FTIR-ATR Identification of a Food Oil Using Chemometrics

Size: px
Start display at page:

Download "FTIR-ATR Identification of a Food Oil Using Chemometrics"

Transcription

1 FTIR-ATR Identification of a Food Oil Using Chemometrics *NOTE: BRING A FLASHDRIVE TO LAB ON THE DAY YOU INTEND TO DO THE EXPERIMENT AS YOU WILL NEED TO TRANSFER THE DATA FILES FOR FURTHER PROCESSING IN EXCEL. 1. Purpose This procedure is designed to determine the identity of a food oil using Fourier-transform infrared (FTIR) spectroscopic data obtained with an attenuated total reflectance (ATR) cell. 2. Background While scientists regularly do quantitative analysis to determine the concentration(s) of some unknown compound(s) in a particular sample, it is also common to do qualitative analysis to ascertain the identity of a sample and perhaps its chemical composition and overall purity. There are two experiments in Chm 466 that are qualitative in nature, both using a food oil as the unknown. Food oils are mixtures of triglycerides that consist of three long fatty acids attached to a glycerol backbone. While there are many different fatty acids, they differ only in the number of carbons and double bonds present in the long fatty acid tails. The figure below shows a triglyceride with three common fatty acids, the top one being the saturated palmitic acid, the middle being the monounsaturated oleic acid, and the bottom being the polyunsaturated α- linolenic acid. Note that the polyunsaturated linolenic acid has three cis- double bonds and so is chemically named all-cis-9,12,15-octadecatrienoic acid. Different food oils have different percentages of the various fatty acids. In the GC-MS experiment this semester, an unknown oil is identified based upon the relative amounts of its constituent fatty acids. Doing so requires chemical derivatization of the oil prior to analysis so that the triglyceride fatty acids are converted to their corresponding fatty acid methyl esters (FAMEs) with boiling points low enough to be done by gas chromatography. The unknown oil is then identified based upon the variable percentages of individual fatty acids as measured by the GC-MS. While this method is very accurate and can provide positive identification of the different fatty acids, it is somewhat time-consuming due to the derivatization and extraction process and subsequent chromatographic analysis. A simpler, though possibly less precise technique for determining the identity of an oil relies upon infrared (IR) spectroscopy. As you learned in organic chemistry, IR wavelengths interact with the vibrational modes of a molecule s bonds, with different bonding pairs absorbing IR radiation at different wavelengths (or wavenumbers, as is traditionally used in IR). In this procedure, you will obtain FTIR spectra of neat oil samples individually analyzed on an attenuated total reflectance (ATR) crystal. Using ATR-FTIR is much simpler than traditional spectroscopic methodologies as there is no sample cell. The minimal sample preparation results in a very simple procedure.

2 Schematic of ATR-FTIR instrument The ATR-FTIR data allow you to quickly fingerprint an unknown oil and then identify it by comparison to a set of standard spectra for various oils. However, since all food oils are very similar in their composition (all consist of the same C-14 to C-22 fatty acids attached to a glycerol backbone, the only difference being the relative amounts of the various fatty acids), the spectral differences are small and often almost indiscernible at least to the naked eye. As such, the data needs to be processed using a basic form of chemometrics you will do using an Excel spreadsheet. 3. Materials and Equipment Oil standards and unknowns obtained from instructor. Plastic or Pasteur transfer pipettes, kimwipes, cotton swabs, and acetone should be available by the instrument. 4. Safety Safety glasses should be worn at all times. Food oils are for analysis only and, like all other laboratory chemicals, are not intended for human consumption. For more general safety in the laboratory, please refer the appendix. 5. Procedure Obtain the set of five oil standards from your instructor along with your unknown and proceed to the instrumental room. You will possibly need to bring along several plastic transfer pipettes and your notebook. A bottle of acetone (wash bottle with the red top) should be near the Nicolet Avatar 370 FTIR instrument located near the north double doors of the laboratory. If there is no acetone, consult your instructor.

3 Picture of Thermo-Nicollet is10 FT-IR Confirm that the computer and instrument are both on (the instrument on light is located at the top front right of the instrument). If the instrument is not on, you will eventually get a computer message: Computer cannot communicate with spectrometer, when you open the OMNIC software. If you see that, please see the instructor immediately. The ATR crystal must be cleaned prior to each spectral measurement and before beginning the experiment. Cleaning is important for your precision and accuracy. Picture of ATR Assembly with the ATR crystal Next, you need to clean the cell itself. Clean around the outside of the crystal by gently rubbing with a Kimwipe and acetone. DO NOT TOUCH THE SMALL CENTER CRYSTAL WITH A KIMWIPE AS THAT CAN SCRATCH AND DAMAGE THE DELICATE CRYSTAL. Put 2-3 drops of acetone on the small shiny center cell and gently wipe in circular fashion with a Q-tip. Once you have wiped off the acetone, you have fully

4 cleaned the apparatus and are ready to begin analysis of the first standard. You will repeat this cleaning procedure for each trial of your standard and unknown. Before starting the software to acquire the data on FT-IR, plug your USB flash drive into a USB port. Once you have your USB flash drive plugged into the computer, open the OMNIC software program (desktop icon with the multicolored triangle), again assuring that the spectrometer is on by noting the green light at the top left-hand corner. Wait for a few seconds until you see a message about an accessory change, simply press the OK button to continue. Expand the window to full-size. Windows desktop showing OMNIC icon Upon starting OMNIC, you may see this window. Simply click OK to continue. Go to the Collect menu, select Experiment Setup. The Experimental Setup window should open. At the bottom of this window, click on the Open command and load/open 466 Lab Food Oil Analysis.exp that has the correct instrument settings for running the oils. Click OK (bottom right of window) to accept the experimental parameters. Open Experiment window. Select 466 Lab Food Oil Analysis.exp. Once you select the correct experiment file, you should see Experiment Setup shown right above.

