User Definable Settings

Similar documents
Next Generation Hematology Analyzer. BeneSphera H33s. Minimum Size, Maximum Capability

BC Auto Hematology Analyzer. Pleasant Experience in Hematology Testing

Medonic M-series M32 Innovation built on total quality

MicrosEmi CRP External Output Manual (English)

INTERFACE SPECIFICATION

ADVIA 360, ADVIA 560, and ADVIA 560 AL Systems

Hematology Program (BC90A/BC90B/BC90C/BC90D/CS90A/CS90B/CS90C/CS90D) Cycle 11: March 2016 February Sample No: 1 Sample Date: 14 Apr 16

Hematology Analyzer. Serial Communication Protocol. 1.7 release

Abacus plus. Hematology Analyzer. User s Manual release

Date Name Signature Issued 06-MAY-2011 Csaba Magyar Checked and approved 17-MAY-2011 Gábor Farkas

World-class veterinary hematology system

OM-A Revision history Rev. Date Edited by Sections affected Approved by

UM-AJV Abacus Junior Vet User s Manual

Hematology Analyzer. User s Manual. Revision 2.17

Boule Page 1/36 Service manual BM800 Doc No Section 9 Version 02

ELECTIVES MODULES. Implement. Electives. Setup. Troubleshooting. Basics

scil Technical Support

Swelab Alfa. A new definition of cell counter precision.

Lab ID 406. Your Parameters. Institution: Analyte. WBC Siemens Advia 120 Siemens 10 9 /L. RBC Siemens Advia 120 Siemens /L

Contents. Preface Safety Instruction Specifications Measuring Principles Parameter Flags... 33

Boule Medical AB CA620-CellGuard Page Usermanual V3.7

RT-7600 Auto Hematology Analyzer User s Manual

Fast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education

Quick Instruction Guide for Operating the scil Vet abc Plus

Creating and Using an Excel Table

COULTER STKS Analyzer with Reticulocyte Analysis. Operator s Guide. PN B (March 1995) COULTER CORPORATION Miami, Florida 33196

Easy Match QC Lesson 9. Setting Up a Print Job

Hemolyzer. 5 NG (closed mode) User Manual. Version 1.35

BA-88A Semi-auto Chemistry Analyzer

Abacus Arcus. Service Manual. Rev. 2.4

Excel Contents. MS Excel /3/2011 JSD#2-Neptune Page 1

UniCel DxH 600 Coulter Cellular Analysis System Training Modules

Hemolyzer 3 NG. User Manual. Version 1.35

Microsoft Excel 2010 Part 2: Intermediate Excel

Step by step instructions for layout for Theology papers Part 1 Setting up margins

Excel Level Three. You can also go the Format, Column, Width menu to enter the new width of the column.

Add Bullets and Numbers

Hematology Analyzer. Service Manual. 3.0 release

DCN Next Generation Simultaneous Interpretation. en Software User Manual LBB 4172/00

HA-86 (89) Auto Hematology Analyzer User s Manual

Administrator s Guide ALYX Component Collection System Codes/ 4R5707 4R5725

Microsoft Excel 2010 Basic

Labels and Envelopes in Word 2013

PART 7. Formatting Pages

Unicel DxH 600 System

MAIL MERGE USING MS WORD 97

Usermanual Release

TopPoint User's Guide

MAILMERGE WORD MESSAGES

Document Formatting with Word

Cell to Cell mouse arrow Type Tab Enter Scroll Bars Page Up Page Down Crtl + Home Crtl + End Value Label Formula Note:

DCN Simultaneous Interpretation. Software User Manual en LBB 3572

Custom Label and Paper Sizes Guide

MultiSite Suite: General Ledger

A warning provides additional information which, if ignored, may result in a risk of personal injury.

MB 780 / MB790. Service Guide OkiData Americas, Inc.

EQUIPMENT MODULAR PATIENT MONITOR VIDEO COLPOSCOPE OBSTERICAL/GYNACOLOGICAL MONITORS ULTRASOUND MACHINES ULTRASONIC POCKET DOPPLER

SymBiot I Sample Workstation

How to Open Excel. Introduction to Excel TIP: Right click Excel on list and select PIN to Start Menu. When you open Excel, a new worksheet opens

Address Bar. Application. The space provided on a web browser that shows the addresses of websites.

INDEX TECHNICAL SERVICE BULLETIN

C813n C823n/C823dn C833n/C833dn C843n/C843dn ES8433 ES8443

Printer Settings. Accessing Power Saver. Selecting Output Options. Phaser 5500 Laser Printer

Getting Started with. Office 2008

Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003.

Unscrew and remove the cap from the waste tank

Intermediate Word for Windows

Microsoft Excel 2007 Creating a XY Scatter Chart

Application of Skills: Microsoft Excel 2013 Tutorial

Open Microsoft Word: click the Start button, click Programs> Microsoft Office> Microsoft Office Word 2007.

