An experimental evaluation of the Agility MLC for motion-compensated VMAT delivery

Similar documents
An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking

Experimental investigation of a general real-time 3D target localization method using sequential kv imaging combined with respiratory monitoring

Using a research real-time control interface to go beyond dynamic MLC tracking

doi: /j.ijrobp

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images

Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: Implications for 4D radiotherapy

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

Dynalog data tool for IMRT plan verification

Electromagnetic Detection and Real-Time DMLC Adaptation to Target Rotation During Radiotherapy

A finite state model for respiratory motion analysis in image guided radiation therapy

Tumor motion during liver SBRT

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

Image Guidance and Beam Level Imaging in Digital Linacs

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

The MSKCC Approach to IMRT. Outline

Calibration and quality assurance for rounded leaf-end MLC systems

State-of-the-Art IGRT

Clinical implementation of photon beam flatness measurements to verify beam quality

Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG)

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Online dose reconstruction for tracked volumetric arc therapy: Real-time implementation and offline quality assurance for prostate SBRT

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Analysis of RapidArc optimization strategies using objective function values and dose-volume histograms

Radiation therapy treatment plan optimization

724. Simulation of the radiation therapy system for respiratory movement compensation

FAST, precise. qa software

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Basic Radiation Oncology Physics

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study

Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion

7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

Computational modelling and analysis: IOP Medical Physics Group

Takayama, Kenji; Mizowaki, Takashi; Nagano, Kazuo; Kamino, Yuichiro; Hi. Rightc 2009 Elsevier Ireland Ltd.

Agility MLC transmission optimization in the Monaco treatment planning system

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Evaluation of mechanical accuracy for couch-based tracking system (CBTS)

VMAT for dummies: Concepts, Clinical Implementation and Treatment Planning. Rajat Kudchadker, Ph.D. Associate Professor

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

arxiv: v1 [physics.med-ph] 26 Oct 2009

Brilliance CT Big Bore.

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Estimating 3D Respiratory Motion from Orbiting Views

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

Dosimetric impact of the 160 MLC on head and neck IMRT treatments

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Raising the Bar in IMRT QA

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

Real-time monitoring of linac performance using RT plans and logfiles

Implementation of Filtered back Projection (FBP) Theory for Intensity Modulated Radiation Therapy (IMRT) Planning

ADVANCING CANCER TREATMENT

arxiv: v3 [physics.med-ph] 6 Dec 2011

Calibration of Video Cameras to the Coordinate System of a Radiation Therapy Treatment Machine

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images

Determination of rotations in three dimensions using two-dimensional portal image registration

TPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1

Lag correction model and ghosting analysis for an indirect-conversion flat-panel imager

Reconstruction of 4D CT data sets acquired during free breathing for the analysis of respiratory motion

Photon beam dose distributions in 2D

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Machine and Physics Data Guide

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Interactive Treatment Planning in Cancer Radiotherapy

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

Trajectory optimization for dynamic couch rotation during volumetric modulated arc radiotherapy

Medical Dosimetry 37 (2012) Medical Dosimetry. journal homepage:

BOSTON UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES. Thesis REAL-TIME 4D TUMOR TRACKING AND MODELING FROM INTERNAL

Use of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles

ADVANCING CANCER TREATMENT

TEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( )

Advanced Radiotherapy

Chapter 9 Field Shaping: Scanning Beam

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

Analysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D) CT Images of Targets Undergoing Regular Motion

Dynamic Image Prediction Using Principal Component and Multi-Channel Singular Spectral Analysis: A Feasibility Study

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

1. Learn to incorporate QA for surface imaging

An Optimisation Model for Intensity Modulated Radiation Therapy

Real-Time 4D Tumor Tracking and Modeling from Internal and External Fiducials in Fluoroscopy

A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans

Comparison of MLC error sensitivity of various commercial devices for VMAT pre-treatment quality assurance

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Transcription:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 An experimental evaluation of the Agility MLC for motion-compensated VMAT delivery G A Davies, P Clowes, J L Bedford, P M Evans, S Webb and G Poludniowski Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Downs Road, Sutton, Surrey, SM2 5PT E-mail: gemma.davies@icr.ac.uk Abstract. An algorithm for dynamic multileaf-collimator (dmlc) tracking of a target performing a known a priori, rigid-body motion during volumetric modulated arc therapy (VMAT), has been experimentally validated and applied to investigate the potential of the Agility (Elekta AB, Stockholm, Sweden) multileaf-collimator for use in motion-compensated VMAT delivery. For five VMAT patients, dosimetric measurements were performed using the Delta 4 radiation detector (ScandiDos, Uppsala, Sweden) and the accuracy of dmlc tracking was evaluated using a gammaanalysis, with threshold levels of 3% for dose and 3 mm for distance-to-agreement. For a motion trajectory with components in two orthogonal directions, the mean gammaanalysis pass rate without tracking was found to be 58.0%, 59.0% and 60.9% and was increased to 89.1%, 88.3% and 93.1% with MLC tracking, for time periods of motion of 4 s, 6 s and 10 s respectively. Simulations were performed to compare the efficiency of the Agility multileaf-collimator with the MLCi multileaf-collimator when used for motion-compensated VMAT delivery for the same treatment plans and motion trajectories. Delivery time increases from a static-tumour to dmlc tracking VMAT delivery were observed in the range 0%-20% for the Agility, and 0%-57% with the MLCi, indicating that the increased leaf speed of the Agility MLC is beneficial for MLC tracking during lung radiotherapy. 26 27 28 29 30 31 32 33 34 35 36 1. Introduction Dynamic multileaf-collimator (dmlc) tracking as a research interest has led to the development of prototype real-time MLC tracking systems for both Siemens (Tacke et al. 2010, Krauss et al. 2011, Krauss et al. 2012) and Varian (Sawant et al. 2008, Keall et al. 2011, Poulsen et al. 2012a) linear accelerators. It is currently not possible to use Elekta linear accelerators for real-time dmlc tracking, however the feasibility of conformal dmlc tracking of a target performing a known a priori motion trajectory has previously been investigated by McQuaid et al. (2009). In the absence of a realtime MLC tracking system for Elekta linear accelerators, an algorithm for tracking motion known a priori has previously been presented (Davies et al. 2011). In the study presented in this paper, the algorithm is applied to and experimentally validated

