Electronic Loads. Installation & Operation Manual EL101/300 EL1000

Size: px
Start display at page:

Download "Electronic Loads. Installation & Operation Manual EL101/300 EL1000"

Transcription

1 Electronic Loads Installation & Operation Manual EL101/300 EL /2015

2 Electronic Loads -1- CAUTION Prevent the inputs of the device from electrostatic discharge This may damage the device. Do not connect active objects such as batteries or fuel cells to the power outputs of the device when the device is off This may damage the device.

3 Electronic Loads -2- Unpacking Zahner products are produced carefully, calibrated and tested to achieve a high quality standard. Also the assembling of the accessories and packing is done with great care. Therefore please check the shipment directly after receipt whether the device and all accessories are undamaged. The shipment must contain the following parts: EL1000 EL1000 cable for connection of the EPC40/41/42 (D-Sub9 - Lemosa) cable for connection of a PC USB port (USB-A D-Sub9) with build-in USB-serial converter twisted sense cable (Lemosa plug, blue & green cables) power cord this manual EL300 EL300 with installed control cable for connection of the EPC40/41/42 (Lemosa plug) 2 thick cables (blue and red) with high current banana plugs ( 12mm) twisted sense cable (Lemosa plug, blue & black cables) power cord this manual EL101 EL101 with installed cable for connection of the EPC40/41/42 (Lemosa plug) 2 cables (blue and red) with banana plugs ( 4mm) twisted sense cable (Lemosa plug, blue & black cables) power cord this manual

4 Electronic Loads -3- Basics Today, dynamic measurements on electrochemical objects are of great interest. Modern instruments for impedance measurements, cyclic voltammograms and pulse response experiments provide a broad frequency range from µhz to MHz. At the same time, they provide a huge impedance range from µω to GΩ. Anyway: for most instruments there is one restriction left, they have a limited current range in the magnitude of some Amperes maximum. In the field of electrochemical power generation for example this is only sufficient for 'small' systems. The electrochemical workstations of the IM6 family provide a current range of ±2,5A (Zennium) and ±3A (IM6) and measure impedances down to some µω. Therefore, we provide external potentiostats (EL, PP-Series and XPOT) which can extend the application field of the IM system. The electronic loads EL101/300 and EL1000 are designed as additional potentiostats to allow dynamic investigations on technical systems up to 25 A (EL101), 100 A (EL300) and 200A (EL1000). Their main applications are discharging tests on batteries, accumulators and fuel cells. The electronic loads are easily integrated in the IM6/Zennium system using EPC40/41/42 controller cards. All functions are controlled directly from the Thales software. Up to 16 electronic loads may be controlled by one IM6/Zennium system using up to 4 EPC40/41/42 cards. With the MultiCell functionality of the IM systems the connected electronic loads can be used in parallel. Up to 16 cells may be investigated at the same time. EPC4 / EPC40 / EPC41 / EPC42 The EPC40/41/42 is able to control up to 4 external potentiostats like the EL101/EL300/EL1000, the XPOT and the PP201/211/241. Up to four EPC40/41/42 for a total of 16 external devices can be installed in an IM6 and the Zennium systems. Each port provides analogue and digital interfaces for the communication of the external device with the IM system. The analogue part of the port feeds the device with the DC potential and the AC amplitude at a resolution of 16 bit. Measured current and potential are sent from the external device to the IM to be treated there in the same way as signals from the internal IM cards. The EPC40/41/42 has a bandwidth of 250 khz. A bi-directional serial communication line allows to digitally control the external potentiostat functions and measuring ranges. Besides, the EL1000 electronic load can be controlled by a Windows -PC. In this case they provide the methods Test Sampling as well as U- and I-curves vs. time. They also work as a LabView Virtual Instrument under the LabView software. Furthermore, you can control the EL1000 in a mixed mode with an IM6/Zennium and a PC in parallel. Both devices can be connected and disconnected in operation. Plug or unplug external devices only when both, IM6/Zennium AND the external device, are switched off. Otherwise the devices may be damaged.

5 Electronic Loads -4- Selecting an external device All external IM devices are directly controlled by the Thales software. Each device has a unique device number which is identical with the EPC40/41/42 port number. So, if a device is connected to EPC port 3, you address it with the device number 3. Device number '0' is reserved for the internal potentiostats of the IM system. If an RMUX (relay multiplexer) card is installed, the device numbers of the ELs start with 17, not with 1, because the device numbers 1 to 16 are reserved for the 16 RMUX channels then. To select a device, call the Test Sampling page and click one the # symbol next to the Device Identity box. Device Main # An input box will appear where you can input the device number. Potentiostat Number: 1 Now the selected device type is shown in the Device Identity box, for example: Device EL300#1 # If no device has been connected to the addressed EPC40/41/42 port an error message is displayed and the software automatically selects the internal potentiostat. If the selected device is present but has not been activated yet the software starts to calibrate it automatically. If an external device is changed the new device has to be calibrated before use. The calibration is carried out only for the selected device. All other calibration data keep unchanged. If a device number other than 0 is selected, the parameters of the Test Sampling page now are valid for this device. The following methods are available for external devices: EIS C/E I/E MIE AS impedance measurement parameter impedance measurement current potential curve recording Multiple parallel current potential curve recording series measurements Specific parameters for these methods are to be edited in the pages related to the method. Changing devices When changing the device number, the now inactive device will hold its DC conditions such as DC potential and on/off status as long as it is selected anew or the system is shut off. On the other hand only the selected device is internally connected to the FRA. Therefore only this device is able to output an AC signal superimposed to the DC potential.

