DIGITAL PEAK HOLDER CSD-709. Instruction Manual EN C

Size: px
Start display at page:

Download "DIGITAL PEAK HOLDER CSD-709. Instruction Manual EN C"

Transcription

1 DIGITAL PEAK HOLDER CSD-709 Instruction Manual EN C

2

3 Introduction Introduction Thank you for your selecting our CSD-709 digital peak holder. This manual describes installation, connection, setting, and operating procedures for your equipment. Please read this manual carefully before using your CSD-709. Also, be sure to transfer this manual to the final user of the CSD-709. The final user of the equipment should read the manual and file it carefully. CAUTION The contents of this manual are subject to change without prior notice in order to reflect improvements to the product. Meanings of the marks and precautions used in this manual This manual offers precautions you must follow to avoid injuries and accidents, to maintainn measurement accuracy, and to ensure correct settings and operations. The precautions are given below. WARNING Failure to observe the instructions can result in injury or accident. Always observe the instructions identified by this mark. CAUTION Be sure to read the descriptions identified by this mark to maintain the accuracy of measurements or prevent malfunctions. POINT Provides supplementary information with respect to the installation, setting, and operation of the CSD-709, as well as its connection to other devices. Read the descriptions identified by this mark whenever necessary. Indicates where related descriptions can be found. 1

4 Introduction In this manual, the headings of sections and paragraphs indicate the accompanying functions used to set various features of the CSD-709. For example, the heading shown below accompanies function [CF-07] in order to set an appropriate value for C function CF-07 when it is necessary to set the excitation voltage. ***************************************************************************************************************** 2.3 Setting the excitation voltage [CF-07] ***************************************************************************************************************** Furthermore, the default of each function is described on an as-needed basis in this manual. For example, the default for C function CF-07 used to set the excitation voltage is described as shown below. ***************************************************************************************************************** Default 10 VDC ***************************************************************************************************************** 2

5 Safety Precautions Safety Precautions To use the CSD-709 safely, be sure to observe the following precautions. Precautions When settings are changed incorrectly during measurement with the CSD-709, damage to peripheral devices. Do not cause impact to the CSD-709, for example, by striking it against a nearby object. Otherwise, the components may be damaged or the electrical circuit may fail. correct results may not be obtained or the equipment may malfunction and cause Do not press the front panel of the CSD-709 with excessive force or with the sharp edge of a screwdriver, for example. Otherwise, panel switches may be damaged or the operability or environmental worthiness of the CSD-709 may be impaired. Do not remove the case lid, remove the panel seat, or disassemble the CSD-709, except in order to replace a fuse. Otherwise, the case or panel seat may be damaged or the operability or environmental worthiness of the CSD-709 may be impaired. Compatible standard The CSD-709 conforms to the following standards: EN : 2013 [Electrical equipment for measurement, control, and laboratory use - EMC requirements Part 1: General requirements] [Immunity requirements for equipment intended to perform functions in a safety-related system (functional safety) in industrial locations] EN : 2010 [Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements] EN50581: 2012 [Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances] (RoHS Directive) CAUTION The CSD-709 conforms to EN and EN when the following conditions are met. Shielding All cables, except for power cables, shall be of the shielded type. For the shielding method, see related sections. Grounding The CSD-709 shall be groundedd based on Type D single grounding using a protective ground connector. 3

6 Revision History Revision History Date Manual No. Revision reason (details) 2017/03 DRW. No. EN ROM Ver /04 DRW. No. EN A 2017/06 DRW. No. EN B 2018/06 DRW. No. EN C Due to ECN No.FN A.11.3 BCD output (Sync type), A.11.4 BCD output (source type) Change from [~1 000, or times/s] [~ 1 000, 2 000, or times/s]. G3. Setting, Function Setting Table F-33 Change from [~, 2 (5 ms)] to [~, 2 (10 ms), 3 (5 ms), 4 (1.25 ms), 5 (0.5 ms), 6 (0.25 ms), 7 (0.125 ms)]. F-34 Change from [~, 5 (1 000 times/s), 6 (4 000 times/s)] to [~, 5 (1 000 times/s), 6 (2 000 times/s),7 (4 000 times/s)] Due to ECN No.FN Change [Confirmation with Check mode] to [Confirmation with Check mode (EzCTS, Version, Options, Excitation voltage, External control input/output, Analog output, BCD output, Monitor mode)] Changing the settings in EzCTS mode Add [EzCTS is a personal computer application (option). The function data of CSD-709 can be easily read-out and written with EzCTS. This software is useful for maintenance and backup of data.] A.10. Accessories Change [CD-ROM(application)] to CD-ROM * (CSD-709D-01, EzCTS) Add [* The CD-ROM attached when the optional additional memory is mounted in the CSD-709 includes both of the dedicated application CSD-709D-01 and the EzCTS.] Due to FN A G1 Wiring Add the note. Only BCD-ENABLE (pin number B15) of the source type becomes effective at the [OPEN]. Use this function after short-circuiting B15 and A1 or B1 when you do not use it. G4. Timing chart Raplace the drawing. Correct mis-writing 4

7 Contents Contents Introduction... 1 Safety Precautions... 3 Revision History... 4 Contents Overview Checking contents of package Connectable components Features of CSD Nomenclatures and functions Mode Basic key operation Procedures to prepare for operation of the CSD Installation and Wiring Installation Wiring Setting the excitation voltage Calibration Overview Changing the calibration number Setting the decimal point display position Calibration based on actual load Calibration based on input of numerical value Fine adjustment Simplified calibration Verifying the registered calibration data Settings Related to Measurement Overview Setting the AD sampling frequency Setting the digital filter Enabling the analog peak detection circuit Settings Related to Detection Setting the detection conditions Detection with multimode Detection result judgment and waveform comparison Settings Related to Comparison Overview Setting the comparison value Setting the comparator operating condition Setting the comparator hysteresis Settings Related to Operation Overview Setting the display update frequency Lock Setting the overload display Confirmation with Check mode (EzCTS, Version, Options, Excitation voltage, External control input/output, Analog output, BCD output, Monitor mode)

8 Contents 7.6 Output of CHECK value Setting the F key operation Zero set and effective range Setting hold targets Initialization of setting Settings Related to Output Control External control input signal External control output signal Troubleshooting Error code list Troubleshooting Replacing the fuse Communications via USB Communication specifications Setting the USB interface Data format Remote calibration via USB A Specifications of CSD A.1 Analog Specifications A.2 Digital Specifications A.3 Functions of Front Panel Sheet Key Switches A.4 External Control Function A.5 Comparator Function A.6 Function A.7 USB Interface (for connection with EzCTS and specialized application) A.8 General Specifications A.9 Standard Shipment Specifications A.10 Accessories A.11 Option B Warranty B.1 Warranty B.2 Repairs B.3 Component Service Lives C Character Display Pattern D Detection Operation and Section D.1 Detection operation D.2 Detection section E Voltage Output Card CSD E.1 Wiring E.2 Setting E.3 Calibration E.4 Adjustment E.5 Output Operation in Hold Mode E.6 Checking on Problems F Current Output Card CSD F.1 Wiring F.2 Setting F.3 Calibration F.4 Adjustment F.5 Output Operation in Hold Mode F.6 Checking on Problems

9 Contents G BCD Card CSD /CSD G.1 Wiring G.2 Output specifications G.3 Setting G.4 Timing chart G.5 Output Operation in Hold Mode G.6 Checking on Problems H RS-232C Card CSD H.1 Wiring H.2 Communication specifications H.3 Setting H.4 Checking on Problems I RS-422/485 Card CSD I.1 Connection I.2 Communication specifications I.3 Setting I.4 Checking on Problems J Communications Based on Modbus Protocol J.1 Communication specifications J.2 Setting J.3 Modbus function forma J.4 Calibration based on Modbus K Serial Interface Card CSD K.1 Connection K.2 Communication specifications K.3 Setting K.4 Checking on Problems L External Control Voltage Input Type Circuit CSD L.1 Connection M Power Supply Voltage CSD M.1 Connection Function Setting Table

10 Contents 8

11 1 Overview 1 Overview 1.1 Checking contents of package Before using the CSD-709, be sure to check if the package contains all the items listed below. CSD-709: 1 Startup Guide (in Japanese and English): 1 set each Midget fuse (2.5 A): 1 Unit seal: 1 External control input/outputt connector: 1 Strain-gage-based transducer connector: 1 CAUTION The following components are contained in the package when options are mounted to the CSD-709. Voltage output plug (component for CSD ): 1 Current output plug (component for CSD ): 1 BCD output plug (component for CSD /CSD ): 1 RS-232C plug (component for CSD ) ): 1 RS-422/485 plug (component for CSD ): 1 Serial interface plug (component for CSD ): 1 CC-Link plug (component for CSD ): 1 Application CD-ROM (medium for CSD ): 1 USB cable (component for CSD ): Connectable components A strain-gage-based transducer, such as a load cell or pressure sensor, can be connected to the CSD-709. A PC and printer can also be connected. Shown below are the components that can be connected to the CSD-709. Strain-gage- based transducer Voltage/ /current output RS-232C RS-422/485 BCD CC-Link SI/F Measuring devices such as a voltmeter, ammeter, or data logger External control devices such as a PLC Printer and display USB PC (for additional memory or PC application) 9

12 1 Overview 1.3 Features of CSD Basic functions The CSD-709 supports the following basic functions. The CSD-709 captures the sharp peak load more accurately by means of high-speed AD sampling of up to times/s, an analog low-pass filter, a digital low-pass filter, and moving-average processing. It also supports an analog peak hold function that enables more rapid peak capturing regardless of sampling rate. The CSD-709 supports a simplified calibration function that starts the calibration process by simply holding down a key. Calibration based on communication through various interfaces is also available. In the pre-work check, you can use preset check values to easily verify that the correct calibration status is maintained. The load currently applied to the strain-gage-based transducer, such as the load cell, is displayed in mv/v, allowing you to separate abnormal values and unstable measured values. Tracks and peaks can be output in analog form. The hold target can be flexibly selected from displayed values and comparative outputs. Up to four different types of calibration data or comparator settings are memorized so that you can select the desired value by means of key operation or external control input. You can select the peak detection mode from 36 modes configured through a combination of nine hold modes (peak hold, bottom hold, peak-bottom hold, peak-to-peak hold, maximum hold, minimum hold, maximum-minimum difference hold, average hold, and four inflection point holds) and four section modes (entire section, specified section, time-specified section, and automatic start time-specified section) Addition of functions using options The following additional functions become available through mounting options. Current and voltage outputs synchronizing with high-speed AD sampling Sink-type and source-type BCD outputs Readout of loads and confirmation/change of settings by means of RS-232C interface Readout of loads and confirmation/change of settings by means of RS-422/485 interface Readout of loads by means of serial interface (S-I/F) Power supply voltage in DC External control input in voltage Readout of loads and confirmation/change of settings by means of CC-Link interface Utilization of PC applications through the addition of memory 10

13 1 Overview 1.4 Nomenclatures and functions Front panel 1. Load window 2. Status window 3. Judgment window 4. Key Indicates the value currently measured in Measurement mode. Also indicates the settings in any mode other than Measurement mode, such as Function mode. Indicates the current condition by turning on the associated LED. SEL1, SEL2 CHECK HOLD PEAK MEAS. END LOCK Either LED is turned on depending on the calibration number selected. This LED is turned on while the check value is being output. This LED is turned on in hold mode. This LED is turned on during detection of a peak. This LED is turned on during detection of a peak in a section other than reset section. This LED is turned on during detection of a peak in the hold mode. This LED is turned on in the key lock mode. Indicates the condition of the selected external control output by turning on the associated LED. Inputs commands necessary for settings and operations. Selects any mode other than Measurement mode, such as Function mode. Holding this key down for at least two seconds triggers simplified numerical value input based calibration. Allows you to move to the next upper digit in the target setting. Holding this key down for at least two seconds triggers simplified zero calibration. Increases the numerical value of the target setting by one. Holding this key down for at least two seconds triggers simplified span calibration. Selects the track or various peak detections. Resets the peak. Also outputs the CHECK value. Executes the F key operation already set. Also registers the setting. 11