MLC tracking with the Agility MLC 2 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 for the latest Elekta multileaf-collimator, Agility (Elekta AB, Stockholm, Sweden), for MLC tracking during volumetric modulated arc therapy (VMAT) (Otto 2008, Yu & Tang 2011). Few studies so far have investigated the capabilities of the Agility MLC (Cosgrove et al. 2009, Bedford et al. 2013) and the authors believe a study of its suitability for MLC tracking is timely. The Agility MLC has 160 leaves, of projected leaf width 5 mm at the isocentre, arranged in two banks of 80 leaves. Each bank of leaves is contained within a dynamic leaf guide (DLG), which can move with the MLC leaves. The maximum velocity of the individual MLC leaves is 35 mm s 1, and for the dynamic leaf guide, 30 mm s 1, therefore the maximum possible velocity, if both the dynamic leaf guide and the MLC leaves are moving in the same direction, is 65 mm s 1. There are two jaws that move perpendicular to the direction of MLC leaf travel, and these have a maximum velocity of 90 mm s 1. These increased leaf and jaw velocities compared with previous Elekta multileaf-collimator models, such as the MLCi with a maximum leaf speed of 20 mm s 1, could confer advantage for dmlc tracking during VMAT delivery, as is investigated in this paper. The aim of this study was to experimentally validate the algorithm presented by Davies et al. (2011) for its ability to successfully predict the machine parameters during static-tumour and MLC-tracking VMAT delivery with the Agility MLC, and then to apply this algorithm to investigate the potential of the Agility MLC for motion-compensated VMAT delivery. Dose measurements were performed to determine the dosimetric accuracy of this MLC-tracking technique using the Delta 4 diode-array phantom, which has previously been evaluated for IMRT and VMAT verification by Bedford et al. (2009). A motion-platform purpose-built in-house for translating the Delta 4 phantom in three-dimensions was used to simulate target motion. Additionally, through simulations, the delivery-time efficiency for tracking with the Agility multileafcollimator compared with the previous generation multileaf-collimator, the MLCi, was evaluated for the same VMAT plans and motion trajectories. 65 66 67 68 69 70 71 72 73 74 75 2. Method 2.1. VMAT treatment plans VMAT treatment plans for five lung cancer patients were prepared for the Agility MLC using AutoBeam v5.2, an in-house treatment planning system (Bedford 2009, Bedford 2013). The plans generated were conformal arcs, i.e. the MLC leaves conform to the planning target volume but the prescribed monitor units per segment (i.e. the interval between two neighbouring control points) vary. A case study of the VMAT planning process using AutoBeam has previously been presented (Bedford et al. 2008). Each plan had a control point spacing of 5, with a 350 arc ranging from 185 to 175 degrees. The collimator angle for each plan was 0. When implemented clinically, a collimator of angle of 2-5 degrees would be used, but this was removed for simplicity

MLC tracking with the Agility MLC 3 76 77 78 79 80 81 82 83 when transforming the MLC leaf positions to account for motion. A summary of the key parameters for each VMAT plan is given in table 1. In order to provide information regarding the complexity of each treatment plan, the minimum and maximum values of gantry speed, dose rate and prescribed monitor units per segment is presented. The variation in gantry speed and dose rate was measured during delivery of each treatment plan on the linear accelerator used for the experiments presented in this study. It should be noted that the nominal maximum gantry speed of the linear accelerator is 6.0 s 1, however the transient gantry speed sometimes exceeds this, as is observed in table 1. Table 1. Key parameters of the VMAT treatment plans used in this study. Values presented are for static-tumour, i.e. without motion-compensation, VMAT delivery. Range (min - max) Patient Prescribed Monitor MU per Gantry speed Dose rate number dose (Gy) units (MU) segment ( s 1 ) (MU s 1 ) 1 2.75 348.0 4.8-5.2 5.7-6.5 5.45-5.98 2 2.75 374.9 4.5-6.2 5.1-6.4 5.17-6.90 3 2.75 408.1 5.2-6.4 5.2-6.5 5.95-7.42 4 1.75 249.1 2.5-4.9 5.3-6.5 2.87-5.62 5 2.75 363.1 1.6-9.1 5.1-6.8 1.78-9.13 Measured data during delivery on linear accelerator. 84 85 86 2.2. Simulation of a respiratory motion trajectory A two-dimensional Lujan et al. (2003) respiratory-motion trajectory was modelled of the form: ( πt z target (t) = z 0 A z cos 6 τ + π ) 2 (1) 87 88 89 90 91 92 93 94 95 96 y target (t) = y 0 + A y cos 6 ( πt τ + 3π 5 where z target (t) is the motion along the z-axis and y target (t) is the motion along the y-axis. z 0 and y 0 are the rest positions of the target motion (taken to be zero in this study), A z is the amplitude of motion in the z direction and is equal to 20 mm and A y is the amplitude of motion in the y direction and is equal to 10 mm. The time period is given by τ, and was varied to be either 10 s, 6 s or 4 s in this set of experiments. A motion platform designed and built in-house for the translation of the Delta 4 phantom in three-dimensions, and accurate to within 0.2 mm using a Galil DMC-4060 (Galil Motion Control, Inc., Rocklin, CA, USA) motion controller, was programmed to move with the trajectory described by (1) and (2). The experimental set-up and the orientation of the axes of motion with respect to the linear accelerator are shown in figure 1. ) (2)

