA/WM 8 Zi-B 10 All others 10 Vista XP USB Key Partition on hard drive System Recovery from USB Key System Recovery from Hard Drive All others 2000 CD Contact Service Note: Some HDO8000 and MDA800 instruments were shipped without the USB recovery key. However, it is crucial to use the USB key recovery procedure. Contact support@teledynelecroy.com for instructions on creating a recovery key. After Recovery... The recovery process produces a replica of the software at the time of manufacture. It does not automatically update the: Oscilloscope firmware Windows operating system Virus scan definition files Drivers for some hardware options and accessories Therefore, after completing the disk image recovery, it is highly recommended to search vendor websites and upgrade the individual components to current revision levels. The latest oscilloscope firmware can be downloaded from: teledynelecroy.com/support/softwaredownload under Oscilloscope Downloads > Firmware Upgrades. Most required hardware drivers can be installed during the firmware upgrade. Application and calibration data is stored on the D: drive. Current calibration constants are not overwritten during recovery of the C: drive. You do not need to restore these. Contact Teledyne LeCroy For the most complete and up-to-date list of offices by country, visit: teledynelecroy.com/support/contact 926284 Rev B, Page 1

HDO4000/HDO6000 Recovery This procedure is for HDO4000/HDO4000A and HDO6000/HDO6000A models only. This procedure will destroy all data on the C: drive. If your drive is still accessible, first back up any files you may want to restore after recovering. Before You Start Connect a keyboard and mouse to the instrument. Power down the instrument. Note the serial number on the label on the back of the instrument. Using True Image Home 1. Power on the instrument. 2. During the startup process, as soon as you see "Starting Acronis Loader...", press the F11 key until the Acronis logo appears momentarily. The Acronis window is displayed. Do not press F11 before you see "Starting Acronis Loader..." If a boot menu dialog appears, press Cancel or Esc. 3. Select Acronis True Image Home (Full Version). 4. On the Acronis True Image Home page, under options for Recover, select My Disks. The Recovery Wizard opens. 5. On the Recovery Wizard, under Archive Selection, select the disk archive with a create date, then click 6. Under Recovery Method, select Recover whole disks and partitions, then click 7. Under What to Recover, select NTFS (SYSTEM) (C:), then click 8. Under Settings of Partition C, in the top section, Partition location (required), select New Location.The Partition Destination window opens. 9. Under New Partition Location, select NTFS (SYSTEM) (D:), then click Accept. This returns you to the Settings of Partition C step. Click 10. A summary window indicates that Acronis True Image is ready to proceed with recovering partition C -> D. Click Proceed to start the recovery. Recovery takes 4-15 minutes. Do not power down or otherwise interrupt the recovery while in process. 11. When recovery is complete, you will see "Recover operation succeeded." Click OK. Oscilloscope System Recovery Instructions 12. Click the close button to exit the Acronis window. The instrument will restart and begin installing the required software. If asked to restart Windows, select Restart Later. Reinstall the X-Stream Application 1. When the X-Stream DSO Setup Wizard appears, click 2. On the License Agreement page, select I Agree. 3. On the Choose Components page, select the default (installation) and click Install. If you see a message that Windows can't verify the publisher of the driver software, choose "Install this driver software anyway" and click Install. 4. When the X-Stream installation is complete, choose to Reboot now and click Finish. Rename Instrument In order to apply option keys to the instrument in the future, restore the correct name (serial number). 1. Minimize the oscilloscope application window. 2. From the Windows Start Menu, choose Control Panel > System. 3. Scroll down to Computer name, domain, and workgroup settings, then click Change Settings. 4. On the System Properties dialog, click Change and enter the oscilloscope serial number in Computer name. 5. Click OK twice to return to the desktop. Reactivate the F11 Startup Utility In order for the system recovery wizard to be accessed again from the boot menu, reactivate the F11 startup utility. It is critical to complete these steps after restarting. 1. From the Windows Start Menu choose All Programs > Acronis > Acronis True Image Home. 2. On the Acronis True Image Home window, towards the top right, click Tools & Utilities. 3. On the Tools & Utilities page, beneath Protection Tools, click Acronis Startup Recovery Manager. 4. On the Acronis Startup Recovery Manager window, click Activate. F11 boot time recovery is now enabled. Note: The process may take several minutes. Page 2

