CT vs. VolumeScope: image quality and dose comparison

Similar documents
Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Computer-Tomography II: Image reconstruction and applications

Digital Image Processing

Ch. 4 Physical Principles of CT

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Effects of the difference in tube voltage of the CT scanner on. dose calculation

DUE to beam polychromacity in CT and the energy dependence

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Enhanced material contrast by dual-energy microct imaging

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

RADIOLOGY AND DIAGNOSTIC IMAGING

Basics of treatment planning II

Image Acquisition Systems

UNCOMPROMISING QUALITY

Spectral analysis of non-stationary CT noise

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Optimisation of Toshiba Aquilion ONE Volume Imaging

BME I5000: Biomedical Imaging

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Tomographic Reconstruction

Scatter Correction Methods in Dimensional CT

A closer look at CT scanning

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

DUAL-ENERGY CT IN PROTON THERAPY

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

Quantitative imaging for clinical dosimetry

Proton dose calculation algorithms and configuration data

Michael Speiser, Ph.D.

Micro-CT Methodology Hasan Alsaid, PhD

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations

GPU implementation for rapid iterative image reconstruction algorithm

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Effect of CT number stopping power conversion uncertainties on dose distributions

Introduction to Biomedical Imaging

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

DUAL energy X-ray radiography [1] can be used to separate

Loma Linda University Medical Center Dept. of Radiation Medicine

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

AIDR 3D Iterative Reconstruction:

ICRP Symposium on Radiological Protection Dosimetry

Non-Stationary CT Image Noise Spectrum Analysis

ML reconstruction for CT

INTERNATIONAL STANDARD

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Empirical dual energy calibration EDEC for cone-beam computed tomography

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Physical bases of X-ray diagnostics

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Computer-Tomography I: Principles, History, Technology

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Some reference material

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

Quantitative image-based spectral reconstruction for computed tomography

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Evaluation of the Elekta Synergy concept for patient positioning in image guided radiotherapy

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Quality control phantoms and protocol for a tomography system

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Geant4 in Brachytherapy

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016

Geant4-DICOM Interface-based Monte Carlo Simulation to Assess Dose Distributions inside the Human Body during X-Ray Irradiation

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0

Cardiac Dual Energy CT: Technique

Agenda : Lung Density Breakout Session

Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005

Financial disclosure. Onboard imaging modality for IGRT

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

Transcription:

CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow, 117997, Russia b Institute for Roentgen Optics, PMZR, 10, 1 st Volokolamsky Pr., Moscow, 123060, Russia ABSTRACT A number of examinations were performed on the VolumeScope Compton scatter tomographic device and a CT scanner (Siemens Somatom Plus 4) using an anthropomorphic tissue-equivalent heterogeneous phantom simulating a human body. Three main parts of the phantom were under examination: head, thorax and pelvis. An absorbed dose in the phantom was measured by LiF thermoluminescent detectors. The CT scanner demonstrated better space resolution in cross-plane and lower statistical noise. On the other hand, absorbed dose due to CT scanning is about 20 times higher at the same points of measurement. Examples of reconstructed images are presented. Keywords: VolumeScope, absorbed dose, digital filter, image quality, anthropomorphic phantom, 3D visualization. 1. INTRODUCTION The VolumeScope (Fig. 1) is an X-ray scanner based on Compton scatter detection from the point of measurement 1. After some mathematical reconstruction of the measured data 2, the device allows to determine an electron density distribution of the object under examination and represent it as a 3D grid of volume elements, voxels. The resulting distributions can be displayed as gray-scale sections of the examined object or as 3D surfaces of internal structures for further analysis. Figure 1. General view of the VolumeScope. To evaluate the image quality, a number of measurements were performed on the VolumeScope and on the CT scanner Siemens Somatom Plus 4 installed in the Russian Oncological Center. As the VolumeScope is a laboratory prototype * E-mail: vnvasil@orc.ru ** E-mail: gam@iroptic.ru *** E-mail: kst@iroptic.ru Proc. of SPIE Vol. 5943 248

