DICOM CONFORMANCE STATEMENT

Similar documents
DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number:

DICOM. Conformance Statement For VINNO Medical Ultrasound System Revision 2.0. Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd.

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

Technical Publications

nstream Version 3.1 DICOM Conformance Statement

DICOM Conformance Statement June 24, 2011 Version 0.4 ImageWorks Internal document number:

Artwork consists of twenty three (23) 8½ x 11 inch pages.

DICOM CONFORMANCE STATEMENT

Punctual Dicom Workstation

DICOM Conformance Statement

REGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version).

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL DRAD-3000A, DRAD-3000E

Raffaello Conformance Statement for DICOM V3.0

DICOM Conformance Statement

Technical Publications

Technical Publications

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0

DICOM Conformance Statement

GE Healthcare. Technical Publications. ConnectR Plus Version 5.0 DICOM CONFORMANCE STATEMENT. GE Healthcare IT. Direction DOC Revision 0.

DICOM Conformance Statement. DSI Release 4.4. Document Number June Copyright Philips Medical Systems Nederland B.V.

DICOM Conformance Statement for FLEXAVISION HB/FD

Printlink III-ID/IV. Revision History. Date Version Description 20/08/2004 Ver First edition

Technical Publications

Technical Publications

TITAN DICOM Conformance Statement

Technical Publications

OASIS V4.0 DICOM Conformance Statement Rev. 3

Whole Body X-ray CT System Supria. DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A

GEMINI DICOM Conformance Statement 16 Power PET/CT Imaging System Configuration

g GE Healthcare Technical Publications encore TM 12.0 CONFORMANCE STATEMENT for DICOM Direction Revision 11

DICOM. Conformance Statement. Panorama Release 6. Panorama 0.6T Panorama 0.23T Panorama 0.23T I/T Panorama 0.23T R/T

Technical Publications

PRO DATA Srl. MicroPrint. DICOM PRINT SCU Conformance Statement (MicroPrint-prnsrv-print-scu.sxw)

Technical Publications

Technical Publications

DICOM Conformance Statement

Philips Medical Systems DICOM Conformance Statement USIT 1.5

DE32-DCM DentalEye 3.2. DICOM conformance statement

Technical Publications

JiveX Enterprise PACS Solutions. JiveX DICOM Worklist Broker Conformance Statement - DICOM. Version: As of

Technical Publications

DICOM 3.0 Conformance Statement for PlatinumOne Systems

Hologic Physician s Viewer 5.0 DICOM Conformance Statement

DICOM Conformance Statement

Philips Medical Systems. Intera Document Number April 2003

KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT

Fluoroscan InSight V5.0.6 DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM

Technical Publications

DICOM 3.0 Conformance Statement DXIMAGE RIS

Technical Publications

DICOM Conformance Statement

Technical Publications

DigiRoebs 3.1. DICOM Conformance Statement V1.0. medigration GmbH. DICOM Conformace Statement Seite 1 / 38.

DICOM Conformance Statement Merge Eye Care PACS v. 4.0

Lumify 1.8.x DICOM Conformance Statement

MediaWorkStation. MWS DICOM Push Interface. DICOM Conformance Statement. Version 2.0. English

DICOM CONFORMANCE STATEMENT FOR CANON DIGITAL RADIOGRAPHY SYSTEM. MODEL DRAD series / MRAD series

Technical Publications

Technical Publications

Uscan. DICOM Conformance Statement

Technical Publications

Technical Publications

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004

Technical Publications

Technical Publications

DICOM Conformance Statement. Forum

DxMM/DxWin DICOM 3.0 Conformance Statement. Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B

Technical Publications

Medical Imaging Consultants, Inc.

SonoWin Picture Archiving System

DICOM Conformance Statement for GALILEI. Software Version V6.0

Philips Medical Systems DICOM Conformance Statement USIT 1.5 L3

DICOM Conformance Statement for DR-200/DRU-200

DICOM Correction Item

Villa Sistemi Medicali DRF 4343

AGFA HEALTHCARE DICOM Conformance Statement

Canon DR Image Viewer System Basic DICOM Conformance Statement

DICOM Conformance Statement FORUM

DICOM Conformance Statement

DICOM Conformance Statement FORUM

Senographe Essential Acquisition System

X-Porte DICOM Conformance Statement

Cardiac Workstation Last Updated January 10,

Technical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG

Technical Publications

BRIT Systems. DICOM Conformance Statement

Technical Publications

AVIA (Dx MM) DICOM 3.0 Conformance Statement

DICOM Conformance Statement for PenFetch

FluoroPRO. DICOM conformance Statement

Technical Publications

KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR /2017/PI-E 1/32

Aloka ProSound DICOM

Sonovision DICOM Conformance Statement

Transcription:

MultiScan G-Arm System Versions DigiArc 100AU/100AE DICOM CONFORMANCE STATEMENT Whale Imaging 1 / 16

REVISION HISTORY REVISION DATE 1.0 August 5.2013 For information contact Headquarters Whale Imaging Inc 200 Reservoir Street, Suite 301 Needham, MA. 02494 USA Tel +1 781 449 7200 Manufacturing Facility Beijing East Whale Imaging Technology Co., Ltd. New City Industry Garden B2-2, No.9 KeChuang 2 Street, YiZhuang Economy Technology Exploitation Section, Beijing, 100023, China Tel: +86-10-51281263 Fax: +86-10-67892105 Copyright 2013. All rights reserved. No part of this document may be reproduced or transmitted without express written permission from Whale Imaging Inc. Whale reserves the right to discontinue or make changes to the communications features described by these DICOM Conformance Statements on any of Whale s products with which they are associated now or in the future. 2 / 16

1. INTRODUCTION... 4 1.1 Purpose of this Manual... 4 1.2 References... 4 1.3 Definitions... 5 1.4 Symbols and Abbreviations... 5 2. IMPLEMENTATION MODEL... 5 2.1 Application Data Flow Diagram... 5 2.2 Functional definitions of AE... 6 2.3 Sequencing of Real-World Activities... 6 3 AE Specifications... 6 3.1 Storage AE specification... 6 3.1.1 Association Establishment Policy... 7 3.1.2 Association Initiation Policy... 7 3.2 Query AE specification... 10 3.2.1 Association Initiation Policy... 10 3.3 Printer AE specification... 12 3.3.1 Association Establishment Policy... 13 3.3.2 Association Initiation Policy... 13 4 Communication Profile... 15 4.1 Supported Communication Stacks... 15 4.2 Physical Media Support... 15 5 Extensions/specializations/privatizations... 15 6 Configuration... 15 6.1 AE Title/Presentation Address Mapping... 15 6.2 Configurable Parameters... 15 7 Support of the Extended Character Sets... 16 3 / 16

1. INTRODUCTION 1.1 Purpose of this Manual This statement accurately describes the DICOM conformance of the DigiArc 100AE/AU X-ray system and includes the DICOM information objects, service classes, and communication protocol supported by the implementation. When reading this statement, please refer to the second part of the DICOM 3.0 protocol (Conformance). This document is intended for those competent with software design and/or system integration. The reader of this document should be familiar with the DICOM Standard and with the terminology and concepts used in that Standard. The use of these DICOM Conformance Statements should be made in conjunction with the DICOM Standards. This document is intended to facilitate communication with Whale imaging equipment. It is not sufficient on its own to ensure that inter-operation will be successful. The user should proceed with caution and consider i) integration of the system beyond the scope of the DICOM 3.0 standards ii) Testing and validation of the device prior to completion iii) future evolution of the DICOM standard with respect to future integrated devices. iv) any deleterious effects on system performance due to integration of the system, which are the responsibility of the user. 1.2 References 2009 version of the DICOM standard. 4 / 16

1.3 Definitions 2009 version of the DICOM standard. 1.4 Symbols and Abbreviations 2009 version of the DICOM standard. 2. IMPLEMENTATION MODEL DigiArc 100A X-ray facility has three kinds of DICOM service: DICOM transmit, WorkList, DICOM print. 2.1 Application Data Flow Diagram DICOM Standard Interface Transmit image Storage AE WorkList/ Query WorkList/ Query AE DICOM Server Print image Print AE When the user selects a study to transmit and clicks the transmit button the storage application 5 / 16

entity will create an association and after the connection is successful will start transmitting images. When all images of this study are transmitted to the server the release request will be sent and the association is closed. When the user opens the WorkList interface or clicks on the query button the WorkList application entity will create an association, and after the connection is successful patient and study information that satisfy the query conditions on the server will be obtained and displayed. Choose one patient item and double-click to enter Fluoroscopy mode. After selecting the images and format to print, click on the print button. The printer application entity will create a connection and after the connection is successful will start sending all the print images. When all images of this study are sent to the server the release request will be sent and the association is closed. 2.2 Functional definitions of AE Storage Application Entity supports the following functions: Initiates a DICOM association to send images. Transmits DICOM images to the DICOM Storage SCP. Printer Application Entity supports the following functions: Initiates a DICOM association to print images. Transmits images data to DICOM Print SCP. Query Application Entity supports the following functions: Initiate a DICOM association to ask for remote patient and study information. Initiates a DICOM Worklist Query to research patient and study information that satisfy the query conditions. 2.3 Sequencing of Real-World Activities No sequencing of Real-World activities is required. 3 AE Specifications 3.1 Storage AE specification This Application Entity provides Standard Conformance to the following DICOM V3.0 SOP 6 / 16