14 1 Overview Rear panel 1. Protective ground connector The grounding cable is plugged in here. 2. Power supply connector board The power cable is plugged in here. 3. Connectors for options 4. External control input/output connector 5. Strain-gage-based transducer connector The cable suitable for the option mounted is plugged in here. The external control input/output connector is plugged in here. The signal cable from the strain-gage-based transducer or the strain-gage-based transducer connector to which the monitor output signal cable is connected is plugged in here. 6. USB interface connector The USB cable is plugged in here. 12

15 1 Overview 1.5 Mode The CSD-709 supports the following measurement and operating modes. Measurement mode Function mode C function mode P function mode CAL selection mode CCAL mode CADJ mode VCAL mode VADJ mode Check mode COMP setting mode Measurement mode: You can display the value currently measured. Measurement mode is further divided into two modes: Track mode and Peak mode. In Track mode, the value, which is input from the measuring unit and changes with the elapse of time, is displayed in the load window. In Peak mode, the maximum and minimum measurements are detected and held. Function mode: You can set the functions prefixed with [F-], including the function related to the number of AD sampling times and the measurement of the digital filter, the function related to comparisons (e.g., operation of the comparator), and the function related to the operation of the key lock and hold target, and check these settings. C function mode: You can set the functions prefixed with [CF-], including the selection of calibration number, decimal point display position, and current calibration data, and check these settings. P function mode: You can set the peak hold related functions prefixed with [PF-], including determination of the detecting section and execution of detection, and check these settings. CAL selection mode: You can select CCAL mode used for calibration, CADJ mode used for fine adjustment of calibration results, and VCAL mode and VADJ mode used for calibration and adjustment of optional current output and voltage output cards. Check mode: You can check CSD-709's ROM version, mounted options, external control operations, and the like. COMP setting mode: You can set comparison values and check comparator settings. 13

16 1 Overview The mode in which multiple detection operations, such as peak detection, maximum detection, and inflection point detection, are combined to obtain the results from the load changing with the elapse of time is called [multimode] and distinguished from operations based on a single detecting sequence ([single mode]). 14

17 1 Overview 1.6 Basic key operation To toggle between Measurement mode and any other mode, use the key. To move to any of COMP setting modes from the Function mode, press the or key to change the display, then press the key. To return to Measurement mode from any other mode, press the key several times. Measurement mode Function mode C function mode P function mode CAL selection mode Check mode COMP setting mode In Function, C function, and P function modes, respectively, the following numerical values prefixed with the codes shown below are displayed. Mode Prefix Representation in this manual Function mode F-xx [F-xx] C function mode CF-xx [CF-xx] P function mode PF-xx [PF-xx] 15

18 1 Overview In these modes, the and keys execute digit movement and value change, respectively. Key Function Allows you to move to the adjacent digit. Changes numerical values 0 to 9. For example, to change the display from to, conduct the following steps. 1. Press key to flash [2]. 2. Press key to change [2] to [4]. 3. Press key to flash [1]. 4. Press key to change [1] to [3]. To input the calibration data, use and as well. 1.7 Procedures to prepare for operation of the CSD-709 Before using the CSD-709, conduct the following setting sequence. 1. Install the CSD-709. (See Section 2.1, "Installation".) 2. Connect cables to the CSD-709. (See Section 2.2, "Wiring".) 3. Calibrate the CSD-709. (See Section 3, "Calibration".) 4. Set the CSD-709 according to the intended application. (See Sections 4, "Settings Related to Measurement" to 7 "Settings Related to Operation".) 5. Start the operation. 16

19 2 Installation and Wiring 2 Installation and Wiring 2.1 Installation Precautions Use the CSD-709 within the temperature and humidity ranges specified below. Ambient temperature -10 C ~ 50 C Ambient humidity 85 %RH, max (No condensation is permitted.) WARNING Do not install the CSD-709 in any of the locations listed below. Locations exposed to direct sunlight or high temperatures Damp locations Locations exposed to vibrations or impact Locations exposed to lots of dust or fine particles Environments containing corrosive gases or salt Locations subject to sudden temperature or humidity fluctuations Close to equipment that generates magnetic or electromagnetic radiation Locations subject to the effects of radiation and radioactive materials Locations such as laboratories where chemical changes may occur Installing the CSD-709 This section describes the external dimensions of the CSD-709 and the appropriate installation procedures. Shown below are its external dimensions and panel cut dimensions. Strain-gage-based transducer connector plugged in: Panel cut dimensions External control input/output connector plugged in: 17

20 2 Installation and Wiring Described below are the installation procedures for the CSD Drill holes in the panel according to the panel cut dimensions. 2. Remove the two screws (A) from the rear panel and detach the two panel fittings (B) from both side panels. 3. Mount the CSD-709 from the front of the panel. 18

21 2 Installation and Wiring 4. Attach the panel fittings (B) from the rear of the panel and fix them with the two screws (A). 19

22 2 Installation and Wiring 2.2 Wiring WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. Before turning on power, check to confirm that the power supply voltage indicated on the CSD-709 and in the specifications agrees with the voltage actually supplied. Using an unconfirmed power supply may result in electric shock or equipment damage. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. 20

23 2 Installation and Wiring Plugging in the strain-gage-based transducer connector The strain-gage-based transducer connector is used to connect signal cables of the strain-gage-based transducer, such as load cell or pressure sensor, to the CSD-709. Also, this connector is used to connect monitor output signal cables. Shown below are terminal locations on the strain-gage-based transducer connector. Terminal symbol Terminal Color code of signal cable A Strain-gage-based transducer connector (+EXC) Red B Strain-gage-based transducer connector (-SIG) Blue C Strain-gage-based transducer connector (-EXC) White D Strain-gage-based transducer connector (+SIG) Green E Strain-gage-based transducer connector (shield) Yellow + Monitor output (+) - Monitor output (-) For example, one load cell is connected to the CSD-709 as shown below. 21

24 2 Installation and Wiring Similar to the connection with one load cell, one load cell and one junction box (B-304) are connected to the equipment as shown below. Junction box This time, connect signal cables as shown below in the extended junction box. To load cell Red Blue White Green Yellow (shield) Conforming crimp terminal: or 2-4 Terminal pitch: 9.5 mm Red Blue White Green Yellow (shield) To CSD-709 CAUTION The signal cable colors shown above are our standard color codes. Each transducer manufacturer employs unique signal colors for strain-gage-based instruction manual and inspection result record of the strain-gage-based transducer to be used before transducers such as the load cell. For this reason, check the connecting signal cables to the CSD-709. When you intend to use a tensionn type or compression and tension type load cell and define the direction of tensionn as a positive direction, swap the signal cables of +SIG and -SIG beforee connection. Arrange the CAB-502 cable between junction box and CSD-709 so that the length of the cable will not exceed 30 m. If the length exceeds 30 m, the input voltage to the CSD-70warranted measurement accuracy. will drop due to cable resistance, creating a risk of deviations from 22

25 2 Installation and Wiring With respect to the excitation voltage to the strain-gage-based, you can select a voltage level from 2.5 VDC, 5 VDC, and 10 VDC. The selected excitation voltage determines the number of parallelly connectable strain-gage-based transducers as shown below. Excitation voltage Number of parallelly connectable transducers 2.5 VDC 8 5 VDC 4 10 VDC 2 For excitation voltage setting procedures, see Section 2.3, "Setting the excitation voltage". Described below are procedures used to plug in signal cables to the strain-gage-based transducer connector of the CSD Remove the sheath at the end of each signal cable by 6 mm to 7 mm. 2. Insert the signal cable into the hole of the strain-gage-based transducer connector (A) and tighten the screw at the top face of the connector. CAUTION In case of stranded cable, lightly twist core wires so that they will not be separated. If core wires are twisted excessively, the cable may be easily disengaged from the connector. 3. Mount the strain-gage-based transducer connector to the CSD-709 and tighten the two screws (A) at the ends of the connector. A A 23

26 2 Installation and Wiring Plugging in the monitor outpu Voltages proportional to the inputs from the strain-gage-based of monitor output. The monitor output can be used transducer, such as the load cell or pressure sensor, can be picked up in the form when you want to observe or record waveforms with a recorder. The monitor output will take any of the values shown below depending on the excitation voltage input to the strain-gage-based transducer. (Note that the monitor output is a voltage per 1 mv/v of input signal.) Excitation voltage Monitor output 2.5 VDC Approx. 0.5 V 5 VDC Approx. 1 V 10 VDC Approx. 2 V CAUTION The monitor output voltage is picked up beforee the input signal from the strain-gage-based transducer is AD-converted. Therefore, the monitor output does not link with the displayed value processed with the digital filter. When you need a voltage or current output that links with the displayed value, use the optional voltage output card or current output card. For excitation voltage setting procedures, see Section 2.3, "Setting the excitation voltage". Plug in monitor outpu signal cables to the [+ +] and [-] terminals of the strain-gage-based transducer connector using the same procedures designed for the signal cables of the strain-gage-based transducer. For more information, see Section 2.2.1, "Plugging in the strain-gage-based transducer connector". 24

27 2 Installation and Wiring Plugging in the external control input/output connector To remotely control the CDS-709 or verify its status, the external control input/output connector is used. Shown below are the terminal locations on this connector. Pin No. A1 A2 A3 A4 A5 A6 A7 A8 B1 B2 B3 B4 B5 B6 B7 B8 Purpose External control input (INPUT1) External control input (INPUT2) External control input (INPUT3) External control input (INPUT4) External control input (INPUT5) External control input (INPUT6) - External control input (COM1) External control output (OUTPUT1) External control output (OUTPUT2) External control output (OUTPUT3) External control output (OUTPUT4) External control output (OUTPUT5) External control output (OUTPUT6) - External control output (COM2) Described below are the plug-in procedures for connection with external control input/output connector. 1. Mount the external control input/output connector to the CSD-709 and tighten the two screws (A) at the ends of the connector. A A 25

28 2 Installation and Wiring Connection of external control input The external control input circuit is classified into two types: no-voltage contact input type ( standard) and voltage input type (option). The no-voltage contact input type circuit inputs signals by short-circuiting/opening the path between the input and COM1 terminals. To conduct short-circuiting in this type, use contacts (e.g., relay and switch) and contactless outputs (e.g., open collector output). Shown below is the equivalent circuit of the no-voltage contact input type circuit. Inside of CSD V IN 1 ~ K COM1 CAUTION When signals are input based on the open collector output, use the sink type (minus common) for the connection. The no-voltage contact input type external control input circuit features internal power supply voltage of 12 VDC and short-circuit current of about 5 ma. For the voltage input type circuit, see Attachment L, "External Control Voltage Input Type Circuit (CSD )". 26

29 2 Installation and Wiring Connection of external control output The external control output circuit supports the photo MOS relay output. Plus common connection or minus common connection can be established. Shown below is the equivalent circuit of the external control output circuit. Inside of CSD-709 COM2 +Vext Surge protector Plus common connection OUT 1 ~ 6 Load Inside of CSD-709 OUT 1 ~ 6 Load +Vext +Vext Surge protector Minus common connection COM2 CAUTION The external control output circuit features a maximum rated voltage of 30 VDC and a maximum rated current of 100 ma per point. 27

