1 UVU444 DEPT on the U400 FL-26JUL94CD PRELIMINARY INFORMATION 1) The DEPT (Distortionless Enhancement by Polarization Transfer) experiment is frequently used to identify the types of carbons (i.e., C, CH, CH 2, and CH 3 ) in a molecule. Depending on the setting of the parameter mult, a different DEPT spectrum (or spectra if mult is arrayed), and consequently, different results can be obtained: mult=0.5 DEPT-45 (gives all protonated carbons) mult=1.0 DEPT-90 (gives CH's only) mult=1.5 DEPT-135 (gives CH's and CH 3 's up and CH 2 's down) Quaternary carbon signals are suppressed in the DEPT experiment, and therefore, can be easily identified upon comparison of any DEPT spectrum with the regular 13 C spectrum. Routinely, only a DEPT-135 is needed to identify all the carbons since CH's and CH 3 's usually have quite different chemical shifts. Occasionally, a DEPT-90 may be needed to distinguish CH's from CH 3 's. The DEPT experiment can also be set up to acquire a set of four spectra (mult=0.5,1,1,1.5 for one DEPT-45, two DEPT-90 and one DEPT-135) for use in automatic spectral analysis and/or editing (see details in Part 4 below). Finally, the DEPT experiment is somewhat more sensitive than the APT or normal 13 C experiment. In addition, the setting of the parameter d1 is governed by the T1's of the proton nucleus instead of the carbon nucleus. This could be advantageous if your carbons' T1's are longer than those of the protons, which is the case for most molecules. 2) pw90 and pp: The default pw90 and pp are usually good enough for this experiment, unless the physical properties of your sample are very different from those of the standard due to, for example, the presence of ionic or paramagnetic materials or the use of a very different solvent. Talk to MSL staff if you are not sure if you need to calibrate these parameters for your sample or don't know how. If new values are determined, remember to change the parameter pw90 before setting up the DEPT experiment, and change the parameter pp before starting the DEPT experiment. 3) d1: By default, d1 is set to 1.5 seconds. In general, d1 is set to 1 to 2 times the value of T1 for the PROTONS of interest. 4) nt: By default, nt is set to half the value used in your regular 13 C spectrum due to its higher sensitivity. 1. COLLECT A 13 C SPECTRUM If you have already collected a 13 C spectrum previously, load the data into an experiment (e.g., exp1) and process it as usual. Note that when set up this way, the DEPT spectrum may have to be phased manually. If no 13 C spectrum has been acquired, acquire one and phase and reference the spectrum as usual. 2. SET UP THE DEPT EXPERIMENT dodept INTRODUCTION: dodept is an interactive macro that will take you through the setup of a DEPT experiment. In this handout, the dialog will appear in this font, and the symbol signifies that a response is required. Most messages will appear in the Status Window, while inputs are entered via the Input Window. Window arrangement on the left side of the screen is, from top to bottom, Status, Input, Menu, Graphics, and Text. 2A. Enter the Macro In the experiment that contains the 13 C spectrum, enter: dodept <rtn> macro to set up a DEPT experiment in exp2 NOTE: If you wish to do the DEPT in a different experiment, enter dodept(exp#), where exp# is the experiment number for the DEPT. Experiment number must be between 1 and 9, excluding 5.

