Clinical implementation of photon beam flatness measurements to verify beam quality

Similar documents
DAILY LINAC QA BEAM QA

A SYSTEM OF DOSIMETRIC CALCULATIONS

ph fax Deming Way Middleton WI USA

Raising the Bar in IMRT QA

Feasibility study of performing IGRT system daily QA using a commercial QA device

FAST, precise. qa software

Dynalog data tool for IMRT plan verification

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

Photon beam dose distributions in 2D

Basic Radiation Oncology Physics

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Machine and Physics Data Guide

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

Verification measurements of an emc algorithm using a 2D ion chamber array

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

MCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR

MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays

Use of a commercial ion chamber detector array for the measurement of high spatial resolution photon beam profiles

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

Implementation of the EGSnrc / BEAMnrc Monte Carlo code - Application to medical accelerator SATURNE43

Validation of GEANT4 Monte Carlo Simulation Code for 6 MV Varian Linac Photon Beam

MapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays

Table of Contents. About AQUA

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA

Comparison of absorbed dose distribution 10 MV photon beam on water phantom using Monte Carlo method and Analytical Anisotropic Algorithm

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

Preface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.

Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code

Quality assurance of a helical tomotherapy machine

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

3D Scanning Comes Full Circle

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Initial Clinical Experience with 3D Surface Image Guidance

Quick Reference Datasheet For All RIT113 Packages

UNCOMPROMISING QUALITY

Calibration and quality assurance for rounded leaf-end MLC systems

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo evaluation of the dosimetric uncertainty in matched 6 MV Elekta and Varian linear accelerators

Megan A. Wood, M.S. Under the direction of Larry DeWerd, Ph.D. University of Wisconsin Medical Radiation Research Center (UWMRRC)

TG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

Output factor comparison of Monte Carlo and measurement for Varian TrueBeam 6 MV and 10 MV flattening filter-free stereotactic radiosurgery system

Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility

4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1

Comparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center

Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility

Development a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Automated Quality Assurance for Image-Guided Radiation Therapy

Proton dose calculation algorithms and configuration data

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Use of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta

Medical Dosimetry 37 (2012) Medical Dosimetry. journal homepage:

TPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1

A software tool for the quantitative evaluation of 3D dose calculation algorithms

Goals. Scope/exclusions. Outline: follow outline of MPPG (plus rationale & some implementation experiences)

A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans

Indrin Chetty Henry Ford Hospital Detroit, MI. AAPM Annual Meeting Houston 7:30-8:25 Mon 08/07/28 1/30

Delta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear

Monitor Unit (MU) Calculation

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

The MSKCC Approach to IMRT. Outline

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS

Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI

Outline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

2D DOSE MEASUREMENT USING A FLAT PANEL EPID

An investigation into the use of MMCTP to tune accelerator source parameters and testing its clinical application

EXTERNAL PHOTON BEAMS: PHYSICAL ASPECTS

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation

Commissioning and verification of the collapsed cone convolution superposition algorithm for SBRT delivery using flattening filter-free beams

Comparison Study of Monte Carlo Simulations and Measurements of Relative Output Factors of 6 MV Photon Beam

Accuracy of treatment planning calculations for conformal radiotherapy van 't Veld, Aart Adeodatus

A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems

A Geant4 based treatment plan verification tool: commissioning of the LINAC model and DICOM-RT interface

Shielding factors for traditional safety glasses

Verification of dose calculations with a clinical treatment planning system based on a point kernel dose engine

