Nitrite photometer. mono. Operating Instructions

Size: px
Start display at page:

Download "Nitrite photometer. mono. Operating Instructions"

Transcription

1 Nitrite photometer mono Operating Instructions

2 Contents 1. Important Safety Information 1.1 Notes and symbols used in these instructions 1.2 Intended use 1.3 Staff qualifications 1.4 Warning notices in these instructions 1.5 Further documents 2. Pay Particular Attention to 2.1 General instructions 2.2 Commissioning 2.3 Operation 2.4 Cleaning 2.5 Disposal 3. Scope of Delivery 3.1 Optional accessories 4. Performance Specifications 4.1 Performance features 5. Instrument View 5.1 Position of the operating element 6. Description of Display and Operating Elements 7. Commissioning 7.1 General instructions 7.2 Settings of the measuring instrument 8. Advance Functions 9. Calibrating 9.1 Inserting the test cell and calibrating 9.2 Calibration error "Err1" 10. Measuring 10.1 Inserting the test cell and measuring 10.2 Possible measuring errors and their remedy 11. Overview of Available Parameters 12. Software Installation and Description 12.1 Installing the software 12.2 IR interface connection 12.3 Starting the software 12.4 Description of functions 13. Replacing the Battery 14. Technical Data 2 / 13

3 1. Important Safety Information 1.1 Notes and symbols used in these instructions K Operator instructions Always observe / warning Please read the operating instructions carefully and thoroughly prior to working with the instrument. These operating instructions must be passed on to the new owner should ionoval mono change hands. Ensure that the operating instructions are always available to all users. Always adhere to hazard warnings and safety tips when using reagents, chemicals and cleaning agents. Please refer to the respective safety data sheet! Download safety data sheets for the supplied reagents at Non-compliance with these instructions can result in personal injuries and/or damage to the instrument and a loss of warranty rights. 1.2 Intended use ionoval mono is a photometer for determining the concentration of water contents (see table under point 11). Always adhere to the limits specified in chapter Technical Data. To ensure correct and intended usage, the instructions, especially the Important Safety Information, have to be read and understood. The instrument is not used as intended if it is used in application areas not stated in these instructions. it is used in conditions which deviate from the ones described in these instructions. K Trouble-free operation of the ionoval mono instrument is only guaranteed when using HEYL reagents! Faults or problems resulting from the use of other reagents will not be accepted as a warranty or fair ealing claim! (Comment: Other reagents can only be calibrated and used with ionoval FM-A!) 1.3 Staff qualifications Only fully qualified staff or authorised individuals supervised by a specialist are allowed to operate the instrument or use chemicals. 1.4 Warning notices in these instructions The warning notices in these instructions warn the user about potential dangers to individuals and property resulting from incorrect handling of the instrument. Warning notices are structured as follows: Description of type and source of the danger Description of the consequences resulting from non-observance Preventive measures. Always follow these preventive measures. 1.5 Further documents 3 / 13

4 Please also observe the operating instructions and handling instructions of the chemicals being used. 2. Pay Particular Attention to 2.1 General instructions Adhere to national health and safety regulations as well as environmental protection regulations in the country of use. Do not modify or tamper with the instrument. Do not carry out any actions at the instrument which are not described in these instructions; failure to adhere to these instructions will negatively affect any warranty claims that you make thereafter. 2.2 Commissioning Remove the switch-on guard (white plastic part at the power supply connection) 2.3 Operation Please remove the battery from the instrument if it is not going to be used for a longer period (several months). You can switch the battery off by inserting the switch-on guard if the instrument is not going to be used for a short period. 2.4 Cleaning Only use a dry, lint-free cloth for cleaning. 2.5 Disposal Dispose of the old batteries and the instrument in accordance with national regulations. 3. Scope of Delivery 1 portable lab 1 ionoval mono measuring instrument (incl. battery) 2 test cells 1 operating instructions 3.1 Optional accessories Optional IR interface with corresponding software Power supply unit Instrument case, power supply unit, IR interface, test cells and reagents 4 / 13

5 4. Performance Specifications ionoval mono is a photometer for determining the concentration of water contents (see table under point 11). The microprocessor-controlled instrument functions photometrically (LED) and is primarily used in the fields of water treatment, sewage water, swimming pools and drinking water. The highly integrated electronics, simple operation and large digital display enable quick and precise analytical determination. The basic version is powered via a 9 V battery. The last 100 measuring results can be stored with a consecutive number, the measuring point, parameters and the time and date. 4.1 Performance features Large, easy-to-read display Simple menu structure Save up to 100 measuring values in a ring buffer (can be switched off) (measurement no., measuring point, parameters, measuring value, time and date) Delete saved measurements View saved data directly at the instrument Read saved data via the IR interface 1 Read and adjust the settings (store, beep, timer) via IR interface 1 Real-time clock (set the time and date directly at the instrument) Transfer the PC system time and date via IR interface 1 Output of a beep when pressing keys (can be switched off) Timer with time display (countdown) and acoustic signal output once the reaction time of the reagent has expired (can be switched off) Display in ppm or mg/l Measuring point (can be set from 1-9) Battery voltage display Energy saving mode (instrument switches off automatically after 5 minutes) Measuring range (see table under point 11) Operation with power supply unit (9V DC), (optionally available) 1 only with optionally available IR interface 5. Instrument View 5.1 Position of the operating elements and connections 5 / 13

