H290 HUMIDITY TRANSMITTER INSTRUCTION MANUAL

Similar documents
MP 101A HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL

HygroClip M22 Series. 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL

A1H Portable Humidity Temperature Indicator Instruction Manual

HC322 HUMIDITY / DEW POINT AND TEMPERATURE TRANSMITTER INSTRUCTION MANUAL

HygroFlex HF3 Transmitters. User Guide

Humidity Temperature Transmitter User Guide

HygroFlex HF7 Humidity Temperature Transmitters. User Guide

HygroFlex HF7 Humidity Temperature Transmitters. User Guide

HygroFlex HF3 Transmitters and Thermohygrostats: User Guide

Humidity goes Digital

HygroPalm. Portable Humidity Temperature Indicator. Instruction Manual

HygroFlex. Humidity Temperature Transmitter. Instruction Manual

HygroFlex M. Meteorological Humidity Temperature Transmitter. Instruction Manual

HygroFlex TF5 Temperature Transmitter. User Guide

Humidity goes Digital

HygroClip DIGITAL Humidity and temperature probe

DWYER INSTRUMENTS, INC. Series AVUL Air Velocity Transmitter. Specifications - Installation and Operating Instructions.

HMT330 Series Humidity and Temperature Transmitters for Industrial Applications

HygroClip2 ADVANCED Humidity Temperature Probes

INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor

Duct Sensors. Building Technologies HVAC Products. Symaro. for relative humidity and temperature with calibration certificates

HygroFlex5-EX Humidity & Temperature Transmitter Instruction Manual

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods

AM3 Humidity. Temperature. Dew Point Indicator INSTRUCTION MANUAL

HM4. Humidity and Temperature Probe with Heated Sensor. Instruction Manual E-M-HM4-V1_12. HM4 Humidity and Temperature Probe with Heated Sensor

CDD4 Duct Carbon Dioxide Transmitter

HygroClip 2 (HC2) Humidity Temperature Probes User Guide

KOBOLD PDA Series Digital Pressure Transmitter

Instruction Manual. M Pump Motor Controller. For file reference, please record the following data:

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

RHT-WM-485-LCD Transmitter and RHT-DM-485-LCD Transmitter

NightStar. Installation, Operation, & Maintenance Instructions M /04/2016

Transmitters RHT-WM-485-LCD, RHT-DM-485-LCD and RHT-P LCD

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

KOBOLD TDA Series Digital Temperature Transmitter

Programmable transmitter of temperature, relative humidity and other derived humidity values with 0 to 10 V outputs Instruction Manual

Model HU-224/225. Technical Information TI.224/ Humidity Transduce r FOR ADDITIONAL INFORMATION SEE HU-224/225 DATA SHEET

EE671 - Miniature Air Flow Transmitter

JUMO PINOS L01. Calorimetric flow sensor. Brief description. Special features. Customer benefits. Data Sheet

OPERATION MANUAL MODEL TT-LP

TOXALERT MODEL AIR 2000

ON/OFF HUMIDIFIER CONTROL (ROOM 0R DUCT SENSOR) WITH OUTDOOR TEMPERATURE RESET OVERVIEW WALL CONTROLLER INSTALLATION (TCY-BH-T-U)

DMT345 and DMT346 Dewpoint Transmitters

HWS Series Wall Mount with Relay Setpoints, LCD, and Humidistat Control

Table of contents: 1. HygroClip 2 in general. 2. The different probes. 3. Features in details. 4. Customer benefits. 5. Replacement overview

Mini Photoelectric Sensor

HygroLog HL20 and HL21 Data Logger. User Guide

HygroLog HL20 and HL21 Data Logger. User Guide

Keysight M8000 Series BER Test Solutions

1 Channel Strobe Controller ORDERCODE 40226

Precision humidity/temperature control/sensing

HygroClip DI Digital Interface for HygroClip Probes v 1.0 Instruction Manual

Room sensors QFA4160 QFA41 D AQF4150 AQF3153. Symaro. for relative humidity and temperature with calibration certificates

Description. Type Product No. Measuring Range PMU-Type max. ambient temperature

CDD4 Series Room CO2 Transmitter Installation Instructions

QXA2000 QXA2001. Condensation Monitor QXA2001 QXA2000 AQX2000

CDD Carbon Dioxide Transmitter

DD1590 Series Proximity Driver. Connection Technology Center, Inc. Vibration Analysis Hardware. DD1590 Series. Proximity Probe Driver.

