Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Size: px
Start display at page:

Download "Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide"

Transcription

1 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins, Member, IEEE, S. Surti, Member, IEEE, M. E. Werner, M. E. Daube-Witherspoon, Senior Member, IEEE, L. Popescu, Member, IEEE, S. Vandenberghe, G. Muehllehner, Fellow, IEEE Abstract A proto-type time-of-flight (TOF) 3D PET scanner based on lanthanum bromide detectors has been developed. The LaBr 3 (5%Ce) Anger-logic detectors in this new scanner use 4x4x30 mm 3 pixels and continuous light-guide coupled to a hexagonal array of 50-mm PMTs. The scanner consists of 24 modules with a 93-cm detector diameter and 25-cm axial fieldof-view. Initial characterization of scanner performance has been performed, including energy and timing performance. We currently measure an overall system energy resolution of 7.5% and a system timing resolution is 460 ps, although we expect these results to improve eventually when the electronics are fully optimized. Since there are not yet standard tests to quantify the benefit of TOF, we designed two phantoms with hot and cold spheres in 27-cm and 35-cm diameter vessels to evaluate the TOF performance as a function of body size. The data from this scanner are reconstructed with a fully 3D listmode iterative TOF algorithm with all data corrections incorporated into the system model. We find that TOF reconstruction reduces the noise and background variability, especially for the larger phantom representing a large patient. In addition, TOF improves detail and contrast of the spheres (lesions), especially the smallest 10-mm sphere. The TOF reconstruction reaches convergence faster than the non-tof reconstruction, and the rate of convergence is seen to be more insensitive to object size. These results indicate that TOF will help improve image quality and potentially reduce scan time with clinical patients. I. INTRODUCTION In this paper we describe the development of a time-offlight (TOF) scanner based on lanthanum bromide. The fundamental improvement with TOF has been shown to result from a variance reduction in the reconstruction of the data by incorporating the TOF information in the forward- and backprojection steps. The variance reduction translates to an effective sensitivity gain which has in the past been described as a ratio of D/Δx [1], where D is the object diameter and Δx = c Δt/2. It is intuitive that the TOF gain is related to this Manuscript received November 11, This work was supported in part by the National Institutes of Health under Grant No. R33-EB001684, and sponsored research agreements with Saint-Gobain and Philips Medical Systems. J. S. Karp is with the Department of Radiology, University of Pennsylvania, Philadelphia, PA USA (tel: , joelkarp@mail.med.upenn.edu). A. Kuhn, S. Surti, M. Werner, M. E. Daube-Witherspoon, L. Popescu, and G. Muehllehner, are with the Department of Radiology, University of Pennsylvania, Philadelphia, PA USA. A. E. Perkins is with Philips Research, Briarcliff Manor, NY USA. S. Vandenberghe was with Philips Research USA and is now with the Medisip/Elis University Ghent, Belgium. ratio, but not necessarily equal to it. In fact, it is too simplistic to characterize the TOF gain as a single value, since it will certainly be task dependent and must also depend on the method of data correction and image reconstruction. For example, Tomitani et al [2] argued that once reconstruction effects are included, the variance reduction (or sensitivity gain) should be D/1.6Δx, thus reducing the impact of TOF compared to the gain proposed in the paper of Budinger et al [1]. We have recently performed a lesion detectability study based on non-pre-whitening match filter analysis and simulated data with 10-mm lesions [3], and find the gain in (SNR-NPW) 2 is comparable to the reduction in variance predicted by [2]. For example, for a 27-cm diameter phantom the improvement is approximately 3.1, 1.9, and 1.4 for assumptions of 300 ps, 600 ps, and 1000 ps (fwhm) timing resolution. Using the formula in [2] we calculate gains of 3.8, 1.9, and 1.1 for the same assumptions, whereas using the formula in [1] we calculate gains of 6.0, 3.0, and 1.8. These differences suggest to us that we must proceed carefully in trying to quantify the TOF gain with measured data, but certainly it will be important to strive for the best possible timing resolution in order to extract the maximum benefit from TOF reconstruction. We have developed the lanthanum bromide PET scanner since we have achieved very good timing resolution and energy resolution with pixelated LaBr 3 detectors. LaBr 3 with 0.5% cerium concentration was first reported by the research group at Delft University, van Loef et al [4] and this work was followed by the group at Radiation Monitoring Devices, Shah et al [5] who found that the rise time, thus timing resolution improves as the cerium concentration increases. In [5] the timing resolution of LaBr 3 vs. BaF 2 is reported to 390 ps fwhm for 0.5% cerium, 260 ps for 5% cerium, and 170 ps for 30% cerium. The light output remains fairly constant, close to 60,000 photons per MeV, and the energy resolution of these small samples is better than 4% fwhm for 511 kev. Compared to BaF 2 the light output is 30x higher, thus with LaBr 3 we can overcome a major limitation of the TOF scanners in the 1980 s that had poor spatial resolution. We have developed arrays of LaBr 3 with 5% cerium concentration by working closely with Saint-Gobain in order to demonstrate that the excellent performance of this scintillator could be translated into a PET detector, which requires good sensitivity and spatial resolution, in addition to energy and timing resolution. In previous work, Kuhn et al [6] is has been shown that an Anger-logic LaBr 3 (5%Ce) /05/$ IEEE 1919

