Verification measurements of an emc algorithm using a 2D ion chamber array

Similar documents
Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm

Clinical implementation of photon beam flatness measurements to verify beam quality

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

FAST, precise. qa software

UNCOMPROMISING QUALITY

Proton dose calculation algorithms and configuration data

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Transitioning from pencil beam to Monte Carlo for electron dose calculations

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

The IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved

VALIDATION OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR CLINICAL ELECTRON BEAMS IN THE PRESENCE OF PHANTOMS WITH COMPLEX HETEROGENEITIES

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Raising the Bar in IMRT QA

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS

Dynalog data tool for IMRT plan verification

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

Preface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Improvements in Monte Carlo simulation of large electron fields

An investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application

A Geant4 based treatment plan verification tool: commissioning of the LINAC model and DICOM-RT interface

Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Basic Radiation Oncology Physics

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Monte Carlo Treatment Planning: Implementation of Clinical Systems

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Basics of treatment planning II

Use of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Michael Speiser, Ph.D.

gpmc: GPU-Based Monte Carlo Dose Calculation for Proton Radiotherapy Xun Jia 8/7/2013

Monitor Unit (MU) Calculation

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

87.55.K-, Gh, km, Ca. Agility, analytic model, Elekta, Monaco, Monte Carlo, virtual source

Outline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications

Quick Reference Datasheet For All RIT113 Packages

A SYSTEM OF DOSIMETRIC CALCULATIONS

A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

DAILY LINAC QA BEAM QA

A software tool for the quantitative evaluation of 3D dose calculation algorithms

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

TPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

Code of practice: Create a verification plan for OCTAVIUS Detector 729 in Philips Pinnacle³

GafChromic Protocol Multi-Channel Film Dosimetry + Gamma Map Analysis

Calculating percent depth dose with the electron pencilbeam redefinition algorithm

Machine and Physics Data Guide

Advanced Radiotherapy

Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system

2D DOSE MEASUREMENT USING A FLAT PANEL EPID

Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator

Calibration and quality assurance for rounded leaf-end MLC systems

Source Model Tuning for a 6 MV Photon Beam used in Radiotherapy

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Photon Energy Spectrum Reconstruction Based on Monte Carlo and Measured Percentage Depth Dose in Accurate Radiotherapy

Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam

Investigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Air Kerma Strength Measurements and Monte Carlo based Dosimetric Characterization of a directional Pd 103 planar source array.

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Medical Dosimetry 37 (2012) Medical Dosimetry. journal homepage:

Commissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free beams

Release Notes for Dosimetry Check with Convolution-Superposition Collapsed Cone Algorithm (CC)

On compensator design for photon beam intensity-modulated conformal therapy

EXTERNAL PHOTON BEAMS: PHYSICAL ASPECTS

Implementation of Monte Carlo Dose calculation for CyberKnife treatment planning

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Verification of a commercial implementation of the Macro- Monte-Carlo electron dose calculation algorithm using the virtual accelerator approach

Use of 3D printers to create a patient-specific 3D bolus for external beam therapy

Suitability Study of MCNP Monte Carlo Program for Use in Medical Physics

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

Three-dimensional dosimetry of small fields

A Customized Bolus Produced Using a 3-Dimensional Printer for Radiotherapy

DUAL-ENERGY CT IN PROTON THERAPY

A fast and accurate GPU-based proton transport Monte Carlo simulation for validating proton therapy treatment plans

Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Verification measurements of an emc algorithm using a 2D ion chamber array Mark D. Wanklyn, 1a Ghirmay Kidane, 2 and Liz Crees 2 Medical Physics Department, 1 Guy s & St. Thomas NHS Foundation Trust, London,UK; Radiotherapy Physics, 2 Barking Havering & Redbridge University Hospitals NHS Trust, Romford, Essex, UK mark.wanklyn@nhs.net Received 20 October, 2015; accepted 7 June, 2016 The aim of this study was to assess the suitability of the Im RT MatriXX 2D ion chamber array for performing verification measurements on the Varian Eclipse electron Monte Carlo (emc) algorithm for a range of clinical energies (6, 12, and 20 MeV) on a Varian 2100iX linear accelerator. Firstly, the suitability of the MatriXX for measuring percentage depth doses (PDD) in water was assessed, including characterization of the inherent buildup found in the MatriXX. Secondly the suitability of the MatriXX for measuring dose distributions in homogeneous and heterogeneous phantoms was assessed using gamma analysis at 3%/3 mm. It was found that after adjusting the PDD curves for the inherent buildup, that the position of R 50,D measured using the MatriXX agreed to within 1 mm to the PDDs generated using the emc algorithm for all energies used in this study. Gamma analysis at 3%/3 mm showed very good agreement (> 95%) for all cases in both homogeneous and heterogeneous phantoms. It was concluded that the Im RT MatriXX is a suitable device for performing emc verification and could potentially be used for routine energy checks of electron beams. PACS number(s): 87.55.km, 87.55.Qr Key words: electron, Monte Carlo, emc, verification, ion chamber array I. INTRODUCTION The electron Monte Carlo (emc) algorithm in Eclipse v.11 (Varian Medical Systems, Palo Alto, CA) is a fast implementation of a Monte Carlo simulation of the dose deposition of highenergy electron beams. The emc algorithm consists of two models; the Initial Phase Space (IPS) model, (1) which models the electron beam emerging from the treatment head, and the Macro Monte Carlo (MMC) Transport model, (2) which calculates the dose deposition in the material as the electron travels through the material. The emc algorithm models each individual primary electron interaction along with secondary electron interactions, and for that reason can require both a large amount of computer processing power and long calculation times in order to achieve an accurate dose distribution. (3) The accuracy of the Eclipse emc algorithm has been evaluated in terms of optimal calculation parameters, (4) accuracy in heterogeneous and homogeneous phantoms, (5) radiochromic film verification, (6) and small field dose calculation. (7) Currently, radiochromic film is used to perform emc verification measurements and has been shown to agree within 2%/1 mm with measurements taken using an ionization chamber, and 2%/2 mm with an emc algorithm. (8) However, using radiochromic film can require significant time to prepare the film for use and a Corresponding author: Mark D. Wanklyn, 26B Browning Street, London, Greater London, United Kingdom, SE17 1LU; phone: (+44) 7920874385; email: mark.wanklyn@nhs.net 320 320

321 Wanklyn et al.: emc verification with 2D ion chamber array 321 to read out the resulting exposure. (9,10) There is also the requirement to calibrate each batch of film and the scanner, as well as to develop a suitable scanning protocol. Another problem with radiochromic film is that each sheet is single-use and therefore high costs can be incurred when using radiochromic film to perform emc verification measurements. Unlike radiochromic film, 2D ion chambers are multi-use, provide an instant readout which can be analyzed on set, and have a constant calibration factor that can be stored, removing the need for repeat calibrations. Most radiotherapy departments will already have a 2D ion chamber array available which is typically used to perform QA measurements, therefore there is no further cost incurred in performing emc verification measurements. This paper assesses the suitability of using a 2D ion chamber array to perform emc verification measurements in homogeneous and heterogeneous solid water phantoms. II. MATERIALS AND METHODS A. The Varian Eclipse emc algorithm In order to configure the IPS and transport models in the treatment planning system (TPS), measured beam data are required. This data include open-field (no electron applicator in place) and applicator measurements. The data required for the configuration of the emc algorithm are outlined in detail in the Varian Eclipse algorithm guide and in the literature. (11) The measurements required by the Eclipse treatment planning system include PDD curves in water for 6, 9, 12, 16, and 20 MeV with all available applicator sizes of 6 6, 6 10, 10 10, 15 15, and 20 20 cm 2. Additionally, point-dose measurements in water for all energy/applicator combinations listed and open-field (no applicator in place) measurements are required for the following: PDDs in water for all energies; in air profiles at 95 cm SAD and point-dose measurements in water for all energies at the reference depth. The configuration process then involves fitting monoenergetic beam data to the measured polyenergetic beam data provided. The emc algorithm will use the fitted monoenergetic beam data in all calculations. The emc algorithm has user-selectable calculation parameters which determine the overall accuracy of the calculation. These have been explored in the literature (4,11) and the parameters used in this study are shown in Table 1. These values are based upon the recommendations set in the paper by Zhang et al. (11) and allow for emc calculations to be performed in a clinically acceptable timeframe whilst maintaining accuracy to ensure a clinically relevant result. Table 1. Calculation parameters used throughout the study for all energies. These parameters allow for accurate dose calculations in a clinically acceptable time. Calculation Parameter Accuracy (%) 1 No. Particle Histories 0 Calculation grid size (cm) 0.2 Smoothing method 3D Gaussian Smoothing level Medium Random generator seed number Default Value