L Series Humidity/-temperature sensors for use in heating, ventilation and air conditioning Technical data Humidity

Advanced Test Equipment Rentals ATEC (2832) HMP45A&HMP45D HUMIDITY AND TEMPERATURE PROBES Operating Manual

E4PA-N. Application Examples. Features. Ultrasonic Displacement Sensor. Now, more advanced with a new mutual interference prevention function.

Thread length. 27 (40) mm. 27 (44) mm. 27 (40) mm. 34 (50) mm. 34 (49) mm. 39 (60) mm. 39 (54) mm. 44 (65) mm. 44 (59) mm

E2K-F. Flat Capacitive Sensor with a Thickness of Only 10 mm. Flat Proximity Sensor. Ordering Information. Sensors [Refer to Dimensions on page 4.

HMT330 Series Humidity and Temperature Transmitters for Industrial Applications

HT Series Wall Mount with Relay Setpoints, LCD, and Humidistat and Thermostat Control

Tempco Instruction Manual

Model 2380 Rack-Mount Kit

Thread length (overall length) material

E4PA. Ultrasonic Displacement Sensor. Ordering Information

PHASETRONICS. SCR Power Control Specialists. EP1 Series Power Control Single Phase SCR Amps OPERATION & SERVICE MANUAL

Operation and Maintenance Manual

Model 7705 Control Module

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION

TPE 1464 Series Pressure Transducer

Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series

Model 2380 Rack-Mount Kit

PTU300 Combined Pressure, Humidity and Temperature Transmitter

Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2

EX-RC1 Remote I/O Adapter

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions

Mini-DVI-WP Series WALL PLATE MINIATURE MULTIMODE FIBER OPTIC DVI TRANSMISSION SYSTEM

E2K-F. Flat Capacitive Sensor with a Thickness of Only 10 mm. Flat Proximity Sensor. Ordering Information. Sensors [Refer to Dimensions on page 4.

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

THERMASGARD TM 65. Immersion screw-in duct temperature measuring transducer, calibratable, with multi-range switching and active output TECHNICAL DATA

INFORMATION AMPLIFIER TA-102 OPERATING INSTRUCTIONS TABLE OF CONTENTS

AC/DC Current Probe Model SL206 USER MANUAL

Installation Sensing distance Connection Output configuration Operation mode NO Operation mode NC

DTM-321 REL HOLD T2 K-TYPE TEMP: 20 C 60 C 140 F C F HUMIDITY: %RH POWER-UP OPTIONS

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions

MODEL 8000MP LEVEL SENSOR

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions

Piccolo2-OEM Fiber-optic Version (Item nr. PICO2-OEM)

Temperature and humidity module DHT11 Product Manual

Scanner 2000 Steam Mass Flow Transmitter

Schneider Electric s SCD series of duct mount sensors measure the levels of CO 2. , RH (if equipped) and temperature of air inside a duct.

IGB Series. Operation Manual. Platform Scale. Kg/Lb Specification IMPORTANT

Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V C30B and V T40B.

Meet requirements of USB 2.0

Transcription:

H290 HUMIDITY TRANSMITTER INSTRUCTION MANUAL

DESCRIPTION 4 OPERATION 4 Power Supply 4 Output Signal 4 Temperature Operating Range and Limits 4 Humidity Limits 5 Temperature Compensation 5 Sensor Protection 5 Optional Temperature Sensor 5 INSTALLATION 6 Mechanical Installation 6 General Recommendations 6 Probe Installation 6 Probe Position 7 Calibration Port 7 Electrical Installation 8 Board Plan 8 Wiring Diagram 8 Cabling 8 Grounding 9 Optional Temperature Sensor 9 MAINTENANCE 10 Cleaning or Replacing the Dust Filter 10 Periodic Calibration Check 10 CALIBRATION 10 CALIBRATION BASICS 10 Calibration Device 10 Humidity Standards 10 Calibration Potentiometers and Test Connector 11 Calibration Procedures 11 One-Point Calibration 11 Full Calibration 12 SPECIFICATIONS 13