2 If the DEPT exp# is the same as the current experiment, the following dialog will appear: Do dept in current experiment? (y/n) If answered no: Enter dept exp#: Then the following menu will be displayed in the text window at the bottom of the screen: DEPT MENU 1) DEPT-135 2) DEPT-90 3) DEPT-135 and DEPT-90 4) DEPT for spectral editing 2B. Choose a Setup Enter your choice of setup at the following prompt in the Input Window: Enter your choice (1-4) from the menu below: If 1 is chosen, a DEPT-135 experiment will be set up (i.e., mult=1.5). If 2 is chosen, a DEPT-90 experiment will be set up (i.e., mult=1). If 3 is chosen, an array will be set up for a DEPT-135 and a DEPT-90 (i.e., mult=1.5,1). If 4 is chosen, an array will be set up to collect a set of four spectra: one DEPT-45, two DEPT-90, and one DEPT-135 (i.e., mult=0.5,1,1,1.5). This setup is recommended for spectral editing and/or analysis only, because the data will take four times as long to acquire as that in choice #1 and 2 and twice as long as choice #3. By default, d1 will be set to 1.5 seconds and nt to half of that of the regular 13 C spectrum. However, you may change these parameters by answering the following question accordingly: Do you wish to change nt and/or d1? (y/n) If answered yes, enter the new values at the following prompts: nt= d1= The dodept macro will terminate at this point. Be sure to check the parameters before starting the experiment. 2C. Start the Experiment You may start the DEPT experiment with the au or ga command: au start DEPT experiment and process the data automatically with the deptproc macro when acquisition is complete. NOTE: If the experiment is terminated with aa or sa (see below), no data processing will occur. ga start DEPT experiment and do a wft when acquisition is complete. NOTE: Once the experiment is started, it may be aborted by clicking the Abort Acq button or typing the aa command if no data is to be retained, such as when you discover that a mistake has been made in the setup of the experiment, or when it has become apparent that the experiment is useless. If data retention is desired, such as when sufficient signal-to-noise has been obtained or when you have run out of time, you should use the sa 2

3 command to stop the acquisition. At this point, you should process the data manually because automation will not work when the experiment is terminated before its completion. When the experiment is finished, save the data and proceed with its processing and plotting as described below. 3. PROCESSING AND DISPLAYING NOTE: If a regular 13 C spectrum is acquired together with the DEPT data, it is most likely that no further phasing of the DEPT spectra will be necessary. In that case, you should use the deptproc macro to process the data. On the other hand, if the 13 C spectrum is collected at another time, the DEPT spectra will probably require some phasing, either auto or manual, depending on the type of data. If your data contains a DEPT-45 (preferred) or a DEPT-90 spectrum (may or may not work), you may use the macro deptpaph for processing. Otherwise, use deptproc to process the data and phase the spectrum manually if necessary. * * * * * A. To process and display DEPT data without autophasing: deptproc <rtn> B. To process and display DEPT data with autophasing: deptpaph <rtn> C. To display DEPT data without any processing: deptds <rtn> 4. PLOTTING A. deptplot for data acquired with menu item #1-3 above. It will plot DEPT data, with or without the regular 13 C spectrum on the same page. See examples in Fig deptplot(c13exp#) <rtn> C13exp# is the experiment number that contains the 13 C spectrum < > If no argument is provided with the deptplot macro, the following dialog will appear: Plot without 13C spectrum? (y/n) If answered yes, the macro will terminate here and only the DEPT spectrum (or spectra) will be plotted (see example in Fig. 1). If answered no, the dialog will continue: Enter exp# containing 13C spectrum: Finally, you are provided with the option to plot the text: Plot with text? Then the plotting will begin. B. adeptpp for data acquired with menu item #4 only. It will plot the edited spectra and print the analysis table, OR plot the unedited spectra. See examples in Fig adeptpp <rtn> enter the plotting macro Then enter your choice of plotting option at the following prompt: Plot 1)edited or 2)unedited spectra? 3

