P09052 Molecular Imaging System Upgrade
|
|
- Sabina George
- 5 years ago
- Views:
Transcription
1 P0905 Molecular Imaging System Upgrade Table of Contents Conference Paper Calibration Procedure.. 7 Depth Calculation. Calibration Results.. 4 Depth Results.. 5 Recommendations. 7 Drawings. 9
2 Calibration Procedure: Test Setup: Place 0 transparencies taped together (or other suitable non-glossy surface) onto platen, and place top on Imaging System. Housing for grid adjustment should be removed until grid has been focused onto image plane. Excitation filters were not used during calibration procedure. Ensure that the Imaging System and the Optigrid motor control are connected to the Imaging Computer. Phase : Scheimpflug djustments - To bring the Optigrid into focus on the image plane with maximum contrast. Manually adjust the lens linear stage along the optical rail to achieve desired image distance.. For Computar C lens setups, position the Optigrid plane approximately cm from the center of the lens.. Manually adjust the lens and Optigrid fixtures until grid lines can be seen on the platen. 4. Using the linear and rotary micrometers, finely focus the grid onto the platen. a. Using the preview feature of the Kodak MI software can help during focusing of the Optigrid Phase : Optigrid Polarization Procedure - To establish consistent phase movements of Optigrid before calibration begins. Ensure US connection to Optigrid motor control is connected to PC, and power on motor control unit.. Start US Demo pp.6. Polarize the Optigrid by entering 0 in the left Phase Voltage region and 40 in the right Phase Voltage region. 4. Click test start. This will toggle the Optigrid phase between 0 and 40 V. 5. Let this run for 0 minutes. 6. djust Grid Voltage to 0 V by entering '0' into the 'Grid Voltage' dialog box on the Optigrid US Demo pp.6 and clicking 'Set Grid Voltage'. of 0.
3 Figure. US Demo pp.6 Phase : Coarse Calibration - To determine approximate voltage ranges for 0 and 40 degree phase shifts. Set camera to desired settings. Sample settings used during prototype calibration can be seen below.. Minimizing ambient light to the system, take image with specified setting at 0 V of grid voltage.. Export data as an 6 bit TIF with no scaling. File Export Image, Save as Type '6-bit TIFF File-No Scaling(*.tif) Use file name 0.tif 4. Repeat steps and in 5 V increments up to and including 0 V, accumulating 7 images 5. Start MTL and change the current directory to location where images are saved. Save a copy of calcphase.m to this directory. 6. Run calcphasenew.m. The output is a phase versus voltage plot. 7. efore running the code, a portion of the image must be selected that shows high contrast and minimal distortion. The code will rotate the image a set amount and must be adjusted to each set of calibration images. of 0.
4 Figure : Sample Camera Settings for Calibration Procedure 7. From the plot determine center voltages for fine calibration. Select a voltage to remain static during fine calibration. 0 V and 0 V are the recommended.. Record the phase at this voltage. dd 0 degrees to the phase.. Locate the two voltages which the new phase lies between. Calculate the approximate 0 phase location using: 4. dd 40 degrees to the phase at chosen static voltage. Repeat step to determine voltage for 40 degree phase shift. 4 of 0.
5 X: 50 Y: Figure. Sample Phase vs. Voltage plot (Output of Coarse Calibration Phase 4: Fine Calibration. Keeping the same camera settings the same as fine calibration.. Select a range of plus or minus volts from the three selected voltages determined in coarse calibration. a. For example, if 0 V, 68 V and 9 V are determined to be the phases separated by 0 o, voltages of 0 V, 65 7 V and V would be selected.. Create a test plan to showing the sequence of captures. test plan for the above example is shown in Figure Capture an image at each location. It is EXTREMELY important the grid be moved in the following matter: a. First LWYS start the grid at 0 V. Then move to the first voltage (0 V) and capture and image. Label this image V_. Then move to the next voltage (65 V) and capture another image. Label this image V_. Then to the next voltage (88 V). Label this image (V_). Then, the grid must be brought back to 0 V before moving back to the next line (0 V). Label this image (V_) and continue on through the entire plan. The naming convention for the files is: Phase _ Image Set Number.tif Where Phase is either V, V or V. 5.Open of CalcOG.m. Change the number of iteration of the for-loop to the number of image sets. 6. Run the code. This will calculate all of the combined images. 7. Open MMeritMap.m. Change the path names in line 8 and 0 to the location of the images. 5 of 0.
6 Image Set Number Phase Voltage (V) Phase Voltage (V) Phase Voltage (V) Figure 4. Sample Test Plan 6 of 0.
7 8. Change the variables V_coarse and V_coarse to reflect the voltages that were used in the test. For this example, Vcoarse= 65::7 and V_coarse=88::9. 9. Run the code and a merit map will be generated. sample Merit Map is shown in Figure 5 0. The images with the lowest score will have the best phase separation and those voltages should be used when acquiring images Fig. 5 Sample Merit Map. 7 of 0.