30 2 Installation and Wiring Connecting the power supply The CSD-709 is classified into AC power type (standard) or DC power type (option). The power input cable must be plugged in the power supply connector board. Described below are plug-in procedures for the power supply. Always ground the equipment. Failure to connect the grounding cable to the CSD-709 may result in electric shock or equipment malfunction. The power supply input for the AC power type equipment features voltage of 100 V to 240 V and frequency of 50/60 Hz. The power consumption is about 10 W (at 1000 VAC) The tightening torque of the set screw on the connector board is as follows. Connector board Tightening torque Power supply 0.6 N m The crimp terminal conforming to the connector board of the equipment is as follows. Connector board Width of crimp terminal Conforming crimp terminal Power supply 6.2 mm, max O-type or Y-type For connectionn of power supply to the DC power type equipment, see Attachment M, "Power Supply Voltage CSD ". 1. Check that power is not applied to the cable. CAUTION 2. Remove the two screws (L and N) from the power supply connector board. Also, remove one screw from the protective ground connector. Protective ground connector 3. Engage power cable terminals with the threaded holes and secure them with the screws. 28

31 2 Installation and Wiring 2.3 Setting the excitation voltage [CF-07] With respect to the excitation voltage to the strain-gage-based transducer, you can select a voltage level of 2.5 VDC, 5 VDC, or 10 VDC. The excitation voltage should be set by using the C function CF-07. Shown below are the procedures for setting excitation voltage. POINT Changing the excitation voltage affects all calibration results stored to calibration numbers 0 to 3. For this reason, reexecute the calibration for numbers 0 to 3, respectively. When you enter Check mode, you can verify the excitation voltage currently set. For more information about Check mode, see Section 7.5, "Verification with Check mode". For more information about calibration procedures, see Section 3, "Calibration". 1. Press once to display. 2. Press once to change the display to. 3. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 4. Using and, change the numerical value to and press. The following numerical value showing the currently set excitation voltage appears. Numerical value Excitation voltage 0 10 VDC 1 5 VDC VDC Default 10 VDC For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set the excitation voltage and press. Whenever is pressed, the excitation voltage is cyclically changed from 0 to 2 in ascending order. 6. Press twice to return to Measurement mode. 29

32 3 Calibration 3 Calibration 3.1 Overview In either of the following cases, always calibrate the CSD-709 before use. Otherwise, you may obtain incorrect measurement resultss or the equipment may malfunction and cause damage to peripheral devices. Use the CSD-709 for the first time. Replace the strain-gage-based transducer or equipment. When incorrect measurement results are obtained, recalibrate the CSD-709. If you continue using the equipment without recalibration, it may malfunction and cause damage to peripheral devices. POINT WARNING When we have calibrated the CSD-709 in combination with a measuring unit (such as a strain-gage-based transducer), recalibration is not necessary. The CSD-709 can be calibrated in the following two ways. Calibration based on actual load Calibration based on input of numerical value In the calibration based on actual load, use a weight to adjust the load so that the input to the CSD-709 from the measuring unit will be correctly displayed on the CSD-709. If the CSD-709 is not calibrated: Load and decimal point display position are not displayed correctly. After the CSD-709 is calibrated: Correct load is displayed and the decimal point is placed in the correct position as set. 30

33 3 Calibration In the calibration based on input of numerical value, do not use a weight. Instead, manually enter the value to be input from the measuring unit to the CSD-709 through the front panel to adjust the value displayed on the equipment. In these two different types of calibrations, you should use the following procedures. Procedures common to actual load based calibration and numerical value input based calibration Changing the calibration number Setting of decimal point display position Setting of unit weight Setting of measuring weight Setting of weight mass Actual load based calibration Zero calibration based on actual load Numerical value input based calibration Entry of zero point Span calibration based on weight Entry of span Registration In these two types of calibrations, the actual load based calibration covers [zero calibration based on actual load] and [span calibration based on weight] and the numerical value input based calibration covers [entry of zero point] and [entry of span]. Which method you select is determined based on these factors: calibration conditions and whether the weight can be prepared. 31

34 3 Calibration Zero calibration based on actual load / Entry of zero point Normal zero calibration method is [zero calibration based on actual load]. Register the zero point with a tare or equivalent weight placed on the measuring unit. Zero calibration based on entry of zero point should be executed as long as all conditions listed below are met. The CSD-709 being used is replaced with another CSD-709. The zero calibration data (mv/v value) of the former CSD-709 is known. It is not possible to establish a no-load condition (initial condition including tare) for the measuring unit. Span calibration based on weight / Entry of span When it is possible to prepare a weight of at least two-thirds the measuring weight, execute [span calibration based on weight]. When there is no weight, execute [entry of span]. POINT When the span is input, the calibration accuracy is about 1/ If the accuracy of higher than 1/1 000 is required, execute [span calibration based on weight]. Whichever method you select, actual load based calibration or numerical value input based calibration, you can fine-tune the results to get more accurate values. 32

35 3 Calibration 3.2 Changing the calibration number [CF-18] The CSD-709 can store calibration results to four calibration numbers (0 to 3), respectively, and allows you to use the desired number by selecting it. For example, after calibration results acquired by multiple measuring units are stored to those calibration numbers, selecting the desired number allows you to execute measurement by using the measuring unit connected to the number. When you intend to use calibration numbers in such a way, check the number to which the calibration result will be stored and provide proper settings before starting the calibration sequence. In the CSD-709, you can select the desired calibration number from four numbers by using any of the following methods: Setting with C function CF-18 External control input Communication by means of USB Communication by means of optional devices (RS-232C card or RS-422/485 card) Which method is used to change the calibration number is determined by entering any of values listed below to the function F-89. Setting Calibration number change method 0 Changes the number with C function CF Changes the number with the external control input. 2 Changes the number through communication by USB. 3 Change the number through communication by means of optional devices. Default 0 (Changes the number with CF-18.) You can check the selected calibration number based on the status of the LED (on or off) in the status window, as shown below. Calibration LED in the status window number SEL1 SEL2 0 (Off) (Off) 1 (On) (Off) 2 (Off) (On) 3 (On) (On) Described below are calibration number change procedures by means of C function CF-18. Regarding the change of a calibration number using an external control input, USB-based communication, and option-based communication, see Section 8.1, "External control input signal", Section 10.3, "Data format", and Section J.4, "Calibration with Modbus", respectively. 1. Press once to display. 2. Press once to change the display to. 3. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 33

36 3 Calibration 4. Using and, change the numerical value to and press. The current calibration number appears. Default 0 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the calibration number to which the calibration results are stored, and press. Whenever is pressed, the calibration number is cyclically changed from 0 to 3 in ascending order. 6. Press twice to return to Measurement mode. 34

37 3 Calibration 3.3 Setting the decimal point display position [CF-01], [CF-42], [CF-43], [CF-44] The CSD-709 can set the number of decimal places from 0 to 4. You should store the decimal point display position to C function CF-01, CF-42, CF-43, or CF-44 depending on the calibration number, as shown below. C function Decimal point display position stored CF-01 Decimal point display position for calibration number 0 CF-42 Decimal point display position for calibration number 1 CF-43 Decimal point display position for calibration number 2 CF-44 Decimal point display position for calibration number 3 Also, you can set the decimal point display position during the actual load based calibration and numerical value input based calibration. Described below are the setting procedures of decimal point display position. 1. Press once to display. 2. Press once to change the display to. 3. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 4. Using and, change the numerical value to any of,,, and, and press. Any of the following numbers showing the current decimal point display position appears. Number Decimal point display position 0 Hide the decimal point place Default 0 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the number showing the decimal point display position and press. Whenever is pressed, the number showing the decimal point display position is cyclically changed from 0 to 4 in ascending order. 6. Press. 7. Press twice to return to Measurement mode. 35

38 3 Calibration 3.4 Calibration based on actual load The calibration based on actual load should be executed after you enter CCAL mode as shown below. Set the decimal point display position, unit weight, and measuring weight, conduct a series of subsequent setting procedures while pressing, and finally register the calibration data. Transition to the CCAL mode To execute calibration, enter CCAL mode. Setting of decimal point display position Display position. and set the decimal point display Setting of unit weight Setting of measuring weight Display and set the unit weight to be defined as the minimum scale. Display and set the measuring weight to be defined as the maximum measuring value. Setting of weight mass Display span calibration. and set the weight mass used in Zero calibration Span calibration based on weight Display and set the zero point with a tare or equivalent weight placed on the measuring unit. Display and apply the load equivalent to the set weight mass to the measuring weight to set the span. Registration With displayed, press the calibration data. to register CAUTION When you intend to execute actual load based calibration, prepare a weight of at least two-thirds of the measuring weight (maximum measuring value). Before the actual load based calibration, check the calibration number to be used. For the verification method of the calibration number, see Section 3.2, "Changing the calibration number". Transition to CCAL mode 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press. CCAL mode is activated. 36

39 3 Calibration Setting of decimal point display position POINT To terminate the calibration, press to leave CCAL mode. When you leave CCAL mode during calibration, all settings established beforehand will be permanently erased. (After the transition to CCAL mode) 1. After confirming is displayed, press. The currently set decimal point display position appears. 2. Change the decimal point display position by using and on an as-needed basis. Whenever is pressed, the decimal point display position is cyclically changed from [0], [0.0], [0.00], and [0.000] to [0.0000], in ascending order. Similarly, whenever is pressed, the decimal point display position is cyclically changed in descending order. For procedures to set the decimal point display position, see Section 3.3, "Setting the decimal point display position". 3. Press. Setting of unit weight (After the setting of decimal point display position) 1. After confirming is displayed, press. The currently set unit weight (minimum scale) appears and flashes. 2. Using and, set the unit weight. Only 1, 2, 5, or 10 can be set. When you have set the number of decimal places, the rightmost figure will be 1, 2, 5, or 10. Whenever is pressed, the unit weight is cyclically changed from 1, 2, 5, and 10, in ascending order. Similarly, whenever is pressed, the unit weight is cyclically changed in descending order. POINT The set unit weight determines the increment of the value displayed in the load window. For example, when [2] is selected, the value in the load window is increased at an increment of 2 (e.g., ). Similarly, when [5] is selected, the value in the load window is increased at an increment of 5 (e.g., ). 3. Press. Setting of measuring weight (After the setting of unit weight) 1. After confirming is displayed, press. The currently set measuring weight (maximum measuring value) appears and the number in the last digit flashes. 2. Using and, set the measuring weight. 37

40 3 Calibration CAUTION The measuring weight does not accept [0]. When you select [0] and press flashing appears and then reappears., For setting procedures with 3. Press. and, see Section 1.6, "Basic key operation". Setting of weight mass (After the setting of measuring weight) 1. After confirming is displayed, press. The currently set weight mass appears, and the number in the last digit flashes. 2. Using and, set the weight mass. CAUTION The weight mass does not accept a value exceeding the measuring weight or [0]. When you select either value and press, flashing appears and then reappears. POINT The weight mass used in the calibration should weigh at least two-thirds of the measuring weight. When it is possible to prepare a weight whose mass is equal to the measuring weight, set the weight mass to a value equal to the measuring weight. 3. Press. Zero calibration (After the setting of weight mass) 1. Confirm is displayed. 2. Place the tare or equivalent weight on the measuring unit. 3. Press. flashes and the zero calibration is initiated. Upon completion of zero calibration, appears. 38

41 3 Calibration CAUTION Be careful that load fluctuation induced by vibration, for example, will not be exerted on the measuring unit during zero calibration. When the input from the measuring unit exceeds 2.5 mv/v, flashing appears. When the input is lower than -2.5 mv/v, flashing These outliers are outside the scope of zero point calibration. appears. Span calibration based on weight (After zero calibration) 1. Confirm is displayed. 2. Apply a load equal to the value determined as weight mass to the measuring unit. 3. Press. flashes and span calibration is initiated. Upon completion of span calibration, appears. CAUTION When the input from the measuring unit exceeds 3.1 mv/v, flashing appears. When the input is lower than the value used in the zero calibration, flashing appears. Those outliers are outside the scope of span calibration. Registration POINT The calibration data will not be registered until you press with displayed. When the calibration is terminated, all settings established beforehand will be permanently erased. When the calibration results are reregistered, already stored fine adjustment and zero setting data will be erased. Therefore, reset these data on an as-necessary basis. For more information about fine adjustment and zero setting, see Section 3.6, "Fine adjustment" and Section 7.8, "Zero setting and effective range", respectively. (After the span calibration based on weight) 1. After confirming is displayed, press. The calibration data is registered, and CCAL mode is completed. 2. Press twice to return to Measurement mode. 39

