A Further Study on Large Size LSO and LYSO Crystal Samples

Size: px
Start display at page:

Download "A Further Study on Large Size LSO and LYSO Crystal Samples"

Transcription

1 October 25, 2005 IEEE NSS, N12-6, Puerto-Rico 1 A Further Study on Large Size LSO and LYSO Crystal Samples Jianming Chen, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology

2 BGO, LSO & LYSO Samples % Ce in LSO/LYSO, 5--10% yttrium fraction in LYSO Cube: 1.7 cm (1.5 X 0 ), Bar: 2.5 x 2.5 x 20 cm (18 X 0 ) SIC BGO CPI LYSO Saint-Gobain LYSO CTI LSO 2

3 Experiment Without any thermal treatment, all samples went through measurement for optical and scintillation properties. Optical properties measured: transmittance, emission and excitation spectra. Light output and light response uniformity were measured with PMT and Si APD readout. Radiation induced phosphorescence intensity was measured for two LYSO long samples under γ ray irradiation, which was used to estimate readout noise with 100 ns gate. 3

4 Excitation, Emission & Transmittance Identical transmittance, emission & excitation spectra Part of emitted light is self-absorbed in long samples 1.7 cm Cube 2.5 x 2.5 x 20 cm Bar 4

5 CTI LSO: longitudinal optical uniformity No longitudinal variation in optical properties Transverse transmittance approaches theoretical limit 5

6 Emission Weighted Q.E. Areas under emission spectra proportional to the LY 59%/75% QE for LSO/LYSO readout with Si PD/APD 6

7 PMT Based Readout with Coincidence Systematic error with repeated mounts & measurements: < 1% 7

8 137 Cs & 22 Na Pulse Height Spectra Cube and bar samples have 8% and 10% FWHM resolution respectively for 137 Cs (0.66 MeV) and 22 Na source (0.51 MeV) CPI LYSO bar has double peak because of poor annealing Cube Bar 8

9 Light Output & Decay Time LSO/LYSO Light yield: a factor of 4/100 of BGO/PWO LSO/LYSO decay time: 42 ns compared to 300 ns of BGO Cube Bar 9

10 APD Based Readout with Coincidence Two Hamamatsu S APD, Canberra 2003 BT preamplifier and ORTEC 673 shaping amplifier with shaping time 250 ns Hamamatsu S APD 10

11 APD Readout Calibration Readout noise: 67 electrons for 2 APD with 250 ns shaping Systematic error with repeated mountings & measurements: 2.4% Calibration with Fe-55 source: 2.38 electrons/adc Channel 11

12 Pulse Height & Light Yield with APDs Na-22 annihilation peak (510 kev) well measured CTI LSO sample has significantly higher LO 12

13 LSO Light Response Uniformity Uniformity depends on end coupled to the PMT/APD, indicating a not uniform light yield along crystal 13

14 LYSO LRU Measured with APD Uniformity depends on which end coupled to the APD, indicating a not uniform light yield along crystal 14

15 γ ray Induced Readout Noise Sample L.Y. F Q 15 rad/h Q 500 rad/h σ 15 rad/h σ 500 rad/h ID p.e./mev μa/rad/h p.e. p.e. MeV MeV CPI 1, x x SG 1, x x γ ray induced PMT anode current can be converted to the photoelectron numbers (Q) integrated in 100 ns gate. Its statistical fluctuation contributes to the readout noise (σ). 15

16 CMS PWO Resolution 77k PWO Designed Resolution Energy Resolution: σ(e)/e < 1% if E > 25 GeV σ(e)/e ~ 0.5% at 120 GeV 2004 SM Beam Test Monitoring: 0.04% 16

17 LSO/LYSO ECAL Performance Less demanding to the environment because of small temperature coefficient. Radiation damage is less an issue as compared to the CMS PWO ECAL. A better energy resolution, σ(e)/e, at low energies than L3 BGO and CMS PWO because of its high light output and low readout noise: % / E 0.55% /E / E 17