4 If option 1 is chosen, spectral editing will be performed and the resulting four new spectra and a spectral analysis table will be plotted and printed respectively. The edited spectra will contain, from bottom to top, all protonated carbons, CH carbons, CH 2 carbons, and CH 3 carbons respectively. If option 2 is chosen, the unedited spectra will be plotted (i.e., from bottom to top, DEPT-45, -90, and -135). You will also have a choice of plotting with the parameters, text only, or none at all: Plot with 1)parameters, 2)text only, or 3)none? NOTE: If option 1 is chosen, wc (width of chart) will be adjusted accordingly to accommodate the parameters. Next, the macro will determine if the data have been edited, and if it has, the data will be retransformed < > If you choose plotting option 1, the threshold will be adjusted and a line list will be printed in the text window. The number of lines found will also be shown in the status window. Check the threshold setting by answering the following question accordingly (NOTE: only protonated carbons should be and is counted!): Is the number of peaks picked correct? If you answer no to the above question, enter 1 if too many peaks are picked or 2 if too few: 1)increase or 2)decrease the threshold? The threshold will then be increased or decreased by 20% and the above dialog will be repeated until the correct number of lines are picked. Then, spectral editing and analysis will be performed on the data and the analysis table will be printed Finally, the DEPT spectra, edited or not, will be plotted. C. adeptplot for data acquired with menu item #4 only. It will plot the edited spectra and the spectrum analysis table, all on the same page. See example in Fig. 5. adeptplot <rtn> enter the plotting macro First, the macro will determine if the data have been edited, and if it has, the data will be retransformed. Then, the threshold will be adjusted and a line list will be printed in the text window. The number of lines found will also be shown in the status window. Check the threshold setting by answering the following question accordingly (NOTE: only protonated carbons should be and is counted!): Is the number of peaks picked correct? If answered no, enter 1 if too many peaks are picked or 2 if too few: 1)increase or 2)decrease the threshold? The threshold will then be increased or decreased by 20% and the above dialog will be repeated until the correct number of lines are picked. Then, spectral editing and analysis will be performed on the data, and the resulting edited spectra and analysis table will be plotted on the same page. D. adeptprint for data acquired with menu item #4 only. Prints spectrum analysis table. After the macro is entered, it will determine if the data has been edited. If it has, the results will be printed. If not, a dialog similar to the one for adeptplot will appear, and at the end, a spectral analysis table will be printed. 4

5 Fig. 1. deptplot: DEPT-135, with 13 C spectrum Fig. 2. deptplot: DEPT-135 and DEPT-90, with 13 C spectrum Fig. 3. adeptpp: edited spectra, with parameters (spectrum analysis table on next page) Fig. 4. adeptpp: unedited spectra, with text only Fig. 5. adeptplot: typical output, with text 5

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 1 Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 Please note: Under no circumstances move the magnet or the automatic sampler table. Do not attempt to use the auto sampler

More information

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy PINMRF Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy INCLUDING: AV-III-400-HD w/ 5mm BBFO SmartProbe 369 WTHR AV-III-500-HD w/ 5mm BBFO

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500 August 9, 2016 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they

More information

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800 Walkup NMR Varian NMR Spectrometer Systems With VNMR 6.1C Software Pub. No. 01-999159-00, Rev. A0800

More information

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

c. Saturday and Sunday: 10 minute time blocks and unlimited time. Scheduling 400 MHz A with 100 slot sample changer: a. Monday to Friday from 8 am to 8 pm: 10 minute blocks, 30min max. b. Monday to Friday from 8 pm to 8 am: 10 minute blocks, unlimited time. c. Saturday

More information

NMR Training VNMRJ 4.2A/VNMRS 400

NMR Training VNMRJ 4.2A/VNMRS 400 NMR Training VNMRJ 4.2A/VNMRS 400 The VNMRS/new 400 MHz NMR uses VNMRJ 4.2A. Although VNMRJ is different than VNMRalmost ALL typed commands that you know from VNMR should work. LOGIN: Your username is

More information

NMR Users Guide Organic Chemistry Laboratory

NMR Users Guide Organic Chemistry Laboratory NMR Users Guide Organic Chemistry Laboratory Introduction The chemistry department is fortunate to have a high field (400 MHz) Nuclear Magnetic Resonance (NMR) spectrometer. You will be using this instrument

More information

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 January 10, 2019 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

2 Dimensional NMR User s Brief Guide

2 Dimensional NMR User s Brief Guide 2D NMR Handout 7/17/18 Weiguo Hu 2 Dimensional NMR User s Brief Guide (This handout presumes your thorough familiarity with the 1D handout commands and their use!) Options There are a vast number of 2D

More information

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy PINMRF Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table of Contents

More information

MNova Version Installation Instructions.

MNova Version Installation Instructions. MNova Version 12.0.0 Installation Instructions. Note: This Document contains information for a new NMR user. Carefully read and accomplish each section before moving to the next. If you can t finish a

More information

ADVANCED DATA PROCESSING

ADVANCED DATA PROCESSING ADVANCED DATA PROCESSING Making Backups The NMR lab periodically backs up data on CD-ROMs for your convenience ([Files] [Oldfids]). If you would like your own personal backups, you can do this via the

More information

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1 Gemini-300, Manual Operation (version 1.0, 4Feb95) 1 I. GETTING STARTED (LOGGING IN, LOGGING OUT, USING THE NMR PROGRAM). A. Logging into the Gemini To login onto the Sun workstations, hit the return key.

