1 6-Series Sonde Quick Start Guide This document will assist an experienced user with the steps required to make your YSI 6-Series product operational. For full instructions, please refer to the 6-Series User Manual. Unpacking and Inspection Inspect the outside of the shipping container(s) for damage. If you see any damage, contact your shipping carrier immediately. Remove the equipment from the shipping box. Some parts are loose in the box, so check the packing material carefully. Check off all items on the packing list and inspect them for damage. Items Needed Sonde Probes Calibration cup Probe guard Batteries (if sonde has battery compartment) DI water Flathead screwdriver Calibration standards Communication or field cable Preparing Sonde for Use To prepare the sonde for operation, you will need to: Install a new membrane on the YSI 6562 DO Probe, if you are using this item. Install probes (sensors) into the connectors on the sonde bulkhead. Provide power for the sonde through batteries or line power. Connect a field cable or PC communication cable.! It is important to calibrate all sensors prior to using the sonde for field measurements. Refer to Section of the 6-Series User Manual. If using 6562 DO probe: 6570 maintenance kit 5775 membrane kit Razor blade
2 Installing Membrane on 6562 DO Probe (Skip to next section if you are not using this probe) The 6562 Rapid Pulse Dissolved Oxygen probe is shipped with a protective dry membrane on the sensor tip. Remove the O-ring and membrane. Handle the probe with care. It is important not to scratch or contaminate the sensor tip.! The DO sensor on the 600R/QS cannot be removed. Open the 6562 DO Probe Kit and follow these steps: 1. Prepare the electrolyte solution as specified on the bottle. 2. Put/keep the protective cap on the connector end of the probe to protect it from contamination. 3. Hold the probe in a vertical position and apply a few drops of KCl solution to tip. The fluid should completely fill the small moat around the electrodes and create a large meniscus on the probe face. Be sure no air bubbles are stuck to the face of the sensor. If necessary, shake off electrolyte and start over. 4. Secure a membrane between your left thumb and the probe body. Touch membrane only at the ends. 5. With the thumb and forefinger of your right hand, grasp the free end of the membrane and gently stretch it up, over, and down the other side of the sensor. The membrane should be stretched below the ridge where the O-ring seats so a good seal is achieved. 6. Secure the end of the membrane under the forefinger of your left hand. 7. Roll the O-ring over the end of the probe, being careful not to touch the membrane surface with your fingers. 8. If bubbles or wrinkles are present, remove the membrane and repeat steps Trim excess membrane with a razor blade. Rinse off any excess KCl solution, but be careful not to get any water in the connector. Refer to Section 2.3 of the 6-Series User Manual for complete instructions and photos.
3 ! When installing probes, be careful not to cross-thread the probe nuts. If you meet resistance while finger-tightening, STOP, back out, and begin again. Installing Probes! Before installing any probe into the sonde bulkhead, be sure that the probe port is free of moisture. Use a Kimwipe or lintfree cloth to dry. 1. Remove the calibration cup from your sonde, if necessary. 2. Using the long end of the probe installation tool supplied in the 6570 Maintenance Kit, remove the port plug from the port. 3. Some probes can only be installed into specific ports. Refer to Section 2.3 of the 6- Series User Manual for port designations. 4. Apply a thin coat of lubricant (Krytox ) to the O-rings on the connector side of the probe.! Make sure there are NO contaminants between the O-ring and the probe. Contaminants will cause leaks when the sonde is deployed. 5. Optical probes (6025 chlorophyll, 6130 Rhodamine, 6131 phycocyanin blue-green algae, 6132 phycoerythrin blue-green algae, 6136 turbidity, and 6150 ROX DO): Insert the probe in the appropriate optical port, seating the pins of the two connectors. Handtighten the probe nut to the bulkhead. Finish tightening the probe using the short end of the installation tool. Do not over-tighten Rapid Pulse DO, 6560 conductivity/ temperature, 6561/6579 ph, or 6565/6569 ph/orp probes: Insert the probe into the appropriate port, seating the pins of the two connectors. Push the probe inward so that it seats into the port. Screw down the probe nut by hand until it is finger tight. Use the long end of the installation tool to tighten the probe a quarter turn or until snug. Do not overtighten. 7. Ammonium, nitrate, and chloride ISE probes do not have slip nuts and should be installed without tools. Seat the probe body directly on the sonde bulkhead. Use your fingers to tighten.