MLC tracking with the Agility MLC 4 Figure 1. Photograph of the experimental set-up: the motion platform carrying the Delta 4 is shown mounted on the linear accelerator couch. The orientation of the axes of motion are labelled in red. 97 98 99 100 101 102 103 104 105 106 107 108 2.3. MLC tracking algorithms A VMAT arc is defined as a series of N control points (n = 1, 2, 3,..., N), and a segment m is defined as the interval between two neighbouring control points, n and n + 1. The total number of segments (m = 1, 2, 3,..., M) is given by N 1. The algorithm presented by Davies et al. (2011) compensates for a priori known motion by transforming the planned MLC leaf positions at each control point, and iteratively reducing the dose rate of each segment of the VMAT arc until the MLC tracking plan is deliverable i.e. machine constraints such as maximum leaf velocity and maximum gantry velocity are not violated. In this study, the algorithm was applied to the Integrity 3.0 control system, in which the dose rate can vary by 256 levels, from 0 to the maximum. For motion parallel to the direction of leaf travel the positions of the k MLC leaves, at control point n, z kn, were transformed using: z k n = z kn + f n (3) 109 where for a collimator angle of 0, the correction required, f n is given by:

MLC tracking with the Agility MLC 5 Figure 2. Schematic illustrating the linear-interpolation technique for transforming the MLC aperture to account for motion perpendicular to the direction of leaf travel. 110 111 112 113 114 115 116 117 118 119 120 121 f n = y target (t = n 1 m=1 t m ) sin θ (4) where t m is the segment time and θ is the gantry angle, which varies from 0 to 359. For motion perpendicular to the direction of leaf travel, which was not considered in the initial study by Davies et al. (2011), the MLC leaf positions were transformed using a linear-interpolation technique. This involved refitting the MLC leaves to a new contour which was derived from transforming the static MLC aperture contour to account for motion perpendicular to the direction of leaf travel. In this study, the centre of each MLC leaf edge was fitted to its intercept with the new, transformed aperture contour; this is illustrated in figure 2. The contour shift at control point n, perpendicular to the direction of leaf travel, f n, is given by: f n = z target (t = n 1 m=1 t m ) The jaws (denoted Y1 and Y2) that move perpendicular to the direction of leaf travel were also transformed to account for motion using: (5) Y jaw n = Y jawn + f n (6) 122 123 124 125 126 127 An example of the aperture transformation, parallel and perpendicular to the direction of leaf travel, at a particular control point for patient 1 is shown in figure 3. It should be noted that in this study, the direction of motion of the MLC leaves was perpendicular to the motion axis with the largest amplitude, as given by (1). Clinical VMAT plans at our centre are planned in this way, with a collimator rotation close to 0. This has the advantage of maximising the number of leaves used to create the treatment

MLC tracking with the Agility MLC 6 128 129 130 131 132 133 134 135 136 137 138 139 aperture as the apertures are typically elongated in the z-direction and therefore the distance between adjacent MLC leaves at each control point is minimised. 2.4. Experimental validation of the MLC tracking algorithm with the Agility MLC The algorithm requires knowledge of certain machine parameters to calculate the statictumour VMAT and dmlc tracking VMAT delivery. These parameters are given in table 2. It should be noted that if all the MLC leaves are moving in the same direction, it is possible for the combined maximum velocity of the MLC leaves and the DLG to be 65 mm s 1. However, the main purpose of the dynamic leaf guide is to increase the travel range of the MLC leaves. Therefore, as the VMAT plans used in this study, both static and tracking, required leaf motion between control points typically less than or equal to 20 mm and of an oscillatory nature, the DLG remained stationary. Therefore the maximum leaf velocity was taken to be 35 mm s 1. Table 2. Machine constraints of the Elekta Synergy linear accelerator used for the measurements. Machine parameter Gantry speed ( s 1 ) 6 Gantry inertial compensation distance ( ) 5 Dose rate (MU s 1 ) 9.25 MLC leaf velocity (mm s 1 ) 35 DLG velocity (mm s 1 ) 30 Jaw velocity (mm s 1 ) 90 Nominal maximum value 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 To determine the ability of the algorithm to correctly predict the linear accelerator parameters during static-tumour and dmlc VMAT tracking delivery, a logging function on the Elekta linear accelerator control system was used to record the dose rate during each static-tumour and dmlc tracking VMAT delivery. The real-time delivery time, i.e. that measured from the logging function, was compared with the simulated delivery time and the time difference in seconds compared. Additionally, the simulated and real-time dose rates were compared as a function of monitor units for each plan. 2.5. Evaluation of the accuracy of dmlc tracking For each VMAT plan six dose distributions were measured with the Delta 4 phantom: (i) Static-tumour VMAT plan with static Delta 4 (ii) Static-tumour VMAT plan with motion platform translating Delta 4 in 1D (iii) Static-tumour VMAT plan with motion platform translating Delta 4 in 2D (iv) 1D MLC tracking VMAT plan with motion platform translating Delta 4 in 1D (v) 1D MLC tracking VMAT plan with motion platform translating Delta 4 in 2D (vi) 2D MLC tracking VMAT plan with motion platform translating Delta 4 in 2D