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson, Louise Alexander, Steve Morgan Brighton and Sussex University Hospitals NHS Trust, Brighton, UK simon.goodall@genesiscancercare.com.au Received 9 April, 2015; accepted 6 July, 2015 This work describes the replacement of Tissue Phantom Ratio (TPR) measurements with beam profile flatness measurements to determine photon beam quality during routine quality assurance (QA) measurements. To achieve this, a relationship was derived between the existing TPR 15/5 energy metric and beam flatness, to provide baseline values and clinically relevant tolerances. The beam quality was varied around two nominal beam energy values for four matched Elekta linear accelerators (linacs) by varying the bending magnet currents and reoptimizing the beam. For each adjusted beam quality the TPR 15/5 was measured using an ionization chamber and Solid Water phantom. Two metrics of beam flatness were evaluated using two identical commercial ionization chamber arrays. A linear relationship was found between TPR 15/5 and both metrics of flatness, for both nominal energies and on all linacs. Baseline diagonal flatness (F DN ) values were measured to be 103.0% (ranging from 102.5% to 103.8%) for 6 MV and 102.7% (ranging from 102.6% to 102.8%) for 10 MV across all four linacs. Clinically acceptable tolerances of ± 2% for 6 MV, and ± 3% for 10 MV, were derived to equate to the current TPR 15/5 clinical tolerance of ± 0.5%. Small variations in the baseline diagonal flatness values were observed between ionization chamber arrays; however, the rate of change of TPR 15/5 with diagonal flatness was found to remain within experimental uncertainty. Measurements of beam flatness were shown to display an increased sensitivity to variations in the beam quality when compared to TPR measurements. This effect is amplified for higher nominal energy photons. The derivation of clinical baselines and associated tolerances has allowed this method to be incorporated into routine QA, streamlining the process whilst also increasing versatility. In addition, the effect of beam adjustment can be observed in real time, allowing increased practicality during corrective and preventive maintenance interventions. PACS number: 87.56.Fc Key words: quality assurance, beam flatness, photon energy, tissue phantom ratio, beam quality I. INTRODUCTION Radiotherapy treatments can only be delivered as planned if the radiation beam produced by a linac matches the model used within the treatment planning system (TPS). Routine linac quality assurance (QA) testing is, therefore, recommended to ensure that the produced physical beam properties are maintained within specified tolerances of the baseline values assumed by the TPS. (1,2) The beam quality, as characterized by its penetrative ability in water, is one such fundamental property. The current standard metric for beam quality measurements is percentage a Corresponding author: Simon Goodall, Genesis Cancer Care, 24 Salvado Rd., Wembley, Perth, WA 6014, Australia; phone: +618 6318 28456; fax: +618 6318 2810 email: simon.goodall@genesiscancercare.com.au 340 340

