STANDARD OPERATING PROCEDURE

Size: px
Start display at page:

Download "STANDARD OPERATING PROCEDURE"

Transcription

1 Document No: LAB/023/2 Page 1 of 7 Uncontrolled copy if not printed on yellow paper. Title: STANDARD OPERATING PROCEDURE Use of the Magtration System 8Lx Document No: LAB/023 Version No: 2 CopyNo/Holder: 09/Jay Brown Issue Date: 07/09/07 Review Date: 07/09/09 Issued by: Jay Brown 1 INTRODUCTION 1.1 The Magtration System 8Lx is a desktop automated nucleic acid purification system that uses magnetic beads to extract DNA from up to 7 ml of whole blood. 1.2 Up to 8 blood samples can be processed simultaneously in 1 hour. 1.3 The Extracted DNA is eluted into 1ml of 10 mm Tris-HCl, ph SCOPE 2.1 This SOP describes the procedures for using the Magtration System 8Lx to extract DNA from up to 7ml of whole blood, which has been treated with an anticoagulant such as EDTA. 3 RESPONSIBILITIES 3.1 It is the responsibility of the laboratory personnel carrying out the procedure to ensure that all steps are completed both competently and safely. 4 RELATED DOCUMENTS 4.1 Disposal of waste is described in the SOP LAB/ Handling of Reagents is described in SOP REA/ Any SOP deviations are recorded as described in the SOP QAU/ Allocating samples to locations is described in SOP LAB/ Maintenance is recorded on Form LAB/ A list of Designated Trainers is recorded on Approved List PER/005C 5 HEALTH AND SAFETY 5.1 All handling of blood samples where there is potential of exposure to blood (e.g. transfer from client tubes) must be performed in a Class II laminar flow cabinet equipped with a drip tray.

2 Document No: LAB/023/2 Page 2 of A lab coat and gloves must be worn. 5.3 All staff working with blood are advised to have Hepatitis B vaccinations. 5.4 Care should be taken when using the Magtration System 8Lx as some components move out and there can be a risk of collision. Ensure that other laboratory users are aware of this before starting the program. 5.5 The reagent kit contains protein denaturants. Wear gloves, and take appropriate safety precautions when using this kit, to avoid direct skin contact. 6 EQUIPMENT/MATERIALS/REAGENTS 6.1 Equipment Equipment Manufacturer CIGMR ID# Magtration System 8LX PSS 0495 Variable volume pipette (P1000) Gilson N/A MagnaRack Invitrogen N/A 6.2 Materials Material Supplier Cat# Blood sample(s) for extraction N/A N/A Syringe PSS F ml Teklab blood tube Teklab Z ml Sarstedt tube Sarstedt Sarstedt elution tube Sarstedt ml cryovial Greiner Reagents Reagent Supplier Cat# Magtration reagent pack PSS E % Gigasept S&M SYSTEM BOOKING 7.1 To book the Magtration System 8Lx, users will send an to eightlxbooking@listserve.manchester.ac.uk stating the time and date they want to book. The number of samples to be processed must be included in the , along with the CIGMR Project Code to be charged. 7.2 Confirmation of the booking will be received via . Reagent packs will be provided on the day of the booking. The number of reagent packs available will be equal to the number of samples specified in the booking .

3 Document No: LAB/023/2 Page 3 of LX users should also book a Class II cabinet in order to perform the blood transfer at an appropriate time (Section 8.1). Booking forms for Cabinet 1 & 2 can be found on the network drive N:\Projects\CIGMR LAB BOOKINGS\Class II Cabinets. 8 DNA EXTRACTION PROCEDURE 8.1 Defrost 1-8 whole blood samples, and in a Class II laminar flow cabinet. Transfer a maximum of 7ml of each blood sample to a labelled 10ml Teklab blood tube. All blood contaminated plastic must be fully submerged in 12% Gigasept overnight, prior to disposal in a Biohazard incineration container 8.2 Before using the Magtration System 8Lx, open the front cover by sliding it upwards, and clean the internal sample stage and tube racks with 70% ethanol as described in section Close the front cover. 8.4 Turn on the Magtration System 8Lx by using the main power switch located on the left side of the machine. 8.5 Open the lower front door located on the front side of the Magtration System 8Lx and turn on the power of the PC by pressing the blue switch located in the centre of the unit. The PC will boot Windows and automatically open the 8Lx system s software. 8.6 A welcome screen will be displayed, showing the following information: User User ID User Level Last Login Last Protocol Last status (displays the error code produced last session, if no errors were produced it displays a 0) Total operating time Last batch ID Total UV time Total Tip On time Total piercing time 8.7 Using the stylus pen on the touch screen select Ok to enter the program. 8.8 Using the stylus pen select Ext.Menu, to display the Ext.Menu screen. 8.9 Select the System Initialization option. The 8Lx sets all the units to their starting home positions Select the Cleaning pins option. The Piercing stage will move to the front of the machine. Open the Front cover and check that each of the piercing pins are in their holes located behind the black magnets. Close the front cover and select Ok. The piercing stage will move back into position.