6 6. Description of Display and Operating Elements Digit Meaning 1,2,3 Not used 4 Not used Battery display 5 Flashing signal indicates the battery must be replaced 6 Not used Multi-functional display 2 7 Displays time, measuring point (displays saved measuring results 8 Not used Timer display 9 Displays that the timer is switched on PC mode 10 Displayed if menu item PC mode is selected Displayed if PC mode is active Display Store (buffer) 11 Displays that measurements are being saved 12 Not used Display CAL 13 Display flashes once the instrument has been switched on and requests calibration Displayed if instrument is in calibration mode 14 Displays the unit 15 Displays the unit Multi-functional display 1 16 Displays measuring results, parameters, date (displays saved measuring results), CAL (during calibration), DATA (if PC mode is active) and error messages (Err 1-5) 6 / 13

7 7. Commissioning After removing the switch-on guard (white plastic part at the power supply connection), the instrument is switched on in battery mode. The instrument is always in measuring mode after switching it on. The instrument will switch off automatically (energy saving mode) (except when in PC mode) if it is not operated or no key is pressed for 5 minutes. If a power supply unit is used, the inserted battery is switched off. Note: If the instrument is operated without an inserted battery but with a power supply unit and is disconnected from the mains, the instrument will be in exactly the same mode as before disconnection once the power is reconnected. If the instrument is in sleep mode (automatic switch off), press any key to switch it on again. 7.1 General instructions Only operate the instrument in the ambient conditions (e.g. temperature) specified in the chapter entitled Technical Data. Always protect the instrument against moisture. It should never come into contact with condensation or splash water. The instrument should only be used for its specified intended purpose. 7.2 Settings of the measuring instrument It is recommended to carry out instrument settings during initial commissioning. These settings are saved and can be changed at any time if required. They are divided into the following settings and advanced functions. 7 / 13

8 8. Advanced Functions 8 / 13

9 9. Calibration Every time the instrument is switched on the CAL symbol on the display flashes as a reminder that calibration must be carried out. 9.1 Inserting the test cell and calibrating Take a test cell from the portable lab, rinse it with the water sample (without reagents) and fill it up to the mark. When using the test cell please observe the following: - The test cell should be clean and dry - When inserting the test cell the gap in the mark must face towards the display 9.2 Calibration error "Err1" If the error message "Err1" is displayed, a calibration error has occurred (e.g. test cell is soiled). The instrument remains in calibration mode and after reinserting the test cell the calibration process can be continued by pressing the START key. 10. Measuring 10.1 Inserting the test cell and measuring After calibration (see 9.1) add 2 dosing spoons full of nitrite reagent and stirr with the red spoon until everything has dissolved. When inserting the test cell into the instrument, ensure that the mark (gap) of the test cell faces towards the display to achieve consistent measuring results. Insert the test cell into the test cell slot and press the START key to commence measuring. After pressing the START key and if the timer has been set, the measured value is automatically displayed once the reaction time (15 minutes) has expired. The display indicates that the reaction time has expired. Press the SET key to interrupt the timer and to cancel the measurement. If the store has been activated, the current storage location number is displayed for each measurement. If the timer has been switched off, a reaction time of 15 minutes has to be adhered to. Once the reaction time has expired, press the START key to trigger measuring. 9 / 13

10 10.2 Possible measuring errors and their remedy If Err2 appears, the reagent being measured is "lighter" than the liquid used for calibration. Please check whether calibration has been carried out with the correct liquid; if necessary, recalibrate and subsequently repeat the measurement. If Err2 is still displayed, check your reagent and, if necessary, renew it. Subsequently carry out a new measurement. If Err4 appears after the measurement, the sample being measured is too dark. The measuring range has been exceeded. In this case, measure the water sample after previous dilution. To achieve this, multiply the measured value displayed by the instrument with the dilution factor. 11. Overview of Available Parameters Parameter Measuring range Instrument article no. Reagent article no. Aluminium 0-1 mg/l Ammonium 0-1 mg/l Bromine 0 5 mg/l Chlorine dioxide 0,01 1 mg/l Chloride mg/ Chromate 0-1 mg/l Iron L 0-1 mg/l Iron H 0-10 mg/l Hydrazine 0-1 mg/l Copper 0-2 mg/l Manganese L 0-0,25 mg/l Manganese H 0-20 mg/l Nitrate 0-10 mg/l Nitrite 0-1 mg/l Ozone 0-1,5 mg/l Phosphate 0-10 mg/l Silicate 0-2 mg/l Sulphite 0-10 mg/l If you require new reagents, please order the corresponding parameter refill packs under the specified reagent article number! 10 / 13

11 12. Software Installation and Description 12.1 Installing the software Please insert the installation CD into your CD-ROM drive to install the data transfer software ionodata. If the installation program does not start automatically, click on Start and then on Run. Enter d:\start.exe and click on OK. The installation wizard will guide you through the installation process IR interface connection Please connect the IR interface to a free serial port (COM port) of your computer Starting the software Before starting the software, activate PC mode at ionoval mono and align the IR interface to the small red window of the ionoval mono. Click on Start, then on Programs and subsequently on ionodata to start the program. The following window should appear: If the IR interface has not been correctly aligned during program start-up or is more than 1 metre away from the instrument, the following message is displayed: 11 / 13