42 3 Calibration 3.5 Calibration based on input of numerical value The calibration based on input of numerical value should be executed after you enter CCAL mode as shown below. Set the decimal point display position, unit weight, and measuring weight, conduct a series of subsequent setting procedures while pressing, and finally register the calibration data. Transition to the CCAL mode To execute calibration, enter CCAL mode. Setting of decimal point display position Display position. and set the decimal point display Setting of unit weight Setting of measuring weight Display and set the unit weight to be defined as the minimum scale. Display and set the measuring weight to be defined as the maximumm measuring value. Setting of weight mass Display and set the weight mass. Entry of zero point Entry of span Registration As long as all conditions listed below are met, display and enter the zero calibration data. The CSD-709 being used is replaced with another CSD-709. The zero calibration data (mv/v value) of the former CSD-709 is known. It is not possible to establish a no-load condition (initial condition including tare) for the measuring unit. Display. Subtract the input from the measuring unit in the zero calibration from the input from the measuring unit to which the load equal to the set measuring weight is applied. Set this calculation result. With displayed, press to registerr the calibration data. CAUTION Before the numerical value input based calibration, check the calibration number to be used. For the verification method of the calibration number, see Section 3.2, "Changing the calibration number". The strain-gage-based transducer is accompanied by the datasheet showing the following parameter values. In numerical value input based calibration, the load indicated in the "Capacity" column and voltage indicated in the "Rated output" column are used for both measuring weight and span. Capacity Rated Output Non-Linearity Hysteresis Input Resistance Output Resistance Zero Balance 40

43 3 Calibration Transition to CCAL mode 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press. CCAL mode is activated. Setting of decimal point display position POINT To terminate calibration, press to leave CCAL mode. When you leave CCAL mode during calibration, all settings established beforehand will be permanently erased. (After the transition to CCAL mode) 1. After confirming is displayed, press. The currently set decimal point display position appears. 2. Change the decimal point display position by using and on an as-needed basis. Whenever is pressed, the decimal point display position is cyclically changed from [0], [0.0], [0.00], and [0.000] to [0.0000], in ascending order. Similarly, whenever is pressed, the decimal point display position is cyclically changed in descending order. For procedures to set the decimal point display position, see Section 3.3, "Setting the decimal point display position". 3. Press. Setting of unit weight (After setting the decimal point display position) 1. After confirming is displayed, press. The currently set unit weight (minimum scale) appears and flashes. 2. Using and, set the unit weight. Only 1, 2, 5, or 10 can be set. When you have set the number of decimal places, the rightmost figure will be 1, 2, 5, or 10. Whenever is pressed, the unit weight is cyclically changed from 1, 2, and 5 to 10, in ascending order. Similarly, whenever is pressed, the unit weight is cyclically changed in descending order. POINT The set unit weight determines the increment of the value displayed in the load window. For example, when [2] is selected, the value in the load window is increased at an increment of 2 (e.g., ). Similarly, when [5] is selected, the value in the load window is increased at an increment of 5 (e.g., ). 41

44 3 Calibration 3. Press. Setting of measuring weight (After setting the unit weight) 1. After confirming is displayed, press. The currently set measuring weight (maximum measuring value) appears, and the number in the last digit flashes. 2. Using and, set the measuring weight. CAUTION The measuring weight does not accept [0]. When you select [0] and press, flashing appears and then reappears. For setting procedures with 3. Press. and, see Section 1.6, "Basic key operation". Setting of weight mass (After setting the measuring weight) 1. After confirming is displayed, press. The currently set weight mass appears, and the number in the last digit flashes. 2. Using and, set the weight mass. CAUTION The weight mass does not accept values exceeding the measuring weight or [0]. When you select either value and press, flashing appears and then reappears. POINT When entering the span, set the weight mass equal to the measuring weight. 3. Press. Entry of zero point (After setting the weight mass) 1. Confirm 2. Press is displayed. to display. Repressing redisplays CAUTION. 42

45 3 Calibration 3. Press. The mv/v value of the currently set zero point appears, and the number in the last digit flashes. 4. Using and, set the mv/v value of the zero point. CAUTION When the value set exceeds 2.5 mv/v, flashing value is set is lower than -2.5 mv/ /V, flashing outside the scope of zero point calibration. appears. When the appears. These outliers are 5. Press. Entry of span (After zero point entry) 1. Confirm 2. Press is displayed. to display. CAUTION Repressing redisplay. 3. Press. The mv/v value of the currently set span appears and the number in the last digit flashes. 4. Using and, set the mv/v value of the span. Subtract the input from the measuring unit in the zero calibration from the input from the measuring unit to which the load equal to the set measuring weight is applied. Set this calculation result as mv/v value of span. CAUTION When the span entered exceeds 3.1 mv/v, flashing appears. When the span entered is lower than the value used in the zero point adjustment, flashing appears. Those outliers are outside the scope of span calibration. 5. Press. 43

46 3 Calibration Registration POINT The calibration data will not be registered until you press with displayed. When the calibration is terminated, all settings established beforehand will be permanently erased. When the calibration results are reregistered, previously stored fine adjustment and zero setting data will be erased. Therefore, reset this data on an as-needed basis. For more information on fine adjustment and zero setting, see Section 3.6, "Fine adjustment" and Section 7.8, "Zero setting and effective range", respectively. (After the entry of span) 1. After confirming is displayed, press. The calibration data is registered, and CCAL mode is completed. 2. Press twice to return to Measurement mode. 44

47 3 Calibration 3.6 Fine adjustment Even after calibration, an error may be generated between actual and measured weights due to the change in the environment in which the CSD-709 is used. In such a case, fine adjustment should be implemented for zero point and span. Described below are the fine adjustment procedures. CAUTION Before fine adjustment, check the calibration number to be used. For the verification method of the calibration number, see Section 3.2, " Changing the calibration number". Transition to CADJ mode To implement the fine adjustment, enter CADJ mode. Do you want to fine-adjust the zero point No Yes Fine adjustment of zero point Display to "0". and set the zero-point load Do you want to fine-adjust the span? No Yes Fine adjustment of span Display, place a weight lower than the measuring weight on the measuring unit, and define the value displayed on the measuring unit as weight mass. Registration With the data. displayed, presss to register POINT You can implement only zero point fine adjustment or span fine adjustment. For example, when it is necessary to only fine-adjust the span with the fine adjustment of zero point omitted, press or in CADJ mode to change the display from to. When it is necessary to complete the work without implementing the fine adjustment of span after the fine adjustment of zero point, press or with displayed to display and press. 45

48 3 Calibration 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press once to change the display to. 5. Press. The CADJ mode is activated and appears. POINT To terminate the fine adjustment, presss to leave CADJ mode. When you leave CADJ mode during the process, all settings established beforehand will be permanently erased. 6. Press. The current zero-point load flashes. 7. Using and, set the zero-point load to [0]. Holding down decreases the value continuously. Similarly, holding down the value continuously. increases Be careful that load fluctuation induced by vibration, for example, will not be exerted on the measuring unit during calibration. When the input from the measuring unit exceeds 2.5 mv/v, flashing appears. When the input is lower than -2.5 mv/v, flashing appears. These outliers are outside the scope of zero point calibration. 8. Press. appears. CAUTION 9. Place the weight lower than the set measuring weight to the measuring unit and presss. The input from the measuring unit flashes. 10. Using and, set the displayed value to the weight mass. CAUTION When the input from the measuring unit exceeds 3.1 mv/v, flashing appears. When the input is lower than the value used in the zero point adjustment, flashing appears. These outliers are outside the scope of span calibration. 11. Press. appears. 46

49 3 Calibration 12. Press. The calibration data is adjusted and CADJ mode is completed. POINT The fine adjustment data will not be registered until you press with displayed. When the fine adjustment is terminated, all settings established beforehand will be permanently erased. 13. Press twice to return to Measurement mode. 47

50 3 Calibration 3.7 Simplified calibration on an as-needed basis. The CSD-709 can execute zero and span calibrations by using simplified procedures. Also, it offers numeral value input based calibration using simplified procedures. This section describes these simplified calibration procedures by dividing calibration into the following processes. Simplified zero calibration Simplified span calibration Simplified numerical value input based calibration CAUTION Before each simplified calibration, check the calibration number to be used. For the verification method of the calibration number, see Section 3.2, "Changing the calibration number". If appears when you attempt to execute simplified calibration, such calibration is locked. In this case, cancel the lock and commence with calibration. For unlocking procedures, see Section 7.3.1, "Locking the simplified calibration" Simplified zero calibration In simplified zero calibration, you should calibrate the zero point based on the output with no load placed on the measuring unit. 1. Hold down to display. 2. With no load placed on the measuring unit, press. flashes. Then, continues flashing to allow you to return to Measurement mode. CAUTION When the input from the measuring unit exceeds 2.5 mv/v, flashing appears. When the input is lower than -2.5 mv/v, flashing These outliers are outside the scope of zero point calibration appears Simplified span calibration In simplified span calibration, you should calibrate the span using the weight prepared. 1. Hold down to flash. 2. Press. The currently set measuring weight appears. 3. Change the measuring weight using and 48

51 3 Calibration CAUTION The measuring weight does not accept [0]. When you select [0] and press, flashing appears and then reappears and flashes. For setting procedures with 4. Press. flashes. and, see Section 1.6, "Basic key operation". 5. Press. The currently set weight mass appears. 6. Similar to the case of measuring weight, change the weight mass on an as-needed basis. The weight mass does not accept the value exceeding the measuring weight or [0]. When you select either value and press, flashing appears and then reappears and flashes. 7. Press. appears. CAUTION 8. Place the load equal to the set weight mass to the measuring unit. 9. Press. flashes. Then, mode. continues flashing to allow you to return to Measurement CAUTION When the input from the measuring unit exceeds 3.1 mv/v, flashing appears. When the input is lower than 0.0 mv/v, flashing appears. These outliers are outside the scope of span calibration. 49

52 3 Calibration Simplified numerical value input based calibration In the simplified numerical value input based calibration, you should check the following parameter settings one by one while pressing, and change the settings on an as-needed basis. Decimal point display position Unit weight Measuring weight mv/vv value of zero point mv/vv value of span 1. Hold down to display and flash. 2. Press. The currently set decimal point display position flashes. 3. Change the decimal point display position by using and on an as-needed basis. Whenever is pressed, the decimal point display position is cyclically changed from [0], [0.0], [0.00], and [0.000] to [ ], in ascending order. Similarly, whenever is pressed, the decimal point display position is cyclically changed in descending order. For setting procedures of decimal point display position, see Section 3.3, "Setting the decimal point display position". 4. Press. flashes. 5. Press. The currently set unit weight flashes. 6. Similar to the case of decimal point display position, change the unit weight on an as-needed basis. Only 1, 2, 5, or 10 can be set. When you have set the number of digits in the decimal place, the rightmost figure will be 1, 2, 5, or 10. Whenever is pressed, the unit weight is cyclically changed from 1, 2, and 5 to 10, in ascending order. Similarly, whenever is pressed, the unit weight is cyclically changed in descending order. 7. Press. flashes. 8. Press. The currently set measuring weight appears. 9. Change the measuring weight by using and on an as-needed basis. CAUTION The measuring weight does not accept [0]. When you select [0] and press, flashing appears and then reappears and flashes. For setting procedures with and, see Section 1.6, "Basic key operation". 50