4 Document No: LAB/023/2 Page 4 of Press the Main Menu button, and then select the Protocol button located in the top left corner of the screen From the new window select Whole Blood for 7ml ver 1.9. Press Ok Select Set Tips. The Z-unit that holds the tips moves forward and the syringe plunger head holder opens up Check on the touch screen that the door is registered as unlocked by looking for a red square in the middle of the screen that says Unlock Open the front of the Magtration System 8Lx. Holding the Syringe tip by the large main body section with the tip pointing down. Insert the lower portion of the syringe tip into the lower hook, then push the upper portion of the syringe tip into the upper hook. Once all tips have been inserted, one for each sample being processed, close the front cover and press Ok. The Z-unit moves to its home position Open the front cover. Use the handle located on the underside of the sample stage to pull the stage forward Insert one 30 ml Sarstedt tube per sample into the heater block (second hole from the back) Remove the sample rack to allow you to insert the pre-packed reagent cartridges. The arrows on the packs must point towards the back of the instrument. Use one cartridge per sample being extracted Remove and retain the lid from a Sarstedt elution tube and insert one tube per sample into the first hole (at the front) on the sample rack Replace the sample racks making sure that R is on the right, and L is on the left. Care must be taken to ensure the sample racks are correctly positioned (in particular, racks must lie completely flat). Remove the lids from the 10 ml Teklab blood tubes and place them into the third hole (towards the back) on the sample rack. If using barcodes line them up with the notches on the sample rack Push the sample stage back into the Magtration System 8Lx, and close the front cover On the touch screen select Scan Info. The barcodes on the samples will now be read. If the barcodes can not be read, or you are not using barcodes, enter sample information manually be selecting the cell you wish to edit on the touch screen, and then select Edit Info Using the On-screen Keyboard input the sample information and press Enter. Repeat 9.22 and 9.23 for other samples if necessary To begin the extraction process press Start. Then select Start Batch. Tips must be completely aligned with the pins. If not, press Pause Batch, open the cover, and gently align the bottom of the tips over the pins. Close the lid and press Start Batch. This can be repeated until the tip covers are ejected. If the tips are not aligned, the tip covers will not eject correctly and the program will abort (characterised by a loud grinding noise). The robot must immediately be switched off and a Designated Trainer notified should this occur.

5 Document No: LAB/023/2 Page 5 of The extraction process will start and is completed in 1 hour Once the method has finished successfully the screen will display Writing data of result date into USB key Memory. Select Ok. This will save the information of the current batch to the hard drive. For retrieving data see 9.39 to On the touch screen select the Unlock option Open the front cover and move the sample stage forwards, as described in Remove the elution tube and screw on the lid. Transfer the tube to the Invitrogen MagnaRack. Leave on the rack for a minimum of 30 minutes to allow any residual magnetic beads to bind to the magnet With the tube still positioned in the MagnaRack, remove the lid and carefully transfer the DNA solution to an appropriate clean tube (e;g 1.8ml cryovial for downstream processing on the Hamilton robot). Tubes must be labelled prior to DNA transfer Remove the 10 ml blood tubes, replace their lids and dispose of them in a Biohazard incineration container Remove the pre-packed reagent cartons and transfer them to a Class II cabinet. Pour their contents into a waste container containing 12% Gigasept and leave overnight before discarding down the sink. Dispose of the empty reagent packs in a Biohazard incineration container. The same procedure is followed for any waste produced in Section Close the sample stage and front cover. On the touch screen select Remove Tips. The Z- unit will move forward Open the front cover Hold the syringe tip below the lower hook. Press the lower hook down and gently pull the syringe tip forward. Then pull the upper section of the syringe tip out of the upper hook. Dispose of the syringe in the Biohazard incineration container Open the cover and remove the 30 ml Sarstedt tubes from the heater block and dispose of them in a Biohazard incineration container. Use caution as the heater blocks will be hot Close the front cover and select Ok. The extraction process is now complete Check for any spillages and clean according to 9.3 if necessary In the event of error codes being produced contact a Designated Trainer for advice To retrieve data from the Magtration System 8Lx, select Ext. Menu. Then select Result File Files are recorded by date (written backwards) and batch e.g _0000.dat would be batch 1 on 1st January Highlight the result file you wish to look at and press Select. The result information for that batch will now be displayed in the table at the bottom of the screen.

6 Document No: LAB/023/2 Page 6 of Before shutting down the system at the end of the day repeat step 10.3 and carry out the sterilization procedure described in SYSTEM MAINTENANCE 9.1 Sterilisation procedure. At the end of each day s use the Magtration System 8Lx needs to be sterilized using the built-in germicidal UV lamp From the main screen select Sterilize The sterilize screen will appear. Use the arrows located in the bottom right corner to adjust the sterilization time to 2 hours Press Start The UV lamp will run for 2 hours, the Magtration System 8Lx will then automatically shut down. The robot should be turned off using the switch on the left-hand side after use To disable the lamp before the sterilization process is complete press Cancel. 9.2 Piercing pin cleaning. Once a month the Piercing pins need to be cleaned and the process must be signed off on the Magtration System 8Lx maintenance log (Form LAB/007) which is located on the side of the Magtration System 8Lx Whilst the front cover is closed go to the main screen and select Ext.Menu Select Cleaning Pins. The pin stage will move forward Open the front cover and remove one piercing pin Use a soft cloth dampened with 70% ethanol to clean the pin Return the pin to its original location Repeat steps to to clean all 8 pins Close the front cover and press Ok. The piercing stage will move back to its home position. 9.3 Stage and magnet cleaning. Once a month the stage and magnets need to be cleaned and the process must be signed off on the Magtration System 8Lx maintenance log (Form LAB/007) which is located on the side of the Magtration System 8Lx Before cleaning the stage and magnets, wait at least 30 minutes after extraction operations have been performed. This will allow the incubation areas to cool Do not allow magnetic cards or metal objects get close to the magnet areas Open the front cover.