12 After confirming this message the following window appears: Please check your serial connection and, if necessary, set the respective COM port. Click on Test to establish the correct connection to ionoval mono Description of functions COM1: Click on COM1 if your IR interface is connected to the first serial port. COM2: Click on COM2 if your IR interface is connected to the second serial port. COM3: Click on COM3 if your IR interface is connected to the third serial port. The window above the Test button can output the following messages: OK: Data transfer was successful. DATA: Data transfer is active. ERROR (1-3): An error has occurred. Detailed description under Help. Test: Click on Test to check the data transfer. Settings Store on/off: Activate/Deactivate the ionoval mono store. Timer on/off: Activate/Deactivate the ionoval mono timer. Beep on/off: Activate/Deactivate acoustic signal when pressing ionoval mono keys. SETUP: Click on SETUP to transfer settings to ionoval mono. Set clock: Click on Set clock to transfer the system time and date from the PC to ionoval mono. Retrieve data: Click on Retrieve data to transfer saved measuring data from ionoval mono to the PC. It is possible to enter a file name for the measurements to be saved in the window above the Save data button. Save data: Click on Save data to save data under your entered file name in the installation directory, subfolder ionoval measuring data. View data: Click on View data to open the saved file with the program displayed in the window below the View data button. The Notepad editor has been preset. A different editor can be selected by replacing the Notepad entry in the file ionodata.ini in the installation directory. The Parameters window displays the parameter of your ionoval mono. Exit: Click on Exit to close the application. 12 / 13

13 13. Replacing the Battery Please loosen the two screws on the base of the instrument to change the battery. Insert the new battery (9V alkaline), always observe correct battery polarity. Subsequently reassemble the housing. Dispose of the old battery in accordance with national regulations. 14. Technical Data Voltage supply : 9V DC battery Power supply unit : 9V DC, optional Measuring range : mg/l ozone Display : 4 digits Measured value store : 100 measured values Real-time clock : yes Timer for reaction time : yes Energy saving mode : yes IR interface : optional Software for interface : optional Ambient temperature : 0-45 C Weight : Approx. 0.9 kg or 1.25 kg (with optional accessories) We reserve the right to make technical changes without notice in the interest of constantly improving our products! ionoval_mono_nitrit_gb100506_c1.doc 13 / 13

Process Photometer for total chlorine ppm with cleaning function

Process Photometer for total chlorine ppm with cleaning function Lenntech Operating Instructions Testomat 2000 CLT self clean Process Photometer for total chlorine 0 2.5 ppm with cleaning function Lenntech info@lenntech.com Tel. +31-152-610-900 www.lenntech.com Fax.

More information

DS400 / DS404. Software Installation and Operations Manual. DS400 Docking Station. (800) (734)

DS400 / DS404. Software Installation and Operations Manual. DS400 Docking Station. (800) (734) DS400 / DS404 33 Docking Stations Software Installation and Operations Manual DS400 Docking Station DS404 Multi-inlet Docking Station 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734)

More information

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230 USER MANUAL VIBRATION CONTROL Version 2.2 1 IMPORTANT NOTES Electrical danger within the meaning of this documentation or the warning labels on the product itself respectively means that death, serious

More information

B63/ NS MS. EtherNet/IP LINK

B63/ NS MS. EtherNet/IP LINK 3 609 929 B63/ IMenip 2008-09 NS MS EtherNet/IP LINK 3 609 929 B63/2008-09 IMenip Bosch Rexroth AG 15/76 Table of Contents About this document................. 16 General safety instructions............

More information

testo 175 Data loggers Instruction manual

testo 175 Data loggers Instruction manual testo 175 Data loggers Instruction manual 2 1 Contents 1 Contents 1 Contents 3 2 Safety and the environment 4 2.1. About this document 4 2.2. Ensure safety 5 2.3. Protecting the environment 5 3 Specifications

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

Dialock DT Lite. Operating instructions...2. Operating instructions

Dialock DT Lite. Operating instructions...2. Operating instructions Operating instructions Dialock DT Lite Operating instructions... Operating instructions. Notes on these operating instructions. Contents and target group of the instructions It is essential to pay attention

More information

DCS200/DCS200-09/DCS DCS300/DCS300-09/DCS355

DCS200/DCS200-09/DCS DCS300/DCS300-09/DCS355 THE SEEKER 200 SERIES & THE SEEKER 300 SERIES VIDEO INSPECTION SYSTEMS DCS200 (shown) DCS300 (shown) DCS200/DCS200-09/DCS200-05 DCS300/DCS300-09/DCS355 USER S MANUAL Please read this manual carefully and

More information

DS404. Software Installation and Operations Manual. Multi-Inlet Docking Station. (800) (734)

DS404. Software Installation and Operations Manual. Multi-Inlet Docking Station. (800) (734) DS404 Multi-Inlet Docking Station Software Installation and Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com 1 Table of Contents INTRODUCTION.

More information

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 100. T-Logg 100 E

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 100. T-Logg 100 E E39.0.0X.6C-03 Data logger for temperature as of version V1.3 Operating Manual T-Logg 100 T-Logg 100 T-Logg 100 E GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 +49 (0) 9402 / 9383-0

More information

Always there to help you. Register your product and get support at AJ6200D/98. Question? Contact Philips.

Always there to help you. Register your product and get support at  AJ6200D/98. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips AJ6200D/98 User manual Contents 1 Important 3 2 Your clock radio 3 What's in the box

More information

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016 Operating instructions AS-i SmartLine module AC3200 AC3201 80237876/00 06/2016 Contents 1 Preliminary note...3 1.1 Symbols used...3 1.2 Warnings used...3 2 Safety instructions...3 2.1 General...3 2.2 Target

More information

Instruction manual of the controller unit. PCS ph-redox-chlorine-temp PCS ph-oxygen-temp PCS ph-redox-temp

Instruction manual of the controller unit. PCS ph-redox-chlorine-temp PCS ph-oxygen-temp PCS ph-redox-temp Instruction manual of the controller unit PCS ph-redox-chlorine-temp PCS ph-oxygen-temp PCS ph-redox-temp SPIS TREŚCI GENERAL INFORMATIONS... 3 Main Functions... 3 Unit Specification... 3 Unit Dimensions......

