Analog Input and Output Options for Syringe Pumps

Similar documents
LabVIEW Pump Control Program Using RS232 and USB

Connecting a Cisco Reader Module

MegaTrak MCU-9000 Harness Wiring

GSV-1A4 M12/2 M12/2. Highlights

7I74 MANUAL 8 CHANNEL RS-422 INTERFACE

Arrakis Systems 6604 Powell Street / Loveland, Colorado 80538

Operating Bulletin. Model BB3-232 and BB3-I-232 USB Multi-Drop Box. The Fastest Flow Controller Company in the World!

Automation Interface Requirements for System I/O Interface of a DPC I Welding System

CHAPTER Wi r i n g NOTICE:

STATUS Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

Chemical Analysis Group Gas Chromatograph Common Cabling \ Pin-outs Document A15707

7I44 MANUAL 8 CHANNEL RS-422 INTERFACE

Option 2. Nil. Note 1) Option 1. Note With waterproof seal and mounting screw. With waterproof seal and mounting screw. 48 conversion adapter

USB-to-I2C. Ultra Hardware User s Manual.

Local Set Point Module Operating Bulletin

Series B9 and C9 Electronic Metering Pumps. Instruction Manual. Manual No : 1796 Rev. : D Rev. Date : 11/2015

716 Converter. Programmable Data Converter

SRI-02 Speech Recognition Interface

SERIE 59. Rear clamp. Frontal clamp HIGH RESOLUTION HOLLOW SHAFT INCREMENTAL ENCODER FOR INDUSTRIAL APPLICATIONS

CFO100 Series. User Manual CRT141 & CRR141. CFO141 User Manual, , rev001

the Interactive Catalog

TEC 2000 INSTRUCTION MANUAL TEC SERIES

Driver/Controller Development Assistance Package For IDEX Health & Science TitanHT TM Driver Board

STYLE 3600 SWITCH INTERFACE TRANSMITTER INSTALLATION AND OPERATING INSTRUCTIONS

PR602LCD v1.0 Installer Manual Rev. A

Model 7701 Multiplexer Module

Parrot MULTICOMM For FORD steering wheel

Incremental Encoder Adapter Page 1 of 8. Description. Features. Differential Cable Driver/Receiver:

MODEL A-30 ALTITUDE DIGITIZER OPERATION MANUAL INSTALLATION MANUAL

Industrial Power Supply Connectors

Issue : 1.1 Date : 19/1/2004. Trigger IO. C-Cam Technologies. a division of. Vector International 1 / 6

6A27 6A40 6A68 6A154 6A175

4.7 Expansion Module MAC00-FS1/FS4

Gotharman s DIY demoon

4.8 Expansion Module MAC00-FS1/FS4

BLU- CIF Interface Board

Instruction Manual. For file reference, please record the following data:

Connecting LEDs to the ADB I/O

HAWK MV-4000 Smart Camera

Installation Instructions for: Channel Thermocouple Amplifier

USB-to-I2C. Professional Hardware User s Manual.

RK-A CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER

955S BRIK. Series 955S Brik. Smart BRIK LDT ABSOLUTE PROCESS CONTROL KNOW WHERE YOU ARE... REGARDLESS. Linear Displacement Transducer

KOBOLD Compact Electronics

DIGITAL DISPLAY. for Industry Applications. Series PAXD. Key-Features:

Connections to PAD-LA Series

FVT/FVR20D2I1C4E INSTALLATION AND OPERATION MANUAL

QUICK INSTALLATION. MSX-E1731 Ethernet multifunction counter system /2016

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement

BehringerMods.com. Instructions for modification of Behringer SRC analog inputs and outputs

Win-I2CUSB Hardware User s Manual

BLU- CIF Interface Board

Microminiature and Subminiature Gauging DVRT Quick Start Guide

ATT2 Installation Guide for PQube 3

Connecting a Cisco Input Module

LAA-5 Pinout, Pins J2.24, J3.24 and J3.25 are internally connected to the main power ground.

Power connections 3) Power from 2024 to Control box (The valve harness connected) Ground (0V) +24 V. The voltage between pins 1 & 2 in JPT# 123

EC07 EC63. Compatible actuators: ERL2 ESD2. How to order EC 07 B. Series. B Installation. method

Lab 4: Introduction to ELVIS II+ Introduction to ELVIS II+

Additional board ZIB2-PDx-N for Plug & Drive motors PDx-N

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies

Digital Keypad Introduction

NEW product. Large digital pressure sensor 31series. Easily viewable LCD dual displays. High level visibility with 3-color display.