7 Document No: LAB/023/2 Page 7 of Dampen a cloth with 70% ethanol and wipe the incubator stage, including inside the incubator holes Wipe the black magnet areas Close the front cover Record that the procedure has been carried out on the Magtration System 8Lx maintenance log (Form LAB/007) which is located on the side of the Magtration System 8Lx. 10 DOCUMENT HISTORY This is the second version of this SOP. The following changes were made: Overall re-write Re-formatting Pin alignment check in 8.22

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Instructions for use of AS9100, AS9101, AS9200 and AS9201 TM529 11/17 Table of Contents 1 Description...1 2...2 2.1. Maxprep Liquid

More information

icolumn 12 Operation Manual

icolumn 12 Operation Manual icolumn 12 Operation Manual AccuBioMed Co., Ltd. 8F.-8, No.5, Wuquan 1st Rd., Xinzhuang Dist., New Taipei City 24892, Taiwan (R.O.C.) Tel: +886-2-2299-5989 Fax: +886-2-2299-2678 www.accubiomed.com AccuBioMed

More information

Maxwell CSC Instrument Operating Manual

Maxwell CSC Instrument Operating Manual TECHNICAL MANUAL Maxwell CSC Instrument Operating Manual Instructions for use of the Maxwell CSC Instrument, Model Number AS6000. Note: Please ensure that all sealing tape and any residual adhesive are

More information

AnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System

AnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System BioChain Institute Inc. Instruction Manual Automated Nucleic Acid Preparation System AnaPrep 12 Catalog No.: Z1321001 Rev. Date: April 30, 2015 Manual No. Z1001 Version: Z1321001UA (MV1.14) For General

More information

Maxwell 16 MDx Instrument

Maxwell 16 MDx Instrument OPERATING MANUAL Maxwell 16 MDx Instrument Instructions for Use of Product AS3000 2800 Woods Hollow Rd. Madison, WI USA Revised 4/15 TM320 Maxwell 16 MDx Instrument Operating Manual All technical literature

More information

Maxwell CSC Instrument Operating Manual

Maxwell CSC Instrument Operating Manual Technical Manual Maxwell CSC Instrument Operating Manual This Technical Manual contains the instructions for use of the Maxwell CSC Instrument, Model Number 9800-000, Catalog Number AS4000. This product

More information

QIAsymphony SP/AS User Manual Operating the QIAsymphony SP

QIAsymphony SP/AS User Manual Operating the QIAsymphony SP April 2012 QIAsymphony SP/AS User Manual Operating the QIAsymphony SP For use with software version 4.0 Sample & Assay Technologies Trademarks QIAGEN, QIAsymphony (QIAGEN Group); Falcon (Becton, Dickinson

More information

QIAsymphony SP/AS Operating the QIAsymphony SP

QIAsymphony SP/AS Operating the QIAsymphony SP December 2017 QIAsymphony SP/AS Operating the QIAsymphony SP For use with software version 5.0 QIAGEN GmbH, QIAGEN Strasse 1, 40724 Hilden GERMANY R1 1107354 Sample to Insight Contents 1 Introduction...8

More information

TABLE OF CONTENTS. Table of Contents. Table of Figures

TABLE OF CONTENTS. Table of Contents. Table of Figures TABLE OF CONTENTS Table of Contents Table of Figures 1 2 1 Introduction 3 2 Installation 4 2.1 Installation Precautions 4 2.2 Environmental Requirements 4 3 Specifications 4 4 Precautions 4 5 Materials

More information

Maxwell 16 IVD Instrument Technical Manual

Maxwell 16 IVD Instrument Technical Manual Maxwell 16 IVD Instrument Technical Manual In Vitro Diagnostic Medical Device MDSS GmbH Schiffgraben 41 30175 Hannover, Germany INSTRUCTIONS FOR USE OF PRODUCT AS3050 Printed in USA 4/15 Part# TM315 1.

More information

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011 QUICK REFERENCE Ion 540 Kit Chef Catalog Numbers A27759, A30011 Pub. No. MAN0014119 Rev. C.0 Note: For safety and biohazard guidelines, see the Safety appendix in the Ion 540 Kit Chef User Guide (Pub.

More information

Premium Continuous Ink Supply Systems.

Premium Continuous Ink Supply Systems. Premium Continuous Ink Supply Systems. Instruction Manual For Epson Printers R200, R210, R230, R300, R310, RX650 RX510, RX630 157 Union Road, Ascot Vale Victoria 3032 Phone: (+613) 9005 5555 2 Installation

More information

To connect the AC adapter:

To connect the AC adapter: Replacing the AC Adapter Replacing the AC Adapter 3 Plug the power cord into a wall outlet. The power indicator turns on. To connect the AC adapter: Connect the power cord to the AC adapter. Power indicator

More information

Maxwell RSC Instrument Operating Manual

Maxwell RSC Instrument Operating Manual Maxwell RSC Instrument Operating Manual Catalog Number AS4500 TM411 3/18 Table of Contents 1 Introduction...5 1.1 Maxwell RSC Instrument Purification Procedure...5 1.2 Product Use...6 1.3 Product Use Limitations...6

More information

Total WBC count and differential WBC count HemoCue WBC DIFF system

Total WBC count and differential WBC count HemoCue WBC DIFF system TLM_PRO_029 and Version No. 001 Page 1 of 6 Department: Clinical Sciences Unit: Tropical Laboratory Medicine Total WBC count and differential WBC count HemoCue WBC DIFF system Project/study: Not applicable

