Biology 105 Laboratory 1: Microscope

Similar documents
The Microscope: A Tool of the Scientist

Lab 5 Microscopy. Introduction. Carrying the Microscope. Depth of Focus. Parts of the Compound Microscope. Magnification

I. Use and Care of the Microscope:

Lab 4 - The Microscope

Introduction to the Microscope. Types of Microscopes Care Parts Focusing

The Microscope. Lab Exercise. Contents. Light Microscope. Objectives. Scanning Electron Microscope. Introduction. Transmission Electron Microscope

Title: Using a Compound Light Microscope and Comparing Cells - Bacteria, Plant Cells and Animal Cells

Introduction to the Microscope. Types of Microscopes Care Parts Focusing

Introduction to the Microscope. Hanan Jafar BDS.MSc.PhD

Introduction to the Microscope. Types of Microscopes Care Parts Focusing

The Microscope. The History. How a Microscope Works. How a Microscope Works 4/28/16. The History. The History. Mr.

Richter Optica. Instructions for Model F-1 Compound Microscope

EXPERIMENT. Microscopy for Microbiology Use and Function

Use of the Binocular Microscope

USE OF THE MICROSCOPE

Microscope Introductory Lab Series

The History. Many people experimented with making microscopes. Was the microscope originally made by accident? (Most people were creating telescopes)

Introduction to Microscopes! Now Seeing IS Believing!

Great Bay Community College Document Number: Corporate Drive Revision Number: 3

INSTRUCTIONS FOR MODEL C-1028 COMPOUND MICROSCOPE

Why were cells not discovered before the 17th century?

MICROSCOPE CARE AND USE

Activity #35 A License To Learn Part One: Earning a License Challenge Question: Initial Thoughts:

Hand sectioning a mushroom

Eco D-ELS-4 Stereo Microscope. Instruction Manual

M40 Microscope User s Manual

Richter Optica. Instructions for Model: MDS1 Compound Microscope

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

Key Terms. Magnification Objective Aperture Diaphragm. Tissue Organism Organ Cell Artificial organ Dialysis

Bell Work: Brainstorm: What s the smallest thing that can be seen with current microscope technology? Please raise your hand when you have your idea!

BIOLOGICAL MICROSCOPE

Operation Manual (Version 1.0.0)

G500 SERIES MICROSCOPE OPERATION MANUAL For use with MFR# G502, G502T, G504, G504T, G505, G505T.

The History. Many people experimented with making microscopes. Was the microscope originally made by accident? (Most people were creating telescopes)

Compound Microscope Instruction Manual

Instruction Manual. B1 Series. Copyright 02/02 Motic Microscopes, European Division

Polarization Microscope. Omano OM349P Series. User Guide

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

Lab Exercise 5: The Smear and Simple Staining

INTRODUCTION - VetScan HDmicroscope

Mi50 Inverted Microscope User s Manual. Fein.

Digital Comprehensive Scope 2. Instruction Manual T-19541C T-19641C

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, TX Phone (210) Fax (210) INSTRUCTIONS FOR

Richter Optica. Instructions for Model: HSX1 High School Microscope

OMM200 and OMM200T. Metallurgical Microscope

Richter Optica. Instructions for: S6-BL, S6-TS, S6-SPS, S6-ILST Models

Teaching Microscope Instruction Manual OMTM85 and OMTM85L

kena Instruction Manual T-1050

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system.

Operation Manual. for Micromas ter I Micros copes. T his manual covers catalog numbers : S 11031, S 11033, and S 11035

Kids Microscope. Instruction Manual for MI-1000STD Kids Microscope MI-1000LED Kids LED Cordless Microscope

Binocular Compound Microscope Instruction Manual

OPTIKA. ECOVISION Series. Entry-level biological microscopes for students. Entry-level biological microscopes for students

Instruction Manual. B3 Series. Copyright 02/02 Motic Microscopes, European Division

Comprehensive Scope 2. Instruction Manual T-19011C - Monocular T-19021C - Dual Viewing T-19031C - Binocular T-19041C - Trinocular

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

M20 Metallurgical Microscope User s Manual. Fein.