8 Depth Calculation: When three phase-shifted images are captured depth information can be extracted from an image. This is accomplished by monitoring each individual points. When a point of an object is captured with three different phase the intensity over those three phases will model that of a sine curve. (see Figure 6). s the source of the intensity moves away from the platen the magnitude of the sine wave decreases. Therefore, depth can be calculated based on the magnitude of the best fit sine wave of the three points. Fitting a sine wave can be difficult so anther correlative property is that of the mean squared error with respect to the mean of the three points. Where m(i,j) is the mean of I,I and I at the location (i,j) Fig 7. High mplitude phase shifted data (Upper Left) nd their corresponding points plotted on a sine curve (Lower Left). Low mplitude phase shifted data (Upper Right) nd their corresponding points plotted on a sine curve (Lower Right). 8 of 0.
9 Calibration Results: Many attempts were made at calibrating the Optigrid. These are the best results that we could attain and the results corresponding to them. The voltages selected were 0 V, V and 7. V. It can be observed that the lines are completely gone in the combined image (Figure #) when compared to the image captured at 0 V (Figure #). Figure # Structured Illumination picture at 0 V 9 of 0.
10 Figure #. Resulting Combined Image. Depth Results: When depth resolution was performed on the following image (Figure #) Figure # resulted. The image capture used minute exposure times. 0 of 0.
11 Figure # Original Image that depth analysis is performed on. Figure #. Depth Estimation of calibration block of 0.
12 Recommendations: ngle djustment: The rail angle adjustment was not tested in the experiments. The steepest angle supported by the system was used for all of the results presented. This helped reduce distortion associated with projection angle. There is a trade off between maximum magnification and maximum angle. s the magnification is increased, the projection angle with respect to the platen is decreased. For the presented prototype, it was deemed more important to have a greater range of magnification. For future testing, it may be beneficial to investigate using a shorter rail with a smaller magnification range. utomation: There are two areas of the presented system whose performance would be greatly enhanced by automation: Optigrid Calibration and Optigrid Warm-Up. Calibration of the Optigrid often requires over 00 images to be taken. Using the current methodology, even with relatively low exposure time, this process is very time consuming and prone to human error. Timing is also very critical to achieving repeatable phase shifts in the Optigrid. utomation of this process would cut down on calibration time and improve on timing consistency. Resulting in more consistent and accurate results. Ensuring the Optigrid is conditioned properly for repeatable results requires the Optigrid to be warmed up and constantly exercised before use. The current motor control allows for the cycling of grid voltage value between values. Modification of this GUI to allow for more voltage values to be included in this cycling would allow for a warm up routine which would better mimic the phase shifting seen during image capture further increasing repeatability of system performance. Integration of the Optigrid motor control GUI into the image capture software proved to be beyond the scope of this project. Establishing communication between these two programs would be the first step in achieving integration, and would help automate the above mentioned processes. Calibration Procedure Modifications: possible way to increase the performance of the Optigrid would be to take four phase separated images at 90o apart rather than three images 0o apart. This could help reduce any of the first harmonic that may reside in the combined image. It also may prove to be more forgiving in that more images would be subtracted possibly canceling out some data that would result from an imperfectly shifted image. dditionally, a fourth point would increase the accuracy of the estimation of the magnitude of the illumination when considering depth. Instead of starting the calibration process at 0 V, starting at 0 V should be considered. The piezo actuator is more linear in the central portions of its region of movement and this shift would move the target voltages more in the linear range. This is important because it decreases the penalty for not having perfectly shifted images. of 0.
13 Mechanical Modifications: dding another clamp to the enclosure would help to keep out more stray light. The current clamp only hold down one side. Putting a shorter optical rail in the box could increase the angle of projection which would cause there to be less distortion. However there would be a trade off in that the range of motion along the rail would be diminished. of 0.
14 4 DRWN Patricia 5//009 CHECKED RIT & Carestream Q TITLE MFG ox Extension Notes: ) Extension Material:.09" Mild Steel ) eroglaze Z06 Flat, diffused black paint (Sherwin Williams UT-J6 black powder coat recommended) 4 PPROVED SIZE SCLE DWG NO oxextension SHEET OF 5 REV 4 of 0.
15 4 Left Extension Right Ledge Flat lank.75 (.000) R (45 ) Right Ledge Formed (.798) (X) R.09 DRWN Patricia 5//009 CHECKED RIT & Carestream Q TITLE MFG ox Extension PPROVED 4 SIZE SCLE DWG NO oxextension SHEET OF 5 REV 5 of 0.
16 4 45 R.50 Right Side Flat lank ase Flat lank ase Formed 0.7 Right Side Formed R (.500) DRWN Patricia 5//009 CHECKED RIT & Carestream Q TITLE 0.75 MFG PPROVED ox Extension 4 R.09 (.500) SIZE SCLE DWG NO oxextension SHEET OF 5 REV 6 of 0.
17 Cover Flat lank.000 Top Flat lank R Top Formed R Cover Formed 7.75 (.548) R.09 (0.76) DRWN Patricia 5//009 CHECKED Q TITLE RIT & Carestream R.09 MFG ox Extension PPROVED SIZE SCLE DWG NO oxextension SHEET 4 OF 5 REV 7 of 0.