6 Electronic Loads -5- EL101 / EL300 The EL101/300 external electronic loads are One-Quadrant-Potentiostats. This means that they are able to sink current but cannot source current. Typical applications are discharging experiments at batteries, accumulators and fuel cells. The EL-series potentiostats can be operated in both potentiostatic and galvanostatic modes, controlled by the Thales software. The output as well as the input are electrically isolated from ground. Whereas the EL101 is air cooled, the EL300 needs water cooling when loaded with more than 25 A. For water cooling you find an inlet and an outlet at the backside of the EL300. The EL300 may get damaged if more than 25 A are applied without water cooling Cell Connections It is important to know that EL potentiostats SINK current from the test object and therefore the cell connections must be as short and as thick as possible. Otherwise the measurements may be faulty and it may even seem that the EL is defective. For this reason, the standard cable set shipped with the ELs should be shortened as far as possible. It is also important to connect the test object with the correct polarity to the EL potentiostat. Whereas typical one-cell-voltages of 0.8V to 1.2V do not damage an EL when being connected with the wrong polarity, cell stacks can do very well. Therefore, we recommend to connect the sense inputs as described in fig. 1 (Full Cell Configuration). Call the Test Sampling page of the Thales software and check for the potential polarity. It must be negative. Now connect the power inputs according to fig.1 with the potentiostat switched off. The displayed potential must not change significantly when connecting the power lines. With the correct polarity the test object may be connected to the EL potentiostat in one of the following ways: 1. Full Cell Configuration (Standard Kelvin Scheme) EL101 This configuration is used with active test objects like batteries, accumulators and fuel cells if a complete cell is to be investigated. sense ZAHNER elektrik power active line The DC potential applied must be negative and greater or equal than the rest potential of the cell. Rest potential set potential 0 V anode pole cathode pole

7 Electronic Loads -6-2.a. Half Cell Configuration Anode EL101 sense This configuration is used with active test objects like batteries, accumulators and fuel cells if only the anodic part of the cell has to be investigated. power active ZAHNER elektrik line The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential anode pole cathode pole Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode. 2.b. Half Cell Configuration - Cathode EL101 This configuration is used with active test objects like batteries, accumulators and fuel cells if only the cathodic part of the cell has to be investigated. sense power active ZAHNER elektrik line The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential anode pole cathode pole Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode.

8 sense active line Electronic Loads Partial Cell Configuration EL101 This configuration may be used, if a certain part of a battery or fuel cell stack has to be investigated. sense power active ZAHNER elektrik line The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential cathode anode Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode. 4. Applications with an additional power supply EL300 power current limit 60 Ampere Volt potential coarse fine 80 This configuration may be used if the potential drop on the power lines is too high to reach the high-current test conditions. In addition it allows experiments on passive objects and batteries under charging conditions and electrolysis cell ZAHNER elektrik output Z The DC potential applied must be higher than the rest potential and must NOT EXCEED U max = +4 V. Rest potential set potential +4 V Built-in buffer amplifier The built-in buffer amplifier may be used to increase the potential range of the EL up to +/-12V. To select the buffer amplifier select any connection scheme 'with buffer' and set the corresponding gain factor. Using the gain factor GAIN=1 the EL will control -4V...+4V. Setting GAIN=0.33 the EL will control a dynamic potential range between -12V and +12V.

9 Electronic Loads -8- EL1000 The EL1000 external electronic load is an One-Quadrant-Potentiostat. This means that it is able to sink current but cannot source current. Typical applications are discharging experiments at batteries, accumulators and fuel cells. The EL-series potentiostats can be operated in both potentiostatic and galvanostatic modes, controlled by the Thales software. The output as well as the input are electrically isolated from ground. The EL1000 is air cooled. Measuring floating objects On the rear of the EL1000 you will find two connectors with a jumper. If the jumper is set, the signal ground is connected to ground via a 100 Ω protective resistor. The EL1000 power stage is NOT connected to signal ground. This is necessary, if the test object is floating (no metallic or electrolytic contact to ground). If any part of the cell is grounded, the jumper must be removed. silver banana jack (protective ground) black banana jack (system ground) When investigating floating objects the jumper must be set. Stand-alone mode Besides the usage of the EL1000 as an external one-quadrant potentiostat for an IM6/Zennium system, it also may be used in a stand-alone mode. The setup is done through an IM system or a PC (Currently not implemented, but will be added with a software update as soon as possible). Normally, the EL1000 should only be connected or disconnected to the IM system if both, the IM6/Zennium and the EL1000 are switched off. If you want to use the EL1000 as a stand-alone device, unselect it in the Test Sampling page of the Thales software (you may e.g. change the device number to the main potentiostat). The inactive device will hold its DC conditions such as DC potential and on/off status as long as it is selected anew or the system is shut off. Now you may unplug the EL1000 at the EPC40/41/42 connector. For regaining access to the EL1000, connect it to the EPC40/41/42 and select it in the Test Sampling page of the Thales software. Cell Connections It is important to know that EL potentiostats SINK current from the test object and therefore the cell connections must be as short and as thick as possible. Otherwise the measurements may be faulty and it may even seem that the EL is defective. It is also important to connect the test object with the correct polarity to the EL potentiostat. When being connected with the wrong polarity, the EL1000 has a protection circuit and will indicate polarity error. Therefore, we recommend to connect the sense inputs as described in fig. 1 (Full Cell Configuration). Call the Test Sampling page of the Thales software and check for the potential polarity. It must be negative. Now connect the power inputs according to fig.1 with the potentiostat switched off. The displayed potential must not change significantly when connecting the power lines. With the correct polarity the test object may be connected to the EL potentiostat in one of the following ways:

10 Electronic Loads Full Cell Configuration (Standard Kelvin Scheme) This configuration is used with active test objects like batteries, accumulators and fuel cells if a complete cell has to be investigated. The DC potential applied must be negative and greater or equal than the rest potential of the cell. Rest potential set potential 0 V 2.a. Half Cell Configuration Anode This configuration is used with active test objects like batteries, accumulators and fuel cells if only the anodic part of the cell has to be investigated. The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode. 2.b. Half Cell Configuration - Cathode This configuration is used with active test objects like batteries, accumulators and fuel cells if only the cathodic part of the cell is to be investigated. The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode.

11 Electronic Loads Partial Cell Configuration This configuration may be used, if a certain part of a battery or fuel cell stack has to be investigated. The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential Depending on the type of the reference electrode the measured potential may be different from the real potential at the reference electrode site. The real potential can be calculated from the measured potential by subtracting the potential of the reference electrode. 4. Applications with an additional DC sink This configuration may be used to sink more than 200A. The total amount of current from the EL1000 and the additional electronic load must not exceed 680A The DC potential applied must be greater or equal than the rest potential of the cell. Rest potential set potential The measured current must be positive I 0 The total current must NOT EXCEED 680A Example: Ensure to switch on the external load before switching on EL1000 Ensure to switch off EL1000 before switching off external load External DC load with 200A DC EL1000 with 10A DC and 5A amplitude Setup Testsampling EL1000: Galvanostat 210A DC, 5A amplitude

12 Electronic Loads Applications with an additional power supply (serial) This configuration may be used if the potential drop on the power lines is too high to reach the high-current test conditions. In addition it allows experiments on passive objects and batteries under charging conditions and electrolysis cell. The DC potential applied must be higher than the rest potential and must NOT EXCEED U max = +4 V. Rest potential set potential +4 V 5.a. Compensation of voltage drops "Zero Volt Option" 5.b. Electrolysis of fuel cells 5.c. Charging batteries

13 Electronic Loads Applications with an additional power supply (external input) This configuration allows experiments under following conditions: charging, SoC and discharging battieries fuel cell and electrolysis cell operation The 3rd party power supply operates as a galvanostat between positiv input and external input. When the EL1000 is turned off the device shows the external current flowing through the connected object (charging / electrolysis). The EL1000 is now able to sink "negative" currents. In this case only a part from the external current is drawn from the EL1000 and the rest flows through the object (charging / electrolysis). When turn on the EL1000 with zero current the external current from the 3rd party supply is flowing completely through the EL1000 and thus the object has zero current (SoC). Examples: When sinking more current with the EL1000 the additional current portion flows from the object (discharging / fuel cell). The external power supply operates as a galvanostat (setting current limit) and provides a higher potential as the object. EL1000 setting DC current: I < 0 charging battery / electrolysis cell I = 0 battery at state of charge I > 0 discharging battery / fuel cell operation

14 Electronic Loads -13- Built-in buffer amplifier The built-in buffer amplifier may be used to increase the potential range of the EL up to +/-100V. Select the potential range at the cell connection scheme page. The corresponding gain factor will be set automatically. The EL1000 will control -4V...+4V and -100V V. Never plug or unplug a EL1000 at the D-Sub9 connector at the backside of the device with the IM switched on. This may damage the system. It is recommended to fix the D-Sub9 connector with the screws to prevent accidental unplugging. If the EL1000 is unplugged while being selected in the Test Sampling page of the Thales software, it will shut off the current. This is a precaution to prevent undefined situations. The EL1000 need a warm-up time of about 15 minutes after power on. A calibration procedure is initiated automatically with the first access by the IM system. If you access a EL1000 during the warm-up time, please do a forced calibration (EIS menu -> Calibrate) after about 15 minutes. As the EL1000 is optimised for high current (some 100 ma to 200 A) an erroneous current display of some ma is acceptable.

15 Electronic Loads -14- Specifications EL101 EL300 EL1000 Operating modes pot/gal pot/gal pot/gal Potential range ±4V / ±12V ±4V / ±12V ±4V / ±100V Pot. accuracy 0.25% / ±2mV ±0.25% / ±2mV 0.1% / ±5mV Current range 0A 25A 0A 100A 0A 200A Current accuracy 0.25% / ±3mA 0.25% / ±3mA 0.25% / ±10mA Power dissipation T a T a 300W water-cooled T a Frequency range * 10µHz - 10kHz 10µHz - 10µHz - 10µHz - 10µHz - Impedance range 1µΩ - 1kΩ ** 1µΩ - 1kΩ ** 1µΩ - 1kΩ ** Amb. temperature 0 o C 25 o C 0 o C 25 o C 0 o C 25 o C System requirements IM6/Zennium + EPC42 IM6/Zennium + EPC42 IM6/Zennium + EPC42 or PC * Please note that the maximum frequency range only can be achieved with an IM6 or an Zennium equipped with EPC41 or EPC42. With an IM5d the maximum frequency range of external devices is 30kHz. ** Impedances below 1mΩ must be measured galvanostatically.

