An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste

Similar documents
Total Measurement Uncertainty Estimation For Tomographic Gamma Scanner R.Venkataraman 1, S.Croft 1, M.Villani 1, R.D. McElroy 1, and R.J.

NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

Improved Detector Response Characterization Method in ISOCS and LabSOCS

Ch. 4 Physical Principles of CT

Identification of Shielding Material Configurations Using NMIS Imaging

Digital Image Processing

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Theoretical Investigations of Tomographic Methods used for Determination of the Integrity of Spent BWR Nuclear Fuel

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

GPU implementation for rapid iterative image reconstruction algorithm

CT vs. VolumeScope: image quality and dose comparison

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

DUAL energy X-ray radiography [1] can be used to separate

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Application of MCNP Code in Shielding Design for Radioactive Sources

FIELD PARADIGM FOR 3D MEDICAL IMAGING: Safer, More Accurate, and Faster SPECT/PET, MRI, and MEG

Radiographic Simulator artist: Version 2

The Determination of Inner Surfaces in Composites by X-Ray Refraction

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System

Recognition and Measurement of Small Defects in ICT Testing

Introduction to Biomedical Imaging

Research Collection. Localisation of Acoustic Emission in Reinforced Concrete using Heterogeneous Velocity Models. Conference Paper.

TOMOGRAPHIC RECONSTRUCTIONS AND PREDICTIONS OF RADIAL VOID DISTRIBUTION IN BWR FUEL BUNDLE WITH PART- LENGTH RODS

Combining Analytical and Monte Carlo Modelling for Industrial Radiology

Medical Imaging BMEN Spring 2016

Enhanced material contrast by dual-energy microct imaging

X-ray simulation and applications

Image Acquisition Systems

THE EVOLUTION AND APPLICATION OF ASYMMETRICAL IMAGE FILTERS FOR QUANTITATIVE XCT ANALYSIS

How does the ROI affect the thresholding?

Evaluation of RayXpert for shielding design of medical facilities

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

Evaluation of Computed Tomography Techniques for Material Identification in Low Level and Intermediate Level Waste

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

Tomographic Reconstruction

Emission Computed Tomography Notes

CHAPTER 4 RAY COMPUTATION. 4.1 Normal Computation

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

BME I5000: Biomedical Imaging

Muon imaging for innovative tomography of large volume and heterogeneous cemented waste packages

Constructing System Matrices for SPECT Simulations and Reconstructions

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

RADIOMICS: potential role in the clinics and challenges

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Scatter Correction Methods in Dimensional CT

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Digital Volume Correlation for Materials Characterization

Automated ADVANTG Variance Reduction in a Proton Driven System. Kenneth A. Van Riper1 and Robert L. Metzger2

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Simulation of Radiographic Testing for POD Assessment

Coordinate Measuring Machines with Computed Tomography

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Spectral analysis of non-stationary CT noise

Computed Tomography for Industry Needs and Status Umesh Kumar

CoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

Technology), Beijing, , China;

DICOM. Supplement 188 Multi-Energy CT Imaging. DICOM Working Group 21 Computed Tomography

Investigation on reconstruction methods applied to 3D terahertz computed Tomography

Introduction to Positron Emission Tomography

Kwabena Arthur. at the. June2017. Certified by: George Barbastathis Professor Thesis Supervisor

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

THREE-DIMENSIONA L ELECTRON MICROSCOP Y OF MACROMOLECULAR ASSEMBLIE S. Visualization of Biological Molecules in Their Native Stat e.

