DICOM Conformance Statement, Biim Ultrasound App Version 1

Similar documents
SonoWin Picture Archiving System

Aloka ProSound DICOM

AVIA (Dx MM) DICOM 3.0 Conformance Statement

This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be

Uscan. DICOM Conformance Statement

Topcon Medical Systems

DICOM Conformance Statement for IMAGEnet 5

SonoWin Picture Archiving System

Technical Publications

DICOM Conformance Statement

DICOM Conformance Statement

XPACS DICOM 3.0 Conformance Statement Document Version October 2004

DICOM Conformance Statement for PenFetch

MediaWorkStation. MWS DICOM Push Interface. DICOM Conformance Statement. Version 2.0. English

QPC XSCAN32 DICOM 3.0 Conformance Statement Document Version October 2004

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4

X-Porte DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT. BrainLAB PatXfer 5

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2

nstream Version 3.1 DICOM Conformance Statement

TOSHIBA AMERICA MEDICAL SYSTEMS

REVISION HISTORY... 4 NOTE... 5 INTRODUCTION... 5 DICOM CONFORMANCE STATEMENT...

DICOM Conformance Statement for GALILEI. Software Version V6.0

DICOM Conformance Statement

SCORE 3D Workstation DICOM CONFORMANCE STATEMENT

DEJARNETTE DICOM/FFP RESEARCH SYSTEMS. DICOM Conformance Statement

TITAN DICOM Conformance Statement

DICOM Conformance Statement

DICOM Conformance Statement

DICOM Conformance Statement

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG

Sonovision DICOM Conformance Statement

Punctual Dicom Workstation

DICOM Conformance Statement FORUM

DICOM. Conformance Statement For VINNO Medical Ultrasound System Revision 2.0. Copyright 2012 by VINNO Technology (Suzhou) Co., Ltd.

Kodak Point of Care CR systems. DICOM Conformance Statement

DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number:

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003

DICOM 3.0 Conformance Statement DXIMAGE RIS

Technical Publications

RamSoft. PowerServer. DICOM Conformance Statement

DICOM Conformance Statement. for ES9410 DICOM

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

DxServer for Windows NT DICOM 3.0 Conformance Statement. Document Reference (Référence du document): 00/ Dec28/ABA/MM100/410A

DICOM 3.0 Conformance Statement for PlatinumOne Systems

DICOM Conformance Statement. Forum

DICOM CONFORMANCE STATEMENT

DICOM Conformance Statement

DICOM Conformance Statement FORUM

DxMM/DxWin DICOM 3.0 Conformance Statement. Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B

CADstream DICOM Conformance Statement

DICOM Conformance Statement RadWorks 4.0 Product Line

DICOM CONFORMANCE STATEMENT STORAGE SCU FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-053A or XIDF-056A (MIIXR0001EAB)

FusionXD 2.1 DICOM Conformance Statement

Retinal imaging control software NM. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement

DICOM Conformance Statement June 24, 2011 Version 0.4 ImageWorks Internal document number:

Surgimap. DICOM Conformance Statement. Revision 2.0

KARL STORZ OR1 FUSION V1.3.1 DICOM Conformance Statement

Delta 4. DICOM Conformance Statement D (7)

STaR. DICOM Conformance Statement. September 2009 Version NO COPIES PERMITTED

BRIT Systems. DICOM Conformance Statement

MediaWorkStation. DICOM Worklist Interface. DICOM Conformance Statement. Version 1.1. English

Technical Publications

FusionXD 2.0 DICOM Conformance Statement

DICOM Conformance Statement (DCS) Rev 1.1. Surgimap Nemaris Inc. 475 Park Ave S, 11 th Fl. New York, NY 10016

Technical Publication. DICOM Conformance Statement. Trauma 3.0. Document Revision 2. October 5 th, 2010

Whole Body X-ray CT System Supria. DICOM Conformance Statement

DICOM Conformance Statement

dicom PACS DX-R Version 3.0

KARL STORZ AIDA V1.3.1 DICOM Conformance Statement PRODUCT INFO OR1 OR /2017/PI-E 1/32

DICOM Conformance Statement. for C911 DICOM PRINTER SERIES : C911/C931 DICOM ES9411/ES9431 DICOM Pro9431 DICOM