53 3 Calibration 10. Pres ss. flashes. 11. Press. The mv/v value of the currently set zero point appears. 12. Similar to the case of measuring weight, change the zero point value on an as-neededd basis. CAUTION When the zero point value exceeds 2.5 mv/v, flashing appears. When the value is lower than -2.5 mv/v, flashing appears. These outliers are outside the scope of zero point calibration. 13. Press. The display changes to. CAUTION When the span exceeds 3.1 mv/v, flashing lower than the mv/v value of zero point, flashing are outside the scope of span calibration. 14. Press. The mv/v value of the currently set span appears. 15. Similar to the case of mv/v value of zero point, change the span on an as-needed basis. appears. When the value is appears. These outliers 16. Press. flashes. 17. Press. flashes and the Measurement mode becomes active. POINT The calibration data will not be registered until you press with flashing. When calibration is erminated, all settings established beforehand will be permanently erased. 51

54 3 Calibration 3.8 Verifying the registered calibration data [CF-50] to [CF-69] You can check the registered calibration data by referencing the following C functions. Calibration data Stored data Referencing C function Unit weight CF-50 Measuring weight CF-51 Calibration data 0 Weight mass CF-52 Zero calibration mv/v value CF-53 Span calibration mv/v value CF-54 Unit weight CF-55 Measuring weight CF-56 Calibration data 1 Weight mass CF-57 Zero calibration mv/v value CF-58 Span calibration mv/v value CF-59 Unit weight CF-60 Measuring weight CF-61 Calibration data 2 Weight mass CF-62 Zero calibration mv/v value CF-63 Span calibration mv/v value CF-64 Unit weight CF-65 Measuring weight CF-66 Calibration data 3 Weight mass CF-67 Zero calibration mv/v value CF-68 Span calibration mv/v value CF-69 For example, to check the data from the unit weight to span calibration mv/v value in the calibration data 0, perform the following steps. 1. Press and in this order to change the display to. 2. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 3. Using and, change the numerical value to and press. The unit weight stored in calibration data 0 appears. For setting procedures with and, see Section 1.6, "Basic key operation". 4. Press again. appears. Whenever is pressed, the contents stored in the functions subsequent to CF-51 appear in order. 5. To end the confirmation, press twice to return to Measurement mode. 52

55 4 Settings Related to Measurement 4 Settings Related to Measurement 4.1 Overview The CSD-709 processes the inputs from the measuring units and transforms the results into calculated values. Analog signal Data sampling and AD conversion Digital low-pass filter (set per AD sampling frequency) Stabilization based on moving average Calculated value The sampling frequency for AD conversion determines the reproducibility of the digital data against the inputs. Sampling frequency for AD conversion Low High Influence of noise Less susceptible Highly susceptible Reproducibility of noise Low High The digital low-pass filter and moving average determine the magnitude of influence of unexpectedly generated noises. Digital low-pass filter Low frequency High frequency Influence of noise Less susceptible Highly susceptible Reproducibility of noise Low High Moving average frequency Low High Influence of noise Less susceptible Highly susceptible Processing time Short Long 53

56 4 Settings Related to Measurement The CSD-709, by enabling the analog peak detection circuit, can avoid the delay induced by digital processing. This time, you should set the analog filter to suppress noises contained in the inputs from the measuring unit. CAUTION When the digital low-pass filter and moving average are incorrectly set, correct measured values will not be generated and peripheral devices may be damaged as a result. 4.2 Setting the AD sampling frequency [CF-02] The measured value is transferred to the CSD-709 in the form of an analog signal. How many sampling times the equipment executes in one second to convert analog signals into digital data is called the [AD sampling frequency]. The AD sampling frequency is set to C function CF-02. Described below are the AD sampling frequency setting procedures. POINT CSD-709 s comparator conversion frequency and the current and voltage output frequencies of optional devices synchronize with the AD sampling frequency. For this reason, to change the comparator conversion frequency, current output frequency, or voltage output frequency, change the AD sampling frequency. 1. Press once to display. 2. Press once to change the display to. 3. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 4. Using and, change the numerical value to and press. The number showing the current AD sampling frequency appears. Number Default AD Sampling Frequency 40 times/s 400 times/s times/s 2 (4 000 times/s) For setting procedures with and, see Section 1.6, "Basic key operation". 54

57 4 Settings Related to Measurement 5. Press to set the number showing the current AD sampling frequency. Whenever is pressed, the number is cyclically changed among 0, 1, and 2, in ascending order. 6. Press. 7. Press twice to return to Measurement mode. 55

58 4 Settings Related to Measurement 4.3 Setting the digital filter [F-05], [F-06], [F-07] The CSD-709 is equipped with two digital filters shown below to suppress the influence of noises contained in the inputs from the measuring unit. Digital low-pass filter Moving average The digital low-pass filter can be set for every AD sampling frequency Setting the digital low-pass filter The digital low-pass filter values for the dataa digitalized according to the AD sampling frequency are set to functions F-05, F-06, and F-07. In each function, the digital low-pass filter value for any of the following AD sampling frequencies is stored. Function Related AD sampling frequency Settable band F times/s 0 (OFF) ~ 4.0 Hz F times/s 0 (OFF) ~ 40 Hz F times/s 0 (OFF) ~ 400 Hz CAUTION According to the AD sampling frequency set to C function CF-02, enter an appropriate value to any of the functions F-05, F-06, and F-07. For example, when CF-02 has been set to [2] (4 000 times/s), enter an appropriate value to function F-07. In this case, values set to F-05 and F-06 will produce no effect. POINT To remove lower frequency noises, reduce the AD sampling frequency to set a lower digital low-pass filter value. Described below are procedures to set the digital filter for the AD sampling frequency. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed with [F-]. 3. According to the currently set AD sampling frequency, use numerical value to any of, and The currently set digital filter value appears. and to change the, and press. Default [F-05]: 40 (4 Hz), [F-06]: 40 (40 Hz), [F-07]: 40 (400 Hz) For setting procedures with and, see Section 1.6, "Basic key operation". 56

59 4 Settings Related to Measurement 4. Using and, set the digital filter value. You can set the value in the following frequency ranges depending on the target function. Function Range of settable value Applicable band F-05 0 ~ 40 0 (OFF) ~ 4.0 Hz (setting 0.1 Hz) F-06 0 ~ 40 0 (OFF) ~ 40 Hz (setting 1 Hz) F-07 0 ~ 40 0 (OFF) ~ 400 Hz (setting 10 Hz) 5. Press. 6. Press twice to return to Measurement mode Setting the moving average [F-01] To gain higher stability, the number of moving average samples should be set for the data processed by the digital low-pass filter. The number of moving average samples can be set to function F-01 by using any value within a range from 1 to 999. Described below are setting procedures of the number of moving average samples. POINT Usually, narrow the frequency band (responsibility) used in the digital low-pass filter to roughly remove noises and stabilize the data by removing minute fluctuations in the moving average. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed with [F-]. 3. Using and, change the numerical value to and press. The currently set number of moving average samples appears. Default 80 For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set the number of moving average samples and press. 5. Press twice to return to Measurement mode. 57

60 4 Settings Related to Measurement 4.4 Enabling the analog peak detection circuit [F-12] To avoid the delay induced by the digital processing of inputs from the measuring unit, the analog peak detection circuit should be enabled. You can enable or disable the analog peak detection circuit with function F-12. Described below are the procedures used to enable the analog peak detection circuit. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed with [F-]. 3. Using and, change the numerical value to and press. The number showing the current status of the analog peak detection circuit appears. Number Status of analog peak detection circuit 0 Disable 1 Enable Default 0 (Disable) For setting procedures with and, see Section 1.6, "Basic key operation". 4. To enable the analog peak detection circuit, press to change the display to [1]. 5. Press. 6. Press twice to return to Measurement mode. 58

61 4 Settings Related to Measurement Setting the analog filter [F-02] While the analog peak detection circuit is being disabled, the analog filter is fixed to 30 khz. After the analog peak detection circuit is enabled, you should set the analog filter to suppress the influence of noise. You can set the desired analog filter with function F-02. Described below are the analog filter setting procedures. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed with [F-]. 3. Using and, change the numerical value to and press. The number showing the current analog filter setting appears. Number Analog filter setting 0 10 Hz 1 40 Hz Hz Hz 4 1 khz 5 4 khz 6 10 khz 7 30 khz Default 7 (30 khz) For setting procedures with and, see Section 1.6, "Basic key operation". 4. Press to set the number showing the analog filter setting. Whenever is pressed, the number is cyclically changed from 0 to 7, in ascending order. 5. Press. 6. Press twice to return to Measurement mode. 59

62 5 Settings Related to Detection 5 Settings Related to Detection 5.1 Setting the detection conditions This section describes the procedures to set detection conditions. To set the detection conditions, determine the detection condition number first. Then, set the detecting section and detecting operation Detection condition number [PF-09] The CSD-709 can store a combination of detecting section and detecting operation by using eight detection condition numbers (0 to 7). You can select and use the desired number. The desired detection condition number can be selected using P function PF-09. Before setting the detecting section and detecting operation, you should determine the detection condition number to which each combination of these parameters will be stored. Described below are the detection condition number setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set detection condition number appears. Default 0 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set the detection condition number. Whenever is pressed, the detection condition number is cyclically changed from 0 to 7, in ascending order. 6. Press. 7. Press twice to return to Measurement mode Setting the detecting section [PF-02] Set the detecting section to P function PF-02. Described below are the detecting section setting procedures. 1. Press once to display. 2. Press twice to change the display to. 60

63 5 Settings Related to Detection 3. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The number showing the currently set detecting section appears. Number Detecting section 0 Entire section 1 Specified section 2 Time-specified section 3 Automatic start time-specified section Default 0 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the value specifying the detecting section. Whenever is pressed, the number showing the detecting section is cyclically changed from 0 to 3, in ascending order. 6. Press. 7. Press twice to return to Measurement mode. Setting the detecting time [PF-40] When you have selected [2] (Time-specified section) or [3] (Automatic start time-specified section) as the desired detecting section, you should set the detecting time to P function PF-40. Described below are the detecting time setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set detecting time appears in [ms]. Default ms For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the detecting time. You can set any value between 0 and ms. 6. Press. 7. Press twice to return to Measurement mode. 61

64 5 Settings Related to Detection Setting the automatic start level [PF-41] When you have selected [3] (Automatic start time-specified section) as the desired detecting section, you should set the threshold automatic start level to P function PF-41. Described below are the automatic start level setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set automatic start level appears. Default For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the automatic start level. You can set any value between and , which will be displayed in the load window. 6. Press. 7. Press twice to return to Measurement mode Setting the detection operation [PF-01] Set the detection operation to P function PF-01. Described below are the detection operation setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 62

65 5 Settings Related to Detection 4. Using and, change the numerical value to and press. The number showing the currently set detection operation appears. Number Detection operation 00 Peak 01 Bottom 02 Peak-bottom 03 Peak-to-peak 04 Minimum 05 Maximum 06 Maximum-minimum difference 07 Inflection point A 08 Inflection point B 09 Inflection point C 10 Inflection point D 11 Average Default 00 (Peak) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the number showing the detection operation. 6. Press. 7. Press twice to return to Measurement mode. Setting the reference point [PF-10] When you have selected [02] (Peak-bottom) as the desired detection operation, you should set the reference point to P function PF-10. Described below are the reference point setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set reference point appears. Default 0 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the reference point. You can set any value between and , which will be displayed in the load window. 6. Press. 7. Press twice to return to Measurement mode. 63

66 5 Settings Related to Detection Setting the detecting data width and detecting magnification [PF-11], [PF-12] When you have selected [05] (Maximum) as the desired detection operation, you should set the detecting data width to P function PF-11. When you have selected [04] (Minimum) or [06] (Maximum-minimum difference) as the desired detection operation, you should set the detecting magnification to PF-12, in addition to setting the detecting data width to PF-11. Described below are the setting procedures for the detecting data width and detecting magnification. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set detecting data width appears. Default 200 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the detecting data width. You can set any value between 0 and , which will be displayed in the load window. 6. Press. 7. When you want to set the detecting magnification, use and to change the numerical value to. When you do not want to set the detecting magnification, press twice to return to Measurement mode. 8. Press. The currently set detecting magnification appears. Default 15 (1.5 times) 9. Using and, set the detecting magnification. You can set any value between 0 and 49 to PF-12. The actual detecting magnification will be the PF-12 value multiplied by Press. 11. Press twice to return to Measurement mode. 64

