CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Similar documents
8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

CT Iterative Reconstruction Techniques

AIDR 3D Iterative Reconstruction:

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions

The Near Future in Cardiac CT Image Reconstruction

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Performance Evaluation of CT Systems

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Optimisation of Toshiba Aquilion ONE Volume Imaging

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Acknowledgments and financial disclosure

AAPM Standard of Practice: CT Protocol Review Physicist

Disclosures 7/21/2014. AAPM 2014 Challenges and opportunities in assessment of image quality

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Computer-Tomography II: Image reconstruction and applications

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

Micro-CT Methodology Hasan Alsaid, PhD

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Philips SPECT/CT Systems

CT vs. VolumeScope: image quality and dose comparison

A noise power spectrum study of a new model-based iterative reconstruction system: Veo 3.0

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

Agenda : Lung Density Breakout Session

Medical Image Processing: Image Reconstruction and 3D Renderings

DUE to beam polychromacity in CT and the energy dependence

Lucy Phantom MR Grid Evaluation

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Non-Stationary CT Image Noise Spectrum Analysis

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017

Q.Clear. Steve Ross, Ph.D.

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

Central Slice Theorem

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Spectral analysis of non-stationary CT noise

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Tomographic Reconstruction

ML reconstruction for CT

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Iterative Reconstructions: The Impact on Dose and Image Quality

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Digital Image Processing

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

Tomography at all Scales. Uccle, 7 April 2014

Motion Compensation from Short-Scan Data in Cardiac CT

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

IMPROVING SENSOR NOISE ANALYSIS FOR CT-SCANNER IDENTIFICATION

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

ADVANCED RECONSTRUCTION FOR ELECTRON MICROSCOPY

Reconstruction Methods for Coplanar Translational Laminography Applications

Equipment Specification

Refined Measurement of Digital Image Texture Loss Peter D. Burns Burns Digital Imaging Fairport, NY USA

Super-resolution Reconstruction of Fetal Brain MRI

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction

INDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Whole Body Submillimeter Scan

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

F3-A5: Model-Based Iterative Reconstruction for Advanced Baggage Screening

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role

Cardiac Dual Energy CT: Technique

Image Acquisition Systems

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Multigrid Reconstruction of Micro-CT Data

Limited view X-ray CT for dimensional analysis

Digital Image Processing COSC 6380/4393

Digital Volume Correlation for Materials Characterization

A new calibration-free beam hardening reduction method for industrial CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Introduction to Positron Emission Tomography

Transcription:

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology Cleveland Clinic Lerner College of Medicine/Case Western Reserve University Cleveland, OH

Objectives Understand why Noise Power Spectrum(NPS) is a more comprehensive descriptor of CT image noise properties Be familiar with methods of evaluating NPS Understand why the NPS of Filtered Backprojection (FBP) differs from Iterative Reconstruction (IR) Be familiar with IR s noise property and its implication on CT protocol optimization

Same Noise Magnitude but Different Texture 120 kv, 1000 eff mas, B70 Noise =16 120 kv, 10 eff mas, B10 Noise =16

Why Noise Power Spectrum (NPS)? Noise (standard deviation) only measures the magnitude of image noise properties. Noise power spectrum measures not only the magnitude but also the spatial correlation of noise properties ( texture ).

NPS for Images w/ Different Noise Texture B30 B70

What information can we get from NPS? Area under NPS curve is equal to the square of noise (magnitude) Mean and peak frequencies are related to the noise texture ( noise grain size ). Fine texture usually indicates NPS has higher mean and peak frequencies Mean freq: 3.2 lp/cm Peak freq: 2.5 lp/cm Mean freq: 5.4 lp/cm Peak freq: 6.3 lp/cm

Evaluating CT Noise Power Spectrum NPS is determined from the Fourier transform of the spatial autocorrelation function of a zero-mean noise image. For CT volumetric image data, NPS can be evaluated in 3D, 2D, or 1D (radially-averaged 2D) 3D 2D (xy-plane) 1D (radially-averaged 2D)

Equations for NPS Evaluation 3D NPS: NPS 3D f x, f y, f z = xδy z N x N y N z < DFT 3D I x, y, z I(x, y, z) 2 > 2D NPS f x f y -plane: 2D NPS f y f z -plane: NPS 2D f x, f y = xδy N x N y < DFT 2D I x, y I(x, y) 2 > NPS 2D f y, f z = yδz N y N z < DFT 2D I y, z I(y, z) 2 > I(x, y, z) is the CT volumetric image data, and I(x, y, z) is the averaged image to remove the DC components and/or de-trending is the ensemble mean 1D NPS is computed through radially-average of 2D NPS. Siewerdsen et al, A framework for noise-power spectrum analysis of multidimensional Images Medical Physics, Vol. 29, No. 11, November 2002

De-trend CT Images To de-trend and remove structured noise: By subtracting a polynomial fitted (2 nd order) image from the original. For repeated acquisitions, use the difference image from two adjacent scans. The NPS computed from the difference image should be divided by 2 To have the better de-trended image, make sure the tube starting angle is the same for the adjacent scans.

NPS Estimate Based on Single Acquisition ROIs distributed radialsymmetrically

NPS Evaluation via Repeat Acquisitions For repeat acquisitions, the ensemble mean of NPS can be computed from images from repeat scans. The ensemble mean NPS tends to be less noisy for better estimate of mean and peak frequencies.