Isolated Voltage Input 3B30 / 3B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Voltage Input 3B30 / 3B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Voltage Input 3B30 / 3B31 FEATURES Interfaces, amplifies, & filtersanalog input voltages. Narrow-bandwidth (3Hz) single-channel single conditioning. Module provides simultaneous precision voltage

More information

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES Interfaces, amplifies, protects& filters wide-bandwidth (h0 khz) single-channel analog voltage inputs. Module provides simultaneous precision voltage

More information

BOP QUICK START GUIDE 100W, 200W, 400W. KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY WARNINGS I INTRODUCTION II SAFETY.

BOP QUICK START GUIDE 100W, 200W, 400W. KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY WARNINGS I INTRODUCTION II SAFETY. QUICK START GUIDE I INTRODUCTION KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the Kepco 100W,

More information

Replacement of analog modules of the IM6/IM6e

Replacement of analog modules of the IM6/IM6e Replacement of analog modules of the /e Initial Note: Please read all steps of this instruction and get familiar with the (e) by following the replacement instructions step by step. Be cautious and do

More information

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM FEATURES Interfaces, amplifies, filters, isolates, & linearizes analog input voltages from a J, K, T, E, R, S or B-type thermocouple Thermocouple input signal is internally linearized High accuracy internal

More information

Isolated Process Current Input 7B32 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Process Current Input 7B32 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Process Current Input 7B32 FEATURES Interfaces, amplifies and filters a process-current input. Module provides a precision output of either +1 V to +5 V or 0 V to +10 V, linear with temperature.

More information

Isolated, Voltage or Current Input 7B30 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Voltage or Current Input 7B30 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated, Voltage or Current Input 7B30 FEATURES Interfaces, amplifies and filters unipolar and bipolar millivolt and voltage inputs. Provides a protected precision output of either +1 V to +5 V or 0 V

More information

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User's Guide. MiniTec TM Series Model MN25 MultiMeter User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM VERSION 2.11 npi 2002 npi electronic GmbH, Hauptstrasse 96, D-71732 Tamm, Germany Phone +49 (0)7141-601534;

More information

BIPOLAR OPERATIONAL POWER SUPPLY

BIPOLAR OPERATIONAL POWER SUPPLY QUICK START GUIDE I INTRODUCTION KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the Kepco 100W,

More information

Mini Digital Multimeter

Mini Digital Multimeter User Manual Mini Digital Multimeter Model MN15A Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech MN15A MultiMeter. The MN15A offers

More information

Autolab/ Electronic Load

Autolab/ Electronic Load Autolab/ Electronic Load Installation Description Picture Courtesy of Nedstack Fuel Cell Technology BV Autolab/Electronic load combination The Autolab range of instruments is limited in terms of maximum

More information

BOP QUICK START GUIDE 100W, 200W, 400W. KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY I INTRODUCTION

BOP QUICK START GUIDE 100W, 200W, 400W. KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY I INTRODUCTION QUICK START GUIDE I INTRODUCTION KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the Kepco 100W,

More information

USER'S MANUAL DCL-650

USER'S MANUAL DCL-650 USER'S MANUAL Clamp on Multimeter - Compact Design True RMS AC/DC DCL-650 CIRCUIT-TEST ELECTRONICS www.circuittest.com Safety International Safety Symbols This symbol, adjacent to another symbol or terminal,

More information

Autoranging True RMS Multimeter User Manual

Autoranging True RMS Multimeter User Manual Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.

More information

Isolated Voltage Input 7B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Voltage Input 7B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Voltage Input 7B31 FEATURES Interfaces, amplifies, and filters unipolar and bipolar voltage inputs. Module provides a precision output of either +1 V to +5 V or 0 V to +10 V, linear with temperature.

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex

More information

FCI4 Fuel Cell Interface. Operator's Manual

FCI4 Fuel Cell Interface. Operator's Manual FCI4 Fuel Cell Interface Operator's Manual Copyright 1997-2005 Gamry Instruments, Inc. All rights reserved. Printed in the USA. Revision 5.01 November 30, 2005 i ii Limited Warranty Gamry Instruments,

More information

Isolated Process Current Input with Loop Power 7B35

Isolated Process Current Input with Loop Power 7B35 Isolated Process Current Input with Loop Power 7B35 FEATURES Single-channel signal conditioning current input module that interfaces with two-wire transmitters. Module provides a precision output of either

More information

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

More information

OPERATING INSTRUCTION. Pen-Type Digital Multimeter

OPERATING INSTRUCTION. Pen-Type Digital Multimeter OPERATING INSTRUCTION Pen-Type Digital Multimeter International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

More information

SIM-9106/9303 High Current Switching Mode Power Supply with C.C., C.V., Remote Control & Sensing User Manual

SIM-9106/9303 High Current Switching Mode Power Supply with C.C., C.V., Remote Control & Sensing User Manual SIM-9106/9303 High Current Switching Mode Power Supply with C.C., C.V., Remote Control & Sensing User Manual 1. INTRODUCTION This family of high power remote sensing, remote control, switching mode CC\CV

More information

Digital Clamp Meter User Manual

Digital Clamp Meter User Manual Compact AC, AC/DC Digital Clamp Meter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents 1.Safety... 3 2.Safety Notes... 3 3.Warnings... 3 4.Cautions...