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

IPCB42501 / IPCB42551

IPCB42501 / IPCB42551 IPCB42501 / IPCB42551 Installation manual Version 03/2017 IPCB42501 IPCB42551 English translation of the original German user manual. Retain for future reference. uk Introduction Dear Customer, Thank you

More information

Installation manual plugs and connectors with screw connection (16/32 A)

Installation manual plugs and connectors with screw connection (16/32 A) EN Installation manual plugs and connectors with screw connection (16/32 60003213 Issue 04.2016 2016-04-01 Table of contents 1 About this manual 3 1.1 Structure of the warnings 3 1.2 Symbols used 4 1.3

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Electronic Rotational Speed Monitoring Device Type EDO Device Identification No.: 93.045 508.001 /.005 /.007 /.012 OPERATING INSTRUCTIONS 2 CE-Sign an Conformity The device meets the requirements of the

More information

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 E39.0.31.6C-02 Data logger for humidity temperature as of version V1.0 Operating Manual T-Logg 160 GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 +49 (0) 9402 / 9383-0 +49 (0) 9402

More information

315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors

315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors 315 /Ion Installation Input Connectors socket (automatic temperature compensation) socket data socket reference socket If you ordered an electrode with the 315, connect it as follows: Corning Incorporated

More information

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02 SP-400 Portable Water Analyzer Operation Manual Rev. A Firmware version 1.02 Pyxis Lab, Inc 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1 Contents 1 General Description...

More information

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines Operating manual Conductivity measuring cell to be installed in pipelines ba55321de02 10/2010 Copyright Weilheim 2010, WTW GmbH Reprinting - even as excerpts - is only allowed with the explicit written

More information

POCKET MULTIMETER Model No: MM18

POCKET MULTIMETER Model No: MM18 INSTRUCTIONS FOR: POCKET MULTIMETER Model No: MM18 Thank you for purchasing a Sealey product. Manufactured to a high standard this product will, if used according to these instructions and properly maintained,

More information

CD DAB+ RADIO WITH BLUETOOTH MUSIC SYSTEM BT-9000

CD DAB+ RADIO WITH BLUETOOTH MUSIC SYSTEM BT-9000 CD DAB+ RADIO WITH BLUETOOTH MUSIC SYSTEM BT-9000 Instruction Manual For information and support www.lenco.eu Important Safety Instructions When used in the directed manner, this unit has been designed

More information

Digital UVC Light Meter

Digital UVC Light Meter UV512C USER S MANUAL Digital UVC Light Meter Thank you for purchasing this Digital UVC Light Meter. Read through this instruction manual before operating the unit. Please also store and retain this instruction

More information

Palintest PHOTOMETER 7500 INSTRUMENTS. INST.7500.C Version 3-01/09

Palintest PHOTOMETER 7500 INSTRUMENTS. INST.7500.C Version 3-01/09 Palintest INSTRUMENTS PHOTOMETER 7500 INST.7500.C Version 3-01/09 Your Palintest direct-reading photometer is designed to give long and trouble-free operation. To ensure you get the best out of this photometer,

More information

Digital Quick-Response Thermometer

Digital Quick-Response Thermometer H56.0.01.6C-06 Digital Quick-Response Thermometer Operating Manual GTH 1170 WEEE-Reg.-Nr. DE93889386 GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D-93128 Regenstauf +49 (0) 9402 / 9383-0

More information

Manual. Portable Smart Soldering Station TS100 12/24V computer-controlled, programmable, adjustable temperature. No Copyright Graupner/SJ GmbH

Manual. Portable Smart Soldering Station TS100 12/24V computer-controlled, programmable, adjustable temperature. No Copyright Graupner/SJ GmbH EN Manual Portable Smart Soldering Station TS100 12/24V computer-controlled, programmable, adjustable temperature No. 94412 Copyright Graupner/SJ GmbH 2 / 2 S8474_PDB_SBEC_V_1_jh Index Introduction...4

More information

RS Stock No Instruction Manual RS Input Data Logging Thermometer

RS Stock No Instruction Manual RS Input Data Logging Thermometer RS Stock No. 730-0458 Instruction Manual RS-1384 4 Input Data Logging Thermometer EN FR IT DE ES TABLE OF CONTENTS / EN TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION FEATURE... 1 2. SPECIFICATIONS... 2

More information

PORTABLE SPEAKER & AMPLIFIER USER MANUAL

PORTABLE SPEAKER & AMPLIFIER USER MANUAL PORTABLE SPEAKER & AMPLIFIER USER MANUAL TABLE OF CONTENTS Safety Instructions Important Safety Instructions...2 USB Charging USB Charging...14 Getting to Know the IGNT-2 What s in the Box...4 Getting

More information

Always there to help you Register your product and get support at SBT300/12. Question? Contact Philips.