322 Wanklyn et al.: emc verification with 2D ion chamber array 322 B. The MatriXX 2D ion chamber array: The MatriXX I mrt (IBA Dosimetry, Schwarzenbruck, Germany) 2D ion chamber array is an array of 1,020 parallel plate chambers (0.08 cm 3 volume) arranged in a 32 32 grid with a detector center-to-center spacing of 7.62 mm, giving a maximum active area of 24.4 24.4cm 2. The MatriXX has an inherent buildup of 3 mm of Tecaran ABS plastic. The software used in conjunction with the MatriXX is Omnipro Im RT (IBA Dosimetry) The MatriXX was calibrated for absolute dose measurements for the range of electron energies used in this study (6, 12, 20 MeV). Restrictions in access to the linear accelerators limited the amount of data that could be collected and therefore these energies were chosen to represent the range of energies available in the clinic, whilst minimizing the need for linear accelerator time. The calibration was carried out by comparing the central-axis reading (interpolated from the readings of the four most central chambers of the MatriXX as it does not have a central axis chamber) on the MatriXX with the absolute dose measured using an NACP chamber. The absolute dose on the NACP chamber was measured for all energies at their respective reference depth (z ref ) in a solid water phantom, with applicator size 10 10cm 2 and 100 cm SSD, (z ref varies with energy, see Table 2). An off-axis calibration on the Im RT MatriXX was made to compensate for nonuniformity of off-axis chamber responses. Table 2. z ref values for each electron energy used in this study. Energy (MeV) Z ref (mm) 6 13.6 12 30.1 20 50.0 C. PDD verification using the MatriXX 2D ion chamber array C.1 Inherent buildup characterization The inherent build-up of the MatriXX 2D ion chamber array has an equivalent path length in water of 3.6 mm for photons. It has been shown that the path length in water of this inherent buildup varies with electron energy and beam quality. (12) For this study, PDD curves were measured for 6, 12, and 20 MeV, using the MatriXX with different thicknesses of solid water buildup (2 mm step size), and normalized to their maximum values. The position of R 100,D, R 80,D, R 50,D, and R 30,D, measured using the MatriXX, were compared to those from PDD curves obtained using an NACP chamber in water. Although the detectors in the MatriXX are not NACP chambers they are similar in design, albeit with a smaller active volume. An approximation to NACP chambers was made, and a solid water to water fluence ratio correction of 1.013 was applied to the depths measured in solid water. This fluence ratio correction was measured when absolute output was measured using NACP chambers in solid water. The scaling factor was derived to account for the difference in fluence between water and solid water.

323 Wanklyn et al.: emc verification with 2D ion chamber array 323 C.2 Comparing R50,D measured using the MatriXX 2D ion chamber array vs. calculated using emc The calculated shift due to the inherent buildup at each energy was applied to the PDD curve obtained using the MatriXX to allow a direct comparison to the PDD generated using emc. In order to obtain the PDD using emc, a plan was calculated on a solid water phantom using the calculation parameters shown in Table 1 and a 10 10 cm 2 applicator for each energy (6, 12, 20 MeV). The central depth-dose plane was exported to OmniPro Im RT and the profile along the center of the dose plane plotted. The depth of R 50,D measured using the MatriXX was compared to the R 50,D calculated using emc. D. Dose plane verification using the MatriXX 2D ion chamber array D.1 Homogeneous solid water phantom Dose planes were measured using the MatriXX in a simple homogeneous solid water phantom, with 5 cm solid water backscatter and 2 cm, 5 cm, 7 cm solid water buildup on top of the MatriXX for 6, 12, and 20 MeV, respectively. 100 monitor units (MUs) were delivered using a 10 10 cm 2 applicator at 100 cm SSD. Equivalent dose planes were planned and calculated using emc and were then exported to the OmniPro Im RT software. Using the gamma analysis tool within OmniPro Im RT at 3%/3 mm and a pass rate of 95% γ 1, the similarity between the emc calculated and MatriXX measured dose planes was assessed. The ROI where the gamma analysis was performed was the area encompassed by the 20% isodose. D.2 Heterogeneous solid water phantom (with lung insert) Dose planes were measured using the MatriXX with 5 cm solid water backscatter and a heterogeneous buildup created using the Standard Imaging (Ref 91235) (Standard Imaging Inc., Middleton, WI) lung phantom with an additional 1 cm solid water for all energies in this study. A 10 10cm 2 applicator at 100 cm SSD was used and 100 MU delivered. The solid water phantom was CT-scanned, imported into Eclipse and the corresponding dose planes calculated using emc. Gamma analysis was performed on the dose planes using the OmniPro Im RT software and the similarity of the dose planes assessed using a criterion of 3%/3 mm and a pass rate of 95% γ 1. The ROI where the gamma analysis was performed was the area encompassed by the 20% isodose.

