Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode)

Similar documents
Operating Procedure PerkinElmer LAMBDA 750

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017

Perkin-Elmer Lambda 3B Spectrophotometer

RSA-PE-20. Introduction Unpacking and Inspection... 2

Fluorimeter User s Booklet

HOWTO Make an Initial Connection to the Office Anywhere Desktop using a Windows 7 based PC v1

Standard Operating Procedure Inhouse Solar Simulator

Installation & Setup Guide

Standard Operating Procedure for the Horiba FluroMax-4

Acquiring Data (Through UV WinLab Software) Procedure for Lambda 35 UV2

1) Installing Bluetooth software for Windows (A) Place installation CD into PC and setup should launch automatically.

UV WinLab Software The Lambda Series

TRIAX Series Spectrometers

User s Guide. Ethernet Module for Barcode Printer

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

INSTALLING OR UPGRADING THE DVR S HDD

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

How to Install, Configure and Use sftp (Windows Version)

DEPOSITPARTNER INSTALLATION GUIDE Desktop Application

Light Sources opmaak.indd :32:50

2.2. Facilities Requirements

TCC College WiFi and Printer Setup 07/11/2018 College WiFi and Printer Setup Guide

Agilent ICP-OES SVS 2 Productivity Package Installation and Upgrade Instructions

PerkinElmer Series 200 HPLC: Operating Procedure

Equipment Overview: Safety:

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

OKIFAX 5650 MFP Option Kit. Overview. Setup Guide

Installing TAM Air Assistant Software TM

Sound Card Installation for Windows 95/98

Practice and Review Activities Software

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction..

Installing the Kurzweil 3000 Web License From the WEBSITE

Unpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place.

U3 Encryption Utility User Guide

Network Management Software ALLNET Zone Controller

67 Series Spectrophotometer PC Software

V-600 Operation. JASCO V-600 for Windows Software Manual

RemoteManager Software

TrendView recorders - Installation Instruction

C101-E137 TALK LETTER. Vol. 14

SE800 Short Manual SENTECH Instruments GmbH

UNICO 2100 SERIES SPECTROPHOTOMETER Model 2100 Spectrophotometer Model UV 2100 Spectrophotometer

Spectro-115E TM and Spectro-115U TM Spectroelectrochemical System. Operator s Manual

The two bullets below provide instructions that will guide you through the process of setting up and connecting to the ILG VPN:

UV / VIS SPECTROPHOTOMETER

Spectrometer Visible Light Spectrometer V4.4

Acton Research Corporation SpectraPro Monochromator Control Software for Windows

Throughput of an Optical Instrument II: Physical measurements, Source, Optics. Q4- Number of 500 nm photons per second generated at source

Spectrograph overview:

2 To display the Administration Home page, click the Administration link. (You may need to provide the Administrator password.)

Step 1. Overview. Goal. System Requirements - 1 x WaveAce passive probe. Setup

Contents. - i - Ver.:2

Multi-Live V1.1. For. Wireless Network Camera

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

SIMPLE USER GUIDE RAMAN INSTRUMENT

Operation Manual for OPM-1315 Optical Power Meter

Cary Fiber Optic Probes

IWH3216 Touch/Touch II

M73 Series Network Projector Operation Guide

INSTALLATION GUIDELINES FOR CLIENTS

IS-BX-12 User Manual

EH6000H series EH6108H + /EH6216H + Quick User Guide

Using Respondus with GeorgiaVIEW Vista

Security Management System - Managing Infinova 307X NVR channels. Video display in Security Management System server software

Legal Library. In this guide we will review how to download and use our Legal Library tool. Commercial Brokers Association May 2018

Interlink Express Desktop Printing Service Installation Guide

RMX 2000/4000 MPMx Migration Procedure

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

WUPOS Installation Guide

RMX 2000/4000 MPMx Migration Procedure

Appserv Internal Desktop Access Mac OS Device with Safari Browser. Enter your Appserv username and password to sign in to the Website

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

EntraPass Installation Guide Kantech-OnBoard systems

e-bridge Color Profile Tool Quick Start Guide

CONTENTS. Using the Emperia Scan app 3. Using the Web Portal to access leads 6 FAQ 9

TRUSTGATE USB TOKEN FIRST TIME INSTALLATION MANUAL GUIDE

Contents Overview... 2 Part I Connecting to the VPN via Windows OS Accessing the Site with the View Client Installing...