5 You are now ready to begin your first trial. On the icons directly above the blank spectrum, click the Col Smp (Collect Sample). OMNIC tool buttons. You will be prompted with a window that says, Enter the spectrum title where you will enter your initials, the date of the analysis, and some identifier of the sample type (e.g., AZ0904Walnut1 or DD1021Unk7121 ). You MUST NOT have any spaces in your file names. Enter the sample name here. DO NOT have space in the sample name here. As you will run your standards and unknown in triplicate, make the last letter of your title 1, 2 or 3 to denote the different trials. Be sure to record the file name(s) in your notebook. After you click on the OK button, you will possibly get a prompt stating, A background spectrum must exist to produce the specified final format. Click OK. You will subsequently see a prompt that says, Background. Please prepare to collect the background spectrum. As the ATR crystal is clean, press OK, at which point the instrument will make 16 scans of the background spectrum. OMNIC will ask to perform the background scan. Click OK buttons to continue with no samples loaded in ATR.

6 A typical background scan should look like the one shown above. Notice that the data is still being acquired (shown as 7/16 on the lower left corner of the window here). You can monitor the progress of the background collection by the black bar icon in the lower left-hand corner. This will be the only blank background run during the entire experiment. After completion of background, the computer will prompt Sample. Please prepare to collect the sample spectrum. Once the background is done, you will be prompted for your sample. Load your sample BEFORE clicking OK button. It is now time to run the first trial of your first standard. The order in which you run the standards is unimportant, but they should be run before your unknown. Use a Pasteur pipette to slowly add 1-2 drops of the oil onto the crystal (the exact amount is not important; add just enough so that a drop of oil covered the crystal) above the FTIR crystal.

7 Picture of ATR crystal with a drop of oil. After adding the oil, run the sample by pressing OK on the computer. The computer will acquire and average 16 scans over the next seconds with the progress again shown by the black bars in the lower left corner. Once the 16 scans have been averaged, you will be prompted to Add to Window 1? Enter Yes. Once the sample spectrum is collect ed, OMNIC will ask if you would like to add the spectrum to the active window. Answer Yes to continue. At this point, you should have only the corrected %Transmittance showing on your screen (usually in red color) with %Transmittance plotted versus Wavenumber with negative peaks. Note that as transmittance and absorbance are inverse functions, the peaks with the lowest %T are at wavelengths where the oil exhibits maximum absorbance. You do not need to print off the spectra, but you MUST save your data.

8 Typical spectrum of an oil sample To save the data, you need to first change the file type to CSV text, a format that is compatible with Excel. To do so, go to File menu, select Save As dialogue box (also can be done by clicking the F12 hot key). DO NOT click on Save button (nor select Save from File menu, as that will NOT save in CSV format). In the Save as type dialogue box and scroll down until you find the CSV Text file type. After choosing the CSV option, click on the Set Filename to Title button and save with the CSV file type. Once again, you MUST NOT have any spaces in your file names. Check the file name before you save. If you have a space in your file name, please remove the space. This will save your data with the same name as before but with a CSV file type extension. Make sure you save each trial with the CSV extension. To save your data, you should go to your USB flash drive. You should also make sure that you are saving the data as a CSV file as shown on right. Once you have saved your data, use a Q-tip to soak up the oil on the crystal and then clean the crystal as described above. As before, the yellow cell crystal should only be cleaned with a

9 Q-tip and acetone. You are then ready to run the second and third trials of your first standard. These and all successive runs will be made with the original blank. Trials should be treated identically as the first trial (save with csv file extension using a slightly different filename, etc.). As such, you will have three data files for each standard and unknown. Again, you do not need to run the blank background other than for the first trial of the first standard. After running each standard and unknown in triplicate, proceed to the data analysis without exiting OMNIC. You may need to re-run a few samples, but you won t know that until you plot the data first. You may not be able to properly eject your USB drive. That s OK. Simply pull it out of the USB port. Leave the instrument and computer both on. Make sure that no oil remains on the crystal and that the station is clean; if the acetone bottle is low, inform the instructor. At this point, you should have 18 CSV files with unique name on your flash drive. If you do not have 18 files, check your work. You should also tabulate all the 18 file names in your lab notebook for a reference. Before leaving, transfer your data files to your flash drive and proceed to the data analysis section.

10 6. Data Analysis Data analysis will be done using Excel and the computers in the balance room. There are four spectral regions that have been deemed particularly useful for determining the identity of the unknown. Those FTIR wavenumber regions are: (Region A identified as being a cis-c=c stretching mode)* (Region B identified as being an asymmetric C-H (-CH 2 -) n stretching mode)* (Region C identified as being a cis-c=c bending mode)* (Region D identified as being a trans-c=c bending mode)* *Lerma-Garcίa, M.J., et al., Food Chem 118 (2010) To process your data, you will need to use one of the computers in the balance room. Begin by inserting your flash drive into a USB port on the back of the desktop tower. You should get a message on the desktop showing that the computer has recognized the flash drive. If you wish, open the My Computer icon on the upper-left corner of the desktop and then open your Removable Disk flash drive. Once the flash drive files are seen, open the Excel file FT-IR on the desktop. This program will combine all the 18 CSV files you have created into one file that can be analyzed in Excel easily. If you do not have 18 files, you should check your work. It will first ask for the files for the standard 1 trials.

11 Upon starting the FT-IR program, it will ask you a series of questions. You will be asked to select your CSV files. Notice the title of the window indicates the file it is asking for. In this example, it is asking for the Standard 1 Trial 1 data. Select the file, then click OK to continue. It will ask more files once you do so. You will have to choose all 15 files for the standard trials and the three unknown trials (18 files total). Once you are done selecting the files, it should ask you for confirmation. Check your notebook for accuracy, making sure the list in your notebook matches with the list shown on the computer screen. If the dialog box does not appear as shown next, ask your laboratory instructor for help.