3025 MICROSCOPE SERIES INSTRUCTIONS

Swift SM100 Series Stereo Microscope

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

Symbio Microscope. L1500 Series. Biological Upright Compound Microscopes. Operating Manual

D.I.Y. Microscope Kit

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

DSU Start-Up instructions

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com

Richter Optica. Instructions for Model: MDS2 Middle School Microscope

User Manual. Binocular Compound LED Microscope. MicroscopeNet.com

R40POL Microscope User s Manual

kena TM Instruction Manual T-1050

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Swift M3600, M3700 Series Digital & Compound Microscope

INSTRUCTION MANUAL POLARIZING & ORE MICROSCOPES. Polarizing Microscopes. Ore Microscopes. Model: NYMC62B00

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

BLIPS perfectly aligned cleaned by any impurity Gently press the lens

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

S-series T-series. english

Operation Manual (Version 1.0.0)

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

The image is virtual and erect. When a mirror is rotated through a certain angle, the reflected ray is rotated through twice this angle.

INSTRUCTION MANUAL for STEREO MICROSCOPE FUTURO EDU CAM

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

S20 STEREO MICROSCOPE

Olympus UNIVERSAL ILLUMINATOR LSD MODEL WITH TRANSFORMER TC

Microscopes. Passion for Science PAGE PAGE PAGE PAGE PAGE

OLYMPUS STEREO MICROSCOPES I INSTRUCTION MANUAL I MODELS VMF,VMT&VMZ OLYMPUS

X-series C-series. english

148 Series. User s Manual

STEREO MICROSCOPE MODEL K400 / K500 / K700 INSTRUCTION MANUAL

EXAMET-4 MANUAL MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

Fisher Scientific Microscopes

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

MICROSCOPE USER MANUAL

Richter Optica. Instructions for Model: S-2 Dual Power Stereo Microscope

LABOMED, INC. APPLICATION

Transcription:

Biology 105 Laboratory 1: Microscope A. The Compound Microscope The microscope is a delicate, precision instrument. It must be treated gently, as the slightest bump or jar may damage the alignment of its lenses. Carry the microscope with care by grasping its arm with one hand, keeping the instrument close to your body and supporting its base with the other hand. Always carry the microscope with two hands. Microscope lenses are to be cleaned before and after each use with lens paper. NEVER wipe a lens with paper towels, a handkerchief, or Kleenex. If you wear corrective glasses for astigmatism, you must use them with the microscope. However, if your glasses are for near-sightedness or far-sightedness, you may use the focus of the microscope rather than your glasses. Parts of the Compound Microscope 1. The ocular is located at the upper end of the inclined body tube. The ocular contains two lenses with a total magnification of 10X. 2. The inclined body and the nosepiece hold the ocular lenses and objective lenses at a set distance from each other. The body stays in place and the stage is raised or lowered for proper focusing. 3. The rotating nosepiece is attached to the lower end of the focusable nosepiece and the three objective lenses are mounted on the rotating nosepiece. The rotating nosepiece may be turned in either direction, so the objective may be changed while viewing a specimen. 4. The three objective lenses, so called because they are closest to the object or specimen, have individual magnifying powers of 4X, 10X, and 40X. The 4X objective lens is called the low power objective, the 10X objective lens is the medium power objective, and the 40X objective lens is the high power objective. When an objective is placed over the specimen, the objective lens will produce a magnified image which will be viewed through, and further magnified by, the ocular at the top of the body tube. A combined magnifying action makes this instrument a compound microscope. The magnifying power of the ocular multiplies or compounds the magnifying power of the objective. The total magnification is the magnification of the ocular multiplied by the magnification of the objective. For example, on low power the total magnification is 4 x 10 = 40. 5. The stage is mounted perpendicularly from the arm and below the nosepiece. This part supports the specimen slide over a hole that admits light for specimen illumination. 6. The mechanical stage has a spring-clip to hold the specimen slide in place. Two control knobs move the mechanical stage (and therefore the specimen slide) for viewing. 1