4 Installing Probe Guard The probe guard protects probes during deployment and some calibration procedures. Once the probes are installed, screw on the guard by aligning it with the threads on the bulkhead and turn the guard clockwise until secure. Be careful not to bump the 6562 probe membrane when installing the probe guard. Installing Calibration Cup Installing the calibration cup is performed in the same manner as installing the probe guard. The calibration/storage cup is used for calibration procedures as well as sonde storage when not in use.! It is important to calibrate all sensors prior to using the sonde for field measurements. Refer to Section of the 6- Series User Manual. Power External power is required for the YSI 600R, 600QS, 600XL, 600XL V2, 6820 V2, and non-battery versions of 600OMS V2 and 600LS sondes. Internal batteries power the YSI 6920 V2, 6920DW, 6600 V2, 6600EDS V2, 600XLM, 600XLM V2, 600DW, and battery versions of 600OMS V2 and 600LS sondes. These can also run on external power. If you have a YSI 650 MDS display or other datalogger, attaching your sonde to it will power the sonde when power sonde is enabled. For lab calibration, a YSI 6067 or 6095 calibration cable and YSI 6038 (110 VAC) or 6037 (220 VAC) power supply is required for sondes without internal batteries. (6095 cable must be used with a field cable.) Connect the 6038 or 6037 power supply to the 6067 or 6095 cable. After attaching the four-pin connector from the power supply to the pigtail, plug the power supply into the appropriate AC outlet.
5 When using the sonde s internal batteries, install batteries prior to test and calibration. When using the 650 as the interface for sondes with internal power, disable the sonde power feature. Conversely, when using the 650 for sondes without batteries, enable the sonde power feature. Expected power supply outputs are: 6VDC = 4AA 12VDC = 8AA 12VDC = 8C If you see significantly differing values from these outputs, the batteries may not have a full charge or may be installed incorrectly. (You can find outputs in Status in the sonde menu; see next page.) Refer to Section 2.2 of the 6-Series User Manual for complete details. Communication Communication to the sonde is started with EcoWatch for Windows, the 650 display/ logger, or any terminal emulator. 1. Place the EcoWatch compact disk in your computer s CD drive. An Installshield Wizard will automatically start on most computers. Follow the prompts to complete the installation. 2. Select the EcoWatch icon on your desktop or from the Windows Program Menu and open the application. 3. Select the sonde icon in the toolbar, and then select the proper Comm port to which your sonde is connected. Refer to the diagram above for Communication Settings. Click OK to open a terminal window. 4. The terminal window is white. Type Menu after the # sign, press Enter, and the sonde Main menu will be displayed. 5. This software is menu-driven. Select functions by typing their corresponding numbers. You do not need to press Enter after choosing a selection. Type the 0 or Esc key to return to the previous menu.
6 6. After establishing connection with EcoWatch, you can configure sonde settings including date/time, sensors, logging, setup, and access stored files. Refer to Section 4 of the 6-Series User Manual or the Help section of the software. Menu Run = set up sonde for discrete or unattended logging applications Calibrate = access sensors for calibration File = view logged files including the calibration record (.glp file) Status = access firmware version, date and time, battery voltage and estimated life, amount of free memory, and logging status Advanced = view calibration constants, choose Auto Sleep configuration, adjust TDS constant and other sonde settings. Refer to Sections 2.5 and of the 6-Series User Manual before making any changes to the settings in the Advanced Menu. Technical Support YSI Technical Support is available Monday through Friday 8 am to 5 p.m. Eastern Time. Tel: option 1 or option 1 environmental@ysi.com Web: System = access sonde communication settings and instrument identification Report = choose parameters to log and units of measurement Sensor = enable sensors and assign probes to ports item # dwg # A69307 rev 01 October 2008
MULTIMETER: YSI-ProDSS rev. 9/23/15
MULTIMETER: YSI-ProDSS rev. 9/23/15 EQUIPMENT NEEDED YSI-ProDSS multimeter data sheet, clipboard, pencil Note: The battery may be damaged if kept in an environment outside the range of -20-60 C (-4-140
More informationUSER MANUAL DOCUMENT # REF. ProDSS USER MANUAL
USER MANUAL DOCUMENT #626973-01REF ProDSS USER MANUAL i This page left intentionally blank Contents Introduction 3 Safety information 3 Precautionary symbols 3 Product components 3 Battery use and battery
More informationYSI Environmental. Operations Manual. YSI 556 MPS Multi Probe System. P u r e D a t a f o r a H e a l t h y P l a n e t.
