The Olis Titrator Series

Similar documents
6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

Manual for the MicroSIA Instrument

FROG-4000TM Quick Start Guide

Unpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place.

870 KF Titrino plus. Installation and tutorial EN

Introduction: Features:

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide

Aqueous GPC Instructions - Detailed

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

PerkinElmer Series 200 HPLC: Operating Procedure

915 KF Ti-Touch. Tutorial EN

CH142: Instructions for the Vernier LabPro

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

Installation Instructions of TitrIC 1

Manual HPLC-4 Alliance-UV

Accessories for

On-Line Instrument Systems, Inc. 130 Conway Drive, Suites A & B, Bogart, GA (800) (706)

Thermo Scientific GENESYS UV-Vis Spectrophotometer Accessories. Make the most of every sample

KM-4800w. Installation Guide

SymBiot I Sample Workstation

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System

Mitos P-Pump. product datasheet

848/877 Titrino plus. Tutorial EN

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

Agilent ICP-OES SVS 2 Productivity Package Installation and Upgrade Instructions

Installation Instructions of TitrIC 3

Application Bulletin 318 e. Installation Instructions of TitrIC 4

4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system

DIMATIX Printer Tutorial and Operating Instructions

QL5-EX Quick Start-up Guide

Removal and Installation8

User s Guide. Pressure Controller for liquid delivery systems

Application Bulletin 321 e. Installation Instructions of TitrIC 7

Zeiss Z.1 Lightsheet. Quick-Start Guide

On-Line Instrument Systems, Inc. 130 Conway Drive, Suites A & B, Bogart, GA (800) (706)

Using the SASS 2300 Wetted-Wall Air Sampler

Canto II Operation Protocol

MICROFLUIDIC EOF KIT 9015

MINISCAN INSTRUCTIONS

Service Manual. Subject Issued By. Date MM/DD/YY 29/06/17. Hybrid Atom Printhead Customer Support Team ITEM PART NUMBER QTY

Blood Hemolysis Detector User Manual

minipcr Mobile App operation (Android devices) Appendix: maintenance and operating conditions

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

The Pro/Classic Series Printhead angle can be set between 0 and 90. Common settings are shown below. 192/32 1 Print Head

RD-SDM02 Contact Angle Measurement Instrument. Operation Manual

HI FREE & TOTAL SULFUR DIOXIDE MINITITRATOR for wine analysis

OPERATING INSTRUCTION MANUAL MODEL P/R60 & P/R65 DIFFERENTIAL PH & ORP PROBES

STANDARD OPERATING PROCEDURE

SOLUTIONS RECERTIFICATION CALIBRATION AND UV/VIS REFERENCE MATERIALS. Photometric and Wavelength Accuracy Reference Material Set

Temperature logger with internal sensor Instruction manual

Accessories for

Premium Continuous Ink Supply Systems.

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Magnetic Immunoassay Reader Operator Manual

Instruction Manual HI Titratable Acidity Mini Titrator & ph Meter for Dairy Products

Equipment Overview: Safety:

DataHub: FAQs/Troubleshooting

- 1 - TITRAMAX OPERATING MANUAL

SPA III Quick Reference Guide for Custom Panels

Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012

Operation of the Toxic Organic Chemical Air Monitor (TOCAM ) with the Ellvin Chromatography Software

ASXPRESS PLUS Rapid Sample Introduction System Quick Installation Guide

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Biogas Batch Fermentation System

Unscrew and remove the cap from the waste tank

HI FORMOL NUMBER MINITITRATOR for Fruit Juices and Wines analysis

NANOSIGHT NS300 OPERATING MANUAL. MAN EN-00 January 2016

Version 1.4 June 2018

1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader

9 Maintenance and Repair

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

VIAFLO Fixed and Adjustable Spacing Electronic Pipettes

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

Waters 600E Multisolvent Delivery System Quick Start Guide

SMART FORCE AND TORQUE SENSORS. User s Guide

Auper Electronic Controls Inc. Eclipse 784 B. Quick User Guide

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

Quick Solutions Packet

Checklist for the SMPS & CPC (5.400 & 5.500) PART 1 Setup of the external Plumbing (for the DMA and the CPC):

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

SMART FORCE AND TORQUE SENSORS. User s Guide

HP Photosmart c3180 Main Circuit Board Replacement

H115i QUICK START GUIDE

Jouse. User s Guide. Revision 1.0. Prepared by: Compusult Limited 40 Bannister Street Mount Pearl, Newfoundland, Canada A1N 3C9