Navigate to Success: A Guide to Microsoft Word 2016 For History Majors

EO+ User s Manual. Swelab AC920 EO+ Swelab AC920 EO+ with cap piercer Swelab AC970 EO+

How to Format Modern Language Association (MLA) Style Papers

Instructions for Formatting MLA Style Papers in Microsoft Word 2010

A warning provides additional information which, if ignored, may result in a risk of personal injury.

CELL-DYN 1800 Field Service Training Workbook

Product Parts Front view... 2 Rear view... 2 Inside the product... 3 Scanner parts... 3 Control panel... 4

Setting Up a Paper in APA Style Using Microsoft Word 2008 for MACs

Excel Basics. TJ McKeon

Links to Activities ACTIVITY 1.1. Links to Activities Links to Activities

What we will learn in Introduction to Excel. How to Open Excel. Introduction to Excel 2010 Lodi Memorial Library NJ Developed by Barb Hauck-Mah

Creating a Template in WordPerfect

SwiftView Quick-Start Guide. Revised for SwiftView

Formatting a Report with Word 2010

Microsoft Excel Lab: Data Analysis

IMPORT. DisplayWord Your Forms and Your Media User Guide. Create forms for probate using DisplayWord. Software

Microsoft Word Advanced Skills

Intelligent Automation

Setting Up a Paper in APA Style Using Microsoft Word 2007

DOCUMENTATION CHANGE NOTICE

Communication Skills Center

Ms excel. The Microsoft Office Button. The Quick Access Toolbar

The Newsletter will contain a Title for the newsletter, a regular border, columns, Page numbers, Header and Footer and two images.

Introduction to Microsoft Word 2010

The HOME Tab: Cut Copy Vertical Alignments

Automatic analyzer Reagent system. Random access STAT (immediate access and execution) Barcode reader. Built-in for reagents, optional for samples ISE

Understanding Word Processing

CELL-DYN Sapphire TM Interface Set-up Procedure for Accelerator DM Application Software Version 1.0

350 SMART MANOMETER OPERATING INSTRUCTIONS

Chapter 15 Printing Reports

Transcription:

User Definable Settings Section 1: User Definable Print Format Settings...1 Section 2: Parameter Unit Selection...7 Section 3: Setup Menu 2...8 Section 4: Setup Menu 3...11 Section 1: User Definable Print Format Settings Description This section describes how to set the user definable print format settings for the Medonic M-Series and Swelab Alfa instruments. Printer Setup Menus Menus To change printer and printout formats press [ADVANCED] from the Main menu, then [SETUP], and then [PRINT SETUP]. There are three Print Setup menus to choose from. Figure 1.1 Figure 1.2 Figure 1.3 Ticket Format This function allows the user to change the column layout of the printout. To change the format, choose the corresponding number from the list below and press [OK] to select. Figure 1.4 1

Section 1: User Definable Print Format Settings (continued) Manual Print Mode This function allows the user to printout a result using the [PRINT] button. To turn this function off select 0, if a printout is preferred select either 1 or 2 to choose if printout will contain a histogram along with the numerical values. See Figure 1.5. Figure 1.5 Figure 1.6 Auto Print Mode This function allows the user to select whether a printout is automatically printed after each sample, background, or control analysis. If no automatic printout is preferred select 0, if a printout is preferred select either 1 or 2 to choose if printout will contain a histogram along with the numerical values. See Figure 1.6. One Ticket Per Page This function allows the user to print more than one analysis per page. Default is set to one analysis per page (activated). Choose [INACTIVATE] if user wants to print more than one analysis per page. The allowable number of analyses per page is dependent on print format setup. See example below to activated or inactivate function. Press this button to inactivate. Press this button to activate. Figure 1.7 2

Section 1: User Definable Print Format Settings (continued) Printer Type To connect the instrument to the printer follow instructions in User s Manual. Once connected, instruct the printer to interface with the instrument by selecting the corresponding number from the list below (Figure 1.8) and press [OK] to select. Figure 1.8 Figure 1.9 Paper Type This function allows the user to choose the type (size) of paper used for the printout. Paper types are printer dependent. Select from the following list (See Figure 1.9): 1. Seiko DPU 411/2 and 414 (not applicable, uses paper rolls) 2. IBM proprinter / Epson compatible 1 = A4 2 = 12 3 = 11 (Letter) 4 = 14 (Legal) 5 = 10.5 (Executive) 6 = Non-specified size 3. HP PCL 3 and 5 protocol compatible 1 = A4 3 = 11 (Letter) 4 = 14 (Legal) 5 = 10.5 (Executive) 6 = Non-specified size Show Flag Texts This function allows the user to choose whether flag text is displayed on printout. Default is set to display flag text (activated). Choose [INACTIVATE] if user does not want to display flag text on printout. See Figures 1.6 above for an example. Note If an option is not available, the number will not be accepted when operator presses [OK]. 3