ES-600 Ozone Controller Operation Manual

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

PURE SINE WAVE INVERTER REMOTE GP-ISW-R-12/24

Centroid ACORN CNC controller Specification and Use Guide Updated 8/3/17. Overview

DIGITAL ANZEIGE. for Industry Applications. Series PAXP, PAXDP. Key-Features:

Trigger I/O Manual. Trigger IO Manual. C-Cam Technologies. a division of. Vector International. Date :12/01/2011 Page 1 / 8 Version 1.

RS-232 Connector Pin Assignments

Easy Arc Integration into a Simatic S7-300 Process Control System

INSTALLATION INSTRUCTIONS Model 930 EntryCheck

CBL Starter Cable Kit for DuraNAV System

Series Watt DC Power Supplies

Accsense Wireless System. Temperature Monitor

Eco Sensors OZONE CONTROLLER Model OS-6 Instructions for Use. General and New Features

Alicat Scientific Operating Bulletin. Alicat Scientific P and PC Series Pressure Gauge and Controller FLOW. Tare PSIG.

BLU- CIF2 Interface Board

GB MIX COMUNICAZIONE - MI. Digital Electro-pneumatic Regulator ER Series

PDS-100 Programmable Dispensing System

NEW product. Large digital pressure sensor 31series. Easily viewable LCD dual displays. High level visibility with 3-color display.

INSTALLATION INSTRUCTIONS. This document describes the installation procedure for the SAB-400 and SAB-401 Station Audio Board.

Lab 5: Access Control Training Station

Operation MANUAL. CARBOMETER - CO2 concentration measurement unit. LAMBDA CARBOMETER - CO 2 concentration measurement unit

RK CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER (ANALOG OUTPUT OR START/STOP)

MC CO MODBUS ADDRESSABLE SENSOR

GPIO CONFIGURATION. Screen Touch-Pads <LOCAL RACK> Selects the GPIO in the Local Rack (16 GPIO channels).

R-DAS Tiny User s Manual

QRPometer Assembly Manual Copyright 2012 David Cripe NM0S The 4 State QRP Group. Introduction

NeXGen Controller Installation Guide for Bennett Equipment

Parrot MULTICAN for Mini cars (part 2)

LVX Control Unit. Features:

TV-Free CI-TF-UCONN22. For Fiat, Chrysler, Dodge Uconnect 8,4 system with 22 pin connector

EAM 58 B / C - 63 A / D / E BIT PARALLEL - SSI

Troubleshooting the Digisolver for Allen Bradley SLC-500 Programmable Logic Controllers

RTU560 Connections and Settings DIN Rail RTU 560CIG10

MouldFlo Wiring Manual

1. Mount the echo and tremolo control switches under the keyboard shelf, in a position convenient for the organist.

Dyna-Trax II User Information

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

Transcription:

Analog Input and Output Options for Syringe Pumps Installation and Operation Instructions Syringe Pumps Technical Bulletin TB03 Overview Teledyne Isco offers several options for external analog control and for output signals to monitor flow rate and pressure: Analog Feature Voltage: 0-10V Current: 4-20mA Input Standard feature, Optional, Internal circuit board External converter box Output Optional, Internal circuit board Optional, Internal circuit board A single controller provides interfacing for up to three pumps. Voltage and current inputs control functions of run/stop, flow setpoint, and pressure setpoint. Voltage and current outputs transmit data for flow rate, pressure, and volume remaining in the cylinder. Analog Input Up to three pumps can be controlled externally for either constant flow or constant pressure. The analog voltage input is a built-in standard feature, and accepts a signal of up to 10 volts. Voltage Input Wiring Connections For analog control of a pump, connect the analog ground wire to the GND terminal under ANALOG INPUT on the ACCESSORIES terminal block on the rear panel of the controller (Figure 1). Connect the input wire to terminal 1. To control a second and third pump, connect input wires to terminals 2 and 3, respectively. A multi-pump system running in continuous flow mode only requires connection to terminal 1. 4-20mA Input Wiring Connections Up to three pumps can be controlled externally for either constant flow or constant pressure. For the 4-20mA converter box, connect the signal source to the 25-pin D connector on the side of the box. Table 1 describes the function of each terminal on the block, and also the converter box wiring (Figure 2, next page). Table 2 describes the pin functions for the 25-pin D input connector on the converter box. The 4-20mA input option requires an external converter box connected to the rear panel of the controller. The box has two external cables, labeled Cable A for control of pump A and Cable B for control of pump B. The output for the converter box is limited to a maximum signal of 8 volts. Table 1: Controller Terminal Connections for 4-20mA Input Terminal Name Function 4-20mA Wiring Digital Input 1 Run/Stop, Pump A Cable A: White Digital Input 2 Run/Stop, Pump B Cable B: Green Digital Input 3 Direction, Pump A Cable A: Green +15 Volt 15VDC to converter box Cable A: Red Digital Ground 15V Rtn Cable A: Black Digital Ground Run/Stop Rtn, Pump A Cable A: Blue Digital Ground Direction Rtn, Pump A Cable A: Yellow Digital Ground Run/Stop Rtn, Pump B Cable B: Yellow Digital Output A2 Run/Stop Status, Pump A Cable A: Orange Digital Output B4 Run/Stop Status, Pump B Cable B: Brown Analog Ground Flow/Pressure Rtn, Pump A Cable A: Gray Analog Ground Flow/Pressure Rtn, Pump B Cable B: Blue Analog Input 1 Flow/Pressure Ctrl, Pump A Cable A: Purple Analog Input 2 Flow/Pressure Ctrl, Pump B Cable B: White 4 ma = Stop 20 ma = Max flow in constant flow mode; max pressure in constant pressure mode Figure 1: Analog voltage input/output terminals pump A pump B pump C Run/Stop open collector active low analog ground pump A pump B pump C 0-10V analog input Note When using the 4-20 ma converter box, if a third pump is added, another cable will be installed and labeled Cable C. 0 Volts = Stop Vmax = Max flow in constant flow mode; max pressure in constant pressure mode

Cable A connections white green red black blue yellow orange gray violet 0-8 volts (with box) 0-10 volts (w/o box) 4-20mA converter box to control signal Refer to Table 2. Cable B connections green yellow brown blue white 0-8 volts (with box) 0-10 volts (w/o box) Cable A Cable B Figure 2: Analog control wiring connections Table 2: 4-20mA Input Pin Functions for Operator Interface Pin Functions Value Pin Functions Value 1 Chassis 14 Flow Pump A Rtn 2 Flow In, Pump A 4mA = 0 ml/min 15 Flow In, Pump B 4mA = 0 ml/min 3 Rtn Flow, Pump B 16 Flow In, Pump C 4mA = 0 ml/min 4 Rtn Flow, Pump C 17 R/S status, Pump C Contact Closure 5 R/S status, Pump C Common 18 R/S status, Pump B Contact Closure 6 R/S status, Pump B Common 19 R/S status, Pump A Contact Closure 7 R/S status, Pump A Contact Closure 20 Direction, Pump C Logic High = Up 8 Run/Stop, Pump C Logic High = On 21 Direction Rtn, Pump C 9 Run/Stop Rtn, Pump C 22 Direction, Pump B Logic High = Up 10 Run/Stop, Pump B Logic High = On 23 Direction Rtn, Pump B 11 Run/Stop Rtn, Pump B 24 Direction, Pump A Logic High = Up 12 Run/Stop, Pump A Logic High = On 25 Run/Stop Rtn, Pump A 13 Direction Rtn, Pump A NOTE: Pump C references are not enabled at this time. Only Pump A and Pump B connections will be used. When Pump C is enabled, another cable from the interface box will be installed and labeled Cable C. Also note that the Direction control for Pump A (pins 24 & 13) will provide direction control for both pump A and B in dual pump mode.