2 detector with 4x4x30 mm 3 pixels can achieve an average timing resolution of 313 ps fwhm and an average energy resolution of 5.1% fwhm at 511 kev. The crystal array is coupled to a continuous light-guide and hexagonal arrangement of 50-mm diameter PMTs. We chose Photonis XP20D0 PMTs for their combination of timing performance and reasonable cost. The choice of 50-mm PMTs is also one of practicality, since larger PMTs lead to a fewer number for the whole system. Since the light output of LaBr 3 is very high the crystal identification is excellent with large PMTs, and since the decay time is short the probability of pulse pile-up is relatively low. Large detectors were constructed with crystal arrays 27 pixels wide by 60 crystals high. A complete scanner consists of 24 detector modules with 432 PMTs; 6 PMT rows in the axial direction and 72 PMT columns around. Each detector module is coupled to four PMT columns with the two edge columns shared with adjacent detector modules on either side in order to identify pixels at the very edge of the module. The performance of the large detector, in terms if energy and timing resolution is nearly identical to the small proto-type arrays mentioned above and described in more detail in [7]. the rotating line source, each detector pixel in the PET scanner is in coincidence with a reasonably large number of different pixels on the opposite side (see figure 2). The assumption of independence greatly reduces the number of counts that we need to collect for a good estimate. We have considered other source configurations, including fixed sources permanently outside the transverse FOV, but have not yet tested these configurations. Given the location of the source, the true TOF difference between the arrival times of each gamma is computed and compared to the measured TOF difference to generate the TOF difference error. The peak (relative bias) of each pixel s TOF difference error histogram is located by curve fitting. The location of the peak is subtracted from the a priori bias estimate (zero without prior information) and the process is iterated until the crystal timing histograms are all centered about zero. The output is a timing offset correction (final bias) for each crystal that is used to correct the measured timing differences between crystals as they are recorded. This correction is written as a look-up table that can be used either as real-time or off-line timing correction for the list-mode data before reconstruction. Fig. 1. LaBr 3 detector module with 1620 pixels each 4x4x30 mm 3 coupled through light-guide to hexagonal arrangement of 50-mm diameter PMTs. Each detector module is coupled to four PMT columns with the two edge columns shared with adjacent detector modules. With these detectors we have designed a PET scanner with a detector diameter of 93 cm and axial FOV of 25 cm. The predicted sensitivity for this fully 3D system is about 6 cps/kbq, which is very competitive with current PET scanners from GE, Siemens, and Philips Medical Systems using denser scintillators (BGO, LSO, or GSO), but generally these scanners have shorter crystals and shorter axial FOV. Our system is also expected to have a high noise-equivalent countrate due to the fast timing which limits random coincidences, and the energy resolution which reduces scatter and randoms. II. TIMING CALIBRATION METHOD In the calibration technique, the data are written in listmode format with the source location known or encoded in the data for the case of a rotating source. For these initial measurements we used a rotating 18-F filled line source at a 20-cm radius. We assume that the absolute time biases for each pixel on the detector are independent, and so the TOFdifference bias for a given pixel pair is simply the difference in the absolute time biases of each pixel. Therefore, it is not a requirement that each pixel on the detector be measured in coincidence with each other pixel on the detector. By using Fig. 2. Each crystal is in coincidence with a large number of opposing crystals in the transverse direction as the line source rotates (left) which aids in averaging out any timing biases from different trigger regions or detectors. Each crystal is also in coincidence with opposing crystals along the entire axial FOV (right). A 2-D map of this timing offset estimate table for four detectors is shown in Fig. 3. The intensity of the map represents the timing offset for each pixel. The PMT gains have been adjusted to minimize energy variations and optimize the energy resolution, but have not been fully optimized. Only small fluctuations are observed from crystal to crystal and the larger variation arises from the spatiallydependent light collection that is a function of the PMT array pattern. The regions between the PMTs need the largest off-set correction. The entire range (minimum to maximum) of offset factors for the entire scanner is about +/- 1ns. The timing offset calibration was applied to a list-mode collection of a Ge-68 point source positioned at the center of the scanner. The timing resolution of the summed histogram over all crystal pairs is 460ps fwhm. The difference between this measured FWHM value and 313 ps previously reported in [6] are due to losses in the electronics that have not yet been fully optimized in the TOF scanner. 1920