Section 1: User Definable Print Format Settings (continued) Customized Print Formats The Customized Print Format allows the user to copy a ticket format, enter their own Header and Footer text, margins, paper length, histogram format, and parameter order for the results on the printout. Step 1 Action Select [ADVANCED] from the Main menu, then [SETUP], and then [PRINT SETUP]. Press [MORE] twice to enter Print Setup C and select one of the Customize ticket buttons 51-54. 2 3 Figure 1.10 Figure 1.11 Choose a ticket format, by copying a pre-established ticket format and press [OK] to accept and then [MORE] to continue to next step. To change the format of the histograms choose [HISTOGRAM FORMAT]. Select one of the formats from the list below, then press [OK] to save: 0 = Default for the ticket format and printer selected 1 = 3 double width histograms placed vertically (same as Printer Type 1 Format) 2 = 3 single width histograms placed horizontally (same as Printer Type 2 Format 4) 3 = PLT and RBC single width histograms side by side and WBC double width histogram in next row 4 = 3 double width histograms placed horizontally (very wide, set left margin to 0) 5 = WBC double width histogram and PLT and RBC single width histograms side by side in next row Figure 1.12 Figure 1.13 4

Section 1: User Definable Print Format Settings (continued) Step Note 4 Action By selecting [PRINT] the user can printout and view the last sample with the new format changes. This selection is only used in case the connected printer is not conforming to the current standard, and allows the user to override the current selections regarding form feed and print form lengths as shown in Paper type section. To change the format of the paper length select [PAPER LENGTH], then select one of the formats from the list below. Press [OK] to save: 0 = Default for the ticket format and printer selected 15 99 = Choose the number of rows to adjust the paper length 5 Figure 1.14 Figure 1.15 Select [HEADER] and then [EDIT]. The Header text places text above the parameters and ID on the printout. Begin by typing in the desired header text. This text will be displayed between the two arrows on the top of the page. To add another line of text select [NEW LINE] and an enter symbol will appear to show where the next line begins. (Suggested to put one NEW LINE at end of the Header text entry) To delete a word or letter, use the arrow buttons to place the cursor before the word or letter to be deleted and then select the [CE] button until deleted. To enter a space select [SPACE BAR]. 6 Figure 1.16 Figure 1.17 When the Header is complete select [OK] and the Header will appear on the display page to view. If no further edits are necessary select [EXIT]. 5

Section 1: User Definable Print Format Settings (continued) Step 7 8 Action The Footer text places text below the parameters and histograms on the printout. Follow Steps 5-6 for customizing the Header, except choose [FOOTER] at Step 5. Left Margins allows the user to edit the left margins for the header, parameters, histograms, information text, and footer for a selected ticket format. Select [LEFT MARGINS]. Choose which margin to change, and select the margin by the number of spaces from the left side of the printout page. (Example: Choosing 12 for the Header will move the header text 12 spaces from the left side of the page.) Select [OK] to save margin spacing. Choose next margin to be changed, and when complete select [EXIT] to save. 9 10 Figure 1.16 Figure 1.17 Parameter Order allows the user to edit how the order of parameters appears on the printout. The number of Columns and Rows is dependent of the copy of the ticket format selected. By selecting the [ Position] and [Position ] buttons the user scrolls through the Rows and Columns. If there is more than one column, the second column can be viewed by selecting [Position ] one more time after the last row of the previous column is displayed. By scrolling through the positions, the current parameter that is in that position is viewed under the Parameter buttons. To change a parameter in a certain position, scroll to the position first and then use the [ Parameter] and [Parameter ] buttons to select the new parameter for that position. The user can change one or multiple parameters at a time. If many parameters will be changed, continue with position and parameter selections until complete, and then press [EXIT]. Parameter order and position can be viewed the same as the Header and Footer by selecting [PRINT]. Note: Skip means that nothing is printed at the selected row and Blank means that an empty field (line) will be printed in that position. Note: If format 4 is chosen, WBC parameters, in second column, are fixed and can not be edited. To choose the new Customized Print Formats as printout default select [MORE] to enter Print Setup A. Select [TICKET FORMAT]. Enter in customized ticket number (example: 51) and press [OK] to save. 6