More information

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Once FTIR data has been acquired for the gases evolved during your DSC/TGA run, you will process using the OPUS software package. Select

More information

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open.

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open. Basic 1D Processing Opening Saved Data Start the VnmrJ software. Click File > Open. This will open a pop-up window. Clicking Home will take you to the data directory within your account. You can create

More information

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS TABLE OF CONTENTS Disclaimer... 2 Conventions... 2 Routine 1D NMR Experiments Procedure for acquiring 1D Proton spectrum...

More information

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy Delta V5 operation Widetron Technologies Corp. Wu, Chung Ying Connect to spectrometer 1. Start Delta V5 (double click ). 2. Open Spectrometer Control (click ). 3. Select spectrometer on Available Instruments.

More information

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005 MestReC Cheat Sheet by Monika Ivancic, July 1 st 2005 This Cheat Sheet is to be used at UW-Madison as a quick guide to processing using the MestReC NMR software. You may find more help at the MestReC homepage

More information

North Carolina State University Department of Chemistry Varian NMR Training Manual

North Carolina State University Department of Chemistry Varian NMR Training Manual North Carolina State University Department of Chemistry Varian NMR Training Manual by J.B. Clark IV & Dr. S. Sankar 1 st Edition 05/15/2009 Section 1: Essential Operations for Basic 1D Spectra Preparing

More information

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University 1. INTRODUCTION...1 1.1. About this Worksheet... 1 1.2. A Very Brief

More information

Processing data with Bruker TopSpin

Processing data with Bruker TopSpin Processing data with Bruker TopSpin This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

Frequently Asked Questions (FAQ)

Frequently Asked Questions (FAQ) Frequently Asked Questions (FAQ) 5/11/07 How can I import a spectrum into a Microsoft Word or PowerPoint document? Choose one: 1) One possibility is to transfer the FID to a computer in the NMR Lab or

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota Varian NMR Instructions - 1D Department of Chemistry NMR Facility University of Minnesota June 17, 2008 Table of Contents 1. General Information...2 2. References...2 3. Rules...3 4. Scheduling...3 5.

More information

1. Log in at the RESLOG terminal by entering your code and select the instrument you want to use.

1. Log in at the RESLOG terminal by entering your code and select the instrument you want to use. Skip to Main Content Home Policies Instruments Manuals Staff Useful Links Operating Instructions for Varian NMR Spectrometers Instructions for Operating Varian Mercury and Unity/Inova NMR Spectrometers

More information

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy PINMRF Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy INCLUDING: Avance DRX500-1 w/ 5mm TBI Probe - 367 WTHR Avance DRX500-2 w/ 5mm BBO or TXI

More information

Galaxie Photodiode Array Software

Galaxie Photodiode Array Software Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Galaxie Photodiode Array Software User s Guide Varian, Inc. 2002-2006 03-914950-00:6 Table of Contents Introduction... 3 General... 3 PDA

More information

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin. TopSpin QuickStart Preliminaries Sign Up Sign up for instrument time at calendar.yahoo.com/hmcnmr The password is nmr. During the summer, 1100-1200 and 1600-1700 are reserved for walk-on use (15 min maximum

More information

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised 4-21-2015 Relaxation times are divided into two types: longitudinal, which concerns change in magnetization along the

More information

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry NMR Spectroscopy with VnmrJ University of Toronto, Department of Chemistry Walk-up interface 1 Logging in 1 Starting VnmrJ 1 Inserting sample into the magnet or sample changer 1 Enter sample information

More information

MassHunter Personal Compound Database and Library Manager for Forensic Toxicology

MassHunter Personal Compound Database and Library Manager for Forensic Toxicology MassHunter Personal Compound Database and Library Manager for Forensic Toxicology Quick Start Guide What is MassHunter Personal Compound Database and Library Manager? 2 Installation 3 Main Window 4 Getting

