1 Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy
2 ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf Collimator (mmlc) generates precise and fine field shapes for state-ofthe-art stereotactic radiation treatments of brain and head and neck tumors, as well as other brain lesions such as arteriovenous malformations. The ModuLeaf mmlc s field size and precision also permit it to be used for extra-cranial 3D conformal and Intensity-Modulated Radiation Therapy (IMRT) treatments. The ModuLeaf mmlc provides unique advantages to radiation therapy departments. With a 2.5 width at the isocenter for all 80 leaves and a maximum field size 1 of 10 cm x 12 cm, ModuLeaf is one of the finest resolution mini MLCs available in the industry today Compatibility with most major linear accelerators and treatment planning systems makes it an economical and easy upgrade on current equipment Sophisticated and cutting-edge design developed in close cooperation with the German Cancer Research Center (DKFZ Heidelberg) to ensure ease of use and solid reliability in the clinical routine Demountable design provides clinical flexibility Features and Clinical Benefits In less than five minutes, the ModuLeaf mmlc is easily mounted onto the linear accelerator accessory holder with a motorized trolley. With this quick step, a Siemens linear accelerator becomes a highly accurate treatment delivery machine that performs similar to dedicated radiosurgery systems. When mounted on a Siemens linear accelerator equipped with a 550 TxT Treatment Table, the ModuLeaf mmlc combines its 2.5 leaf projection with exceptional isocentric specifications and provides a sharp tool to irradiate the tumor while sparing the healthy surrounding tissues. The ModuLeaf is a versatile add-on mmlc for 3D conformal IMRT and stereotactic applications. The ModuLeaf mmlc also supports high-resolution IMRT delivery. Its large over-travel and outstanding 0.5 leaf positioning accuracy ensure that the dose distribution effectively delivered to a patient matches the planned dose distribution. Extremely low leakage minimizes unnecessary dose to healthy tissue. A field size of 10 cm x 12 cm enables the ModuLeaf mmlc to be used in a wide variety of anatomical sites. 80 leaves 2.5 leaf width at isocenter when mounted on a Siemens linear accelerator 10 cm x 12 cm maximum field size at isocenter Low interleaf leakage Fast mounting process Note: all specifications at isocenter vary with the actual sourceto-collimator distance. 1. When mounted on a Siemens linear accelerator.
3 ModuLeaf Mini Multileaf Collimator Treatment Applications The ModuLeaf mmlc is a versatile and powerful tool that delivers a precise radiation dose to the target volume while sparing critical tissue. The design allows it to be used in a wide array of clinically valuable treatment applications offering a high precision treatment to a larger patient population. 3D Conformal High leaf resolution and precise leaf positioning enable the user to create more conformal shapes over a range of possible field sizes. In combination with minimal leakage, it provides a superior technique for conformal radiation therapy. IMRT The ModuLeaf mmlc was designed for IMRT. The extensive over-center travel of the leaves, fast positioning of a shape, and low leakage make it ideal for unique field shapes required by IMRT. The ModuLeaf mmlc mounts on the accessory holder of a Siemens linear accelerator. Stereotactic The fine leaf width and accurate leaf positioning allow precise beam shaping for small intracranial and extracranial target volumes, where highest precision is required. The ModuLeaf mmlc is ideal for stereotactic radiotherapy and stereotactic radiosurgery applications. Dose distribution achieved with IMRT, using the ModuLeaf mmlc = Creating an intensity-modulated beam with a series of irregular fields.