18 4 ssembly Weld Joint Top Left Extension Right Ledge Welded Joint to ox Right Side Weld Joint Weld Joint to ox Weld Joint to ox Cover ase Weld Joint DRWN Patricia 5//009 CHECKED RIT & Carestream Q TITLE MFG ox Extension PPROVED 4 SIZE SCLE DWG NO oxextension SHEET 5 OF 5 REV 8 of 0.
19 4 D Drill and Tap for /4-0 SHCS 0.6 [.400].00 [.000] 8.55 [.50] 45. [5.78] 9.70 [7.587] 7.66 [8.569] [.76] [.600] D.00 [.000] 0.6 [.400] [0.000] 5. [.056] (58.57 [.06]) (R6.5 [.50]) (87.64 [.450]) 6.5 [.50]. [.48] (.69 [5.4]) [.746] [4.5] 44.6 [5.680] C (57.5 ) 6.69 [8.5].5 [8.800] (9.87 [9.050]) 7.49 [9.50] C 5.40 [6.000] [.500] (64.97 [6.495]).09 [8.50] (8.44 [8.600]) 6.50 [.500] Drill and Tap for a 0- Screw 05.8 [4.45] 76.0 [.000] 5.40 [6.000] 47. [.855].5 [5.50] 9.05 [.750] 5.40 [6.000] DRWN Patricia 4/8/009 CHECKED Carestream-RIT Q TITLE MFG L-racket Notes: ) Material: ase:.5" luminium Side:.75" luminium 4 PPROVED SIZE C SCLE DWG NO Lracket SHEET OF 9 of 0. REV -
20 4 DRWN Patricia Heneka 5//009 CHECKED RIT & Carestream Q TITLE MFG Phantom lock Notes: ) Material: /" Lexan (Polycarbonate) ) Size: 0 sheets of 5x8 plastic with (4) /4" holes.5" from the edge ) Hardware: (4) /4-0.5" Screw and /4-0 Nut 4 PPROVED SIZE SCLE DWG NO Phantomlock SHEET OF REV 0 of 0.
IR Integration with Vic-Snap and Combining IR Data with Vic-3D. Procedure Guide
IR Integration with Vic-Snap and Combining IR Data with Vic-3D Procedure Guide Introduction This guide lays out the steps for combining 3D DIC data with temperature data from an IR camera. The basic parts
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationMinimizing Noise and Bias in 3D DIC. Correlated Solutions, Inc.
Minimizing Noise and Bias in 3D DIC Correlated Solutions, Inc. Overview Overview of Noise and Bias Digital Image Correlation Background/Tracking Function Minimizing Noise Focus Contrast/Lighting Glare
More informationPrism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis
Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in
More informationQuick Scope. CNC/Manually-operated Vision Measuring Microscope
Bulletin No. 1579 Quick Scope CNC/Manually-operated Vision Measuring Microscope Quick Scope Non-contact Vision Measuring Microscopes Offer High Accuracy, Excellent Affordability and Powerful Capabilities!
More informationAccurately measuring 2D position using a composed moiré grid pattern and DTFT
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 Accurately measuring 2D position using a composed moiré grid pattern and DTFT S. Van
More informationLaserLyte-Flex Red & Green Alignment System
LaserLyte-Flex Red & Green Alignment System LaserLyte-Flex The LaserLyte-Flex is an entry-level laser alignment system. It uses interchangeable optics and a semiconductor laser diode to project a highly
More informationIndustrial Metrology. Multisensor Measuring Machines O-INSPECT 322/442
Industrial Metrology Multisensor Measuring Machines O-INSPECT 322/442 The moment the answer is right in front of your eyes. This is the moment we work for. // Certainty Made By Zeiss 2 Table of Contents
More informationHan PCB termination. Han-Fast Lock PCB adapter for Han DD PCB adapter for Han DDD module
Han termination Contents Page Han-Fast Lock... 20.11 adapter for Han DD... 20.13 adapter for Han DDD module... 20.16 adapter for Han 40 A Axial module... 20.18 adapter for Han E... 20.20 adapter for Han
More informationDIGIPLAN presents. Portable Folding Chair. Unfolded View. Folded Side View. 2004/01/02 rev
WWW..DIGIPLN..C DIGIPLN presents Portable Unfolded View Folded Side View 00/0/0 rev.0 WWW..DIGIPLN..C We are offering these plans for no charge as an example of what you can expect when you purchase other
More informationCh 22 Inspection Technologies
Ch 22 Inspection Technologies Sections: 1. Inspection Metrology 2. Contact vs. Noncontact Inspection Techniques 3. Conventional Measuring and Gaging Techniques 4. Coordinate Measuring Machines 5. Surface
More informationLaboratory 2: 2D Strain Measurement October 12/13, 2005 BIOEN 5201 Introduction to Biomechanics Instructor: Jeff Weiss TA: Trevor Lujan
Laboratory 2: 2D Strain Measurement October 12/13, 2005 BIOEN 5201 Introduction to Biomechanics Instructor: Jeff Weiss TA: Trevor Lujan Lab Quiz: A 10 point lab quiz will be given at the beginning of lab
More informationBear Power Supplies Engineering Department Smart Power Supply Test Bench
The purpose of this document is to validate that the 1.0 Fundamental Overview The smart power supply test bench is an automated test fixture for a power supply and control board assembly. It will fully
