Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

Similar documents
Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

An investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application

Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

Monte Carlo evaluation of the dosimetric uncertainty in matched 6 MV Elekta and Varian linear accelerators

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Improvements in Monte Carlo simulation of large electron fields

87.55.K-, Gh, km, Ca. Agility, analytic model, Elekta, Monaco, Monte Carlo, virtual source

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)

Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system

Comparison of measured and Monte Carlo calculated electron beam central axis depth dose in water

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam

A virtual photon energy fluence model for Monte Carlo dose calculation

Monte Carlo Methods for Accelerator Simulation and Photon Beam Modeling

Evaluation of latent variances in Monte Carlo dose calculations with Varian TrueBeam photon phase-spaces used as a particle source

A Geant4 based treatment plan verification tool: commissioning of the LINAC model and DICOM-RT interface

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Radiation Beam Characterization and Dosimetry of Theratron Equinox-80 Telecobalt Machine Using BEAMnrc Monte Carlo Simulation Code

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements

Monte Carlo Treatment Planning: Implementation of Clinical Systems

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital

Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode

BRAZILIAN JOURNAL OF RADIATION SCIENCES (2018) 01-14

FAST, precise. qa software

Photon beam dose distributions in 2D

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014

Verification measurements of an emc algorithm using a 2D ion chamber array

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems

Preface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.

Michael Speiser, Ph.D.

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Implementation of Monte Carlo Dose calculation for CyberKnife treatment planning

UNIVERSITY OF CALGARY. Monte Carlo model of the Brainlab Novalis Classic 6 MV linac using the GATE simulation. platform. Jared K.

2D DOSE MEASUREMENT USING A FLAT PANEL EPID

Transitioning from pencil beam to Monte Carlo for electron dose calculations

Basics of treatment planning II

UNCOMPROMISING QUALITY

Basic Radiation Oncology Physics

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Monte Carlo investigation of collapsed versus rotated IMRT plan verification

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Dosimetric characteristics of LinaTech DMLC H multi leaf collimator: Monte Carlo simulation and experimental study

Application of efficient Monte Carlo photon beam simulations to dose calculations in voxellized human phantoms

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

AN ABSTRACT OF THE DISSERTATION OF

Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo Code

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006

Suitability Study of MCNP Monte Carlo Program for Use in Medical Physics

MPEXS benchmark results

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013

Clinical implementation of photon beam flatness measurements to verify beam quality

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations

On compensator design for photon beam intensity-modulated conformal therapy

EXTERNAL PHOTON BEAMS: PHYSICAL ASPECTS

Monte Carlo simulations. Lesson FYSKJM4710 Eirik Malinen

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Monte Carlo simulations

Monte Carlo Simulations of. Chemical Vapour Deposition. Diamond Detectors

BEAMDP Users Manual. C.-M. Ma and D.W.O. Rogers Ionizing Radiation Standards National Research Council of Canada, Ottawa

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Shielding factors for traditional safety glasses

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

2007 Institute of Physics Publishing. Reprinted with permission.

A software tool for the quantitative evaluation of 3D dose calculation algorithms

Comparison of Predictions by MCNP and EGSnrc of Radiation Dose

I Introduction 2. IV Relative dose in electron and photon beams 26 IV.A Dose and kerma per unit incident fluence... 27

A SYSTEM OF DOSIMETRIC CALCULATIONS

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

VALIDATION OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR CLINICAL ELECTRON BEAMS IN THE PRESENCE OF PHANTOMS WITH COMPLEX HETEROGENEITIES

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved

Interface Dosimetry for Electronic Brachytherapy Intracavitary Breast Balloon Applicators

The importance of accurate linear accelerator head modelling for IMRT Monte Carlo

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus

Commissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free beams

Geant4 in Brachytherapy

Transcription:

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Danial Seifi Makrani 1, Hadi Hasanzadeh 2*, Tayyeb Allahverdi Pourfallah 3, Arash Ghasemi 4, Majid Jadidi 1, Hamed Babapour 5 1 Department of Medical Physics, Faculty of Medical Sciences, Semnan University of Medical Sciences, Semnan, Iran 2 Cancer Research Center, Department of Medical Physics, Semnan University of Medical Sciences, Semnan, Iran 3 Department of Biochemistry and Biophysics, Faculty of Medicine, Mazandaran University of Medical Sciences, Sari, Iran 4 Department of Radiology and radiation oncology, Faculty of Medicine, Mazandaran University of Medical Sciences, Sari, Iran 5 Guilan Oncology center (GOC), Rasht, Iran *Corresponding Author: email address: Hasanzadeh.h@semums.ac.ir (H. Hasanzadeh) ABSTRACT The Monte Carlo method can be used to describe any technique that approximates solutions to quantitative problems through statistical sampling. This method is considered to be the most accurate method for dose calculation in radiotherapy. For complete modeling of a linear accelerator, it is required that the manufactured information covers all data, but some data such as primary electron energy must be indicated. The purpose of this study was to determine the best primary electron energy for 15 MV photon beam with varying the energy and FWHM. A Monte Carlo model for photon-beam output of a Siemens primus linear accelerator was validated by plotting the energy spectrum of photon beam and calculating the percentage depth dose (PDD) and beam profiles for 10 10 cm 2 field. Square 10 10 cm 2 field was validated by measurements in water by a farmer chamber. Linac head simulation was performed with BEAMnrc and dose calculation and 3D dose file were produced by DOSXYZnrc. The results were analyzed using MATLAB. It was found a good agreement between calculated PDD and beam profile for 15 MV photon beam using Monte Carlo simulation with primary electron energy of 11 MV and FWHM of 0.4 with maximum dose difference of 1.2% in PDD curves. In conclusion, using primary electron energy of 11 MV and FWHM of 0.4 has very good accuracy in calculating of dose distribution for 15 MV photon beam and it can be considered as a promising method for patient dose calculations. Key words: BEAMnrc; Monte Carlo; Linac simulation; Beam Parameters; FWHM. INTRODUCTION Radiation therapy uses high-energy radiation to shrink tumors and kill cancer cells [1]. The main goal of radiation therapy is to deliver the highest dose to the tumor while maintaining minimum dose to the surrounding healthy tissues. To do so, the dose distribution must be computed and verified with an accurate method. Monte Carlo method is widely accepted as the most accurate method for modeling radiotherapy treatments [2-4] and has become more accessible due to technological advances in computer systems. To run Monte Carlo (MC) code for clinical applications, it is essential to define all the 75 characteristics of photon beam such as: initial electron energy and full width of half maximum (FWHM) of the intensity distribution of these primary electrons that stimulate the target of linac. The accelerated primary electron beam starts from the flight tube dominating an angular and spatial distribution, subsequently. This electron beam punches the high-z metal target resulting to production of bremsstrahlung photons. It is notable that in Monte Carlo simulation of clinical linacs, usually no electron beam modeling is performed preceding to exiting the flight tube; so numerous works have been performed on determination of primary electron