System Recovery from USB Key This procedure is for MDA800/MDA800A, HDO8000/HDO8000A, and DDA/SDA/WaveMaster 8 Zi-B (not Zi/Zi-A) models. If your Acronis recovery key is lost, contact support for instructions on creating a replacement key. This procedure will destroy all data on the C: drive. If your drive is still accessible, first back up any files you may want to restore after recovering. You must re-activate Windows after recovery is complete, either by Internet connection to Microsoft's website or by telephone. Have your Windows Product Key number handy for Windows reactivation. Before You Start Note the Windows Product Key number and serial number, usually listed on a sticker on the back of the instrument. If you intend to reactivate Windows through the Internet, connect the instrument to your network. Power down the instrument. Connect a keyboard to the instrument. Remove any mouse that is already connected. Do not connect a mouse until instructed later in the procedure. Boot from USB Key 1. Connect the supplied USB key to the instrument. 2. Power on the instrument. 3. During the startup process, as soon as you see "F2 for System Utilities," press the F2 key to enter the BIOS menu. If you miss the F2 prompt and begin to boot into Windows, shut down and begin again. 4. Enter the BIOS password lcrsys (lower case). 5. When on the BIOS menu, press the Cursor Right key until Boot menu is selected. 6. On the Boot menu, press the Cursor Down key until USB HDD is selected. 7. Press the Plus + key to move USB HDD to the top of the list. 8. Press the F10 key, then press Enter to save. The instrument will boot into the Acronis environment. 9. Once at the Acronis loader page, press the Cursor Up key to select Acronis True Image Home 2015, then press Enter. Oscilloscope System Recovery Instructions 10. When you see the Acronis True Image Home window, connect a mouse to the instrument. Run the Recovery Wizard 1. On the Acronis Home page, under options for Recover, select My Disks. The Recovery Wizard opens. 2. On the Recovery Wizard, under Archive Selection, select the disk archive with a create date, then click 3. Under Recovery Method, select Recover whole disks and partitions, then click 4. Under What to Recover, select NTFS (SYSTEM) (C:), then click 5. Under Settings of Partition C, in the section Partition location (required), select New Location. The Partition Destination window opens. 6. Under New Partition Location, select NTFS (SYSTEM) (C:), then click Accept. This returns you to the Settings of Partition C. Click 7. A summary window indicates that Acronis True Image is ready to proceed with recovering C: > C:. Click Proceed to start the recovery. Recovery takes 4-15 minutes. Do not power down or otherwise interrupt the recovery while in process. 8. When recovery is complete, you will see "Recovery operation succeeded." Click OK. 9. Click the close button to exit the Acronis window. The instrument will begin to shut down and restart. Restore the Boot Sequence 1. During the startup process, as soon as you see "F2 for System Utilities," press the F2 key to enter the BIOS menu. If you miss the F2 prompt and begin to boot into Windows, shut down and begin again. 2. Enter the BIOS password lcrsys (lower case). 3. When on the BIOS menu, press the Cursor Right key until Boot menu is selected. 4. On the Boot menu, press the Cursor Down key until ATA HDD 0 is selected. 5. Press the Plus + key to move ATA HDD 0 to the top of the list. 6. Press the F10 key, then press Enter to save. The instrument will begin to reboot. Page 3

Oscilloscope System Recovery Instructions Reinstall the Oscilloscope Application 1. When the X-Stream DSO Setup Wizard appears, click 2. On the License Agreement page, select I Agree. 3. On the Choose Components page, select the default (installation) and click Install. If you see a message that Windows can't verify the publisher of the driver software, choose "Install this driver software anyway" and click Install. 4. When the X-Stream installation is complete, choose to Reboot now and click Finish. 5. When prompted, enter your Windows Product Key number and follow the Microsoft procedure to reactivate Windows. Note: The Windows re-activation may occur before or after the X-Stream installation. The order does not matter. Rename Instrument In order to apply option keys to the instrument in the future, restore the correct name (serial number). 1. Minimize the oscilloscope application window. 2. From the Windows Start Menu, choose Control Panel > System. 3. Scroll down to Computer name, domain, and workgroup settings, then click Change Settings. 4. On the System Properties dialog, click Change and enter the oscilloscope serial number in Computer name. 5. Click OK twice to return to the desktop. Page 4