Preparation Before programming the controller to accept the analog signal, you must know the value of the maximum flow rate (constant flow) or the maximum pressure (constant pressure). Note The high and low limits you enter cannot exceed the pump specifications: 4-20mA maximum input = 8V; Normal input = 10V. To determine the analog voltage range, use the formula explained below: Setting Up External Control or (CP or CF and selected pump are displayed) (EXTERNAL) (On/Off) where: V FS --- U FS = Maximum flow rate or Maximum pressure (Full Scale): This is the same value entered when setting LIMITS, as described in Section 3 of your user manual, for MAX PRESS or MAX FLOW. V = Volts per unit of flow or pressure U = Incremental unit of flow or pressure Vmax = Maximum input voltage Calculation Examples = Vmax Flow Rate If MAX FLOW is set at 25 ml/min in the LIMITS menu and the desired scale factor is 5.0 volts per 20 ml/ min, do the following to determine the analog voltage range: 5 25 ----- 20 = 6.25 (MORE) (SET RANGE) Enter Vmax value Figure 3: Button sequence for external control setup Analog Outputs Analog outputs are optional and installed inside the controller for monitoring flow rate (up to 3 pumps), and volume (up to 2 pumps). Voltage Analog Output Output voltage for pressure is located on the back panel of the pump module via standard banana jacks. The output connection is the female 25-pin Sub-D connector located on the rear panel of the controller, under ANALOG OUTPUT. In this example, the analog voltage range would be 0.0 6.25 volts. Pressure If MAX PRESS is set at 510.2 ATM in the LIMITS menu and the desired scale factor is 2.0 volts per 100 ATM, do the following to determine the analog voltage range: 25-pin cable Controller rear panel 2 510.2 -------- 100 = 10.204 In this example, the analog voltage range would be 0.0 10.204 volts. Figure 4: Analog voltage output connection, rear panel Voltage analog outputs are pre-installed at the factory. This option has a total of four outputs for monitoring flow rate with selectable ranges from 0 to 5V, -5 to +5V, and 0 to 10V. The default range at installation is 0 to 10V. To adjust the range, remove the top cover of

the controller, (Figure 5), and move the jumper for each output to a new range (Figure 6). Each output can be set to a different range. The top of the selected range is the same maximum value entered for MAX FLOW when setting LIMITS, as described in Section 3 of your user manual. For example: if the output range is 5V and the preferred output scale is 1 volt per 10 ml/min, MAX FLOW is set at 50 ml/min. MAX FLOW cannot exceed the pump specification. A controller running one to two pumps can also output cylinder volume on the remaining outputs. The volume output range is not adjustable. Full scale is equal to one pump stroke. Output #2 Output #1 Output #4 Output #3 Analog output board Jumper +/- 5V + 10V + 5V Figure 5: Accessing the analog output board (Voltage output board shown) Figure 6: Adjusting the voltage output ranges Table 3: Voltage Output Signal/Sub-D Pins Pin # Description One/Two Pump Function Three Pump Function 1 Chassis (earth) 2 Analog common 3 Analog common 4 Analog common 5 Analog common 14 Output #1 Flow Rate A Flow Rate A 15 Output #2 Flow Rate B Flow Rate B 16 Output #3 Volume A Flow Rate C 17 Output #4 Volume B Volume B

4-20mA Analog Output The 4-20mA analog output board is pre-installed at the factory, but can also be purchased in an upgrade kit. The board initially contains three output cards, for flow rate, pressure, and volume for Pump A. Additional cards can be purchased to output data for one or two more pumps. Table 4: Analog Output Signal/Sub-D Pins for 4-20mA Channel Data Type/Pump Pin 1 Pump A Flow Rate 1 1 Pump A Flow Rate (Signal Return) 14 2 Pump A Pressure 2 2 Pump A Pressure (Signal Return) 15 3 Pump A Volume Remaining 3 3 Pump A Volume Remaining (Signal Return) 16 4 Pump B Flow Rate 4 4 Pump B Flow Rate (Signal Return) 17 5 Pump B Pressure 5 5 Pump B Pressure (Signal Return) 18 6 Pump B Volume Remaining 6 6 Pump B Volume Remaining (Signal Return) 19 7 Pump C Flow Rate 7 7 Pump C Flow Rate (Signal Return) 20 8 Pump C Pressure 8 8 Pump C Pressure (Signal Return) 21 9 Pump C Volume Remaining 9 9 Pump C Volume Remaining (Signal Return) 22 Part Numbers for Analog Options Part numbers are listed in the table below. Contact Teledyne Isco for additional details and ordering information. Table 5: Part Listing of Analog Options Description Part Number 4-20mA analog input converter box 68-1247-119 Pump controller w/ analog voltage outputs 68-1240-046 Analog voltage installation kit 68-1247-070 4-20mA analog output for one pump (3 outputs) Additional output cards 68-1247-120 68-1247-121 Last modified January 4, 2010 Teledyne Isco, Inc. P.O. Box 82531, Lincoln, Nebraska, 68501 USA Toll-free: (800) 775-2965 Phone: (402) 464-0231 Fax: (402) 465-3001 E-mail: IscoService@teledyne.com Teledyne Isco is continually improving its products and reserves the right to change product specifications, replacement parts, schematics, and instructions without notice.