67 5 Settings Related to Detection Setting the inflection point [PF-20] to [PF-35] When you have selected any of numbers between [07] (Inflection point A) and [10] (Inflection point D) as the desired detection operation, you should set the following P functions. Described below are the setting procedures for inflection point A. For the setting of inflection point B, C, or D, use P function connected to the relevant point in the procedures below. Set inflection poin A B C D 1. Press once to display. Necessary P function PF-20 (Detecting width data), PF-21 (Detecting time A), PF-22 (Detecting time B), PF-23 (Fine adjustment in detection) PF-24 (Detecting width data), PF-25 (Detecting time A), PF-26 (Detecting time B), PF-27 (Fine adjustment in detection) PF-28 (Detecting width data), PF-29 (Detecting time A), PF-30 (Detecting time B), PF-31 (Fine adjustment in detection) PF-32 (Detecting width data), PF-33 (Detecting time A), PF-34 (Detecting time B), PF-35 (Fine adjustment in detection) 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set detecting data width appears. Default 100 For setting procedures with and see Section 1.6, "Basic key operation". 5. Using and, set the detecting data width. You can set any value between and , which will be displayed in the load window. 6. Press. 7. After confirming is displayed, press. The currently set detecting time A width appears in [ms]. Default 100 ms 8. Using and, set the detecting time. You can set any value between 0 and 450 ms. 9. Press. 10. After confirming is displayed, press. Set the detecting time B using Step 8 and press. Default 100 ms 65

68 5 Settings Related to Detection 11. After confirming is displayed, press. The currently set fine adjustment in detection appears. Default Using and, set the fine adjustment in detection. You can set any value between 0 and 10, which will be displayed in the load window. 13. Press. 14. Press twice to return to Measurement mode. Setting the average update time [PF-42] When you have selected [11] (Average) as the desired detection operation, you should set the average update time to P function PF-42. Described below are the average update time setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set average update time appears in [ms]. Default 50 ms For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the average update time. You can set any value between 50 ms and ms. 6. Press. 7. Press twice to return to Measurement mode Setting example of detecting condition This section gives examples of setting four detecting conditions below. Detection of a peak in the entire section Detection of a bottom in the specified section Detection of a maximum in the time-specified section Detection of an inflection point in the automatic start time-specified section 66

69 5 Settings Related to Detection Detection of a peak in the entire section To detect a peak with the entire section defined as the desired detecting section, set the following values to P functions PF-01 and PF-02, respectively. P function PF-01 (Detecting operation) PF-02 (Detecting section) Setting 00 (Peak) 0 (Entire section) Under these settings, the peak is determined as shown below according to the change in load. When the detection operation initiated through PEAK/TRACK input is generated by, for example, pressing, point A is determined as a peak. When the load changes thereafter and exceeds the value of point A, the peak changes according to the track, and, when the load reaches point B, the peak will be the value of point B. Similarly, when the load exceeds the value of point B, the peak changes according to the track. When the RESET input is generated by, for example, pressing, the process enters into the reset section and the displayed load becomes [0]. When the process leaves the reset section, it reenters the detecting section and the value of point D will be the peak until the PEAK/TRACK input is ended by, for example, pressing. 67

70 5 Settings Related to Detection Detection of a bottom in the specified section To detect a bottom with the specified section defined as the desired detecting section, set the following values to P functions PF-01 and PF-02, respectively. P function PF-01 (Detecting operation) PF-02 (Detecting section) Setting 01 (Bottom) 1 (Specified section) Under these settings, the bottom is determined as shown below according to the change in load. When the detection operation initiated through of PEAK/TRACK input is generated by, for example, pressing, point A is determined as a bottom. When the load changes thereafter and falls below the value of point A, the bottom changes according to the track, and, when the process enters into the hold section by, for example, pressing, the bottom will be the value of point B. When the process returns to the detecting section from the hold section, the value of point C is determined as a bottom. When the RESET input is generated by, for example, pressing in the detecting section, the process enters the reset section and the displayed load becomes [0]. When the RESET input is ended, the process reenters the detecting section, and, when the PEAK/TRACK input is generated by, for example, pressing, the process enters the hold section. When the RESET input is generated in the hold section, the detection process ends. 68

71 5 Settings Related to Detection Detection of a maximum in the time-specified section To detect a maximum with the time-specified section defined as the desired detecting section, set the following values to P functions PF-01, PF-02, PF-11, and PF-40, respectively. P function PF-01 (Detecting operation) PF-02 (Detecting section) PF-11 (Detecting data width) PF-40 (Detecting time) Setting 05 (Maximum) 2 (Time-specified section) (Display in the load window) (ms) Under these settings, the maximum is determined as shown below according to the change in load. When the detecting operation is initiated through input of PEAK/TRACK by, for example, pressing, the section within the detecting time already set is defined as the detecting section. When a maximum (point A) is detected in the detecting section, the subsequent section will be the hold section. When the RESET input is generated by, for example, pressing in the detecting section, the process enters the reset section and the displayed load becomes [0]. When the RESET input is ended, the process reenters the detecting section. When the RESET input is generated in the hold section, the detection process ends. 69

72 5 Settings Related to Detection Detection of an inflection point in the automatic start time-specified section To detect an inflection point with the automatic start time-specified section defined as the desired detecting section, set the following values to P functions PF-01, PF-02, PF-40, and PF-41, and PF-20 to PF-35, which are used depending on the sharp change of the load to be detected, respectively. P function Setting PF-01 (Detecting operation) 07 (Inflection point A) PF-02 (Detecting section) 3 (Automatic start time-specified section) PF-20 (Detecting data width) 100 (Display in the load window) PF-21 (Detecting time A) 100 (ms) PF-22 (Detecting time B) 100 (ms) PF-23 (Fine adjustment in detection) 0 (Display in the load window) PF-40 (Detecting time) (ms) PF-41 (Automatic start level) (Display in the load window) Under these settings, the inflection point is determined as shown below according to the change in load. When the track exceeds the automatic start level, the detection operation is initiated and the section within the detecting time already set is defined as the detecting section. When an inflection point (point A) is detected in the detecting section, the subsequent section will be the hold section. When the RESET input is generated by, for example, pressing in the detecting section, the process enters the reset section and the displayed load becomes [0]. When the RESET input is ended, the process reenters the detecting section. When the RESET input is generated in the hold section, the detection process ends. 70

73 5 Settings Related to Detection Setting the detection prohibit time [PF-43] To exclude the impact load generated in the pressure fitting from the parameters subject to peak detection, set the detection prohibit time to P functions PF-43. If this time is set, the detection is not implemented until the specified time counted after the initiation of detection process has expired. After this time elapses, parameters such as peaks are detected. Described below are the detection prohibition time setting procedures. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently set detection prohibit time appears. Actual time will be the displayed value multiplied by 0.1. Default 00 (0.0 s) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the detection prohibit time. You can set any value between 00 and 10 that show 0.0 to 1.0 s, respectively. 6. Press. 7. Press twice to return to Measurement mode. 71

74 5 Settings Related to Detection 5.2 Detection with multimode The CSD-709 can combine detecting sections and detecting operations, all of which are identified by eight detecting condition numbers, in order to obtain an arbitrary detection result from the track changing with the elapse of time. Up to four combinations are achievable. In the multimode where multiple detecting conditions are combined in such the manner, the following preparation is needed. (1) Assign unique detecting condition numbers to detecting sections and operations used in the multimode, respectively. (2) Change the mode from standard single mode to multimode. (3) Assign the established detecting condition numbers to Areas 1 to 4 in the execution order. For detecting condition number setting procedures, see Section 5.1.1, "Detecting condition number". For more information about detection methods in the multimode, see the instruction manual attached to the CSD Entering into the multimode [PF-03] To change the detection under single condition to the detection based on the multimode, you should set [1] to P function PF-03. Described below are setting procedures to enter multimode. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to and press. The currently setting appears. For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set [1] showing the multimode and press. 6. Press twice to return to Measurement mode. 72

75 5 Settings Related to Detection Assigning the execution order [PF-90], [PF-91], [PF-92], [PF-93] In the multimode, the detecting conditions respectively assigned to Areas 1 to 4 are executed in order. To execute the detection in the multimode, set the detecting condition number to Area 1 at minimum. In the multimode, the detecting conditions are executed from Area 1 in order. When the process using the detecting condition set to Area 4 is completed or the process enters into the area having no detecting condition, the multimode is ended immediately. To actually assign detecting conditions to those areas, set detection condition numbers to the following P functions PF-90 to PF-93, respectively. P function Area PF-90 Area 1 PF-91 Area 2 PF-92 Area 3 PF-93 Area 4 Described below are the setting procedures for detection condition execution orders. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to any of,,, and, and press. The currently set detection condition number appears. Default [PF-90]: 0, [PF-91], [PF-92], [PF-93]: -1 (not used) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set the detecting condition number to be executed to any value between 0 and 7. To exclude the area from multimode, set [-1]. 6. Press. 7. To set detecting condition numbers to other areas, repeat Steps 4 to Press twice to return to Measurement mode. 73

76 5 Settings Related to Detection Example of measurement in multimode This section shows a setting example where the following four detecting conditions are assigned to Areas 1 to 4 to execute operations in multimode. Detection of a peak in the entire section Detection of a bottom in the specified section Detection of a maximum in the time-specified section Detection of an inflection point in the automatic start time-specified section First, set the respective detection conditions while setting the detection condition numbers as described below. Set [0] to PF-09 and establish the conditions necessary to detect a peak in the entire section. Set [1] to PF-09 and establish the conditions necessary to detect a bottom in the specified section. Set [2] to PF-09 and establish the conditions necessary to detect a maximum in the time-specified section. Set [3] to PF-09 and establish the conditions necessary to detect an inflection point in the automatic start time-specified section. Upon completion of these settings, set the following values to PF-03, PF-90, PF-91, PF-92, and PF-93, respectively. P function PF-03 (Change to multimode) PF-90 (Area 1) PF-91 (Area 2) PF-92 (Area 3) PF-93 (Area 4) Setting 1 (Multimode) 0 (Detecting condition number) 1 (Detecting condition number) 2 (Detecting condition number) 3 (Detecting condition number) When you give the PEAK/TRACK input by, for example, pressing connected to the detection condition set to Area 1 after the establishment of settings, the detection is initiated. This time, operations triggered by the search conditions set to these areas are the same as those executed individually by means of the respective detection conditions. POINT During measurement in multimode, the external control output, EXCUTE, is turned on. When [7] has been set to function F-55 beforehand, you can forcibly terminate the process by pressing during measurement in multimode. For setting procedures of, see Section 7.7, "Setting the F key operation". 74

77 5 Settings Related to Detection 5.3 Detection result judgment and waveform comparison The detection result judgment feature allows you to check that the peak in the detecting section is in a range from the upper limit set to P functionn P-51 to the lower limit set to P-52. Also, the waveform comparison feature allows you to check thatt the total value (track) from the start to end of measurement is in a range from the upper limit set to P-51 to the lower limit set to P-52 at the end of the detecting section or the completion of multimode. CAUTION The comparison of waveforms can only be executed on the CSD equipped with an optional memory card Detection result judgment [P-50], [P-51], [P-52] The detection result judgment feature can be used in the following sections. Entire section Specified section Time-specified section Automatic start time-specified section To execute the detection result judgment, set [1] (Detection result judgment) to P function P-50 and the upper and lower limits of judgment to P-51 and P-52, respectively. The result of the detection result judgment is obtained in the form of external control output by setting any of the following numerical values to F-70 to F-75. Numerical value Function 07 Shows the result of the detection result judgment is acceptable. 08 Shows the result of the detection result judgment is not acceptable. 09 Shows whether the result of the detection result judgment exceeds the upper limit. For settings related to the external control output, see Section 8.2, "External control output signal". When the detection is actually initiated, external control outputs to which [07] to [09] are set are turned offf between the start and end of the detecting section. When the detection is completed, results are output from those external control outputs to which [07] to [09] are set. In the following example, the result does not deviate from both the upper and lower limits and thus only the external control output to which [07] is set will be turned on after the detection. 75