Classes as an SCU: SOP Class Name Digital X-Ray Image Storage -For Presentation SOP Class UID 1.2.840.10008.5.1.4.1.1.1.1 3.1.1 Association Establishment Policy 3.1.1.1 General The maximum length PDU for the DICOM Storage SCU is: 16384 3.1.1.2 Number of Associations DICOM Storage SCU AE can receive only one association at one time. 3.1.1.3 Asynchronous Nature DICOM Storage SCU AE does not support asynchronous mode. All operations will be performed synchronously. 3.1.1.4 Implementation Identify Information Implementation Version Name: WHALE Beluga v2.0 Implementation Class UID: 1.2.156.8772657669.7145658277 3.1.2 Association Initiation Policy clicked. The Storage AE will initiate an association when the transmit button in the user interface is 3.1.2.1 Real -World Activity 3.1.2.1.1 Associated Really-World Activity The DICOM Storage AE initiates an association for the user to send studies. 3.1.2.1.2 Proposed Presentation Context Table Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Negotiation 7 / 16

Name Digital X-Ray Image Storage For Presentation UID 1.2.840.10008.5.1.4.1.1.1.1 Name List Implicit VR Little Endian UID List 1.2.840.10008.1.2 SCU None 3.1.2.1.3 SOP Specific Conformance to Digital X-Ray Image Storage SOP Class When the SCU invokes a C-STORE requirement, the server will send the success or failure status, the SCU take diffident action according to these statuses. The following IOD Modules tables specify information about Real-World Objects. IE Module Usage Patient Patient M Study General Study M Patient Study U Series General Series M CR Series M Equipment General Equipment M General Image M Image Pixel M Image CR Image M VOI LUT U SOP Common M The following Module Attribute tables contain the supported elements for each module in the Module table above. Name Tag Type VR Patient's Name 0010,0010 2 PN Patient ID 0010,0020 2 LO Patient's Birth Date 0010,0030 2 DA 8 / 16

Patient's Sex 0010,0040 2 CS Study Instance UID 0020,000D 1 UI Study Date 0008,0020 2 DA Study Time 0008,0030 2 TM Study ID 0020,0010 2 SH Accession Number 0008,0050 2 SH Referenced SOP Class UID 0008,1150 1C UI Referenced SOP Instance 0008,1155 1C UI UID Modality 0008,0060 1 CS Manufacturer 0008,0070 2 LO SOP Class UID 0008,0016 see PS3.3 C.12.1.1.1 UI SOP Instance UID 0008,0018 see PS3.3 C.12.1.1.1 UI Series Instance UID 0020,000E 1 UI KVP 0018,0060 2 DS Exposure Time 0018,1150 2C IS X-ray Tube Current 0018,1151 2C IS View Position 0018,5101 2C CS Pixel Data 7FE0,0010 1 OW/OB The following Status tables contain the status codes that are more specifically processed when receiving messages from a Storage SCP equipment: Service Status Status Codes Further Meaning Application Behavior When receiving Status Codes Related Fields Processed if received Refused A7XX Out of resources Association is closed with Remote AE. Appropriate message is displayed to the user. 9 / 16

Error A9XX CXXX Data Set does not match SOP Class Cannot Understand General Warning message is logged. Association is not closed with Remote AE. Error message is displayed to the user. General Warning message is logged. Association is not closed with Remote AE. Error message is displayed to the user. (0000,0901) (0000,0901) B000 Coercion of Data Elements General Warning message is logged. Association is not closed. (0000,0901) Warning B007 Data Set does not match SOP Class General Warning message is logged. Association is not closed. (0000,0901) B006 Elements Discarded General Warning message is logged. Association is not closed. (0000,0901) Success 0000 None 3.2 Query AE specification This Application Entity provides Standard Conformance to the following DICOM V3.0 SOP Classes as an SCU: SOP Class Name SOP Class UID Study Root Query SOP Class 1.2.840.10008.5.1.4.1.2.2.1 3.2.1 Association Initiation Policy 10 / 16