341 Goodall et al.: Flatness to determine beam quality 341 depth dose (PDD) in water. During a commissioning process, a high resolution PDD plot will be captured using a full water phantom. From this PDD, a tissue phantom ratio (TPR) value may be derived as a beam quality constancy check parameter. This commonly adopted method is often used during routine QA due to its simplicity and practicability, with only dual-depth, isocentric, ionization chamber measurements being required. At Brighton and Sussex University Hospitals (BSUH), routine measurements are made at 5 cm and 15 cm depth within a Solid Water-equivalent plastic (WEP) phantom, and a TPR 15/5 constancy parameter is derived from their ratio, as described in the Materials & Methods section A. During the production of a clinically useful X-ray photon beam, a flattening filter is typically used within the linac to produce a beam profile which is flat at a given depth in water. Due to the conical shape of the flattening filter, the energy spectrum of the beam varies across the beam profile with a unique dependence upon the individual filter. If changes arise within the input beam energy spectrum, the interactions with this filter will also vary and modify the flatness of the profile at a given depth. It has been shown that these changes in flatness can be used to monitor changes in beam quality with an increased sensitivity over traditional PDD techniques. (3,4) At BSUH, two IC PROFILER (Sun Nuclear Corporation (SNC), Melbourne, FL) multiaxis ion chamber array devices, (5) referred to locally as ICP1 and ICP2, are currently used to monitor the symmetry of photon beams when located within a linac gantry head attachment. If the beam quality could also be validated from the profile flatness data acquired during these measurements, the additional time and equipment required for TPR measurements could be removed, allowing the QA process to be streamlined. In addition, the placement of the ICP within the linac head attachment allows the ICP surface to remain perpendicular to the central beam axis as the gantry rotates. This, in turn, permits an assessment of the beam flatness to be made with ease at any gantry angle or during arc deliveries. The aim of this work was therefore to establish the relationship between TPR 15/5 values and beam flatness for the four clinical Elekta linacs at BSUH. From this relationship, baseline flatness values may be derived alongside tolerances which equate to those currently implemented. This will, in turn, allow TPR 15/5 measurements to be replaced by flatness measurements on a routine basis, streamlining the QA process and improving versatility. II. MATERIALS AND METHODS Measurements were made on four matched linacs: linacs 1 3 (Precise, Elekta Ltd., Crawley, UK) and linac 4 (Synergy, Elekta Ltd.). Linacs 2 4 were dual-photon energy with 6 and 10 MV photon beams available, whilst linac 1 was single-photon energy with only a 6 MV photon beam available. Measurements of TPR 15/5 were made using a 0.6 cm 3 ionization chamber positioned within a WEP phantom, and beam flatness was measured using one of the two available ICP devices. Each ICP was calibrated using the recommended wide-field calibration technique and associated software. (6) The additional steps suggested by Goodall and Morgan (7) were also applied to improve cross-device calibration consistency. To capture flatness and TPR 15/5 measurements for a range of photon beam qualities extending beyond clinically acceptable values, the bending magnet system was progressively adjusted. The bending coarse and bending fine magnets were both altered from the clinical values, maintaining the relative ratio between them and hence the achromatic setup. For each new set of bending magnet currents, the electron gun current servo and automatic frequency control systems were optimized so as to restore a maximum dose rate, and beam steering corrections were made to achieve optimal symmetry. Measurements were then made on each linac of TPR 15/5, the diagonal and the axial normalized flatness for all variations around the two nominal energies.

342 Goodall et al.: Flatness to determine beam quality 342 A. TPR 15/5 Measurements of TPR 15/5 were made based on the techniques described in the IPSM Code of Practice for high-energy photon therapy dosimetry. (8) The ionization chamber was placed at the isocenter, defining a constant source to chamber distance of 100 cm, during measurements at depths of 15 cm and 5 cm in WEP. Readings were taken at each depth with a symmetric 10 10 cm 2 field and the TPR 15/5 was then calculated as (1) where and are the average ionization chamber readings at depths of 15 cm and 5 cm, respectively. The temperature, pressure, and ion recombination corrections were assumed to be equal at both depths. At our institution, the baseline values for TPR 15/5, when measured using the technique described above, are 0.703 and 0.752 for 6 and 10 MV beams, respectively. On monthly testing of the TPR 15/5, a clinical tolerance of ± 0.5% is applied locally to surpass the recommendations of the guidance. (1,2) B. Beam flatness The ICP was positioned with an SSD of 100 cm and a 1 cm slab of Perspex was added to provide buildup in addition to the intrinsic 0.9 cm within the device. (5) The source-to-chamber distance was therefore set to 100.9 cm. All measurements were made by integrating the dose of a 30 30 cm 2 field over a 100 MU exposure. Two definitions of the beam flatness were investigated: the diagonal normalized flatness, F DN, as described by Gao et al., (3) and the axial normalized flatness, F AN. From the captured profile the beam flatness values were calculated as and (2) where are the corrected readings from the chambers situated on the positive and negative diagonal axes at ± 17.0 cm, are the readings from the chambers situated on the x- and y-axes at ± 12.0 cm, and R CAX is the reading from the central beam axis ion chamber. For all readings, the derived chamber specific calibration factors were applied and the background was accounted for. III. RESULTS For each linac and nominal energy value, both F DN and F AN were found to vary linearly, with the TPR 15/5 percentage error arising from an investigative range of bending magnet settings. Figure 1 shows the relationship between TPR 15/5 percentage error and flatness for the measurements performed at 6 MV on linac 3, extending across the clinically acceptable range of beam quality. When varying the bending magnets settings, a greater response was displayed by the diagonal flatness data (F DN ) for every linac and energy combination when compared to the equivalent axial flatness data (F AN ). For the 6 MV beams, the mean rate of change in F DN across all four linacs was calculated to be -4.2% ± 0.2% per 1% change in the TPR 15/5 percentage error. For F AN, this value was calculated to be -3.3% ± 0.2% per 1% change in the TPR 15/5 percentage error.