DICOM CONFORMANCE STATEMENT

Canon DR Image Viewer System Basic DICOM Conformance Statement

Technical Publications

Hologic Physician s Viewer 5.0 DICOM Conformance Statement

DICOM Conformance Statement RadWorks 2.1 Product Line

Dx Server for Windows NT DICOM 3.0 Conformance Statement

DICOM Conformance Statement AIDA HD Connect

DICOM Conformance Statement. report 42. Version 3.2. Reporting Application for Cardiovascular MR and CT

PRO DATA Srl. MicroPrint. DICOM PRINT SCU Conformance Statement (MicroPrint-prnsrv-print-scu.sxw)

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

ETIAM Nexus. DICOM Conformance Statement.

Dx Server for Windows NT DICOM 3.0 Conformance Statement

g GE Ultrasound Technical Publications Vscan Extend (Version 1.0.3) CONFORMANCE STATEMENT for DICOM Direction DOC Revision 3

ISG Technologies Inc. Viewing Stations DICOM Conformance Statement VR 3.1 for Windows Document Number: Revision: 1.

Med X Change DRSHD Digital Recording System High Definition DICOM Conformance Statement Document Version: 1.5

DICOM CONFORMANCE STATEMENT

DICOM 3.0 Conformance Statement. For. DCM v2.93. Revision 1.0 October 23, 2003

DICOM Conformance Statement (DCS) Rev 1.2. Surgimap Nemaris Inc. 475 Park Ave S, 11 th Fl. New York, NY 10016

DICOM CONFORMANCE STATEMENT. BrainLAB PatientBrowser 1.0

SYNAPSE 3D Conformance Statement FUJIFILM SYNAPSE 3D V3.0. Conformance Statement. Revision Z45N

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM. Conformance Statement. HD3 Release Rev A. Koninklijke Philips Electronics N.V All rights are reserved.

Patterson DICOM Imaging. DICOM Conformance Statement

g GE Medical Systems Ultrasound

DICOM V3.0 Conformance Statement. SenoIris 1SP2

DICOM Conformance Statement Application: Linkverse DICOM Provider 2.0

DICOM Conformance Statement. CharruaPACS

Transcription:

Biim Ultrasound, AS Title: Document: DICOM Conformance Statement, Biim Ultrasound App Version 1 D00085 Rev B DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 1 of 10

Table of Contents 1 Purpose and Scope... 3 2 Conformance statement overview... 3 3 Definitions... 3 4 Reference Documents... 4 5 Networking... 4 5.1 Implementation Model... 4 5.1.1 Application Data Flow... 4 5.1.2 Functional Definition of AEs... 5 5.1.3 Sequencing of Real-World Activities... 5 6 AE Specifications... 5 6.1 Store AE Specification... 5 6.1.1 SOP Classes... 5 6.1.2 Association Establishment Policies... 5 7 Communication... 6 7.1 TCP/IP Stack... 6 8 Configuration... 7 8.1 Local Device Settings... 7 8.2 Remote Device Settings... 7 9 Support of Character Sets... 7 10 Security... 7 10.1 Secure Connection Profile... 7 11 Information Object Definition... 8 11.1 Patient Module of Created SOP Instances... 8 11.2 Study Module of Created SOP Instances... 8 11.3 General Series Module of Created SOP Instances... 8 11.4 Equipment Module of Created SOP Instances... 9 11.5 Image Pixel Module of Created SOP Instances... 9 11.6 Image Pixel Module of Created SOP Instances... 10 DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 2 of 10

1 PURPOSE AND SCOPE The purpose of this document is to provide the DICOM conformance statement for the Biim Ultrasound App Version 1. The document assumes that the reader has a basic understanding of DICOM standard and functionality. 2 CONFORMANCE STATEMENT OVERVIEW The Biim Ultrasound App Version 1 implements the necessary DICOM services to save acquired ultrasound images to the digital image archives. The Biim Ultrasound App Version 1 product is based on NEMA DICOM version number 3.0 2007. Table 1 Networking Services provides an overview of the networking services supported by the Biim Ultrasound App Version 1. Table 1 Networking Services Networking SOP Classes User of Service (SCU) Provider of Service (SCP) Transfer Ultrasound Image Storage Yes No General Verification Yes No 3 DEFINITIONS Table 2 contains definition of terms used or referenced in this document. Table 2. Definitions and Acronyms Term Definition AE DICOM Application Entity AES AET DICOM FDA Advanced Encryption Standard Application Entity Title Digital Imaging and Communications in Medicine Food and Drug Administration DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 3 of 10