4 System Specifications MLC Unit Number of Leaves 80 Leaf Positioning Speed 2 20 /sec at isocenter Leaf Material Tungsten Leaf Over-center Travel 1 55 at isocenter Leaf Height 7 cm Positioning Accuracy 0.5 Max Field Size 1,2 10 cm x 12 cm at isocenter Housing Material Carbon Fiber Leaf Width 1,2 2.5 at isocenter 1. All specifications at isocenter vary with the actual source-to-collimator distance. Details are specified in the isocentric geometric specifications. 2. When the ModuLeaf mmlc is mounted on a Siemens linear accelerator. Weight kg Length Width Height Dimensions Focusing Characteristics of Leaves The ModuLeaf mmlc employs single-focused leaves. The cross section perpendicular to leaf travel direction is focused (i.e., the leaf side surfaces follow the beam divergence). To reduce leakage between leaves (interleaf leakage), the leaves are slightly de-focused (i.e., the leaf side surfaces are focused toward a virtual source offset by a few millimeters from the X-ray source). The leaf ends are rounded to achieve a position-independent penumbra over the entire travel range. Isocentric Geometric Specifications Linear Accelerator Maximum Field Size X Maximum Field Size Y Maximum Overtravel Leaf Width Positioning Accuracy Siemens (All Models) Varian Clinac 1 with MLC Varian Clinac without MLC Clinac 600C, 1800C, 2100C Adaptation Plates Linear Accelerator Manufacturer Linear Accelerator Model Leaf Travel Direction (IEC 61217) Weight kg Siemens All Models X 1.7 Varian All Models except 6/100 X 4.6
5 ModuLeaf Mini Multileaf Collimator Isocenter Clearance Siemens Linear Accelerators Varian Linear Accelerators without MLC 30 cm 35 cm The clearance between the MLC and the isocenter depends on the linear accelerator and its accessory holder. The clearance is between 30 and 35 cm. Data for specific linear accelerators available upon request. mmlc Trolley Dimensions Weight kg Length Width Height ~ In less than five minutes, the ModuLeaf mmlc is mounted and ready for use. Dosimetric Specifications and Measurements Transmission Through Leaves (intraleaf) % Transmission Between Leaves (interleaf) % 80/20% Penumbra (5 cm x 5 cm field) Specification <2 <2.5 <4 Typical Value <2 ~3 1. Measurements were performed on a PRIMUS Linear Accelerator from Siemens at 6 MV, reference: Hartmann, G. and Föhlisch, F. Dosimetric characterization of a new miniature multileaf collimator. Phys. Med. Biol. 47: N , leaf width at isocenter for all 80 leaves provides a maximum field size of 10 cm x 12 cm.
6 Interfaces to Linear Accelerators The ModuLeaf mmlc is compatible with most coercially available linear accelerators. Depending on the linear accelerator model and V&R system, there are different levels of integration available. The usage of Basic Modes have no limitations on linear accelerator models, but requires more interaction from the user during a treatment session. The Automatic and Synchronized Modes have been developed for IMRT treatments. This feature allows verification of treatment setup between ModuLeaf mmlc and the linear accelerator, as well as the execution and verification of step and shoot IMRT treatments. Siemens Linear Accelerators (All Models) Supported R&V Supported Treatment Mode Requirements Treatment Plan Export To Supported File Formats (Basic Mode) Supported File Formats (Automatic Mode and Synchronized Mode) LANTIS 5.2 or above Basic and Automatic Step and Shoot DMIP 6 or 9, SIMTEC -AFS PRIMEVIEW 1.5 or higher, Control Console software version 6.5/7.2 or higher ModuLeaf RTPLink DICOM RT RTPLink DICOM RT (with release 2.3) Sophisticated and cutting-edge, designed in cooperation with the German Cancer Research Center (DKFZ) in Heidelberg, Germany.
7 ModuLeaf Mini Multileaf Collimator ModuLeaf Controller Specifications Controller Workstation Power Supply Computer PC with integrated real time computer Mains Voltage 115/230 V~ Processor 32 bit CPU (Intel PIII, 1.6 G Hz or more) Supply Current 6 A/3 A Network Card 10 baset 10/100 Mbit Primary Fuses 6 AT/3.15 AT Network Protocol TCP/IP Secondary Fuses 3 AT Network Cable BNC, RJ 45 Supply Frequency 50/60 Hz Screen Resolution Min x 1024 pixels Monitor 1.5 A Trolley Charger 0.5 A MLC Controller Workstation Dimensions Weight kg Length Width Height ~ Environmental Conditions The ModuLeaf mmlc may be stored for a maximum period of 10 weeks. Transportation and Storage Ambient Temperature Range Relative Humidity Atmospheric Pressure -10 C to +40 C (14 F to 104 F) 10% to 75% 500 hpa to 1060 hpa Operation +10 C to +40 C 30% to 75% 700 hpa to 1060 hpa Mounted on a Siemens linear accelerator, the ModuLeaf mmlc provides a 30 cm isocenter clearance.