YSI Environmental P u r e D a t a f o r a H e a l t h y P l a n e t. YSI 556 MPS Multi Probe System Operations Manual Contents 1. Safety... 1 1.1 General Safety Information... 1 2. General Information...
More informationIntroduction: Features:
Index Features:... 3 Specifications:... 4 Accessories:... 5 Functions of Keyboard:... 7 Preparation:... 8 ... 8 ... 8 Calibration:... 8 ... 8 ... 9 ...
More informationUser Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English
User Manual DOCUMENT #601119REF EcoSense ph1000a ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL English Item #601119REF Rev C, October 2017 For the latest version of this manual, visit ysi.com
More informationOperations Manual EcoSense ph100a
Operations Manual EcoSense ph100a Portable ph, mv and Temperature Instrument English CONTENTS GENERAL INTRODUCTION... 3 INITIAL INSPECTION... 3 THE INSTRUMENT... 3 BATTERY INSTALLATION... 3 Battery Disposal...
More informationDatalogging Conductivity/ TDS Meter. Instruction Manual
Datalogging Conductivity/ TDS Meter 850039 Instruction Manual 1 TABLE OF CONTENTS 1. INTRODUCTION...3 2. PANEL DESCRIPTION...4 3. MEASURING PROCEDURE...5 3-A General Measurement...5 3-B Auto and Manual
More informationFIELD OPERATION GUIDE YSI 610 DISPLAY/LOGGER
Envir onm ental Monit oring Syst ems FIELD OPERATION GUIDE YSI 610 DISPLAY/LOGGER CONNECTING TO A SONDE Since YSI 6-Series sondes do not have displays or controls, users must communicate to them with a
More informationProDSS CATALOG MULTIPARAMETER WATER QUALITY FIELD INSTRUMENT. W86-02 Rev. A
ProDSS MULTIPARAMETER WATER QUALITY FIELD INSTRUMENT CATALOG W86-02 Rev. A YSI.com/ProDSS Customize your Professional Digital Sampling System to meet the demands of your water quality monitoring application.
More informationSTORM 3 BROWSER-BASED SOLUTION FOR DEPENDABLE DATA LOGGING. Tech Note D SIM CARD REPLACEMENT
STORM 3 BROWSER-BASED SOLUTION FOR DEPENDABLE DATA LOGGING SIM CARD REPLACEMENT Tech Note D114 0117 Storm 3 Data Logger Replacing the SIM Card Tech Note D114 0117 Article written by Matthew Longfield,
More informationYSI EXO Series Multi-Parameter Water Quality Sondes
YSI EXO Series Multi-Parameter Water Quality Sondes Introduction Yellow Spring Incorporated (YSI) is a leading manufacturer of water quality sensors and instruments. The EXO-series of water quality sondes
More informationYOUR EYES IN THE FIELD
YOUR EYES IN THE FIELD WATER QUALITY Instruments www.in-situ.com APPLICATIONS RUGGED RELIABILITY, INTUITIVE SOFTWARE, AND CONFIDENCE YOUR SENSORS WILL BE ACCURATE FROM CALIBRATION TO FIELD DEPLOYMENT,
More informationPROBES FOR YSI MULTILAB METERS
PROBES FOR YSI MULTILAB METERS These IDS sensors automatically store their unique serial number and calibration data. They maintain their calibration data even if moved to another meter. The meter monitors
More informationMultiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures
Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98194 Features HI98194 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,
More informationOperation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M
Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS
More informationSingle Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations
Single Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations Digital electrodes edge performs measurements through its unique
More informationWater Quality Meter. Water Quality Meter Kit. Instruction Manual
Water Quality Meter 850081 Water Quality Meter Kit 850086 Instruction Manual Contact : EASTERN ENERGY CO.,LTD. (HEAD OFFICE) 40/4 Vitoondumri Rd., Banbueng, Banbueng, Chonburi 20170 Tel: 0-3844-6117, Fax:
More informationOperations Manual EcoSense ph100
Operations Manual EcoSense ph100 Portable ph, mv and Temperature Instrument English Français Español Deutsch Italiano WARRANTY The EcoSense ph100 Instrument is warranted for one year from date of purchase
More informationEXO. advanced water quality monitoring platform E102-02
EXO advanced water quality monitoring platform E102-02 Breaking the sonde barrier. EXO represents the intersection of the Environment and Observation and a new generation of monitoring technology. 2 Features
More informationHI 9828 Multiparameter Water Quality Meter...
