Blood Hemolysis Detector User Manual
|
|
- Barnaby Brooks
- 6 years ago
- Views:
Transcription
1 Blood Hemolysis Detector User Manual Written by: Kayla Marquis, Jordan Seekins, Hanna Anderson, Marie-Louise Doe April 2017
2 Table of Contents OVERVIEW... 3 GETTING STARTED... 4 CONTENTS IN BOX... 4 ADDITIONAL SUPPLIES OR EQUIPMENT NECESSARY FOR OPERATION... 5 SETUP... 6 MICROFLUIDICS SETUP... 6 OPTICAL UNIT SETUP... 6 LABVIEW CALIBRATION... 7 PHONE APPLICATION SET-UP... 7 OPERATION... 9 SAMPLE SEPARATION... 9 OPTICAL ANALYSIS-LABVIEW UNIT OPTICAL ANALYSIS-PHONE CAMERA ASSEMBLY MAINTENANCE CLEANING PART REPLACEMENT TROUBLESHOOTING MICROFLUIDICS OPTICS SPECIFICATIONS MICROFLUIDICS OPTICS SCHEMATICS OVERALL SYSTEM MICROFLUIDIC CHANNEL CIRCUIT SCHEMATIC LABVIEW BLOCK DIAGRAM BREADBOARD CIRCUIT OPTICAL ANALYSIS CHAMBER PHONE FILTER ATTACHMENT SAMPLE CARTRIDGE HOUSING GLOSSARY OF TECHNICAL TERMS
3 Overview 3
4 When red blood cells rupture, free hemoglobin is released. If not immediately detected, severe health complications such as acute kidney or heart failure could result. Traditional free hemoglobin detection methods require blood to be mixed with chemicals as well as the use of expensive detection equipment. For a fraction of the time, money, and sample size of conventional methods, this point-of-care hemolysis detector will indicate the concentration of free hemoglobin while maintaining the integrity of the blood sample. This hemolysis detector consists of a blood plasma separator and an optical detector. The plasma separator uses a microfluidic system to isolate the plasma from the red blood cells. Free hemoglobin is contained in the plasma, permitting efficient detection using a photoresistor and LED to determine the concentration of free hemoglobin in the sample. A second option for optical analysis for a mobile phone is also provided, where a phone analysis unit and instructions for the ImageJ analysis software are included. Getting Started Contents in box Microfluidic chamber with connected tubing and luer lock fittings 3mL luer lock syringe Glass slide capillary tube Optical analysis chamber with cover and circuitry in breadboard 4
5 Hemolysis Analyser LabView VI disc Phone analysis unit: Phone filter attachment Sample cartridge housing Additional supplies or equipment necessary for operation Syringe pump Amazing Goop, or other waterproof sealant 70% ethanol cleaning solution Computer NI mydaq console, terminal connector and USB cable Set of calibration slides Galaxy S7 Edge cellphone LabVIEW ImageJ Microsoft Excel or other spreadsheet program 5
6 Setup Microfluidics Setup 1. No setup is required. Optical Unit Setup 1. Attach the breadboard wires into the black NI mydaq terminal connector by securing them into the indicated channels. Green: +15V Yellow: AGND (AO) Orange: 0+(AI +/-10V) Red: 0 (DIO 0-5V) Brown: DGND 2. Connect the NI mydaq console to computer. 3. Open the LabView VI titled Hemolysis Analyzer from the disc provided. 6
7 LabVIEW Calibration 1. Run LabView VI by clicking the arrow in the toolbar. 2. Turn LED on by pressing LED On/Off button. For this part, disregard instrument hemolysis indications 3. Insert 0 mg/ml calibration slide into analyser. 4. Record the voltage shown in the Calibration Control Panel. 5. Do this for remaining 5 calibration slides. 6. Using Excel, plot the voltage vs. the concentration. 7. Fit a linear trendline to the data. 8. Input the slope and y-intercept of this line into the controls in the calibration panel. 9. VI will now display concentration data based on the calibration line determined. Phone Application Set-up 1. Slide the phone filter attachment over the top half of the phone. 7
8 2. Make sure the slots for the camera and flash are aligned properly. 3. Attach the sample box to the device. 8
9 Operation Sample Separation 1. Acquire blood sample, minimum of 100μL 2. Put blood sample in 3 ml luer lock syringe 3. Attach syringe luer lock to tubing 4. Insert the red blood cell outlet tubing into a small collection container. 9
10 5. Put plasma outlet tubing into contact with glass slide capillary tube to fill with plasma. 6. Set syringe pump flow rate to 25 ml/hr to initiate whole blood separation. 7. Once slide reservoir is full stop syringe pump. 8. Seal the slide assembly with waterproof sealant to prevent leakage of sample. 9. Choose LabVIEW unit or phone camera optical analysis. 10