78 5 Settings Related to Detection In the following example, the result exceeds the upper limit and thus external control outputs to which [08] and [09] are set will be turned on after the detection. In the following example, the result deviates from the lower limit and thus the external control output to which [08] is set will be turned on after the detection. As shown above, the detection result judgment allows you to judge that the peak is in the range determined by the upper and lower limits of the detecting section. Described below are setting procedures of the detection result judgment. 76

79 5 Settings Related to Detection 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 4. Using and, change the numerical value to The number showing the current setting appears. and press. Number Default Setting details Disable Detection result judgment Waveform comparison 0 (Disable) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set [1] and press. 6. Press. 7. After confirming is displayed, press The currently set upper limit appears.. Default Using and, set the upper limit. You can set any value between and Press. 10. After confirming is displayed, press The currently set lower limit appears.. Default Using and, set the lower limit. You can set any value between and Press twice to return to Measurement mode Waveform comparison [P-50], [P-51], [P-52] CAUTION The comparison of waveforms can only be executed on the CSD equipped with the optional memory card. 77

80 5 Settings Related to Detection The waveform comparison feature can be used in the following sections. Time-specified section Automatic start time-specified section To execute the waveform comparison, set [2] (Waveform comparison) to P function P-50 and the upper and lower limits of judgment to P-51 and P-52, respectively. The result of the waveform comparison is obtained in the form of external control output by setting any of the following numerical values to any function between F-70 and F-75. Numerical value Function 07 Shows the result of the detection result judgment is acceptable. 08 Shows the result of the detection result judgment is not acceptable. 09 Shows whether the result of the detection result judgment exceeds the upper limit. For the setting relative to the external control output, see Section 8.2, "External control output signal". The maximum for the data used in the waveform comparison is Therefore, the longest time that can be used in the comparison is determined by the AD sampling frequency set to C function CF-02, as shown below. Setting of CF-02 Setting details 0 (40 times/s) 100 s 1 (400 times/s) 10 s 2 (4 000 times/s) 1 s Default 0 (Disable) For this reason, when the detecting time of the time-specified section or automatic start time-specified section exceeds the maximum time, the data will be updated whenever the longest time has elapsed. When detection is actually initiated, external control outputs characterized by [07] to [09] that are set to any functions between F-70 and F-75 are turned off from the start to the end of the detecting section. When the detection is completed, results are output from those external control outputs to which [07] to [09] are set. In the following example, the result does not deviate from both the upper and lower limits and thus only the external control outputs to which [07] is set will be turned on after the detection. 78

81 5 Settings Related to Detection In the following example, the result exceeds the upper limit and thus external control outputs to which [08] and [09] are set will be turned on after the detection. In the following example, the result deviates from the lower limit and thus the external control output to which [08] is set will be turned on after the detection. As shown above, the waveform comparison allows you to judge that the total value (track) is in the range determined by the upper and lower limits of the detecting section. Described below are the setting procedures of the waveform comparison. 1. Press once to display. 2. Press twice to change the display to. 3. Press. The P function mode is activated, and the display changes to a two-digit numerical value prefixed with [PF-]. 79

82 5 Settings Related to Detection 4. Using and, change the numerical value to The value showing the current setting appears. and press. Numerical value Default Setting details Disable Detection result judgment Waveform comparison 0 (Disable) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using, set [2] and press. When [0] (Entire sect appears. CAUTION ion) or [1] (Specified section) has been set to P function PF-02, After [2] (Waveform comparison) is set to PF-50, section) will automatically set PF-50 to [0] changingg the setting of PF-02 to [0] (Entire section) or [1] (Specified (Disable). 6. Press. 7. After confirming is displayed, press The currently set upper limit appears.. Default Using and, set the upper limit. You can set any value between and Press. 10. After confirming is displayed, press The currently set lower limit appears.. Default Using and, set the lower limit. You can set any value between and Press twice to return to Measurement mode. 80

83 6 Settings Related to Comparison 6 Settings Related to Comparison 6.1 Overview The CSD-709 is equipped with a comparator that compares the track or peak to three preset values to turn on the open collector output when the track or peak deviates from this setting. You can set values used in the comparator in three stages: S0, S1, and S2. In addition, settings with respect to the comparator can be stored using four brand codes connected to four calibration numbers, 0 to 3. In the comparison with the comparator, use F function F-55 to set the F key function to the COMP setting mode and press to enter comparison values to S0 to S2, respectively. Select the track or peak as the target to be compared to S0 to S2 and set the operating conditions. For example, when you select the peak as the target to be compared to S1 and provide a setting such that the open collector output will be turned on when the displayed load becomes equal to or greater than S1, the open collector output is configured as shown below: Displayed load: Low S1 setting Displayed load: High S1 judgment display S1 open collector output ON OFF In the comparison with the comparator, you can set the delay based on the data width and duration. This delay suppresses the influence of noise to ensure an accurate comparison process. For example, [1] (Turn on the open collector output when the comparison target is equal to or greater than the comparison value) has been set, you can delay actual turn-on timing using the data width and duration. This operation is called [ON delay] By setting the ON delay appropriately, you can configure different open collector outputs, as shown below. Display Hysteresis data width S1 setting Hysteresis duration Time S1 open collector output when data width and duration are not set S1 open collector output when data width and duration are set ON OFF ON OFF 81

84 6 Settings Related to Comparison Similarly, if [2] (Turn on the open collector output when the comparison target is equal to or lower than the comparison value) has been set, you can delay the actual turn-on timing using the ON delay feature. If [1] (Turn on the open collector output when the comparison target is equal to or greater than the comparison value) or [2] (Turn on the open collector output when the comparison target is equal to or lower than the comparison value) has been set, you can also delay actual turn-off timing using data width and duration. This operation is called [OFF delay] By setting the OFF delay appropriately, you can configure different open collector outputs, as shown below: Hysteresis duration Display Hysteresis data width S1 setting Time S1 open collector output when data width and duration are not set S1 open collector output when data width and duration are set ON OFF ON OFF The operation based on the ON delay or OFF delay is called [comparator hysteresis]. 82

85 6 Settings Related to Comparison 6.2 Setting the comparison value To use the comparator to compare the track or peak to preset values, set the comparison values to be used. You can set up to three comparison values in the form of S0 to S2 to comparator brand codes connected to calibration numbers 0 to 3. This time, the comparison values S0 to S2 are stored to the connected calibration numbers, respectively. Therefore, it is necessary to select the calibration number to be connected to the comparison value before the setting work. Described below are the setting procedures for comparison values S0 to S2. For calibration number change procedures, see 3.2, "Changing the calibration number". 1. Press once to display. 2. Press five times to change the display to. 3. Press. The COMP selection mode is activated, and flashes. Once the F key function is set to the COMP setting mode, you can shift to the COMP setting mode from Measurement mode by simply pressing. For the setting procedures of the F key, see Section 7.7, "Setting the F key operation". 4. Press. The value currently set to S0 appears. 5. Using and, set the value of S0. You can set any value between and , which will be displayed in the load window. For setting procedures with and, see Section 1.6, "Basic key operation". 6. Press. 7. Check that is flashing. Press to set the value of S1 in the same way as in Step 5, then press. 8. Check that is flashing. Press to set the value of S2 in the same way as in Step 5, then press. 9. Press twice to return to Measurement mode. 83

86 6 Settings Related to Comparison 6.3 Setting the comparator operating conditionn [F-21], [F-22] Which parameter, track, or peak to be defined as a comparison target in the comparator is set to Function F-21. You should set the post-comparison action to Function F-22. Described below are the setting procedures for comparison targets and operating conditions. CAUTION If you fail to correctly set the on/off operation of the open collector output for the comparison value, peripheral devices may malfunction. POINT When or is observed in the load window, the displayed value used in the comparison by the comparator is handled as [+ (infinity)] or [- (infinity)]. 1. Press once to display. 2. Press. Function mode is activated, and the display changes to two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. The three-digit number showing the currently set comparison target appears. Digit Comparison target Third digit S2 Second digit S1 First digit S0 The comparison target is indicated by [0] (Track) or [1] (Peak). Default 000 For setting procedures with and, see Section 1.6 "Basic key operation". 4. Using and, set the comparison target. 5. Press. 84

87 6 Settings Related to Comparison 6. After confirming is displayed, press. The three-digit number showing the currently set operating condition appears. Digit Third digit Second digit First digit Operating condition S2 S1 S0 The operating condition is indicated by [0] (Always turn off the open collector output), [1] (Turn on the open collector output when the comparison target is equal to or greater than the comparison value), or [2] (Turn on the open collector output when the comparison target is equal to or lower than the comparison value). Default Using and, set the operating condition. 8. Press. 9. Press twice to return to Measurement mode. 85

88 6 Settings Related to Comparison 6.4 Setting the comparator hysteresis [ F-24], [F-25], [F-26] To delay the open collector turn-on or turn-off timing by setting the data width or duration, you should set the comparator hysteresis. Setting comparator hysteresis prevents chattering and other undesirable phenomena in the photo MOS relay. You should set the comparator hysteresiss using the following three functions: Function Setting details F-24 Operating condition of comparator hysteresiss F-25 Data width of comparator hysteresis F-26 Duration of comparator hysteresis Described below are comparatorr hysteresis setting procedures. CAUTION The comparator hysteresis setting determines the on/off condition of the open collector output. If the setting is incorrect, the on/off condition of the open collector output will be incorrect, and peripheral devices may malfunction. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. The number showing the currently set comparator hysteresis operating condition, "0" (ON delay) or "1" (OFF delay), appears. Default 0 (ON delay) For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using, set the operating condition. 5. Press. 6. After confirming is displayed, press. The data width of the currently set comparator hysteresis appears. Default Using and, set the data width. You can set any value between 00 and 99, which will be displayed in the load window. 8. Press. 86

89 6 Settings Related to Comparison 9. After confirming is displayed, press. The duration of the currently set comparator hysteresis appears. Default Using and, set the data width. You can set any value between 00 and 99, which will be displayed in the load window. The actual duration is the displayed value multiplied by Press. 12. Press twice to return to Measurement mode. 87

90 7 Settings Related to Operation 7 Settings Related to Operation 7.1 Overview This section describes the following items related to the operation of the CSD-709. Setting display update frequency Lock Setting overload display Confirmation with Check mode Output of CHECK value Setting F key operation Zero set and effective range Setting hold targets Initialization of setting WARNING Before implementing any of above operations, terminate the current measurement operation. When settings are changed incorrectly during measurement with the CSD-709, correct results may not be obtained or the equipment may malfunction and cause damage to peripheral devices. 88

91 7 Settings Related to Operation 7.2 Setting the display update frequency [F-04] To update the frequency of the value displayed in the load window, use function F-04. Described below are the display update frequency setting procedures. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. The number showing the currently set display update frequency appears. Number Update frequency 0 4 times/s 1 20 times/s 2 50 times/s times/s Default 1 (20 times/s) For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using, set the update frequency. Whenever is pressed, the number is cyclically changed from 0 to 3 in ascending order. 5. Press. 6. Press twice to return to Measurement mode. 89