More information

9 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module. InfinityLab LC Series Diode Array Detectors User Manual

9 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module. InfinityLab LC Series Diode Array Detectors User Manual InfinityLab LC Series Diode Array Detectors User Manual 9 Maintenance Introduction to Maintenance 158 Warnings and Cautions 159 Overview of Maintenance 161 Cleaning the Module 162 Remove and Install Doors

More information

Instruction Manual. Automated Nucleic Acid Purification System. AmpliX 12. Cat # 8084

Instruction Manual. Automated Nucleic Acid Purification System. AmpliX 12. Cat # 8084 Instruction Manual Automated Nucleic Acid Purification System AmpliX 12 Cat # 8084 Catalog No.: ZP01001 Rev. Date: 22 Oct. 2013 Manual No. Z1001 Version: 1.16 For General Laboratory Use Only Wellkang Limited

More information

X940e, X945e. Maintenance Guide

X940e, X945e. Maintenance Guide X940e, X945e Maintenance Guide July 2010 www.lexmark.com Contents...3 Storing supplies...3 Checking the status of supplies...3 Checking the status of supplies from the control panel... 3 Printing a menu

More information

Datacard CR500 Instant Issuance System. User Reference Guide. July Rev B

Datacard CR500 Instant Issuance System. User Reference Guide. July Rev B Datacard CR500 Instant Issuance System User Reference Guide July 2015 527495-001 Rev B Datacard CR500 User Reference Guide The CR500 Instant Issuance System Contents The CR500 Instant Issuance System The

More information

New Ricoh 1060 Copier. August 2004

New Ricoh 1060 Copier. August 2004 New Ricoh 1060 Copier August 2004 Step By Step Turn Main Power Switch ON Press operation switch to make ON indicator light up You may turn just the operation switch off at night DO NOT turn main power

More information

USING ABSOLUTE BLACK INKS FOR MAKING SCREEN POSITIVES ON THE EPSON STYLUS PRO 7700, 7890, 7900, 9700, 9890, & 9900

USING ABSOLUTE BLACK INKS FOR MAKING SCREEN POSITIVES ON THE EPSON STYLUS PRO 7700, 7890, 7900, 9700, 9890, & 9900 USING ABSOLUTE BLACK INKS FOR MAKING SCREEN POSITIVES ON THE EPSON STYLUS PRO 7700, 7890, 7900, 9700, 9890, & 9900 The following instructions explain how the i2i Absolute Black ink for making screen positives

More information

Ink Cartridge and Print Head Maintenance Manual

Ink Cartridge and Print Head Maintenance Manual Ink Cartridge and Print Head Maintenance Manual Hewlett Packard D135/145/155/175 CP1160/Color Copier 610 And other printers with similar cartridge design. Contents Page 3 4 5 6 7 8 9 10 12 15 21 25 26

More information

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10 Karl Suss SB6 Bonder Operation Tool Id: 119.020 The technical content of this document is the responsibility of UCLA CNSI Integrated Systems Nanofabrication Cleanroom Los Angeles, CA Karl Suss SB6 Bonder

More information

REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800

REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800 REFILLABLE CARTRIDGE SYSTEM FOR THE EPSON STYLUS PRO 7800 & 9800 Thank you for purchasing an Ink2image bulk ink feed system. You have just taken the first step to enjoying the cost saving and superior

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Page 1 of 43 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the DCA 2000 Analyser in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University

More information

HARMONi G3. Quick Start Guide for HARMONi G3. imac Processor/FireWire Upgrade

HARMONi G3. Quick Start Guide for HARMONi G3. imac Processor/FireWire Upgrade HARMONi G3 imac Processor/FireWire Upgrade imac and Operating System Compatibility The HARMONi G3 imac processor/firewire upgrade is compatible only with imac 233, 266, and 333 MHz models (Revisions A-D);

More information

PENfriend2 labelling PEN

PENfriend2 labelling PEN PENfriend2 labelling PEN DL110 Please retain these instructions for future reference. General description Record information onto a self-adhesive voice label and attach to a wide range of items in and

More information

9 Maintenance and Repair

9 Maintenance and Repair Agilent 1200 Infinity II Series RID User Manual 9 Maintenance and Repair Introduction to Maintenance 120 Warnings and Cautions 121 Overview of Maintenance 123 Cleaning the Module 124 Storage of the Detector

More information

Upgrading LVDS Cables Instruction Sheet

Upgrading LVDS Cables Instruction Sheet Upgrading LVDS Cables Instruction Sheet INTRODUCTION Use the following instructions to replace the LVDS cables in CP2000-M/MR projectors. The new cables are slightly longer in length and allow for better

More information

Portal Access Software Technical Manual

Portal Access Software Technical Manual Portal Access Software Technical Manual Instructions for using the Portal Access Software with the Maxprep, Maxwell RSC and Maxwell RSC 48 Instruments TM549 Revised 1/19 Table of Contents 1 Description...1

More information

C935 Series. Maintenance Guide

C935 Series. Maintenance Guide C935 Series Maintenance Guide April 2007 www.lexmark.com Contents...3 Cleaning the exterior of the printer...3 Storing supplies...3 Conserving supplies...3 Checking the status of supplies...4 Checking

More information

Demonstrated Protocol: How to Set Up the Ion OneTouch 2 Instrument When Switching Between the Ion PGM System and Ion Proton System

Demonstrated Protocol: How to Set Up the Ion OneTouch 2 Instrument When Switching Between the Ion PGM System and Ion Proton System USER BULLETIN Demonstrated Protocol: How to Set Up the Ion OneTouch 2 Instrument When Switching Between the Ion PGM System and Ion Proton System Publication Number MAN0007719 Revision 2.0 Overview.............................................................