energy in radiotherapy accelerators [5-8]. Simulating the linac head with Monte Carlo codes is the most accurate and detailed method to obtain the influence of different parameters on dose distribution [9]. The purpose of this work was to compute the best primary electron energy for 15 MV photon beam linac (Siemens primus, USA) varying the energy and FWHM using Monte Carlo method. The BEAMnrc and DOSXYZnrc codes were used to model 15 MV Primus linac head and measure the PDD and beam profile in the modeled water phantom. The data were validated by measurements in water phantom using a farmer chamber. MATERIALS AND METHODS In this study, we explored full width at half maximum (FWHM) of the intensity distribution of primary electrons from the target of Siemens Primus linac stimulated by BEAMnrc and DOSXYZnrc. Both programs are based on electron gamma shower user code (EGSnrc) that come as a package under license of the National Research Council of Canada (NRC) [10]. All the materials and the dimensions for the Linear accelerator head were built based on manufacturer s specification datasheet provided from Siemens Healthcare Company, USA. The primus accelerator components are shown in Figure 1, including the exit window, target, primary collimator, flattening filter, monitor chamber, Y jaws and MLC. PEGS4 (EGS preprocessor) cross-section data for the specific materials in the accelerator were from 700 ICRU PEGS4data file. In simulation of radiation transport using MC methods, the history of a particle is specified as a sequence of tracks where each track ends with an interaction event where the particle can change its direction and lose energy. The history ends when it leaves the region of interest or when its energy is lower than the predefined cutoff energy [11]. In this study, the number of histories for Monte Carlo calculations was 5 10 8 particles, resulting from 10 8 particles in a phase space after the primus linac head. This was done to ensure reliable statistics in the phase space file generated by the BEAMnrc simulation [12]. The number of the primary electrons that strike the target on top of the linac head is similar to the number of history. The global cut off energies used in the simulations were 700 KeV for electron cutoff energy (ECUT) and 10 KeV for photon global cutoff (PCUT) [13]. Monte Carlo simulations were performed for monoenergetic beams ranging from 11 to 15 MeV and FWHM varied from 0.3 cm to 0.4 cm for 15 MV beam. Figure 1. Simulated linear accelerator head (XY view) The primary output of the BEAMnrc simulation for the head of linac is a file called phase space. The phase space contains information such as energy, position, direction, etc. of millions of particles (photons, electrons, positrons). From this space, beam quality factors including photon and electron spectra and two-dimensional energy distribution was obtained to analyze beam production mechanism. This phase space was scored in a plane upright to the beam axis at 100 cm distance from the target. 76

For PDD and dose profiles measurements at the maximum dose depth in square fields, farmer chamber was used in the water tank for 15 MV X ray beam. The water phantom was created using DOSXYZnrc code with voxel size of 0.5 0.5 0.2 cm 3 at source to surface distance (SSD) of 100 cm. The simulated PDD and beam profile were compared with that obtained by the farmer chamber. The primary electron parameters such as energy and FWHM were changed to reduce difference between calculated and measured values to less than 2%. RESULTS We evaluated four different modes to achieve the accurate PDD values and profiles. All PDD and profiles were calculated at different conditions (including energy and FWHM) using Monte Carlo method and were compared to experimental values using ion chamber. Figure 2 shows the PDD obtained from simulation for the field size of 10 10 cm 2 in 15 MV beam. The solid line shows the PDD obtained from Monte Carlo simulation and the dots represent the measured data points using ion chamber. Figure 2(a, b and d) illustrate the discrepancy between the calculated and measured values. To reduce this discrepancy, the initial energy of the primary electron beam was decreased and finally after several probations, the value at which the simulation and measurements matched was 11 MeV (Figure 2(c)). The agreement between calculated results by Monte Carlo modeling and direct measurement was obtained to be 1.2%. (a) (b) (c) (d) Figure 2. Percentage depth dose at 15 MV beam with a field size of 10 10cm 2 ; (a) 13 MeV (b) 15 MeV (c) 11 MeV (d) 12 MeV 77

Figure 3. Calculated beam profile at 15 MV. (green) 11 MeV and FWHM of 0.3, (blue) 11MeV and FWHM of 0.4 and (red) the experimental data As anticipated, with an electron beam with primary energy of 11 MeV and FWHM of 0.4, there exists a good agreement between 15 MV dose profile simulated with Monte Carlo modeling and those measured with ionization chamber (Figure 3). DISCUSSION Determination of electron beam parameters is an essential part of simulating radiation transport using Monte Carlo methods. This helps one to find out some real parameters of a beam with nominal energy including real energy and FWHM of the exit beam. Javier Pena et al. [6] presented a method for commissioning photon beams that employs depth doses and lateral profiles. By simultaneous comparison of measurements and simulations for several mean energy/radial FWHM combinations, one is able to determine the values that yield best matching. In this study, PDDs were obtained from simulation for the field size of 10 10 cm 2 in 15 MV beam with an electron energy of 11, 12, 13 and 15 MeV with 78 FWHM of 0.3 and 0.4 for a Siemens primus linac. The difference between the calculated and measured doses for both energies were about 1.2%, well below 2% which was chosen as a standard to set the useful results for modeling linac [8]. At 15 MV beam, a good agreement was obtained in 11 MeV primary electron energy with 0.4 FWHM of the intensity distribution. In this study, the significant components of primary electron beam in final results were obtained and it was shown that a small change in electron beam properties has strong effects on deposited dose in the water phantom. A Monte Carlo simulation of a Siemens primus linac was performed the results of which will be used for future studies. The results of phase space can be used for MLC leakage and calculation of scattering due to them. Jabbari et al. [8] found out a good agreement for 6 MV beam in 6.5 MeV primary electron energy with 0.31 FWHM of the intensity distribution and at 18 MV in 15 MeV primary electron energy with 0.29 FWHM of the intensity distribution.