System Recovery from Hard Drive Do not use this procedure for HDO4000s, HDO6000s, HDO8000s, MDA800s, DDA/SDA/WaveMaster 8 Zi-Bs, or any instruments running Windows 2000. This procedure will destroy all data on the C: drive. If your drive is still accessible, first back up any files you may want to restore after recovering. You must reactivate Windows after recovery is complete, either by Internet connection to Microsoft's website or by telephone. Have your Windows Product Key number handy for Widows reactivation. Important! Depending on the year and model, your oscilloscope may be installed with Acronis True Image Home, Acronis True Image Echo Workstation, or Phoenix cme PC Edition. It is critical to check which application you are using, as following the incorrect procedure can cause system failure. Before You Start If you intend to reactivate Windows through the Internet, connect the instrument to your network. Connect a keyboard and mouse to the instrument. Note the Windows Product Key number and serial number, usually listed on a sticker on the back of the instrument. Note: In LabMasters, the Windows Product Key is found on the removable hard drive. Power down the oscilloscope before removing the hard drive to find the product key. Replace the drive before powering on again at the start of the recovery. Power down the instrument. Determine the Recovery Product 1. Power on the instrument. 2. During the startup process, when you see "Starting Acronis Loader...", press the F11 key until the Acronis logo appears momentarily. The Acronis window is displayed. Note: Do not press F11 before you see "Starting Acronis Loader..." If a boot menu dialog appears, press Cancel or Esc. Check the Acronis window to determine which product you are using. Follow Using True Image Home or Using Echo Workstation accordingly. Or Oscilloscope System Recovery Instructions If you do NOT see "Starting Acronis Loader..." during your boot sequence, press the F4 key when the LeCroy logo appears and follow Using Phoenix cme. Using True Image Home 1. Select Acronis True Image Home (Full Version). 2. On the Acronis True Image Home page, under options for Recover, select My Disks. The Recovery Wizard opens. 3. On the Recovery Wizard, under Archive Selection, select the disk archive with a create date, then click 4. Under Recovery Method, select Recover whole disks and partitions, then click 5. Under What to Recover, select NTFS (SYSTEM) (C:), then click 6. Under Settings of Partition C, in the top section, Partition location (required), select New Location.The Partition Destination window opens. 7. Under New Partition Location, select NTFS (SYSTEM) (D:), then click Accept. This returns you to the Settings of Partition C step. Click 8. A summary window indicates that Acronis True Image is ready to proceed with recovering partition C -> D. Click Proceed to startrecovery. Recovery takes 4-15 minutes. Do not power down or otherwise interrupt the recovery while in process. 9. When recovery is complete, you will see "Recover operation succeeded." Click OK. 10. Click the close button to exit the Acronis window. The instrument will restart and begin installing the required software. Continue to reinstall the X-Stream application. Note: If asked to restart Windows, select Restart Later. Using Echo Workstation 1. On the Pick a Task page, select Recovery. The Restore Data Wizard opens. 2. Click Next on the Welcome page. 3. On Backup Archive Selection, choose Acronis Secure Zone (this is where the recovery data is located on the instrument). Click Page 5

Oscilloscope System Recovery Instructions 4. On Restoration Type Selection, select Restore disks or partitions, then click 5. On Partition or Disk to Restore, select SYSTEM (C:) as the source, then click 6. On Restored Partition Location, select SYSTEM (C:), then click 7. On Restored Partition Type, select Active, then click Note: If asked if you want to buy Acronis products, check "Do not show this message again" and click OK. 8. On Restored Partition Size, simply click 9. On Next Selection, select No, I do not (want to restore another partition), then click 10. On Restoration Options, click Next to use the default selections. 11. A summary window indicates that Acronis True Image is ready to proceed with the recovery of the C: partition. Click Proceed to start the recovery. Recovery takes 4-15 minutes. Do not power down or otherwise interrupt the recovery while in process. 12. Once the recovery is complete, an Information dialog indicates that the Data was successfully restored. Click OK. 13. From the Acronis menu bar, choose Operations > Exit. The oscilloscope will restart and begin installing the required software. Using Phoenix cme 1. When the cme console End User License Agreement is shown, read the agreement and click Accept. 2. On the Phoenix cme Console main page, click the link "Click here to start recover". The recovery starts shortly after the FirstWave flash page, and the FirstWare Progress screen is displayed. No further selections are required.the recovery takes about 10 minutes. Note: The screen occassionally goes blank for prolonged periods. This is normal and not an indication of malfunctioning. Reinstall the X-Stream Application 1. When the X-Stream DSO Setup Wizard appears, click 2. On the License Agreement page, select I Agree. 3. On the Choose Components page, select the default (installation) and click Install. Note: If you see a message that Windows can't verify the publisher of the driver software, choose "Install this driver software anyway" and click Install. 4. When the X-Stream installation is complete, choose to Reboot now and click Finish. 5. When prompted, enter your Windows Product Key number to re-activate Windows. Rename Instrument In order to apply option keys to the instrument in the future, restore the correct name (serial number). 1. Minimize the oscilloscope application window. 2. From the Windows Start Menu, choose Control Panel > System. 3. Scroll down to Computer name, domain, and workgroup settings, then click Change Settings. 4. On the System Properties dialog, click Change and enter the oscilloscope serial number in Computer name. 5. Click OK twice to return to the desktop. Reactivate the F11 Startup Utility Phoenix and Echo Workstation users can skip this procedure and go on to update software to current levels. In order for the system recovery wizard to be accessed again from the boot menu, reactivate the F11 startup utility. It is critical to complete these steps after restarting. 1. From the Windows Start Menu choose All Programs > Acronis > Acronis True Image Home. 2. On the Acronis True Image Home window, towards the top right, click Tools & Utilities. 3. On the Tools & Utilities page, beneath Protection Tools, click Acronis Startup Recovery Manager. 4. On the Acronis Startup Recovery Manager window, click Activate. F11 boot time recovery is now enabled. Note: The process may take several minutes. Page 6

Oscilloscope System Recovery Instructions 2017 Teledyne LeCroy, Inc., Inc. All rights reserved. Teledyne LeCroy is a trademark of Teledyne LeCroy, Inc. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice. April, 2017 Page 7