HI 9828 Multiparameter Water Quality Meter... Backlit LCD can display up to 12 Measurements at a time Feature Highlights Display up to 12 parameters Graphic LCD with backlight Meter features IP67 and probe
More informationportable waterproof multiparameter meter
water resistant battery powered interchangeable electrode DIN connection replaceable electrode portable waterproof multiparameter meter Professional testing and analysis out in the field is supported by
More informationBenchtop Multi-parameter Water Quality Meter
LMPWM-A Series Benchtop multi-parameter water quality meter LMPWM-A series is comes with different measurement modes for quality and testing of water on different parameters. It measures ph, Conductivity,
More informationPortable Conductivity Meter M1410. User Manual 1.002
Portable Conductivity Meter M1410 User Manual 1.002 General information This manual describes the operation of the Portable Conductivity Meter M1410. Please read this entire manual before operating this
More informationDissolved Oxygen Meter Kit DOK. Instruction Manual
Dissolved Oxygen Meter Kit 850081DOK Instruction Manual Contact : EASTERN ENERGY CO.,LTD. (HEAD OFFICE) 40/4 Vitoondumri Rd., Banbueng, Banbueng, Chonburi 20170 Tel: 0-3844-6117, Fax: 0-3844-6200 Email:
More informationDataHub: FAQs/Troubleshooting
On the Ward s DataHub... = Select key = On/Off and ESC key = Scroll key 1. How do I pair my DataHub with a Windows computer? 2. How do I pair my DataHub with an Apple Macintosh (Mac) computer? 3. How do
More informationResolution selectable from 0.01 and ph. Range ph. Accuracy ± ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
More informationTDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter
TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT Pool Professional s Meter TRACER EC/TDS/SAL POCKETESTER TM CODE 1749-KIT TABLE OF CONTENTS Introduction... 4 Specifications... 4 Contents... 5 Parts &
More informationWK7200. Water Quality Test Kit USER S MANUAL
WK7200 USER S MANUAL Water Quality Test Kit INTRODUCTION: Thank you for selecting model WK7200 microprocessor-based waterproof Water Quality Test Kit. It is used to measure a wide range of ph, ORP, Conductivity,
More informationHanna Instruments HI9819X Multiparameter series
Hanna Instruments HI9819X series www.hannainst.com HI98194 Waterproof Meter ph / mv, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen, Atmospheric Pressure and Waterproof IP67 rated waterproof,
More information4010-2/3 ph Adapter DIN / BNC
OPERATIONS MANUAL ba76144e01 02/2013 4010-2/3 ph Adapter DIN / BNC PH ADAPTER FOR MULTILAB 4010-2 AND MULTILAB 4010-3 4010-2/3 ph Adapter For the most recent version of the manual, please visit www.ysi.com.