More information

Replacing the RAID Battery Backup Unit Assembly on Series 3 FireSIGHT 3500 Defense Centers, Version 5.x

Replacing the RAID Battery Backup Unit Assembly on Series 3 FireSIGHT 3500 Defense Centers, Version 5.x Replacing the RAID Battery Backup Unit Assembly on Series 3 FireSIGHT 3500 Defense Centers, Version 5.x Last Updated: December 4, 2014 Use these instructions to replace the RAID battery backup unit (BBU)

More information

Precautions The XRD uses x-ray radiation. The instrument is designed to prevent radiation exposure.

Precautions The XRD uses x-ray radiation. The instrument is designed to prevent radiation exposure. 1 Operating Procedure for X-Ray Diffraction Purpose and Scope This document describes the procedures and policies for using the MSE Bruker D8 Focus XRD The scope of this document is to establish user procedures.

More information

PROCEDURE FOR TENSILE TESTING OF REINFORCING BARS

PROCEDURE FOR TENSILE TESTING OF REINFORCING BARS 1. PROCEDURE OVERVIEW This FSEL test standard is used to determine the tensile properties of reinforcing bars. The testing is in general conformance with ASTM A370-15 Standard Test Methods and Definition

More information

Instructions for SVC-KIT-0020

Instructions for SVC-KIT-0020 Kaleidescape, Inc. July 22, 2010 Instructions for SVC-KIT-0020 Title Time to complete 1U Server Power Supply Replacement 1 hour Procedure to complete Locate Parts and Tools Service Kit Parts Power supply

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Operating & Maintaining the Connex500/350 3-D Printer

Operating & Maintaining the Connex500/350 3-D Printer Operating & Maintaining the Connex500/350 3-D Printer Starting the Connex Printer... 2 Loading Model and Support Cartridges... 4 Producing Models... 5 Printer Interface Color Key... 7 Printing Indicators...

More information

120Ra-1 Pentium III Processor Installation Insert

120Ra-1 Pentium III Processor Installation Insert 120Ra-1 Pentium III Processor Installation Insert PN: 455-01614-000 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,

More information

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC MultiTherm Shaker Operations Manual Item: H5000-H and H5000-HC Foreword Thank you for purchasing a MultiTherm Shaker. This operations manual contains instructions for the care of this Instrument. In order

More information

QIAsymphony SP/AS Consolidated Operating Guide

QIAsymphony SP/AS Consolidated Operating Guide December 2017 QIAsymphony SP/AS Consolidated Operating Guide For use with software versions 4.0 and 4.1 Version 1 R2 1112127 QIAGEN GmbH, QIAGEN Strasse 1, 40724 Hilden GERMANY Sample to Insight Contents

More information

Replacing/Upgrading Mac Pro Processors (Early 2008)

Replacing/Upgrading Mac Pro Processors (Early 2008) Replacing/Upgrading Mac Pro Processors (Early 2008) The steps in this procedure illustrate how to remove the lower processor (CPU B). The instructions are the same for removing the upper processor (CPU

More information

Solana User Manual 1

Solana User Manual 1 Solana User Manual 1 Quidel Corporation Worldwide Headquarters 12544 High Bluff Drive San Diego, CA 92130 United States quidel.com 858.552.1100 MDSS GmbH Schiffgraben 41 30175 Hannover Germany 2 Table

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Automatic Tissue Processor MTP INS5500GB 2011-07-08 Instructions MTP 1 Instructions MTP 2 CONTENTS 1. INTENDED USE... 4 2. SYMBOLS... 4 3. SAFETY NOTES... 4 4. COMPONENTS... 5 5.

More information

Removal and Installation8

Removal and Installation8 8 Screw Types 8-4 Top Cover Assembly 8-5 Left Hand Cover 8-6 Right Hand Cover 8-10 Front Panel Assembly 8-14 Left Rear Cover 8-15 Right Rear Cover 8-16 Extension Cover (60" Model only) 8-17 Media Lever

More information

Open Desktop WorkStation System Components Items Labeled in RED are ESD Sensitive

Open Desktop WorkStation System Components Items Labeled in RED are ESD Sensitive Page 1 of 21 Open Desktop WorkStation System Components Items Labeled in RED are ESD Sensitive Back Plane Pegasos IDE Cable Hard Drive Video Card G4 CPU IDE Cable CDROM Page 2 of 21 Remove top of case

More information

User Manual. Trinocular Stereo Microscope

User Manual. Trinocular Stereo Microscope User Manual Trinocular Stereo Microscope Model V434 XV434 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Component Illustration... 3 2. Installation 4 3. Operation....6

More information

These items can be purchased separately or as a kit from your distributor.