8 On account of certain regional limitations of sales rights and service availability, we cannot guarantee that all products included in this brochure are available through the Siemens sales organization worldwide. Availability and packaging may vary by country and are subject to change without prior notice. Some/All of the features and products described herein may not be available in the United States. The information in this document contains general technical descriptions of specifications and options, as well as standard and optional features which do not always have to be present in individual cases. Siemens reserves the right to modify the design, packaging, specifications, and options described herein without prior notice. Please contact your local Siemens sales representative for the most current information. Note: Any technical data contained in this document may vary within defined tolerances. Original images always lose a certain amount of detail when reproduced. ModuLeaf, 550 TxT, PRIMEVIEW, SIMTEC, PRIMUS, and LANTIS are trademarks of Siemens AG. Siemens Medical Solutions USA, Inc. Oncology Care Systems 4040 Nelson Avenue Concord, California USA Headquarters Siemens Medical Solutions USA 51 Valley Stream Parkway Malvern, PA USA Tel.: Contact Addresses Germany Siemens Medical Solutions Oncology Care Systems - Heidelberg Hans-Bunte-Str. 10 D Heidelberg Germany Contact Address Siemens Medical Solutions USA, Inc. Oncology Care Systems 4040 Nelson Avenue Concord, California USA Tel.: (US and Canada) , Siemens AG Order No. A91004-M2690-G A00 Printed in USA OCS-059 VH KL CL 1.5M
Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More informationThe MSKCC Approach to IMRT. Outline
The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification
More informationDosimetric impact of the 160 MLC on head and neck IMRT treatments
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Dosimetric impact of the 160 MLC on head and neck IMRT treatments Prema Rassiah-Szegedi, a Martin Szegedi, Vikren Sarkar, Seth Streitmatter,
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationAn Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy
An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationCalibration and quality assurance for rounded leaf-end MLC systems
Calibration and quality assurance for rounded leaf-end MLC systems Maria N. Graves, a) Antoinette V. Thompson, b) Mary K. Martel, Daniel L. McShan, and Benedick A. Fraass Department of Radiation Oncology,
More informationUse of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta
Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationMAMMOMAT 1000 / 3000 Nova. Accessories. Accessories (optional) Compression plate with low edge 18 cm x 24 cm #
MAMMOMAT 1000 / 3000 Nova Accessories Accessories (optional) Compression plate with low edge 18 cm x 24 cm # 44 94 378 Standard compression plate with 40 mm high front edge. Compression plate with high
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationEvaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG)
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG) Lalith K. Kumaraswamy,
More informationQalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004
Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence
More informationInvestigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers
Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy
More informationDose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them
Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them
More informationA method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy a
A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy a Mark R. Arnfield, b) Jeffrey V. Siebers, Jong O. Kim, Qiuwen Wu, Paul J. Keall, and
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationComputational modelling and analysis: IOP Medical Physics Group
Computational modelling and analysis: IOP Medical Physics Group Using Bespoke Matlab Software for the Collection and Processing of Dynalog Files from the Varian Millennium 120 MLC system following delivery
More informationPyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system
Journal of Physics: Conference Series OPEN ACCESS PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system To cite this article: Aitang Xing et al 2014
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationAssesing multileaf collimator effect on the build-up region using Monte Carlo method
Pol J Med Phys Eng. 2008;14(3):163-182. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0014-0 website: http://www.pjmpe.waw.pl M. Zarza Moreno 1, 2, N. Teixeira 3, 4, A. P. Jesus 1, 2, G. Mora 1 Assesing multileaf
More informationDX-D 300. Flexible Direct Radiography System
DX-D 300 Flexible Direct Radiography System MUSICA 2 processing provides superior contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationMED MI 1005 Network Isolators
MED MI 1005 Network Isolators General Network isolators protect patients and medical devices from excessive leakage currents that may flow through the data network. Every device connected to the network
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More informationPosition accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon
Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell
More information7/29/2017. Making Better IMRT Plans Using a New Direct Aperture Optimization Approach. Aim of Radiotherapy Research. Aim of Radiotherapy Research
Making Better IMRT Plans Using a New Direct Aperture Optimization Approach Dan Nguyen, Ph.D. Division of Medical Physics and Engineering Department of Radiation Oncology UT Southwestern AAPM Annual Meeting
More informationAccuracy study. Elekta Synergy S. High precision radiation therapy using Elekta Synergy S. UMC, Utrecht, Netherlands. Institution: Purpose:
S T E R E O T A C T I C R A D I A T I O N T H E R A P Y Elekta Synergy S High precision radiation therapy using Elekta Synergy S Institution: UMC, Utrecht, Netherlands Purpose: The primary application
More informationAnalysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code
American Journal of Biomedical Engineering 216, 6(4): 124-131 DOI: 1.5923/j.ajbe.21664.3 Analysis of Radiation Transport through Multileaf Collimators Using BEAMnrc Code Ankit Kajaria 1,*, Neeraj Sharma
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Evaluation of dosimetric effect of leaf position in a radiation field of an 80-leaf multileaf collimator fitted to the LINAC
More informationAdvanced Radiotherapy
Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationIterative regularization in intensity-modulated radiation therapy optimization. Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006.