device, the purpose was to recognize its most significant problems and limitations in comparison with a commercial medical instrument for further improvement. Another task was to determine an absorbed dose inside the patient under examination as one of important characteristics of any X-ray diagnostic equipment. 2. METHODS The measurements were performed with the ATOM anthropomorphic phantom 3 (CIRS Inc., USA) simulating a human body. The phantom is made of tissue-equivalent materials and includes skeleton, lungs and brain models. All photon interaction cross-sections of the tissue substitutes are close to the cross sections of corresponding biological tissues within 1% down to the photon energy 30 kev. The phantom is dissected for 2.5 cm slices and has a grid of 5 mm holes inside to place thermoluminescent detectors for absorbed dose measurements. When not used, the holes are filled with cylindrical tissue-equivalent inserts. Three typical regions of the phantom were investigated: head, thorax and pelvis. Properties of the phantom materials are presented in Table 1. Physical properties of phantom materials Table 1 Tissue substitutes Physical density, Electron density, g/cm 3 10 23 per cm 3 Soft tissue 1.05 3.434 Spinal Cord 1.07 3.448 Spinal Disks 1.15 3.694 Lung 0.21 0.681 Average Bone 1.60 5.03 Brain 1.07 3.47 Before examining the anthropomorphic phantom, a preliminary measurement with a homogeneous water sample was performed for the estimation of a statistical noise level. A set of Plastic Water 4 plates and a liquid water sample were used with the CT scanner and the VolumeScope and the noise distribution parameters were determined by accompanying software, the DICOM viewer efilmlite 1.5.2 and the VolumeScope viewer respectively. The VolumeScope X-ray tube operated with the voltage 150 kv and anode current 4.5 ma. A primary X-ray beam of 2 mm in diameter was filtered by 1 mm Al and 0.5 mm Cu. The scanning speed was 90 mm/s and the voxel size was 2 mm along each coordinate axis. As scanning volume of the VolumeScope was limited by 200 x 200 x 70 mm due to the collimator design, front and back sides of the phantom were scanned separately. The CT scanner tube voltage was 120 kv, the anode current and voxel size are shown in Table 2. The CT scanner operation parameters Table 2 Phantom Anode current, ma Voxel size, mm PW Cube 150 0.967 x 0.967 x 5 Head 185 0.455 x 0.455 x 8 Thorax 110 0.807 x 0.807 x 8 Pelvis 170 0.807 x 0.807 x 8 3. RESULTS A standard deviation of the statistical noise was calculated for numerous regions of interest (ROIs) in the Plastic Water sample by the DICOM viewer efilmlite. Typical value was about 25 Hounsfield units i.e. about 2.5% of full signal range. The anthropomorphic phantom was examined with greater voxel volume (pelvis, thorax) or higher anode current (head) and demonstrated lesser noise standard deviation obtained in homogeneous ROIs typically 1-1.5%. The statistical noise standard deviation calculated by the VolumeScope viewer for liquid water sample was equal to 3.6%. To reduce the noise, we used a 2D low-pass digital filter with cutoff frequency 0.3 cycles/mm. This filter effectively suppressed the noise but saved high frequency details of the image. Using the filter, the standard deviation of Proc. of SPIE Vol. 5943 249

the noise in the water sample decreased to 1.6% and was approximately equal to the CT values. Currently, this filter is a standard tool of the data reconstruction VolumeScope software. Nevertheless, all 2D images below are presented without any postprocessing. Typical sections of the phantom head are shown in Fig. 2. The images demonstrate all internal bone structures, skull, spine and cartilage. The measured relative electron density of bones is in reasonable agreement with the data tabulated by the phantom manufacturer. A comparison of cross-plane sections of the phantom head obtained by the VolumeScope and the CT scanner is presented in Fig 3. The VolumeScope images demonstrated a significant blurring of boundaries and higher statistical noise in comparison with the CT data. This is resulted from better spatial resolution of the CT scanner in cross plane. On the other hand, some blurring of skull surface is visible on the CT image (Fig. 3f) due to significant voxel size and density averaging in the longitudial direction. The VolumeScope results demonstrate some electron density underestimation under the bone structure concerned with additional photoabsorption in bone due to its higher atomic number. a) b) c) d) Figure 2. Typical sections of the phantom head and neck (VolumeScope data): a frontal face section, b frontal section along the spine, c cross-plane neck section, d frontal section of the skull top. Proc. of SPIE Vol. 5943 250

a) b) c) d) e) f) Figure 3. Comparison of the VolumeScope (left) and the CT scanner (right) data for the phantom head cross-plane sections The VolumeScope program allows to reconstruct surfaces of internal structures inside the object of examination. External and skull surfaces of the phantom face region are presented as 3D objects in Fig. 4 on the base of VolumeScope data. As the figure demonstrates, most significant parts of the skeleton are successfully reconstructed. The objects in front of the eye-sockets and in the middle of forehead are bone-equivalent cylindrical inserts of 5 mm in diameter and 25 mm in length used for dosimetry. Proc. of SPIE Vol. 5943 251

Figure 4. External and skull surfaces of the phantom face reconstructed on the base of VolumeScope data. Examples of front and back parts of the thorax are presented in Fig. 5. Lungs, sternum, spine, spinal cartilage, ribs and clavicles are visible. The sternum and spine can be separated from the surrounding tissues and reconstructed as 3D objects (Fig. 6). a) b) c) d) Figure 5. The VolumeScope (left) and the CT (right) images of front (up) and back (bottom) parts of the thorax. Proc. of SPIE Vol. 5943 252