More informationUser English Manual for Sputnik Stereo Camera
User English Manual for Sputnik Stereo Camera STEREOSCOPIC SET Stereoscopic photography in contrast to conventional to-dimensional photography makes it possible to obtain pictures which give correct three-dimensional
More informationCalibration of Recorded Digital Counts to Radiance
Graduate Radiometry, 1051-719 Laboratory No. 2 Fall Quarter, 2006 Calibration of Recorded Digital Counts to Radiance INTRODUCTION The goal of this experiment is to calibrate your digital camera, using
More informationComputed tomography (Item No.: P )
Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography
More informationSTEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR. TECHSPEC Optical Cage System
STEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR TECHSPEC Optical Cage System INTRODUCTION 2 What is an autocollimator? The Autocollimator is a precision instrument composed of an achromatic objective,
More informationComputed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms
Related topics Beam hardening, artefacts, and algorithms Principle The CT principle is demonstrated with the aid of simple objects. In the case of very simple targets, only a few images need to be taken
More informationPolarization of Light
Polarization of Light Introduction Light, viewed classically, is a transverse electromagnetic wave. Namely, the underlying oscillation (in this case oscillating electric and magnetic fields) is along directions
More informationQUICK SCOPE. Quick Scope Non-contact Vision Measuring Machine Offer High Accuracy, Excellent Affordability and Powerful Capabilities!
Vision Measuring Systems QUICK SCOPE Catalog No. E4142-359 Quick Scope Non-contact Vision Measuring Machine Offer High Accuracy, Excellent Affordability and Powerful Capabilities! Quick Scope has been
More informationwire guards PLAST D FUSERS FLUORESCENT LIGHTING FIXTURE PROTECTION Wire Guards / Fluorescent lighting fixture protection / Product Description:
Wire Guards FLUORESCENT LIGHTING FIXTURE PROTECTION Wire Guard units offer an inexpensive way to protect light fixtures from excessive abuse. Wire Guard units can be fabricated to fit most shapes of light
More informationTRUCK-MOUNT ARROW BOARDS
WSD-1015 10 2014 TRUCK-MOUNT ARROW BOARDS MODELS WB, WFB, WFP, WLP & WVGB PRODUCT SPECIFICATIONS OCTOBER 2014 Truck- Mount Arrow Boards Page 2 of 15 1. DESCRIPTION 1.1. Description Arrow boards direct
More informationTN425: A study of fluorescence standards confirms that OptiGrid confocal images are suitable for quantitative microscopy
TN425: A study of fluorescence standards confirms that OptiGrid confocal images are suitable for quantitative microscopy Introduction The OptiGrid converts the illumination system of a conventional wide
More informationGP-B Telescope Image Divider Assembly (IDA) Position Left and Right Relay Lenses on IDA Channel B P0294 Rev -
Stanford University W.W. Hansen Experimental Physics Laboratory Gravity Probe B Relativity Mission Stanford, California 94305-4085 SUGP-B P0294 Rev - GP-B Telescope Image Divider Assembly (IDA) Position
More informationPolarization of light
Polarization of light TWO WEIGHTS RECOMENDED READINGS 1) G. King: Vibrations and Waves, Ch.5, pp. 109-11. Wiley, 009. ) E. Hecht: Optics, Ch.4 and Ch.8. Addison Wesley, 00. 3) PASCO Instruction Manual
More information5. Tools. Blue SHARK 123SHARK
5. Tools 5.1 Edit DSD Presets 5.2 Machine Adjustments 5.3 Custom DSDs 5.4 DSD Specific Adjustments 5.4.1 Cut Styles 5.5 Decode 5.6 Special Function 5.7 Security 31 5.1 Edit DSD Presets The Tools section
More informationCARD PRINTER PRINTHEAD REPLACEMENT INSTRUCTIONS
CARD PRINTER PRINTHEAD REPLACEMENT INSTRUCTIONS CAUTION: The discharge of electrostatic energy that accumulates on the surface of the human body or other surfaces can damage or destroy the printhead. Please
More informationLinear Luminaire with LED Series 6012/5
> For use in Zones 1 & 2, 21 & 22 > Latest LED technology with a high luminous efficacy and a long service life > Extra flat version for low mounting depth > Powder-coated sheet-steel or brushed stainless
More informationChief Architect X10 New Feature List
PRODUCTIVITY Saved Plan Views. Create and save multiple plan views (similar to saved cameras). Each view retains Layer Set, Annotation Set and Active Defaults. Open multiple Plan Views in Windows or Tabs,
More informationBOTTOM SIDE ALIGNMENTSOP
BOTTOM SIDE ALIGNMENTSOP October 2013 The wafer is aligned to the mask using the bottom side alignment microscope (BSA). 1. Load mask.please refer to Step 1-3 in the previous SOP for loading a mask. 2.