343 Goodall et al.: Flatness to determine beam quality 343 Additionally for all linacs, when adjusting the bending magnet current for the 10 MV nominal energy setting, a more rapid change in flatness with TPR 15/5 percentage error was demonstrated as compared to the 6 MV results given above. For 10 MV, the overall mean rate of change in F DN across all four linacs was calculated to be -6.1% ± 0.2% per 1% change TPR 15/5 percentage error. Figure 2 shows the individual data for linac 2. Using ICP1, the mean baseline F DN value was calculated across all four linacs to be 103.0% for a nominal energy of 6 MV with a range of 102.5% to 103.8%. For a nominal energy of 10 MV, the mean F DN value was calculated to be 102.7% with a range of 102.6% to 102.8%. When using the ICP2 device, the rates of change of F DN with TPR 15/5 percentage error for 6 and 10 MV were found to be equal to those measured using ICP1 within the experimental error of a ± 0.2% variation in the rate of change of F DN with TPR 15/5 percentage error. Small discrepancies were measured between the baseline F DN values derived for each ICP which were greater than the F DN repeatability of ± 0.2%. These variations were found to have a mean magnitude of 0.5% and a maximum magnitude of 0.8%. Fig. 1. The variation of diagonal normalized flatness (solid line R 2 = 0.9924) and axial normalized flatness (dashed line R 2 = 0.9908) with TPR 15/5 error (%) for linac 3 at a nominal energy of 6 MV. Fig. 2. The variation in diagonal normalized flatness with TPR 15/5 error (%) for nominal energies of 6 MV (solid line R 2 = 0.9982) and 10 MV (dashed line R 2 = 0.9972) using linac 2. The dotted lines show the ± 0.5% clinical tolerance applied locally to the TPR 15/5 measurements.