More information

How to analyze complex NMR signals through spin simulation/iteration

How to analyze complex NMR signals through spin simulation/iteration How to analyze complex NMR signals through spin simulation/iteration 1. Introduction When a NMR signal s multiplicity exceeds the common doublet, triplet, quartet or doublet of doublet structure, analysis

More information

Agilent ChemStation Plus

Agilent ChemStation Plus Agilent ChemStation Plus Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004, 2006-2008 No part of this manual may be reproduced in any form or by any means (including electronic

More information

[ WHITE PAPER ] System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind INTRODUCTION

[ WHITE PAPER ] System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind INTRODUCTION System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind Explore how Empower Chromatography Data Software System Policies and privileges can be used to control

More information

SequencePro Data Analysis Application. User Guide

SequencePro Data Analysis Application. User Guide SequencePro Data Analysis Application User Guide SequencePro Data Analysis Application User Guide DRAFT October 31, 2001 12:52 pm, Title_page.fm Copyright 2001, Applied Biosystems. All rights reserved.

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

Quick Start Guide to nmrpipe Processing of 15 N-HSQC data

Quick Start Guide to nmrpipe Processing of 15 N-HSQC data Quick Start Guide to nmrpipe Processing of 15 N-HSQC data Data conversion: We make the basic assumption here that the data set was recorded on an INOVA-class or later model of NMR instrument running VNMRJ.

More information

General Information for Automated Inova 500 MHz NMR (Indy)

General Information for Automated Inova 500 MHz NMR (Indy) General Information for Automated Inova 500 MHz NMR (Indy) Primary Contacts: NMR Facility Manager: David VanderVelde, x3004, davidv@caltech.edu GLA: Pamela Tadross, x6131, 646-220-5789 (cell), 204 Church,

More information

15. HOMONUCLEAR COSY

15. HOMONUCLEAR COSY Homonuclear COSY Page 67 15. HOMONUCLEAR COSY by cg fry: created 12/27/94 updated 11/26/00 In COrrelational SpectroscopY (COSY), a fast 2D spectrum is usually obtained first. This initial set of data takes

More information

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods? Retention Time Locking with the MSD Productivity ChemStation Technical Overview Introduction A retention time is the fundamental qualitative measurement of chromatography. Most peak identification is performed

More information

QUANTAX EDS SYSTEM SOP

QUANTAX EDS SYSTEM SOP QUANTAX EDS SYSTEM SOP December 2017 Energy-Dispersive X-Ray Spectroscopy (EDS, EDX, EDXS or XEDS), is an analytical technique used for the elemental analysis or chemical characterization of a sample.

More information

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12 DOSY Diffusion Ordered SpectroscopY. NMR diffusion experiments, like DOSY, are used to determine the diffusion coefficients of solute

More information

Reference Manual MarkerView Software Reference Manual Revision: February, 2010

Reference Manual MarkerView Software Reference Manual Revision: February, 2010 Reference Manual MarkerView 1.2.1 Software Reference Manual - 1 - Revision: February, 2010 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB

More information

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167 Chapter TOCSY 15 Introduction 15.1 Goto TOtal Correlation SpectroscopY provides a different mechanism of coherence transfer than COSY for 2D correlation spectroscopy in liquids. In TOCSY, cross peaks are

More information

Instruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4

Instruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4 Instruction manual for T3DS calculator software Release 2.4 T3DS calculator v2.4 16/02/2018 www.batop.de1 Table of contents 0. Preliminary remarks...3 1. Analyzing material properties...4 1.1 Loading data...4

More information

Agilent G6854 MassHunter Personal Pesticide Database

Agilent G6854 MassHunter Personal Pesticide Database Agilent G6854 MassHunter Personal Pesticide Database Quick Start Guide What is MassHunter Personal Pesticide Database? 2 Installation 3 Main Window 4 Getting Started 11 Database operations 12 Searching

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

To Create Multiple Columns in a Word Document

To Create Multiple Columns in a Word Document To Create Multiple Columns in a Word Document To convert your standard single- (or no-) column Word document into one with multiple columns, simply choose the column format you want from the Columns menu.

