A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

Similar documents
Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines

Significance of time-dependent geometries for Monte Carlo simulations in radiation therapy. Harald Paganetti

Determination of rotations in three dimensions using two-dimensional portal image registration

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

ROBUST OPTIMIZATION THE END OF PTV AND THE BEGINNING OF SMART DOSE CLOUD. Moe Siddiqui, April 08, 2017

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Estimating 3D Respiratory Motion from Orbiting Views

The MSKCC Approach to IMRT. Outline

Image Guidance and Beam Level Imaging in Digital Linacs

Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion

Calibration and quality assurance for rounded leaf-end MLC systems

A software tool for the quantitative evaluation of 3D dose calculation algorithms

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

Brilliance CT Big Bore.

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Using a research real-time control interface to go beyond dynamic MLC tracking

2005 IEEE Nuclear Science Symposium Conference Record M10-2

Basic Radiation Oncology Physics

Dynalog data tool for IMRT plan verification

IMRT site-specific procedure: Prostate (CHHiP)

Tumor motion during liver SBRT

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved

Coverage based treatment planning to accommodate organ deformable motions and contouring uncertainties for prostate treatment. Huijun Xu, Ph.D.

A finite state model for respiratory motion analysis in image guided radiation therapy

Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: Implications for 4D radiotherapy

Iterative regularization in intensity-modulated radiation therapy optimization. Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006.

Fast dose algorithm for generation of dose coverage probability for robustness analysis of fractionated radiotherapy

Motion artifact detection in four-dimensional computed tomography images

Development of a deformable lung phantom for the evaluation of deformable registration

Chapter 9 Field Shaping: Scanning Beam

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Reconstruction of 4D CT data sets acquired during free breathing for the analysis of respiratory motion

Monaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application

Michael Speiser, Ph.D.

How would, or how does, the patient position (chin extended) affect your beam arrangement?

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Proton dose calculation algorithms and configuration data

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Imad Ali 1 Nesreen Alsbou 2 Justin Jaskowiak 1 Salahuddin Ahmad 1. Abstract RADIATION ONCOLOGY PHYSICS

Dosimetric impact of the 160 MLC on head and neck IMRT treatments

Quantitative imaging for clinical dosimetry

Image-based Monte Carlo calculations for dosimetry

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

NEW METHOD OF COLLECTING OUTPUT FACTORS FOR COMMISSIONING LINEAR ACCELERATORS WITH SPECIAL EMPHASIS

On compensator design for photon beam intensity-modulated conformal therapy

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

An automated method for comparing motion artifacts in cine four-dimensional computed tomography images

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

State-of-the-Art IGRT

FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Effects of the difference in tube voltage of the CT scanner on. dose calculation

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

Analysis of 4D CT cine images for the characterization of organ motion due to breathing

Radiation therapy treatment plan optimization

Robust Lung Ventilation Assessment

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

Evaluation of latent variances in Monte Carlo dose calculations with Varian TrueBeam photon phase-spaces used as a particle source

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

FAST, precise. qa software

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

Dose Calculation and Optimization Algorithms: A Clinical Perspective

15 Dose Calculation Algorithms

Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

CHAPTER 9 INFLUENCE OF SMOOTHING ALGORITHMS IN MONTE CARLO DOSE CALCULATIONS OF CYBERKNIFE TREATMENT PLANS: A LUNG PHANTOM STUDY

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

BOSTON UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES. Thesis REAL-TIME 4D TUMOR TRACKING AND MODELING FROM INTERNAL

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI

A design of a DICOM-RT-based tool box for nonrigid 4D dose calculation

Monte Carlo verification of IMRT dose distributions from a commercial treatment planning

Basics of treatment planning II

Takayama, Kenji; Mizowaki, Takashi; Nagano, Kazuo; Kamino, Yuichiro; Hi. Rightc 2009 Elsevier Ireland Ltd.

Transcription:

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel L. McShan, James M. Balter, Benedick A. Fraass, and Randall K. Ten Haken Department of Radiation Oncology, The University of Michigan, Ann Arbor, Michigan 48109-0010 Received 18 February 2003; accepted for publication 22 April 2003; published 25 June 2003 We describe the implementation of a fluence convolution method to account for the influence of superior-inferior SI respiratory induced motion on a Monte Carlo-based dose calculation of a tumor located in the liver. This method involves convolving the static fluence map with a function describing the SI motion of the liver the motion function has been previously derived from measurements of diaphragm movement observed under fluoroscopy. Significant differences are noted between fluence-convolved and static dose distributions in an example clinical treatment plan; hot and cold spots on the order of 25% are observed in the fluence-convolved plan at the superior and inferior borders of the liver, respectively. This study illustrates that the fluence convolution method can be incorporated into Monte Carlo dose calculation algorithms to account for some of the effects of patient breathing during radiotherapy treatment planning, thus leading to more accurate dose calculations. 2003 American Association of Physicists in Medicine. DOI: 10.1118/1.1581412 The standard computation of dose distributions in conformal radiotherapy is based on a single instance of patient anatomy. However, as a consequence of patient breathing and the associated changes in organ anatomy, differences will result between the planned dose distribution and that actually delivered. Therefore, the incorporation of uncertainties due to organ motion resulting from respiration is a very important requirement for accurate dose calculation within a given treatment plan. 1 5 One particular method to account for respiratory-induced organ motion is to convolve the static dose distributions with functions that approximate the breathing. 4,5 A study by Lujan et al. 4 has shown that applying a single convolution to the static dose distribution, for patients undergoing irradiation of the liver, is sufficient to predict the dose distribution for the given treatment plan; intrafraction effects were found to average out over the course of many fractions. A potential limitation of the dose convolution approach is that it is based on the assumption that the dose distribution is spatially invariant, i.e., the convolution of dose is conducted in an assumed homogeneous medium without cognizance of the variation in patient tissue densities. In a recent study, Beckham et al., 6 have applied a fluence-convolution method to incorporate random setup error in the dose calculation. Specifically, Beckham et al. 6 use the Monte Carlo method to sample the photon fluence that is first convolved with a 2D Gaussian random setup error kernel. Beckham et al. 6 point out that fluence convolution is more accurate than dose convolution in heterogeneous media since fluence convolution correctly models the spatial variation of dose resulting from movement due to setup error. In this study, we extend the concept of fluence convolution to account for respiratory motion in a clinically realistic treatment plan for a lesion located within the liver. We compare calculations using the fluence convolution method with those from the static no motion case, as well as with a treatment plan generated using the dose convolution method. The treatment planning study has been conducted using the UMPlan University of Michigan, Ann Arbor, MI treatment planning software. Dose calculations for both fluence and dose-based convolution have been performed using the D ose P lanning M ethod DPM Monte Carlo code system, 7,8 which has been integrated within UMPlan. The Monte Carlo virtual source model used for patient-specific dose calculations is similar to that developed by Chetty et al., 9 and is described at length in that paper. However, a brief description of the source model is necessary here. The treatment head components of a Varian 21EX linear accelerator Varian Associates, Palo Alto, CA were simulated using the BEAMnrc Monte Carlo code CNRC, Ottawa, CN. A virtual source model was reconstructed from the phase space distribution to sample the source particle s position, energy, and direction. Arbitrary field shapes are simulated by multiplying the uncollimated fluence map by a matrix describing the multileaf collimator MLC configuration the composite fluence matrix includes a correction for leaf edge penumbral effects. 9 The fluence convolution method to correct for respiratoryinduced motion of the liver was implemented by convolving the static fluence distribution, static, with a probability distribution function associated with breathing induced diaphragmatic displacement, labeled F motion in this paper, to generate a composite fluence map, motion, which is used to sample the source particle s position. static represents the MLC-shaped field static fluence distribution and is located in a plane perpendicular to the beam central axis, between the accelerator and the patient. For a point, r, that includes the motion, we then have motion r F motion static r F motion r r static r dr. 1 1776 Med. Phys. 30 7, July 2003 0094-2405Õ2003Õ30 7 Õ1776Õ5Õ$20.00 2003 Am. Assoc. Phys. Med. 1776

1777 Chetty et al.: Monte Carlo fluence convolution respiratory motion liver 1777 FIG. 1. Percentage dose difference map fluence convolved dose static dose in the sagittal view. Both treatment plans have been normalized to 100% at the isocenter. Included are the isodose lines corresponding to the 20% differences solid yellow lines and the 10% differences dashed yellow lines. Nonhighlighted differences dark regions are within 2%. F motion has been developed by Lujan et al., 4 based on the correlation between the liver and the diaphragm movements as observed under fluoroscopy by Balter et al. 10 and is given by F motion z na z 0 z a 1 z 0 z a (2n 1)/2n 1/n for z 0 a z z 0, 1/2 1 where z 0 is the position at exhale, a the amplitude of motion patient specific, 1.5 cm in our study, and n 3, a parameter that describes the general shape of the model. In the Monte Carlo implementation, F motion is divided into 15 equally spaced probability bins, from the position at exhale (z 0 )to the position at inhale (z 0 a). The position (x, y at a fixed z location and energy for each particle starting from the virtual source is determined by first sampling static. The source particle s incident direction is determined from the position coordinates assuming that the particle originated from a point, 9 i.e., u x/r, v y/r, w z/r, where R x 2 y 2 z 2. To account for the motion in the SI direction, F motion is sampled to determine the positional translation, z; the particle s z coordinate is then translated according to the relation: z z z. Note that the particle s x and y coordinates remain unchanged as F motion incorporates motion in only the z SI dimension. The following relation may then be used to describe the coordinate transformation from the static fluence distribution static, in the unprimed coordinates, to the motion convolved fluence, motion, in the primed coordinates: x x, y y, and z z z. 3 2 The direction cosine vectors for each source particle are also appropriately modified as there is a change in R from the translation of z, u x R, y v R, and w z R, where R x 2 y 2 z 2. The fluence convolution method described here is different from that of Beckham et al., 6 where the x, y position coordinates of each source particle in the phase space file are resampled according to a 2D Gaussian random setup error kernel. In addition to differences in the positional sampling and in the treatment head description phase space versus virtual source model, Beckham et al. 6 do not recalculate the direction vectors for each translated source particle, as we do in our implementation. The influence of respiratory motion on the dose distribution was also evaluated using a dose convolution method, 4 performed by convolving the static dose distribution, D static calculated using static ) with the function, F motion, described above. The dose at a point r, that includes the motion, is calculated as follows: D motion r F motion D static r F motion r r D static r dr, F motion is cast in the form of a 1D discrete matrix along the z SI dimension of the patient, and is convolved with D static so that the integral shown in Eq. 1 is reduced to a summation. The treatment plan beam configuration consisted of 15 MV anterior, lateral and oblique fields, combined with segmental fields directed from the same angles to produce a dose distribution of 100 5% within the planning target vol- 4 5

