Wide Guy: Inverted Widefield Microscope

Similar documents
Setting up Micro-Magellan for Device control

Micro-Magellan Install and User Guide

Start-up system. Nikon Wide Field Microscope Ee1078. Complete manual

Olympus DeltaVision Microscope Start-Up and Shut-Down Instructions.

Zeiss Efficient Navigation (ZEN) Blue Edition Standard Operation Protocol

Biology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System

Zeiss Z.1 Lightsheet. Quick-Start Guide

Spinning Disk Protocol

Cell Imaging Unit UIC Bárbara Fekete ZEISS AXIOIMAGER MICROSCOPE MANUAL

Multi-Photon Training

Nikon A1-B Confocal Operating Manual. Start-up. Microscope

Multi Time Series Rev

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Zeiss Elyra Microscope Standard Operation Protocol Basic Operation

Heit/Kerfoot/Dikeakos. Leica DMI6000 B. User Manual

Confocal 6 Guide. Switching On Components of the system should be switched on in the following order: 1. Lasers Argon laser (488nm and 514nm)

3-D. Here red spheres show the location of gold nanoparticles inside/around a cell nucleus.

Confocal Application Notes April 2010

Manual ANGSTROM grid confocal microscope

Operating Procedure for Horiba Raman Microscope

LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes

Primary Use. Operating Principle

DSU Start-Up instructions

ANDOR/OLYMPUS SPINNING DISK USER GUIDE

XRADIA microxct Manual

Indiana Center for Biological Microscopy. BioRad MRC 1024 MP Confocal & Multi-Photon Microscope

Olympus IX-70 Imaging Protocol

CoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis

University of Minnesota Nano Fabrication Center Standard Operating Procedure

Digital Microscopes Zoomy 2.0 Digital Microscope What software/devices will I need to use a Zoomy? Connecting Zoomy 2.0

The Paper Project Guide to Confocal Imaging at Home & in the Classroom

Basic Users Guide for the LSM 510 Confocal Microscope

CoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis

Olympus IX-70 Imaging Protocol

Olympus BX51 on Aura. Introduction to the NRI-MCDB Microscopy Facility BX51 Upright Microscope

Fluorimeter User s Booklet

All-In-One. digital inverted microscope. High-quality Imaging Has Never Been Easier. evos-ca.com

SmartSoft AES Operator s Guide

OIC. User Guide to Confocal Microscope: Zeiss LSM Pascal

Zeiss AxioImager.Z2 Fluorescence Protocol

LEXT 3D Measuring LASER Microscope

1. Getting Started: Brief Step-By-Step Guide (PDF File)

The walkthrough is available at /

123D Catch - Tutorial

Living Image Software

ADPRO FastTrace 2 USER GUIDE. Provided By:

with an incubator for study of live

Lumascope Startup Guide -- Lumascope 720

HOW TO SETUP AND RUN SHIVA...

LSM510 Confocal Microscope Standard Operation Protocol Basic Operation

Imaris 4.2 user information

Photogrammetry Guide to Creating Digital 3D Models

Creating and Displaying Multi-Layered Cross Sections in Surfer 11

Confocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by

Links to Activities ACTIVITY 1.1. Links to Activities Links to Activities

NANOSPEC 4150 STANDARD OPERATING PROCEDURES

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Image Analysis begins with loading an image into GenePix Pro, and takes you through all the analysis steps required to extract data from the image.

Colin Paul Updated 14 November Preparation of publication-quality videos using ImageJ

Quick Start Guide - cellsens

Gallios TM Quick Reference

SMART CLASSROOM TRAINING MANUAL

Zeiss LSM 510 Meta inverted confocal manual

Caustics - Mental Ray

XPS1 Automated Multi-Sample Run Procedure

Software for Observation and Measurement SGMMS V2.6. Users Manual

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017

IMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1

Blender Notes. Introduction to Digital Modelling and Animation in Design Blender Tutorial - week 1 The Blender Interface and Basic Shapes

UNIT ONE: The Worksheet. Workbook Window Excel Worksheet Fill handle Automatic fill Column widths Opening a file Saving a file

CCD Acquisition Function

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1

Volocity ver (2013) Standard Operation Protocol

Image Processing Guideline for TMU 7T MRI

ND Processing Tools in NIS-Elements

WINDOWS NT BASICS

Leica Microsystems Intelligent Structured Illumination Microscopy

Contributing Lists in Activity Center

User s Guide to the LMD Laser Micro-dissection. System

Tutorial: Importing VMOD/MODFLOW Models

Table of Contents Introduction... 4

Icy Training - Level 1 - Introduction

13. Albums & Multi-Image Printing

Standard Operating Procedure for the Horiba FluroMax-4

Installing and Using Prior Software

Premiere Pro Desktop Layout (NeaseTV 2015 Layout)

Visual Physics Introductory Lab [Lab 0]