MLC tracking with the Agility MLC 7 100 80 60 40 Y2 Position (mm) 20 0 20 X1 X2 40 Y1 y x 60 Target 80 motion axes z 100 100 80 60 40 20 0 20 40 60 80 100 Position (mm) 100 80 (a) 60 Y2 40 Position (mm) 20 0 20 40 X1 y x 60 Target 80 motion axes z 100 100 80 60 40 20 0 20 40 60 80 100 Position (mm) Y1 X2 (b) Figure 3. A diagram showing the beam s-eye-view of the MLC aperture, at a gantry angle of 90, for patient 1. The MLC leaves are shown in blue, and the position of the jaws that move perpendicular to the direction of leaf travel, Y1 and Y2, are shown in red. (a) shows the MLC aperture for the static-tumour VMAT plan and (b) shows the MLC aperture for the 2D MLC tracking VMAT plan with 10 s time period, where the aperture has been transformed for motion along the z-axis and the y-axis. The axes of target motion are labelled in green. The Agility MLC has a total area of 400 mm 400 mm, but only the inner 200 mm 200 mm area is shown here.

MLC tracking with the Agility MLC 8 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 where static-tumour VMAT plans are those created by the treatment planning system, 1D MLC tracking VMAT plans are those where the MLC leaf positions have been adjusted to follow the primary axis of motion in this study (given by (1)), and 2D MLC tracking VMAT plans refer to plans following motion along both axes of motion (given by (1) and (2)). Dose distributions (ii) and (iv) were acquired with the motion platform translating the Delta 4 with the motion trajectory given by (1) and dose distributions (iii), (v) and (vi) were acquired with the motion platform translating the Delta 4 with the motion trajectory given by both (1) and (2). Dose distributions (ii) - (vi) were compared to the reference dose distribution (i) by means of a gamma analysis with a threshold of 3% and 3 mm. Detectors that were outside of the 5% isodose level for the reference dose distribution (dose distribution (i)) were not included in the gamma analysis. This was repeated for each of the three time periods of motion investigated in this study. For reference, the dose distribution calculated by the treatment planning system (TPS) for the Delta 4 geometry was also compared with dose distribution (i) and evaluated using a gamma analysis with threshold levels of 3% and 3 mm, excluding detectors outside of the 20% isodose level, as per clinical practice. There is a finite start-up time between the user initiating the VMAT arc on the control system, and the radiation beam. Therefore, in order to synchronise the movement of the Delta 4 phantom with gantry movement and thus MLC leaf positions, an optical gating interface was used. The start-up time was determined to be 1.5 s, and therefore as the user initiated the VMAT arc, the motion of the Delta 4 was triggered, but delayed by 1.5 s to remove this latency. 2.6. Evaluation of the delivery efficiency of MLC tracking with the Agility and MLCi multileaf-collimators Simulations were performed to compare the increase in delivery time from a statictumour VMAT delivery to a 1D and 2D MLC tracking delivery for the Agility MLC and the MLCi MLC. The MLCi has a lower maximum leaf velocity of 20 mm s 1, and a lower maximum jaw velocity. The MLCi jaws that move parallel to the direction of leaf travel have a maximum velocity of 20 mm s 1 and those that move perpendicular have a maximum velocity of 15 mm s 1. Another major difference is that the MLCi has a leaf width at isocentre of 10 mm, double that of the Agility. However, it was not possible to repeat the experiments performed with the Agility to determine the dosimetric effect of this increased leaf width, as a linear accelerator with an MLCi multileaf-collimator and a version of the Integrity control system was not available at our centre.

MLC tracking with the Agility MLC 9 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225 226 227 228 3. Results 3.1. Experimental validation of the MLC tracking algorithm with the Agility MLC The simulated versus real-time dose rate as a function of monitor units for two of the patients is shown in figure 4. The static-tumour VMAT deliveries are shown in figures 4(a) and 4(c), and the corresponding 2D MLC tracking deliveries, for a time period of 4 s, are shown in figures 4(b) and 4(d). It can be seen that in general, the algorithm correctly predicts the dose rate bin for each segment of the VMAT plan, and the same was observed for all plans measured. The simulated versus real-time delivery time for all plans measured in this set of experiments is shown in figure 5. The agreement with the line of equality as given by Lin s concordance correlation coefficient (Lin 1989) was found to be 0.914, indicating strong correlation between the algorithm s predicted value and the measured value. However, it can be seen that the algorithm always underestimates the real-time delivery time. This is due to the fact that the algorithm assumes that the required dose rate bin is reached instantaneously and that the dose rate remains constant at this value for the duration of the segment without fluctuations, which in practice is not the case. Additionally, unscheduled pauses in the treatment beam during the VMAT arc introduce residual time differences between the simulated and the real-time delivery. Such pauses can be seen in figures 4(a) and 4(c), at around 10 MU, where the dose rate drops to zero. In the absence of a real-time control system this leads to a degree of desynchronisation between actual motion and the resulting motion-compensation if this occurs during the MLC tracking VMAT deliveries. 3.2. Evaluation of the accuracy of MLC tracking The results of the gamma analysis with threshold levels of 3% for dose and 3 mm for distance-to-agreement are shown in figure 6. Additionally, the results averaged over the five patient plans investigated in this study are presented in table 3. Figure 6 shows that for all patient plans, the use of MLC tracking, either 1D or 2D, has improved the gammaanalysis pass rate compared to the pass rate without tracking, and the improvement is most pronounced for the 10 s time period motion trajectory. This result is expected, as the differences in the simulated and real-time delivery due to the limitations of the model (discussed in the previous section) will be less important for the longer time period. For example, a 1 s time difference corresponds to a tenth of a breathing cycle for the 10 s tracking plans but will correspond to a quarter of a breathing cycle for the 4 s tracking plans. The results of the gamma-analysis reflect this; the mean pass rate for 1D tracking with 1D motion and 2D tracking with 2D motion is 93.0% and 93.1% respectively for a 10 s time period, and is reduced to 84.2% and 89.1% for a 4 s time period. Additionally, it was found that the gamma analysis pass rate for the dose distribution calculated by the TPS for the Delta 4 compared with the measured static dose distribution was found to be 100% for all patients. In general, the highest gamma-analysis pass rate is seen for the 1D tracking with