Appendix C. Vernier Tutorial

AD Leak Tester INSTRUCTION MANUAL 1WMPD

USER MANUAL INCLUDED IN THE BOX

INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor

Model : LUV-300 Please read this manual carefully before using the instrument Labnics Equipment

PX Series Balances. Quick Start Guide. Please download the user manual from

HITACHI LACHROM ELITE

Upchurch Scientific Nano Flow Sensor User s Guide

GEMINI. Compact Microplate Processor. System Guide

Installation instruction for MVA-20: 894 Professional CVS fully automated for CVS determinations. Content

A. About this Manual

STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

User Manual. Coulometric KF Titrator C10S/C20S/C30S

eprep Sample Preparation Workstation

831 KF Coulometer. Never before has coulometric Karl Fischer water determination been so economical!

ONCORE Manual Automated Slide Staining System

Transcription:

The Olis Titrator Series Perfect Titrations into any type of cuvette With considerable effort, one can do precise titration studies by hand. Or, one can let the OLIS titrators do the technician s work over and over again, perfectly. There is a great deal of flexibility in this computer controlled titration accessory. Volumes added can be from 0 to 100% of the volume of the syringe, in increments to 0.0001th of the syringe volume. The syringes in the titrator can be the same or different sizes; volumes can be 0.25, 0.5, 1.0, 2.5, 5, or 10 ml. Thus, stepping resolution (minimum volume moved) is 0.025 ul to 1 ul. Stirring in the cuvette is not required, so cuvettes of any pathlength and geometry can be used. The cuvette can be cylindrical or rectangular, long pathlength or short, as the measurement requires. Any number of scans can be made of each sample preparation, so one can test for temporal stability, too. Ambient and Peltier versions are available. Do your protein folding/unfolding, equilibrium ligand binding experiments, and acid/base titrations by hand. Over and over. Or free yourself with the OLIS Titrator. It is to hire a perfect and tireless pipetter.

d. Titrator i. Setting up the titrator 1. Place the titrator on a sturdy location as close as possible to the sample chamber. 2. Determine if the titrator has its own control box. Some will use the instrument control box while others will have their own. A control box will be provided with the instrument if necessary. 3. Connect the four motor cables from the titrator to either the titrator control box or instrument control box. If a separate control box is present, connect the RS232 connection to an appropriate com (or USB) on the computer. 4. Connect the titrator to the sample chamber using the tubing included with the titrator. There is a flow valve on the 92

titrator which determines whether the output is waste (front) or to the sample (back). Connect the tubing to the flow position by turning the tubing connector clockwise. Connect the other end of the tubing to the connector on the sample chamber. If no connector is present contact Olis for assistance. 5. Inside the sample chamber, connect the flow cell to the tubing port on the inside of the chamber. 6. If relatively large volumes are to be added to the cell during the titration (as is often the case with urea or guanidine), there must be an overflow connection made from the flow cell. This could simply be a piece of tubing large enough to hold the excess volume of the titration. Often, Olis will provide an overflow connection in the sample chamber. 93

7. Fill two disposable syringes with pure water (preferably filtered and degassed). Attach these to the fill valves on the titrator. Put the fill valves in the Fill position. 8. Turn on the instrument. Start SpectralWorks. On the Repeated Scans page, choose Repeated Scans as a function of Titration script. If this option is not available, contact Olis for assistance. The titrator will calibrate, which involves moving the syringe plunger until a physical sensor is reached. 9. Click on the Manual Control form, click on the Park All button. 94

10. The syringes will move to the half full position. If any bubbles are present remove the thumb screw holding the plunger to the motor. Move the plunger by hand in and out until any bubbles are removed. Reattach the thumb screw. 11. Click the Empty All syringes button. ii. Setting up the Peltier temperature controlled titrators 1. After unpacking the Titrator assembly, connect cables labeled 1 and 2 to the titrator and to the back of the Olis Control box. 95