3 Fig D map of the timing offset for 4 detectors. The intensity of the map represents the timing offset for each pixel. Only small fluctuations are observed from crystal to crystal and the larger variation arises from the spatially-dependent light collection that is due to the PMT array pattern. III. SCATTER FRACTION The energy calibration method is similar to that previously developed for non-tof scanners. Using a point source in the center of the FOV, we measure the energy off-sets for each pixel and generate correction table that is applied on-line before writing the list-data to disk. With correction, the energy spectrum for 511 kev is shown in Fig. 4 and demonstrates an overall energy resolution of 7.5% fwhm. Again, the differences between this result and the bench-top measurements with these detectors reported in [6] are due to losses in the electronics that have been adapted for the TOF scanner. The cause of these losses is under investigation. Fig. 5. Scatter fraction vs. lower level energy threshold for the NEMA NU- 2 scatter phantom, 20-cm x 70-cm, as well as two similar phantoms but with 27-cm and 35-cm diameter. The scatter fraction was calculated using the prescribed NEMA NU-2 analysis. Since the scanner operates as a fully 3D scanner and has a large axial FOV of 25 cm, the scatter fraction would be considerably higher if not for the excellent energy resolution, as seen in Fig. 4. This also has impact on the accuracy of the scatter correction. We use a model-based scatter correction method [8] which is based on a single-scatter simulation, and the accuracy of this assumption improves as the lower level energy threshold is raised. At 475 kev the relative number of multiple-scatter events is sufficiently small that they may be scaled from a single-scatter estimate with reasonable accuracy. Fig 6. shows the scatter estimate (SSS) using this model for two water-filled phantoms, 27-cm and 35-cm diameter. 2.9 cm Fig. 4. Energy spectrum of point source in air, representing sum of all pixels in scanner after correction energy offsets. An ELLD of 475 kev corresponds to the edge of the photo-peak. Fig. 4 shows that we can raise the lower level energy discriminator (ELLD) to at least 475 kev. Thus, we performed the NEMA NU-2 scatter measurement [8] with the line source within the 20-cm x 70-cm phantom using this energy threshold. In addition, we performed two similar measurements using annular sleeves surrounding the 20-cm diameter phantom to first increase the diameter to 27-cm and then to 35-cm diameter. This was done to estimate the scatter fraction for larger patients, since the 20-cm diameter phantom is considered to be representative of a thin patient, whereas the 27-cm diameter phantom represents an average patient and the 35-cm diameter phantom represents a large patient. These results are shown in Fig. 5. At 475keV the scatter fraction is 23%, 30%, and 36% for the 20-cm, 27-cm, and 35-cm phantoms, respectively. Fig. 6. Radial profile of water-filled phantom distributions, (left) 27-cm and (right) 35-cm diameter, with calculated scatter estimate (SSS). IV. LESION PHANTOM MEASUREMENTS We performed measurements with two water-filled cylindrical phantoms of diameters 27-cm (vol. = 24 l) and 35- cm (vol. = 53 l) representing an average and heavy patient respectively. A ring of six spheres with diameters 37, 28, 22, 17, 13, and 10-mm was placed at radial distance of 7-cm in the two phantoms and centered axially in scanner for data acquisition. The two large spheres were cold in all measurements. Two data sets were collected with each phantom for activity concentrations of 8:1 and 4:1 in the four small spheres with respect to the background. Additional uniform cylinders were placed next to the lesion phantoms (axially) during measurement with the same activity concentration as that in the lesion phantom in order to simulate activity outside the field-of-view. This configuration is very similar to the IEC image quality phantom prescribed by NEMA NU-2, except for the size of the phantom itself. Data were collected in list-mode with an ELLD = 475 kev applied before image reconstruction. Image reconstruction was 1921