Always there to help you Register your product and get support at   SBT300/12. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips SBT300/12 User manual Contents 1 Important 3 Safety 3 Notice 3 2 Your Bluetooth Speaker

More information

7. Cleaning and maintenance Disposal What if there are problems? Technical data... 21

7. Cleaning and maintenance Disposal What if there are problems? Technical data... 21 WL 32 D Lichtwecker Gebrauchsanweisung...2 12 G Wake up light Instructions for use... 13 22 F Réveil lumineux Mode d emploi... 23 33 E Luz para despertar Manual de instrucciones... 34 43 I Luce sveglia

More information

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -...

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -... E39.0.1X.6C-01 Data logger for standard signals as of version 1.0 Operating Instruction T-Logg 120... T-Logg 120 W -... T-Logg 120 K -... GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße

More information

Always there to help you. Register your product and get support at AJ5300D. Question? Contact Philips.

Always there to help you. Register your product and get support at   AJ5300D. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips AJ5300D User manual Contents 1 Important 2 Safety 2 English 2 Your Docking Entertainment

More information

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter Instruction Sheet PN 51A-6081C/rev.A June 2009 Model 6081-C Wireless Contacting Conductivity Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures,

More information

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no PV Remote Unit A 1378 Instruction manual Version 2.1, Code no. 20 752 010 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si

More information

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS

More information

Operating instructions: STABILA Electronic-IP65

Operating instructions: STABILA Electronic-IP65 TECH Manual Operating instructions: STABILA 196-2 Electronic-IP65 ON/OFF Illumination MODE Hold Key lock sound calibration REF TECH 196E Manual Revision IP65 Code: International Protection Rating. 6: Dust

More information

WELCOME TO YOUR NEW exact idip. Let us show you around. CONTENTS WELCOME

WELCOME TO YOUR NEW exact idip. Let us show you around. CONTENTS WELCOME QUiCK start guide 2 CONTENTS Welcome...2 idip Overview...3 Let s Get Started Install Batteries...4 Download App...4 Home Screen...5 Test Result Screen...5 Select, Fill, Dip, Read Select Customer...6 Turn

More information

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating manual PSA06

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating manual PSA06 PKP Prozessmesstechnik GmbH Borsigstrasse 24 D-65205 Wiesbaden-Nordenstadt Tel: 06122 / 7055-0 Fax: 06122 / 7055 50 Operating manual PSA06 Electronical pressure switch Content Page 19-34 1. Foreword 19

More information

PX Series Balances. Quick Start Guide. Please download the user manual from

PX Series Balances. Quick Start Guide. Please download the user manual from Scan QR Code with your WebCam for downloading the user instruction manual. PX Series Balances Quick Start Guide Please download the user manual from www.ohaus.com. EN-1 1. INSTALLATION 1.1 Select the

More information

Always there to help you. Register your product and get support at AJ6200DBZ. Question? Contact Philips.

Always there to help you. Register your product and get support at   AJ6200DBZ. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/support Question? Contact Philips AJ6200DBZ User manual Contents 1 Important 2 2 Your clock radio 3 Introduction 3 What's

More information

Operation Manual IMR Environmental Equipment, Inc Central Ave. St. Petersburg, FL Phone: E-mai:

Operation Manual IMR Environmental Equipment, Inc Central Ave. St. Petersburg, FL Phone: E-mai: Portable Multi-Gas Detector Operation Manual 3634 Central Ave. St. Petersburg, FL 33711 Phone: 727-328-2818 E-mai: info@imrusa.com SAFETY INFORMATION Incorrect operation or unsuitable using environment

More information

GMH Operating Manual for Digital Quick-Response Thermometer

GMH Operating Manual for Digital Quick-Response Thermometer H58.0.01.6C-02 Operating Manual for Digital Quick-Response Thermometer GMH 1170 WEEE-Reg.-Nr. DE93889386 GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D-93128 Regenstauf +49 (0) 9402 / 9383-0

More information

Always there to help you. Register your product and get support at AJ3275D. Question? Contact Philips.

Always there to help you. Register your product and get support at   AJ3275D. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips AJ3275D User Manual Contents 1 Important 3 Risk of swallowing batteries 3 2 Your clock

More information

Operating instructions Optical distance sensor. O5D15x / / 2014

Operating instructions Optical distance sensor. O5D15x / / 2014 Operating instructions Optical distance sensor O5D15x 706307 / 01 08 / 2014 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...3 3 Functions and features...4 3.1 Applications...4

More information

INSTRUCTION MANUAL CAT ecount

INSTRUCTION MANUAL CAT ecount Electron Microscopy Sciences INSTRUCTION MANUAL CAT. 70309-01 ecount Safety precautions The ecount is used specifically in bioscience, medical, and chemistry laboratories. It may come in contact with biohazardous

More information

User manual BT7900. Register your product and get support at

User manual BT7900. Register your product and get support at User manual BT7900 Register your product and get support at www.philips.com/support Contents 1 Important 1 Safety 1 Notice 1 2 Your Bluetooth speaker 3 Introduction 3 What's in the box 3 Overview of the

More information

Manual Sound level indicator PCE-353

Manual Sound level indicator PCE-353 PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com PCE Instruments UK Ltd. Units 12/13 Southpoint Business

More information

BORESCOPE INSTRUCTIONS INSPECTION CAMERA RANGE RBS100 RBS200 RBS300. Retain these instructions for future reference

BORESCOPE INSTRUCTIONS INSPECTION CAMERA RANGE RBS100 RBS200 RBS300.  Retain these instructions for future reference BORESCOPE INSPECTION CAMERA RANGE RBS100 RBS200 RBS300 INSTRUCTIONS Retain these instructions for future reference www.ringautomotive.co.uk 1 Instructions for Ring Automotive Borescope Inspection Camera

More information

NXT4 Three phase meter for commercial, industrial and advanced residential applications

NXT4 Three phase meter for commercial, industrial and advanced residential applications EMH metering GmbH & Co. KG Neu-Galliner Weg 1 19258 Gallin GERMANY Tel. +49 38851 326-0 Fax +49 38851 326-1129 E-Mail info@emh-metering.com Web www.emh-metering.com Tel. +49 38851 326-1930 (Technical Support)

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

CENTER 521 INSTRUCTION MANUAL 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S)

CENTER 521 INSTRUCTION MANUAL 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S) CENTER 521 INSTRUCTION MANUAL 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S) CONTENTS 1. GENERAL DESCRIPTION... 1 2. SAFETY INFORMATION... 2 3. FEATURES... 2 4. SPECIFICATIONS...