MLC tracking with the Agility MLC 10 9 8 Simulated delivery Real time delivery 9 8 Simulated delivery Real time delivery 7 7 Dose rate (MU s 1 ) 6 5 4 3 Dose rate (MU s 1 ) 6 5 4 3 2 2 1 1 0 0 50 100 150 200 250 300 350 Monitor units (a) 0 0 50 100 150 200 250 300 350 Monitor units (b) 9 8 Simulated delivery Real time delivery 9 8 Simulated delivery Real time delivery 7 7 Dose rate (MU s 1 ) 6 5 4 3 Dose rate (MU s 1 ) 6 5 4 3 2 2 1 1 0 0 50 100 150 200 250 300 350 400 Monitor units (c) 0 0 50 100 150 200 250 300 350 400 Monitor units (d) Figure 4. Dose rate as a function of monitor units for simulated and real-time VMAT delivery. The dose rate for the static-tumour VMAT plans for patient 1 and patient 2 are shown in (a) and (c) respectively, and the corresponding 2D MLC tracking VMAT plans (time period of 4 s) are shown in (b) and (d). 229 230 231 232 233 234 235 236 237 238 239 1D motion delivery and 2D tracking with 2D motion delivery, with some exceptions. During the experiments, if an unscheduled pause occurred in the VMAT arc, no attempt was made to repeat the arc until it delivered without pauses. The VMAT arc was only repeated if a fault brought the treatment to a premature end. Therefore any unscheduled pauses in the radiation beam resulted in the desynchronisation of the MLC leaf positions and the motion of the Delta 4, and the acquired dose distribution still displayed minor motion-blurring. Therefore the gamma-analysis pass rate for certain tracking plans was reduced. For example, for patient 1 for a time period of 6 s, the 1D tracking with 1D motion delivery had the lowest gamma-analysis pass rate of all the tracking deliveries. The time difference between real-time and simulated delivery was 1.9 s for 1D tracking with 1D motion, 0.9 s for 2D motion with 1D tracking and 0.6 s for the 2D tracking

MLC tracking with the Agility MLC 11 72 70 Real time delivery time (s) 68 66 64 Static 1D dmlc: τ = 10 s 2D dmlc: τ = 10 s 62 1D dmlc: τ = 6 s 2D dmlc: τ = 6 s 60 1D dmlc: τ = 4 s 2D dmlc: τ = 4 s Line of equality 58 58 60 62 64 66 68 70 72 Simulated delivery time (s) Figure 5. Comparison of the real-time delivery time and the simulated delivery time for both static-tumour and MLC tracking plans for all patients. 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 255 256 257 258 259 with 2D motion delivery. Therefore, the first delivery with the largest time difference received the lowest gamma-analysis pass rate. For patient 5, for a 6 s time period, 1D tracking with 2D motion received the highest pass rate, and the same is observed for patient 2 for a 10 s time period. For patient 5, the time difference between simulated and real-time was 1.6 s for 1D tracking with 1D motion, 0.6 s for 1D tracking with 2D motion and 1.0 s for 2D tracking with 2D motion and the analogous time differences were 1.4 s, 0.6 s and 1.1 s for patient 2. This explains why 1D tracking with 2D motion had the highest pass rate for these examples. Despite these explicable discrepancies, the results show that when the VMAT arc is delivered smoothly without interruptions, the algorithm works well and a pass rate over 90% can be achieved, but unscheduled pauses are detrimental to the tracking dose distribution. For the sinusoidal motion studied here, with real-time control of the radiation beam, it would be possible to correct for the desynchronisation caused by these unscheduled pauses. For example, the breathing phase of the motion platform, and the corresponding breathing phase of the MLC leaf positions could be continually monitored, and if substantial desynchronisation is observed, the radiation beam could be paused until the correct phase is reached by the motion platform, at which point the beam could be reinitialised. However, such real-time control of the radiation beam was not available for this set of experiments, but would be desirable for further work with this algorithm.