12 Once you have selected all 18 files, it will show you which ones you have selected. Check what you have in your lab notebook against what is being shown BEFORE you click OK. If it does not match, click Cancel and start the data analysis part over from the beginning. If the list does not look shown above (if the Trial column does not look as shown above), consult your lab instructor immediately. Any time during this part, if you have made any mistakes, click Cancel button. The program will stop, then, you will have to start over with the data analysis (starting from selecting the standard 1 trial 1 file). One you have all the 18 files selected correctly, the program will combine all the 18 CSV files into one singular file called FT-IRdata.CSV in the same directory as where you have all of your 18 CSV files. Click OK to continue. Your data should be saved in the same directory as where your 18 CSV files are saved (should be on your flash drive). After that, Microsoft Excel should automatically start. Note the worksheet tabs at the bottom of the spreadsheet entitled Graph, Trials, and Averages.

13 Notice the different worksheet listed in this spreadsheet. There should be a button called Import CSV on the Graph worksheet, on A2-A3 cells. Click that button. Click Import CSV button shown above. Once you do, it will ask for your name. Enter your name (this is NOT an optional. You have to enter your name here to continue). You will be asked to enter your name. Enter your name. This is NOT optional, and the only way to get through is to enter your name. Once you enter your name, your CSV file should be imported automatically into the Trials worksheet (and the Excel should switch the active worksheet to the Trials worksheet). If this does not work, consult your laboratory instructor for help. You are now ready to extract and average (necessary because the data is noisy, making discrimination between the different oils impossible with merely the peak minimum) the data for the four spectral regions of interest. Go to the Averages worksheet tab at the bottom of the FTIR Template spreadsheet. If everything has been entered correctly, the headings in the first row will match the names of your standards and unknown. Go to Cell B3 to begin averaging the data for the first region by typing: =average(trials!b20351:b20524). You should now see a single value in Cell B3 that is the average of the %T for the wavenumber region from cm -1. Enter the equation as shown above. You need to average the three other regions for your first standard trial in the same way. Cell B4 of the Averages folder will contain the average %T for the wavenumber between

14 cm -1. To do this, go to Cell B4 and type: =average(trials!b19687:b19861). Once entered, you should see a single value representing the average %T. Subsequently, go to Cell B4 and type: =average(trials!b3011:b3184), which will give the average %T for the wavenumber region from cm -1. Finally, go to Cell B6 and type: =average(trials!b3426:b3599), which will give the average %T for the wavenumber region from cm -1 You should be entering formulas shown below in each cells: Cell B3 B4 B5 B6 Formula =average(trials!b20351:b20524) =average(trials!b19687:b19861) =average(trials!b3011:b3184) =average(trials!b3426:b3599) To complete this part of the data analysis for the other trials, you can simply copy the formulas from Cells B3-B6 to all the adjacent cells to the right (C3-C6, D3-D6, etc.). To do so, use the mouse to highlight Cells B3-B6, move the mouse pointer to the right bottom corner of the selected cells until the pointer changes from a white cross to a black cross (see the figures below). You should see the mouse pointer changes to a black cross if you move the pointer over the small black square in the right bottom corner of the selected cells. Once you see the pointer changes to the black cross, while holding the left mouse button, scroll the mouse to the right so that all the remaining cells are highlighted. Once you release the left button, all averaged values should be shown on the Averages folder. OnceB3-B6 have formula entered, you can populate the rest of the worksheet as shown above. Notice the mouse pointer changes to a black cross from white cross on the right. Click the mouse button and drag to right as shown below.

15 If you drag it correctly, all the cells should be populated with the same formula. Alternatively, select cells B3-B6, go to Edit, select Copy from the menu, then, select cells C3-S6, and go to Edit select Paste from the menu. This should populate the range of the cells with the correct formula. At this point, click the Graph worksheet tab at the bottom of the FTIR Template worksheet. You should see your averaged data for all standards and unknowns. There should also be a graph of the region ratios (A/B on the x-axis and C/D on the y-axis). If you did the experiment correctly, you should see relatively close groupings of each standard and your unknown. Print the worksheet (be sure your printout contains both the table and the graph). Based on the A/B vs. C/D graph you obtain, determine the identity of the oil unknown. Points will be based upon not only making the correct choice, but also upon thoroughly explaining your choice and also on the overall quality of your data (graphical precision and accuracy). The grading rubric is shown in the next section. You should probably save your work. Your data will be saved automatically on your flash drive along side of the CSV files as FTIR Template.xls. 7. Reporting Requirements and Grading If you do not identify the oil correctly, you will receive a TFO. If you identify the oil correctly, you will receive points for this and you must also describe your analysis and defend your choice. Go back and re-read the introduction to the experiment, then start on your write-up. Your write up will be fairly brief, but should succinctly describe these things. Write it in your notebook neatly and legibly. Points will be deducted from messy or illegible write-ups. Grading Rubric: Precision of each set of data Data Analysis (address points below) Description of what data was collected. 80 pts 20 pts Describe why you are using this particular method of analysis, rather than looking at the raw data (spectra). A description of your graph, particularly what data you are graphing. Your reasoning for your oil choice.