18 Summary Ce doped LSO & LYSO crystals have fast (42 ns) and high (4 X BGO) light output. The light output of 2.5 x 2.5 x 20 cm LSO and LYSO samples, excited by 0.51 MeV γ ray, can be readout by Hamamatsu S APD (2 x 0.25 cm 2 ) with electronic noise of 70 electrons (< 35 kev). Longitudinal light response uniformity of long LSO/LYSO samples seems affected by Ce distribution along crystal. The radiation induced phosphorescence in 2.5 x 2.5 x 20 cm LYSO causes ~0.2 MeV 15 rad/h. An LSO/LYSO crystal calorimeter would provide very good energy resolution down to MeV range for future HEP experiments. To be understood: why LSO long sample has higher LO as compared to long LYSO samples with APD readout. 18

19 Possible Origin of Non Uniformity C. Melcher: LO in LSO is a function of Ce concentration B. Chai: LO in LYSO is a function of atomic fraction of Yttrium 19

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

The Progress of TOF on BESIII

The Progress of TOF on BESIII The Progress of TOF on BESIII Yuekun Heng 1, Chong Wu 1, Zhijia Sun 1, Li Zhao 1,2, Cheng Li 2, Qi An 2, Shubin Liu 2, Jinjie Wu 2, Yuda Zhao 1,3, Fengmei Wang 1,4, Xiaojian Zhao 1, Feng Shi 1, Zhenghua

More information

SiPMs for Čerenkov imaging

SiPMs for Čerenkov imaging SiPMs for Čerenkov imaging Peter Križan University of Ljubljana and J. Stefan Institute Trends in Photon Detectors in Particle Physics and Calorimetry, Trieste, June 2-4, 2008 Contents Photon detectors

More information

AX-PET : A novel PET concept with G-APD readout

AX-PET : A novel PET concept with G-APD readout AX-PET : A novel PET concept with G-APD readout Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet

More information

HHS Public Access Author manuscript Phys Med Biol. Author manuscript; available in PMC 2016 February 19.

HHS Public Access Author manuscript Phys Med Biol. Author manuscript; available in PMC 2016 February 19. Monte Carlo calculations of PET coincidence timing: single and double-ended readout Stephen E Derenzo, Woon-Seng Choong, and William W Moses Life Sciences Division, Lawrence Berkeley National Laboratory,

More information

The LHCb upgrade. Outline: Present LHCb detector and trigger LHCb upgrade main drivers Overview of the sub-detector modifications Conclusions

The LHCb upgrade. Outline: Present LHCb detector and trigger LHCb upgrade main drivers Overview of the sub-detector modifications Conclusions The LHCb upgrade Burkhard Schmidt for the LHCb Collaboration Outline: Present LHCb detector and trigger LHCb upgrade main drivers Overview of the sub-detector modifications Conclusions OT IT coverage 1.9

More information

Scintillators and photodetectors. 1. Generation of Optical Photons 2. Transport of Optical Photons 3. Detection of Optical Photons

Scintillators and photodetectors. 1. Generation of Optical Photons 2. Transport of Optical Photons 3. Detection of Optical Photons Scintillators and photodetectors 1. Generation of Optical Photons 2. Transport of Optical Photons 3. Detection of Optical Photons 1) Generation of Optical Photons A) Organic (molecular) scintillators Advantages:

More information

Development of LYSO Detector Modules for an EDM Polarimeter at COSY. for the JEDI Collaboration

Development of LYSO Detector Modules for an EDM Polarimeter at COSY. for the JEDI Collaboration Mitglied der Helmholtz-Gemeinschaft Development of LYSO Detector Modules for an EDM Polarimeter at COSY for the JEDI Collaboration February 28, 2018 DPG Spring Meeting, PhD @ SMART EDM_Lab, TSU, Georgia

More information

TORCH: A large-area detector for precision time-of-flight measurements at LHCb

TORCH: A large-area detector for precision time-of-flight measurements at LHCb TORCH: A large-area detector for precision time-of-flight measurements at LHCb Neville Harnew University of Oxford ON BEHALF OF THE LHCb RICH/TORCH COLLABORATION Outline The LHCb upgrade TORCH concept

