Non-Stationary CT Image Noise Spectrum Analysis

Similar documents
Spectral analysis of non-stationary CT noise

Scaling Calibration in the ATRACT Algorithm

Edge-Preserving Denoising for Segmentation in CT-Images

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

2D Fan Beam Reconstruction 3D Cone Beam Reconstruction. Mario Koerner

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

DUE to beam polychromacity in CT and the energy dependence

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Estimation of Missing Fan-Beam Projections using Frequency Consistency Conditions

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

Iterative CT Reconstruction Using Curvelet-Based Regularization

THE FAN-BEAM scan for rapid data acquisition has

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Radon Transform and Filtered Backprojection

Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences

Feldkamp-type image reconstruction from equiangular data

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Suitability of a new alignment correction method for industrial CT

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

An approximate cone beam reconstruction algorithm for gantry-tilted CT

Translational Computed Tomography: A New Data Acquisition Scheme

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Introduction to Biomedical Imaging

GE s Revolution CT MATLAB III: CT. Kathleen Chen March 20, 2018

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Index. aliasing artifacts and noise in CT images, 200 measurement of projection data, nondiffracting

CT vs. VolumeScope: image quality and dose comparison

Adaptive algebraic reconstruction technique

Revisit of the Ramp Filter

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

The n-pi-method for Helical Cone-Beam CT

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Central Slice Theorem

Frequency-based method to optimize the number of projections for industrial computed tomography

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Image Reconstruction from Projection

APPLICATION OF RADON TRANSFORM IN CT IMAGE MATCHING Yufang Cai, Kuan Shen, Jue Wang ICT Research Center of Chongqing University, Chongqing, P.R.

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

GPU-based cone-beam CT reconstruction with extended FOV

Cover Page. The handle holds various files of this Leiden University dissertation

Quantitative image-based spectral reconstruction for computed tomography

Henry Ford NERS/BIOE 481. Lecture 11 B Computed Tomography (CT)

Accelerated C-arm Reconstruction by Out-of-Projection Prediction

A Novel Two-step Method for CT Reconstruction

USING cone-beam geometry with pinhole collimation,

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Modern CT system generations Measurement of attenuation

Reconstruction in CT and hints to the assignments

2D Vessel Segmentation Using Local Adaptive Contrast Enhancement

Portability of TV-Regularized Reconstruction Parameters to Varying Data Sets

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

High Performance GPU-Based Preprocessing for Time-of-Flight Imaging in Medical Applications

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

Noise weighting with an exponent for transmission CT

Ch. 4 Physical Principles of CT

Digital Image Processing

Scatter Correction Methods in Dimensional CT

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

SPECT (single photon emission computed tomography)

Total Variation Regularization Method for 3D Rotational Coronary Angiography

A practical software platform of integrated reconstruction algorithms and forward algorithms for 2D industrial CT

Basics of treatment planning II

CoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Nondestructive tests of cylindrical steel samples using the ultrasonic projection method and the ultrasound transmission tomography method

Corso di laurea in Fisica A.A Fisica Medica 4 TC

On the Characteristics of Helical 3-D X-ray Dark-field Imaging

Geometry-Based Optic Disk Tracking in Retinal Fundus Videos

Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System

Ultrasonic Multi-Skip Tomography for Pipe Inspection

Ray Contribution Masks for Structure Adaptive Sinogram Filtering

A prototype table-top inverse-geometry volumetric CT system

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

Computer-Tomography II: Image reconstruction and applications

An FDK-like cone-beam SPECT reconstruction algorithm for non-uniform attenuated

Cover Page. The handle holds various files of this Leiden University dissertation

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

MEDICAL IMAGING 2nd Part Computed Tomography

X-ray tomography. X-ray tomography. Applications in Science. X-Rays. Notes. Notes. Notes. Notes

Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data

Artifact analysis of approximate helical cone-beam CT reconstruction algorithms

A method and algorithm for Tomographic Imaging of highly porous specimen using Low Frequency Acoustic/Ultrasonic signals

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay

Quality control phantoms and protocol for a tomography system

Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Transcription:

Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de Abstract. We investigate the spatial dependency of noise characteristics in CT images. The perceived image quality depends on the noisegranularity. Especially in low-dose applications the noise granularity influences the diagnostic value. A model is presented, that provides two-dimensional, stationary noise realizations for arbitrary image pixel locations from which two-dimensional Noise Power Spectrum estimates can be computed. It fully incorporates the CT reconstruction process for (indirect) fan-beam reconstruction, the quarter offset of the detector channels and the detector noise characteristics. It can be used with simulated and measured data and allows for the assessment of spectral noise characteristics for arbitrary objects. Introduction The noise in the reconstructed CT image is influenced by various parameters. When planning components of a Computed Tomography scanner, like detectors or X-ray tubes, it is necessary to tune their parameters in a way that leads to the best achievable objective image quality. In terms of software aspects, different reconstruction techniques and several noise-reduction algorithms like sinogram pre-processing exist, which have to be evaluated not only in terms of total noise reduction but also on their influence on the image quality. These tasks require methods to assess the image noise characteristics. Fig. illustrates the typical noise characteristics of a CT image. The noise texture shows an isotropic grain structure at the center. Towards the boundary it decreases and becomes increasingly oriented. This effect can be observed in both images of different water phantoms. There are models which analytically compute the noise variance propagation for the inverse Radon transform [], more precise noise propagation models also include local correlations introduced by various reconstruction steps []. Additionally, several approaches have been made, which model the noise propagation from the detector noise transfer function to the image noise power spectrum (NPS) [3], [4] but these models are limited by the fact, that the assumptions on the noise transfer functions (like detector- or focus-noise transfer function) are only valid for the image center. Objective quality measures of a CT image, however, rely on an assessment of the noise characteristics throughout the whole image.

9 Balda et al. Materials and Methods We develop a model to generate two-dimensional realizations of the image noise and NPS estimates for arbitrary image positions. It uses a projection noise model. For any pixel position and projection angle, an ensemble of projection noise values at the associated detector position is generated. This local detector noise signal then undergoes all steps of the reconstruction algorithm that influence the noise characteristics. Back-projection of the according noise signals for each angle yields a stationary, two-dimensional noise realization for this image location. This model gives an insight into the noise shaping properties of the filtered back-projection (FBP). Unlike analytical noise propagation methods, it can be easily adapted to any FBP-based reconstruction algorithm and various effects can be incorporated into the model without greatly increasing its overall complexity. The characteristics of this noise realization will differ throughout the image and resemble the CT image noise at a selected location p = (p x,p y ) T. The value of an image pixel depends on the attenuation of all beams passing through this pixel. These beams have associated projection angles φ [; π). The model creates a stationary local noise image patch i p (x) using the following steps for all projection angles φ ([5] for a concise description of the FBP):. Find measured attenuation value: The detector coordinate γ(p, φ) can be computed as follows: γ(p,φ) = ( ( atan r(p) cos(φ φ T (p)) d SOC +r(p) sin(φ φ T (p)) ) + β ) β a β () The complete fan angle is β, β is the fan angle between two neighboring detector channels, β a is the alignment angle due to the quarter offset (β a = for no quarter offset). The distance between X-ray source and origin is denoted d SOC, the polar coordinates of p are denoted r(p) for the distance to Fig.. Excerpts (5mm 5mm) of CT images of water phantoms: 3mm diameter (left), 4 mm diameter (right).

CT Noise Spectrum Analysis 93 the origin and φ T (p) for the polar angle. Using γ(p,φ), the according attenuation value µ(γ(p, φ), φ) can be interpolated from the measured projection data.. Initialize projection noise model: The projection noise model provides a noise distribution for the attuenuation values µ(γ(p, φ), φ). The noise in the attenuation values measured by the detector consists of electronics noise, quantization noise and quantum noise. The quantum noise depends on the number of incoming X-ray quanta. A detailed description on how quantum noise affects the detected attenuation values can be found in []. In practice, the noise effects can either be modeled or assessed by calibration measurements. Usually, the detector noise signal can be assumed to be uncorrelated and white. In practice minor correlations between the detector channels are introduced by optical cross-talk. 3. Generate virtual detector noise signals: A virtual, stationary detector noise signal is generated, that has constant noise distribution parameters throughout all channels. In case of rebinning to parallel beam geometry, an additional reading is generated with identical parameters. It represents the projection from the opposite direction. If the detector noise cross-talk behavior should be considered, these signals have to be convoluted with the impulse response of the detector. 4. Rebin virtual noise signal: The rebinning also influences the NPS. It consists of two steps. First the beams are reordered to parallel beam geometry with inhomogeneous ray distances. This requires a -D interpolation in the sinogram. In case of a quarter offset, the beams from opposite projection directions are then sorted into one single reading with the double amount of channels, finally the readings are resampled to homogeneous ray distances which requires a -D interpolation. In case of direct FBP, this step is omitted. 5. Filtered back-projection: The rebinned virtual noise signal is filtered and back-projected. The resulting noise pattern is stationary and resembles noise characteristics as this specific location in the image.. Algorithm The algorithm can be summarized as follows.. Set projection angle φ = π/+ φ r and φ = π/n r. Calculate γ(x,φ) according to () and interpolate reading d φ [c] at γ to get attenuation value µ(φ). 3. Initialize detector noise generator with µ(φ). 4. Generate two vectors of noise signal realizations d n,i [c] with c [;N d ] and i {,}. 5. Perform rebinning on d n, and d n, [c] to get ˆd n [c p ]. 6. Back-project d t [c p ] onto the noise patch i x. 7. if r < N r : Set r = r + and GOTO ;