16 Instrumental Analysis Laboratory Safety Rules A. Instructions: Carry out all manipulations in accordance with instructions and the safety rules and procedures given herein. B. Eye Protection: All students and staff working in the laboratory must wear safety glasses at all times. If a student needs to be reminded more than three times to wear goggles, she/he will be dismissed from lab for the remainder of the day, and will not be given an opportunity to make up the work. C. Apparel: The clothes you wear in lab are an important part of your safety equipment, and should offer protection from splashes/spills. Closed toed shoes (sneakers are fine), Full-length pants or a full-length skirt, and A shirt that completely covers your torso (i.e. at minimum, a t-shirt). In other words, you must NOT wear shorts to lab. You must NOT wear flip-flops, sandals, or crocs. You must NOT wear tank tops, halter tops, spaghetti-strap tops, or low cut jeans to lab. Exposed abdomens, hips, and backs are not safe in the lab. D. Gloves: Gloves are an important part of personal protection. Gloves will be available at all times in the laboratory. Your instructor will require their use when appropriate. E. Food: Food, drinks, and gum are not allowed in lab. None at all, not even water bottles. F. Sanitation Issues: Be sure to wash your hands before leaving lab, before you eat anything outside of lab, and before you answer your cell phone. G. Music: Individual headphones are not allowed. Your may choose to play music for the entire class. H. Cell Phones and Other Electronic Devices: Cellular phones and other electronic devices that you do not need to perform your laboratory work should be put away. I. Other: All students are explicitly prohibited from: 1. conducting any unauthorized experiments. 2. removing chemicals or apparatus from the laboratory for any reason. 3. working in the lab alone, or at other than regularly scheduled lab periods. 4. smoking in the laboratory or within 20 feet of any doorway. 5. impeding movement in aisles or through doorways with bags, skateboards, etc.

FTIR-ATR Identification of a Food Oil Using Chemometrics

FTIR-ATR Identification of a Food Oil Using Chemometrics FTIR-ATR Identification of a Food Oil Using Chemometrics *NOTE: BRING A FLASHDRIVE TO LAB ON THE DAY YOU INTEND TO DO THE EXPERIMENT AS YOU WILL NEED TO TRANSFER THE DATA FILES FOR FURTHER PROCESSING IN

More information

Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013

Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013 Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013 Adapted from JP Blitz and DG Klarup, "Signal-to-Noise Ratio, Signal

More information

Nicolet is50 FTIR User s Booklet

Nicolet is50 FTIR User s Booklet Nicolet is50 FTIR User s Booklet I. Getting started 1) Start your session by launching the LSA Chemistry Recharge software. This software keeps track of the amount of time that you are using the instrument.

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

FTIR Spectroscopy. (Brucker Vector 22)

FTIR Spectroscopy. (Brucker Vector 22) FTIR Spectroscopy (Brucker Vector 22) Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. 1.0 PURPOSE: In infrared

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Once FTIR data has been acquired for the gases evolved during your DSC/TGA run, you will process using the OPUS software package. Select

More information

Agilent MicroLab Quant Calibration Software: Measure Oil in Water using Method IP 426

Agilent MicroLab Quant Calibration Software: Measure Oil in Water using Method IP 426 Agilent MicroLab Quant Calibration Software: Measure Oil in Water using Method IP 426 Application Note Environmental Authors John Seelenbinder and Dipak Mainali Agilent Technologies, Inc. Introduction

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

Nicolet FT-IR Procedure

Nicolet FT-IR Procedure Nicolet FT-IR Procedure Getting Started Jerry Hu (Lab Manager), ext. 7914 or 7940 jghu@mrl.ucsb.edu Jaya Nolt (Lab Tech), ext. 4997 jaya@mrl.ucsb.edu Materials Research Laboratory University of California,

More information

Automating standard operating procedures using simple scripting

Automating standard operating procedures using simple scripting Automating standard operating procedures using simple scripting Technical Overview Advantage statement Easy to create, customized software scripts created using Resolutions Pro Software reduce user training

More information

Microsoft Excel Using Excel in the Science Classroom

Microsoft Excel Using Excel in the Science Classroom Microsoft Excel Using Excel in the Science Classroom OBJECTIVE Students will take data and use an Excel spreadsheet to manipulate the information. This will include creating graphs, manipulating data,

More information

67 Series Spectrophotometer PC Software

67 Series Spectrophotometer PC Software 67 Series Spectrophotometer PC Software Instruction Manual 670 004/Rev D/04-17 IMPORTANT Please ensure the SD Card is fitted into the socket as detailed in the image below. -------------------------------------------------------------------------------------------------------------------------

More information

EXCEL SPREADSHEET TUTORIAL

EXCEL SPREADSHEET TUTORIAL EXCEL SPREADSHEET TUTORIAL Note to all 200 level physics students: You will be expected to properly format data tables and graphs in all lab reports, as described in this tutorial. Therefore, you are responsible

More information

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI Laser Spectroscopy Labs Jasco 4700 FTIR Operation instructions Turn the system on 1. Turn on the computer in case it has been switched off. Normally computer stays on. 2. The instrument should be always

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation The GVSU chemistry department owns 2 NMR spectrometers. A JEOL Eclipse 300 MHz and a Varian Inova 400 MHz. Due to their strong, constant magnetic

More information

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS April 16, 2018 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

Using Microsoft Excel

Using Microsoft Excel About Excel Using Microsoft Excel What is a Spreadsheet? Microsoft Excel is a program that s used for creating spreadsheets. So what is a spreadsheet? Before personal computers were common, spreadsheet

More information

OCEAN OPTICS Chem 2000 Spectrophotometer

OCEAN OPTICS Chem 2000 Spectrophotometer OCEAN OPTICS Chem 2000 Spectrophotometer prepared for: Department of Chemistry and Biochemistry New Mexico State University Las Cruces, NM 88003 last updated May 2003 R.. Cowley TABLE OF CONTENTS I. Introduction...2

More information

Intermediate Microsoft Excel (Demonstrated using Windows XP) Using Spreadsheets in the Classroom