More informationDynamic Perception Stage Zero Dolly Owners Manual
matthias a. uhlig Dynamic Perception Stage Zero Dolly Owners Manual camera obscura verlag matthias a. uhlig Matthias A. Uhlig Dynamic Perception Stage Zero Dolly Owners Manual Waschow: Camera Obscura Verlag,
More informationCellular Shades MOTORIZED SKYLIGHT. Simplicity with rechargeable batteries. Installation & Care Instructions
Cellular Shades MOTORIZED SKYLIGHT Simplicity with rechargeable batteries Installation & Care Instructions 152741B 7/2/2018 GETTING STARTED A few simple tools are required: - Measuring tape - Power drill,
More informationLPKF Proto-Mat S103 Milling Operation Instructions
LPKF Proto-Mat S103 Milling Operation Instructions Created By Evan Clow ATAMI Oregon State University Created June 20, 2018 Revision Date Description Curator 0 7/24/2018 Initial Document Evan Clow LPKF
More informationLaboratory 2: 2D Strain Measurement October 18/ BIOEN 5201 Introduction to Biomechanics Instructor: Jeff Weiss TA: Heath Henninger
Laboratory 2: 2D Strain Measurement October 18/19 2006 BIOEN 5201 Introduction to Biomechanics Instructor: Jeff Weiss TA: Heath Henninger Lab Quiz: A 10 point lab quiz will be given at the beginning of
More informationIntroduction to the FTA4000
Introduction to the FTA4000 This document will help you get started with your new FTA4000. The FTA4000 is optimized for contact angle measurements using very small droplets on small samples. It uses a
More informationPhysics 1CL WAVE OPTICS: INTERFERENCE AND DIFFRACTION Fall 2009
Introduction An important property of waves is interference. You are familiar with some simple examples of interference of sound waves. This interference effect produces positions having large amplitude
More informationHigh Resolution Indoor Dome Camera
INSTRUCTION MANUAL High Resolution Indoor Dome Camera Model: DBD14 Please read this manual thoroughly before use and keep it handy for future reference. -2- INTRODUCTION Features Model Features: 1/3 Sony
More informationSpring 2011 Workshop ESSENTIALS OF 3D MODELING IN RHINOCEROS February 10 th 2011 S.R. Crown Hall Lower Core Computer Lab
[1] Open Rhinoceros. PART 1 INTRODUCTION [4] Click and hold on the Boundary Lines in where they form a crossing and Drag from TOP RIGHT to BOTTOM LEFT to enable only the PERSPECTIVE VIEW. [2] When the
More informationInstalling the Focus Motor
The MicroTouch Wireless Autofocuser is designed to work with Feathertouch Focusers from Starlight Instruments. It allows automatic focusing with CCD and DSLR cameras. Included is FocusMax software to automatically
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More informationPolarization Microscope. Omano OM349P Series. User Guide
Polarization Microscope Omano OM349P Series User Guide Table of Contents Getting Started... 3 Components... 3 Technical Specifications... 5 Setup Instructions... 6 Basic Operation... 7 Changing the Light
More informationLinear Luminaire with LED Series 6412/5
> For use in Zones 2, 21 & 22 > Latest LED technology with a high luminous efficacy and a long service life > Extra flat version for low mounting depth > Powder-coated sheet-steel or brushed stainless
More informationElecraft W1 SWR/Wattmeter Enclosure by W8FGU
Elecraft W1 SWR/Wattmeter Enclosure by W8FGU The W1 enclosure is made of Lexan, a polycarbonate, which is very strong. It also has a UV blocking coating on one side and was assembled carefully with this
More informationMultisensor Coordinate Measuring Machines ZEISS O-INSPECT
Multisensor Coordinate Measuring Machines ZEISS O-INSPECT Having all the necessary options for reliable measurements. ZEISS O-INSPECT // RELIABILITY MADE BY ZEISS 2 The O-INSPECT multisensor measuring
More informationPRODUCT INFORMATION. Visionary-T 3D SNAPSHOT WIDE RANGE OF APPLICATIONS FOR INDOOR USE. 3D vision
PRODUCT INFORMATION Visionary-T 3D SNAPSHOT WIDE RANGE OF APPLICATIONS FOR INDOOR USE 3D vision FULL FLEXIBILITY FOR INDOOR USE The streaming cameras in the Visionary-T 3D vision product family deliver
More informationSamba Hair System User manual Jefferson Blvd. Culver City, CA 90230, USA Tel: (310) Fax: (310)
Samba Hair System User manual 11922 Jefferson Blvd. Culver City, CA 90230, USA Tel: (310) 577-8110 Fax: (310) 943-3280 http://www.bossanovatech.com Page 1/32 About This Manual This Manual describes SAMBA
More informationDiagonal FOV 18.5mm - 2.7mm ( inch); Video Magnification 30x-210x on 22" Monitor.