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b

DIFFUSION TENSOR IMAGING ANALYSIS. Using Analyze

Quick Guide ZEN 2. First steps with ZEN

Leica DFC3000 G. Crisp fluorescence documentation for routine experiments

Microsoft Excel 2010 Basic

Samba Hair System User manual Jefferson Blvd. Culver City, CA 90230, USA Tel: (310) Fax: (310)

Handout Objectives: a. b. c. d. 3. a. b. c. d. e a. b. 6. a. b. c. d. Overview:

Creating a Presentation

Excel 2013 for Beginners

Leica DMI4000 B Leica DMI6000 B. Automated Inverted Microscopes for Life Sciences

Bridgeware Systems War Board Documentation

Condensed AFM operating instructions:

LET S FOCUS ON FOCUSING

Transcription:

Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3 2.2 Software Initialization and Sample Alignment............................ 5 3 Data Acquisition 7 3.1 Micro-Manager............................................ 7 3.2 μmagellan............................................... 7 3.2.1 Explore Mode......................................... 7 3.2.2 Acquisition........................................... 10 4 Shut Down Procedure 11 5 Trouble Shooting 12 1

1 Introduction Colloquially known as Wide Guy, the microscope is a conventional widefield microscope set upon the inverted Zeiss Axioscope 200M body. The illumination source is a Sutter Lambda XL with appropriate excitation and emission filters for the common DAPI, FITC, TRITC, and Cy5 lines. Due to the configuration of the microscope, 7-color imaging (plus brightfield) can be achieved when taking advantage of the Brilliant Violet dye series. The microscope also includes components and features such as: Andor Clara CCD Fast acquisition rates with large array resolution Motorized microscope stage Controlled by either joystick or software z-axis piezo Software controlled for 3D image stacks Sample incubator Available for cells in culture imaging μmagellan Dynamic device control and acquisition 2

2 Initial Setup Sign up for microscope time using the MyCores microscope scheduler. 1. This can be done up to 3 weeks in advance for the owners consortium. 2. For non-owners, reservations can be made beginning Friday for time during the following week. 2.1 Hardware Startup Note: If you want to image cells in culture, turn ON the incubator 2 hours ahead of time and place the chamber on the microscope to allow the system to stabilize (Figure 1). Figure 1: Cell Incubator on Stage 1. If the computer is not on, press the external power button taped to the outside of the case (Figure 2). 2. Turn ON the power strip integrated into the shelf behind the microscope (Figure 3). 3. Turn on the Lambda XL (Figure 4). (a) Flip the power switch on the back of the unit. (b) On the display press the Local button. (c) Press 1 to start lamp. 3

Figure 2: Computer Power Button Figure 3: Power Strip for Components 4

Figure 4: Lamp for Microscope 2.2 Software Initialization and Sample Alignment All hardware components must be powered on before software can be loaded. 1. Double-click on the Micro-Manager icon, which is located on the Desktop, to start the data acquisition software. 2. Upon start-up, select the configuration file. Note: This config file should already be loaded, and the user should only have to select OK. Bonus Note: If you become a heavy user, talk to a BIDC employee about making a custom config file to speed up your imaging time. 3. In the software choose the correct objective. 4. Align the sample focus by eye: (a) In the software under Mode choose Eyes - color where color is the emission range of your dye. ProTip: Use the brightest, most robust dye in your sample for initial alignment. Additionally, Red has lots of background due to the autofluorescence of most sample substrates. (b) Click the live button in the software to send the open shutter command to the instrument. (c) Look through the eye-piece and bring your sample into focus using the focus mechanism. Note: Rotate the focus towards you, as this will move the objective away from your sample and align the focal plane. (d) Once you are satisfied with the focus, hit the stop button to close the shutter. 5. Parameter setup within the software: Note: The path length is slightly different between your eyes and the camera, so the focus will need to be adjusted by a few microns. (a) Under Channel choose the appropriate dye related excitation/emission combination. (b) Choose an exposure time (typically between 100-500 ms). Note: The appropriate exposure time is sample and dye dependent. Take note of final parameters for future reference. 5

(c) Click the live button to get a continuous stream of your sample. (d) Use the scroll wheel of the mouse to finely adjust the focus of the objective until your sample is in view. Note: The focal change response time is dependent on exposure time. It is possible to scroll through your sample plane without it appearing on the screen. Note: Typically, when the sample is brightest the focus is at the middle plane of your sample. 6. Repeat the adjusting of exposure time for each dye you wish to image. Note: The Brightfield lamp intensity can be adjusted as well as the exposure time. 6