More information

Using only HyCal. Graph. Graph

Using only HyCal. Graph. Graph Live charge weighted ep yield from all the 2GeV empty target runs Scattering angle from.7 to.9 deg, background dominated by upstream collimator (8%) Notice that here uncertainty from the live charge measurement

More information

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies Suleman Surti Physics & Instrumentation Group Department of Radiology University of Pennsylvania

More information

In-vivo dose verification for particle therapy

In-vivo dose verification for particle therapy In-vivo dose verification for particle therapy D.R. Schaart, NCS Lustrum, 5-Oct-2012 HollandPTC 1 Protons vs. photons Photons Protons Dennis R. Schaart Delft University of Technology 2 Protons: the promise

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998 Studies of 1 µm-thick silicon strip detector with analog VLSI readout arxiv:physics/981234v1 [physics.ins-det] 18 Dec 1998 T. Hotta a,1, M. Fujiwara a, T. Kinashi b, Y. Kuno c, M. Kuss a,2, T. Matsumura

More information

Extremely Fast Detector for 511 kev Gamma

Extremely Fast Detector for 511 kev Gamma Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,

More information

Front end electronics and system design for the NUSTAR experiments at the FAIR facility. FEE 2006 Workshop

Front end electronics and system design for the NUSTAR experiments at the FAIR facility. FEE 2006 Workshop Front end electronics and system design for the NUSTAR experiments at the FAIR facility Presented by Ian Lazarus on behalf of NUSTAR collaboration FEE 2006 Workshop Ian Lazarus NPG, CCLRC Daresbury Overview

More information

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs) 8//6 The EPID Strikes Back: Future EPID Technology and Applications Larry E. Antonuk Department of Radiation Oncology University of Michigan, Ann Arbor Acknowledgements: Youcef El-Mohri, Qihua Zhao (U.

More information

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System

Large Plastic Scintillation Detectors for the Nuclear Materials Identification System Large Plastic Scintillation Detectors for the Nuclear Materials Identification System J.S. Neal, J.T. Mihalczo, M. T. Hiatt, J. D. Edwards Oak Ridge National Laboratory P. O. Box 2008, Oak Ridge, Tennessee

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Digital Filters in Radiation Detection and Spectroscopy

Digital Filters in Radiation Detection and Spectroscopy Digital Filters in Radiation Detection and Spectroscopy Digital Radiation Measurement and Spectroscopy NE/RHP 537 1 Classical and Digital Spectrometers Classical Spectrometer Detector Preamplifier Analog

More information

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,

More information

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES

CHAPTER 11 NUCLEAR MEDICINE IMAGING DEVICES CHAPTER 11 M.A. LODGE, E.C. FREY Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins University, Baltimore, Maryland, United States of America 11.1. INTRODUCTION Imaging

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays

A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays I. Rech, A. Ingargiola, R. Spinelli, S. Marangoni, I. Labanca, M. Ghioni, S. Cova Dipartimento di Elettronica ed Informazione Politecnico

More information

X-Ray fluorescence and Raman spectroscopy

X-Ray fluorescence and Raman spectroscopy X-Ray fluorescence and Raman spectroscopy Advanced physics laboratory (nd part) 4CFU Catalini Letizia, De Angelis Giulia Vittoria, Piselli Verdiana Abstract In this paper we report about two different

More information

Development of Ultrafast CXRS system in Heliotron J. Graduate School of Energy Science Kyoto University LU XIANGXUN 03/15/2016

Development of Ultrafast CXRS system in Heliotron J. Graduate School of Energy Science Kyoto University LU XIANGXUN 03/15/2016 1 Development of Ultrafast CXRS system in Heliotron J Graduate School of Energy Science Kyoto University LU XIANGXUN 03/15/2016 2 Outline 1. Introduction 2. Charge exchange Recombination Spectroscopy (CXRS)

More information

Work in Tbilisi. David Mchedlishvili (SMART EDM_lab of TSU) GGSWBS , Tbilisi. Shota Rustaveli National Science Foundation