More informationElectrodes ph EC DO /14 dge 2 P/L e
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand Manual log on stability Interval logging Temperature readout ( C or
More informationProDSS. Multiparameter Water Quality Field Instrument CATALOG
ProDSS Multiparameter Water Quality Field Instrument CATALOG ProDSS Digital Sampling System Parameters: Dissolved Oxygen (Optical) Turbidity ph ORP/Redox Conductivity Specific Conductance Salinity Total
More informationUser's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD
User's Guide ExStik EC400 Conductivity / TDS / Salinity / Temperature Meter MODE HOLD Introduction Congratulations on your purchase of the ExStik EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity
More informationUser's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500
User's Guide ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Introduction Congratulations on your purchase of the ExStik EC500 ph/conductivity/total Dissolved Solids (TDS) / Salinity
More informationVFS 2 Video Fiber Scope User s Guide. Test & Inspection
VFS 2 Video Fiber Scope User s Guide Test & Inspection VFS 2 Video Fiber Scope User s Guide Test & Inspection 2004-2010, AFL Telecommunications, all rights reserved. VFS2-00-1000 Revision B 2010-07-30
More informationVisit us at Back to the Extech DO600 Product Page
User's Guide II ExStik DO600 Dissolved Oxygen Meter RECALL Dissolved Oxygen DO600 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com
More informationDO Meter Kit. Instruction Manual
DO Meter Kit 850048 Instruction Manual DO Meter Kit 850048 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any
More informationDDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD
DDS-12DW Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD DDS-12DW Benchtop Conductivity Meter 1 Introduction Thank you for selecting the DDS-12DW benchtop conductivity meter.
More informationINSTRUCTION MANUAL. Handy Polaris 2
INSTRUCTION MANUAL Handy Polaris 2 **This page is left blank** Table of Contents 1 General... 4 2 Menus... 5 2.1 Calibrate... 5 2.2 Set Salinity... 5 2.3 View Log... 5 2.4 Setup Menus... 5 2.4.1 User Interface...5
More informationand Multiparameter Meters
and Multiparameter Meters 54650-23 ph Menu Options Table 1 How to Change ph Menu Options Setup Setup 1 Time Setup 2 Date Setup 3 Year Setup 4 Temperature Units Setup 5 Display Lock Setup 6 Measurement
More informationResolution selectable from 0.01 and ph. Range ph. Accuracy ± ph
ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or F) Automatic
More informationUser Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter
User Guide ExStik II EC400 Conductivity / TDS / Salinity / Temperature Meter Introduction Congratulations on your purchase of the ExStik II EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity /
More informationHI98194, HI98195, HI98196
HI98194, HI98195, HI98196 Multiparameter Meters ph/mv, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen, Atmospheric Pressure and Temperature INSTRUCTION MANUAL Dear Customer, Thank you
More informationMulti Water Quality Checker U-50 Series Instruction Manual CODE:GZ B
Multi Water Quality Checker U-50 Series Instruction Manual CODE:GZ0000230552B Preface This manual describes the operation of the Multi Water Quality Checker, U-50 Series. Be sure to read this manual before
More informationINTERFACE & SOFTWARE GUIDE
TM INTERFACE & SOFTWARE GUIDE Wireless Remote Display USB Converter Battery Sensor Setup DataLogger Software r e m o t e i n t e r f a c e escape enter status w w w. s k y s t r e a m e n e r g y. c o
More informationUser Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at
User Manual ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the
More information24-hour Continuous Field Monitors
Instruction Manual 24-hour Continuous Field Monitors ph, ORP, and Conductivity Technology Made Easy... 68X254901 Rev. 5 08/03 Preface This manual serves to explain the use for all models of the Field Monitors
More informationResolution selectable from 0.01 and ph. Range to ph. Accuracy ± ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 to 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand Manual log on stability Interval logging Temperature readout ( C
More informationUsing the SASS 2300 Wetted-Wall Air Sampler
Wetted-Wall Air Sampler Front View Features Rear View Features Setting Up the SASS 2300 System Connect Wall Power Transformer Connect System Electrical Power The SASS 2300 requires 12-24 VDC from either
More informationMultiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures
Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98195 Features HI98195 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,
More informationHI 98194, HI 98195, HI 98196
Instruction Manual HI 98194, HI 98195, HI 98196 Multiparameter Meters w w w. h a n n a i n s t. c o m 1 Dear Customer, Thank you for choosing a HANNA Instruments product. Please read this instruction manual
More informationHI 98194, HI 98195, HI 98196
Instruction Manual HI 98194, HI 98195, HI 98196 Multiparameter Meters TABLE OF CONTENTS Dear Customer, Thank you for choosing a HANNA Instruments product. Please read this instruction manual carefully
More informationUser s Manual. March 30, 2010 P/N: Revision E TURNER DESIGNS 845 W. Maude Ave. Sunnyvale, CA Phone: (408) FAX: (408)
User s Manual March 30, 2010 P/N: Revision E TURNER DESIGNS 845 W. Maude Ave. Sunnyvale, CA 94085 Phone: (408) 749-0994 FAX: (408) 749-0998 Table of Contents Table of Contents... 2 1.0 Introduction...