Intermediate Microsoft Excel (Demonstrated using Windows XP) Using Spreadsheets in the Classroom (Demonstrated using Windows XP) Using Spreadsheets in the Classroom Adapted from Taskstream Word Tutorial (2003) < http://www.taskstream.com > Updated 4/05 by Dr. Bruce Ostertag What Can Microsoft Excel

More information

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro CVI SPECTRAL PRODUCTS 111 Highland Drive Putnam CT, 06260 (860) 928-5834 SM32Pro 2.8.28 Table of Contents Warranty and Liability 1 Quick Start Installation Guide 2 System Requirements 3 Requirements for

More information

Excel 2016 Basics for Mac

Excel 2016 Basics for Mac Excel 2016 Basics for Mac Excel 2016 Basics for Mac Training Objective To learn the tools and features to get started using Excel 2016 more efficiently and effectively. What you can expect to learn from

More information

How to use Excel Spreadsheets for Graphing

How to use Excel Spreadsheets for Graphing How to use Excel Spreadsheets for Graphing 1. Click on the Excel Program on the Desktop 2. You will notice that a screen similar to the above screen comes up. A spreadsheet is divided into Columns (A,

More information

LEGENDplex Data Analysis Software Version 8 User Guide

LEGENDplex Data Analysis Software Version 8 User Guide LEGENDplex Data Analysis Software Version 8 User Guide Introduction Welcome to the user s guide for Version 8 of the LEGENDplex data analysis software for Windows based computers 1. This tutorial will

More information

Excel 2003 Tutorial II

Excel 2003 Tutorial II This tutorial was adapted from a tutorial by see its complete version at http://www.fgcu.edu/support/office2000/excel/index.html Excel 2003 Tutorial II Charts Chart Wizard Chart toolbar Resizing a chart

More information

Computer Essentials Session 1 Lesson Plan

Computer Essentials Session 1 Lesson Plan Note: Completing the Mouse Tutorial and Mousercise exercise which are available on the Class Resources webpage constitutes the first part of this lesson. ABOUT PROGRAMS AND OPERATING SYSTEMS Any time a

More information

Using Scantron ParLAN 6.5 for the First Time:

Using Scantron ParLAN 6.5 for the First Time: Page 1 of 19 Using Scantron ParLAN 6.5 for the First Time: Last updated: Monday, December 02, 2013 Nine Basic Steps To request a Scantron Account, please contact the Academic Technology Center. Step One:

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

When you first start OneNote, it creates a sample notebook for you. You can use this notebook or quickly create your own.

When you first start OneNote, it creates a sample notebook for you. You can use this notebook or quickly create your own. Basic tasks in Microsoft OneNote 2013 OneNote is a digital notebook that provides a single place for all of your notes and information everything you need to remember and manage in your life at home, at

More information

Perkin-Elmer Lambda 3B Spectrophotometer

Perkin-Elmer Lambda 3B Spectrophotometer Written by Mark Csele Written by Jeff Brown Operation Turn on the Lambda 3B and the Computer Remove all samples from the sample chamber. Turn on the lamp for the range being investigated using the switches

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

Visual Physics Introductory Lab [Lab 0]

Visual Physics Introductory Lab [Lab 0] Your Introductory Lab will guide you through the steps necessary to utilize state-of-the-art technology to acquire and graph data of mechanics experiments. Throughout Visual Physics, you will be using

More information

For technical assistance, please contact: Technical Support Thermo Fisher Scientific 5225 Verona Road Madison WI U.S.A.

For technical assistance, please contact: Technical Support Thermo Fisher Scientific 5225 Verona Road Madison WI U.S.A. The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Fisher Scientific shall not be liable for

More information

Introduction to Excel

Introduction to Excel Introduction to Excel Written by Jon Agnone Center for Social Science Computation & Research 145 Savery Hall University of Washington Seattle WA 98195 U.S.A. (206)543-8110 November 2004 http://julius.csscr.washington.edu/pdf/excel.pdf

More information

INFRARED SPECTROSCOPY. MATTSON FTIR INSTRUCTIONS FOR THE CLASSROOM TEACHER (Revised: )

INFRARED SPECTROSCOPY. MATTSON FTIR INSTRUCTIONS FOR THE CLASSROOM TEACHER (Revised: ) MATTSON FTIR INSTRUCTIONS FOR THE CLASSROOM TEACHER (Revised: 1-23-93) BEGINNING A SESSION 1. Make sure the IR path is clear (Sample compartment empty). 2. Access the software by entering UFIRST on the

More information

Microsoft Excel Level 2

Microsoft Excel Level 2 Microsoft Excel Level 2 Table of Contents Chapter 1 Working with Excel Templates... 5 What is a Template?... 5 I. Opening a Template... 5 II. Using a Template... 5 III. Creating a Template... 6 Chapter

More information

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1 New software features of the 7010 Particle Size May 5, 2009 Analyzer 2.0.0 Danielle Chamberlin May 5 2009 Page 1 Introduction to the new 7010 software Instrument operation New look and feel easy to operate

More information

Cmpt 101 Lab 1 - Outline

Cmpt 101 Lab 1 - Outline Cmpt 101 Lab 1 - Outline Instructions: Work through this outline completely once directed to by your Lab Instructor and fill in the Lab 1 Worksheet as indicated. Contents PART 1: GETTING STARTED... 2 PART

More information

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s

More information

DataStudio Starter Manual

DataStudio Starter Manual Instruction Manual Manual No. 012-08107 DataStudio Starter Manual Manual No. 012-08107 Manual No. 012-08107 DataStudio Starter Manual Introduction What is DataStudio? DataStudio is a data acquisition,

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Using HyperChem for CH 463. Chun Park

Using HyperChem for CH 463. Chun Park Using HyperChem for CH 463 Chun Park Using HyperChem for CH 463 1 st Edition Written by Chun Park Oregon State University Table of Contents Introduction 1 Chapter 1 Getting Started 2 Starting HyperChem...