324 Wanklyn et al.: emc verification with 2D ion chamber array 324 III. RESULTS AND DISCUSSION A. PDD verification using ion chamber array A.1 Inherent buildup characterization It can be seen from Fig. 1 that there is a shift between the PDD curve measured using the MatriXX compared to the PDD curve measured using the NACP chamber. The average difference in R 100,D, R 80,D, R 50,D, and R 30,D position (in mm) between the MatriXX measurements in solid water and the NACP measurements in water was calculated for each energy (6, 12, and 20 MeV) and applied as the inherent buildup to the PDD curve measured using the MatriXX. This shift was found to be energy dependent at 4.9 mm, 4.9 mm, and 5.1 mm for 6, 12, and 20 MeV, respectively. Fig. 1. Percentage depth-dose curves obtained using the MatriXX in solid water (with fluence ratio correction applied) and NACP chamber in water for 6, 12, and 20 MeV showing a shift in PDD curve. The zero position of the MatriXX is not a real zero position but is used to show 0 cm solid water buildup used.

325 Wanklyn et al.: emc verification with 2D ion chamber array 325 A.2 Comparing R50,D measured using the MatriXX 2D ion chamber array vs. calculated using emc The depth of R 50,D was measured from the curves shown in Fig. 2. It was found that the position of R 50,D measured using the MatriXX agreed to within 1 mm against emc values for all energies (-0.5 mm, 0.6 mm, and -0.9 mm for 6, 12, and 20 MeV, respectively). When compared to the commissioning values measured using an NACP chamber in water, it is found that the position of R 50,D measured using the MatriXX (with the inherent buildup correction applied) agrees to within 0.5 mm (-0.3 mm, -0.4 mm, and -0.4 mm for 6, 12, and 20 MeV, respectively). Fig. 2. Percentage depth-dose curves obtained using the MatriXX in solid water and the emc-calculated PDD curves in solid water at 6, 12, and 20 MeV. The energy-dependent inherent buildup shifts calculated are applied to the PDD curves obtained with the MatriXX.

326 Wanklyn et al.: emc verification with 2D ion chamber array 326 B. Dose plane verification using the MatriXX 2D ion chamber array Using the homogeneous and heterogeneous phantoms described above, dose planes were measured using the MatriXX and compared with the equivalent dose planes calculated using the Eclipse emc algorithm. Gamma maps can be seen in Fig. 3 for 6 MeV (3(a)), 12 MeV (3(b)), and 20 MeV (3(c)). The gamma analysis results (3%/3 mm) are shown in Table 3. Fig. 3. Gamma map comparing the dose distributions at 6 (a), 12 (b), and 20 (c) MeV in a homogeneous solid water phantom (left) and a heterogeneous solid water phantom with a lung insert (right).