More informationIntegrated Instruction Manual WP 600 Series Meters
Integrated Instruction Manual WP 600 Series Meters ph 600/ 610/ 620, COND 600/610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Rev. 1 Oct 07 Part of Thermo Fisher Scientific Technology Made Easy...
More informationAppendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips
Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to
More informationHI 9828 Multiparameter Water Quality Meter...
HI 9828 Multiparameter Water Quality Meter... Backlit LCD can display up to 12 Measurements at a time Feature Highlights Display up to 12 parameters Graphic LCD with backlight Meter features IP67 and probe
More informationModbus Signal Output Adapter Quick Start Guide
TM Modbus Signal Output Adapter Quick Start Guide Adapter Overview: Mini USB Connector Used to configure adapter settings, provide power to the adapter, and passthrough communication to the attached sonde.
More informationOperation Manual. Hand-held ph/mv/temperature Meter. ph6810
Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4
More informationINSTRUCTION MANUAL. AquiStar GDL. Dissolved Oxygen Datalogger
INSTRUCTION MANUAL AquiStar GDL Dissolved Oxygen Datalogger Table of Contents 1 Introduction... 3 What is a GDL?... 3 Initial Inspection and Handling... 3 Do s and Don ts... 3 Getting Started... 4 GDL
More informationFOCIS Flex - No Wireless Fiber Optic Connector Inspection System User Guide
Test & Inspection FOCIS Flex - No Wireless Fiber Optic Connector Inspection System User Guide www.aflglobal.com or (800) 321-5298, (603) 528-7780 Table of Contents Safety Information....3 How to View Device
More informationVIS Video Inspection Scope
User s Guide VIS 300 - Video Inspection Scope VIS1-00-2000, Rev A, 6-09-10 Specifications subject to change without notice Contents Section 1 - General Overview A. Contact ODM for customer service B. Unpacking
More informationIoline 300/350HF System
Quick Start Guide Ioline 300/350HF System User Notice Trademarks Ioline is a trademark of Ioline Corporation. Other product names, logos, designs, titles, words or phrases mentioned within this publication
More informationFLOW MONITOR & TRANSMITTER
FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other
More informationOperation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R
Operation Manual MODEL 3173/3173R MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R INITIAL INSPECTION..... 2 GENERAL INTRODUCTION.. 2 USING THE METER. 3 THE DISPLAY....
More informationContents. Communications, Modular (excludes 17B, 17C, MK 21 and KMB BandMasks ) 1.1 Daily Maintenance. 1.2 Monthly Maintenance (or
Communications, Modular (excludes 17B, 17C, MK 21 and KMB BandMasks ) Daily Maintenance Communications, Modular (excludes 17B, 17C, MK 21 and KMB BandMasks ) Contents COM-1 COM-1 1.1 Daily Maintenance
More informationUSER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe
USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting
More informationUSER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe
USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe ADWA HUNGARY Kft. Alsókikötõ sor 11, 6726 Szeged, Hungary Tel. +36 62 317 878 Fax +36 62 550 610 www.adwainstruments.com MANAD630
More informationINSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01
INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01 27618366 / 74 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: CHAPTER 1 INTRODUCTION Contech CC-01 measures Conductivity, Total Dissolved
More informationUnpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place.