4 performed for different count levels using a fully 3D list-mode TOF reconstruction with blob basis functions [9, 10]. Fig. 7 shows reconstructed images for the two phantoms with 8:1 activity concentration ratio for a clinical equivalent wholebody scan time of 3 mins/frame, which was determined by scaling the activity concentration in the phantom to a corresponding patient dose of 10 mci. Visually these images show that the TOF images converge faster to high contrast values (about 5 iterations), while the non-tof images require more iterations (at least 10 iterations) to converge to similar contrast values but at the expense of increased noise. Note that the reconstruction algorithm [9] uses geometrically ordered subsets, so that the number of iterations required is significantly reduced. For the 35-cm phantom, it is very clear the 10-mm diameter hot sphere has very low contrast and is at the limit of detectability even after 10 iterations in the non- TOF image, while TOF image shows a good contrast and detectability. In fact, the TOF reconstruction images of the 35- cm are similar in quality to the non-tof images of the 27-cm phantom, indicating that with TOF we can achieve results with heavy patients that would not be achievable in a non- TOF scanner. In Fig. 8 we show representative images for data acquired with 4:1 activity concentration ratio for the hot spheres with respect to the background for varying counts represented again as clinical equivalent scan time per frame. In the non- TOF image the 10-mm lesion is barely visible in the 27-cm cylinder but clearly visible in the TOF images. For the 35-cm phantom, the 13-mm lesion is not visible in the non-tof images but clearly visible in the TOF images at least at 3 mins/frame or greater. TOF information clearly leads to better detectability of small lesions, faster convergence, and improved background variability. We can also argue that the TOF allows one to reduce scan time for example, TOF images acquired with 3 mins/frame (or less) are better than the non-tof images acquired with 6 mins/frame. These results are borne out in the quantitative measurement of contrast and background variability measurements performed for the 10-mm sphere using the NEMA NU image quality measurements techniques [8]. A plot of these measurements is shown in Fig. 9. SUMMARY We have developed a proto-type whole-body TOF scanner based on lanthanum bromide that is fully 3D and has an axial FOV of 25 cm. The system is currently operating in a laboratory environment. Thus, it does not yet have gantry covers or a patient bed, so have only performed phantom measurements to date. We will, however, complete the system in the near future so that we can evaluate the TOF benefit with patients. In addition, some aspects of the system are still in development, such as the electronics, so we have not yet achieved the full potential of the TOF performance of these detectors. We have demonstrated that the excellent intrinsic performance of LaBr 3 (5%Ce) can be translated to TOF PET, whereby a favorable combination of sensitivity and spatial resolution are achieved in addition to excellent energy resolution and timing resolution. To date, we have measured an overall system energy resolution of 7.5% fwhm and a system timing resolution of 460 ps fwhm. With the lower level energy threshold set to 475 kev, the scatter fraction is 23% for the NEMA NU-2 phantom which is 20-cm diameter, and have also measured 30% and 36% for 27-cm and 35-cm diameter phantoms, respectively. These additional measurements were acquired to estimate the scatter fraction for average (27-cm) and heavy (35-cm) patients, as well as the thin (20-cm) patient. We have acquired and reconstructed lesion phantoms to begin evaluation of scanner performance with TOF. The data are reconstructed with a list-mode fully 3D iterative algorithm with data corrections included in the system model. We designed the phantoms to be similar to the NEMA NU-2 image quality phantom, but inserted the spheres into a body of 27-cm diameter and 35-cm diameter. As expected we find that TOF reduces the noise and background variability, especially for the larger phantom representing a large patient. In addition, TOF improves detail and contrast of the spheres (lesions), especially the smallest 10-mm sphere. The TOF reconstruction reaches convergence faster than the non-tof reconstruction, and the rate of convergence is seen to be more insensitive to object size. These results indicate that TOF will help improve image quality and potentially reduce scan time with clinical patients. REFERENCES Fig. 9. Plot of contrast vs background variability for the 10-mm hot sphere in the two phantoms with 4:1 activity concentration ratio with respect to background. The points along a line represent increasing number iterations starting with iteration 1 in the bottom left corner (lowest contrast and lowest background variability). [1] T. F. Budinger, "Time-of-Flight positron emission tomography - status relative to conventional PET," J. Nucl. Med., vol. 24, pp , [2] T. Tomitani, "Image-reconstruction and noise evaluation in photon time-of-flight assisted positron emission tomography," IEEE Trans. Nucl. Sci., vol. 28, pp ,