Iterative regularization in intensity-modulated radiation therapy optimization Carlsson, F. and Forsgren, A. Med. Phys. 33 (1), January 2006. 2 / 15 Plan 1 2 3 4 3 / 15 to paper The purpose of the paper
More informationRadiation therapy plan checks in a paperless clinic
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Radiation therapy plan checks in a paperless clinic R. Alfredo Siochi, (a) Edward C. Pennington, Timothy J. Waldron, John E.
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationAn experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking
An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking G A Davies, P Clowes, D McQuaid, P M Evans, S Webb and G Poludniowski Joint Department of Physics, Institute
More informationDX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300
DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM MUSICA processing provides excellent contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for
More informationMonaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application
Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationA fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study
A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationFocalSim DICOM Conformance Statement. Table of Contents:
FocalSim DICOM Conformance Statement The following is the statement of conformance for the CMS FocalSim product for DICOM 3.0. This conformance statement does not apply to other products. If you have any
More informationAn Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields
Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
Technical Note Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K.
More informationFLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300
FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 The DX-D 300 DR system unites excellent image quality with complete convenience. It offers top-of-the-line technology, a single detector and a fully-motorized
More informationDX-D 300. Flexible Direct Radiography System
DX-D 300 Flexible Direct Radiography System MUSICA² processing provides outstanding contrast detail and consistently excellent image quality Universal, flexible and affordable modality combines a single
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationCommissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility
0 0 0 0 JMP R Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K. D.
More informationHidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.
2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More informationMCNP4C3-BASED SIMULATION OF A MEDICAL LINEAR ACCELERATOR
Computational Medical Physics Working Group Workshop II, Sep 3 Oct 3, 7 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (7) MCNP4C3-BASED SIMULATION
More informationCreating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience
DVH Estimates Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience Karen Chin Snyder, MS, DABR AAMD Region V Meeting October 4, 2014 Disclosures The Department of Radiation Oncology
More informationData. MAGNETOM Symphony, A Tim System Matrix Coils
Data MAGNETOM Symphony, A Tim System Matrix Coils Head Matrix Coil The Head Matrix coil is part of the standard system configuration. 12-element design with 12 integrated preamplifiers, two rings of 6
More informationTHE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE SIMULATION OF
More informationTEPZZ Z754_7A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: A61N 5/10 ( )
(19) TEPZZ Z74_7A_T (11) EP 3 07 417 A1 (12) EUROPEAN PATENT APPLICATION (43) Date of publication: 0..16 Bulletin 16/ (1) Int Cl.: A61N / (06.01) (21) Application number: 16163147.8 (22) Date of filing:
More informationSuggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 3, Summer 2010 Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems Sanaz Hariri,
More informationDetermination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: Implications for 4D radiotherapy
Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: Implications for 4D radiotherapy K. Wijesooriya, C. Bartee, J. V. Siebers, S. S. Vedam, and P. J. Keall a Department
More informationThe Dose Junction Issue Associated with Photon Beams for Large Volume Radiation Therapy and the Sensitivity to Set-up Error
Research Article imedpub Journals http://www.imedpub.com/ Journal of Medical Physics and Applied Sciences The Dose Junction Issue Associated with Photon Beams for Large Volume Radiation Therapy and the
More informationMeasurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code
reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 64 68 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Original article Measurement of depth-dose of linear accelerator
More informationDigital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy
Author manuscript, published in "Medical Imaging 008: Physics of Medical Imaging, San Diego, CA, USA : United States (008)" DOI : 10.1117/1.768668 Digital phantoms for the evaluation of a software used
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Measurement and Monte Carlo simulation for energy- and intensity-modulated electron radiotherapy delivered by a computer-controlled
More informationDX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM
DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM The DX-D 300 DR system unites excellent image quality with the ultimate convenience. It offers top-of-the-line technology: a fullymotorized positioner enabling
More informationAgility MLC transmission optimization in the Monaco treatment planning system
Received: 16 January 2018 Revised: 19 April 2018 Accepted: 1 June 2018 DOI: 10.1002/acm2.12399 RADIATION ONCOLOGY PHYSICS Agility MLC transmission optimization in the Monaco treatment planning system Michael