11 Optical Analysis-LabVIEW Unit 1. Once the Hemolysis Analyzer VI is open and calibration is complete, use the Threshold Concentration control on the VI front panel to input a desired threshold value for hemolysis concentration (omit units when entering; program reads as mg/dl). 2. Insert sealed capillary slide containing the plasma sample in the slot of the optical analysis chamber. 11
12 3. Place the chamber cover over the sample to block out external light. 4. Press the white arrow in the top lefthand corner of the VI front panel, indicated here by the red circle, to run the analysis. 5. Press the LED On/Off button to turn on the LED in the chamber. 6. Allow analysis to run until result is displayed in the Sample Concentration box, indicating the concentration of hemoglobin in the sample (mg/dl). 12
13 7. If sample concentration is equal to or higher than the threshold value input in (1), the green Normal Sample indicator will turn off, and the red Hemolyzed Sample indicator will turn on. 8. Press the red STOP icon to halt the VI when sample has been analyzed. Optical Analysis-Phone Camera Assembly 1. Insert sample slide into slot on the right side of the sample cartridge housing. 2. Photograph sample slide with phone s integrated camera application. 3. Transfer image to computer and open file in ImageJ. 4. Go to Image > Open and select appropriate picture. 13
14 5. Go to Analyze > Set Measurements > select Mean gray scale. 6. Make sure Mean Gray Value is selected. Click OK to close menu. 7. Using a selection tool, select sample area in image. 14
15 8. Go to Analyze > Measure. Mean Gray Value will be displayed in pop-up box. 9. Use calibration curve to determine sample concentration. Maintenance Cleaning 1. Clean microfluidic portion of device by rinsing with 70% ethanol, run rinse to remove color and then continue rinsing for 1 minute after to ensure complete cleaning. 15
16 2. Following cleaning with 70% ethanol, rinse system with 1x phosphate buffered saline. 3. When the microfluidic chamber becomes discolored it will need to be discarded and an additional unit ordered. Part replacement Refer to Specifications-Optics for part details if replacement becomes necessary. Contact manufacturers if new 3D-printed parts are needed. Troubleshooting Microfluidics If microfluidic separation is blocked or leaking, replace with new component. Optics If LabVIEW VI doesn t run upon initiation of the Hemolysis Analyzer program, check that both the terminal connector on the NI mydaq console and the USB cable are secured properly. If a higher than reasonable concentration occurs, ensure that the LED control has been clicked to the ON position, indicated by a green icon on the dashboard. If the above problem occurs but the LED is visibly on, check the connection of the photoresistor. Photoresistor may need to be replaced (see specifications below). Specifications Microfluidics Minimum blood sample size: 100μL Maximum blood flow rate: 25 ml/hr Tubing size: Inner Diameter x Outer Diameter x Width Materials: Tubing: polytetrafluoroethylene (PTFE) Microfluidic channel: polydimethylsiloxane (PDMS) Adhesive for tubing microfluidic channel connection: Gorilla Clear Epoxy Adhesive Optics Optical Analysis Chamber: 3D-printed polylactic acid (PLA) chamber Dimensions: Photoresistor: Illuminance 16
17 P/N: PDV-P9008-ND Luna Optoelectronics Light-emitting diode: LEDWE-15 Ultra-Bright White LED Thor Labs Center Wavelength Tolerance (CWL): nm Total Optical Power: 13.0 mw at 20mA Breadboard: 500k Ohm Resistors 50k Ohm Resistors Hardware and Software: Labview NI 2013 or newer required, National Instruments NI mydaq compatibility and software Phone Specifications: Samsung Galaxy S7 Edge Schematics Overall System 17
18 Microfluidic Channel Circuit Schematic 18
19 LabVIEW Block Diagram Breadboard Circuit Optical Analysis Chamber (Dimensions in mm) 19
20 Phone Filter Attachment (Dimensions given in mm) 20
21 Sample Cartridge Housing (Dimensions given in mm) 21
22 22