Unpacking and Installing the Flora 2512 UV Printer Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place. Step 2: Unscrew 10mm bolts holding the end panels. On the
More informationOPERATING INSTRUCTION MANUAL MODEL P/R60 & P/R65 DIFFERENTIAL PH & ORP PROBES
OPERATING INSTRUCTION MANUAL MODEL P/R60 & P/R65 DIFFERENTIAL PH & ORP PROBES N116-19 REV. 10 Water Analytics, Inc. 100 School Street Andover, MA 01810 Tel: (978) 749-9949 (855) 747-7623 Fax: (978) 749-9961
More informationEcoSense DO200A EcoSense DO200M Dissolved oxygen & temperature instruments
User Manual DOCUMENT #606035REF EcoSense DO200A EcoSense DO200M Dissolved oxygen & temperature instruments USER MANUAL English, Français, Español, Português WARRANTY The EcoSense DO200A/DO200M Instrument
More informationXL 15 Meter: Table of Contents
XL 15 Meter: Table of Contents Introduction...1 Unpacking the Meter...2 XL Series Meters Specifications... 3-4 Getting Started XL Series Meters Connectors... 5-8 Start-up and Shut down procedures...9 ph
More informationManual for the MicroSIA Instrument
Manual for the MicroSIA Instrument Version 1.1 1 FIAlab Instruments, Inc., 2151 N. Northlake Way, Suite 100, Seattle, WA 98103 USA Table of Contents 1 INTRODUCTION... 3 1.1 GENERAL... 3 1.2 OPERATION PRINCIPLE...
More informationTracer ph/tds/salt PockeTester TM Code 1766
Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity
More informationZebra-Tech UW-Calipers
Zebra-Tech UW-Calipers Operation Manual Contents 1. Your UW-Calipes at a glance... 1 Features... 2 2. UW-Caliper specifications... 3 3. Charging your UW-Calipers... 4 4. Operating your UW-Calipers... 5
More informationStandard Operating Procedure Remote Water Quality Monitoring Network. Site Selection Data Sonde Datalogger Data Correction
Standard Operating Procedure Remote Water Quality Monitoring Network Site Selection Data Sonde Datalogger Data Correction 2017 Table of Contents I. SITE SELECTION...1 A. Region...1 B. In-house Site Selection
More informationUSER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170
USER GUIDE Salinity Meter Pen Style Water Quality Meter Model EC170 Introduction Congratulations on your purchase of the Extech Pen Style Water Quality instrument; the Model EC170 measures Salinity and
More informationph and Electrochemistry
Portable Meters Portable Meters At-a-Glance Selection Guide Take your measurements into the great outdoors. Our Orion Star A series portable meters are built rugged and waterproof for dependable performance.
More informationCOMMONWEALTH OF PENNSYLVANIA SPECIFICATION
COMMONWEALTH OF PENNSYLVANIA SPECIFICATION Department of Environmental Protection 400 Market Street Harrisburg, PA 17105 Shall Meet: One (1) YSI Turnkey Hub Station NEMA 4X fiberglass enclosure CR310 measurement
More informationBTU 4 & Sensors. Installation and Operation Manual
BTU 4 & Sensors Installation and Operation Manual Table of Contents Introduction... 1 Sensors... 1 BTO Outdoor Sensor... 1 BTI Indoor Sensor... 2 BTS Stack Sensor... 2 Installation... 3 BTU 4 Burk Temperature
More informationINSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus
INSTRUCTION MANUAL Portable ph and ph/ise Meters Models 210Aplus 230Aplus 250Aplus 290Aplus Table Of Contents Chapter I Introduction................................1 Chapter II General Information.........................3
More informationSmart Multivariable Transmitter (SMV 3000) Electronics Module Replacement Kit Instruction
Smart Multivariable Transmitter (SMV 3000) Electronics Module Replacement Kit Instruction Electronics Module (Part number 51404208 503, -513) Document Form: 34-SM-33-01 Effective: 09-01 Supersedes: 34-SM-33-01,
More informationInstruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.
Instruction Manual HI 9033 HI 9034 Waterproof Multi-Range EC and TDS Meters for Field Applications www.hannainst.com PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments product.
More informationAirPro Surveyor 2 Manual
AirPro Surveyor 2 Manual AirPro Surveyor Specifications Table of Contents Size 3/8 x 7 1/2 x 4 3/8 Weight 4.6 lbs. 2094 g Dynamic Range 1-1000 ml/min. total flow/constant flow Flow Capacity (8 Hrs.) 1000
More informationYSI Professional Plus Water Quality Instrument. Specifications. Tougher. Smarter. More Versatile.