94 Balda et al. The length of the detector noise signal N d has to be chosen according to the desired size and field of view (FOV) of the noise resulting image patch i x so that the image patch is fully covered by the back-projected values of d t [c], n r is the number of desired rotation angle samples. 3 Results and Discussion The algorithm was tested with several measured and simulated phantoms. Fig. (a) shows the reconstruction of a simulated phantom with marked voxel locations. This phantom is especially suited to demonstrate the properties of the proposed method, since the two strongly attenuating bone structures have a dominating influence on the noise structure of the whole image. Fig. 3 displays the respective stationary noise patches and the corresponding NPS estimates. Dominating noise components in directions of strong attenuation (between the two bones) can be observed. This leads to an increasingly directed noise grain structure. For the pixel close to the border, there is no such dominant influence on the noise direction, this leads to a more homogeneous noise structure. The overallnoiseisstrongerforpixelsclosetothecenterand/ornearstrongattenuators. For testing the noise model on measured data, an elliptic water phantom of 45mm width and 5mm height was used. Fig. (b) shows the measured noise standard deviation from a circular ROI of radius 5 mm shifted along the longer axis of the ellipse. The corresponding model estimates are generated with the introduced method, an analytic forward projector and a calibrated noise model. The measurement and the values computed with the model on simulated data show a good agreement with relative deviations below 5%. Noise standard deviation of attenuation values (mm ).4.3...9 x 3 Model Measurement 6 4 8 6 3 34 38 4 Pixel coordinate in horizontal direction Fig.. Left: Reconstruction of an elliptical water phantom with two femur bone structures. Noise patches are provided in Fig. 3 for marked voxel locations; Right: Comparison of noise standard deviation in from measured data and model.

CT Noise Spectrum Analysis 95 Fig.3. Top row: Noise patches for voxels marked in Fig. (a) (window center:, width: 4. HU); bottom row: normalized NPS estimates computed from noise patches above. 5 5 5 5 5 5 5 5 5 3 3 3 Vertical frequency (mm ) 35.5.4...4 5 5 5 3 35 Center voxel.5.5.5 Horizontal frequency (mm ) x 8.5.5.5 Vertical frequency (mm ) 35.5.4...4 5 5 5 3 35 Voxel at (55, 38).5.5.5 Horizontal frequency (mm ) 5.5 5 4.5 4 3.5 3.5.5 35 5 5 5 3 35 9 Voxel at (55, 34) x.5 6 Vertical frequency (mm ).4...4.5.4.5.5 Horizontal frequency (mm ) x 9..8.6.4..8.6 4 Conclusion We have introduced a method to generate stationary noise patches and Noise Power Spectrum estimates that resemble the noise characteristics of arbitrary voxel positions of CT images. The NPS estimates show the influence of object structures and reconstruction properties on the image noise structures. The model can be easily adapted to various FBP-based reconstruction techniques. This method can be used as a valuable tool to evaluate the influence of reconstruction parameters such as reconstruction kernels or noise adapted sinogramfilters on the quality of the reconstructed image. References. Buzug TM. Einführung in die Computertomographie: Mathematisch-physikalische Grundlagen der Bildrekonstruktion. Berlin: Spinger; 5.. Borsdorf A, Kappler S, Raupach R, et al. Analytic noise propagation for anisotriopic denoising of CT images. In: IEEE Nucl Sci Symp Conf Rec; 8. p. 5335 38. 3. Riederer SJ, Pelc NJ, Chesler DA. The noise power spectrum in computed X-ray tomography. Phys Med Biol. 978;3(3):446 54. 4. Kijewski MF, Judy PF. The noise power spectrum of CT images. Phys Med Biol. 987;3(5):565 75. 5. Kak AC, Slaney M. Principles of Computerized Tomographic Imaging. Soc Ind Appl Math Publishing;.