23 Glossary of Technical Terms Absorbance - Ability of a material or substance to absorb light. Bandpass - A type of filter that permits the transmission of one wavelength with a narrow surrounding range indicated by the center wavelength tolerance. Blood plasma - Yellowish colored liquid component of blood that aids in the suspension of the red blood cells in whole blood. Makes up about 55% of the total blood volume. Breadboard - A board for making an experimental circuit. Calibration curve - Method for determining the concentration of an unknown sample based on known values. Free hemoglobin - When red blood cells rupture, hemoglobin is released, or freed. Free hemoglobin is very reactive outside the red blood cells, potentially resulting in harmful bodily reactions and processes. Hemoglobin - Contained within red blood cells. Hemolysis - The rupturing of red blood cells that releases the cellular contents into the plasma. Luer lock - Type of tubing connection, useful for fast connections and disconnections. LED - Light-Emitting Diode. Microfluidics - The movement of fluids through a microscopic system. Microfluidic separation chamber - Aids in separation of red blood cells and plasma by varying channel dimensions and angles. Photoresistor - A resistor where the resistance changes depending on the intensity of the surrounding light. 23
RD-SDM02 Contact Angle Measurement Instrument. Operation Manual
RD-SDM02 Contact Angle Measurement Instrument Operation Manual The components of the RD-SDM02 system are not resistant against acids, bases or aggressive organic solvents. RD Support does not provide for
More informationEducational Microfluidic Starter Kit
Educational Microfluidic Starter Kit Product datasheet Page Description 4 Benefits 5 System Specifications 5 Typical System setup 6 Parts List 7 IP License 19 MAR-000189_v.A.11 Page 1 of 20 Low Pressure
More information9 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module. InfinityLab LC Series Diode Array Detectors User Manual
InfinityLab LC Series Diode Array Detectors User Manual 9 Maintenance Introduction to Maintenance 158 Warnings and Cautions 159 Overview of Maintenance 161 Cleaning the Module 162 Remove and Install Doors
More informationMICROFLUIDIC EOF KIT 9015
MICROFLUIDIC EOF KIT 9015 User Manual - Version 1.1 Visiting address Colosseum 15 7521 PV Enschede Postal address P.O. Box 214 7500 AE Enschede The Netherlands Phone +31 53 850 6 850 Fax +31 53 850 6 851
More informationVersion 1.4 June 2018
1 Version 1.4 June 2018 Table Of Contents 1. Introduction 3 2. Unpacking the Instrument 4 2.1. Standard Accessories 2.2. Optional Accessories 3. Specifications 4 4. Installation 5 4.1. Quick 4-20 ma Start
More informationMitos P-Pump. product datasheet
product datasheet page Product description 2 Main benefits 3 Why choose the Mitos P-Pump? 3 Closed-loop flow control 4 Product specifications 5 Accessories - overview 6 Mitos P-Pump Basic 7 Mitos P-Pump
More informationInstallation and operation manual ReciFlow Gas
Installation and operation manual ReciFlow Gas 1 1. Measurement principle... 3 2. Installation... 5 3. Operation... 7 4. Electrical interfaces... 11 5. Communication protocol... 14 6. Software update and
More informationStandard Operating Procedure for the Horiba FluroMax-4
Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.
More informationManual for the MicroSIA Instrument
Manual for the MicroSIA Instrument Version 1.1 1 FIAlab Instruments, Inc., 2151 N. Northlake Way, Suite 100, Seattle, WA 98103 USA Table of Contents 1 INTRODUCTION... 3 1.1 GENERAL... 3 1.2 OPERATION PRINCIPLE...
More informationStrain and Force Measurement
NORTHEASTERN UNIVERSITY DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 0-MEASUREMENT AND ANALYSIS Strain and Force Measurement OBJECTIVES The primary objective of this experiment
More informationIssue : 1.1 Date : 19/1/2004. Trigger IO. C-Cam Technologies. a division of. Vector International 1 / 6
Trigger IO C-Cam Technologies a division of Vector International 1 / 6 1 Trigger IO connector Trigger IO connector Camera with LS interface Trigger IO connector Camera with USB interface 2 / 6 1.1 Pin
More informationAppendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
More informationThe Olis Titrator Series
The Olis Titrator Series Perfect Titrations into any type of cuvette With considerable effort, one can do precise titration studies by hand. Or, one can let the OLIS titrators do the technician s work
More informationMontgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System
Page 1 of 9 Approvals Preparer: Jason McMillan Reviewer: Dr. Maggie Bryans Date: 18DEC14 Date: 18DEC14 1. Purpose: Operation of the BioLogic LP Chromatography System. 2. Scope: Applies to the BioLogic
More informationQ Exactive HF-X QuickStart Guide
Q Exactive HF-X QuickStart Guide About this Guide Q Exactive HF-X Tune This Q Exactive HF-X QuickStart Guide gives an introduction on setting up and using the Thermo Scientific Q Exactive HF-X mass spectrometer.