Three-dimensional nondestructive evaluation of cylindrical objects (pipe) using an infrared camera coupled to a 3D scanner

Computational Medical Imaging Analysis

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Basics of treatment planning II

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

Modelling I SPECIFIC REQUIREMENTS ON COMPUTER MODELING OF RADIOGRAPHIC INSPECTIONS OF NUCLEAR COMPONENTS

Continuation Format Page

SHIELDING DEPTH DETERMINATION OF COBALT PHOTON SHOWER THROUGH LEAD, ALUMINUM AND AIR USING MONTE CARLO SIMULATION

SESSION 5: INVESTIGATING LIGHT. Key Concepts. X-planation. Physical Sciences Grade In this session we:

3D TeraHertz Tomography

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

SUPPLEMENTARY INFORMATION

Modeling Strategies for Dynamic Finite Element Cask Analyses

Development of a multi-axis X-ray CT for highly accurate inspection of electronic devices

specular diffuse reflection.

A Source Localization Technique Based on a Ray-Trace Technique with Optimized Resolution and Limited Computational Costs

Bias-Variance Tradeos Analysis Using Uniform CR Bound. Mohammad Usman, Alfred O. Hero, Jerey A. Fessler and W. L. Rogers. University of Michigan

Deformation of granular texture media studied by X-ray CT & 3D DIC at the continuous and microstructure scales

Module 7 VIDEO CODING AND MOTION ESTIMATION

RADIOGRAPHIC LEAST SQUARES FITTING TECHNIQUE ACCURATELY

Quality control phantoms and protocol for a tomography system

GENERAL AUTOMATED FLAW DETECTION SCHEME FOR NDE X-RAY IMAGES

Shielding factors for traditional safety glasses

X-ray tomographic study of wetting bentonite

Development of Electrical Capacitance Volume Tomography (ECVT) and Electrostatic Tomography (EST) for 3D Density Imaging of Fluidized Bed System

Limitations in the PHOTON Monte Carlo gamma transport code

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

NKS GammaUser

Transcription:

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste S.C. Kane 1, S. Croft 1, P. McClay 1, W.F. Mueller 1, R. Venkataraman 1, M.F. Villani 1 and R.J Estep 2 1 Canberra Industries Inc. 800 Research Parkway Meriden, CT 06450 USA ABSTRACT 2 PO Box 1663 Advanced Nuclear Technology, Group N-2, MS B228 Los Alamos National Laboratory Los Alamos, NM 87545 USA Accurate assay of dense heterogeneous waste forms by high-resolution gamma-ray spectroscopic techniques is made especially challenging by severe and non-homogenous attenuation of the waste material and/or a non-uniform source distribution. The development of the Tomographic Gamma Scanner (TGS) arose out of the desire to address these difficulties. By using first generation computed tomography techniques, an attenuation map of the item is constructed. This attenuation map, combined with passive emission scan data, is used to reconstruct attenuation and activity-location compensated assay values. The approach has been successfully used for nondestructive assay of low to medium density drummed waste. In extending the method to higher densities, one faces the problem of poor counting statistics in both the transmission and emission data. The reconstructed images, which are a prerequisite to forming a representative item-specific matrix correction from the composite data, become diffuse and mottled due to poor precision and are, correspondingly, less valuable. In this work, the formation of the linear attenuation map is specifically examined using an analytical and an experimental approach. The analytical approach examined the sampling frequency of the voxel grid to understand the effects of voxel position on the attenuation map. Experimentally, a set of blocks of various materials was used to construct test patterns closely representing the mathematical model embedded in the TGS image reconstruction software. Results are presented to demonstrate the behavior of contrast resolution and spatial resolution. The dual intensity transmission technique is also discussed as a means for extending the practical density range over which the TGS can be deployed. INTRODUCTION In the Non-Destructive Assay (NDA) of radioactive waste, the common assumptions of both uniformity in the matrix material and distribution of radionuclides are sometimes inaccurate. Often, NDA is challenging due to severe non-uniform attenuation, for example, due to concrete or metal scrap present in the waste form. The Tomographic Gamma Scanner (TGS) can address these issues through an approach similar to computed tomography to determine the attenuation map of the waste matrix and the spatial distribution of radioactivity present. Combining this with high-resolution gamma spectroscopy, which allows identification of the discrete gamma-rays, enables energydependent attenuation correction and thus quantification of the gamma-emitting radionuclides present.