Term IEC JPEG NEMA SCP SCU SOP TLS UID US VR Definition International Electrotechnical Commission Joint Photographic Experts Group National Electrical Manufacturers Association DICOM Service Class Provider (server) DICOM Service Class User (client) Service Object Pair Transport Layer Security Unique Identifier Ultrasound Value Representation 4 REFERENCE DOCUMENTS NEMA DICOM Standard Version 3.0 2007. 5 NETWORKING 5.1 Implementation Model 5.1.1 Application Data Flow Figure 1 Application Data Flow Diagram describes application data flow when sending ultrasound images. Storage AE sends images to a single remote AE. Each set of images will be sent separately using their own association to Remote AE. Images will have been captured and saved by a user function like Save Image. Sending of the images will have been initiated by a user activity Send to DICOM. There is a standard DICOM interface between Storage AE and Remote AE. Figure 1 Application Data Flow Diagram DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 4 of 10

5.1.2 Functional Definition of AEs 5.1.2.1 Functional Definition of the Storage Application Entity The Biim Ultrasound App Version 1 sends images to the Remote AE by opening an association first. Then it sends the images in the loop. The user can select multiple images and the Biim Ultrasound App Version 1 loops through all images selected for sending. Then the Biim Ultrasound App Version 1 closes the association. 5.1.3 Sequencing of Real-World Activities The following steps are repeated for sending images: A-Associate C-ECHO command C-STORE Image instance (loop through all images) A-Release The Biim Ultrasound App Version 1 offers the possibility to check the DICOM connection by using the verify application entity procedure. It has the following steps: A-Associate C-ECHO command A-Release 6 AE SPECIFICATIONS 6.1 Store AE Specification 6.1.1 SOP Classes The Store AE provides conformance to the DICOM V3.0 SOP Classes that have been listed in Table 3 Store AE SOP Class Support. Table 3 Store AE SOP Class Support SOP Class Name SOP Class UID Conformance Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Standard Verification 1.2.840.10008.1.1 Standard 6.1.2 Association Establishment Policies 6.1.2.1 General The Biim Ultrasound App Version 1 always proposes a standard application context name for DICOM 3.0: DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 5 of 10

Table 4 DICOM Application Context for AE Storage Application Context Name 1.2.840.10008.3.1.1.1 Maximum PDU size offered to SCP: 16 384 bytes 6.1.2.2 Number of Associations The Biim Ultrasound App Version 1 opens always one Association to the destination when a connection is requested. Table 5 Maximum number of Association for AE Storage Maximum number of simultaneous associations 1 6.1.2.3 Asynchronous Nature The Biim Ultrasound App Version 1 uses synchronous communication. It has only one transaction ongoing simultaneously. 6.1.2.4 Implementation Identifying Nature Implementation Class UID 1.2.276.0.7230010.3.0.3.6.1 Implementation Version Name biim L 12-4 6.1.2.5 Proposed Presentation Contexts Table 6 Proposed Presentation Contexts for Store and Verification Presentation Context Table Abstract Syntax Transfer Syntax Role Extended Negotiation Name UID Name List UID List Verification 1.2.840.100 08.1.1 Ultrasound Image Storage 1.2.840.100 08.5.1.4.1.1. 6.1 7 COMMUNICATION 7.1 TCP/IP Stack Implicit VR Little Endian 1.2.840.10008.1.2 Implicit VR Little Endian Explicit VR Little Endian JPEG Baseline 1.2.840.10008.1.2 1.2.840.10008.1.2.1 1.2.840.10008.1.2.4.50 SCU SCU None None The Biim Ultrasound App Version 1 product uses TCP/IP communication that is supported by the underlying operating system. The system supports all physical interfaces that are available in the device. The Biim Ultrasound App Version 1 does not configure or control the network interfaces. DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 6 of 10