The Query AE will initiate an association when the WorkList interface is opened or the Query button is clicked. 3.2.1.1 Real -World Activity 3.2.1.1.1 Associated Really-World Activity The DICOM Query AE initiates an association for the user to query a remote database. 3.2.2.1.2 Proposed Presentation Context Table Name Modality Worklist SOP Class Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role UID 1.2.840.10008.5.1.4.31 Name List Implicit VR Little Endian UID List Extended Negotiation 1.2.840.10008.1.2 SCU None The following table contains the keys contained in the C-FIND Request. Name Tag Type VR Patient's Name 0010,0010 2 PN Patient ID 0010,0020 2 LO Patient's Birth Date 0010,0030 2 DA Patient's Sex 0010,0040 2 CS Scheduled Procedure Study Date 0040,0002 2 DA Scheduled Procedure Study Time 0040,0003 2 TM Accession Number 0008,0050 2 SH Modality 0008,0060 1 CS The following Status tables contain the status codes when receiving messages from a Query SCP: Service Status Status Codes Further Meaning Application Behavior When receiving Status Codes Related Fields Processed if received 11 / 16

Refused A700 Out of resources Association is closed. Appropriate message is displayed to the user. Failed A900 Cxxx Identifier does not match SOP Class Unable to process Association is closed. Error message is displayed to the user. Association is closed. Error message is displayed to the user. (0000,0901) (0000,0901) Cancel FE00 Matching terminated due to cancellation Association is closed. Error message is displayed to the user. None Success Pending 0000 FF00 FF01 Matching is complete - No final identifier is supplied Matches are continuing-current Match is supplied and any Optional Keys are supported in the same manner as Required Keys. Matches are continuing Warning that one or more Optional Keys were not supported for existence and/or matching for this Identifier None Identifier Identifier 3.3 Printer AE specification This Application Entity provides Standard Conformance to the following DICOM V3.0 SOP 12 / 16

Classes as an SCU: SOP Class Name Basic Grayscale Print Management Meta SOP Class SOP Class UID 1.2.840.10008.5.1.1.9 3.3.1 Association Establishment Policy 3.3.1.1 General The maximum length PDU for the DICOM Printer SCU is: 16384 3.3.1.2 Number of Associations DICOM Printer SCU AE can receive only one association at one time. 3.3.1.3 Asynchronous Nature DICOM Printer SCU AE does not support asynchronous mode. All operations will be performed synchronously. 3.3.1.4 Implementation Identify Information Implementation Version Name: WHALE Beluga v2.0 Implementation Class UID: 1.2.156.8772657669.7145658277 3.3.2 Association Initiation Policy The Printer AE will initiate an association when you click the print button in user interface. 3.3.2.1 Real -World Activity 3.3.2.1.1 Associated Really-World Activity The DICOM Printer AE initiates an association for the user to print image. 3.3.2.1.2 Proposed Presentation Context Table Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Negotiation Name UID Name UID List 13 / 16

The following Module Attribute tables contain the supported elements for each module in the Module table above. List Basic Implicit Grayscale Print 1.2.840.10008.5 VR Management 1.2.840.10008.1.2 SCU None.1.1.9 Little Meta SOP Endian Class 3.3.2.1.3 SOP Specific Conformance to Basic Grayscale Print Management Meta SOP Class Name Tag VR Usage SCU/SCP Number of Copies 2000,0010 IS U/M Print Priority 2000,0020 CS U/M Annotation Display Format ID 2010,0030 CS U/U Border Density 2010,0100 CS U/U Empty Image Density 2010,0110 CS U/U Film Orientation 2010,0040 CS U/M PORTPAIT Film Size ID 2010,0050 CS U/M Image Display Format 2010,0010 ST M/M Magnification Type 2010,0060 ST U/M Referenced SOP Class UID 0008,1150 UI M/M Referenced SOP Instance UID 0008,1155 UI M/M Image Position 2020,0010 US M/M Pixel Data 7FE0,0010 OW/OB M/M 14 / 16

4 Communication Profile 4.1 Supported Communication Stacks DICOM Upper Layer (PS 3.8) is supported using TCP/IP. 4.2 Physical Media Support The product is provided with a 1000 Mb/s auto-sensing Ethernet interface. 5 Extensions/specializations/privatizations Not use 6 Configuration 6.1 AE Title/Presentation Address Mapping The default local AE Title is WHALE and is user configurable. The modality is DXW and is not user configurable. 6.2 Configurable Parameters The user can change the following configuration in the DICOM setup interface. AE Title DICOM Port Number IP Address Film Size Image Layout The user can specify the following Query keys. Patient Name 15 / 16

Patient ID Study Time Modality 7 Support of the Extended Character Sets All applications will support only the ISO_IR 100 as extended characteristic sets. Any incoming SOP instance that is encoded using another extended character set will not be validating. 16 / 16