As the TGS technique is applied to higher densities, stronger transmission sources and/or longer count times are needed to maintain good counting statistics. Poor counting statistics in both the transmission and emission data result in poor reconstruction of the attenuation and emission maps (quantitative images) [1]. In these cases, the quantitative images become mottled and diffuse. The mottled appearance reflects non-meaningful random voxel-to-voxel scatter as the image is only defined by the available scan data in a global, or average, sense according to the algorithms to reconcile what is an over-determined problem. Radial assay bias can also emerge as the data can no longer tightly define the position of the source (e.g. to within the few voxels at the center of the item or close to the edge which is a boundary in the solution space). The result is that the activity image has a tendency to spread out into the zone of intermediate radii [2]. To address these problems at high densities, one could simply increase the intensity of the transmission beam to improve counting statistics. One way to achieve this goal in an automated general-purpose machine is known as Hybrid TGS. In addition, two alternative approaches have been investigated to further understand the impact of high densities. These were an analytical investigation of the sampling of the voxel grid and a physical experiment to observe the spatial resolution of the TGS technique. This paper discusses each of these steps and presents preliminary results. HYBRID TGS The Hybrid TGS technique uses dual-intensity transmission beams with a triple-pass scan protocol to improve TGS performance at high densities. The transmission source in Hybrid TGS is of much greater activity, at least an order of magnitude, than for a typical TGS. To prevent saturating the detector for regions of the item which are less attenuating, the beam is reduced in intensity by use of a moving filter on the source store. In the triple-pass scan protocol, transmission data for each segment is collected with the low-intensity beam and then with the high-intensity beam in addition to the emission pass. The most suitable transmission data for a given view is chosen at the time of image reconstruction as selection on the fly would be too slow and incur too big a parasitic overhead given the short duration of each data grab. A full description of the Hybrid TGS technique can be found elsewhere [3]. The advantage of the Hybrid TGS technique lies in the use of a stronger transmission source, which provides better counting statistics at higher densities. Additionally the penetrability and utility of the transmission source is improved by selecting a nuclide whose emissions are concentrated in a few high-energy photons. These strategies, combined in the Hybrid TGS approach, make it possible to assay drums with fill densities beyond the historical limit of ~1g.cm -3. The disadvantage is the extra time needed to perform the additional transmission pass. As discussed in Reference 3, higher counting statistics enable better attenuation correction maps to be reconstructed from the transmission data. This can be seen in Figure 1, the attenuation maps for a 55 US gallon drum filled uniformly with sand (of density ~1.6g.cm -3 ). The image in Figure 1a, where only low-intensity transmission data was used, displays the pronounced mottledcheckerboard effect that is not present in Figure 1b where both sets of transmission data were used for image reconstruction. These plots are dithered to improve appearance for qualitative viewing: this involves applying a bi-cubic interpolation between the voxels which amounts to each voxel appearing as if it were composed of 3x3x3 sub-voxel. 2

a) b) Figure 1: Attenuation Maps of a Sand (SiO 2 ) Filled Drum using a) Low-Intensity and b) Hybrid Transmission [3]. The upper potion of each plot is a representative cross-section while the lower part is a side projection. ANALYTICAL INVESTIGATION OF SAMPLING Counting statistics are only one aspect of the TGS technique that can affect assay precision. The scan pattern and resulting sampling frequency of the drum will also influence the reconstruction of emission and transmission data. Due to the design of the collimation system and scan pattern, it has generally been assumed that the sampling of the drum is essentially rotationally uniform. To investigate whether this assumption is strictly accurate, an analysis was performed by reconstructing the rays between detector and transmission source at the beginning of each data grab. Figure 2 demonstrates the ray at the beginning of each data grab for all 150 data grabs in a single layer overlaid on the 10 by 10 voxel grid, as these are typical settings for a TGS assay of a 55 US gallon drum. The voxel grid is shown as it would be at the beginning of an assay. Detector Transmission Source Figure 2: Representation of the Views in a Typical TGS Scan. 3