7. The arm supports the body tube. The arm is connected to the base and is used to carry the microscope. 8. The base bears the weight of the microscope. 9. The light source or in-base illuminator, built into the base, illuminates the specimen. The illuminator has a curved outer lens that directs light through the stage opening. The light switch is located on the base. 10. The iris diaphragm is located on the underside of the stage. It may be opened or closed to regulate the amount of light reaching the specimen. The best detail is seen with the lowest light intensity. 11. The under-stage condenser further directs the light from the illuminator to the specimen. You will see a knob for the condenser under the stage. Normally you will not ever have to adjust the condenser. 12. The coarse focus adjustment knob moves the focusable nosepiece up or down in initial focusing. Since a small turn of the coarse focus moves the focusable nosepiece a large distance, it is used only to bring the specimen into view, not into fine focus. The coarse focus knob is NEVER used when the 40X objective is in position above the specimen. If the coarse focus is used with the 40X objective, the objective could ram into the slide causing damage to the slide, or even worse, damage to the objective lens itself. 13. The fine focus adjustment knob permits exact focusing by moving the focusable nosepiece up or down very slightly. The fine focus also allows focusing at different levels within the specimen. All slide material has depth, no matter how thin the specimen; this third dimension can only be seen by continually focusing up or down through the material. This is a critical aspect of focusing, and a very important technique in good microscopy. Focusing Turn the low power objective lens (green band) into position. With the coarse focus adjustment knob, slowly move the objective downward as far as it will go. Note the distance between the point of the objective and the stage. You will save yourself time by adjusting the focus to this approximate level at the beginning of each operation. With the object in perfect focus under low power magnification, turn the high power objective into position and perfect the focus with the fine focus adjustment knob. Do not use the coarse focus knob with the high power objective lens. Note how close the high power objective lens is to the slide. There is no "stop" when focusing with this objective and too great a downward movement may result in cracking of the slide or scratching of the lens. 2

THE PARTS OF THE MICROSCOPE Please complete the following table in your lab notebook: Objective Lens 4X Total Magnification? 10X 40X 3

Please complete the following procedures and answer the questions or provide the requested drawings in your lab notebook. Procedure 1: Examining the Letter e Step 1: Place a slide of the letter "e" on the stage. Step 2: Focus the slide first on low power, using the coarse and fine focus knobs, then increase the power and refocus using ONLY the fine focus knob. Question 1: How does the position of the letter e look when viewed under the microscope? For example, is the position of the letter under the microscope the same as when viewed on the slide with your naked eye? If not, how is the position different? Question 2: Can you see more area of the slide on the low or high power? Why? Procedure 2: Examining Colored Threads Step 1: Obtain four prepared slides of colored threads. Step 2: Focus each of these slides as described above, and answer the following question for each of the slides. Question 1: Which color is on the top and which color is on the bottom for each of your slides? Procedure 3: Examining Prepared Slides Step 1: Obtain at least four prepared slides. Step 2: Focus these slides as described above. Step 3: Sketch each of the slides in your lab notebook. Procedure 4: Preparing a Wet Mount Slide Step 1: Obtain a clean, blank microscope slide and a coverslip. Step 2: With an eye-dropper, add two drops of pond water to the center of the slide. Step 3: Holding the coverslip at a 45 degree angle, slide it over to contact the drop of water and gradually let it fall into place over the water drop. If the coverslip is dropped from too flat 4

an angle, many air bubbles will form and interfere with accurate observations. When viewed under a microscope, air bubbles usually appear as perfectly round with dark black edges. Step 4: View your slide under low and high powers, and note the variety of organisms present in the pond water. Sketch the slide in your lab notebook. 5