HD803-BSM22 HD803 3D HD High Definition 1080p Rotational Microscope with Detent Working Distance. Diagonal FOV 18.5mm - 2.7mm (0.72-0.10 inch); Video Magnification 30x-210x on 22" Monitor. HD 1080P 60
More informationTABLE OF CONTENTS FEATURES. Read manual carefully before using this product! Features 1. Overview 2. Specifications 3. Package Contents 3
NO/LRB-7 TABLE OF CONTENTS Features 1 Overview 2 Specifications 3 Package Contents 3 Installation of batteries 4 Operating instructions 4-5 Storage and maintenance 5 Troubleshooting 6 Special light-powerful
More informationCalibration of a portable interferometer for fiber optic connector endface measurements
Calibration of a portable interferometer for fiber optic connector endface measurements E. Lindmark Ph.D Light Source Reference Mirror Beamsplitter Camera Calibrated parameters Interferometer Interferometer
More informationParts and Service Manual IMCO
XXTREME ADVANTAGE Gimbal Parts and Service Manual IMCO 510 East Arrow Highway San Dimas, CA 91773 (800) 899-8058 (909) 592-6162 Fax (909) 592-6052 www.imcomarine.com email info@imcomarine.com TABLE OF
More informationAP Physics Problems -- Waves and Light
AP Physics Problems -- Waves and Light 1. 1975-4 (Physical Optics) a. Light of a single wavelength is incident on a single slit of width w. (w is a few wavelengths.) Sketch a graph of the intensity as
More informationXL 4000 metal free-standing enclosures and cable compartments
87045 LIMOGES Cedex Telephone : 05 55 06 87 87 Fax : 05 55 06 88 88 XL metal free-standing enclosures and CONTENTS PAGE 1. General characteristics...1 2. Range...2 3. Overall dimensions...2 4. Equipment...3
More informationAgilent 10706B High Stability Plane Mirror Interferometer
7D Agilent 10706B High Stability Plane Mirror Interferometer Description Description The Agilent 10706B High Stability Plane Mirror Interferometer (see Figure 7D-1) is an improved version of the Agilent
More informationDCLM Datasheet. Digitally Controlled Laser Module
DCLM Datasheet Digitally Controlled Laser Module DCLM The DCLM Lyte-MV development kit for Global Laser combines all the functionality of Global Laser MV development kit with the addition of a market leading
More informationThe Anfatec Level AFM a short description. Atomic Force Microscopy - approved devices for affordable prices
The Anfatec Level AFM a short description Atomic Force Microscopy - approved devices for affordable prices Our system is complete for almost all typical applications. It provides all basic modes as: contact
More informationLab 5 Microscopy. Introduction. Carrying the Microscope. Depth of Focus. Parts of the Compound Microscope. Magnification
Lab 5 Microscopy Introduction The microscope is an instrument that contains one or more lenses and is used to view objects that are too small be seen with the unaided eye. A magnifying glass is a simple
More informationChapter 6 : Results and Discussion
Refinement and Verification of the Virginia Tech Doppler Global Velocimeter (DGV) 86 Chapter 6 : Results and Discussion 6.1 Background The tests performed as part of this research were the second attempt
More informationMeasuring object. Fig. 1: Optical scheme of focus-touching (concentricity measurement with measuring ball)
H Focus - Touching At this kind of interferometer the light of the measuring arm of the interferometer can focalized by means of a converging lens directly on a measuring object. This makes it possible
More informationExperiment 9. Law of reflection and refraction of light
Experiment 9. Law of reflection and refraction of light 1. Purpose Invest light passing through two mediums boundary surface in order to understand reflection and refraction of light 2. Principle As shown
More informationAnd. Modal Analysis. Using. VIC-3D-HS, High Speed 3D Digital Image Correlation System. Indian Institute of Technology New Delhi
Full Field Displacement And Strain Measurement And Modal Analysis Using VIC-3D-HS, High Speed 3D Digital Image Correlation System At Indian Institute of Technology New Delhi VIC-3D, 3D Digital Image Correlation
More informationSTEP-BY-STEP INSTRUCTIONS FOR BUILDING A FLUORESCENCE MICROSCOPE. TECHSPEC Optical Cage System
STEP-BY-STEP INSTRUCTIONS FOR BUILDING A FLUORESCENCE MICROSCOPE TECHSPEC Optical Cage System INTRODUCTION 2 What is a Digital Fluorescence Microscope? Unlike traditional microscopes, which utilize an
More informationMAGNIFIERS & TASK LAMPS FOR INDUSTRY
2017 MAGNIFIERS & TASK LAMPS FOR INDUSTRY 1 Daylight Catalogue Professional Magnifiers & Task Lamps 16 great reasons to use Daylight Daylight Simulation Daylight 6,000 K colour temperature for perfect
More informationSetup Information Panosaurus May 3, 2011
Setup Information Panosaurus 2.0 www.gregwired.com May 3, 2011 Please take the time to read all of the setup information to ensure success and ease of use of this tripod head. Much of the setup is a one