Work in Tbilisi. David Mchedlishvili (SMART EDM_lab of TSU) GGSWBS , Tbilisi. Shota Rustaveli National Science Foundation Mitglied der Helmholtz-Gemeinschaft David Mchedlishvili (SMART EDM_lab of TSU) Work in Tbilisi GGSWBS 18 23.08.2018, Tbilisi JEDI: Charged-Particle EDM Search Main principle: Inject polarized particles

More information

Timing properties of MCP-PMT

Timing properties of MCP-PMT Photon Detector Workshop at Kobe, 27-29 June 27 Timing properties of MCP-PMT - Time resolution - Lifetime - Rate dependence K.Inami (Nagoya university, Japan) Introduction Photon device for TOP counter

More information

Goals of this presentation

Goals of this presentation Transferring HEP Innovative concepts of electronics, trigger and DAQ Architecture for future large experiments to state of the art applications in medical imaging. P. Le Dû p.ledu@ipnl.in2p3.fr DEC_2010

More information

Interpolating Silicon Photomultipliers

Interpolating Silicon Photomultipliers Interpolating Silicon Photomultipliers Peter Fischer, Heidelberg University, Germany (Presenter) Claudio Piemonte, FBK, Italy We present the novel Interpolating Silicon PhotoMultiplier (ISiPM) topology

More information

Measuring Reflective Properties of PTFE for Applications in Liquid Xenon Detectors

Measuring Reflective Properties of PTFE for Applications in Liquid Xenon Detectors Measuring Reflective Properties of PTFE for Applications in Liquid Xenon Detectors Catherine Lussenhop Columbia University, Department of Physics August 6, 2010 Abstract The XENON100 experiment uses 161kg

More information

Delayline Detectors. Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution. e - I +

Delayline Detectors. Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution. e - I + Delayline Detectors Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution e - I + Delayline Readout of MCPs - The Technical Approach - Microchannel-Plate (MCP) detectors provide

More information

This section gives a brief description of the MatrixB system outlining its main features and applications.

This section gives a brief description of the MatrixB system outlining its main features and applications. Modular Scintillator Readout Solution for Nuclear Medicine Document Overview This document provides the user with a comprehensive description of the hardware and software provided in the SensL Matrix readout

More information

Positron Emission Tomography

Positron Emission Tomography Physics 656 Seminar on Physical Fundamentals of Medical Imaging Positron Emission Tomography Ahmed Qamesh Outline What is PET? PET mechanism Radionuclide and its synthesis Detection concept and Development

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

The AMS-02 Anticoincidence Counter. Philip von Doetinchem I. Phys. Inst. B, RWTH Aachen for the AMS-02 Collaboration DPG, Freiburg March 2008

The AMS-02 Anticoincidence Counter. Philip von Doetinchem I. Phys. Inst. B, RWTH Aachen for the AMS-02 Collaboration DPG, Freiburg March 2008 I. Phys. Inst. B, RWTH Aachen for the AMS-02 Collaboration DPG, Freiburg March 2008 Cosmic Rays in the GeV Range world average SUSY DM KK DM good agreement of data and propagation models, but some unexplained

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,500 108,000 1.7 M Open access books available International authors and editors Downloads Our

More information

Section 2 PROGRESS IN LASER FUSION

Section 2 PROGRESS IN LASER FUSION Section 2 PROGRESS IN LASER FUSION 2.A Optical Fiducials for X-Ray Streak Cameras X-ray streak cameras are the primary instruments for studying the transient nature of laser-produced plasmas. However,

More information

Supporting information for: A highly directional room-temperature single. photon device

Supporting information for: A highly directional room-temperature single. photon device Supporting information for: A highly directional room-temperature single photon device Nitzan Livneh,, Moshe G. Harats,, Daniel Istrati, Hagai S. Eisenberg, and Ronen Rapaport,, Applied Physics Department,

More information

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Effects of system geometry and other physical factors on photon sensitivity of high-resolution Home Search Collections Journals About Contact us My IOPscience Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography This article

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Project GAMMA: Adaptive Methods for Medical Applications