More informationMercury Manual No. IM-M1000 Rev G
Mercury 1000 Analog Output Encoder System Installation Manual and Reference Guide Manual No. IM-M1000 Rev G Introduction MicroE Systems was founded to advance encoder technology to a level never before
More informationNAME EET 2259 Lab 3 The Boolean Data Type
NAME EET 2259 Lab 3 The Boolean Data Type OBJECTIVES - Understand the differences between numeric data and Boolean data. -Write programs using LabVIEW s Boolean controls and indicators, Boolean constants,
More informationCH142 Spring Spectrophotometers with Vernier Data Acquisition Software
Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample
More informationStandard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System. Last Modified October 19, 2012
Standard Operating Procedure for the Beckman PA/CE MDQ Electrophoresis System Last Modified October 19, 2012 Considerations Prior to Beginning Experiment All solutions must be filtered (using 0.45 m or
More information180 DEGREE PEEL TESTER
180 DEGREE PEEL TESTER MODEL PA-2000-180 OPERATING INSTRUCTIONS CHEMINSTRUMENTS 510 COMMERCIAL DRIVE FAIRFIELD, OHIO 45014 (513) 860-1598 www.cheminstruments.com Revision 1.0 November 5, 2018 CONTENTS
More informationCR9 LIGHT SOURCE FOR FORENSIC CRIME SCENE INVESTIGATION LIGHT SOURCE FOR FORENSIC CRIME SCENE INVESTIGATION LATEST LED TECHNOLOGY
Headquarters, R&D, Manufacturing, Domestic and International sales Obelux Oy Kutomotie 6B FI-0030 Helsinki Finland Tel +35 9 671 600 Fax +35 9 621 551 E-mail info@obelux.com www.obelux.com LATEST LED TECHNOLOGY
More informationSpraybase Technical Specifications Sheet
The Spraybase electrospinning device offers a complete solution to users investigating electrospraying and electrospinning applications. Designed to be compact and easy to use, the Spraybase electrospinning
More informationAgilent Splitter Kits
Agilent Splitter Kits Installation and Configuration This technical note describes the installation and configuration of the Agilent Splitter Kits in an Agilent 1260 Infinity Series System or an Agilent
More informationLab 4: Introduction to ELVIS II+ Introduction to ELVIS II+
Page 1 of 12 Laboratory Goals Introduction to ELVIS Lab 4: Introduction to ELVIS Familiarize students with the National Instruments hardware ELVIS Identify the capabilities of ELVIS Make use of the built
More informationCh En 475: Introduction to Instrumentation and Signal Processing with Labview
Ch En 475: Introduction to Instrumentation and Signal Processing with Labview Measurement Instrumentation Rapid, on-line measurement of temperature, pressure, liquid level, flow rate and composition is
More informationTECH 3821 Lab #2 Relay Driver with Computer Control
TECH 3821 Lab #2 Relay Driver with Computer Control Name: Background: One of the most basic controls in industry is the ability to turn things on and off. As we saw in Lab #1, a relay is often used to
More informationCalibration of Recorded Digital Counts to Radiance
Graduate Radiometry, 1051-719 Laboratory No. 2 Fall Quarter, 2006 Calibration of Recorded Digital Counts to Radiance INTRODUCTION The goal of this experiment is to calibrate your digital camera, using
More informationNanosight NS300 NBTC User Instructions
Nanosight NS300 NBTC User Instructions Preliminary notes: -The laser module is delicate and expensive. Handle it with care and do not leave it on the bench unattended. Only leave it in place in the instrument
More informationS800 Spectrawave Visible Diode Array Spectrophotometer
6 SCANNING VISIBLE INSTRUMENT FOR EDUCATION OPTIMISED FOR THE TEACHING LABORATORY S800 Spectrawave Visible Diode Array Spectrophotometer ABSORBANCE, % TRANSMISSION, CONCENTRATION AND KINETICS LARGE, EASY
More informationUser Operation Manual
User Operation Manual Manual Version: 1.3 For Single Cell Bioanalyzer (SCB-402) Doc# 102-SCB-402-1.3 Table of Contents 1. General Information... 1 1.1 How to Use This Manual... 1 1.2 Operation Safety...