More information

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751) Dissimilar This manual is intended to be only a very brief introduction for using the Bruker 400 MHz NMR spectrometer (PS751) at the California State University LA NMR Facility. For complete information

More information

Tutorial 7: Automated Peak Picking in Skyline

Tutorial 7: Automated Peak Picking in Skyline Tutorial 7: Automated Peak Picking in Skyline Skyline now supports the ability to create custom advanced peak picking and scoring models for both selected reaction monitoring (SRM) and data-independent

More information

VNMRJ 2.1B Quick overview and BioPack features

VNMRJ 2.1B Quick overview and BioPack features VNMRJ 2.1B Quick overview and BioPack features VNMRJ 2.1B interface is a much more intuitive user interface than the VNMR6.1C. The Java interface is a layer on top of the conventional VNMR so all commands

More information

Evex NanoAnalysis EDS Acquisition

Evex NanoAnalysis EDS Acquisition NanoAnalysis EDS Acquisition 1. Proper Kv for the sample 2. Correct working distance for EDS 3. Good Image correct magnification & in focus 4. Make note if sample is tilted 5. Adjust beam current for proper

More information

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data Last updated: July 15, 2011 by Josh Kurutz This manual describes how to process NMR data on the offline processing computer in the IMSERC lab

More information

Agilent EZChrom Elite. PDA Analysis

Agilent EZChrom Elite. PDA Analysis Agilent EZChrom Elite PDA Analysis Notices Copyright Scientific Software, Inc 1997-2003 Agilent Technologies, Inc. 2006. No part of this manual may be reproduced in any form or by any means (including

More information

Part 1: Calculating amplitude spectra and seismic wavelets

Part 1: Calculating amplitude spectra and seismic wavelets Geology 554 Environmental and Exploration Geophysics II Generating synthetic seismograms The simple in-class exercise undertaken last time illustrates the relationship between wavelet properties, interval

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

NMR Experiment Procedure

NMR Experiment Procedure NMR Course 2009 February 19, 2009 2009 NMR Research Center NMR Experiment Procedure NMR Research Center 2/25/2009 61 Experimental Guide for Varian MR Spectrometers February, 2009 Introduction: NMR Experiment

More information

IconNMR Operating Procedures

IconNMR Operating Procedures IconNMR Operating Procedures Revised: 7-1-2015 This guide provides basic instructions for acquiring NMR spectra using the IconNMR automation program. 1. Insert your sample(s) in the SampleXpress sample

More information

JASCO CANVAS PROGRAM OPERATION MANUAL

JASCO CANVAS PROGRAM OPERATION MANUAL JASCO CANVAS PROGRAM OPERATION MANUAL P/N: 0302-1840A April 1999 Contents 1. What is JASCO Canvas?...1 1.1 Features...1 1.2 About this Manual...1 2. Installation...1 3. Operating Procedure - Tutorial...2

More information

Copyright 2008 Casa Software Ltd.

Copyright 2008 Casa Software Ltd. Quantification of Thermo Scientific K- Alpha Data The last ten years have seen the XPS technique develop from a technique used predominantly by experts of XPS to a tool used by application oriented scientists.

More information

Vanderbilt Small Molecule NMR Center

Vanderbilt Small Molecule NMR Center Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample

More information

Bruker Topspin NMR Training Manual

Bruker Topspin NMR Training Manual Bruker Topspin NMR Training Manual Rev 180521 Perry Pellechia University of South Carolina NMR Facility Subject Page Number Overview of Operation... 1 Gauging Samples in Spinners... 2 Starting and logging

More information

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer Dr. Zhenming Du Email: zdu@gsu.edu Phone: 3-5538 Revised 04/02/2015 LC=Left click; DC=double

More information

Bruker Avance 400 MHz Instructions

Bruker Avance 400 MHz Instructions Bruker Avance 400 MHz Instructions General Policy: 1) All users must pass the training course before they will obtain their own access to the spectrometer. 2) A key to Bruker Avance 400 Room is accessible

More information

Alpy 600 Mixed Mode Calibration

Alpy 600 Mixed Mode Calibration Alpy 600 Mixed Mode Calibration Paul Luckas October, 2016 1 Introduction Alpy 600 Calibration using the mixed mode involves the following basic steps. 1. Use a sequence of spectra from a type A or B star