More information

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Manual No. 9835 Revision 0-401 MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Please read this instruction manual carefully before using the instrument. This manual will provide you with all

More information

Version 12/2013 EN. User Manual Radon Scout/Radon Scout Plus Radon logging

Version 12/2013 EN. User Manual Radon Scout/Radon Scout Plus Radon logging Version 12/2013 EN User Manual Radon Scout/Radon Scout Plus Radon logging 1 CONTENT 1. General 3 2. Front panel elements 3 3. Power supply 4 4. Start of measurement 5 5. Interrupt a measurement 6 6. To

More information

Helpline No. UK/Northern Ireland Rep. Ireland Web Support Model Number 82069

Helpline No. UK/Northern Ireland Rep. Ireland Web Support   Model Number 82069 Introduction 1 Congratulations on choosing to buy a Tevion product. By doing so you now have the assurance and peace of mind that comes with purchasing a product made by one of Europe s leading manufacturers.

More information

Content. Sontheim Industrie Elektronik GmbH Page 2 of 15 07/2016 Installation instructions Version 1.2

Content. Sontheim Industrie Elektronik GmbH Page 2 of 15 07/2016 Installation instructions Version 1.2 Content Content... 2 1. Preliminary note... 3 1.1. Symbols used... 3 1.2. Warning signs used... 3 2. Safety instructions... 4 2.1. General... 4 2.2. Target group... 4 2.3. Electrical connection... 4 2.4.

More information

English. Operating Manual. Conductive point level switch MLR120. Retain for later reference. Companies / brands of GHM

English. Operating Manual. Conductive point level switch MLR120.  Retain for later reference. Companies / brands of GHM English Operating Manual Conductive point level switch MLR120 Companies / brands of GHM www.ghm-group.de Retain for later reference Table of contents Page 1. Proper use (application areas)... 3 Safety

More information

Installation Instructions

Installation Instructions LUCCI Slim Line Fan Remote Control SKU# 210012 Rated Voltage 220-240V~ 50Hz Thank you for purchasing this quality Lucci product. To ensure correct function and safety, please read and follow all instructions

More information

Display Panel BAT 100 LSN. en Operation Guide

Display Panel BAT 100 LSN. en Operation Guide Display Panel BAT 100 LSN en Operation Guide Display Panel Table of Contents en 3 Table of contents 1 Safety 4 2 Short information 4 3 System overview 5 4 Installation 6 5 Connection 11 6 Maintenance

More information

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual SIMPLE LOGGER RMS CLAMP-ON MODULE CL600 E N G L I S H User Manual Owner s Record The serial number for the Simple Logger Model CL600 is located on the back of the instrument. Please record this number

More information

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80 User Guide Indoor Air Quality Meter/Datalogger Model EA80 Introduction Congratulations on your purchase of the Extech EA80 Indoor Air Quality Meter. This meter measures Carbon Dioxide (CO 2, ppm) levels,

More information

BT6700. User manual. Register your product and get support at

BT6700. User manual. Register your product and get support at BT6700 User manual Register your product and get support at www.philips.com/support Contents 1 Important 1 Safety 1 Notice 1 2 Your Bluetooth speaker 2 Introduction 2 What's in the box 3 Overview of the

More information

HH-521BT 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S) Instruction Manual

HH-521BT 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S) Instruction Manual HH-521BT 4-Channel Thermocouple Data Logger (with Bluetooth) (Types K, J, T, N, E, R, S) Instruction Manual CONTENTS 1. GENERAL DESCRIPTION... 2 2. SAFETY INFORMATION... 2 3. FEATURES... 2 4. SPECIFICATIONS...

More information

Operating manual. GTL - Configuration tool. Please keep the manual for future use.

Operating manual. GTL - Configuration tool. Please keep the manual for future use. Operating manual GTL - Configuration tool Please keep the manual for future use. V1.00-01 GREISINGER Electronic GmbH Hans-Sachs-Str. 26 93128 Regenstauf Germany Fon +49(0)9402-9383-0 Fax +49(0)9402-9383-33

More information

F-712.PM1 Optical Power Meter

F-712.PM1 Optical Power Meter MP165E, applies to F-712.PM1 F-712.PM1 Optical Power Meter Optical Power Meter for 400-1550 nm Wavelength Range, to 1 ma Input Current, 20 khz Signal Bandwidth, Logarithmic Output ±5 V, Benchtop Device,

More information

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Operating and Installation Manual EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Type EM-TRF-USV for 230 V AC mains voltage; provides uninterruptible power supply (UPS) The art of handling

More information

Hygro-/Thermometer. Operating Manual GFTH 95

Hygro-/Thermometer. Operating Manual GFTH 95 H53.0.X1.6C-07 Hygro-/Thermometer Operating Manual GFTH 95 WEEE-Reg.-No. DE93889386 GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D-93128 Regenstauf +49 (0) 9402 / 9383-0 +49 (0) 9402 / 9383-33

More information

Register your product and get support at www.philips.com/welcome DS3600 EN User manual Contents 1 Important 3 2 Your Docking Speaker 4 Introduction 4 What s in the box 4 Overview of the main unit 5 Overview

More information

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual Colorado Series CR-10 Portable ph / mv / Temperature Meter Operation Manual 2 Contents Page General Introduction and Initial Inspection.............................. 4 Safety Infomation..................................................