Figure 2 reveals the view pattern of the drum is, in fact, not symmetric. This is because the drum rotates in the same direction regardless of the direction of translation. To further explore this realization, the frequency with which each voxel was sampled was determined. Figure 3 shows the sampling frequency of each voxel in the form of an iso-surface plot. The lighter the area, the less frequently that voxel is sampled. The two voxels in the center of the voxel map (on the side of the detector in the Figure) are the least frequently sampled. Detector 1 2 3 4 5 6 7 8 9 32-34 30-32 28-30 26-28 24-26 22-24 20-22 18-20 10 A B C D E F G H I J Transmission Source Figure 3: Sampling Frequency of Voxels. The grid is oriented as at start of an assay. The analysis above does not take into account how much (chord length) of the voxel is sampled in a given view. The so-called thickness matrix, which is calculated by the TGS_FIT code [4], provides the sum of the path lengths traveled by a transmission photon through a given voxel during a segment scan. Figure 4 displays these values as a gray-scale iso-surface plot where the total thickness is greatest where the iso-surface is darkest. Detector 1 2 3 4 5 6 7 8 120-130 110-120 100-110 90-100 80-90 70-80 60-70 9 10 A B C D E F G H I J Transmission Source Figure 4: Iso-surface Representation of the Total Thickness (cm) Presented by a Given Voxel. The grid is oriented as at start of an assay. 4

Again, it can be seen that the figure is not rotationally symmetric. Not all regions of the drum are interrogated by the transmission beam equally. On the other hand, not all voxels are equally important in the emission image reconstruction. So this observation alone is not necessarily accuracy limiting for a given assay. It will, nonetheless, partially govern the spatial sensitivity map, which in turn feeds into the detection limit map and overall uncertainty estimate. EXPERIMENTAL INVESTIGATION OF RESOLUTION To explore the limits of the TGS technique further, an experiment was undertaken to examine the spatial and density resolution. The system used has been described elsewhere [2 and references therein] and made use of a 2.4 inch (60.96 mm) diamond-shaped collimator. Blocks of low-density polyethylene foam (which is virtually transparent with a density of 0.02 g.cm -3 ), particle board (a reconstituted wood of density 0.6 g.cm -3 ), and concrete (of density 2.3 g.cm -3 ) were manufactured to approximately the same size as a typical voxel for the routine assay of a 55 US gallon drum. Enough blocks were made so that a full 10x10 voxel pattern of foam and a full voxel map of wood could be constructed on the TGS platform in such a way as to represent the mathematical model of a segment used in the image reconstruction process. Using the representative voxel maps of foam or wood, blocks were selectively removed or replaced by a block of a different matrix, and a two-segment TGS assay was performed with the traditional (one-pass) counting time of 112.5 seconds for transmission per segment. The grid was numbered so changes to the composition of the voxel grid could be recorded. The numbering is as shown in Figure 5 where 1 through 10 represent the edge closest to the transmission source at the start of the assay and 91 through 100 represent the side closest to the detector. The grayed blocks on the edge are not used in the TGS reconstruction as these blocks lie outside of the cylindrical drum. Figure 5: Voxel Map and Numbering Convention. Many permutations were performed by removing a single block from a voxel map of foam, removing two blocks, replacing a single block of foam with wood or concrete, or replacing two blocks with wood. The following figures are the transmission (linear attenuation coefficient) maps of segment 1 of 2 for a few examples of these configurations. Blocks that were removed or exchanged for a different material are indicated by boxes. The maps are based on the 122 kev transmission data from a 152 Eu transmission source. This modest energy line might be expected to have better contrast sensitivity for the light low atomic number matrices than one of the alternative 5

higher energy lines. Figure 10 shows the attenuation coefficient colorbar for Figure 6 through Figure 15. Figure 6: Foam Layer with Block Missing in Position 55. Figure 8: Foam Layer with Wood Block in Position 85. Figure 10: Attenuation Key. Figure 7: Foam Layer Missing Blocks in Positions 23 and 24. Figure 9: Foam Layer with Wood Blocks in Positions 77 and 78. Figure 11: Foam Layer with Concrete Block in Position 29. The same procedure was performed using the wood blocks as the base voxel map. The following figures show the 122 kev attenuation images for a few examples of these configurations. Again, blocks that were removed or exchanged for a different material are indicated by boxes. 6