5 [3] S. Surti, J. S. Karp, L. M. Popescu, M. E. Daube-Witherspoon, and M. Werner, Investigation of time-of-flight benefit for fully 3D PET, submitted to IEEE Trans. Med. Imag. [4] E. V. D. van Loef, P. Dorenbos, C. W. E. van Eijk, K. Kramer, and H. U. Gudel, "High-energy-resolution scintillator: Ce3+ activated LaBr3," App Phys Lett, vol. 79, pp , [5] K. S. Shah, J. Glodo, M. Klugerman, W. W. Moses, S. E. Derenzo, and A. J. Weber, "LaBr 3 : Ce scintillators for gamma-ray spectroscopy," IEEE Trans. Nucl. Sci., vol. 50, pp , [6] A. Kuhn, S. Surti, J. S. Karp, P. S. Raby, K. S. Shah, A. E. Perkins, and G. Muehllehner, "Design of a lanthanum bromide detector for timeof-flight PET," IEEE Trans. Nucl. Sci., vol. 51, pp , [7] A. Kuhn, S. Surti, J. S. Karp, G. Muehllehner, F. M. Newcomer, R. van Berg, Performance assessment of pixilated LaBr 3 detector modules for TOF PET, presented at 2004 IEEE Nuclear Science Symposium and Medical Imaging Conference, Rome, Italy, [8] "NEMA Standards Publication NU , Performance Measurements of Positron Emission Tomographs," National Electrical Manufacturers Association, Rosslyn, VA [9] L. M. Popescu, "Iterative image reconstruction using geometrically ordered subsets with list-mode data," presented at IEEE Nuclear Science Symposium and Medical Imaging Conference, Rome [10] L. M. Popescu and R. M. Lewitt, "Tracing through a grid of blobs," presented at IEEE Nuclear Science Symposium and Medical Imaging Conference, Rome, Italy, Fig. 7. Representative reconstructed images for the two phantoms for a fixed acquisition time of 3 mins/frame and an activity concentration ratio of 8:1 in the hot lesions with respect to the background. Fig. 8. Representative reconstructed images for the two phantoms for varying acquisition time and an activity concentration ratio of 4:1 in the hot lesions with respect to the background. 1923

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Time-of-Flight Technology

Time-of-Flight Technology Medical Review Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems INTRODUCTION Improving the care for every patient while providing a high standard care to

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9. NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2010 February ; 51(2): 237. doi:10.2967/jnumed.109.068098. An Assessment of the Impact of Incorporating Time-of-Flight

More information

An educational tool for demonstrating the TOF-PET technique

An educational tool for demonstrating the TOF-PET technique Nuclear Instruments and Methods in Physics Research A 471 (2001) 200 204 An educational tool for demonstrating the TOF-PET technique T.Bȧack a, *, J. Cederkȧall a, B. Cederwall a, A. Johnson a, A. Kerek

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

Performance Evaluation of the Philips Gemini PET/CT System

Performance Evaluation of the Philips Gemini PET/CT System Performance Evaluation of the Philips Gemini PET/CT System Rebecca Gregory, Mike Partridge, Maggie A. Flower Joint Department of Physics, Institute of Cancer Research, Royal Marsden HS Foundation Trust,

More information

Performance Evaluation of radionuclide imaging systems

Performance Evaluation of radionuclide imaging systems Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application

More information

Extremely Fast Detector for 511 kev Gamma

Extremely Fast Detector for 511 kev Gamma Extremely Fast Detector for 511 kev Gamma V. Sharyy, D. Yvon, G. Tauzin, E.Delagnes, Ph. Abbon, J P. Bard, M. Kebbiri, M. Alokhina, C. Canot IRFU, CEA D. Breton, J. Maalmi LAL,IN2P3 Journée 2015 du Labex

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015 Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!