YSI Professional Plus Water Quality Instrument Specifications Tougher. Smarter. More Versatile. YSI Professional Plus The YSI Professional Plus is not a refinement. It came straight out of the imagination
More informationEXO. advanced water quality monitoring platform E102
EXO advanced water quality monitoring platform E102 Breaking the sonde barrier. EXO represents the intersection of the Environment and Observation and a new generation of monitoring technology. 2 Features
More informationINSTRUCTION MANUAL ORP-METER MODEL COR-01
INSTRUCTION MANUAL ORP-METER MODEL COR-01 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: 022-61944000 CHAPTER 1 INTRODUCTION CONTECH ORP meter measures ph and mv, can used for Oxidation
More informationPointWrite User Manual
PointWrite User Manual Table of contents Shipping contents... 3 Attaching the PointWrite to the projector... 4 Connecting the PointWrite to PC... 5 Pen action... 5 Exchanging the pen tip...5 Driver Installation...
More informationA150 Benchtop Conductivity Meter Instruction Manual
A150 Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO., LTD A150 Benchtop Conductivity Meter 1 Introduction Thank you for selecting the A150 benchtop conductivity meter. This manual
More informationField Equipment for Measuring Electrical Conductivity, Humidity and ph of Soil
Field Equipment for Measuring Electrical Conductivity, Humidity and ph of Soil User Manual This instrument is designed for measuring the soil s electrical conductivity, humidity, temperature and ph. It
More informationInstruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter
Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter Technology Made Easy... 68X248904 Rev 3 01/04 Preface This manual serves to explain the use of the Waterproof CON 400/410
More informationC100 Handheld Cond./TDS/Sal. Meter. Instruction Manual
C100 Handheld Cond./TDS/Sal. Meter Instruction Manual Table of Contents 1. Introduction ---------------------------------------------------------------------------------------------------------------------------
More informationM5 and M10 Routers Power Supply and Power Cord Component Replacement Instructions
M5 and M10 Routers Power Supply and Power Cord Component Replacement Instructions Part No. 530-003244-01 Revision 1 27 July 2000 This document describes how to remove and replace the AC and DC power supplies,
More informationINSTRUCTION MANUAL. ph/mv CONTROLLER/RECORDER
INSTRUCTION MANUAL ph/mv CONTROLLER/RECORDER LIMITED ONE YEAR WARRANTY Manufacturer warranties all instruments (excluding batteries, damage caused by batteries, probes, standards, buffers) against defects
More informationSYSTEM MAINTENANCE MANUAL TACTICAL ENGAGEMENT SIMULATION SYSTEM (TESS) TARGET TRAINING SYSTEM (TTS) OCONUS FIXED LOCATION INSTALLATION
SYSTEM MAINTENANCE MANUAL TACTICAL ENGAGEMENT SIMULATION SYSTEM (TESS) TARGET TRAINING SYSTEM (TTS) OCONUS FIXED LOCATION INSTALLATION INITIAL SETUP Preconditions Coordinate the TTS setup with the training
More informationThermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet
English ph / ISE / COND / RDO / DO Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the
More informationInstruction Manual HI Multiparameter Meter. With available GPS, logging probe, turbidity and ion measurements
Instruction Manual HI 9829 Multiparameter Meter With available GPS, logging probe, turbidity and ion measurements w w w. h a n n a i n s t. c o m 1 2 Dear Customer, Thank you for choosing a HANNA instruments
More informationHQd DIGITAL METERS. A customized solution for water quality testing that takes the guesswork out of your measurements. Easy to Use in the Field
HQd DIGITAL METERS & IntelliCAL PROBES A customized solution for water quality testing that takes the guesswork out of your measurements. Easy to Use in the Field The wide variety of rugged probe options
More informationReferenceMate portable reference source REF2500 operation manual
ReferenceMate portable reference source REF2500 operation manual Caution: This guide should be read carefully before use. Meggitt Sensing Systems 20511 Seneca Meadows Parkway, Germantown MD 20876, USA
More informationBTU-4D/BTU-4D-I Digital Temperature Unit & Sensors. Installation and Operation Manual
BTU-4D/BTU-4D-I Digital Temperature Unit & Sensors Installation and Operation Manual BTU-4D/BTU-4D-I Installation and Operation Manual Rev E Table of Contents Introduction... 1 TEMP OUTDOOR Sensor... 1
More information