. 2. Connect the cable labeled QTTR1 to the titrator and Temperature control box. Connect the QTTR1 cable to the temperature control box. 3. Connect the RS232 cable from the black Olis box to the com port of the computer. 4. Connect power cables to the Olis and Temperature control boxes. 5. Connect the water tubing from the Out connector on the Olis Heat Exchange pump to the In connector on the titrator by pressing the tubing in until a click is heard. The tubing can be removed by pressing the release on the connector and removing the tubing. The tubing will seal to prevent leaking. DO NOT remove tubing with the pump running as this may damage the pump. Connect the Out of the titrator to the plastic bottle provided with the instrument. Connect the other connection on the bottle to the In connection of the pump. Fill the bottle with approximately 300 milliliters of deionized water, and plug in the heat exchange pump. Allow the pump to run for several minutes. 96

iii. Using the titrator 1. Remove the lids on the syringe chambers by first unscrewing the four thumbscrews on each chamber. Fill the chamber with deionized water within 5 mm of the top of the chamber. Replace the lids and thumbscrews. You should be able to clearly see the syringes through the lids. 2. Attach reservoir syringes containing buffer or solvent to the luer lock connectors on the titrator. These syringes may be glass or plastic. It is advisable to degas the buffer or solvent prior to use to prevent the formation of bubbles in the titrator. 97

3. Turn on the instrument, titrator, temperature controller, and plug in the heat exchange pump. It is important that the pump be running while the temperature controller is on or errors will occur. iv. Determining the cuvette volume of the titrator 1. Fill the reservoir syringes with water or buffer. 2. Place the Syringe valves in the Fill position and the Flow valve in the Flow position. 98

3. Using the Manual Control Form (Repeated Scans page) fill syringe 1 to a volume exceeding the expected cuvette volume by clicking on the syringe in the control and sliding the volume control up (see screen shot in the Manual control form in software manual). The volume is displayed below the syringe. If air enters the system, remove it as described in Removing bubbles from the Titrator section. 4. Move the syringe valves to the Flow position. 5. Record the volume in syringe 1. 6. Incrementally empty syringe 1 by dragging the volume control down slowly. Syringe 1 should empty into the cuvette. Stop after an acceptable volume has filled the cuvette. An acceptable volume fills the cuvette so that it is above the level required for measurement. 7. Record the volume in syringe 1. The cuvette volume is the difference between the volume recorded in step 6 and the current syringe 1 volume. This volume is the minimum volume that should be used in the Cuvette Volume section of the titration script. v. Changing syringes on the titrator 1. With the instrument on, remove one (or both) of the thumbscrews holding the motor bracket to the plunger syringe. 99

2. Using the manual control form on the Repeated Scans page, select the desired syringe(s) and click the Fill Syringe button. This will remove the motor bracket from out of the way. 3. Remove the two thumbscrews on the bracket holding the syringes in place. Set the bracket aside. 4. Using the tool shown, gently unscrew the syringe from the fill valve. Place the syringe aside. 100

e. Sample changers 5. Insert the new syringe into the fill valve. Rotate the syringe clockwise into the threaded fitting. Use the tool to tighten the syringe, but do not overtighten. This can damage the Teflon washer and cause leakage. 6. Replace the syringe bracket. 7. Click on the Empty syringe button on the Manual Control form to move the motor to the empty position. 8. Replace the thumbscrews holding the motor brackets to the syringe plungers. i. Using the sample changer 1. The sample changer will calibrate upon starting the instrument. This involves automatically changing the position of the cuvette holder until a physical sensor is reached. This ensures the changer is calibrated to its home position. 2. If the sample changer is associated with a Peltier cell, there will be an associated control box. Pressing a position on the control box will move the turret to that position. Holding a button in for five sec or more will initiate calibration. The turret can also be totally controlled through the use of the software. 101

Accessories 7. Repeat the stopped flow shot and repeat the adjustment if necessary. vi. Changing the number of data points 1. When collecting stopped flow data at a single wavelength the number of data points in the dataset is typically fixed to 1000. Note that the number of scans collected in the rapid scanning mode of RSM is determined by the averaging mode and the data collection time. 2. To change the dafault number of single wavelength stopped flow data points, simultaneously press ctrl+shift+alt+a to access the administration screen. 3. Locate the A/D node from the those listed on the administration screen. This will have the label GaGe, Datel, or Bitwise. Click on this node to open it. 4. Locate the parameter called NumberOfSFPoints. Click on the parameter and change it to the desired number of points. Press enter and close the administration screen. vii. Digital filtering (RC Filter) of time data d. Titrator 1. Application of a digital RC filter can be achieved by clicking on the Apply RC Filter checkbox. 2. On the Live Display page, click on the checkbox labeled Apply RC Filter. Choose the appropriate strength of the filter in the list box next to the RC Filter control. The units of this value is milliseconds. It is advisable for this value not to exceed 5% of the total time of data collection. i. General description of titrator 1. The Olis titrator consists of two to four motor-driven syringes connected to a flow cell inside the sample chamber. One of the syringes is designated in the software as the sample syringe. The sample to be measured is loaded into this syringe and the titrant(s) in the other syringe(s). During the titration a specified volume form the titrant syringe(s) is added to the cell containing the contents of the sample syringe. The sample is mixed by drawing the sample into the sample syringe, and pushed into the cuvette for measurement. All of the titrator controls, except movement of the valves, is controlled through the software. 51