Figure 6. Sternum (with clavicles) and spine reconstructed as 3D surfaces from the VolumeScope data. 4. DOSIMETRY An absorbed dose inside the patient under examination was measured by thermoluminescent detectors LiF TLD-100 of size 3.2 x 3.2 x 0.9 mm and a TLD reader model 3500 (Harshaw, USA). The detectors were calibrated in a Secondary Standard Dosimetry Laboratory, total uncertainty of the detector response was 6.6% at the 95% confidence level. 29, 49 and 39 detectors were placed inside the head, thorax and pelvis parts of the phantom during the CT examination. As scanned volume on the VolumeScope was limited by 200 x 200 x 70 mm, the number of detectors in this case was 19, 30 and 28 in the head, thorax and pelvis respectively. TLD positions inside the phantom as well as details of the detector response correction are presented in Ref. 5. Minimum, maximum and mean absorbed doses are presented in Table 3. Absorbed dose per examination in the anthropomorphic phantom Table 3 CT scanner VolumeScope Min Max Mean Min Max Mean Head 22.7 50.7 37.3 0.82 3.05 1.43 Thorax 8.0 23.1 15.1 0.60 1.20 0.90 Pelvis 11.0 26.2 16.7 0.62 1.08 0.81 As the table demonstrates, the patient dose on the VolumeScope is about 20 times less than on the CT scanner. The CT dose data are in good agreement with the results by Giacco et al. 6 measured inside pediatric phantoms with similar CT scanner, Siemens Somatom Plus 4, tube voltage 120 kv and current 170 ma and obtained the CTDI value 30.9 mgy. 5. CONCLUSIONS The performed measurements demonstrated that a 3D electron density distribution of biological objects can be obtained by the VolumeScope within the volume 200 x 200 x 70 mm with voxel size 2 mm. The statistical noise level is higher than in the CT data, nevertheless, it can be reduced to the CT level by simple 2D nonadaptive digital filter. There are good possibilities for further noise reduction by using more advanced adaptive 3D filters designed on the base signal and noise power spectrum densities analysis. Equal voxel size along each coordinate axis allows a convenient 3D surface generation of internal structures for further visual analysis. Proc. of SPIE Vol. 5943 253

A significant advantage of the VolumeScope is low absorbed dose in the patient, 1-3 mgy. This is approximately 20 times less than on the CT scanner. Limited scanning volume equal to 200 x 200 x 70 mm and long scanning time are most significant disadvantages of the VolumeScope and must be improved. Some boundaries blurring was observed on all VolumeScope images and resulted from limited spatial resolution 7. This problem can be solved by improvement of the collimator design and deconvolution of the measured data with the point spread function as a kernel. Both approaches are in work. Some reconstruction artifacts were found in the VolumeScope data that appeared as density underestimation in deep slices and are most pronounced under bone structures. These effects are resulted from additional photoabsorption in bone due to its higher atomic number and can be reduced by the reconstruction algorithm improvement. REFERENCES 1. M.A. Kumakhov, A.F. Gamaliy, V.N. Vasiliev, M.Yu. Zaytsev, K.V. Zaytseva, A.A.Markelov, and Yu.V. Ozerov, Scattered X-rays in medical diagnostics, in 2-nd International conference on X-ray and neutron capillary optics. 22-26 September 2004, Zvenigorod, Proc. SPIE 5943, pp. 210-218, 2005. 2. V.N. Vasiliev and K.V.Zaytseva, An algorithm and program for data processing from Compton scatter imaging device, in 2-nd International conference on X-ray and neutron capillary optics. 22-26 September 2004, Zvenigorod, Proc. SPIE 5943, pp. 232-241, 2005. 3. Atom Adult Male Phantom, Model 701-D, Handling Instructions, CIRS Inc. 4. Plastic Water. CIRS Catalog, p.83-84, 2002. 5. V.N. Vasiliev and M.Yu. Zaytsev, An absorbed dose due to examination on the VolumeScope X-ray device, in 2- nd International conference on X-ray and neutron capillary optics. 22-26 September 2004, Zvenigorod, Proc. 5943, pp. 242-247, 2005. 6. G. Giacco, V. Cannata, C. Furetta, F.Santopietro, and G. Fariello, On the use of pediatric phantoms in the dose evaliation during computed tomography (CT) thorax examination, Med. Phys. 28, pp. 199-204, 2001. 7. V.N. Vasiliev and M.Yu. Zaytsev, Frequency characteristics of images measured on the VolumeScope scanning device, in 2-nd International conference on X-ray and neutron capillary optics. 22-26 September 2004, Zvenigorod, Proc. SPIE 5943, pp. 228-231, 2005. Proc. of SPIE Vol. 5943 254