More informationBL121 Pool & Spa Controller. ph/orp Swimming Pool Controller with Built-In Chemical Feed Pumps
BL121 Pool & Spa Controller ph/orp Swimming Pool Controller with Built-In Chemical Feed Pumps Introduction The BL121 is an all-in-one solution for automatic control of ph and chlorine levels in swimming
More informationMicroreactor Chips. product datasheet
Unit 1, Anglian Business Park, Orchard Road, Royston, Hertfordshire, SG8 5TW, UK T: +44 (0)1763 242491 F: +44 (0)1763 246125 E: sales@dolomite-microfluidics.com W: www.dolomite-microfluidics.com Microreactor
More informationLed Illuminator and Light Stimulator. Model # LSD-1. Ver
Led Illuminator and Light Stimulator Model # LSD-1 Ver. 2.0.0 5.16 ALA Scientific Instruments Inc. 60 Marine Street, Farmingdale, NY 11735 Voice: +1 631.393.6401 fax: +1 631.393.6407 www.alascience.com
More informationPHYC 500: Introduction to LabView. Exercise 16 (v 1.2) Controlling hardware with DAQ device. M.P. Hasselbeck, University of New Mexico
PHYC 500: Introduction to LabView M.P. Hasselbeck, University of New Mexico Exercise 16 (v 1.2) Controlling hardware with DAQ device This exercise has two parts. First, simulate a traffic light circuit
More informationUNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY
UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY HIGH TEMPERATURE & HIGH FREQUENCY TESTING SYSTEM LABORATORY USER MANUAL 4/29/2011 Overview
More informationSepacore X10 / X50 system Technical data sheet
Sepacore X10 / X50 system Technical data sheet Sepacore X10 and X50 flash chromatography systems address most requirements for the purification of organic compounds. Whether a crude synthesis mixture or
More informationSupplemental guide on Camera Fabrication
Supplemental guide on Camera Fabrication This document describes the steps for fabricating your camera and lighting system. 1 Materials needed Gather all your materials. You will need the small lexan sheet
More informationPrinting Your First Page. Attaching the Paper Support. Plugging in the Printer. Checking the Printer
Printing Your First Page Attaching the Paper Support Checking the Printer Plugging in the Printer Installing the Ink Cartridges Installing the Printer Software Connecting the Printer 4011307 XXX-00 Attaching
More information-bio. Densitometer DEN-1B Operating instructions
-bio Densitometer Contents 1 Safety...4 2 General Information...6 3 Getting started...8 4 Operation of...10 5 Maintenance...12 6 Specifications...13 7 Guarantee and service...14 Version 1.01 - May 2012
More informationAppendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips
Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to
More informationImages Scientific OWI Robotic Arm Interface Kit (PC serial) Article
Images Scientific OWI Robotic Arm Interface Kit (PC serial) Article Images Company Robotic Arm PC Interface allows real time computer control and an interactive script writer/player for programming and
More informationTube Liquid Sensor OPB350 / OCB350 Series (Calibration Circuit Available)
Features: Can identify if liquid is present in clear tubes that have an outside diameter of 1/16 [1.6mm], 1/8" [3.2mm], 3/16" [4.8 mm] or 1/4" [6.3 mm] Opaque plastic housing enhances ambient light rejection
More informationInstructions for using Go!Link Data Acquisition with the Pharmacia UV-1
Instructions for using Go!Link Data Acquisition with the Pharmacia UV-1 Introduction The Go!Link interface, the Instrumentation Amplifier, and the Logger Lite software permit data acquisition from just
More informationTrigger I/O Manual. Trigger IO Manual. C-Cam Technologies. a division of. Vector International. Date :12/01/2011 Page 1 / 8 Version 1.
Trigger IO Manual C-Cam Technologies a division of Vector International Date :12/01/2011 Page 1 / 8 Version 1.6 1 Introduction This manual describes the trigger input-output feature of cameras produced
More informationArterial pressure transducer APT300
Arterial pressure transducer APT300 inexpensive reliable accurate stopcocks included easy to fill simple holder for rod mount The APT300 transducer is an inexpensive pressure transducer which can be used
More information9 Maintenance and Repair
Agilent 1200 Infinity II Series RID User Manual 9 Maintenance and Repair Introduction to Maintenance 120 Warnings and Cautions 121 Overview of Maintenance 123 Cleaning the Module 124 Storage of the Detector
More informationRadnoti. Radnoti. Radnoti. Precise digitally controlled temperature. Illuminated chamber improves sample viewing
Radnoti Myography Systems Precise digitally controlled temperature Illuminated chamber improves sample viewing Featuring an ergonomically designed sliding wrist rest Modular designed allowing for easy
More informationGlass Pipette & PEEK Tubing Connectors
Glass Pipette & PEEK Tubing Connectors Quick and reliable connection of glass capillary needles and PEEK tubing to most Removable Needle (RN) and Knurled Hub (KH) terminations. Mass Spectrometer Tuning
More informationMinimum < 0.2 ms, maximum 2 ms, typical 1 ms; On request: Delays or holding times for bubble events % relative humidity (not condensing)
The air bubble detector is used to detect air or gas bubbles in plastic tubes filled with liquid and is intended to prevent air infusions. However, it can also be used as a wet/dry sensor. The sensor has
More informationUltrospec 10 User Manual. English. Deutsch. Français. Español. Italiano
English Deutsch Ultrospec 10 User Manual Français Español Italiano Biochrom US Telephone: 1-508-893-8999 84 October Hill Road Toll Free: 1-800-272-2775 Holliston, MA Fax: 1-508-429-5732 01746-1388 support@hbiosci.com
More information3.5x2.8mm SURFACE MOUNT LED LAMP. Features. Description. Package Dimensions. Part Number: AA3528SES/J4. Super Bright Orange
3.5x2.8mm SURFACE MOUNT LED LAMP Part Number: AA3528SES/J4 Super Bright Orange Features Single color. Suitable for all SMT assembly and solder process. Available on tape and reel. Ideal for backlighting.