Novo-Haze TX Transmission Hazemeter

Step 1: Check Contents & Installation Requirements 3. Step 2: Install the Hard Drive 6. Step 3: Setup Wizard 8

Computer Based Examinations

SenCom Jan. 11, 2013 Demonstration on Windows 7 Desktop, Windows Explorer & File Mgmt., and using Thumb/Flash Drives These topics by Mike Smith

Installation Guide. Website: Advanced Card Systems Ltd. Website:

PC Agent Quick Start. Opening the Agent. Logging in to the Agent. Connected Backup. Version 8.6. Revision 0

Documentation. Installing and Maintaining the MeasureNet Laboratory

Installing ProjectSpec from the ProjectMatrix DVD

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro

AnySite. Australia. Install Guide. AnySite v9.1 Install Guide

PLA Series. User s Guide. Quick Start Guide. Powerline Ethernet Adapters. PLA4101, PLA4111, PLA4201, PLA4201 v2, PLA5205, PLA5215, PLA5206, PLA5405

SensorWATCH Basic RH - Wireless Setup Users Help Guide Part Number: A Revision: 1.1.0

Wireless Presentation System User s Manual

Digidoc Icon Camera Installation

RMX 2000/4000 MPMx Migration Procedure

Color Data Software. SpectraMagic DX. Ver E Instruction Manual. Before using this software, please read this manual.

ROC Site Instructions for using Medtronic CODE STAT software

WORKFLOW TRAINING. Reviewing an Agenda Item Through Laserfiche Client. Updated May 2017

FluoroSELECT Fluorometer User s Manual

Multi-NVR Manager. Quick Start Configuration Usage

EntraPass (W10) Installation Guide Kantech-OnBoard systems

AVerMedia IWH3216 Touch Quick Installation Guide

20W Tungsten-Halogen Light Source. Operation Manual

Transcription:

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode) National Centre for Photovoltaic Research and Education (NCPRE) Indian Institute of Technology- Bombay Powai, Mumbai-400076 INDIA

Operating Procedure 1. First Check if the AC (Air Conditioner) unit and lights in the lab is on, if not switch ON. 2. Before switching on the UV/Vis system, open the sample compartment cover and make sure that the beam paths are free from obstruction. This means that no objects project in to the beam paths, no samples are in the sample compartment, and accessories are properly installed. 3. Check if the suitable detector (Integrating mode detector) is placed in the detector compartment. If not follow the Appendix I to change the detector. 4. Check if reflectance standard is placed in sample reflectance port, if not place it. 5. Switch ON the power supply for the system. 6. Switch ON the lamp of the UV/VIS which is located on the top-right of the system.

7. Wait for 20 to 30 minutes after switching on the lamp of the system as stated in step 6. 8. Switch on the PC monitor and CPU 9. Enter the login ID and password disclosed during the training session. 10. Once the desktop is up, click the software icon PerkinElmer UV WinLab to get the page <UV WinLab Explorer> For absorption/transmission measurement: Keep the transmission port open during auto zero process and during your sample measurement put your samples one by one at transmission port without disturbing the sample reflectance port as set earlier and for this follow from Step 11. For Total Reflectance measurement: Keep the transmission port open during auto zero process and also during your sample measurement and put your samples one by one at sample reflectance port by replacing Reflectance standard from sample reflectance port after auto zero process is complete and for this follow from Step 11. For Diffused Reflectance measurement: Keep the transmission port open and also the specular exclusion port open by removing the specular port plug during auto zero process and also during your sample measurement, put your samples one by one at sample reflectance port by replacing Reflectance standard from sample reflectance port after auto zero process is complete and for this follow from Step 11. 11. Goto FILE NEW METHOD to create a new method (if method is not already saved), (Note: if method already available then load your saved method & go to step-18 directly).

12. Once you are in the New Method Wizard, select High Performance UV/VIS NIR Instrument, and click next as shown in above snap shot. 13. Choose the instrument Lambda 950, and click next. 14. Choose the method Scan, and click next 15. Then you will be in the Select Accessories page, click next (no changes to be made) 16. Then click Save/Finish, to get another pop-up page Save Method

17. Enter your method name in the given format <NAME_Sample Name> e.g. Ramesh_SiN/Silicon, and click OK. 18. Now you will be in the UV WinLab Edit- <METHOD NAME> page. Expand all the Folder list and edit them in sequence. 19. Click Data Collection, and edit the method settings as follows: Range Higher wavelength to lower wavelength, eg: From (2500 nm is max limit) to to (250 nm is min limit) Ordinate Mode Select either A (Absorption) or %T (transmission) or %R (reflectance)