MLC tracking with the Agility MLC 12 Table 3. Gamma analysis results averaged over five patient plans. Numbers are percentages, presented as mean ± standard deviation. Gamma analysis percentage pass rate (3% and 3 mm) 1D motion 2D motion 1D motion 2D motion 2D motion Time period (s) No tracking No tracking 1D tracking 1D tracking 2D tracking 4 60.3 ± 8.6 58.0 ± 6.9 84.2 ± 7.5 75.1 ± 2.8 89.1 ± 4.3 6 61.5 ± 7.9 59.0 ± 7.3 85.4 ± 9.9 88.3 ± 3.9 88.3 ± 7.4 10 62.9 ± 6.3 60.9 ± 7.4 93.0 ± 3.9 89.7 ± 4.2 93.1 ± 3.4 Gamma analysis (3%/3 mm) pass rate (%) 100 90 80 70 60 50 40 30 τ = 4 s τ = 6 s τ = 10 s 1D motion: no tracking 2D motion: no tracking 1D motion: 1D tracking 2D motion: 1D tracking 2D motion: 2D tracking 1 2 3 4 5 1 2 3 4 5 1 2 3 4 5 Patient number Figure 6. Gamma analysis results from the dosimetric accuracy experiments with the Delta 4 radiation detector. A threshold level of 3% for dose and 3 mm for distanceto-agreement was used, and detectors outside of the 5% isodose level in the reference dose distribution were not included in the analysis. 260 261 262 263 264 265 266 267 268 269 3.3. Evaluation of the delivery efficiency of MLC tracking with the Agility and MLCi multileaf-collimator The results of the simulations performed to compare the delivery times of the MLCi multileaf-collimator and Agility multileaf-collimator for the same VMAT plans are shown in figure 7. It can be seen that the static-tumour VMAT delivery times are comparable for both the MLCi and the Agility MLC, but the Agility MLC is able to deliver the tracking plans in a shorter treatment time. In particular, for patients 3 to 5, the Agility MLC was able to deliver the tracking plans without an increase in delivery time, which was not possible with the MLCi MLC. For both multileaf-collimators, the largest percentage time increase was observed for the 2D MLC tracking VMAT plans

MLC tracking with the Agility MLC 13 100 100 90 90 80 80 70 70 Delivery time (s) 60 50 40 Delivery time (s) 60 50 40 30 20 10 0 Static 1D dmlc: τ = 10 s 2D dmlc: τ = 10 s 1D dmlc: τ = 6 s 2D dmlc: τ = 6 s 1D dmlc: τ = 4 s 2D dmlc: τ = 4 s Patient 1 Patient 2 Patient 3 Patient 4 Patient 5 30 20 10 0 Static 1D dmlc: τ = 10 s 2D dmlc: τ = 10 s 1D dmlc: τ = 6 s 2D dmlc: τ = 6 s 1D dmlc: τ = 4 s 2D dmlc: τ = 4 s Patient 1 Patient 2 Patient 3 Patient 4 Patient 5 (a) (b) Figure 7. Simulated absolute delivery times for the static and tracking plans used in this study for (a) the MLCi multileaf-collimator and (b) the Agility multileafcollimator. 270 271 272 273 274 275 for the shortest time period motion trajectory. The greatest percentage time increase observed from a static-tumour to MLC tracking delivery for the MLCi was 57%, for patient 1, and the analogous time increase for the Agility was 11%. It is clear that the Agility MLC leaf speed of 35 mm s 1 and increased jaw velocity, compared with the MLCi leaf speed of 20 mm s 1 and lower jaw velocity, is advantageous for MLC tracking, particularly with decreased time period of motion. 276 277 278 279 280 281 282 283 284 285 286 287 288 289 290 291 292 4. Discussion In this study, a theoretical algorithm has been applied, experimentally, to evaluate MLC tracking during VMAT delivery on Elekta linear accelerators for the first time. VMAT MLC tracking experiments have also been performed and presented for Varian linear accelerators (Sawant et al. 2008, Zimmerman et al. 2009, Falk et al. 2010, Keall et al. 2011, Poulsen et al. 2012b). However, these experiments were performed using a real-time MLC tracking system, which is not possible with the Elekta system used here, and limited the success of the results presented in this paper. Keall et al. (2011) performed real-time dmlc IMAT (no dose rate or gantry speed modulation) tracking measurements of realistic lung traces with the Varian MLC tracking system, and found, on average, that 1.6% of points failed the 3%/3 mm gamma analysis with tracking. Additionally, they found no loss of treatment efficiency for the two lung plans investigated, but these were for IMAT plans without modulation of gantry speed and dose rate, whereas all plans used in the study presented here featured modulation of the gantry speed and dose rate. Zimmerman et al. (2009) have presented dmlc tracking results for one lung VMAT plan on a Varian accelerator, also using the Delta 4 detector, and showed an increase in the gamma-analysis pass rate from 87% without