More informationSIMULATION AND VISUALIZATION IN THE EDUCATION OF COHERENT OPTICS
SIMULATION AND VISUALIZATION IN THE EDUCATION OF COHERENT OPTICS J. KORNIS, P. PACHER Department of Physics Technical University of Budapest H-1111 Budafoki út 8., Hungary e-mail: kornis@phy.bme.hu, pacher@phy.bme.hu
More informationAstronomy Lab Lenses and Telescopes
Astronomy Lab Lenses and Telescopes OBJECTIVES: Recognize a meter, a centimeter, and a millimeter. Correctly measure distances in mm, cm, and m. Describe the appearance of both a converging lens and a
More informationAppendix E: Software
Appendix E: Software Video Analysis of Motion Analyzing pictures (movies or videos) is a powerful tool for understanding how objects move. Like most forms of data, video is most easily analyzed using a
More informationKCS Motion. Video Motion Analysis Software
Video Motion Analysis Software Software and supporting material is property of G. Mason, Seattle University, 2007 Overview Overview KCS Motion tracks moving objects in a video clip and analyzes their position,
More information45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique
Phase gradient microscopy in thick tissue with oblique back-illumination Tim N Ford, Kengyeh K Chu & Jerome Mertz Supplementary Figure 1: Comparison of added versus subtracted raw OBM images 45 µm polystyrene
More informationTELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing
Edmund Optics BROCHURE TELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing Contact us for a Stock or Custom Quote Today! UK: +44 (0) 1904
More informationTELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing
Edmund Optics BROCHURE TELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing Contact us for a Stock or Custom Quote Today! USA: +1-856-547-3488
More informationQuality control for the foil industry -the modular solutions-
Quality control for the foil industry -the modular solutions- The observance of layer thicknesses or flawless layer contacts partly determine the quality of films in all application areas. The fast and
More informationAdvanced Stamping Manufacturing Engineering, Auburn Hills, MI
RECENT DEVELOPMENT FOR SURFACE DISTORTION MEASUREMENT L.X. Yang 1, C.Q. Du 2 and F. L. Cheng 2 1 Dep. of Mechanical Engineering, Oakland University, Rochester, MI 2 DaimlerChrysler Corporation, Advanced
More informationKMIC. AV300+ AV350+ AV300+ Micro
KMIC EZseries MVRseries AVseries AVRseries + + + Micro Vision Systems LFseries av150 FPO Starrett Vision Systems Rugged, Accurate & easy to use The Starrett automatic vision (or video-based) metrology
More informationFlat-Field Mega-Pixel Lens Series
Flat-Field Mega-Pixel Lens Series Flat-Field Mega-Pixel Lens Flat-Field Mega-Pixel Lens 205.ver.0 E Specifications and Lineup Full Full Full Full 5MP MP MP MP Image Model Imager Size Mount Focal Length
More informationGHI Systems Rotary Vibration Machine Manual
GHI Systems Rotary Vibration Machine Manual Version 1.0 Release Date January 2002 Copyright 2002 By GHI Systems, Incorporated 916 N. Western Ave. Suite 201 San Pedro, California No part of this publication
More informationLIGHT CLASS X STUDY MATERIAL & QUESTION BANK:
LIGHT CLASS X STUDY MATERIAL & QUESTION BANK: 1. REFLECTION OF LIGHT: The phenomenon of light coming back into the same medium after it strikes a smooth surface is called reflection. 2. LAWS OF REFLECTION:
More informationDV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure
H I G H - P E R F O R M A N C E E M C C D & C C D C A M E R A S F O R L I F E S C I E N C E S DV2 This document provides a straightforward, step-by-step outline of the alignment procedure for the Photometrics
More informationAir curtains for enclosures with 2 or 3 windows including flow regulation valve, pneumatic fitting for 4mm tube.
autovimation offers various solutions to prevent dust and dirt on the front window of the camera enclosure. So the cameras can be used even under unfavourable conditions. Air curtains for enclosures with
More informationLab 4 Projectile Motion
b Lab 4 Projectile Motion What You Need To Know: x = x v = v v o ox = v + v ox ox + at 1 t + at + a x FIGURE 1 Linear Motion Equations The Physics So far in lab you ve dealt with an object moving horizontally
More informationLENS INSTALLATION GUIDE
Paladin Lens Phoenix Lens LENS INSTALLATION GUIDE The Paladin and Phoenix lens systems represent the latest generation in patented anamorphic optical solutions to engage the anamorphic modes available
More informationZeiss AxioImager.Z2 Fluorescence Protocol
Zeiss AxioImager.Z2 Fluorescence Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationBiology 105 Laboratory 1: Microscope
Biology 105 Laboratory 1: Microscope A. The Compound Microscope The microscope is a delicate, precision instrument. It must be treated gently, as the slightest bump or jar may damage the alignment of its
More informationUser Manual CS-SA Series 'SUPER ANALOG' Time Fade Stations JOHNSON SYSTEMS INC.