(Here note that for Integrating sphere mode the preferred parameters for Slits, PMT is fixed and slit width 4 nm and also the InGaAs in servo mode. Also in detector setting the preferred gain for InGaAs is 9 and response time 0.12 or 0.2 (s)). The typical window snap is shown below:) (N.B. Do not change any other parameter without permission except the rounded regions in the snapshot shown above as suggested above). 20. Goto Correction, edit as per your requirement. The preferred selection is shown in the snap shot below:

21. Goto Sample Info, put the number of samples you have for measurement and press Enter. Now you can name your samples in the Sample ID column. 22. Now Goto Output, Tick the Print to file and Output to file. Here click the setup option next to the Output to File. You will now get a pop-up page Export Data. Please tick all the boxes as shown. Please change the Spectrum Export File(Raw) and Spectrum Export File (Processed) from.sp to.asc Now browse the Data Export file to your destination folder. Your destination folder will usually be My Computer/USER s Documents/<Folder with your name> and press OK.

Also, in report frequency select when task is complete instead of on demand option. 23. Your software is now ready. 24. Load your samples in the compartments as mentioned above whether you want transmittance, reflectance or diffused reflectance measurement. Make sure you sample is fully covering the slit window. 25. Now click Start, which is located on the top- middle of the software window. You will now get a pop-up message Remove Sample and press OK to perform a 100% T/0A correction (Autozero). Click OK to proceed with the experiment. Once the Auto correction is done, replace the samples in the SAMPLE compartment one by one as per asked by popup window. Please do not remove your sample from the compartment unless prompted by the software. After running all your sample, there will be a pop-up message saying All sample have been run, then press OK. 26. Close this page by click the X (located on top-right corner). Now you will be prompted to save your method. Press YES, and then click save in the following Pop-up. Your page will now close. 27. Take your data by writing the data from computer on your CD. 28. Now switch off the PC monitor and CPU, followed by switching OFF the system by pressing the button located on the top-right corner of the system. 29. Now switch off power supply button for your system, and the AC (AC if no other person is in the lab using any other instrument) before you leave. 30. Maintain the cleanliness/tidiness of the area near the instrument.

Appendix I How to change the Detector Module 1. Switch the Lambda spectrometer OFF and disconnect the line power chord. 2. Open the sample and current fixed detector module compartment cover. 3. To remove the current detector, undo the two retaining screw located under the blanking caps on the top of the detector. (The screws are accessible using the long positive screwdriver) The two Screws 4. Remove the detector compartment cover by carefully lifting vertically using the two hand holds. 5. Lift the required detector module which you need by the handles and place the unit into the detector compartment so the positioning pins on the detector compartment floor connect with the corresponding holes in the bottom of the detector module base plate. 6. Secure the detector module to the instrument by tightening the thumb screws at front left and rear locations on the base plate. 7. Reconnect the line power chord. 8. Open the sample compartment cover and make sure that beam paths are free from obstruction. 9. Close the sample compartment and turn ON the spectrometer power switch located at the top right corner of the instrument. 10. Wait until all the instrument initialization is complete before starting the UV Winlab software. (Wait for 20 to 30 minutes after switching on the lamp of the system) 11. Switch on the PC monitor and CPU 12. Enter the login ID and password disclosed during the training session. 13. Once the desktop is up, click the software icon PerkinElmer UV WinLab to get the page <UV WinLab Explorer>

Appendix II The optical system of the lambda 950 spectrometer

Three detector module: Integrating Sphere detector module:

Appendix - III UV-Visible Lambda 950 Quick Summary For STD Detector 175 nm 319.20 nm 860 nm 3300 nm UV region Visible region NIR region D 2 Lamp Tungsten Lamp PMT Detector Slit width Variation Range 0.08 nm-5 nm PbS Detector (Gain range 1-9, 3 preferred) Slit width Variation Range 0.08 nm-20 nm Preferred mode Servo i.e. automatic Detector change over wavelength For Integrating Sphere mode 220 nm 319.20 nm 860 nm 2500 nm UV region Visible region NIR region D 2 Lamp Tungsten Lamp PMT Detector Slit width Variation Range 0.08 nm-5 nm InGaAs Detector (Gain range 1-20, 9 preferred) Preferred mode Servo i.e. automatic Detector change over wavelength