327 Wanklyn et al.: emc verification with 2D ion chamber array 327 Table 3. Gamma analysis results at 3%/3 mm and 2%/2 mm comparing the dose planes measured using the with the dose planes calculated by the emc algorithm in Eclipse. 3%/3 mm 2%/2 mm Heterogeneous Heterogeneous Solid Water Solid Water Homogeneous Phantom Homogeneous Phantom Energy Solid Water (with lung Solid Water (with lung (MeV) Phantom insert) Phantom insert) 6 97.0% 98.9% 90.1% 92.4% 12 99.3% 98.4% 95.7% 95.1% 20 100% 99.1% 99.3% 92.4% It can be seen from Table 3 that the gamma analysis results at 3%/3 mm show very good agreement between the dose distributions measured using the MatriXX and those calculated using emc, for both heterogeneous and homogeneous phantoms. However at 2%/2 mm, which is the level at which the Gafchromic film shows good agreement with the emc algorithm, due to the limitation of the spatial resolution of the MatriXX it is evident that there isn t as good agreement. These results demonstrate that the MatriXX does not perform as well as the Gafchromic film in verifying dose distributions calculated by the emc algorithm. The gamma map shown in Fig. 3(a) for the 6 MeV solid water map, shows a large area where the gamma analysis > 1. It was found upon inspection of the solid water used, that there was a defect in the solid water slab which meant that there was a large disagreement in the area of the defect. This isn t seen in all of the dose maps as the solid water phantom was made up of smaller slabs and only the very superficial slab had the defect. It can also be seen from Table 3 that the percentage of points passing the gamma analysis is poorer for lower energies. This could be due to the depth at which the measurement was taken and a lack of scatter due to the shallow depth of measurement. IV. CONCLUSION Once suitably calibrated with the inherent buildup characterized, the MatriXX Im RT 2D ion chamber array has been shown to be suitable for performing the verification measurements required to assess the functionality of the Eclipse emc algorithm. However, it does not perform as well as Gafchromic film when verifying dose distributions in homogeneous and heterogeneous phantoms in standard setups. Limitation in the spatial resolution of the MatriXX may have caused poor pass rate for the criterion of 2%/2 mm. For this reason it is suggested that Gafchromic film remains the gold standard for TPS commissioning verification measurements, but in centers which may not have a Gafchromic service available to them, the results show that the MatriXX can provide a suitable alternative to perform the measurements required to verify the emc algorithm. The ease of setup and use makes the MatriXX suitable not only for initial verification of the emc algorithm but also for periodic QA checks, as well as verification measurements following a TPS upgrade. ACKNOWLEDGMENTS Mark D. Wanklyn would like to thank all of the staff at Queen s Hospital, Romford, for all of their help and support throughout his time there, without which this publication would not have been possible.

328 Wanklyn et al.: emc verification with 2D ion chamber array 328 COPYRIGHT This work is licensed under a Creative Commons Attribution 3.0 Unported License. REFERENCES 1. Janssen JJ, Korevaar EW, van Buttum LJ, Storchi PR, Huizenga H. A model to determine the initial phase space of a clinical electron beam from measured beam data. Phys Med Biol. 2001;46(2):269 86. 2. Neuenschwander H, Mackie TR, Reckwerdt PJ. MMC a high-performance Monte Carlo code for electron beam treatment planning. Phys Med Biol. 1995;40(4):543 74. 3. Fragoso M, Pillai S, Solberg TD, Chetty IJ. Experimental verification and clinical implementation of a commercial Monte Carlo electron beam dose calculation algorithm. Med Phys. 2008;35(3):1028 38. 4. Popple RA, Weinberg R, Antolak JA, et al. Comprehensive evaluation of a commercial macro Monte Carlo electron dose calculation implementation using a standard verification data set. Med Phys. 2006;33(6):1540 51. 5. Ding GX, Duggan DM, Coffey CW, Shokrani P, Cygler JE. First macro Monte Carlo based commercial dose calculation module for electron beam treatment planning new issues for clinical consideration. Phys Med Biol. 2006;51(11):2781 99. 6. Hu YA, Song H, Chen Z, Zhou S, Yin F. Evaluation of an electron Monte Carlo dose calculation algorithm for electron beam. J Appl Clin Med Phys. 2008;9(3):1 14. 7. Xu Z, Walsh SE, Telivala TP, Meek AG, Yang G. Evaluation of the eclipse electron Monte Carlo dose calculation for small fields. J Appl Clin Med Phys. 2009; 10(3):75 85. 8. Chan E, Lydon J, Kron T. On the use of Gafchromic EBT3 films for validating a commercial electron Monte Carlo dose calculation algorithm. Phys Med Biol. 2015;60(5):2091 102. 9. Niroomand-Rad A, Blackwell CR, Coursey BM, et al. AAPM Report No.63 Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys. 1998;25(11):2093 115. 10. Aubry JF, Bouchard H, Bessières I, Lacroix F. Validation of an electron Monte Carlo dose calculation algorithm in the presence of heterogeneities using EGSnrc and radiochromic film measurements. J Appl Clin Med Phys. 2011;12(4):3392. 11. Zhang A, Wen N, Nurushev T, Burmeister J, Chetty IJ. Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm. J Appl Clin Med Phys. 2013;14(2):4062. 12. Zhang M, Li S, Deng H, Zhou S. Technical Note: the uses of I mrt MatriXX in electron beams. IJMPCERO. 2013;2(1):15 18.