These items can be purchased separately or as a kit from your distributor. Unpack Your Printer 1 Unpack Your Printer The following items are included in the standard BMP 53 Mobile Printer package: BMP 53 printer MC-1500-595-WT-BK cartridge USB connection cable Product CD Brady

More information

Customer Upgrade Procedure

Customer Upgrade Procedure Customer Upgrade Dell EMC Unity Family Dell EMC Unity All Flash and Unity Hybrid Adding an optional 80-drive DAE disk drive 302-003-780 REV 01 July 2017 This document describes how to add an optional 80-drive

More information

Quick Solutions Packet

Quick Solutions Packet Pub. No. J33039_EN 2008-11-15 Quick Solutions Packet ORTHO AutoVue Innova ORTHO AutoVue Ultra Topic No. Title 1 Interrupted Liquid Flow Error (ASPM255) 2 Replace the Air Filter 3 Defragment the Hard Drive

More information

Lexmark X203n and X204n Series. Maintenance Guide

Lexmark X203n and X204n Series. Maintenance Guide Lexmark X203n and X204n Series Maintenance Guide February 2011 www.lexmark.com Contents...3 Cleaning the exterior of the printer...3 Cleaning the scanner glass...4 Cleaning the ADF separator rollers...5

More information

Reagent Transfer User Guide

Reagent Transfer User Guide Reagent Transfer User Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 Utility description on page 3 Before you start

More information

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory 6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory Overvoltage Category II Pollution Degree 2 Equipment Class III Safety information Introduction Before using this accessory, you must read

More information

Dell Inspiron XPS and Inspiron 9100 Service Manual

Dell Inspiron XPS and Inspiron 9100 Service Manual Dell Inspiron XPS and Inspiron 9100 Service Manual Dell Inspiron XPS and Inspiron 9100 Service Manual Before You Begin Memory Module, Mini PCI Card, and Devices System Components Subwoofer Bluetooth Card

More information

RISC OS 3 software upgrade fitting instructions

RISC OS 3 software upgrade fitting instructions RISC OS 3 software upgrade fitting instructions Contents Page number Introduction 2 Packing list 2 An overview of the software upgrade procedure 4 Identifying your upgrade 4 Upgrading your computers hardware

More information

Customer Replacement Procedure

Customer Replacement Procedure Customer Replacement Dell EMC Unity Family Dell EMC Unity All Flash and Unity Hybrid Replacing a faulted 80-drive DAE cooling module 302-003-777 REV 01 July 2017 This document describes how to replace

More information

Insulin Autoantibody Assays

Insulin Autoantibody Assays Insulin Autoantibody Assays 24 May 2011 Page 1 of 13 Contents Initial dilution of Insulin Label.. 3 IAA Plating.....4 ciaa Plating... 5 Diluting IAA Label........6 Diluting ciaa Label...... 7 Washing PAS,

More information

1. Review the printer safety information in the Regulatory Compliance document provided with your printer.

1. Review the printer safety information in the Regulatory Compliance document provided with your printer. QUICK REFERENCE This Quick Reference contains supply loading information and general care and maintenance procedures for the Monarch Pathfinder Ultra Platinum 6039 printer. For more detailed information,

More information

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System Page 1 of 9 Approvals Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 18DEC14 Date: 18DEC14 1. Purpose: Operation of the BioLogic LP Chromatography System. 2. Scope: Applies to the BioLogic

More information

ft Aspen Fir Item # Commercial ITEM Lights

ft Aspen Fir Item # Commercial ITEM Lights 7.57.5 ft Aspen Fir Item #142027 Commercial ITEM Lights PARTS LIST A. BOTTOM B. LOWER MIDDLE C. UPPER MIDDLE D. TOP E. TREE STAND NUMBER OF PERSONS RECOMMENDED FOR ASSEMBLY: 2 SKU 142027 7.5 ASPEN FIR

More information

Automated Liquid Handling. Smart Benchtop System for Your Lab GeneTheatre

Automated Liquid Handling. Smart Benchtop System for Your Lab GeneTheatre Automated Liquid Handling Smart Benchtop System for Your Lab 2 3 Automated pipetting routines: easy and fast. The use of the greatly simplifies all pipetting routines within the low to medium throughput

More information

Reconstitution Application Guide

Reconstitution Application Guide Reconstitution Application Guide For Research Use Only. Not for use in diagnostic procedures. This guide contains the following topics: About this guide on page 2 App description on page 3 Before you start

More information

Nanosight NS300 NBTC User Instructions

Nanosight NS300 NBTC User Instructions Nanosight NS300 NBTC User Instructions Preliminary notes: -The laser module is delicate and expensive. Handle it with care and do not leave it on the bench unattended. Only leave it in place in the instrument

More information

GPS163R CD MW/FM Stereo Cassette Player/Recorder

GPS163R CD MW/FM Stereo Cassette Player/Recorder GPS163R CD MW/FM Stereo Cassette Player/Recorder INSTRUCTION MANUAL Before operating this unit, please read this Owner s Manual carefully WARNING: UNDER NO CIRCUMSTANCES MUST THE LIVE OR NEUTRAL WIRES

More information

Automated Liquid Handling. Smart Benchtop System for Your Lab GeneTheatre

Automated Liquid Handling. Smart Benchtop System for Your Lab GeneTheatre Automated Liquid Handling Smart Benchtop System for Your Lab GeneTheatre 2 GeneTheatre Automated pipetting routines: easy and fast. The use of the GeneTheatre greatly simplifies all pipetting routines

More information

1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd 2

1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd 2 Pipette Controller Operator's Manual En ligne; Manuel de l opérateur Online; Bedienungshandbuch En línea; Manual del operador In linea; Manuale d uso www.argos-tech.com RoHS 1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd

More information

Ultrospec 10 User Manual. English. Deutsch. Français. Español. Italiano

Ultrospec 10 User Manual. English. Deutsch. Français. Español. Italiano English Deutsch Ultrospec 10 User Manual Français Español Italiano Biochrom US Telephone: 1-508-893-8999 84 October Hill Road Toll Free: 1-800-272-2775 Holliston, MA Fax: 1-508-429-5732 01746-1388 support@hbiosci.com