More information

RТTH DUAL ROOM SWITCH

RТTH DUAL ROOM SWITCH DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA

More information

Instruction Manual. 24" and 48" Electronic Digital Level Model Nos &

Instruction Manual. 24 and 48 Electronic Digital Level Model Nos & 7123H_Manuals 8/6/13 11:51 AM Page 1 24" and 48" Electronic Digital Level Model Nos. 1876-2400 & 1876-4800 Instruction Manual Congratulations on your choice of this Digital Level. We suggest you read this

More information

-bio. Densitometer DEN-1B Operating instructions

-bio. Densitometer DEN-1B Operating instructions -bio Densitometer Contents 1 Safety...4 2 General Information...6 3 Getting started...8 4 Operation of...10 5 Maintenance...12 6 Specifications...13 7 Guarantee and service...14 Version 1.01 - May 2012

More information

Operating manual. VARIO ph. Hand-held ph meter

Operating manual. VARIO ph. Hand-held ph meter Operating manual VARIO ph Hand-held ph meter ba75373e04 07/2006 VARIO ph Accuracy when going to press The use of advanced technology and the high quality standard of our instruments are the result of continuous

More information

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC MultiTherm Shaker Operations Manual Item: H5000-H and H5000-HC Foreword Thank you for purchasing a MultiTherm Shaker. This operations manual contains instructions for the care of this Instrument. In order

More information

If you have questions about your KCD-2000 and its operation, please contact technical support at

If you have questions about your KCD-2000 and its operation, please contact technical support at This manual contains important information on safety measures and operational features of the KCD-2000 4-way counterfeit detector. Please read it carefully before operating your machine and keep it for

More information

RS Stock No Instruction Manual RS Input Data Logging Thermometer

RS Stock No Instruction Manual RS Input Data Logging Thermometer RS Stock No. 730-0458 Instruction Manual RS-1384 4 Input Data Logging Thermometer EN FR IT DE ES TABLE OF CONTENTS / EN TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION FEATURE... 1 2. SPECIFICATIONS... 2

More information

Always there to help you. Register your product and get support at DS3480. Question? Contact Philips.

Always there to help you. Register your product and get support at   DS3480. Question? Contact Philips. Always there to help you Register your product and get support at www.philips.com/welcome Question? Contact Philips DS3480 User manual Contents 1 Important 2 Safety 2 Notice 2 2 Your Docking Speaker 3

More information

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions SCHOTT FLM 4 Fiber Lighting Modul Operating Instructions Contents 1. Important information 3 2. Scope of Delivery 3 3. Intended use 3 4. Safety information 4 5. Operation 5 5.1. LED Optics Module 5 5.2.

More information

GENERAL SAFETY GUIDELINES

GENERAL SAFETY GUIDELINES This manual contains important safety information about installation and use of this equipment. Ignoring this information could result in injuries or damages. It is strictly forbidden to use this equipment

More information

Indoor Air Quality Meter/Datalogger

Indoor Air Quality Meter/Datalogger User Manual Indoor Air Quality Meter/Datalogger Model EA80 CC Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech EA80 Indoor Air

More information

Type Quickstart (from device Version 2)

Type Quickstart (from device Version 2) Type 8022 Flow Transmitter / Pulse divider Durchflusstransmitter / Impulsteiler Transmetteur de débit / Diviseur d impulsions We reserve the right to make technical changes without notice. Technische Änderungen

More information

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS

More information

Solar Installation PV100 Operating Instructions

Solar Installation PV100 Operating Instructions Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW ENGLAND Tel: +44(0)191 5863511 www.seaward.co.uk sales@seaward.co.uk service@seaward.co.uk Part Number 388A552 Revision 1.2 March 2011

More information

Please read and retain these instructions for future reference v002: 14/03/17. Cradle, Charger, Instruction manual, Lithium-ion Battery 3.7 V d.c.

Please read and retain these instructions for future reference v002: 14/03/17. Cradle, Charger, Instruction manual, Lithium-ion Battery 3.7 V d.c. Power Details: AC Charger Input voltage: 100-240V ~ 50/60Hz, 200mA Cradle Input Voltage: DC 5V, 500mA Cradle Output Voltage: DC 5V, 500mA Accessories: Cradle, Charger, Instruction manual, Lithium-ion Battery

More information

Operating Instructions

Operating Instructions Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW ENGLAND Tel: +44(0)191 5863511 www.seaward.co.uk sales@seaward.co.uk service@seaward.co.uk Part Number 344A550 Revision 1 2006 Seaward

More information

User Guide. ExStik FL700. Fluoride Meter

User Guide. ExStik FL700. Fluoride Meter User Guide ExStik FL700 Fluoride Meter Introduction The model FL700 is a system specifically designed for the quick and accurate measurement of fluoride ions in drinking water and other aqueous samples.