1778 Chetty et al.: Monte Carlo fluence convolution respiratory motion liver 1778 FIG. 2. Isodose lines: 95%, 50%, and 20%, calculated using the fluence convolution solid blue lines and dose convolution dashed yellow lines methods, in the coronal view. Both treatment plans have been normalized to 100% at the isocenter. ume PTV. The CT data set used for treatment planning was acquired at exhale, under voluntary breath hold. The treatment planning volumes: gross tumor volume GTV, clinical target volume CTV, and planning target volume PTV are those defined by the ICRU Report No. 50, 11 and are derived in this work from our current liver treatment protocol. 12 The CTV is formed by a uniform, 1 cm expansion of the GTV. The PTV includes a uniform 0.5 cm margin for setup uncertainties, as well as a 0.3 cm margin superiorly added to account for the reproducibility of the exhale CT scan and a 1.5 cm margin inferiorly to account for patient breathing these expansions are performed relative to the CTV. 4 DPM calculations were conducted using a voxel size of 3 3 1 mm 3 in the x, y, z dimensions, a2mmstep size, FIG. 3. Dose volume histograms for the CTV shown for the static solid black line, fluence convolved dashed blue line and dose convolved solid orange line treatment plans. The inset represents a magnification of the high gradient region of the DVH, between the 95% and 105% dose values.

1779 Chetty et al.: Monte Carlo fluence convolution respiratory motion liver 1779 FIG. 4. Dose volume histograms for the CT-scanned right lung volume shown for the static solid black line, fluence convolved dashed blue line and dose convolved solid orange line treatment plans. and low energy electron and photon cutoff values of 200 and 50 kev, respectively. For each treatment plan, 6 billion histories were simulated 1 billion histories per field, resulting in 1 statistics of roughly less than 1.5% in calculated dose, if we combine the inherent uncertainty in the reconstructed fluence from the virtual source of 1%. The time required for these simulations was approximately 10 hours per 1 billion particles, running on a single 1 GHz, VMS-based, Alpha processor. Figure 1 illustrates a percentage difference map in the sagittal view between fluence convolved and static dose distributions. The difference map was calculated by subtracting the static dose from the fluence convolved dose for each point in the dose distribution. The fluence convolved and static treatment plans were normalized to 100% at the isocenter located within the PTV, as is routinely done for patients planned on our liver treatment protocol. 4,12 The difference between the absolute doses in the static and dose convolved plans at the isocenter, located within the central homogeneous region of the PTV, is negligible. The red color wash regions in Fig. 1 represent areas of positive dose differences indicating that the dose in the fluence convolved plan is higher than that in the static plan. Similarly, the blue color wash corresponds to regions where the fluence convolved doses are lower than those in the static treatment plan. Also, presented in Fig. 1 are the isodose regions corresponding to differences of 20% in the solid lines and 10% in the dashed lines. Maximum point differences of 25% are observed in the region superior to the PTV, located within the lung. The differences observed in Fig. 1 are a consequence of the motion of the liver due to breathing. Given that the CT scan was acquired at the exhale position, it is expected that, during inhale, the expansion of the lung will force the liver to move inferiorly. This means that the region of lung tissue located superiorly to the PTV will be exposed to radiation, while the inferior portion of the PTV will be blocked and will tend to be under-dosed, as noted in Fig. 1. Presented in Fig. 2 are the 95%, 50%, and 20% isodose lines for the treatment plans generated using the fluence convolution solid blue lines and the dose convolution dashed yellow lines methods, in the coronal view. The differences between these two plans are within 2% on average, however, the differences might well be larger had the PTV not extended outside the liver into the lung; the influence of the PTV, protruding into the lung, tends to minimize differences between the spatially invariant dose distribution an assumption required for dose convolution and the spatially variant dose distribution as calculated with fluence convolution. The reason for this is that the Monte Carlo static dose calculation is expected to correctly account for the spread of dose into the lung at the superior edge of the PTV ; convolving the static dose in this region is therefore likely to be as accurate as the fluence convolved calculation. It is clear, however, that a much more extensive investigation involving many treatment plans is necessary before conclusions can be drawn regarding the dosimetric differences between the fluence and dose convolution approaches used in accounting for breathing-induced organ motion in the vicinity of the liver. Analyses of dose volume histograms DVHs for the CTV and the right lung were conducted for the static, fluence and dose convolved treatment plans. For the CTV, DVH differences between the static and the motion convolved fluence or dose convolved dose distributions are relatively small as noted in Fig. 3 indicating that the margin between CTV and PTV adequately accounts for the motion. However, in this particular example, differences in the PTV DVHs between the static and motion convolved doses were also found to be negligible. A likely reason for this is that only the very su-