Ensemble Mean NPS of Repeated Scans NPS via ensemble mean of repeat acquisitions has smoother appearance than those from the single acquisition. Ensemble mean via repeat scans NPS via single scan Yu et al, Measurement and analysis of 3D NPS of an iterative reconstruction method in CT RSNA 2012

Location-dependency of noise pattern Yu et al, Measurement and analysis of 3D NPS of an iterative reconstruction method in CT RSNA 2012

Inter-vendor Comparison of Recon Kernels For a CT site with more than one CT vendor, it is difficult to find matching recon kernels: GE: soft, standard, detail, chest, lung, bone, bone+, edge Siemens: B10f, B20f, B22f, B23f, B26f, B30f, B31f, B35f B80f. Solomon et al. used similarity of the NPS to compare CT recon kernels between GE and Siemens Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Phantom and ROI Placement Using module 3 (uniform section) of the ACR CT phantom Four ROIs in each slice and multiple slices in each acquisition Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

NPS Comparison Steps Normalizing the 1D NPS by its integrated area -> nnps Filtering nnps with a human visual response function Root Mean Square Difference (RMSD) and Peak Frequency Difference (PFD) of NPSs as the metrics of similarity Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Human Visual Response Function Human visual response function is used to account for the variable perception of noise by human at different spatial frequencies nnps f (r) is normalized NPS by its area, V( ) is the human visual response function Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Normalized NPS Normalized NPS for GE recon kernels Normalized NPS for Siemens recon kernels Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Relative Noise and Peak Frequencies Relative noise and peak frequencies for GE recon kernels Relative noise and peak frequencies for Siemens recon kernels Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Closest matched Siemens kernel for a GE kernel Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

Noise Texture Comparison for Matching Kernels GE Siemens GE Siemens Soft B35f Chest B41f Standard B43f Bone+ B75f Detail B46f Lung B80f Solomon et al, Quantitative Comparison of Noise Texture Across CT Scanners Medical Physics, Vol. 39, No. 10, October 2012

CT Iterative Reconstruction Is a feature that uses the information acquired during the scan and repeated reconstruction steps to produce an image with less noise or better image quality (e.g., higher spatial resolution or decreased artifacts) than is achievable using standard reconstruction techniques (FBP)

CT Iterative Reconstruction Iterative Reconstruction may be completed using data in Image Space, Projection (raw CT data) or a Model Based Approach. Projection or Model based approach is much better implementation but increases recon time.

Model-based Iterative Reconstruction Current estimate of object System Optics X-ray source model Geometry model Synthesized projection System statistics Statistics model Updated object Detector model Measured projection Object model Repeat the loop until measured and synthesized projections converge

Standard FBP Images Axial Image CTDIvol =3.7mGy Coronal MPR Images courtesy of GE Healthcare

Model Based Iterative Recon (Veo) Images Axial Image CTDIvol =3.7mGy Coronal MPR Images courtesy of GE Healthcare

Nonlinear Statistics Model in IR In the statistics model of IR algorithm, there is an adaptive regularization term which depends on local contrast gradient, i.e., it can reduce the noise in homogeneous regions and preserve details at edges. The nonlinear regularization term may be more aggressive for noisier data.

Noise Comparison between FBP and IR Chen et al, Assessing CT Noise and Resolution for Nonlinear Reconstruction Medical Physics, Vol. 41, No. 7, July 2014

FBP: NPS vs. Dose The shape of normalized NPS is similar at different dose levels for FBP images NPS at different dose levels NPS normalized Li et al, Statistical Model Based Iterative Reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance Medical Physics, Vol. 41, No. 4, April 2014

IR: NPS vs. Dose The shape of normalized NPS is not similar at different dose levels for IR images NPS at different dose levels NPS normalized Li et al, Statistical Model Based Iterative Reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance Medical Physics, Vol. 41, No. 4, April 2014

Peak Frequency for FBP and IR (Veo) Li et al, Statistical Model Based Iterative Reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance Medical Physics, Vol. 41, No. 4, April 2014

Mean Frequency for FBP and IR (Veo) Li et al, Statistical Model Based Iterative Reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance Medical Physics, Vol. 41, No. 4, April 2014

Noise vs. Dose for FBP and IR (Veo) Li et al, Statistical Model Based Iterative Reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance Medical Physics, Vol. 41, No. 4, April 2014

Implications for Protocol Optimization in IR For IR protocols, the conventional relationship between noise and dose (σ = α (mas) β, = 0.5) may not be valid anymore. Re-assessment of this important relationship under different dose levels is needed. In Veo (GE), is between 0.21 and 0.25, which means the noise standard deviation is much less sensitive to the dose change. A 50% dose reduction will result in only 15% to 19% increase in noise. In FBP, that amount of dose reduction will lead to 40% noise increase.

Implications for Protocol Optimization in IR Noise texture may impact the detectability of low contrast lesion, especially using high strength IR. In general, IR will not degrade the high contrast resolution due to the fact that most IR uses adaptive regularization.

Conclusion NPS is a comprehensive descriptor for CT image noise properties, specifically related to noise texture Nonlinear iterative reconstruction tends to shift peak frequency of NPS curve towards lower value, but it doesn t necessarily mean degradation of high contrast resolution Noise vs. dose relationship for FBP may not be valid for IR due to its nonlinear regularization used in the statistical model Optimization of CT IR-based protocols should consider the implication of its nonlinearity on NPS