More information

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

Recent advances and future advances in time-of-flight PET

Recent advances and future advances in time-of-flight PET ARTICLE IN PRESS Nuclear Instruments and Methods in Physics Research A 580 (2007) 919 924 www.elsevier.com/locate/nima Recent advances and future advances in time-of-flight PET William W. Moses Lawrence

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET

Evaluation of Centrally Located Sources in. Coincidence Timing Calibration for Time-of-Flight PET Evaluation of Centrally Located Sources in Coincidence Timing Calibration for Time-of-Flight PET by Richard Ryan Wargo Graduate Program in Medical Physics Duke University Date: Approved: Timothy G. Turkington,

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

In-vivo dose verification for particle therapy

In-vivo dose verification for particle therapy In-vivo dose verification for particle therapy D.R. Schaart, NCS Lustrum, 5-Oct-2012 HollandPTC 1 Protons vs. photons Photons Protons Dennis R. Schaart Delft University of Technology 2 Protons: the promise

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET

Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies. TOF scanners. Clinical benefit of TOF PET Future Prospects of Time-of-Flight PET: Scintillator, Detector and Depth-of- interaction (DOI) Technologies Suleman Surti Physics & Instrumentation Group Department of Radiology University of Pennsylvania

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

Positron Emission Tomography

Positron Emission Tomography Physics 656 Seminar on Physical Fundamentals of Medical Imaging Positron Emission Tomography Ahmed Qamesh Outline What is PET? PET mechanism Radionuclide and its synthesis Detection concept and Development

More information

ISO ISO ISO OHSAS ISO

ISO ISO ISO OHSAS ISO ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,

More information

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT 2005 IEEE Nuclear Science Symposium Conference Record M03-271 Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT Stéphan Blinder, Marie-Laure Camborde, Ken R. Buckley,

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997

218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 218 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 2, APRIL 1997 Compton Scatter and X-ray Crosstalk and the Use of Very Thin Intercrystal Septa in High-Resolution PET Detectors Craig S. Levin, Member,

More information

Simulations in emission tomography using GATE

Simulations in emission tomography using GATE Simulations in emission tomography using GATE Irène Buvat buvat@imed.jussieu.fr Laboratory of Functional Imaging, U678 INSERM, Paris, France Outline Emission tomography and need for simulations GATE short

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

Quantitative capabilities of four state-of-the-art SPECT-CT cameras

Quantitative capabilities of four state-of-the-art SPECT-CT cameras Seret et al. EJNMMI Research 2012, 2:45 ORIGINAL RESEARCH Open Access Quantitative capabilities of four state-of-the-art SPECT-CT cameras Alain Seret 1,2*, Daniel Nguyen 1 and Claire Bernard 3 Abstract

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 Count-Rate Dependent Component-Based Normalization for the HRRT Mario Rodriguez, Jeih-San Liow, Member, IEEE, Shanthalaxmi Thada, Merence

More information

Addition of Noise by Scatter Correction Methods in PVI

Addition of Noise by Scatter Correction Methods in PVI IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 4, AUGUST 1994 1551 Addition of Noise by Scatter Correction Methods in PVI.J. Scott Barneyt, Ronald Harropl and M. Stella Atkinsi t Div. of Nuclear Medicine,

More information

Simultaneous measurement of noise and spatial resolution in PET phantom images

Simultaneous measurement of noise and spatial resolution in PET phantom images IOP PUBLISHING Phys. Med. Biol. 55 (2010) 1069 1081 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/55/4/011 Simultaneous measurement of noise and spatial resolution in PET phantom images Martin

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY

More information

3/19/2013. Outline. Objective. SPECT System Description. Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system

3/19/2013. Outline. Objective. SPECT System Description. Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system Calibration of an MR-Compatible, CZT Detector Based Stationary Small Animal SPECT system Andrew Rittenbach 1, Jingyan Xu 1, AbdEl-Monem El-Sharkawy 1, William Edelstein 1, Kevin Parnham 2, James Hugg 2,

More information

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy Physics Department, Faculty of Applied Science,Umm Al-Qura

More information

A Further Study on Large Size LSO and LYSO Crystal Samples

A Further Study on Large Size LSO and LYSO Crystal Samples October 25, 2005 IEEE NSS, N12-6, Puerto-Rico 1 A Further Study on Large Size LSO and LYSO Crystal Samples Jianming Chen, Liyuan Zhang, Ren-yuan Zhu California Institute of Technology BGO, LSO & LYSO Samples