More information

Multivariate Calibration Quick Guide

Multivariate Calibration Quick Guide Last Updated: 06.06.2007 Table Of Contents 1. HOW TO CREATE CALIBRATION MODELS...1 1.1. Introduction into Multivariate Calibration Modelling... 1 1.1.1. Preparing Data... 1 1.2. Step 1: Calibration Wizard

More information

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Pipettes allow transfer of accurately known volume of solution from one container to another. Volumetric or transfer pipettes deliver a

More information

DOING MORE WITH EXCEL: MICROSOFT OFFICE 2013

DOING MORE WITH EXCEL: MICROSOFT OFFICE 2013 DOING MORE WITH EXCEL: MICROSOFT OFFICE 2013 GETTING STARTED PAGE 02 Prerequisites What You Will Learn MORE TASKS IN MICROSOFT EXCEL PAGE 03 Cutting, Copying, and Pasting Data Basic Formulas Filling Data

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

KINETICS CALCS AND GRAPHS INSTRUCTIONS

KINETICS CALCS AND GRAPHS INSTRUCTIONS KINETICS CALCS AND GRAPHS INSTRUCTIONS 1. Open a new Excel or Google Sheets document. I will be using Google Sheets for this tutorial, but Excel is nearly the same. 2. Enter headings across the top as

More information

Making an entry into the CIS Payments workbook

Making an entry into the CIS Payments workbook Making an entry into the CIS Payments workbook By now you should have carried out the CIS Payments workbook Setup. If you have not done so you will need to do this before you can proceed. When you have

More information

Marks Management & ScanExam Introduction

Marks Management & ScanExam Introduction Marks Management & ScanExam Introduction 1. Install Marks Management Software (MMS) and ScanExam First uninstall your old version of MMS. Then, use a web browser to visit ssnds.uwo.ca/softwaredownloads.html

More information

Opening and Saving Spectra and Reports Opening a spectrum or report Saving a spectrum or report...112

Opening and Saving Spectra and Reports Opening a spectrum or report Saving a spectrum or report...112 The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Fisher Scientific shall not be liable for

More information

Overview. Experiment Specifications. This tutorial will enable you to

Overview. Experiment Specifications. This tutorial will enable you to Defining a protocol in BioAssay Overview BioAssay provides an interface to store, manipulate, and retrieve biological assay data. The application allows users to define customized protocol tables representing

More information

Chemistry 30 Tips for Creating Graphs using Microsoft Excel

Chemistry 30 Tips for Creating Graphs using Microsoft Excel Chemistry 30 Tips for Creating Graphs using Microsoft Excel Graphing is an important skill to learn in the science classroom. Students should be encouraged to use spreadsheet programs to create graphs.

More information

DOING MORE WITH EXCEL: MICROSOFT OFFICE 2010

DOING MORE WITH EXCEL: MICROSOFT OFFICE 2010 DOING MORE WITH EXCEL: MICROSOFT OFFICE 2010 GETTING STARTED PAGE 02 Prerequisites What You Will Learn MORE TASKS IN MICROSOFT EXCEL PAGE 03 Cutting, Copying, and Pasting Data Filling Data Across Columns

More information

Chemistry Excel. Microsoft 2007

Chemistry Excel. Microsoft 2007 Chemistry Excel Microsoft 2007 This workshop is designed to show you several functionalities of Microsoft Excel 2007 and particularly how it applies to your chemistry course. In this workshop, you will

More information

Spreadsheet definition: Starting a New Excel Worksheet: Navigating Through an Excel Worksheet

Spreadsheet definition: Starting a New Excel Worksheet: Navigating Through an Excel Worksheet Copyright 1 99 Spreadsheet definition: A spreadsheet stores and manipulates data that lends itself to being stored in a table type format (e.g. Accounts, Science Experiments, Mathematical Trends, Statistics,

More information

252 APPENDIX D EXPERIMENT 1 Introduction to Computer Tools and Uncertainties

252 APPENDIX D EXPERIMENT 1 Introduction to Computer Tools and Uncertainties 252 APPENDIX D EXPERIMENT 1 Introduction to Computer Tools and Uncertainties Objectives To become familiar with the computer programs and utilities that will be used throughout the semester. You will learn

More information

Budget Exercise for Intermediate Excel

Budget Exercise for Intermediate Excel Budget Exercise for Intermediate Excel Follow the directions below to create a 12 month budget exercise. Read through each individual direction before performing it, like you are following recipe instructions.

More information

A cell is highlighted when a thick black border appears around it. Use TAB to move to the next cell to the LEFT. Use SHIFT-TAB to move to the RIGHT.

A cell is highlighted when a thick black border appears around it. Use TAB to move to the next cell to the LEFT. Use SHIFT-TAB to move to the RIGHT. Instructional Center for Educational Technologies EXCEL 2010 BASICS Things to Know Before You Start The cursor in Excel looks like a plus sign. When you click in a cell, the column and row headings will

More information

Error Analysis, Statistics and Graphing

Error Analysis, Statistics and Graphing Error Analysis, Statistics and Graphing This semester, most of labs we require us to calculate a numerical answer based on the data we obtain. A hard question to answer in most cases is how good is your

More information

Comfort Optimizer Installation Instructions

Comfort Optimizer Installation Instructions Congratulations on choosing the best all around load calculation program available. Along with this instruction manual should be the installation CD. Before we can install the Comfort Optimizer, we have

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

COPYRIGHTED MATERIAL INTRODUCTION TO ASPEN PLUS CHAPTER ONE

COPYRIGHTED MATERIAL INTRODUCTION TO ASPEN PLUS CHAPTER ONE CHAPTER ONE INTRODUCTION TO ASPEN PLUS Aspen Plus is based on techniques for solving flowsheets that were employed by chemical engineers many years ago. Computer programs were just beginning to be used,

More information

Tricking it Out: Tricks to personalize and customize your graphs.