-3- PLEASE, READ THIS FIRST Check the product for any physical damage that may have occurred during shipment. We carefully pack and routinely insure all shipments. If any damage has occurred, it is your responsibility to file a claim with the carrier, prior to returning the damaged product. Please note that our warranty does not cover damage during shipment. Prior to installation, get fully familiarized with the operating limits of the product and with the installation instructions provided in this manual. Do not unnecessarily remove the sensor protection (dust filter or slotted cap) from the probe. Both sensors (humidity and temperature) can be mechanically damaged by careless removal of the protection. The ROTRONIC HYGROMER humidity sensor looks like a small white paper tag. Do not remove from the probe! Each ROTRONIC instrument is carefully calibrated before shipment. No further adjustments should be required before installation. If you have any question or problem, please call our service department at 631/427-3898 and press 5 (or ask for extension 21).

-4- DESCRIPTION The H290 consists of a humidity probe that is hard wired to a small open board with a 6 ft (2 meter) cable. The probe has provisions for adding an RTD temperature sensor and is pre-wired for a direct 3- wire RTD connection with an external temperature measurement device. The H290 series was specifically designed for OEM applications that require humidity measurement within the temperature range of -5 to 320 F (-20 to 160 C) and can survives exposure to temperatures in the range of -100 to 390 F (-75 to 200 C) at the probe. To maintain accuracy, the H290 series features a fully temperature compensated humidity output signal. Because of this, the H290 can be directly used with most makes of controllers. Humidity is measured with the ROTRONIC HYGROMER C94 capacitive humidity sensor. This sensor offers fast response and exceptional stability, even in high humidity environments. Sensor durability has proven to be excellent in a wide range of industrial applications. This fact is reflected in the 3-year full warranty that covers the H290. Reliability is further enhanced by the easy-to-perform field calibration. The H290 is a 3-wire transmitter that can operates either from a DC power supply (10-35 VDC) or from a 24 VAC power supply. The H290 can be ordered with either a 4..20 ma current output or a 0..5 V voltage output. OPERATION Power Supply The H290 can operate with an unregulated supply voltage between 10 and 35 VDC or with 24 VAC. When powered with a DC voltage, the H290 requires a maximum of 30 ma. Output Signal The H290 can be ordered with one of the following output signals: 4 to 20 ma = 0 to 100%RH 0 to 5 V = 0 to 100 %RH. The humidity signal is linearized. Temperature Operating Range and Limits The H290 was designed to operate within -5 to 320 F (-20 to 160 C) at the probe. Temperature up to 390 F (200 C) or below 5 F (-20 C) will not harm the probe. Temperature at the electronics (board) should be within 20 to 122 F (-5 to 50 C). Operating either the board or the probe of the H290 outside of the temperature limits can result in permanent damage.

-5- Humidity Limits Depending on temperature, the maximum humidity to which the probe can be subjected is as follows:. 100 %RH up to 185 F (85 C). 90 %RH at 212 F (100 C). 60 %RH at 260 F (125 C). 25 %RH at 320 F (160 C) Exceeding the above limits may "shift" the humidity sensor and require a new calibration. At temperatures below 185 F (85 C), the C94 sensor can operate for long periods of time in environments up to 100%RH. Direct condensation does not damage the sensor, however, this can produce an overflow reading (an output signal of more than 100 %RH) for as long as condensation is present on the humidity sensor. To ensure meaningful readings in high humidity applications, be sure to provide tight temperature control so as to prevent the sensor from becoming colder than its environment. Temperature Compensation Practically every make of relative humidity sensor requires a temperature compensation in order to measure accurately over a wide range of temperature conditions. In the specific case of an instrument using a capacitive sensor, compensation is required because the dielectric characteristics of both the water molecule and the hygroscopic polymer used in the sensor vary with temperature. The H290 uses the temperature measured by an RTD located next to the humidity sensor to provide the necessary compensation to the humidity output signal. Within the temperature range of -5 to 320 F (-20 to 160 C), this compensation maintains the temperature error to ±2%RH or less. Sensor Protection As a standard, the sensors of the H290 are protected with a PPS slotted cap with a stainless steel wire mesh insert. When the application is clean and requires the fastest probe response as possible to humidity and temperature changes, the standard wire mesh insert may be replaced with a larger mesh size screen insert to increase air flow at the sensor (consult factory). A typical example of such an application would be a fast ramping environmental test chamber. Do not unnecessarily remove the protective cap from the probe and never use the H290 without protecting the sensors. Optional Temperature Sensor The probe of the H290 has a 6-pin bulkhead. Two of these pins (pins 1 and 6) are free and can be used to add an RTD temperature sensor to the probe. The probe cable and the circuit board are already prewired for a direct 3-wire RTD connection (see INSTALLATION).