More information

Open Access Manual for Superusers

Open Access Manual for Superusers NMR-Service LOC Tel: +41 1 632 29 33 Email: nmrservice@org.chem.ethz.ch Open Access Manual for Superusers 1. Sample preparation Make a homogeneous solution of ca. 20mg in exactly 0.7ml of deuterated solvent,

More information

UniPoint System Software User s Guide

UniPoint System Software User s Guide UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample

More information

Troubleshooting Acquisition Related Problems

Troubleshooting Acquisition Related Problems Troubleshooting Acquisition Related Problems This handout contains pertinent information to help you when a problem arises. Please look through the topics below to identify and find resolutions to the

More information

Thermo Xcalibur Getting Started (Quantitative Analysis)

Thermo Xcalibur Getting Started (Quantitative Analysis) Thermo Xcalibur Getting Started (Quantitative Analysis) XCALI-97207 Revision B September 2010 2010 Thermo Fisher Scientific Inc. All rights reserved. Xcalibur, Surveyor, and Accela are registered trademarks

More information

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES. 08/8/2005 rajan paranji

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES. 08/8/2005 rajan paranji NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES 08/8/2005 rajan paranji CONTENTS Views of XWINNMR INTERFACE 4 Operating the Bruker Spectrometer FLOW CHART 5 LOGON

More information

NMR Experiment Procedure

NMR Experiment Procedure 2013 NMR Experiment Procedure NMR Center Emory University 6/22/2013 Experiment Guide for Varian NMR Spectrometers June, 2013 Introduction: NMR Experiment Procedures...1 Exp. 1: 1D proton and Basic Operations...2

More information

MATERIALS PLUS Segmentation Measurement

MATERIALS PLUS Segmentation Measurement Example: Segmentation MATERIALS PLUS Segmentation is a method of image partitioning based on the intensity / gray scale range of its components. Since a phase is detected and its area is estimated on the

More information

Getting Started. Environmental Analysis Software

Getting Started. Environmental Analysis Software Getting Started Environmental Analysis Software Environmental Software Overview The Environmental software is a data acquisition and data analysis package designed to assist you in complying with United

More information

SPECTRAL IMAGING VIEWER SOFTWARE MANUAL

SPECTRAL IMAGING VIEWER SOFTWARE MANUAL 103 Quality Circle, Suite 215 Huntsville, AL 35806 Phone: (256) 704-3332 Fax: (256) 971-2073 E-Mail: info@axometrics.com Website: http://www.axometrics.com SPECTRAL IMAGING VIEWER SOFTWARE MANUAL 2012

More information

Acquiring Data in VnmrJ 4.2A

Acquiring Data in VnmrJ 4.2A Acquiring Data in VnmrJ 4.2A Linux Primer Initial Steps Changing Password Disabling Screen Lock Reservations Reservation Terminals Rules and Proper Etiquette System Identification Working with VnmrJ 4.2A

More information

NMR Course /13/ :31:47 PM Emory University

NMR Course /13/ :31:47 PM Emory University 1 Introduction: NMR Experiment Procedures Step 1. Login Varian UNITY INOVA NMR spectrometer is controlled by a host Sun computer. For security and administration reasons, all research groups at Emory University

More information

TopSpin. Multiplet Analysis Tutorial

TopSpin. Multiplet Analysis Tutorial TopSpin Multiplet Analysis Tutorial Contents Automatic Multiplet Definition Manual Single-Level Multiplet Definition Manual Multi-Level Multiplet Definition Working with Multiplets Report dialog and export

More information

ME scope Application Note 18 Simulated Multiple Shaker Burst Random Modal Test

ME scope Application Note 18 Simulated Multiple Shaker Burst Random Modal Test ME scope Application Note 18 Simulated Multiple Shaker Burst Random Modal Test The steps in this Application Note can be duplicated using any Package that includes the VES-4600 Advanced Signal Processing

More information

Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013

Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013 Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013 Adapted from JP Blitz and DG Klarup, "Signal-to-Noise Ratio, Signal

More information

EQNMR Facility Rules of Use

EQNMR Facility Rules of Use New User Package for East Quad 600 NMR Date: 01/25/08 Welcome to the East Quad NMR facility. This document contains information for new users using the 600MHz NMR at our facility. In this package, you