Project GAMMA: Adaptive Methods for Medical Applications Project GAMMA: Adaptive Methods for Medical Applications V. Solovov, A. Morozov, V. Chepel, V. Domingos, R.Martins, L. Pereira, F. Neves, A. Lindote, F. Fraga, I.Lopes, F. Alves LIP-Coimbra and Coimbra

More information

PSA Performance Analysis and Optimization

PSA Performance Analysis and Optimization AGATA Week Madrid 2014 IKP Cologne 23. Januar 2014 Content Introduction 1 Introduction 2 3 Introduction PSA performance within specifications but: Systematic errors Non isotropic distribution of hits with

More information

FPGA based charge fast histogramming for GEM detector

FPGA based charge fast histogramming for GEM detector FPGA based charge fast histogramming for GEM detector Krzysztof T. Pozniak a, A. Byszuk a, M. Chernyshova b, R. Cieszewski a, T. Czarski b, W. Dominik c, K. Jakubowska b, G. Kasprowicz a, J. Rzadkiewicz

More information

SCINTILLATORS-SiPM COUPLING

SCINTILLATORS-SiPM COUPLING SCINTILLATORS-SiPM COUPLING G.FELICI LNF-INFN SCINTILLATOR ßàSIPM 1 st PROBLEM: EJ-200 peak emission spectrum do not match S18825 peak sensitivity AMPLITUDE 1.0 0.8 0.6 0.4 0.2 EJ-200 EMISSION SPECTRUM

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420.

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420. EURONS I3 506065 JRA9 RHIB Report made during stay IEM-CSIC Madrid december 2006 MINISTERIO DE ASUNTOS EXTERIORES Y DE COOPERACIÓN AECI VICESECRETARÍA GENERAL The Monte Carlo simulation of a Package formed

More information

Development of automated ultraviolet laser beam profiling system using fluorometric technique

Development of automated ultraviolet laser beam profiling system using fluorometric technique Development of automated ultraviolet laser beam profiling system using fluorometric technique BB Shrivastava*, NS Benerji, P Bhatnagar, HS Vora a and U Nundy Chemical and Excimer Laser Section a Laser

More information

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN PH EP/2004 050 19 August 2004 Novel Geometrical Concept of a High Performance Brain PET Scanner - Principle, Design and Performance Estimates J. Séguinot

More information

hsfc pro 12bit ultra speed intensified imaging

hsfc pro 12bit ultra speed intensified imaging hsfc pro 12bit ultra speed intensified imaging four MCP-image intensifier camera modules ultra fast shutter down to 3ns (optional 1.5ns) in single mode excellent sensitivity of the system allows single

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range

PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range PMT Candidates for SSD Challenge for PMTs in air-shower detection highest linear dynamic range Karl-Heinz Becker, Karl-Heinz Kampert, Alex Kääpä, Christian Pauly, Sven Querchfeld, Julian Rautenberg OBSERVATORY

More information

Obtainment of Prototypical images and Detector performance simulations. Guillaume Potdevin for the XFEL-HPAD-Consortium

Obtainment of Prototypical images and Detector performance simulations. Guillaume Potdevin for the XFEL-HPAD-Consortium Obtainment of Prototypical images and Detector performance simulations Overview of the analysis Outlook Prototypical images: Single object imaging & XPCS short presentation (reminder ) Presentation of

More information

UCS-20. Operating Manual Macintosh Edition. Spectrum Techniques, Inc. Oak Ridge, Tennessee USA. Universal Computer Spectrometer.

UCS-20. Operating Manual Macintosh Edition. Spectrum Techniques, Inc. Oak Ridge, Tennessee USA. Universal Computer Spectrometer. UCS-20 Universal Computer Spectrometer Operating Manual Macintosh Edition April 2004 Spectrum Techniques, Inc. Oak Ridge, Tennessee USA. TABLE OF CONTENTS Overview Software Disk Menus File Edit Mode Calculations

More information

Progress in FDIRC & TOF

Progress in FDIRC & TOF Progress in FDIRC & TOF J. Va vra, SLAC Content FDIRC prototype - Activities related to analysis of the last test beam run TOF detector - Summary of test results Next steps Appendix - PID electronics summary