-6- INSTALLATION Do not remove the protective slotted cap from the probe. The sensors can easily be damaged when not protected. The ROTRONIC HYGROMER humidity sensor has the appearance of a small white paper tag. Do not remove from the probe! Mechanical Installation General Recommendations Relative humidity is extremely dependent on temperature. Accurate measurement of relative humidity requires that the probe and sensors be always at the temperature of the environment to be measured. When mounting the probe, use the following guidelines to obtain good instrument performance: a) Select a representative location: install the probe at a location where humidity, temperature and pressure conditions are representative of the environment to be measured. b) Provide good air movement at the probe: air velocity of at least 200 ft/ minute (1 meter/second) facilitates adaptation of the probe to changing temperature. c) Avoid the following: (1) Close proximity of the probe to a heating element, a cooling coil, a cold or hot wall, direct exposure to sun rays, etc. (2) Close proximity of the probe to a steam injector, humidifier, direct exposure to precipitation, etc. (3) Unstable pressure conditions resulting from excessive air turbulence. d) Immerse as much of the probe as possible in the environment to be measured. Probe Installation The probe of the H290 was designed for through-wall installation. To facilitate probe installation and removal, we recommend using a probe holder mod. QMA-15. This holder is a mounting flange that is equipped with a compression fitting. QMA-15 Wall

-7- Make sure that 9 to 9 1/2" (230-240mm) are immersed in the environment to be measured (see diagram at the end of this manual). Probe immersion depth is critical. Insufficient probe immersion may result in a difference between the temperature of the sensors and that of the environment. This will create a large error of measurement and/or a malfunction. OK 9 to 9.5" immersion NG Small Diameter Duct NG OK Probe Position Install the probe so as to prevent the accumulation of condensation water at the level of the sensor leads. Install the probe so that the probe tip is looking downward. If this is not possible, install the probe horizontally. Calibration Port We recommend providing a calibration port (diameter 0.5 or 13 mm) next to the probe of the H290. During maintenance, this port can be used to introduce the probe of a PFC150 calibrator (available from ROTRONIC). For more details, see One-Point Calibration.

-8- Electrical Installation Board Plan R A S +VDC +VAC +RH 0 Make sure that you can correctly identify the function of each terminal. Applying power (+VDC / +VAC) to the output signal terminal (+RH) can severely damage the unit. HM H80 H35 Wiring Diagram + 10..35 VDC +VDC +VAC 24 VAC (Phase) +VDC +VAC VDC + 4..20mA = 0..100%RH or 0..5V = 0..100%RH RH VAC + 4..20mA = 0..100%RH or 0..5V = 0..100%RH RH - GND 0 Neutral / GND 0 Cabling Avoid running the cable connecting the unit in the same conduit as 110 VAC power cables. If this cannot be avoided, a shielded cable or a cable with twisted wires may be required to prevent interference due to electromagnetic induction. Regarding the maximum cable length, use the following guidelines:

-9- - Current Output: The maximum cable length is limited by the fact that the maximum load is 500Ω. - Voltage Output: The current consumption of the H290 is less than 30 ma. This current flows through the return wire which is common to the output signal and the supply voltage. With a 12 AWG wire, the maximum cable length should not exceed 20 to 25 feet in order to minimize the error created by the voltage drop in the common (0) wire. Grounding We generally recommend grounding the H290, especially if the electronics will be subjected to a low humidity environment (35 %RH or less). Optional Temperature Sensor The probe of the H290 is pre-wired to accept an optional RTD temperature sensor. The RTD should be soldered to pins 1 and 6 located on the probe bulkhead. Pin 1 is connected by means of a green wire to the solder pad labeled S on the electronics board (see Board Plan). Pin 6 is connected by means of 2 wires: brown to the solder pad labeled R and violet to the solder pad labeled A. The board of the H290 has provisions for adding a terminal block to facilitate a direct RTD connection. RTD 3-Wire Connection: R +VDC R S (Board) Pin 1 (Pbe) Pt100 RTD Pin 6 (Pbe) 100 Ω R (Board) A (Board) GND

-10- MAINTENANCE Cleaning or Replacing the Dust Filter The dust filter should be cleaned from time to time, depending on the conditions of measurement. Cleaning should be done without removing the filter from the probe. Gently wipe the filter. If this does not remove most of the stains, the filter should be replaced. To do this, unscrew the filter from the probe. Before putting on a new dust filter, check the alignment of both sensors with the probe. The wires that connect the sensors to the probe are very thin and bend easily. If this happens, correct the alignment by holding the sensor very gently with a pair of small flat nosed pliers. Periodic Calibration Check As long as the probe is operated in an environment free of chemical contaminants and within the operating limits, the long term stability of the humidity sensor is typically better than 1 %RH per year. For maximum accuracy, calibration of the unit may be verified every 6 to 12 months. Applications where the unit is exposed to significant levels of chemical contaminants may require more frequent verifications. The calibration procedure is described in detail in this manual. CALIBRATION CALIBRATION BASICS When calibrating a relative humidity instrument, temperature stability is the single most important requirement. Calibration should be done at room temperature (60 to 80 F or 15 to 26 C) and temperature should be stable to ±0.25 C or better during the period of time required for each calibration point. Do not calibrate close to an air vent or a heater, in direct exposure to sun rays, etc. Calibration Device The ER15 calibration device is a small airtight container that slips over the probe of the H290 and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard (usually an aqueous salt solution). Tip: During calibration, temperature stability may be increased by placing the tip of the probe and calibration device inside an insulating box filled with sand. Humidity Standards The ROTRONIC certified humidity standards permit calibration by non-skilled personnel. These standards are available in boxes of 5 glass ampoules of the same value, which can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions that are precisely titrated at our factory for the right concentration. The 0 %RH humidity standard is made of small granules of a highly porous ceramic that have been dried at a high temperature. A Material Safety Data Sheet is available for each standard. Since humidity standards other than the 0 %RH standard are a salt solution, parts which have come in contact with the liquid should be cleaned after each use.

-11- Calibration Potentiometers and Test Connector TEST CONNECTOR Test Connector: used for a 1-Point Calibration (Requires A2C Calibrator) H35: Offset Adjustment Potentiometer H80: Slope Adjustment Potentiometer HM HM: Potentiometer used for calibration at 5 or 10 %RH. H80 H35 Calibration Procedures One-Point Calibration When the application is over a narrow range of humidity conditions, a one-point calibration may be used to facilitate maintenance and to save time. This consists in adjusting the output of the H290 against the display of a reference instrument (A2C calibrator). The output of the H290 can be read without interrupting the output signal by connecting the test cable of the A2C to the test connector of the H290.(see calibration potentiometers and test connector as well as the manual provided with the A2C). During the procedure, it is important to satisfy the following requirements: The humidity and temperature conditions that are prevalent during calibration must be reasonably stable. Over a period of 15 min., temperature should not vary by more than 1 F (0.5 C). The probe of both instruments should be ventilated or placed in the same stream of moving air. Air velocity must be at least 200 ft/min (1m/s). Usually, it is temperature equilibrium that takes the longest time. Depending on the initial conditions, equilibration can take from a few minutes to as long as 15..20 minutes. If the initial temperature difference between the two probes is more than a few degrees, be sure to wait at least 15 min. before calibrating. When both the A2C calibrator and the H290 provide stable readings, use the H35 potentiometer to adjust the signal from the H290 so as to match the reading of the A2C.