1780 Chetty et al.: Monte Carlo fluence convolution respiratory motion liver 1780 perior and inferior regions of the PTV are influenced by the motion, so that the resulting influence on the DVH is small. Figure 4 illustrates DVHs for the scanned right lung volume generated from the static and motion convolved treatment plans. Significant differences are noted for the motionconvolved plans in comparison to the static treatment plan. For example, the volume of lung receiving doses greater than 50% of the isocenter dose is 17% higher in the fluence convolved treatment plan relative to the static case. Such differences may be clinically relevant. In this study, we have implemented a fluence convolution method to account for the influence of liver motion in the superior-inferior direction due to breathing on the 3D dose distribution. We have tested this method on a clinical treatment plan and have found significant differences between the fluence convolved and static dose distributions, indicating the importance of organ motion in treatment planning. Although the fluence convolution method assumes a rigid-body approximation and ignores intrafraction effects much like the dose convolution approach, it can be incorporated to provide a more accurate dose distribution. The focus of future work will be to conduct an extensive treatment planning study of the fluence convolution method, involving lesions located at various locations within the liver as well as in other anatomical regions, such as the lung. ACKNOWLEDGMENTS This work has been supported in part by NIH Grant No. P01-CA59827 and by a University of Michigan Cancer Center grant funded by the John and Suzanne Munn Endowment. a Electronic mail: indrin@med.umich.edu 1 G. S. Mageras et al., A method of incorporating organ motion uncertainties into three dimensional conformal treatment plans, Int. J. Radiat. Oncol., Biol., Phys. 35, 335 342 1996. 2 M. A. Moerland et al., The influence of respiration induced motion of the kidneys in the accuracy of radiotherapy treatment planning, a magnetic resonance imaging study, Radiother. Oncol. 30, 150 154 1994. 3 J. M. Balter et al., Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing, Int. J. Radiat. Oncol., Biol., Phys. 36, 167 174 1996. 4 A. E. Lujan et al., A method for incorporating organ motion due to breathing into 3D dose calculations, Med. Phys. 26, 715 720 1999. 5 S. D. McCarter et al., Evaluation of the validity of a convolution method for incorporating tumour movement and setup variations into the radiotherapy planning system, Phys. Med. Biol. 45, 923 931 2000. 6 W. A. Beckham et al., A fluence-convolution method to calculate radiation therapy dose distributions that incorporate random set-up error, Phys. Med. Biol. 47, 715 720 1999. 7 J. Sempau et al., DPM, a fast, accurate Monte Carlo code optimized for photon and electron radiotherapy treatment planning calculations, Phys. Med. Biol. 45, 2263 2291 2000. 8 I. J. Chetty et al., Photon beam relative dose validation of the DPM Monte Carlo code in lung-equivalent media, Med. Phys. 30, 563 573 2003. 9 I. J. Chetty et al., A virtual source model for Monte Carlo modeling of arbitrary intensity distributions, Med. Phys. 27, 166 172 2000. 10 J. M. Balter et al., Improvement of CT-based treatment planning models of abdominal targets using static exhale imaging, Int. J. Radiat. Oncol., Biol., Phys. 41, 939 943 1998. 11 ICRU Report 50. International Commission on Radiation Units and Measurements: Prescribing, recording and reporting photon beam therapy, Bethesda, Maryland, 1993. 12 T. S. Lawrence et al., Treatment of cancers involving the liver and porta hepatis with external-beam irradiation and intraarterial hepatic fluorodeoxiuridine, Int. J. Radiat. Oncol., Biol., Phys. 20, 555 561 1991.