More information

Maxwell RSC Tablet PC Configuration Manual for use with Windows 8 Operating System

Maxwell RSC Tablet PC Configuration Manual for use with Windows 8 Operating System Maxwell RSC Tablet PC Configuration Manual for use with Windows 8 Operating System TM537 1/18 Table of Contents 1 Maxwell RSC Instrument and Tablet PC...1 2 Glossary...3 3 Adjusting Volume Settings...5

More information

QL5-EX Quick Start-up Guide

QL5-EX Quick Start-up Guide QL5-EX Quick Start-up Guide 1. UNPACK AND INSPECT NOTE: When unpacking, check each component for damage. If you find evidence of damage, inform the ECI Technology Customer Service Representative about

More information

CF3000 Dealer Diagnostic Tool Instruction Manual

CF3000 Dealer Diagnostic Tool Instruction Manual CF3000 Dealer Diagnostic Tool Instruction Manual Table of Contents: About the CF3000......3 Important Precautions......4 Components....5 Charging the CF3000......7 Licensing the CF3000.......8 Updating

More information

& Technical Specifications

& Technical Specifications User Manual & Technical Specifications User manual Contents Pidion BM-170 Technical specifications... 2 Micro Rolltalk basic package... 3 Micro Rolltalk functions and buttons... 3 Preparing Micro Rolltalk...

More information

Lexmark C54x Maintenance Guide

Lexmark C54x Maintenance Guide Lexmark C54x Maintenance Guide July 2009 www.lexmark.com Contents...3 Storing supplies...3 Checking the status of supplies...3 Checking the status of supplies from the printer control panel...3 Checking

More information

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY Page: 1 of 8 1. Risk Assessment: This Risk Assessment is to be used as a general guide and as such, cannot accommodate all the varying factors

More information

The Olis Titrator Series

The Olis Titrator Series The Olis Titrator Series Perfect Titrations into any type of cuvette With considerable effort, one can do precise titration studies by hand. Or, one can let the OLIS titrators do the technician s work

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

Katana CD Player User Manual

Katana CD Player User Manual Katana CD Player User Manual Contents General section Introduction Katana A.1 Introduction A.2 Features A.3 Installation Precautions CD placement B.1 Connections Connections options C.1 Playback CD operation

More information

Sun StorageTek SL48 Tape Library Getting started

Sun StorageTek SL48 Tape Library Getting started Sun StorageTek SL48 Tape Library Getting started IMPORTANT: Shipping Lock: The shippinglockmustberemovedfor the robotics to work properly. A robot move error is displayed if the shipping lock is not removed.

More information

Aquamatic 5200 FARM Operation Manual

Aquamatic 5200 FARM Operation Manual Operation Manual AM 5200 FARM - 26714v 1.0 August 2013 Aquamatic 5200 FARM Operation Manual 1 (31) 1 Safety 1.1 Warning Ensure that the transport locks are disengaged before switching the instrument on.

More information

HT100 HACCP Touch Manual

HT100 HACCP Touch Manual HT100 HACCP Touch Manual Contents Page Section Title 2 Section 1.0 Introduction and Set Up 3 Section 1.1 Running the Software 4 Section 2.0 Database Management 7 Section 3.0 Supplier Management 9 Section

More information

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system.

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system. PriorLux 1 1. Introduction The PriorLab and PriorLux upright compound microscopes are high quality instruments equipped with high resolution, chromatically corrected optics for excellent image quality.

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

Inspiron Service Manual. 2-in-1. Computer Model: Inspiron Regulatory Model: P69G Regulatory Type: P69G001

Inspiron Service Manual. 2-in-1. Computer Model: Inspiron Regulatory Model: P69G Regulatory Type: P69G001 Inspiron 13 5000 2-in-1 Service Manual Computer Model: Inspiron 13-5378 Regulatory Model: P69G Regulatory Type: P69G001 Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps

More information

TONER CARTRIDGE REMANUFACTURING INSTRUCTIONS HP LASERJET 4600 TONER CARTRIDGE

TONER CARTRIDGE REMANUFACTURING INSTRUCTIONS HP LASERJET 4600 TONER CARTRIDGE HP LASERJET 4600 TONER CARTRIDGE REMANUFACTURING INSTRUCTIONS HP LASERJET 4600 TONER CARTRIDGE REMANUFACTURING THE HP LASERJET 4600 TONER CARTRIDGE By Javier Gonzalez and the Technical Staff at UniNet

More information

BD FACSPresto Near-Patient CD4 Counter

BD FACSPresto Near-Patient CD4 Counter BD FACSPresto Near-Patient CD4 Counter 651000 Instructions For Use IVD 23-12808-02 11/2014 EC REP Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA Benex Limited Pottery

More information

TrendView recorders - Installation Instruction

TrendView recorders - Installation Instruction TrendView recorders - Installation Instruction Analogue In / Analogue Out / Pulse Input cards / Alarm Relay and Digital Input / Output cards QX Recorder SX Recorder QXe Recorder DR Graphic Recorder This

More information

2. REFERENCE DOCUMENTS 90R0047 Safety Guidelines for Handling Equipment 90F0112 Safe Handling and Cleaning Procedures for Serviceable ZOLL Product

2. REFERENCE DOCUMENTS 90R0047 Safety Guidelines for Handling Equipment 90F0112 Safe Handling and Cleaning Procedures for Serviceable ZOLL Product ZOLL Document Number: 90R0008 Page 3 of 11 1. INTRODUCTION This document defines the handling and cleaning process for WCD LifeVest System Hardware returned from the field. This document also defines acceptable

More information

Moving your printer 1. Remove cables. Printing Moving the printer. Using color. Paper handling. 1 Turn the printer off. Maintenance.