More information

Mk II 3DR MULTI-LINE LASER INSTRUCTION MANUAL

Mk II 3DR MULTI-LINE LASER INSTRUCTION MANUAL Mk II 3DR MULTI-LINE LASER INSTRUCTION MANUAL SAFETY Read the following safety instructions before attempting to operate this product. Keep these instructions in a safe place or store in the carry case

More information

PROFESSIONAL STOPWATCH. professional stoppuhr. User manual ENGLISH EN-1

PROFESSIONAL STOPWATCH. professional stoppuhr. User manual ENGLISH EN-1 PROFESSIONAL STOPWATCH professional stoppuhr User manual ENGLISH EN-1 Thanks for choosing the professional stopwatch Motus Trainer MT38 with lap time measurement and storage at 1/100 second precision.

More information

Type MS04. Redox potential sensor cube. Operating Instructions

Type MS04. Redox potential sensor cube. Operating Instructions Redox potential sensor cube Operating Instructions We reserve the right to make technical changes without notice. Technische Änderungen vorbehalten. Sous réserve de modifications techniques. Bürkert SAS,

More information

Foot switch. Installation, operation and maintenance manual

Foot switch. Installation, operation and maintenance manual Installation, operation and maintenance manual ST Foot switch Read this manual before installing, operating or maintaining this actuator. Failure to follow safety precautions and instructions could cause

More information

Manual. VIPA Accessories USB-MPI adapter 950-0KB31. Order No.: VIPA HB158E_PC-AG Reference: RE_950-0KB31 Rev. 08/25

Manual. VIPA Accessories USB-MPI adapter 950-0KB31. Order No.: VIPA HB158E_PC-AG Reference: RE_950-0KB31 Rev. 08/25 Manual VIPA Accessories USB-MPI adapter 950-0KB31 Order No.: VIPA HB158E_PC-AG Reference: RE_950-0KB31 Rev. 08/25 This manual is relevant for: Product Order number as of state: HW SW USB-MPI adapter VIPA

More information

Deckblatt von Marketing / Foto fehlt

Deckblatt von Marketing / Foto fehlt R-Biopharm AG RIDA QUICK SCAN II (V2.5) Deckblatt von Marketing / Foto fehlt Bedienungsanleitung 2017-11-04 2 All rights reserved. This manual may be used only for its intended purpose. It may not be reproduced

More information

INSPECTION TOOL. Inspection Video Camera with Color LCD Monitor. EU Environmental Protection. Model: 8802LE,8803LE. Model: 8802LE,8803LE

INSPECTION TOOL. Inspection Video Camera with Color LCD Monitor. EU Environmental Protection. Model: 8802LE,8803LE. Model: 8802LE,8803LE INSPECTION TOOL Inspection Video Camera with Color LCD Monitor Model: 8802LE,8803LE EU Environmental Protection Waste electrical products should not be disposed of with household waste. Please recycle

More information

Operating Instructions

Operating Instructions PRIMETEST 100 PRIMETEST 100 Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW ENGLAND Tel: +44(0)191 5863511 www.seaward.co.uk sales@seaward.co.uk service@seaward.co.uk Part Number

More information

LSL250. Laser light! Do not stare into the beam or view directly with magnifiers. Class 2 laser EN :

LSL250. Laser light! Do not stare into the beam or view directly with magnifiers. Class 2 laser EN : Lufkin LSL250 Professional rotary laser in extremely rugged design horizontal levelling even under tough conditions. Automatic rotary laser 635 nm, self-levelling range ± 5 horizontal, accuracy 1 mm /

More information

Technical Manual Room temperature controller

Technical Manual Room temperature controller 1473-1-8563 03.07.2014 Technical Manual Room temperature controller RTC-F-1 ABB-free@home Content 1 Notes on the instruction manual 3 2 Safety 4 2.1 Used symbols 4 2.2 Intended use 5 2.3 Improper use 5

More information

FAZE 800 FT PRO USER MANUAL. Version: 1.2. ROBE Lighting s.r.o. Czech Republic

FAZE 800 FT PRO USER MANUAL. Version: 1.2. ROBE Lighting s.r.o. Czech Republic FAZE 800 FT PRO USER MANUAL Version: 1.2 ROBE Lighting s.r.o. Czech Republic www.robe.cz FAZE 800 FT PRO Table of contents 1. Introduction 4 2. Cautions 4 3. Opening the package and inspecting the machine

More information

USER MANUAL. For Flic. English

USER MANUAL. For Flic. English USER MANUAL For Flic English Welcome to the Flic family! Take control and simplify your life with the push of a Flic. - Love, Flic Team Contents Setup guide 1 Pairing your Flic 2 Bluetooth connection 3

More information

Bluetooth/USB Data Logger USER S MANUAL. Hantek 365A/B/C/D/E/F V

Bluetooth/USB Data Logger USER S MANUAL. Hantek 365A/B/C/D/E/F V Bluetooth/USB Data Logger USER S MANUAL Hantek 365A/B/C/D/E/F V 1.0.3 www.hantek.com Content General Safety Summary... 1 Chapter 1 Getting Start... 3 1.1 General Check... 4 1.2 The User interface... 5

More information

English. Operating manual. Universal transmitter UT125. Save for later reference. Companies / brands of GHM

English. Operating manual. Universal transmitter UT125.  Save for later reference. Companies / brands of GHM English Operating manual Universal transmitter UT125 Companies / brands of GHM www.ghm-messtechnik.de Save for later reference. Table of contents page 1. Intended use (areas of application)... 3 1.1 Safety

More information