More information

UCS 10. Operating Manual. Spectrum Techniques, Inc. Oak Ridge, Tenn. USA. Universal Computer Spectrometer. August 1996

UCS 10. Operating Manual. Spectrum Techniques, Inc. Oak Ridge, Tenn. USA. Universal Computer Spectrometer. August 1996 UCS 10 Universal Computer Spectrometer Operating Manual August 1996 Spectrum Techniques, Inc. Oak Ridge, Tenn. USA. TABLE OF CONTENTS Overview Software Disk Menus File Edit Mode Calculations Tools Settings

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information

Modeling the ORTEC EX-100 Detector using MCNP

Modeling the ORTEC EX-100 Detector using MCNP Modeling the ORTEC EX-100 Detector using MCNP MCNP is a general-purpose Monte Carlo radiation transport code for modeling the interaction of radiation with materials based on composition and density. MCNP

More information

Update on Energy Resolution of

Update on Energy Resolution of Update on Energy Resolution of the EMC Using µµγ Sample David Hopkins Royal Holloway, University of London EMC Reconstruction Workshop, December 5 th, 2004 Outline Study of photon energy resolution Compare

More information

Uniformity scan of the 6cm tubes. Jingbo Wang Argonne National Laboratory, Lemont, IL

Uniformity scan of the 6cm tubes. Jingbo Wang Argonne National Laboratory, Lemont, IL Uniformity scan of the 6cm tubes Jingbo Wang Argonne National Laboratory, Lemont, IL wjingbo@anl.gov Update of the test system Realized the computer control for the multi-channel HV supply, the 2D translation

More information

NEAR-IR BROADBAND POLARIZER DESIGN BASED ON PHOTONIC CRYSTALS

NEAR-IR BROADBAND POLARIZER DESIGN BASED ON PHOTONIC CRYSTALS U.P.B. Sci. Bull., Series A, Vol. 77, Iss. 3, 2015 ISSN 1223-7027 NEAR-IR BROADBAND POLARIZER DESIGN BASED ON PHOTONIC CRYSTALS Bogdan Stefaniţă CALIN 1, Liliana PREDA 2 We have successfully designed a

More information

Time-of-Flight Technology

Time-of-Flight Technology Medical Review Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems INTRODUCTION Improving the care for every patient while providing a high standard care to

More information

Analogue, Digital and Semi-Digital Energy Reconstruction in the CALICE AHCAL

Analogue, Digital and Semi-Digital Energy Reconstruction in the CALICE AHCAL Analogue, Digital and Semi-Digital Energy Reconstruction in the AHCAL Deutsches Elektronen Synchrotron (DESY), Hamburg, Germany E-mail: coralie.neubueser@desy.de Within the collaboration different calorimeter

More information

5 th HERD workshop CERN. Presentation of scmos in IsCMOS Sub-System. Jiarui GAO XIOPM, CAS

5 th HERD workshop CERN. Presentation of scmos in IsCMOS Sub-System. Jiarui GAO XIOPM, CAS 5 th HERD workshop CERN Presentation of scmos in IsCMOS Sub-System Jiarui GAO XIOPM, CAS 20171011 1 Outline Background scmos Design Next step 2 Background scmos is the readout module of IsCMOS. IsCMOS

More information

Chapter 7. Widely Tunable Monolithic Laser Diodes

Chapter 7. Widely Tunable Monolithic Laser Diodes Chapter 7 Widely Tunable Monolithic Laser Diodes We have seen in Chapters 4 and 5 that the continuous tuning range λ is limited by λ/λ n/n g, where n is the index change and n g the group index of the

More information

GEM TPC Anode Pad Shape Studies

GEM TPC Anode Pad Shape Studies GEM TPC Anode Pad Shape Studies a,b, S. Kappler a,c, B. Ledermann a, Th. Müller a, M. Ronan d TPC-Application Workshop Berkeley, 04/08/2006 a b SLAC, Menlo Park, USA c RWTH, Aachen d LBNL, Berkeley, USA