Moving your printer 1. Remove cables. Printing Moving the printer. Using color. Paper handling. 1 Turn the printer off. Maintenance. Moving the printer Moving your printer 1 If you re going to move the printer, you must remove the supplies and attached options to prevent damage to the printer. Use the following steps to remove the options

More information

Copyright 2012 i-sens, Inc. All Rights Reserved

Copyright 2012 i-sens, Inc. All Rights Reserved i-sens, Inc. 465-6 Wolgye4-dong, Nowon-gu, Seoul, 139-845, Korea Medical Technology Promedt Consulting GmbH, Altenhofstrasse 80, D-66386 St. Ingbert, Germany Copyright 2012 i-sens, Inc. All Rights Reserved

More information

CAVS USA Inc. Quick Start Manual For Karaoke Jukebox CAVS JB-99RX. Version and up

CAVS USA Inc. Quick Start Manual For Karaoke Jukebox CAVS JB-99RX. Version and up Quick Start Manual For Karaoke Jukebox CAVS JB-99RX Version 1.0.0.261 and up 1 Contents Page 1. Unpacking 3 2. Installation 4 2.1 Touch Screen monitor 4 2.2 Mixer-Amplifier 5 2.3 Microphone 5 3. Booting

More information

Turntable player with USB/SD/PC recording function. Model no: TT620 User s Manual

Turntable player with USB/SD/PC recording function. Model no: TT620 User s Manual Turntable player with USB/SD/PC recording function Model no: TT620 User s Manual READ THIS INSTRUCTION CAREFULLY BEFORE USE AND STORE IN A SAFE PLACE FOR FUTURE REFERENCE 1 Features Two-built-in stereo

More information

Tecan crystallisation robot user guide

Tecan crystallisation robot user guide Tecan crystallisation robot user guide The TECAN crystallisation robot in RRB S9 provides an automated facility for setting up crystallisation trays. This user guide is meant as online support for trained

More information

Loading paper and specialty media Learning about the printer. Stop all printer activity. 7 Back button Return to the previous screen.

Loading paper and specialty media Learning about the printer. Stop all printer activity. 7 Back button Return to the previous screen. Quick Reference Use the 6 Stop or Cancel button To Stop all printer activity. Loading paper and specialty media Learning about the printer Using the printer control panel 0 Use the 9 To Display View the

More information

Upgrading and Servicing Guide

Upgrading and Servicing Guide Upgrading and Servicing Guide Copyright Information The only warranties for Hewlett-Packard products and services are set forth in the express statements accompanying such products and services. Nothing

More information

3M E-A-Rfit Validation System. Version M E-A-Rfit Validation System. Quick Guide Fit-Test Procedure for Non-Custom Earplugs

3M E-A-Rfit Validation System. Version M E-A-Rfit Validation System. Quick Guide Fit-Test Procedure for Non-Custom Earplugs 1 3M Occupational Health and Environmental Safety 3M E-A-Rfit Validation System 3M E-A-Rfit Validation System Version 4.2 Quick Guide Fit-Test Procedure for Non-Custom Earplugs 2 Purpose: The purpose of

More information

Swing-Arm Reading Lamp

Swing-Arm Reading Lamp Guardian-Rho OPERATING INSTRUCTIONS Swing-Arm Reading Lamp MODEL: L5210 thank you Thank you for the purchase of this quality LLOYTRON product. Used carefully and in accordance with the instructions enclosed,

More information

Replacing a 3.5-inch disk drive

Replacing a 3.5-inch disk drive EMC VNXe Replacing a 3.5-inch disk drive 302-000-213 REV 02 July, 2015 This document describes how to replace a disk drive in a supported 12-slot disk enclosure: 12-slot disk processor enclosure (DPE)

More information

Start Here. Remove all tape and lift display. Locate components USB

Start Here. Remove all tape and lift display. Locate components USB HP Photosmart 2600/2700 series all-in-one User Guide Start Here 1 USB Important: Do not connect the USB cable until this guide instructs you to or the software may not install properly. If you have problems

More information

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS Department of Microbiology Version: 1.0 CURRENT 1 of 28 Prepared by QA Committee Issued by: Laboratory Manager Revision Date: 10/2/2017 Approved by Laboratory Director: Annual Review Date: 5/1/2018 Microbiologist-in-Chief

More information

Confocal 6 Guide. Switching On Components of the system should be switched on in the following order: 1. Lasers Argon laser (488nm and 514nm)

Confocal 6 Guide. Switching On Components of the system should be switched on in the following order: 1. Lasers Argon laser (488nm and 514nm) Confocal 6 Guide The system is kept at 37⁰C. If you require the system at a lower temperature book at least one hour before the beginning of your session to allow the system to cool down. It is your responsibility

More information

Nuclease-free water (e.g. ThermoFisher, cat # AM9937) Freshly-prepared 70% ethanol in nucleasefree water (optional)

Nuclease-free water (e.g. ThermoFisher, cat # AM9937) Freshly-prepared 70% ethanol in nucleasefree water (optional) Before start checklist Materials Consumables Equipment Rapid Barcoding Sequencing Kit (SQK- RBK004) 1.5 ml Eppendorf DNA LoBind tubes Ice bucket with ice Flow Cell Priming Kit (EXP-FLP001) 0.2 ml thin-walled

More information