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

Model A Mini AC/DC Clamp Meter. User's Guide

Model A Mini AC/DC Clamp Meter. User's Guide Model 380950 80A Mini AC/DC Clamp Meter User's Guide Introduction Congratulations on your purchase of the Extech 80A Mini AC/DC Clamp Meter. The Model 380950 measures AC/DC Current, AC/DC Voltage, Resistance,

More information

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Wide Bandwidth Strain Gage Input 3B18 FEATURES Wideband (20 khz) single-channel signal conditioning module. Module Bandwidth is user-selectable between 20 khz and 100Hz, with user-supplied filter caps

More information

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CATⅡ)

More information

Model 500 Series Laser Diode Driver User's Manual

Model 500 Series Laser Diode Driver User's Manual Model 500 Series Laser Diode Driver User's Manual .0 INTRODUCTION The Model 500 Series Laser Diode Drivers are low noise, highly stable current sources for use with laser diodes. Selectable output current

More information

OPERATION MANUAL MODEL TT-LP

OPERATION MANUAL MODEL TT-LP TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors

More information

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER OWNER S MANUAL 9908-TE HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER IMPORTANT! Read and understand this manual before using the instrument. Failure to understand and comply with

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at User Manual 400Amp AC Clamp Meter + NCV Model MA430 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of this Extech MA430 Clamp Meter. This

More information

ECHO AIO TEST INTERFACE

ECHO AIO TEST INTERFACE ECHO AIO TEST INTERFACE Product Manual June 2017 2017 Echo Digital Audio Contents INTRODUCTION... 2 AIO CONFIGURATIONS... 3 AIO DIMENSIONS AND WEIGHT (ALL MODELS):... 5 SAFETY INSTRUCTIONS... 6 INSTALLATION...

More information

Mini Digital Multimeter

Mini Digital Multimeter User's Guide Mini Digital Multimeter Model MN15 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech MN15/MN16 Series

More information

User Guide True RMS Multimeter Extech EX205T

User Guide True RMS Multimeter Extech EX205T User Guide Extech EX205T True RMS Digital Multimeter Extech EX210T True RMS Digital Multimeter IR True RMS Multimeter Extech EX205T Introduction Thank you for selecting the Extech EX205T True RMS Auto-ranging

More information

Automation Interface Requirements for System I/O Interface of a DPC I Welding System

Automation Interface Requirements for System I/O Interface of a DPC I Welding System - 1 - Automation Interface Requirements for System I/O Interface of a DPC I Welding System The DPC I welding system offer several features that are intended to communicate with automation. These features

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM VERSION 3.8 npi 2016 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230; Fax: +49

More information

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies ADC7520 SERIES 1600W Battery Chargers and Power Supplies Wide output adjustment range 0 72VDC Analog control by external 0-5VDC voltage Temp.comp charging, sense as on option Power fail relay alarm Master-Slave

More information

Hybrid AC Driver [GCNC-1110]

Hybrid AC Driver [GCNC-1110] Page 1 Installation Manual and Datasheet Page 2 Key Features Smooth and quiet operation at all speeds and extremely low motor heating Industrial grade performance for an alternating current servo motor

More information

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL Issue 2011 ASL Intercom BV DESIGNED & MANUFACTURED BY: ASL Intercom B.V. Zonnebaan 42 3542 EG Utrecht The Netherlands Tel: +31 (0)30 2411901 Fax: +31 (0)30

More information

OPERATING INSTRUCTIONS PA AMPLIFIER P-1812

OPERATING INSTRUCTIONS PA AMPLIFIER P-1812 OPERATING INSTRUCTIONS PA AMPLIFIER P-1812 Please follow the instructions in this manual to obtain the optimum results from this unit. We also recommend that you keep this manual handy for future reference.

More information

RangerBOSS Network Ready Constant Monitor Model CM2800

RangerBOSS Network Ready Constant Monitor Model CM2800 RangerBOSS Network Ready Constant Monitor Model CM2800 Instruction Manual Contents 1 Description CM2800 1 Features 1 2 Installation Installation Instructions 2 Installation Diagram 3 3 Operation Wrist

More information

Mini Digital Multimeter

Mini Digital Multimeter User's Guide Mini Digital Multimeter Model MN15 Introduction Congratulations on your purchase of the Extech MN15 MultiMeter. The MN15 offers AC/DC Voltage, AC/DC Current, Resistance, Diode, and Continuity

More information

HCS-3600 / 3602 / 3604 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control

HCS-3600 / 3602 / 3604 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control HCS-3600 / 3602 / 3604 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control 1. INTRODUCTION User Manual This family of efficient, upgraded SMPS with small form factor,

More information

CheckSum. FUNC-2B Functional Test Module. The FUNC-2B features:

CheckSum. FUNC-2B Functional Test Module. The FUNC-2B features: CheckSum FUNC-2B Functional Test Module FUNC-2B Functional Test Module Made in U.S.A. The CheckSum Model FUNC-2B Functional Test System is designed to be used as an extension to a CheckSum ICT System.