8 CONFIGURATION The Biim Ultrasound App Version 1 allows the user to update the DICOM configuration at any time. The user can save different sets with defined names representing each configuration set. Each set contains local and remote device parameters. 8.1 Local Device Settings Local AE Title can be provided by the user. 8.2 Remote Device Settings The following parameters can be defined: - Remote AE Title - Remote server address - Remote port - Secure connection enabled/disabled 9 SUPPORT OF CHARACTER SETS The Biim Ultrasound App Version 1 system supports the ISO 8859 Latin 1 (ISO-IR 100) character set family. 10 SECURITY 10.1 Secure Connection Profile The Biim Ultrasound App Version 1 supports secure DICOM connection. In this case the association shall be opened as a TLS secured connection. The supported profile is the AES TLS Secure Transport Connection Profile. Table 7 TLS Features describes TLS features supported by Biim Ultrasound App Version 1. Table 7 TLS Features Supported TLS Feature Entity Authentication Exchange of Master Secrets Data Integrity Privacy Mechanism RSA based certificates RSA SHA DES CBC3 SHA The Biim Ultrasound App Version 1 doesn t verify server certificates against a certificate authority. Table 8 Integrity Check Failure Handling describes the system s response to failures. Table 8 Integrity Check Failure Handling Service Status Meaning AE Behavior A-P-ABORT Integrity check has failed during TLS transport Association terminated. User notified. DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 7 of 10

11 INFORMATION OBJECT DEFINITION 11.1 Patient Module of Created SOP Instances Table 9 Patient Module Patient s Name (0010,0010) Manually entered patient last name, first name and middle name. Patient ID (0010,0020) Manually entered or auto generated patient identifier. Patient s Birth Date (0010,0030) Manually entered date of birth. Value may be empty Patient s Sex (0010,0040) Manually entered patient gender. 11.2 Study Module of Created SOP Instances Table 10 Study Module Study Instance UID (0020,000D) Auto generated Study Date (0008,0020) Study date Study Time (0008,0030) Study time Study ID (0020,0010) Automatically generated value starting from 1 Study Description (0008,1030) Entered manually as an Exam Type Value may be empty Accession number (0008,0050) Manually entered value Value may be empty Referring Physician s Name (0008,0090) Manually entered clinician s name Value may be empty 11.3 General Series Module of Created SOP Instances Table 11 General Series Module Modality (0008,0060) US Series Instance UID (0020,000E) Auto generated Series Number (0020,0011) A unique number within the Study, value is 1. Laterality (0020,0060) Zero length Value is empty DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 8 of 10

Series Date (0008,0021) Procedure date Series Time (0008,0031) Procedure time Performing Physician Name (0008,1050) Manually entered clinician s name Value may be empty Protocol Name (0018,1030) Zero length Value is empty Series Description (0008,103E) Entered manually as an Exam Type Operator s Name (0008,1070) Zero length Value is empty 11.4 Equipment Module of Created SOP Instances Table 12 Equipment Module Manufacturer (0008,0070) Biim Ultrasound Station Name (0008,1010) The local AE title of biim L 12-4 device. The user can configure this text manually. Manufacturer s model name (0008,1090) biim L 12-4 Software Version (0018,1020) This is multi valued tag that contains SW part number, SW version and build number. Device Serial Number Value may be empty (0018,1000) Value is 0 11.5 Image Pixel Module of Created SOP Instances Table 13 Image Module Instance Number (0020,0013) Auto generated Patient Orientation (0020,0020) Zero length Value is empty Content Date (0008,0023) Procedure date Content Time (0008,0033) Procedure time Lossy Image Compression (0008,2110) 01 = Lossy Compressed DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 9 of 10

11.6 Image Pixel Module of Created SOP Instances Table 14 Image Pixel Module Samples Per Pixel (0028,0002) Value is 3 Photometric Interpretation (0028,0004) Value is YBR_FULL_422 Rows (0028,0010) Value varies depending on tablet or smartphone screen resolution Columns (0028,0011) Value varies depending on tablet or smartphone screen resolution Bits Allocated (0028,0100) Value is 8 Bits Stored (0028,0101) Value is 8 High Bit (0028,0102) Value is 7 Pixel Representation (0028,0103) Value is 0 Lossy Image Compression Method (0028,2114) Value is ISO_10918_1 Planar Configuration (0028,0006) Value is 0 Pixel Data (7FE0,0010) DICOM Conformance Statement, Biim Ultrasound App Version 1 Page 10 of 10