More informationTOMOHD. Product Specifications
TOMOHD Treatment System Product Specifications 1. Complete Treatment System Dimensions Height Width Length Weight 8 ft., 3.25 in. (252.1 cm) 9 ft., 2.50 in. (280.6 cm) 15 ft., 2.25 in. (463.0 cm) 10,000
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationAn experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors
More informationA DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL
A DOSIMETRIC MODEL FOR SMALL-FIELD ELECTRON RADIATION THERAPY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE MASTER OF
More informationDX-D 45. The easiest way to go instant DR
DIGITAL AED DETECTOR DX-D 45 The DX-D 45 Digital Detector with Automatic Exposure Detection (AED) offers a fast and effective way for radiography facilities to benefit from high quality digital imaging
More informationUsing multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 1, WINTER 2007 Using multileaf collimator interleaf leakage to extract absolute spatial information from electronic portal imaging device images
More informationI. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)
More informationPlanmed Sophie Classic Mobile. English. dedicated for mobile mammography
Sophie Classic Mobile English at its best Sophie Classic Mobile proven mammography on wheels The Sophie Classic Mobile system takes transportable mammography to a new level. Generously equipped with integrated
More informationUsing a research real-time control interface to go beyond dynamic MLC tracking
in partnership with Using a research real-time control interface to go beyond dynamic MLC tracking Dr. Simeon Nill Joint Department of Physics at The Institute of Cancer Research and the Royal Marsden
More informationSIPLUS S7-200 PS 203. Overview
SIPLUS S7-200 PS 203 Overview Design and functionality of the power supply are optimally adapted to the SIPLUS S7-200 micro PLC Slim design Particularly suitable for low cabinet depths Note: SIPLUS extreme
More informationMonte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Monte Carlo modeling of a Novalis TX Varian 6 MV with HD-120 multileaf collimator Luis Alberto Vazquez-Quino, 1 Brian Massingill,
More informationRadiation therapy treatment plan optimization
H. Department of Industrial and Operations Engineering The University of Michigan, Ann Arbor, Michigan MOPTA Lehigh University August 18 20, 2010 Outline 1 Introduction Radiation therapy delivery 2 Treatment
More informationInvestigation of photon beam output factors for conformal radiation therapy Monte Carlo simulations and measurements
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N133 N143 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)32395-9 NOTE Investigation of photon beam output factors for conformal radiation therapy
More information8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA
PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More informationGPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC
GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Product: Incorrect Rotation of the Collimator or Couch ERGO Scope: Sites affected will be those: 1. Running ERGO version 1.7.3 and higher and, 2. Using a Multileaf Collimator (MLC) device for
More informationOn compensator design for photon beam intensity-modulated conformal therapy
On compensator design for photon beam intensity-modulated conformal therapy Steve B. Jiang a) and Komanduri M. Ayyangar Department of Radiation Therapy, Medical College of Ohio, 3000 Arlington Avenue,
More informationChapter 9 Field Shaping: Scanning Beam
Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement
More informationIAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques
IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures
More informationOverview. Digital X-ray imaging
Overview Digital X-ray imaging OR Technology Headquarters The OR Technology competence group: Your competent partner for digital X-ray imaging Producing digital X-ray technology and developing image management
More informationDetermination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation
Determination of primary electron beam parameters in a Siemens Primus Linac using Monte Carlo simulation Danial Seifi Makrani 1, Hadi Hasanzadeh 2*, Tayyeb Allahverdi Pourfallah 3, Arash Ghasemi 4, Majid
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationA Mimetic Algorithm for Simultaneous Multileaf Collimator Aperture Shape and Dosimetric Optimization in CyberKnife Robotic Radiosurgery
A Mimetic Algorithm for Simultaneous Multileaf Collimator Aperture Shape and Dosimetric Optimization in CyberKnife Robotic Radiosurgery Matthew R. Witten, Owen C. Clancey, Division of Radiation Oncology,
More informationMech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University
Mech. Engineering, Comp. Science, and Rad. Oncology Departments Schools of Engineering and Medicine, Bio-X Program, Stanford University 1 Conflict of Interest Nothing to disclose 2 Imaging During Beam
More informationDX-D 600 DX-D 600 HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY, DIRECT RADIOGRAPHY SYSTEM WITH STATE-OF-THE-ART DESIGN IN THREE CONFIGURATION OPTIONS.
D i r e c t r a d i o g r a p h y s y s t e m DX-D 600 HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY, DIRECT RADIOGRAPHY SYSTEM WITH STATE-OF-THE-ART DESIGN IN THREE CONFIGURATION OPTIONS. Two-detector, high-productivity,
More informationImplementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions
More informationINTEGRATED DATA MIGRATION:
Technical Perspectives Integrated data migration INTEGRATED DATA MIGRATION: Patient record transfer with discrete RT data elements ARIA Database DATA MIGRATION Third-party Database The purpose of this
More information