Accessories 2. The titrator may consist of two, three, or four syringes. The four syringe titrator allows addition of up to three different titrants for each measurement. Temperature control using Peltier cells is also available. The default is ambient temperature. 3. The titrator is activated by selecting Repeated Scans as a Function of Titrator Script. The titrator will be subsequently calibrated. A message will appear indicating that reservoir syringes containing solvent or buffer should be connected and the Fill/Flow valves should be in the Fill position. At this point the syringes will fill to a home position and return to empty. ii. Manual control functions Allows for software controll of the Olis titrator. Access to this screen is in the Repeated Scans tab. Script File This designates the current script file name. It is identical to that shown in the Repeated Scans tab. 4. The titrant addition volumes, number of mix cycles, and incubation time after each mix are specified by a script file. The details of the script file format are described in Appendix III. Data collection will begin automatically after each addition and incubation time, and collection parameters are given on the Live Display page. 5. The volume resolution is limited by the volume of the syringe (0.5 ml to 10 ml) and the number of motor steps in traversing the syringe (10,000). Thus the volume resolution ranges from 0.05 µl/step to 1 µl/step. Execute Line Click on this button to execute a line of the script without collecting data. The software will prompt for which line to execute. The syringes will move, but no data will be collected. Execute Script Click on this button to execute an entire script without collecting data. The syringes will move, but no data will be collected. 52

Accessories Select Script This button allows the user to select a valid titration script. Changes in the script in this window will change it in the others as well. View/Print Script This button allows the user to open the script for editing or printing. Note: Do not attempt to change the Default Titrator.osf script as it is a read only file and is necessary as a backup script file. Com Port Number Ensure that this number matches the com port number that the titrator is actually using. Syringe Size (ul) Ensure that the syringe size in this display matches the actual syringe size for the selected syringe. To select a syringe, click on the desired syringe in Syringe Control. The text box below the syringe will become yellow, indicating it has been selected. Changing the syringe size changes the size of only the selected syringe. Syringes are selected using the Syringe Control. Calibrate Syringe These buttons initiate a calibration procedure. During calibration, the syringe fills completely until a sensor is reached, whereupon the syringes empty. It is important to have the valve in Fill position and have enough sample in the fill syringe. If samples are precious, it is important that calibration is done using water or buffer. Calibrate Selected only calibrates the selected syringe, while Calibrate All initiates the calibration procedure for all syringes. Syringes are selected using the Syringe Control. The keystroke shortcuts for Calibrate Selected and Calibrate all are S and A, respectively. Park Syringe When one of these buttons is pressed, on or more of the syringes fills or empties to half full. Park Selected parks only the selected syringe, while Park All moves all of them. Syringes are selected using the Syringe Control. Rinse Syringe Rinsing syringes involves alternating filling and emptying them. When clicking on of these buttons, the software will prompt to ask the number of desired rinses (1-99). It is important to have the valve in Fill position and have enough sample in the fill syringe. If samples are precious, it is important that calibration is done using water or buffer. Rinse Selected rinses only the selected syringe, while Rinse All initiates the procedure for all of them. Syringes are selected using the Syringe Control. The keystroke shortcuts for Rince Selected and Rinse all are R and N, respectively. 53