User Manual CS-SA Series 'SUPER ANALOG' Time Fade Stations JOHNS SYSTEMS INC. Contents Warranty...2 User Manual Copies...2 Introduction...3 Features...3 CS-SA Series Control Station Diagrams...4, DIM and
More informationGuided Problem Solving
-1 Guided Problem Solving GPS Student Page 57, Exercises 1 1: Match each rule with the correct translation. A. (x, y) (x, y 1 ) I. P(, 1) P (3, ) B. (x, y) (x 1 3, y) II. Q(3, 0) Q (3, ) C. (x, y) (x 1,
More informationDraft SPOTS Standard Part III (7)
SPOTS Good Practice Guide to Electronic Speckle Pattern Interferometry for Displacement / Strain Analysis Draft SPOTS Standard Part III (7) CALIBRATION AND ASSESSMENT OF OPTICAL STRAIN MEASUREMENTS Good
More informationHARP-NEF Front End Assembly
Science Fibers HARP-NEF Front End Assembly Calib Fibers HARPN Front End Assembly ( w/ Partial Cover Plates ) HARPN Front End Assembly 1 HARP-NEF Front End Assembly Roger Eng December 6, 2007 2 HARP-NEF
More informationSTEP-BY-STEP INSTRUCTIONS FOR BUILDING A MICHELSON INTERFEROMETER. TECHSPEC Optical Cage System
STEP-BY-STEP INSTRUCTIONS FOR BUILDING A MICHELSON INTERFEROMETER TECHSPEC Optical Cage System INTRODUCTION 2 What is a Michelson Interferometer? A Michelson Interferometer is a simple interferometric
More informationStructured light 3D reconstruction
Structured light 3D reconstruction Reconstruction pipeline and industrial applications rodola@dsi.unive.it 11/05/2010 3D Reconstruction 3D reconstruction is the process of capturing the shape and appearance
More informationActivity P34: Polarization - Verify Malus Law (Light Sensor)
Activity P34: Polarization - Verify Malus Law (Light Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Polarization P34 Malus Law.DS (See end of activity) (See end of activity) Equipment
More information>> NX. Metaform Boundary Conditions. Go to Table of Contents
Metaform Boundary Conditions In this article, we will discuss the Boundary Condition Constraint options used in the building of a Metaform feature and conclude with some Tips and Tricks that may further
More informationChapter 26 Geometrical Optics
Chapter 26 Geometrical Optics 26.1 The Reflection of Light 26.2 Forming Images With a Plane Mirror 26.3 Spherical Mirrors 26.4 Ray Tracing and the Mirror Equation 26.5 The Refraction of Light 26.6 Ray
More informationHDAF800 HD 1080p Digital Microscope USB drive for Image Capture
HDAF800 HD 1080p Digital Microscope USB drive for Image Capture HDAF800-US Auto-Focus HD Camera System 1080p including Universal Stand with 15" Vertical Pole and integrated LED lighting and controls panels.
More informationDiscover the new world of handheld digital microscopy
Discover the new world of handheld digital microscopy Designed by Euromex in the Netherlands Q-scope digital microscopes high quality real time images The Q-scope product range of digital microscopes offer
More informationVisionGauge OnLine Spec Sheet
VisionGauge OnLine Spec Sheet VISIONx INC. www.visionxinc.com Powerful & Easy to Use Intuitive Interface VisionGauge OnLine is a powerful and easy-to-use machine vision software for automated in-process
More informationSource Four CE Mini ETC. Source Four Mini Series GENERAL INFORMATION ORDERING INFORMATION
GENERAL INFORMATION The Source Four CE Mini can go where no full-sized Source Four has gone before. ETC s 12V 50 watt Source Four Mini features the same crisp optics, user-friendly functionality and sleek
More informationOptilia Instruments. Empowering Your Vision!
Optilia Instruments Empowering Your Vision! Product Review: Optilia BGA Inspection Systems Cutting edge technology in optical inspection of BGA, µbga, CSP and FlipChip soldering! RevB, November-2013 Optical
More informationMMA845xQ Sensor Toolbox User s Guide
Freescale Semiconductor Document Number: MMA845xQSTUG User s Guide Rev. 1, 02/2012 MMA845xQ Sensor Toolbox User s Guide 1 Introduction The Freescale MMA845xQ sensor toolbox accelerometer kit provides hardware
More informationLC-1: Interference and Diffraction
Your TA will use this sheet to score your lab. It is to be turned in at the end of lab. You must use complete sentences and clearly explain your reasoning to receive full credit. The lab setup has been
More informationDSU Start-Up instructions
DSU Start-Up instructions Always: - start with the 10x objective - properly center the stage around the current objective before changing to another objective - when done, leave the 10x objective in standby
More informationSource Four CE Mini ETC. Source Four Mini Series GENERAL INFORMATION ORDERING INFORMATION. Source Four CE Mini Luminaire (Body only)
GENERAL INFORMATION The tungsten luminaire can go where no fullsized Source Four has gone before. ETC s 12V 50 watt Source Four Mini features the same crisp optics, user-friendly functionality and sleek
More information