Tricking it Out: Tricks to personalize and customize your graphs. Tricking it Out: Tricks to personalize and customize your graphs. Graphing templates may be used online without downloading them onto your own computer. However, if you would like to use the templates

More information

STANDARD OPERATING PROCEDURES SOP : 1736 PAGE: 1 of 35 REV: 0.0 DATE: 09/25/06 MEASUREMENT OF AIR CHANGES USING THE WOLFSENSE IAQ PROBE

STANDARD OPERATING PROCEDURES SOP : 1736 PAGE: 1 of 35 REV: 0.0 DATE: 09/25/06 MEASUREMENT OF AIR CHANGES USING THE WOLFSENSE IAQ PROBE PAGE: 1 of 35 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Excel 2016 Basics for Windows

Excel 2016 Basics for Windows Excel 2016 Basics for Windows Excel 2016 Basics for Windows Training Objective To learn the tools and features to get started using Excel 2016 more efficiently and effectively. What you can expect to learn

More information

Section 1 Microsoft Excel Overview

Section 1 Microsoft Excel Overview Course Topics: I. MS Excel Overview II. Review of Pasting and Editing Formulas III. Formatting Worksheets and Cells IV. Creating Templates V. Moving and Navigating Worksheets VI. Protecting Sheets VII.

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

Advice for How To Create a Film Project in Windows MovieMaker

Advice for How To Create a Film Project in Windows MovieMaker Advice for How To Create a Film Project in Windows MovieMaker This document was compiled to provide initial assistance to teachers and/or students to create a movie project using the Windows MovieMaker

More information

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW Objectives: The goal of this exercise is to introduce the Laboratory Virtual Instrument Engineering Workbench, or LabVIEW software. LabVIEW is the primary software

More information

Lastly, in case you don t already know this, and don t have Excel on your computers, you can get it for free through IT s website under software.

Lastly, in case you don t already know this, and don t have Excel on your computers, you can get it for free through IT s website under software. Welcome to Basic Excel, presented by STEM Gateway as part of the Essential Academic Skills Enhancement, or EASE, workshop series. Before we begin, I want to make sure we are clear that this is by no means

More information

Experiment 3 Microsoft Excel in Scientific Applications I

Experiment 3 Microsoft Excel in Scientific Applications I Experiment 3 Microsoft Excel in Scientific Applications I OUTCOMES After completing this experiment, the student should be able to: demonstrate basic computer survival skills (opening, closing, printing

More information

The Microscope: A Tool of the Scientist

The Microscope: A Tool of the Scientist Name Class Date Microscope Lab The Microscope: A Tool of the Scientist Background Information (You may refer to page 61 of your textbook.) Problem One of the most important tools of a life scientist is

More information

Assignment 4: Molecular Dynamics (50 points)

Assignment 4: Molecular Dynamics (50 points) Chemistry 380.37 Fall 2015 Dr. Jean M. Standard November 2, 2015 Assignment 4: Molecular Dynamics (50 points) In this assignment, you will use the Tinker molecular modeling software package to carry out

More information

Manual for Jasco-IR-spectrometer

Manual for Jasco-IR-spectrometer Manual for Jasco-IR-spectrometer Do not record spectra of hazardous volatile liquids! Do not record spectra of corrosive compounds! Do not record spectra of hard and sharp edged material (sand, broken

More information

Cell to Cell mouse arrow Type Tab Enter Scroll Bars Page Up Page Down Crtl + Home Crtl + End Value Label Formula Note:

Cell to Cell mouse arrow Type Tab Enter Scroll Bars Page Up Page Down Crtl + Home Crtl + End Value Label Formula Note: 1 of 1 NOTE: IT IS RECOMMENDED THAT YOU READ THE ACCOMPANYING DOCUMENT CALLED INTRO TO EXCEL LAYOUT 2007 TO FULLY GRASP THE BASICS OF EXCEL Introduction A spreadsheet application allows you to enter data

More information

Using Excel for Graphical Analysis of Data

Using Excel for Graphical Analysis of Data Using Excel for Graphical Analysis of Data Introduction In several upcoming labs, a primary goal will be to determine the mathematical relationship between two variable physical parameters. Graphs are

More information

Using Excel for Graphical Analysis of Data

Using Excel for Graphical Analysis of Data EXERCISE Using Excel for Graphical Analysis of Data Introduction In several upcoming experiments, a primary goal will be to determine the mathematical relationship between two variable physical parameters.

More information

Links to Activities ACTIVITY 1.1. Links to Activities Links to Activities

Links to Activities ACTIVITY 1.1. Links to Activities Links to Activities EXCEL Analyzing Data Using Excel Section 1 Skills Start Excel and identify features in the Excel window Enter labels and values Use the fill handle to enter a series Enter formulas Create a formula using

More information

Excel Core Certification

Excel Core Certification Microsoft Office Specialist 2010 Microsoft Excel Core Certification 2010 Lesson 6: Working with Charts Lesson Objectives This lesson introduces you to working with charts. You will look at how to create

More information

EXCEL BASICS: PROJECTS

EXCEL BASICS: PROJECTS EXCEL BASICS: PROJECTS In this class, you will be practicing with three basic Excel worksheets to learn a variety of foundational skills necessary for more advanced projects. This class covers: Three Project

More information

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific Spectronic 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

NMR Users Guide Organic Chemistry Laboratory

NMR Users Guide Organic Chemistry Laboratory NMR Users Guide Organic Chemistry Laboratory Introduction The chemistry department is fortunate to have a high field (400 MHz) Nuclear Magnetic Resonance (NMR) spectrometer. You will be using this instrument

More information

Introduction to Powerpoint. Create it. Opening a presentation

Introduction to Powerpoint. Create it. Opening a presentation Introduction to Powerpoint Create it You will make a short presentation to your class to let them know what to expect in your class this year. Adding some overheads to your presentation will keep their

More information

UW Department of Chemistry Lab Lectures Online

UW Department of Chemistry Lab Lectures Online Introduction to Excel and Computer Manipulation of Data Review Appendix A: Introduction to Statistical Analysis. Focus on the meanings and implications of the calculated values and not on the calculations.