More information

Isolated, Voltage or Current Input 7B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Voltage or Current Input 7B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated, Voltage or Current Input 7B41 FEATURES Interfaces, amplifies and filters unipolar and bipolar voltage inputs. Module provides a precision output of either +1 V to +5 V or 0 V to +10 V. All 7B41

More information

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

Model INSTRUCTION MANUAL DIGITAL MULTIMETER Model 57040 INSTRUCTION MANUAL DIGITAL MULTIMETER SAFETY INFORMATION This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CAT

More information

MTP INSTRUCTION MANUAL

MTP INSTRUCTION MANUAL DT-118B MTP INSTRUCTION MANUAL Pocket Autoranging Digital Multimeter 3 in 1 Model MTP-1025 Auto Ran ging DMM Hz% A OFF V AU TO PO WER OFF MTP Instruments Table of Contents Introduction Page 1 Features

More information

INSTRUCTION MANUAL MODELS: 1760A, 1761, Triple Output DC POWER SUPPLY With Dual 4-Digit LED Displays

INSTRUCTION MANUAL MODELS: 1760A, 1761, Triple Output DC POWER SUPPLY With Dual 4-Digit LED Displays INSTRUCTION MANUAL MODELS: 1760A, 1761, 1762 Triple Output DC POWER SUPPLY With Dual 4-Digit LED Displays TEST INSTRUMENT SAFETY WARNING Normal use of test equipment exposes you to a certain amount of

More information

Installation & Getting Started

Installation & Getting Started Installation & Getting Started Zahner 03/2012 Installation & Getting Started - 2 - Installation & Getting Started - 3-1. General Safety Information...5 2. Unpacking...6 3. Hardware Installation...7 3.1

More information

Wilcoxon Research P702B/P702BF-1 operating guide Power unit/amplifier for piezoelectric transducers with internal amplifiers

Wilcoxon Research P702B/P702BF-1 operating guide Power unit/amplifier for piezoelectric transducers with internal amplifiers Wilcoxon Research P702B/P702BF-1 operating guide Power unit/amplifier for piezoelectric transducers with internal amplifiers Caution: This manual should be read carefully before installation. Meggitt Sensing

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

Isolated, Process Current Output 7B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Process Current Output 7B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated, Process Current Output 7B39 FEATURES Interfaces, isolates and filters a 0 V to + 10 V or +1 V to +5 V input signal. Provides an isolated process current output of 0 ma to 20 ma or 4 ma to 20

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE HVA-100 HIGH VOLTAGE AMPLIFIER MODULE FOR EPMS SYSTEMS VERSION 1.7 npi 2015 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230;

More information

User's Guide. Digital Multimeter. Model MN42

User's Guide. Digital Multimeter. Model MN42 User's Guide Digital Multimeter Model MN42 Introduction Congratulations on your purchase of the Extech MN42 MultiMeter. The MN42 offers AC/DC Voltage, DC Current, and Resistance testing. Proper use and

More information

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3

More information

Isolated Process Current Output 3B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Process Current Output 3B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated Process Current Output 3B39 FEATURES Interfaces, isolates, & filters a 0 to +10V or -10V to +10V input signal and provides an isolated process current output across loads of 0 to 850Ω.. Intended

More information

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389 User s Guide 600A True RMS AC/DC Clamp Meter Model 38389 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further

More information

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL: 61-795 The Service Information provides the following: Precautions and safety information Specifications Performance test procedure Calibration and calibration

More information

User's Guide. Extech AM A AC Analog Clamp Meter

User's Guide. Extech AM A AC Analog Clamp Meter User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.

More information

Autoranging Mini Multimeter

Autoranging Mini Multimeter User Manual Autoranging Mini Multimeter Model MN16A Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech MN16A Autoranging Multimeter.

More information

NI PXIe-1062Q Power Supply Shuttle

NI PXIe-1062Q Power Supply Shuttle USER GUIDE NI PXIe-1062Q Power Supply Shuttle The NI PXIe-1062Q power supply shuttle is a replacement part for the NI PXIe-1062Q chassis. Caution This power supply is not compatible with any other National

More information

7032 Digital-Analog Multimeter

7032 Digital-Analog Multimeter 7032 Digital-Analog Multimeter OPERATOR S MANUAL CONTENTS: 1. Safety precautions and procedures 1 1.1. Preliminary 1 1.2. During Use 2 1.3. After Use.. 2 2. General Description. 3 3. Preparation for Use..

More information

Electrochemical Quartz Crystal Microbalance Quick-Start Guide

Electrochemical Quartz Crystal Microbalance Quick-Start Guide Electrochemical Quartz Crystal Microbalance Quick-Start Guide Installation Sequence is Important for Success The Gamry eqcm 10M is a plug-and-play device. Therefore, setup is easiest if you use the sequence

More information

Page 2 262/264 LaserMount User s Manual. Table of Contents

Page 2 262/264 LaserMount User s Manual. Table of Contents Page 2 262/264 LaserMount User s Manual Table of Contents Introduction... 3 Safety Terms and Symbols... 3 Installation and Use... 4 Connector Pin-Outs... 7 Technical Specifications... 9 Mechanical Drawings...