Accessories Fill Syringe These buttons cause syringes to completely fill. It is important to have the valve in Fill position and have enough sample in the fill syringe. If samples are precious, it is important that calibration is done using water or buffer. iii. Filling and rinsing syringes 1. On the Repeated Scans Page click on the Show Manual Control Form button. If this button is not active, select Repeated Scans as a Function of Titrator Script to activate the titrator. Fill Selected rinses only the selected syringe, while Fill All initiates the procedure for all of them. Syringes are selected using the Syringe Control. Empty Syringe These buttons cause syringes to completely empty. It is important to have the valve in the desired position so the contents of the syringe are not inappropriately expelled. If samples are precious, it is important that calibration is done using water or buffer. 2. Ensure both fill valves are in the Fill position and ensure that enough liquid is present in the reservoir syringes to fill each drive syringe half way. Empty Selected rinses only the selected syringe, while Empty All initiates the procedure for all of them. Syringe Control This control allows the user to manually manipulate the syringe volume. The current volume is displayed in the box at the bottom of each virtual syringe. This volume can be changed by entering in a new value (in ul) or dragging the sidebar control to the desired approximate volume. Filling of the syringe is graphically illustrated by the blue shading. To select a syringe, click on the body of the syringe. A yellow text box will indicate that the syringe is selected. 3. Click on the desired syringe to highlight it. Fill the syringe by dragging the cursor for the syringe. Alternatively, the new volume can be entered in the text box associated with the syringe. 4. To rinse the syringe, click on the Rinse Selected button and enter the number of rinses. The syringe will repeatedly empty and fill one plus the number of extra rinses requested. iv. Creating a titrator script 1. Use Windows Explorer to browse to c:\program Files\OLIS GlobalWorks\ Utilities folder and find the Script Editor. exe program. There is also a help file in the folder entitled Script Editor Help.doc. 54

Accessories 2. Double click the Script Editor program and click on the Titrator Script tab. 4. Enter the Syringe Fill Volumes and Mix Fill Percentage, and Cuvette volume. See the help file in the c:\program Files\OLIS GlobalWorks\Utilities for descriptions of these parameters. 3. Click on Generate Script. Choose the two syringe, three syringe (constant sample), or four syringe. Enter the desired volumes to be included in the script. Enter the number of mixes and incubation times. Click OK. 5. Click on Check Script Volumes. 6. Choose Save As from the File menu. Choose this file when running the titration script. 7. Close the Script Editor program. 55

Accessories v. Running a titrator script 1. On the Repeated Scans page, ensure the Repeat Scans as a function of. is set to Titration Script. This will activate the titrator. 2. Ensure both fill valves are in the Fill position and ensure that enough liquid is present in the reservoir syringes to fill each drive syringe half way. 3. Click on the Park All button. 4. Remove the thumb screw connecting the syringe plunger to the motor bracket. Empty and fill the syringe by hand repeatedly until all bubbles have been forced into the reservoir syringes. Replace the thumb screw. 5. Click on Empty All to empty the syringes. 6. Close the Manual Control Form after the syringes have emptied. e. Enzyme Assay Device 2. Click on Browse and point to the desired titration script. Olis script files use the file extension.osf. 3. On the Live Display page enter the desired wavelength Scan Range, Number of Increments, and Integration time. 4. Click on Collect Data. vi. Removing air bubbles from the syringes 1. On the Repeated Scans Page click on the Show Manual Control Form button to open this form. i. General Description of the Enzyme Assay Device 1. The enzyme assay device is a modification of the Olis titrator. This device typically consists of four syringes each which passes though a mixer to a flow cell. The cell is located in the sample chamber of the fluorometer. The enzyme assay device operates differently from the titrator in that the volume specified are always pushed into the cell. No sample syringe is designated. 2. The option of syringe temperature control is available through use of a Peltier controller. Typically, one syringe is kept in separate control as the others allowing the enzyme to be stored at a different temperature than the assay temperature. 3. The Enzyme Assay Device is activated by selecting Repeated Scans as a Function of Enzyme Assay Script. The device will be subsequently calibrated. 56

Free the students and technicians to focus on sample isolation and data analysis instead of the tedium of serial sample titrations. Overnight runs can be made with no supervision. To own an OLIS Titrator is to have a tireless and error-free pipetter! Protein denaturations, equilibrium titrations, acid-base titrations, enzymatic assays, and more! Specify... cuvette volume sample volume titrant volume number of mix cycles [see footnote, page 3] and incubation time before data collection...then Walk Away! For more information on this and other Olis products: Visit Write Call Tour www.olisweb.com sales@olisweb.com 1-800-852-3504 in the US & Canada 1-706-353-6547 worldwide On-Line Instrument Systems, Inc. 130 Conway Drive, Suites A, B & C Bogart, GA 30622