MLC tracking with the Agility MLC 14 293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309 310 311 312 313 314 315 316 317 318 319 320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 tracking to 97% with tracking, but the efficiency of the treatment was not mentioned. Poulsen et al. (2012b) also demonstrated successful image-based MLC tracking of two lung VMAT plans with dose rate and gantry modulation on a Varian system (mean gamma pass rate of 98.3%), but again the efficiency was not mentioned. None of the measurements performed in the study in this paper achieved the same accuracy level for VMAT MLC tracking, although the mean gamma-analysis pass rate with 2D motioncompensation was on the border of clinical acceptability (Ezzell et al. 2009) for all time periods investigated. One of the contributing factors towards this result, was that it was only possible to transform the MLC leaf positions to follow a target motion at the control points (every 5 for these plans) and therefore the time interval at which the transformation took place varied with gantry speed. Such coarse and irregular sampling of the breathing cycle left a residual error between the position of the target and the transformed position of the MLC aperture, as between control points, the MLC motion was linear but the target motion, non-linear. It is possible to deliver a VMAT arc on Elekta linear accelerators with a finer control point spacing, e.g. 2, which has been shown theoretically to improve the accuracy of tracking compared with coarser control points (Davies et al. 2011). However, for this experimental study, plans of control point spacing 5 were used as the ability of the algorithm to correctly predict delivery parameters during a VMAT arc was shown to increase with increased control point spacing (Davies et al. 2011). The limitations of transforming the leaf positions only at the control points compromised the tracking accuracy the most for the trajectory with the shortest time period, and therefore the highest gamma-analysis pass rates were seen for the motion trajectory with 10 s time period, where the breathing cycle was sampled most frequently. However, as shown by Seppenwoolde et al. (2002) in an analysis of the characteristics of lung-tumour motion, the average time period is 3.6 s. Therefore, for accurate MLC tracking of realistic respiratory motion it would be desirable to be able to transform the leaf positions at a specified time interval, as seen in prototype real-time MLC tracking systems (Keall et al. 2006, Tacke et al. 2010), rather than be constrained solely to the control points. The tracking algorithm presented here reduces the dose rate until the MLC tracking plan is deliverable. However, Bedford & Warrington (2009) have shown that at low dose rates beam symmetry can be degraded, and recommend that dose rates less than 1.25 MU s 1 should only be used for small fractions of the VMAT delivery. For the plans used in this study, the majority of the deliveries, both static-tumour and MLC tracking, were delivered at a dose rate greater than 3 MU s 1 with the lowest observed dose rate being 1.75 MU s 1 for the plans for patient 5, for a duration of approximately 10 MU. Therefore beam symmetry was not an issue in these experiments, but is something to be cautious of when implementing this algorithm. The main limitation of the algorithm used in this study, is the requirement for a priori knowledge of the patient s respiratory motion. Respiration can be unpredictable (Seppenwoolde et al. 2002, Shirato et al. 2006), and therefore there could be difficulty in translating this algorithm to a clinical setting, where the motion may vary from that

MLC tracking with the Agility MLC 15 335 336 337 338 339 340 341 342 343 344 345 346 347 348 349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 seen during the radiotherapy planning process. It is possible to account for changes in breathing frequency using adaptive dose-rate regulation as demonstrated by Yi et al. (2008), which could be used to maintain synchronisation between the planned MLC motion (prescribed as a function of monitor units) and the tumour motion. However, amplitude variations and baseline shifts are more problematic, and difficult to correct with this method. The ideal solution to such a problem is detection of the motion in real-time and transformation of the MLC leaf positions accordingly, allowing pauses in the radiation beam if the tumour motion is too fast for the MLC leaves to track, as described by Sawant et al. (2008). However, real-time MLC tracking is not trivial; accounting for system latency and reliable detection of tumour position are some of the difficulties faced. The plans investigated in this study all had gantry speed and dose rate modulation, and the greatest increase in delivery time observed was 20%, and for three of the patients, there was a negligible increase in delivery time. The results therefore demonstrate that the Agility MLC has the potential to be a powerful and improved tool for MLC tracking due to its increased leaf and jaw speeds. Although the Agility MLC design has the additional benefit of the dynamic leaf guide, for this set of experiments and the VMAT plans selected, the DLG did not prove to be useful. Whether this is the case for all VMAT plans, or for other treatment modalities cannot be concluded from the experiments performed, and it is possible that for step-and-shoot IMRT delivery, or conformal delivery, the DLG would be a desirable feature for MLC tracking. It was not possible in this study to evaluate the dosimetric benefit of the thinner leaf width of the Agility compared with the MLCi multileaf collimator, but as shown by Poulsen et al. (2012b), the areas of underdose and overdose are directly correlated with the gamma-analysis pass rate, and therefore one would expect that an MLC with a thinner leaf width would be more suitable for MLC tracking, where it is often necessary to transform the MLC apertures parallel to the direction of leaf travel. Although the leaf speed of Agility MLC is one of the fastest commercially available, it should be noted that this increased speed enables greater leaf shifts which, if justified, must be modelled correctly in the treatment planning process. As illustrated in a review by Yu & Tang (2011), plans with longer leaf travel can result in greater discrepancies between the planned and delivered dose, and therefore it is vital that sufficient beam sampling is allowed in the dose calculation to ensure that the treatment planning system calculation is representative of the planned delivery. 369 370 371 372 373 374 5. Conclusion An algorithm for MLC tracking VMAT delivery on Elekta linear accelerators has been experimentally validated for the Agility MLC with the Integrity 3.0 control system, and dosimetric measurements have been performed to evaluate the accuracy of MLC tracking with this technique. It has been shown that an a priori method can offer accurate and efficient motion-compensation, with the majority of MLC tracking deliveries carried