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Luminex Prima Theory of Operation Luminex 200 technology is based on flow cell
More informationMost Reliable Sensor Technology. - electric field - inductive - capacitive - optical
Most Reliable Sensor Technology - electric field - inductive - capacitive - optical DIRECTORY {LEVEL_2_2} Touch Interfaces and Indicators switch and indicator at once touch activated - no tiring no wear
More informationOwner s Manual. Model 6672A. 600 Series ipad Conversion Kit 6672A-16 1/17
Owner s Manual Model 6672A 600 Series ipad Conversion Kit 6672A-16 1/17 Harvest Tec 6672A Table of Contents Page Introduction 3 Installation of Bluetooth Receiver 3-6 Large Square Baler Power Cable and
More informationReading: Graphing Techniques Revised 1/12/11 GRAPHING TECHNIQUES
GRAPHING TECHNIQUES Mathematical relationships between variables are determined by graphing experimental data. For example, the linear relationship between the concentration and the absorption of dilute
More informationUser's Guide. Video Borescope. Models BR200 and BR250
User's Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was
More informationINFINITY-CORRECTED TUBE LENSES
INFINITY-CORRECTED TUBE LENSES For use with Infinity-Corrected Objectives Available in Focal Lengths Used by Thorlabs, Nikon, Leica, Olympus, and Zeiss Designs for Widefield and Laser Scanning Applications
More informationSPM Zero dead volume dilutor DATASHEET
Environmental conditions Operating temperature Humidity Max pressure 5-40 C (41-104 F) 20-80%, non-condensing 5 bars (72 psi) Fluidic characteristics Tube port fittings Wetted materials Channel diameter
More informationConnecting LEDs to the ADB I/O
Application Note AN-2 By Magnus Pettersson September 26 1996 Connecting LEDs to the I/O Introduction The following notes are for those of you who are a bit inexperienced with hardware components. This
More informationC101-E137 TALK LETTER. Vol. 14
C101-E137 TALK LETTER Vol. 14 Diffuse Reflectance Measurement of Powder Samples and Kubelka-Munk Transformation ------- 02 Application: Measuring Food Color ------- 08 Q&A: What effect does the scan speed
More informationMicroprocessor Resistivity & Conductivity Monitor/Controller EC410 / EC430. Operation Manual
Microprocessor Resistivity & Conductivity Monitor/Controller EC410 / EC430 Operation Manual INDEX 1 Specifications 2 Assembly and installation 2.1 Precautions for installation 2.2 Installation of controller
More informationUHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 04
Introduction UHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 04 Prizmatix UHP-M Dual-LED light engine is designed to substitute the Metal-Halide and Mercury lamps in routine
More informationNCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02
1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See
More informationCoaxial Electrospinning & Electrospray StartUp Lab Device V2.0
Doxa Microfluidics offers a wide variety of Coaxial Electrospinning and Electrospray Lab Equipment for the design and production of Core-Shell and hollow, and plain nanofibers and particles. All the Lab
More informationMicroprocessors B Lab 3 Spring PIC24/24LC515 EEPROM Interface Using I 2 C
PIC24/24LC515 EEPROM Interface Using I 2 C Lab Report Objectives Materials See separate report form located on the course webpage. This form should be completed during the performance of this lab. 1) To
More informationPerkinElmer Series 200 HPLC: Operating Procedure
PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering
More informationAqueous GPC Instructions - Detailed
Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must
More informationInfrared Add-On Module for Line Following Robot
1 Infrared Add-On Module for Line Following Robot January 3, 2015 Jeffrey La Favre The infrared add-on module allows multiple line following robots to operate on the same track by preventing collisions
More informationSelf-contained Photoelectric Sensors. PZ-V/M Series
Self-contained Photoelectric Sensors PZ-V/M Series Unaffected by Angle, Colour, or Dirt Introducing Multi-reflective Photoelectric Sensors PZ-V Provide Solutions to All the Problems Experienced with Conventional
More informationUser Guide. Video Borescope. Models BR200 and BR250
User Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was designed
More informationBio Compatibility Kit
General Information 1 Installing the 2 How to Use the 5 Important Safety and Solvent Compatibility Considerations 5 Setting the Minimum or Maximum Operating Pressure in ChemStation 7 Setting the Minimum
More informationCHAPTER 2: THREE DIMENSIONAL TOPOGRAPHICAL MAPPING SYSTEM. Target Object
CHAPTER 2: THREE DIMENSIONAL TOPOGRAPHICAL MAPPING SYSTEM 2.1 Theory and Construction Target Object Laser Projector CCD Camera Host Computer / Image Processor Figure 2.1 Block Diagram of 3D Areal Mapper
More informationKINETICS CALCS AND GRAPHS INSTRUCTIONS
KINETICS CALCS AND GRAPHS INSTRUCTIONS 1. Open a new Excel or Google Sheets document. I will be using Google Sheets for this tutorial, but Excel is nearly the same. 2. Enter headings across the top as
More informationOne step ahead in performance and mounting ease
95 Micro Photoelectric Amplifier Built-in PM-5 SERIES PM-45 SERIES PM-5 SERIES General terms and conditions... F- guide... P.9~ Related Information Glossary of terms / General precautions...p.1549~ / P.155~
More informationThe mystem Project Board Quick Start
The mystem Project Board Quick Start This activity will introduce you to the basic functions of the mystem Project Board to get you started. Introduction to the mystem The mystem Project Board is designed
More informationPRODUCT SHEET. NON-INVASIVE SMALL ANIMAL TAIL BLOOD PRESSURE SYSTEMS NIBP250 Blood Pressure Amplifier NIBP200A Blood Pressure System
NON-INVASIVE SMALL ANIMAL TAIL BLOOD PRESSURE SYSTEMS NIBP250 Blood Pressure Amplifier NIBP200A Blood Pressure System NIBP Amplifiers with built-in pump automatically inflate the tail cuff to occlude the
More informationWireless Inspection Camera with recording LCD monitor. Product code: SRCAMV6 CAMERA NUMBER:
User Manual Super Cam V6 SRCAMV6 Wireless Inspection Camera with recording LCD monitor. Product code: SRCAMV6 CAMERA NUMBER: Please read this user manual carefully before using this product. CONTENTS Welcome...
More informationBruker BioSpin FLOW NMR. Racks & Accessories Catalog. Version. NMR Spectroscopy. think forward
Bruker BioSpin FLOW NMR Racks & Accessories Catalog Version 004 think forward NMR Spectroscopy Copyright by Bruker BioSpin GmbH All rights reserved. No part of this publication may be reproduced, stored
More informationMu lt i s p e c t r a l
Viewing Angle Analyser Revolutionary system for full spectral and polarization measurement in the entire viewing angle EZContrastMS80 & EZContrastMS88 ADVANCED LIGHT ANALYSIS by Field iris Fourier plane
More information1500V - Vacuum Rated
Preliminary 8/20/2008 Mercury Digital Output Encoder System Installation Manual and Reference Guide 1500V - Vacuum Rated Customer Controller Vacuum Wall Manual No. IM-M1500V Rev S7 Introduction MicroE
More informationModels 1417 and 1437 User s Manual. High-Speed Photodetector Modules
Models 1417 and 1437 User s Manual High-Speed Photodetector Modules Handling Precautions The detector is sensitive to electrostatic discharges and could be permanently damaged if subjected even to small
More informationSBS Software. Signals By Spreadsheet. Figure 1, SBS System Configuration
Figure 1, SBS System Configuration Internet (www.signalsbyspreadsheet.com) Interface Distributed Input/Output DIO Power Detectors Signals Positions In this guide, you will: Download (Initial Install for
More informationMODEL 8000MP LEVEL SENSOR
1 MODEL 8000MP LEVEL SENSOR INSTRUCTIONS FOR INSTALLATION, OPERATION & MAINTENANCE VISIT OUR WEBSITE SIGMACONTROLS.COM 2 SERIES 8000MP LEVEL SENSOR 1. DESCRIPTION The Model 8000MP Submersible Level Sensor
More informationLAMP /F2C4-FHNO/R23
1224-15/F2C4-FHNO/R23 Features High luminous power Typical chromaticity coordinates x=0.258, y=0.228 according to CIE1931 Bulk, available taped on Ammo. ESD-withstand voltage: up to 4KV The product itself
More informationCh En 475: Introduction to Instrumentation and Signal Processing with Labview
Ch En 475: Introduction to Instrumentation and Signal Processing with Labview Measurement Instrumentation Rapid, on-line measurement of temperature, pressure, liquid level, flow rate and composition is
More informationUNC Flow Cytometry Core Facility LSR II Operation 2018
Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration).