More information

6-1 LECTURE #6: OPTICAL PROPERTIES OF SOLIDS. Basic question: How do solids interact with light? The answers are linked to:

6-1 LECTURE #6: OPTICAL PROPERTIES OF SOLIDS. Basic question: How do solids interact with light? The answers are linked to: LECTURE #6: OPTICAL PROPERTIES OF SOLIDS Basic question: How do solids interact with light? The answers are linked to: Properties of light inside a solid Mechanisms behind light reflection, absorption

More information

Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations

Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations V. Rossi Albertini, B. Paci & P. Perfetti Istituto di Struttura

More information

Results of Radiation Test of the Cathode Front-end Board for CMS Endcap Muon Chambers

Results of Radiation Test of the Cathode Front-end Board for CMS Endcap Muon Chambers Results of Radiation Test of the Cathode Front-end Board for CMS Endcap Muon Chambers B. Bylsma 1, L.S. Durkin 1, J. Gu 1, T.Y. Ling 1, M. Tripathi 2 1 Department of Physics, Ohio State University, Columbus,

More information

ISOTOPE IDENTIFICATION AND SOURCE DIRECTIONAL DETERMINATION USING GAMMA RADIATION PROFILING

ISOTOPE IDENTIFICATION AND SOURCE DIRECTIONAL DETERMINATION USING GAMMA RADIATION PROFILING University of Nebraska - Lincoln DigitalCommons@University of Nebraska - Lincoln Theses, Dissertations, and Student Research from Electrical & Computer Engineering Electrical & Computer Engineering, Department

More information

Dark Photon Search in e+e- Annihilation. MMAPS = Missing Mass A-Prime Search

Dark Photon Search in e+e- Annihilation. MMAPS = Missing Mass A-Prime Search Dark Photon Search in e+e- Annihilation MMAPS = Missing Mass A-Prime Search Cornell University J. Alexander, D. Cassel, M. Perelstein, D. Rubin, P. Wittich University of Minnesota Y. Kubota, B. Wojtsekhowski

More information

LASCAD Tutorial No. 2: Modeling a laser cavity with side pumped rod

LASCAD Tutorial No. 2: Modeling a laser cavity with side pumped rod LASCAD Tutorial No. 2: Modeling a laser cavity with side pumped rod Revised January 19, 2009 Copyright 2006-2009 LAS-CAD GmbH Table of Contents 1 Table of Contents 1 Starting LASCAD and Defining a Simple

More information

LECTURE 15. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 15. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 15 Dr. Teresa D. Golden University of North Texas Department of Chemistry Typical steps for acquisition, treatment, and storage of diffraction data includes: 1. Sample preparation (covered earlier)

More information

Monte Carlo Simulations for the Purpose of Efficiency Curve Calibration for the Fastscan Whole Body Counter

Monte Carlo Simulations for the Purpose of Efficiency Curve Calibration for the Fastscan Whole Body Counter Monte Carlo Simulations for the Purpose of Efficiency Curve Calibration for the Fastscan Whole Body Counter by Hannah Robyn Graham A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Accurate LED Source Modeling using TracePro

Accurate LED Source Modeling using TracePro Accurate LED Source Modeling using TracePro Presented by : Lambda Research Corporation 25 Porter Rd. Littleton, MA 01460 Moderator: Mike Gauvin Vice President of Sales and Marketing Lambda Research Corporation

More information

LASERS FOR ANALYTICAL INSTRUMENTATION

LASERS FOR ANALYTICAL INSTRUMENTATION CW Lasers Combiners Q-switch Lasers >20 s MM/SM/PM fiber Single-frequency options 405 nm to 660 nm 4 wavelengths Open beam or Fiber 1029 nm / 514.5 nm High pulse energy Single Longitudinal Mode LASERS

More information

FST s status on EUV Pellicle & Inspection System Development

FST s status on EUV Pellicle & Inspection System Development FST s status on EUV Pellicle & Inspection System Development OCT.04, 2015 EUV Pellicle TWG @ Imec, nl. Donwon Park FST (Korea) http://www.fstc.co.kr FST Business Segments Division Pellicle TCU (Temperature