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,500 108,000 1.7 M Open access books available International authors and editors Downloads Our

More information

Effects of system geometry and other physical factors on photon sensitivity of high-resolution

Effects of system geometry and other physical factors on photon sensitivity of high-resolution Home Search Collections Journals About Contact us My IOPscience Effects of system geometry and other physical factors on photon sensitivity of high-resolution positron emission tomography This article

More information

PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS

PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS NEMA NU 2 PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS NEMA Standards Publication NU 2-2001 Performance Measurements of Positron Emission Tomographs Published by National Electrical Manufacturers

More information

AX-PET : A novel PET concept with G-APD readout

AX-PET : A novel PET concept with G-APD readout AX-PET : A novel PET concept with G-APD readout Matthieu Heller CERN - PH/DT Marie Curie network MC-PAD Matthieu.heller@cern.ch On behalf of the AX-PET collaboration https://twiki.cern.ch/twiki/bin/view/axialpet

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Q.Clear. Steve Ross, Ph.D.

Q.Clear. Steve Ross, Ph.D. Steve Ross, Ph.D. Accurate quantitation (SUV - Standardized Uptake Value) is becoming more important as clinicians seek to utilize PET imaging for more than just diagnosing and staging disease, but also

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

2007 IEEE Nuclear Science Symposium Conference Record M26-296

2007 IEEE Nuclear Science Symposium Conference Record M26-296 2007 IEEE Nuclear cience ymposium Conference Record M26-296 A Global and a egmented Plane catter Calibration: Improving the Quantitative Accuracy of Frames with High Random Fraction and/or Low Number of

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction 3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg, Member, IEEE, and Jeffrey A. Fessler, Member, IEEE Abstract- we

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies 2005 IEEE Nuclear Science Symposium Conference Record M07-218 Comparison of 3D PET data bootstrap resampling methods for numerical observer studies C. Lartizien, I. Buvat Abstract Bootstrap methods have

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm

Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm 548 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 5, MAY 2000 Unmatched Projector/Backprojector Pairs in an Iterative Reconstruction Algorithm Gengsheng L. Zeng*, Member, IEEE, and Grant T. Gullberg,

More information

GPU-Accelerated Forward and Back-Projections with Spatially Varying Kernels for 3D DIRECT TOF PET Reconstruction

GPU-Accelerated Forward and Back-Projections with Spatially Varying Kernels for 3D DIRECT TOF PET Reconstruction 1 GPU-Accelerated Forward and Back-Projections with Spatially Varying Kernels for 3D DIRECT TOF PET Reconstruction S. Ha, S. Matej, Senior Member, IEEE, M. Ispiryan, Member, IEEE and K. Mueller, Senior

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 641 (211) 121 127 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

State of the art and challenges of time-of-flight PET

State of the art and challenges of time-of-flight PET Physica Medica (2009) 25, 1e11 available at www.sciencedirect.com journal homepage: http://intl.elsevierhealth.com/journals/ejmp REVIEW PAPER State of the art and challenges of time-of-flight PET Maurizio

More information

The great interest shown toward PET instrumentation is

The great interest shown toward PET instrumentation is CONTINUING EDUCATION PET Instrumentation and Reconstruction Algorithms in Whole-Body Applications* Gabriele Tarantola, BE; Felicia Zito, MSc; and Paolo Gerundini, MD Department of Nuclear Medicine, Ospedale

More information

Detection of Lesions in Positron Emission Tomography

Detection of Lesions in Positron Emission Tomography Detection of Lesions in Positron Emission Tomography Bachelor Thesis Nina L.F. Bezem Study: Physics and Astronomy Faculty of Science Supervised by: Dr. Andre Mischke Utrecht University, Institute for Subatomic

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

Emission Computed Tomography Notes

Emission Computed Tomography Notes Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon

More information

QIBA PET Amyloid BC March 11, Agenda

QIBA PET Amyloid BC March 11, Agenda QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998

arxiv:physics/ v1 [physics.ins-det] 18 Dec 1998 Studies of 1 µm-thick silicon strip detector with analog VLSI readout arxiv:physics/981234v1 [physics.ins-det] 18 Dec 1998 T. Hotta a,1, M. Fujiwara a, T. Kinashi b, Y. Kuno c, M. Kuss a,2, T. Matsumura