Figure 12 Wood Layer Missing Figure 13 Wood Layer with Block in Position 56. Concrete Block in Position 83. Lastly, to investigate how higher densities on the outside of the voxel map would affect the attenuation images, concrete blocks were placed on the outside of the wood block layer. Figure 14 shows this scenario. Figure 15 shows a similar setup except that all but the concrete blocks have been removed. Figure 14 Concrete Blocks Surrounding Wood Layer. Figure 15 Concrete Blocks Surrounding Air. The higher energy transmission images were also examined for Figures 14 and 15. The same images were present in the higher energy lines as are shown in the 122 kev transmission images, only the contrast between low and high density was poorer due to smaller attenuation coefficients for the higher energy gamma-rays. OBSERVATIONS AND CONCLUSIONS From these experiments, which are not confounded by activity inside the item, it appears that the attenuation correction spreads when a difference in attenuation is present towards the edge of a drum. This can be seen in almost all Figures presented in the previous section. The spreading is indicative of the radial bias reported elsewhere [2]. The transmission source used in the present experiments was a 6 mci 152 Eu source, which is weaker than typically used in TGS or Hybrid TGS systems. As a result, the attenuation images for the wood block layer already begin to show the mottled-checkerboard effect. The lower counting statistics due to attenuation can be seen in Figures 12 through 15. This affect is worse in Figures 14 and 15 where concrete surrounds a lower-density matrix. The concrete acts as a type of shield, preventing adequate interrogation of the inner matrix with this transmission source. Had a Hybrid TGS system been available for experiment, the mottled-checkerboard effect would not have been present at the densities investigated in this experiment. Due to the magnitude of the transmission source, the counting statistics would be significantly improved leading to a 7

higher fidelity attenuation image. A major benefit results by simply changing the transmission source from 152 Eu to 60 Co, regardless of increasing the activity of the source, as the latter emits 130% more photons in the principle high energy lines (1173 kev and 1332 kev) than the former (1112 kev and 1408 kev). The low energy lines do not transmit yet consume the straight through beam (unattenuated) counting rate dynamic range. The experiment somewhat supports the assumption that the spatial resolution of the transmission map is just greater than the size of a voxel in the mathematical grid. Further experimental and simulation evaluation of the data is required before an overall conclusion can be drawn on the spatial and density resolution of the TGS technique for integrated transmission and emission mapping. For instance, although it has been shown that the transmission scan does not uniformly sample the layer, it is also true that not all voxels contribute equally to the modulation of a particular source distribution. The combined effects are correlated and, consequently, difficult to unravel in a generic sense, requiring further investigation. REFERENCES 1. R. Venkataraman, M. Villani, S. Croft, P. McClay, R. McElroy, S.C. Kane, W. Mueller, and R.J. Estep, An Integrated Tomographic Gamma Scanning System for Non-Destructive Assay of Radioactive Waste, Accepted for Publication in Nuclear Instrumentation and Methods in Physics Research A. 2. R. Venkataraman, S. Croft, M. Villani, and R.J. Estep, Performance Study of the Tomographic Gamma Scanner for the Radioassay of Drums, Proceedings of 45 th Annaual INMM Meeting, Orlando, FL, 18-22 July 2004. 3. S. Croft, S.C. Kane, P. McClay, R.J. Estep, W.F. Mueller, M.F. Villani, and R. Venkataraman. Extending the Dynamic Range of the TGS Through the Use of a Dual Intensity Transmission Beam. INMM 47 Annual Meeting, Nashville, TN, July 2006. 4. R.J. Estep, User s Manual for TGS_FIT version 2.0. LANL Report LA-UR-00-4792. 8