More information

NOVA Cyclic voltammetry tutorial

NOVA Cyclic voltammetry tutorial Version 1.11.0 NOVA Cyclic voltammetry tutorial 1 Cyclic voltammetry staircase Cyclic voltammetry is the most popular electrochemical method. It provides both qualitative and quantitative information as

More information

TH2683 Insulation Resistance Meter. User s Mannual

TH2683 Insulation Resistance Meter. User s Mannual TH2683 Insulation Resistance Meter User s Mannual 1 CONTENTS Chapter 1 Genernal Information... 3 1.1 Feature Overview... 3 1.2 Operating Environment... 3 1.3 Dimensions and Weight... 3 1.4 Unpacking Inspection...

More information

DPM Digital DC Power Meter with Data logging capability. User Manual

DPM Digital DC Power Meter with Data logging capability. User Manual DPM-3232 Digital DC Power Meter with Data logging capability User Manual Introduction As a digital DC power meter, it measures the real time DC Voltage (5-60V), DC Current (0-60A), Watt, and it also displays

More information

Switched Mode Power Supply Operating manual PAP800

Switched Mode Power Supply Operating manual PAP800 Features: Switched mode power supply Wide output 0 144Vdc Analog control by an external 0 5Vdc Power failure alarm output Master-slave connection Powerfinn PAP series is a high power, lightweight, advanced

More information

Resolver to Digital Expansion Board

Resolver to Digital Expansion Board Resolver to Digital Expansion Board Catalog No. EXB009A01 Installation and Operating Manual 6/98 MN1313 Table of Contents Section 1 General Information............................. 1-1 Introduction....................................

More information

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español Owner's Manual True RMS Multimeter Model No. 82023 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español

More information

LDD M SERIES INSTRUCTION MANUAL LDD M SERIES

LDD M SERIES INSTRUCTION MANUAL LDD M SERIES TM LDD M SERIES LDD M SERIES INSTRUCTION MANUAL P O Box Bozeman, MT 9 Phone (0) -90 Fax (0) -9 email sales@wavelengthelectronics.com www.wavelengthelectronics.com TABLE OF CONTENTS Features... Customer

More information

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314 Instruction Manual 2in1 LAN Tester & Multimeter Model: 57314 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.

More information

DPM Bi-directional Digital DC Power Meter with built-in USB data logger & adapter External shunt model. User Manual

DPM Bi-directional Digital DC Power Meter with built-in USB data logger & adapter External shunt model. User Manual DPM-3321 Bi-directional Digital DC Power Meter with built-in USB data logger & adapter External shunt model User Manual Bi-Directional DC Power Meter with built-in USB data logger & adapter. The new DPM-3321

More information

MS2115B Dual Display Digital Clamp Meter Operation Manual ZERO MIN RANGE HOLD

MS2115B Dual Display Digital Clamp Meter Operation Manual ZERO MIN RANGE HOLD MS2115B Dual Display Digital Clamp Meter Operation Manual REL MAX B.L ZERO MIN RANGE HOLD B CONTENTS 1. Safety Information...1 1.1 Preparation...1 1.2 Usage...2 1.3 Mark...3 1.4 Maintenance...4 2. Description...4

More information

Cantata m100 Amplifier

Cantata m100 Amplifier Cantata m100 Amplifier Getting Started Guide www.resolutionaudio.com +1.415.553.4100 Safety Information CAUTION RISK OF ELECTRICAL SHOCK DO NOT OPEN CAUTION: TO REDUCE THE RISK OF ELECTRICAL SHOCK, DO

More information

Thermoelectric Cooler Controller TED1000

Thermoelectric Cooler Controller TED1000 Thermoelectric Cooler Controller TED1000 Operating Instructions MANUAL-TED1000-1.0 Aug 2015 Rev.1 2 Contents 1 General... 4 1.1 Warranty and Assistance... 4 1.2 Maintenance... 4 1.3 General Safety Considerations...

More information

Model R5010. TRMS Digital Multimeter. Instruction Manual

Model R5010. TRMS Digital Multimeter. Instruction Manual INSTRUMENTS Model R5010 TRMS Digital Multimeter Instruction Manual Table of Contents Safety...3-4 IEC1010 Overvoltage Installation Category... 3 Warnings... 4 Features... 5 Specifications...5-8 Instrument

More information

Bluetooth/USB Data Logger USER S MANUAL. Hantek 365A/B/C/D/E/F V

Bluetooth/USB Data Logger USER S MANUAL. Hantek 365A/B/C/D/E/F V Bluetooth/USB Data Logger USER S MANUAL Hantek 365A/B/C/D/E/F V 1.0.3 www.hantek.com Content General Safety Summary... 1 Chapter 1 Getting Start... 3 1.1 General Check... 4 1.2 The User interface... 5

More information

Programmable DC power supplies

Programmable DC power supplies 3-Year Warranty Series 2260B 360W, 720W, and 1080W versions with voltages up to 800V and currents up to 108A Programmable voltage or current rise and fall times prevents damage to low impedance loads from