More information

Vnmrj 3.1 New and improved features

Vnmrj 3.1 New and improved features Vnmrj 3.1 New and improved features Automation& Hardware Printing USB printers are supported. Using CUPS, a generic print dialog and JPG, PDF, & other formats have replaced the previous method of sending

More information

Working with Spreadsheets in Target

Working with Spreadsheets in Target Working with Spreadsheets in Target This How-To Guide walks you through displaying, editing and processing Geosoft database data in the Spreadsheet view. Spreadsheets When you create or open a database,

More information

Eft Spectrometer Guides

Eft Spectrometer Guides Eft Spectrometer Guides Anasazi Instruments, Inc. 4101 Cashard Ave. Suite 103 Indianapolis, IN 46203 Tel: 317-783-4126 866-494-9369 (Toll Free) Fax: 317-783-7083 Email: aiinmr@aol.com PN 000103-000 Rev.

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Varian Solution NMR Procedure

Varian Solution NMR Procedure System Tool Bar Command Line Vertical Panels Protocols Menu System User Tool Bar NMR Data Display Graphics Tool Bar NMR Graphics Area Study Queue Horizontal Panels Hardware Bar Varian Solution NMR Procedure

More information

Position Table Creation

Position Table Creation Position Table Creation The Position Table creation process is illustrated by following a practical example to carryout sequential ESCA analyses on each of three different sample areas. The samples are

More information

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic

More information

Microsoft Word 2007 on Windows

Microsoft Word 2007 on Windows 1 Microsoft Word 2007 on Windows Word is a very popular text formatting and editing program. It is the standard for writing papers and other documents. This tutorial and quick start guide will help you

More information

Multivariate Calibration Quick Guide

Multivariate Calibration Quick Guide Last Updated: 06.06.2007 Table Of Contents 1. HOW TO CREATE CALIBRATION MODELS...1 1.1. Introduction into Multivariate Calibration Modelling... 1 1.1.1. Preparing Data... 1 1.2. Step 1: Calibration Wizard

More information

Smaart v8.3 Release Overview

Smaart v8.3 Release Overview Smaart v8.3 Release Overview Copyright 2018 Rational Acoustics Table of Contents 8.3 Release Notes... 3 Roland OCTA-CAPTURE Control Integration... 4 Gain Tracking...4 A note about Gain Tracking accuracy...

More information

SIMPLE TEXT LAYOUT FOR COREL DRAW. When you start Corel Draw, you will see the following welcome screen.

SIMPLE TEXT LAYOUT FOR COREL DRAW. When you start Corel Draw, you will see the following welcome screen. SIMPLE TEXT LAYOUT FOR COREL DRAW When you start Corel Draw, you will see the following welcome screen. A. Start a new job by left clicking New Graphic. B. Place your mouse cursor over the page width box.

More information

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1) Menu System System Tool Bar Command Line Graphics Tool Bar Vertical Panels Protocols NMR Data Display NMR Graphics Area Study Queue Horizontal Panels Activity Monitoring Varian Solution NMR Automation

More information

SELECTION OF A MULTIVARIATE CALIBRATION METHOD

SELECTION OF A MULTIVARIATE CALIBRATION METHOD SELECTION OF A MULTIVARIATE CALIBRATION METHOD 0. Aim of this document Different types of multivariate calibration methods are available. The aim of this document is to help the user select the proper

More information

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows Model FP-6500 Spectrofluorometer Instruction Manual FP-6500 for Windows P/N: 0302-9999 April 2000 Contents Safety Considerations...i Regulatory Statements... iii Preface... iv Installation Conditions...v

More information

DOING MORE WITH WORD: MICROSOFT OFFICE 2013

DOING MORE WITH WORD: MICROSOFT OFFICE 2013 DOING MORE WITH WORD: MICROSOFT OFFICE 2013 GETTING STARTED PAGE 02 Prerequisites What You Will Learn USING MICROSOFT WORD PAGE 03 Viewing Toolbars Adding and Removing Buttons MORE TASKS IN MICROSOFT WORD

More information