More information

Turns your Wallbox into a Complete Jukebox

Turns your Wallbox into a Complete Jukebox JukeMP3 Wallbox Controller Turns your Wallbox into a Complete Jukebox JukeMP3 Features: 1. The JukeMP3 kit includes everything you need to turn your wallbox into a complete jukebox, except speakers and

More information

Pre-Lab Excel Problem

Pre-Lab Excel Problem Pre-Lab Excel Problem Read and follow the instructions carefully! Below you are given a problem which you are to solve using Excel. If you have not used the Excel spreadsheet a limited tutorial is given

More information

The LabScribe Tutorial

The LabScribe Tutorial The LabScribe Tutorial LabScribe allows data to be accumulated, displayed and analyzed on a computer screen in a format similar to a laboratory strip chart recorder. Equipment Setup 1 Place the iworx unit

More information

Nicolet RAMAN Procedure

Nicolet RAMAN Procedure Nicolet RAMAN Procedure Jerry Hu (Lab Manager), ext. 7914 or 7940 jghu@mrl.ucsb.edu Jaya Nolt (Lab Tech), ext. 4997 jaya@mrl.ucsb.edu Materials Research Laboratory University of California, Santa Barbara

More information

DATA PROCESSING PROCEDURES FOR UCR EPA ENVIRONMENTAL CHAMBER EXPERIMENTS. Appendix B To Quality Assurance Project Plan

DATA PROCESSING PROCEDURES FOR UCR EPA ENVIRONMENTAL CHAMBER EXPERIMENTS. Appendix B To Quality Assurance Project Plan DATA PROCESSING PROCEDURES FOR UCR EPA ENVIRONMENTAL CHAMBER EXPERIMENTS Appendix B To Quality Assurance Project Plan DRAFT Version 1.3 April 25, 2002 William P. L. Carter Atmospheric Processes Group CE-CERT

More information

Installation 3. PerTrac Reporting Studio Overview 4. The Report Design Window Overview 8. Designing the Report (an example) 13

Installation 3. PerTrac Reporting Studio Overview 4. The Report Design Window Overview 8. Designing the Report (an example) 13 Contents Installation 3 PerTrac Reporting Studio Overview 4 The Report Design Window Overview 8 Designing the Report (an example) 13 PerTrac Reporting Studio Charts 14 Chart Editing/Formatting 17 PerTrac

More information

Exercise 5 Animated Excel Charts in PowerPoint

Exercise 5 Animated Excel Charts in PowerPoint Exercise 5 Animated Excel Charts in PowerPoint This is a fun exercise to show you how even graphs can become even more interesting if you use multimedia. You will write some slides on your own to summarize

More information

Intro to Excel. To start a new workbook, click on the Blank workbook icon in the middle of the screen.

Intro to Excel. To start a new workbook, click on the Blank workbook icon in the middle of the screen. Excel is a spreadsheet application that allows for the storing, organizing and manipulation of data that is entered into it. Excel has variety of built in tools that allow users to perform both simple

More information

Reading: Graphing Techniques Revised 1/12/11 GRAPHING TECHNIQUES

Reading: Graphing Techniques Revised 1/12/11 GRAPHING TECHNIQUES GRAPHING TECHNIQUES Mathematical relationships between variables are determined by graphing experimental data. For example, the linear relationship between the concentration and the absorption of dilute

More information

Working with Excel CHAPTER 1

Working with Excel CHAPTER 1 CHAPTER 1 Working with Excel You use Microsoft Excel to create spreadsheets, which are documents that enable you to manipulate numbers and formulas to quickly create powerful mathematical, financial, and

More information

Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET

Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET After reading through the Introduction, Purpose and Principles sections of the lab manual (and skimming through the procedures), answer the following

More information

Spectrum 10 Spectroscopy Software

Spectrum 10 Spectroscopy Software PRODUCT Note Infrared Spectroscopy Spectrum 10 Spectroscopy Software PerkinElmer Spectrum 10 is the newly released 2010 software for the latest range of PerkinElmer FT-IR spectrometers. Designed for busy

More information

N2KExtractor. NMEA 2000 Data Extractor Software. User s Manual

N2KExtractor. NMEA 2000 Data Extractor Software. User s Manual N2KExtractor NMEA 2000 Data Extractor Software User s Manual Revision 1.0.0 Copyright 2012 Maretron, LLP All Rights Reserved Maretron, LLP 9014 N. 23 rd Ave #10 Phoenix, AZ 85021-7850 http://www.maretron.com

More information

Version 1.1. COPYRIGHT 1999 Tufts University and Vernier Software. ISBN (Windows) ISBN (Macintosh)

Version 1.1. COPYRIGHT 1999 Tufts University and Vernier Software. ISBN (Windows) ISBN (Macintosh) Logger Pro Tutorials Version 1.1 COPYRIGHT 1999 Tufts University and Vernier Software ISBN 0-918731-92-5 (Windows) ISBN 0-918731-91-7 (Macintosh) Distributed by Vernier Software 8565 S.W. Beaverton-Hillsdale

More information