More information

In the short span since the introduction of the first

In the short span since the introduction of the first Performance Characteristics of a Newly Developed PET/CT Scanner Using NEMA Standards in 2D and 3D Modes Osama Mawlawi, PhD 1 ; Donald A. Podoloff, MD 2 ; Steve Kohlmyer, MS 3 ; John J. Williams 3 ; Charles

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61675-2 First edition 1998-01 Radionuclide imaging devices Characteristics and test conditions Part 2: Single photon emission computed tomographs Dispositifs d imagerie par radionucléides

More information

Delayline Detectors. Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution. e - I +

Delayline Detectors. Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution. e - I + Delayline Detectors Imaging Detection of Electrons, Ions & Photons with Picosecond Time Resolution e - I + Delayline Readout of MCPs - The Technical Approach - Microchannel-Plate (MCP) detectors provide

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 621 (2010) 590 594 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging The Hybrid Parallel Plates Gas Counter for Medical Imaging F. Anulli, G. Bencivenni, C. D Ambrosio, D. Domenici, G. Felici, F. Murtas Laboratori Nazionali di Frascati - INFN, Via E. Fermi 40, I-00044 Frascati,

More information

SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder. Colleen Marie Allen. Graduate Program in Medical Physics Duke University

SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder. Colleen Marie Allen. Graduate Program in Medical Physics Duke University SUV Analysis of F-18 FDG PET Imaging in the Vicinity of the Bladder by Colleen Marie Allen Graduate Program in Medical Physics Duke University Date: Approved: Timothy Turkington, Supervisor Terence Wong

More information

Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall

Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu (fcdh@stanford.edu), CS 229 Fall 2014-15 1. Introduction and Motivation High- resolution Positron Emission Tomography

More information

Incorporating HYPR de-noising within iterative PET reconstruction (HYPR-OSEM)

Incorporating HYPR de-noising within iterative PET reconstruction (HYPR-OSEM) Home Search Collections Journals About Contact us My IOPscience Incorporating HYPR de-noising within iterative PET reconstruction (HYPR-OSEM) This content has been downloaded from IOPscience. Please scroll

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS

FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS Romanian Reports in Physics, Vol. 62, Nos. 1, P. 27 36, 2010 FUZZY CLUSTERING ALGORITHM FOR GAMMA RAY TRACKING IN SEGMENTED DETECTORS G. SULIMAN, D. BUCURESCU Horia Hulubei National Institute for Physics

More information

SINGLE-PHOTON emission computed tomography

SINGLE-PHOTON emission computed tomography 1458 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 4, AUGUST 2012 SPECT Imaging With Resolution Recovery Andrei V. Bronnikov Abstract Single-photon emission computed tomography (SPECT) is a method

More information

A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms

A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms A Weighted Least Squares PET Image Reconstruction Method Using Iterative Coordinate Descent Algorithms Hongqing Zhu, Huazhong Shu, Jian Zhou and Limin Luo Department of Biological Science and Medical Engineering,

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System Shinji Ohyama, Masato Mukouyama Graduate School

More information

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 20, NO. 7, JULY 2001 633 The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data Ron Levkovitz, Dmitry Falikman*, Michael Zibulevsky,

More information

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques IMAGING A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques Lai K. Leong, Randall L. Kruger, and Michael K. O Connor Section of Nuclear Medicine, Department

More information

The PET components in the most recent generation

The PET components in the most recent generation Impact of Time-of-Flight on PET Tumor Detection Dan J. Kadrmas 1, Michael E. Casey 2, Maurizio Conti 2, Bjoern W. Jakoby 2 4, Cristina Lois 5, and David W. Townsend 3 1 Department of Radiology, University

More information

y z x SNR(dB) RMSE(mm)

y z x SNR(dB) RMSE(mm) a b c d y z x SNR(dB) RMSE(mm) 30.16 2.94 20.28 6.56 10.33 13.74 0.98 25.37 Supplementary Figure 1: System performance with different noise levels. The SNRs are the mean SNR of all measured signals corresponding

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2009 December 1.

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2009 December 1. NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2009 August ; 50(8): 1315 1323. doi:10.2967/jnumed.109.063016. Impact of Time-of-Flight on PET Tumor Detection Dan J.

More information