More information

4170 POS System Installation Guide

4170 POS System Installation Guide 4170 POS System 4170 Installation Guide Thank you for selecting UTC RETAIL s innovative Model 4170 Point of Sale solution! This Installation Guide will help you efficiently install the 4170 POS. The document

More information

INSTRUCTION MANUAL. Triple Output DC POWER SUPPLY MODELS 1651A & 1652 MODELOS 1651A & El Manual de la Instrucción

INSTRUCTION MANUAL. Triple Output DC POWER SUPPLY MODELS 1651A & 1652 MODELOS 1651A & El Manual de la Instrucción INSTRUCTION MANUAL El Manual de la Instrucción MODELS 1651A & 1652 MODELOS 1651A & 1652 Triple Output DC POWER SUPPLY FUENTES DE PODER De Triple Salida DC Test Equipment Depot - 800.517.8431-99 Washington

More information

ATE QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION

ATE QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION QUICK START GUIDE KEPCO An ISO 9001 Company. ATE VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the

More information

DPM-3221 Digital DC Power Meter with data logging capability.

DPM-3221 Digital DC Power Meter with data logging capability. Introduction DPM-3221 Digital DC Power Meter with data logging capability. User Manual As a digital DC power meter, it measures the real time DC Voltage (5-60V), DC Current (0-200 A with the correct shunt),

More information

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc. DLA DMX512 Analyzer DLA DLA-HH 1 Table Of Contents IMPORTANT SAFEGUARDS... 2 DLA OVERVIEW... 3 CONNECTION... 3 OPERATION... 3 HARDWARE SETUP... 4 DLA-HH (PORTABLE) LAYOUT... 4 CHASSIS LAYOUT... 4 DLA MENU

More information

5.6" Multi-function Monitor

5.6 Multi-function Monitor 5.6" Multi-function Monitor User s Manual Please read this Manual carefully before use of this product, and keep it handy for future reference. I. Packing List.. 2 II. Product Appearance... 3-5 III. Product

More information

Model 7700 Multiplexer Module

Model 7700 Multiplexer Module Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 tek.com/keithley Model 7700 Multiplexer Module Instructions for use with DAQ6510 Introduction The 7700 20-Channel Differential

More information

IMP-2MC ELECTRODE IMPEDANCE TESTER MULTICHANNEL

IMP-2MC ELECTRODE IMPEDANCE TESTER MULTICHANNEL MicroProbes for Life Science IMP-2MC ELECTRODE IMPEDANCE TESTER MULTICHANNEL User Instructions Version: January, 2016 18247-D Flower Hill Way Gaithersburg, MD 20879 Phone: +1 301 330-9788 Fax: +1 301 330-9667

More information

TETRIS 2500 High Impedance Active Probe. Instruction Manual

TETRIS 2500 High Impedance Active Probe. Instruction Manual TETRIS 2500 High Impedance Active Probe Instruction Manual Copyright 2018 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

High Performance Oscilloscopes DPO7000, DPO/DSA70000 and DPO/DSA70000B Series

High Performance Oscilloscopes DPO7000, DPO/DSA70000 and DPO/DSA70000B Series Best Practices High Performance Oscilloscopes DPO7000, DPO/DSA70000 and DPO/DSA70000B Series Contacting Tektronix Tektronix, Inc. 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA For product

More information

MARTINDALE INSTRUCTIONS MM35 DIGITAL MULTIMETER ELECTRIC. Trusted by professionals

MARTINDALE INSTRUCTIONS MM35 DIGITAL MULTIMETER ELECTRIC. Trusted by professionals Martindale Electric will carry out routine calibration (on a chargeable basis) if the instrument is returned, carriage paid, to the address on the final page of this document. Alternatively, a chargeable

More information

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810 Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4

More information

3700 SERIES USER MANUAL

3700 SERIES USER MANUAL SAFETY GUIDE This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with

More information

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL August 2016 This product is designed and manufactured by: ASL Intercom B.V. Zonnebaan 42 3542 EG Utrecht The Netherlands Phone: +31 (0)30 2411901 Fax: +31

More information

JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776

JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776 JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT JR3, Inc. 22 Harter Ave. Woodland, CA 95776 5963B 13 October, 2003 TABLE OF CONTENTS CHAPTER 1 OVERVIEW General...1-1 Sensor...1-1 Electronic System...1-1

More information

HERCULES-5000 INSTRUCTION MANUAL

HERCULES-5000 INSTRUCTION MANUAL IndustrieAlpine Allee 1 D - 94513 Schönberg Tel.:+49 (0) 85 54/9609-0 Fax:+49 (0) 85 54/96 09 20 Mail: sales@tetelectronics.de INSTRUCTION MANUAL HERCULES-5000 High-performance power supplies Page 1 von

More information

Available for low voltage and high current High slew rate with 100A/μs

Available for low voltage and high current High slew rate with 100A/μs Low voltage input and high slew rate electronic load Available for low voltage and high current High slew rate with 100A/μs Low voltage input and high slew rate electronic load Kenwood TMI Corporation

More information

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax: OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,

More information

AC/DC Current Probe Model SL201

AC/DC Current Probe Model SL201 AC/DC Current Probe Model SL201 USER MANUAL I ZERO 1 V/A 1 mv/a OFF Limited Warranty The AC/DC Current Probe Model SL201 is warranted to the owner for a period of two years from the date of original purchase

More information