More information

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Alexander Korn, Juergen Giersch a and Martin Hoheisel b a Physikalisches Institut Universitaet Erlangen-Nuernberg,

More information

Fluorescence. Requirements. Introduction. Models: FluorescenceExampleBegin.oml. Properties: FluorescenceExampleProperties.txt

Fluorescence. Requirements. Introduction. Models: FluorescenceExampleBegin.oml. Properties: FluorescenceExampleProperties.txt Fluorescence Requirements Models: FluorescenceExampleBegin.oml Properties: FluorescenceExampleProperties.txt Editions: TracePro Expert Introduction TracePro Expert is capable of modeling fluorescent material.

More information

GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI.

GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI. Status of GEANT4 in LHCb S. Easo, RAL, 30-9-2002 The LHCbexperiment. GEANT4 is used for simulating: RICH testbeam data, HCAL testbeam data. GAUSS Project: LHCb Simulation using GEANT4 with GAUDI. Summary.

More information

Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP

Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP C.STOECKL, G. FISKEL, R. K. JUNGQUIST, P. M. NILSON, AND W. THEOBALD University of Rochester, Laboratory for Laser Energetics Spherical Crystal

More information

IDM-200 -V. Interchangeable HPGe Detector Module (IDM) for Systems Integrators

IDM-200 -V. Interchangeable HPGe Detector Module (IDM) for Systems Integrators IDM-200 -V Interchangeable HPGe Detector Module (IDM) for Systems Integrators Free from liquid nitrogen, free from maintenance, free from integration complexities. Convenient-to-mount, all-in-one integrated

More information

Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax

Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD 20723 (301) 953-6871 (301) 953-6670 fax Understand the instrument. Be able to convert measured counts/pixel on-orbit into

More information

FT Cal and FT Hodo DAQ and Trigger

FT Cal and FT Hodo DAQ and Trigger FT Cal and FT Hodo DAQ and Trigger Outline FT-Cal and FT-Hodo read-out electronics FT-Cal and FT-Hodo DAQ and trigger FADC250 firmware CTP firmware for FT-Cal and FT-Hodo FT-Cal and FT-Hodo crates and

More information

TPC digitization and track reconstruction: efficiency dependence on noise

TPC digitization and track reconstruction: efficiency dependence on noise TPC digitization and track reconstruction: efficiency dependence on noise Daniel Peterson, Cornell University, DESY, May-2007 A study of track reconstruction efficiency in a TPC using simulation of the

More information

Development of a high resolution animal PET with continuous crystals and SiPMs

Development of a high resolution animal PET with continuous crystals and SiPMs Development of a high resolution animal PET with continuous crystals and SiPMs 1, John Barrio1, Jorge Cabello1,2, Ane Etxebeste1, Carlos Lacasta1, Josep F. Oliver1, Magdalena Rafecas1, Carles Solaz1, Vera

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector K. Karafasoulis, K. Zachariadou, S. Seferlis, I. Papadakis, D. Loukas, C. Lambropoulos, C. Potiriadis Abstract Spectroscopic

More information

A full digital approach in Physics Applications

A full digital approach in Physics Applications A full digital approach in Physics Applications March 2015 Carlo Tintori ISO 9001:2008 CERT. N. 9105.CAEN CAEN Digitizers Highlights VME, NIM, Desktop form factors Up to 64 channels in a single board 8,

More information

WAVELENGTH MANAGEMENT

WAVELENGTH MANAGEMENT BEAM DIAGNOS TICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS MONITORS Camera Accessories WAVELENGTH MANAGEMENT UV CONVERTERS UV Converters

More information

Text for the class, Pump and probe technique for picosecond time-resolved x-ray diffraction at the Cheiron School

Text for the class, Pump and probe technique for picosecond time-resolved x-ray diffraction at the Cheiron School Yoshihito Tanaka, Kiminori Ito Oct. 3-4, 2011 Text for the class, Pump and probe technique for picosecond time-resolved x-ray diffraction at the Cheiron School 1. Introduction 1-1. Purpose The pulsed nature

More information