MLC tracking with the Agility MLC 16 375 376 377 out with a modest, or zero, time increase compared with the static-tumour delivery. However, a real-time system is necessary for increased accuracy and to take full advantage of the Agility MLC s capabilities. 378 379 380 381 382 Acknowledgements The authors would like to thank Chris Bunton for collaborating with Peter Clowes on the design and construction of the motion platform and Elekta for useful discussions regarding the behaviour of the Agility MLC. This work was supported by an ICR PhD studentship and Cancer Research UK (grant number C46/A10588). 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415 416 417 418 References Bedford J L 2009 Treatment planning for volumetric modulated arc therapy. Med Phys 36(11), 5128 5138. Bedford J L 2013 Sinogram analysis of aperture optimization by iterative least-squares in volumetric modulated arc therapy. Phys Med Biol 58(5), 1235 1250. Bedford J L, Lee Y K, Wai P, South C P & Warrington A P 2009 Evaluation of the Delta4 phantom for IMRT and VMAT verification. Phys Med Biol 54(9), N167 76. Bedford J L, Nordmark Hansen V, McNair H A, Aitken A H, Brock J E C, Warrington A P & Brada M 2008 Treatment of lung cancer using volumetric modulated arc therapy and image guidance: a case study. Acta Oncol 47(7), 1438 1443. Bedford J L, Thomas M D R & Smyth G 2013 Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator. J Appl Clin Med Phys 14(2), 172 185. Bedford J L & Warrington A P 2009 Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat Oncol Biol Phys 73(2), 537 545. Cosgrove V, Thomas M, Weston S, Thompson M, Reynaert N, Evans C, K.J. B, De Wagter C, Thwaites D & Warrington A 2009 Physical Characterization of a New Concept Design of an Elekta Radiation Head with Integrated 160-leaf Multi-leaf Collimator Int J Radiat Oncol Biol Phys 75(3), S722 S723. Davies G A, Poludniowski G & Webb S 2011 MLC tracking for Elekta VMAT: a modelling study. Phys Med Biol 56(23), 7541 54. Ezzell G A, Burmeister J W, Dogan N, LoSasso T J, Mechalakos J G, Mihailidis D, Molineu A, Palta J R, Ramsey C R, Salter B J, Shi J, Xia P, Yue N J & Xiao Y 2009 IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 36(11), 5359 73. Falk M, af Rosenschöld P M, Keall P, Cattell H, Cho B C, Poulsen P, Povzner S, Sawant A, Zimmerman J & Korreman S 2010 Real-time dynamic MLC tracking for inversely optimized arc radiotherapy. Radiother Oncol 94(2), 218 23. Keall P J, Cattell H, Pokhrel D, Dieterich S, Wong K H, Murphy M J, Vedam S S, Wijesooriya K & Mohan R 2006 Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking system. Int J Radiat Oncol Biol Phys 65(5), 1579 84. Keall P J, Sawant A, Cho B, Ruan D, Wu J, Poulsen P, Petersen J, Newell L J, Cattell H & Korreman S 2011 Electromagnetic-Guided Dynamic Multileaf Collimator Tracking Enables Motion Management for Intensity-Modulated Arc Therapy. Int J Radiat Oncol Biol Phys 79(1), 312 320. Krauss A, Fast M F, Nill S & Oelfke U 2012 Multileaf collimator tracking integrated with a novel x-ray imaging system and external surrogate monitoring. Phys Med Biol 57(8), 2425 39.

MLC tracking with the Agility MLC 17 419 420 421 422 423 424 425 426 427 428 429 430 431 432 433 434 435 436 437 438 439 440 441 442 443 444 445 446 447 448 449 450 451 452 453 454 455 Krauss A, Nill S, Tacke M & Oelfke U 2011 Electromagnetic real-time tumor position monitoring and dynamic multileaf collimator tracking using a Siemens 160 MLC: geometric and dosimetric accuracy of an integrated system. Int J Radiat Oncol Biol Phys 79(2), 579 87. Lin L I 1989 A concordance correlation coefficient to evaluate reproducibility. Biometrics 45(1), 255 68. Lujan A E, Balter J M & Ten Haken R K 2003 A method for incorporating organ motion due to breathing into 3D dose calculations in the liver: sensitivity to variations in motion. Med Phys 30(10), 2643 2649. McQuaid D, Partridge M, Symonds-Tayler J R, Evans P M & Webb S 2009 Target-tracking deliveries on an Elekta linac: a feasibility study. Phys Med Biol 54(11), 3563 3578. Otto K 2008 Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 35(1), 310 317. Poulsen P R, Carl J, Nielsen J, Nielsen M S, Thomsen J B, Jensen H K, Kjærgaard B, Zepernick P R, Worm E, Fledelius W, Cho B, Sawant A, Ruan D & Keall P J 2012a Megavoltage image-based dynamic multileaf collimator tracking of a NiTi stent in porcine lungs on a linear accelerator. Int J Radiat Oncol Biol Phys 82(2), e321 7. Poulsen P R, Fledelius W, Cho B & Keall P 2012b Image-based dynamic multileaf collimator tracking of moving targets during intensity-modulated arc therapy. Int J Radiat Oncol Biol Phys 83(2), e265 71. Sawant A, Venkat R, Srivastava V, Carlson D, Povzner S, Cattell H & Keall P 2008 Management of three-dimensional intrafraction motion through real-time DMLC tracking. Med Phys 35(5), 2050 2061. Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque J V & Miyasaka K 2002 Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int J Radiat Oncol Biol Phys 53(4), 822 834. Shirato H, Suzuki K, Sharp G C, Fujita K, Onimaru R, Fujino M, Kato N, Osaka Y, Kinoshita R, Taguchi H, Onodera S & Miyasaka K 2006 Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy. Int J Radiat Oncol Biol Phys 64(4), 1229 36. Tacke M B, Nill S, Krauss A & Oelfke U 2010 Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC Med Phys 37(2), 753 761. Yi B Y, Han-Oh S, Lerma F, Berman B L & Yu C 2008 Real-time tumor tracking with preprogrammed dynamic multileaf-collimator motion and adaptive dose-rate regulation. Med Phys 35(9), 3955 3962. Yu C X & Tang G 2011 Intensity-modulated arc therapy: principles, technologies and clinical implementation. Phys Med Biol 56(5), R31 R54. Zimmerman J, Korreman S, Persson G, Cattell H, Svatos M, Sawant A, Venkat R, Carlson D & Keall P 2009 DMLC motion tracking of moving targets for intensity modulated arc therapy treatment: a feasibility study. Acta Oncol 48(2), 245 250.