CONCLUSION The percentage depth dose (PDD) and beam profile were calculated using Monte Carlo simulation and compared to the measurement performed by a farmer chamber in water phantom. Respectable agreement between the calculated PDD and beam profile using MC simulation with that calculated using chamber was detected. The results showed that 11 MeV primary electron energy with FWHM of 0.4 has very good agreement in calculating dose REFERENCES 1. Lawrence T, Ten Haken R, Giaccia A. Principles of Radiation Oncology. In: DeVita V, Lawrence T, Rosenberg S, editors. Cancer: Principles and Practice of Oncology. 8 th ed. Philadelphia: Lippincott Williams and Wilkins; 2008. 2. Rogers D. Fifty years of Monte Carlo simulations for medical physics. Phys Med Biol. 2006; 51 (13): R287. 3. Rogers D, Walters B, Kawrakow I. BEAMnrc users manual. NRC Report PIRS. 2001; 509. 4. Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. Med Phys. 2000; 27 (3): 485-98. 5. Siantar CH, Walling R, Daly T, Faddegon B, Albright N, Bergstrom P, et al. Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom. Med Phys. 2001; 28 (7): 1322-37. 6. Pena J, González-Castaño DM, Gómez F, Sánchez-Doblado F, Hartmann GH. Automatic determination of primary electron beam parameters in Monte Carlo simulation. Med Phys. 2007; 34 (3): 1076-84. 7. Sheikh-Bagheri D, Rogers D. Sensitivity of megavoltage photon beam Monte Carlo distribution for 15 MV photon beam. A Monte Carlo model of primus linear accelerator built in this study can be used to calculate dose distribution in physical phantoms. ACKNOWLEDGMENTS This work was a part of Mr. Seifi thesis for the degree of M.Sc. in Medical Physics supported by deputy of research and technology of Semnan University of Medical Sciences. simulations to electron beam and other parameters. Med Phys. 2002; 29 (3): 379-90. 8. Jabbari K, Anvar HS, Tavakoli MB, Amouheidari A. Monte carlo simulation of siemens oncor linear accelerator with beamnrc and dosxyznrc code. J Med Signals Sens. 2013; 3 (3): 172. 9. Chetty IJ, Curran B, Cygler JE, DeMarco JJ, Ezzell G, Faddegon BA, et al. Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlobased photon and electron external beam treatment planning. Med Phys. 2007; 34 (12): 4818-53. 10. Kawrakow I, Rogers D. The EGSnrc code system. NRC Report PIRS-701, NRC, Ottawa. 2000. 11. Oliveira AC, Vieira JW, Santana MG, Lima FR. Monte Carlo simulation of a medical linear accelerator for generation of phase spaces. INAC; 2013; Recife, PE,Brazil. p. 1-15. 12. Sawkey D, Faddegon B. Simulation of large x-ray fields using independently measured source and geometry details. Med Phys. 2009; 36 (12): 5622-32. 13. Ceberg S, Gagne I, Gustafsson H, Scherman JB, Korreman SS, Kjær-Kristoffersen F, et al. RapidArc treatment verification in 3D using polymer gel dosimetry and Monte Carlo simulation. Phys Med Biol. 2010; 55 (17): 4885. 79