-12- Full Calibration Full calibration of the H290 series requires a 3-point calibration which should be done exactly in the sequence indicated in this manual. The first calibration adjustment should be at 35 %RH or at a value close to that. Slip an ER15 calibration device over the probe and make sure it seals tightly on the probe. The receptacle of the calibration device (or solution holder) should be under the probe tip. Remove the receptacle from the calibration device. Power the H290 and read the output signal (see INSTALLATION). Set the H80 and HM potentiometers in mid position. Place one fiber disc (with the exception of the EA00 standard, each box of ROTRONIC humidity standards includes 5 discs) in the receptacle of the calibration device. The purpose of this disc is to prevent accidental spilling of the solution inside the calibration device or on the humidity sensor. Tap the top of one ampoule of 35 %RH solution so that all liquid drops to the bottom of the ampoule. Snap off top and empty contents on fiber disc. Since the ampoule is made of glass, exercise proper caution (gloves, safety glasses) when snapping off the top. Put the receptacle back on the calibration device and make sure that the solution does not come in contact with the sensor: The solution inside the calibration device should never be on top of the sensors. Allow at least 60 minutes to insure that the calibration device, the solution and the sensor are in a state of equilibrium. This is verified by monitoring the output signal. At equilibrium (stable output signal), adjust the reading of the voltmeter with the H35 potentiometer. Remove the receptacle from the calibration device. Throw away the wet disc (non reusable). Thoroughly wash and dry the receptacle, removing all traces of the humidity standard. Use 80 %RH as the second calibration value as this provides the best overall accuracy over the full range of measurement. Repeat the procedure used for the 35 %RH adjustment with an 80 %RH standard. Allow at least 60 minutes for equilibrium. At equilibrium, adjust the probe output with the H80 potentiometer Remove the receptacle from the calibration device and clean thoroughly. Use 5 or 10 %RH as the last calibration value. Repeat the procedure with a 5 of 10 %RH standard. Allow at least 60 minutes for equilibrium. At equilibrium, adjust the probe output with the HM potentiometer Remove the receptacle from the calibration device and clean thoroughly.

-13- SPECIFICATIONS Humidity Sensor ROTRONIC HYGROMER C94 Operating Temperature at Electronics 20..122 F (-5..50 C) Humidity Measuring Range 0..100 %RH Operating Temperature at Sensors -5 to 320 F (-20 to 160 C) Survival Temperature Limit at Sensors -100 to 390 F (-75 to 200 C) Humidity Limits at Sensors 100 %RH up to 185 F (85 C) 90 %RH at 212 F (100 C) 60 %RH at 260 F (125 C) 25 %RH at 320 F (160 C) Output Signal (specify when ordering) 4..20mA (max. load 500 Ω) 0..5VDC (min. load 1000 Ω) Accuracy at 68..77 F (20..25 C) ± 2%RH from 0 to 100%RH Repeatability ± 0.5%RH Humidity Sensor Stability better than 1%RH over a year Response Time (without filter) 10 seconds Calibration Potentiometers 35, 80 and 10%RH Supply Voltage 10-35 VDC/30 ma or 24 VAC Electrical Connections 3 Wires Terminals 12 AWG Probe Cable Length 6 feet (2 meter), including probe Sensor Protection Slotted Cap with SS wire mesh Probe Dimension/Material 9.84 x 0.59" (250 x 15 mm) / PPS Board Dimensions 3.54 x 2.83 (90 x 72 mm) Weight 0.5 lbs (220g) Probe Holder QMA15 (Order Separately) Calibration Device ER15 (Order Separately) 4 x Mounting Hole Dia. 0.12 / 3 mm 3 x Mounting Hole Dia. 0.12 / 3 mm Component Height: 0.8 / 20.3 mm Clearance under the Board: 0.30 / 7.5 mm 2.28 / 58 mm 1.14 / 29 mm 0.59 / 15 mm 2.83 / 72 mm 1.37 / 35 mm 3.14 / 81 mm 3.54 / 90 mm Modifications Reserved