344 Goodall et al.: Flatness to determine beam quality 344 IV. DISCUSSION These results provide similar findings for Elekta Precise and Synergy linacs to those which were shown for a Varian Medical Systems Clinac 2100 by Gao et al. (3) The F DN was shown to always provide a more sensitive measure of photon beam quality than either F AN or the traditional measurement of TPR 15/5. As a result of this work, it has been possible to implement this measurement technique on a routine basis, and streamline the routine and corrective QA processes by maximizing the amount of information which can be captured from a single exposure. In addition, the photon beam quality may easily be assessed with respect to clinically relevant tolerances under more varied conditions, including alternative gantry angles and during arc deliveries. For the linacs that were investigated, it can be suggested that, when measuring at the 80% width points for a 30 30 cm 2 field, applying F DN tolerances of ± 2% or ± 3%, for 6 or 10 MV photons, respectively, would ensure that the TPR 15/5 was correct within the local tolerance of ± 0.5%. It can also be estimated that, if the beam profile maintains less than a ± 1% variation from the baseline value, the TPR 15/5 will remain correct within approximately ± 0.25% or ± 0.20% for 6 and 10 MV photons, respectively. A device such as the ICP also allows the user to view the beam profile in a real-time mode. When carrying out corrective action, any changes to the beam quality may therefore be immediately observed and assessed without the need to complete time-consuming ionization chamber measurements within a phantom. Although the rate of change of F DN with TPR 15/5 percentage error was shown to be comparable between linacs, offsets were measured between the baseline values. These differences may arise from the accuracy with which individual linacs are matched to one another and within the manufacturing process of the linac components, such as the flattening filter. This technique must, therefore, be commissioned for each individual linac and energy combination. Variations in the baseline values were also observed between ICP devices, which may be due to residual uncertainties within the calibration process, despite the additional steps to minimize these. (7,9) Considerations should therefore be given to establishing separate baselines for each device, or to reduce tolerances, if interchangeability is desired. Using the methods described here, it is suggested that approximately 90 min would be required, per energy, to capture a sufficient quantity of TPR 15/5 data and corresponding profiles in order to derive the required relationships for two ICP devices. Once baseline and tolerance values have been established, two key points should be addressed before implementing this technique on a routine basis. Firstly, although the F DN measurement is largely unaffected by errors within the beam symmetry, (3) it is suggested that a value of TPR 15/5 is only interpreted from the F DN value when the beam symmetry is within locally applied tolerances; at BSUH this tolerance is ± 2%, based on guidance. (1,2) Secondly, it is important to ensure the ongoing consistency and reproducibility of the beam profiling device. To ensure this is adequate, an assessment of the accuracy of each profiling device in use should be considered against a full water phantom on an annual basis or following a device recalibration. V. CONCLUSIONS A strong linear correlation was shown between TPR 15/5 and F AN, and also between TPR 15/5 and F DN values for 6 and 10 MV photon beams across multiple Elekta linacs. Measurements of F DN were shown to have an increased sensitivity to changes in beam quality compared to F AN which, in turn, was more sensitive than the current constancy check parameter of TPR 15/5. The use of flatness measurements in place of TPR 15/5 may, therefore, allow beam quality tolerances to be reduced, whilst simultaneously providing a means to more easily achieve them through direct observation of live beam profiles during corrective action. In addition, flatness measurements

345 Goodall et al.: Flatness to determine beam quality 345 have an increased versatility when compared to TPR 15/5 determination by allowing the beam quality to be validated under varying conditions, such as alternative gantry angles. Following an initial commissioning time investment of approximately 90 min per nominal energy for two profiling devices, this technique can allow streamlining of the QA process by minimizing the number of exposures and quantity of equipment which is required. The method also presents opportunities for maintaining an improved level of consistency in machine performance due to its enhanced sensitivity and improved practicality, as compared with traditional methods of beam quality determination. REFERENCES 1. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: Quality assurance of medical accelerators. Med Phys. 2009;36(9):4197 212. 2. Mayles WP, Lake R, Mckenzie A, et al. Physics aspects of quality control in radiotherapy. IPEM Report No. 81. York, UK: The Institute of Physics and Engineering in Medicine; 1999. 3. Gao S, Balter PA, Rose M, Simon WE. Measurement of changes in linear accelerator photon energy through flatness variation using and ion chamber array. Med Phys. 2013;40(4):042101. 4. Ritter TA, Gallagher I, Roberson PL. Using a 2D detector array for meaningful and efficient linear accelerator beam property validations. J Appl Clin Med Phys. 2014;15(6):46 58. 5. Simon TA, Kozelka J, Simon WE, Kahler D, Li J, Liu C. Characterization of a multi-axis ion chamber array. Med Phys. 2010;37(11):6101 11. 6. Simon WE, Shi J, Iannello CA. Wide field calibration of a multi-sensing array. U.S. patent 6125335. Sun Nuclear Corporation; 2000. 7. Goodall S and Morgan S. Technical Note: A method for improving the calibration reproducibility of an ionization chamber detector array. Med Phys. 2014;41(9):091704. 8. Lillicrap SC, Owen B, Williams JR, Williams PC. Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service. Phys Med Biol. 1990;35:1355 60. 9. Simon TA, Simon WE, Kahler D, Li J, Liu C. Wide field array calibration dependence on the stability of measured dose distributions. Med Phys. 2010;37(7):3501 09.