3 Data Acquisition There are two effective ways to collect data on the Wide Guy; through the Micro-Manager interface or the plugin extension known as μmagellan. Both have their pros and cons and the choice ultimately comes down to the user. 3.1 Micro-Manager Data collected from the Micro-Manager software itself produces image stacks at each location with the color channels incorporated into the file. The files are easily loaded and analyzed in ImageJ/FIJI. 1. When you are ready to acquire data, click on the Multi-D Acq button. 2. To perform a z-stack: (a) Check the z-stacks (slices) field and switch to absolute z. (b) Set the interval for your z-stack: Manually enter the start and end positions, or While in live mode, scroll the focus to the top and bottom plane of your sample. Set each position appropriately. 3. To use multiple channels: (a) Check the Channels field and select Channel for Channel group. (b) Select Create Multiple Channels and click New to add additional channels. 4. To image multiple x-y positions: (a) Check the Multiple positions (XY) field, and click the edit positions list button. (b) Select Create Grid near the bottom right. (c) Set positions while in live mode and choose 10% overlap. 5. Select the acquisition order for your Multi-D acquisition. Note: Channel/slice is much faster than slice/channel. 6. Click Acquire!. 3.2 μmagellan The Micro-Manager plugin μmagellan was written by the former BIDC employee Henry Pinkard. It has many built-in features designed to speed up sample exploration, region of interest defining, and sample acquisition time while limiting sample irradiance. Data is recorded in a special folder that includes image stacks of the regions of interest in TIFF format along with a.xml file that contains the metadata. To stitch the files together for data analysis in either ImageJ/FIJI or Imaris, the folder needs to run through a separate plugin known as Imaricμmpiler. Below is a condensed user s guide. A more in depth article can be found at: 3.2.1 Explore Mode https://www.micro-manager.org/wiki/micromagellan Explore mode allows the user to quickly find regions of interest, define z-stack thickness, and create grids for image acquisition. 1. Initial Setup: (a) Assign the saving directory to your user folder. 7

Figure 5: Explore Settings (b) Assign a saving name (μmagellan will save your exploration areas). (c) Enter a z-step (this can be changed before data acquisition). (d) Change Channel Group to channel. (e) Enter a Tile overlap percentage (10% is a typical amount). (f) Press Explore! (a new window will open). 2. Check the channels you wish to use in the upper right. 3. Click on the black space (a magenta box will appear). 4. Click on the magenta box to confirm image area (you should now see an imaged ROI in the colors you selected). 5. Use the look up tables (LUTs) on the right to adjust contrast for each color. 6. Use the scroll wheel to zoom in and out in the explore window and/or drag the screen to move around your sample. 7. You can click and drag to select multiple fields of view to be imaged. Figure 6: Explore Window 8. Adjust the z-limits by either pressing the arrow keys (which will step by the pre-defined step size) or dragging the scroll bar. Imaging a region will now include multiple image planes. 9. Define regions to be imaged: (a) Click on the Grid tab in the upper right. (b) Press New Grid. (c) Define the number of rows and column in the grid coresponding to fields of view. 8

(d) Drag the grid on the screen to the desired region on interest. (e) Repeat for as many areas you would like imaged. Figure 7: Explore Grids Image 9

3.2.2 Acquisition There are many advanced features in the acquisition setting. Please follow the link above for more information. 1. Navigate back to the main μmagellan window. 2. Remove any unwanted grids from the auto-populated list. 3. Press the Saving tab to direct data to the appropriate folders. 4. Press the Time tab to set up a time lapse. 5. Press the Space tab, select Simple Z Stack and input the appropriate z-limits for each grid. 6. To run a single grid, press the Run Acquisition button at the bottom of the window, otherwise press the Run all button under the Setup multiple acquisitions. More options for acquisition include: 1. Surfaces (3D grids) 2. Individual time lapse for each grid 3. Space (To reduce collection of blank data) 4. Covaried Setting (Changing hardware parameters based on imaging position) 5. Drift complensation Please read the online guide or talk to a BIDC employee for more information. 10

4 Shut Down Procedure One your data is saved, you can now shut down the instrument. Please note: Due to limited hard drive space on the computer, user s data will periodically be purged to make room available. The best practice is to immediately transfer data to a personal hard drive while you shut down and clean the area. 1. Exit the software. 2. Lower the objective and take off your sample. 3. Shut down the Lambda XL. (a) Press Local then 2 to shut down lamp followed by 1 to confirm. (b) Turn off the power at the back of the unit. Note: The lamp should be on for 30 minutes before turning off and off for 30 minutes before turning on. If someone is scheduled after you, please leave the lamp on. 4. Press the power switch on the shelf to turn off components. 5. Log off the computer. 6. Clean up the area. 11

5 Trouble Shooting μmagellan Explore! mode launches with error Check that the Channels option is set to Channel. Scroll wheel does not change focus in Micro-Manager 1) Select the hand from the ImageJ window and try again. 2) The z-piezo might be at its limits. Go into the Multi-D Acq window and set the z-positions to zero. Manually bring you sample back into focus. Micro-Manager launches with an error Check that all the components are turned on. 12