More informationNOTE: Read the instructions completely through before beginning actual installation.
CONSOLE CONNECTOR KIT 7850 INSTALLATION INSTRUCTIONS FOR USE WITH: Various Two-channel Organs LESLIE Speaker Models 705, 705C, 720 KIT CONTENT Console Connector 138488 Switch Assembly, Cable Assembly,
More informationChromatography Software Training Materials. Contents
Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will
More informationContents Contents 2 Preface 3 Introduction 4 Components 5 Operation 6 Timer 7 Lamp Replacement 9 Specification 10 Ordering Information 11 2
SPOT-LITE CURING SYSTEM USER S GUIDE http://www uvp com UVP, Inc Ultra-Violet Products Ltd 2066 W 11 th Street Unit 1, Trinity Hall Estate, Nuffield Road Upland, CA 91768 USA Cambridge, CB4 1TG, UK (800)
More informationDual channel temperature logger with two voltage inputs 0-5V Instruction Manual
LOGGER S0541 Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual Instruction Manual for use of S0541 logger Instrument is designed for measurement and record of temperature
More informationLab 5: Diffraction and Interference
Lab 5: Diffraction and Interference Light is a wave, an electromagnetic wave, and under the proper circumstances, it exhibits wave phenomena, such as constructive and destructive interference. The wavelength
More informationSpecifications: LED Ring Lights. Product Number
1 LED Ring Lights provide direct illumination with high and even intensity on specular or opaque surfaces. Beam angle is factory adjusted according to customers demand on working distance (WD) and illuminated
More informationMODELING LED LIGHTING COLOR EFFECTS IN MODERN OPTICAL ANALYSIS SOFTWARE LED Professional Magazine Webinar 10/27/2015
MODELING LED LIGHTING COLOR EFFECTS IN MODERN OPTICAL ANALYSIS SOFTWARE LED Professional Magazine Webinar 10/27/2015 Presenter Dave Jacobsen Senior Application Engineer at Lambda Research Corporation for
More informationUnaffected by Angle, Colour, or Dirt
Cat No. PZ-C2 Unaffected by Angle, Colour, or Dirt Introducing Multi-reflective Photoelectric Sensors PZ-V Provide Solutions to All the Problems Experienced with Conventional Reflective Types. Conventional
More informationPanorama rudder indicator TRI-2 Analogue or CAN input Approved according to MED Three extra large scales Long-life LED illumination Built-in dimmer
USER'S MANUAL Panorama rudder indicator TRI-2 Analogue or CAN input Approved according to MED Three extra large scales Long-life LED illumination Built-in dimmer DEIF A/S Frisenborgvej 33 DK-7800 Skive
More informationINTRODUCTION TO LABVIEW
INTRODUCTION TO LABVIEW 2nd Year Microprocessors Laboratory 2012-2013 INTRODUCTION For the first afternoon in the lab you will learn to program using LabVIEW. This handout is designed to give you an introduction
More informationDimensions. Electrical connection
VCS0-K Dimensions Connector 89 Connector 8 6 8 7 6 Connector 8 Connector Light guide connection Electrical connection Release date: 0-09- : Date of issue: 0-0- 8_eng.xml M VCS 0 8 09 Model Number VCS0-K
More informationHarvatek International 0.4 Quad Digit Clock Display HCD89010
Harvatek International.4 Quad Digit Clock Display Official Product Customer Part No. 573187-4-99 Data Sheet No. **************** **************** Oct. 17. 213 Version of 1.8 Page 1/7 Revision History Revision
More informationSPEC NO: DSAH3956 REV NO: V.7 DATE: NOV/26/2008 PAGE: 1 OF 6 APPROVED: WYNEC CHECKED:
3.5x2.8 mm SMD CHIP LED LAMP PRELIMINARY SPEC Part Number: AA3529SES/L Hyper Orange Features Single color. Suitable for all SMT assembly and solder process. Available on tape and reel. Ideal for backlighting.
More informationVE-53 Sensor Uniaxial / Triaxial Surface and Downhole Velocity Sensor. Operation Manual
VE-53 Sensor Uniaxial / Triaxial Surface and Downhole Velocity Sensor Operation Manual Company: Author: Checked: Approved: Level: Distribution: GeoSIG Ltd Wiesenstrasse 39, 8952 Schlieren, Switzerland,
More information