Section 2: Parameter Unit Selection Description This section describes how to set the parameter units for the Medonic M-Series and Swelab Alfa instruments. The parameter units are in settable groups. Parameter Unit Selection To change the parameter units press [ADVANCED] from the Main menu, then [SETUP], then [SETUP 2], then [REGIONAL SETUP], and then [MORE] until Regional Setup C menu is displayed. Select Parameter Units Group A, B, C, or D, then enter selection number to choose that parameter unit group, and then press [OK] to select new group. Regional Setup C Figure 2.1 Figure 2.2 Parameter Units Group A: Selection 1: Selection 2: RBC 10^12/L RBC 10^6/uL PLT 10^9/L PLT 10^3/uL WBC 10^9/L WBC 10^3/uL LYM,MID,GRAN 10^9/L LYM,MID,GRAN 10^3/uL Parameter Units Group B: Selection 3: RBC 10^6/mm^3 PLT 10^3/mm^3 WBC 10^3/mm^3 LYM,MID,GRAN 10^3/mm^3 Selection 1: MCV fl MPV fl RDW(a) fl PDW fl absolute Parameter Units Group C: Selection 2: MCV um^3 MPV um^3 RDW(a) um^3 PDW um^3 absolute Selection 1: Selection 2: HGB g/dl HGB g/l MCH pg MCH pg MCHC g/dl MCHC g/l Parameter Units Group D: Selection 3 HGB mmol/l MCH fmol MCHC mmol/l Selection 1: HCT % Selection 2: HCT L/L 7

Section 3: Setup Menu 2 Description This section describes User Definable Settings in Setup Menu 2 that were not discussed in the User s Manual. Memory Setup Storing Raw Data Activate Flash Memory Figure 3.1 Figure 3.2 The default for Storing Raw Data is Inactive, as this function takes up a larger amount of memory when activated and will decrease memory space. The main purpose for activating this function is for in-house studies and service use. (A special program is needed when transferring this data to a computer. Contact local service representative for more details.) Choose [ACTIVATE] if function is needed. The default for Flash Memory Mode is Inactive. Contact local service representative for more details on activating this mode. Blood Detector Setup Figure 3.3 These functions allow the user to enable and disable the blood detector. Setting this function to 0 enables the blood detector function. When enabled, aspiration stops when blood is detected by blood detector sensor. A fixed aspiration time, can also be selected (from 0.1 to 10 seconds) if needed. (Setting blood detector to any value except 0 disables the blood detector.) 8

Section 3: Setup Menu 2 (continued) Set Standby Timeouts Figure 3.4 These functions allow the standby timeouts be changed to user preference. 1. Minutes before Displaysaver can be set from 2 to 120 minutes. 2. Minutes before Standby can be set from 10 to 240 minutes. 3. Background Count after Standby is a user friendly function that when activated will perform an automatic background count when instrument comes out of Standby. To change default, select [INACTIVATE]. 4. Enable Startup Sequence walks the operator through the beginning of the day startup routine for the analyzer. To change default, select [ACTIVATE]. 5. Maintain Diluent Levels maintains a consistent liquid level in diluent and lyse pipettes during standby. Regional Setup A Figure 3.5 Figure 3.6 9

Section 3: Setup Menu 2 (continued) Keyboard Layout This function allows user to select external keyboard type for interface to instrument. See user manual for more details on setup. 0 = No external keyboard 1 = English external keyboard 2 = Swedish external keyboard International Parameter Names Figure 3.7 Figure 3.8 Many countries have different names and acronyms for hematology parameters. This functional allows the user to set language dependent parameter names. (Example: France vs Switzerland, both use the French language but they use different parameter names.) 1 = International 2 = Swedish 3 = French 10

Section 4: Setup Menu 3 Description This section describes User Definable Settings in Setup Menu 3 that were not discussed in the User s Manual. Setup Menu 3 Color Setup Figure 4.1 Figure 4.2 This function allows the user to choose a black and white display instead of a color display. Choose [ACTIVATE] to change to black and white display. * Only available in 16 and 20 parameter models. Delay Predilute Setup Figure 4.3 The default setting for the pre-dilute start plate is a five second delay, a precautionary measure so that it is not accidentally pressed when running Open tube samples. Select [INACTIVATE] if no delay is needed. 11

Section 4: Setup Menu 3 (continued) PLT Offset Setup Figure 4.4 The function of the Platelet Offset is to set a background count value for PLT. It is recommended to keep Platelet Offset value at 0. The PLT offset can be set from 0 50, if necessary. (This function should not be used for the purpose of forcing QC background count acceptance.) High Altitude Setup Figure 4.5 Figure 4.6 The default setting for this function is 0, no compensation. This function only needs to be activated if various HF, HH, HL, or HN Indicators repeatedly appear (see Section 9.2 in User Manual), then mode may need to be changed to Moderate or Maximum compensation in higher elevations. By choosing 1 or 2, the software incorporates some minor timing sequences for the wash cycles, no other functions are affected. 12

Section 4: Setup Menu 3 (continued) Instrument ID Setup Figure 4.7 If multiple instruments are used in the same laboratory, a specific ID number can be entered here for ease of identification. 13