92 7 Settings Related to Operation 7.3 Lock For the CSD-709, the following three locks are available to prevent malfunctions: Simplified calibration lock Individual operation key lock Comprehensive key lock After the simplified calibration lock is enabled, you cannot start the simplified calibration using and. The individual operation key lock can disable operations of and on the front panel on a one-by-one basis. In addition, you can lock all keys on the front panel via an external control input. This section describes how the simplified calibration lock and individual operation key lock are set. For more informationn on external control input, see Section 8.1, "External control input signal" Simplified calibration lock To protect zero point and span from accidental changes in the simplified calibration process, set the simplified calibration lock. Described below are simplified calibration lock setting procedures. CAUTION As a default, the simplified calibration lock is disabled. 1. Hold down. and simultaneously to display a single-digit numerical value starting from 2. Using, select [0] (Disable) or [1] (Enable) to disable or enable the simplified calibration lock, then press. flashes, and Measurement mode is reactivated. For setting procedures with, see Section 1.6, "Basic key operation" Individual operation key locks [F-08], [F-09] To avoid unnecessary key operation during measurement, lock the keys on the front panel individually. To lock the keys, use functions F-08 and F-09. Described below are individual operation key lock setting procedures. POINT After a key is locked, the " LOCK" LED in the statuss window turns on. When has been locked, pressing and simultaneously allows you to transition to other modes apart from Measurement mode. When the comprehensive key lock has been enabled by the external control input, you cannot enable the individual operation key lock. 90

93 7 Settings Related to Operation 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. A four-digit numerical value showing the current lock statuses of through appears. Digit Forth digit Third digit Second digit First digit Key The lock status is indicated by [0] (Disable lock) or [1] (Enable lock). Default 0000 For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set the lock statuses of individual keys. 5. Press. 6. After confirming is displayed, press. Two-digit numerical value showing the current lock statuses of and appears. Digit Second digit First digit Key Default Using and, set the lock statuses of individual keys. 8. Press. 9. Press twice to return to Measurement mode. 91

94 7 Settings Related to Operation 7.4 Setting the overload display [CF-03] If the gross load exceeds the measuring weight or an overflow occurs in the AD conversion, the CSD-709 displays the following message in the load window: Overload status Overload Overflow Display in load window OL or -OL OVF or -OVF For the overload display, you can set any of the following conditions: Overload message is displayed when [measuring weight + 9 scales < gross load] or [gross load < negative measuring weight - 9 scales] is established. Overload message is displayed when [measuring weight 110 % < gross load] or [gross load < negative measuring weight 110 %] is established. Overload message is displayed when [measuring weight + 9 scales < gross load] or [gross load < -20 scales] is established. Set any of these overload display conditions to C function CF-03. Described below are the overload display setting procedures. 1. Press once to display. 2. Press once to change the display to. 3. Press. The C Function mode is activated, and the display changes to a two-digit numerical value prefixed by [CF-]. 4. Using and, change the numerical value to, then press. The number showing the current overload display condition appears. Number Overload display condition OL is displayed when [measuring weight + 9 scales < gross load] is established. -OL is displayed when [gross load < negative measuring weight - 9 scales] is established. OL is displayed when [measuring weight 110 % < gross load] is established. -OL is displayed when [gross load < negative measuring weight 110 %] is established. OL is displayed when [measuring weight + 9 scales < gross load] is established. -OL is displayed when [gross load < -20 scales] is established. Default 1 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the number showing the desired overload display condition. Whenever is pressed, the number is cyclically changed from 0 to 2 in ascending order. 6. Press. 7. Press twice to return to Measurement mode. 92

95 7.5 Confirmation with Check mode (EzCTS, Version, Options, Excitation voltage, External control input/output, Analog output, BCD output, Monitor mode) In the CSD-709, you can conduct the following operations after entering Check mode: Changing the settings in the EzCTS mode Checking the firmware version Checking the option mounted in the CSD-709 Checking the excitation voltage Checking the external control input Checking the external control output Checking the analog output Checking the BCD output Checking the current input in Monitor mode CAUTION 7 Settings Related to Operation You can check the analog output when the current output card or voltage output card is mounted in the CSD-709. You can check the BCD output when the BCD card is mounted in the CSD-709. Described below are setting procedures for each of the above operations in Check mode Changing the settings in EzCTS mode EzCTS is a personal computer application (option). The function data of CSD-709 can be easily read-out and written with EzCTS. This software is useful for maintenance and back-up od data. 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press. flashes, and EzCTS mode is activated. In EzCTS mode, settings can be transferred between the CSD-709 on EzCTS, see the [Ez and PC using the Ez Communication Tool Software (EzCTS). For more information Communication Tool Software Instruction Manual]. POINT In EzCTS mode, the CSD-709 only communications based on EzCTS are executed. For more information on communications with the PC, see Section 10, "Communications via does not send acknowledgments for commands not used in EzCTS. In EzCTS mode, USB". When EzCTS is used in any mode other than EzCTS, error code 01 is sent from the CSD

96 7 Settings Related to Operation 5. To complete EzCTS mode, press. When you press instead of, EzCTS mode is completed, and used to check the firmware version appears. 6. Press twice to return to Measurement mode Checking the firmware version 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press once to change the display to, then press. The current firmware version appears and flashes. 5. Press to complete checking the version. When you press instead of, used to check the option mounted appears. 6. Press twice to return to Measurement mode Checking the option mounted in the CSD Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press twice to change the display to, then press. The character showing the mounted option appears and flashes. Displayed character P07 P15 P16 P29 P73 P74 P76 P77 Option Current output card BCD card (sink type) BCD card (source type) Voltage output card CC-Link card RS-232C card RS-422/485 card Serial interface card When a memory card is mounted, [91] is displayed in addition to any of the above characters. 94

97 7 Settings Related to Operation 5. Press to complete checking the option. When you press instead of, used to check the excitation voltage appears. 6. Press twice to return to Measurement mode Checking the excitation voltage 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press three times to change the display to, then press. The current excitation voltage appears and flashes. 5. Press to complete checking the excitation voltage. When you press instead of, used to check the external control input appears. 6. Press twice to return to Measurement mode Checking the external control input 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press four times to change the display to, then press. flashes, and the LED connected to the external control input turns on. External control input INPUT1 is turned on. INPUT2 is turned on. INPUT3 is turned on. INPUT4 is turned on. INPUT5 is turned on. INPUT6 is turned on. LED SEL1 SEL2 CHECK HOLD PEAK MEAS. 5. Press to complete checking the external control input. When you press instead of, used to check the external control output appears. 6. Press twice to return to Measurement mode. 95

98 7 Settings Related to Operation Checking the external control output 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press five times to change the display to, then press. flashes and the [OUT1] LED in the status window turns on. 5. Using and, turn on any of the following LEDs to turn on the connected external control output. LED OUT1 OUT2 OUT3 OUT4 SEL1 SEL2 External control input OUTPUT1 is turned on. OUTPUT2 is turned on. OUTPUT3 is turned on. OUTPUT4 is turned on. OUTPUT5 is turned on. OUTPUT6 is turned on. 6. Press to complete checking the external control output. When you press instead of, used to check the current input appears. When the current output card or voltage output card is mounted, pressing displays used to check the analog output. If the BCD card is mounted, pressing displays used to check the BCD output. 7. Press twice to return to Measurement mode. 96

99 7 Settings Related to Operation Checking the analog output CAUTION You can check the analog output when the current output card or voltage output card is mounted in the CSD-709. POINT When the current output card is mounted, any of the following currents is output: Output value Indicationn of output 4 ma LOW 12 ma MID 20 ma HI When the voltage output card is mounted, any of the following voltages is output: Output value Indicationn of output 0 V LOW 5 V MID 10 V HI 0 V -LOW -5 V -MID -10 V -HI 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press six times to change the display to, then press. flashes, and the current or voltage is output. 5. Press to change the current or voltage to be output. To invert the potential in the voltage output card, press. 6. Press 7. Press to complete checking the analog output. twice to return to Measurement mode. 97

100 7 Settings Related to Operation Checking the BCD output CAUTION You can check the BCD output when the BCD card is mounted in the CSD Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press six times to change the display to, then press. appears and [0] flashes in the first digit. 5. Using Whenever and, shift flashing to the next digit. is pressed, a flashing "0" times to the next highest digit. Whenever pressed, a flashing [0] times to the next lowest digit. is 6. Press 7. Press to complete checking the BCD output. twice to return to Measurement mode Checking the current input in Monitor mode 1. Press once to display. 2. Press four times to change the display to. 3. Press. Check mode is activated, and the display changes to. 4. Press to change the display to The current input (mv/v) appears and flashes., then press. 5. Press to complete checking the current input. When you presss mode appears. instead of, used to change the setting in the EzCTS 6. Press twice to return to Measurement mode. 98

101 7 Settings Related to Operation 7.6 Output of CHECK value [F-13] By pressing to display the already determined value in the load window, you can check whether the currently selected calibration value is correct. This time, the value determined based on the voltage set to function F-13 as the CHECK value appears in the load window. Described below are the CHECK value setting procedures. CAUTION If you display the CHECK value output by pressing status window goes on. To terminate output of the CHECK value, press, the [CHECK] LED in the once again. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to The currently set CHECK value appears., then press. For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set the CHECK value. You can set any value between 00 and 30. The CHECK value actually output in mv/v is the displayed value multiplied by 0.1. Default 3 (0.3 mv/v) POINT If [00] is set as CHECK value, even pressing will not change the value displayed in the load window, and the [CHECK] LED will go on. This is because the voltage connected to the set CHECK value is 0.0 mv/v. The set CHECK value is applied to all calibration results for calibration numbers 0 to Press. 6. Press twice to return to Measurement mode. 99

102 7 Settings Related to Operation 7.7 Setting the F key operation Any of the following operations can be assigned to number to function F-55. Operation [F-55]. To assign an operation, set the connected Number No assignment 0 Display value hold 1 Zero set 2 Zero set clear 3 Printing with serial interface 4 Transition to COMP setting mode 5 Confirmation of CC-Link station number and RS-422/485 ID 6 Forced termination of multimode 7 Described below are the F key operation setting procedures. 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. The number showing the current F key operation appears. Default 0 (No assignment) For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using, set the F key operation. Whenever is pressed, the number is cyclically changed from 0 to 7 in ascending order. 5. Press. 6. Press twice to return to Measurement mode. 100

103 7 Settings Related to Operation 7.8 Zero set and effective range [CF-11] When zero set has been assigned as F key operation and the track displayed in the load window is within the preset range, you can press to store that track to the CSD-709 as [0]. You should set the effective range of this zero set to C function CF-11. Described below are the procedures used to set the effective range of the zero set. POINT If the track is within the following range when you press, the value is stored as [0]. (Measuring weight Effective range (%)) Track (Measuring weight Effective range (%)) The zero set value stored to the CSD-709 can be erased by pressing clear operation is assigned to the F key after the zero set 1. Press once to display. 2. Press once to change the display to. 3. Press. The C Function mode is activated, and the display changes to a two-digit numerical value prefixed by [CF-]. 4. Using and, change the numerical value to, then press. The currently set effective range appears as a percentage. Default 2 For setting procedures with and, see Section 1.6, "Basic key operation". 5. Using and, set the effective range. You can set any value between 00 and Press. 7. Press twice to return to Measurement mode. 101

104 7 Settings Related to Operation 7.9 Setting hold targets [F-18] If the display value hold has been assigned as an F key operation, use function F-18 to determine whether updating the value displayed in the load windoww will be terminated after is pressed. Additionally, use this function to set whether the comparison results will be held when the open collector is turned on during comparison with the comparator. F-18 can set whether any of the following hold targets should be held: Dataa display in the load window Result of comparison with the comparator Output of load via optional device (voltage output card, current output card, BCD card, serial interface card, or CC-Link card) Described below are the setting procedures for the hold target. CAUTION For the optional voltage output card, you can set the voltage output target via C functions CF-30, CF-33, CF-36, and CF-39. When [0] (Display interlock) is set to any of these C functions, whether the output of load from the voltage output card is held is determined by whether hold is executed for the data display in the load window. For operations in hold mode triggered by the voltage output card, see Section E.5, "Output operations in hold mode" ". For the optional BCD card, you can set the output target of the BCD via function F-31. When [0] (Display interlock) is set to this function, whether outpu of load from the BCD card is held is determined by whether hold is executed for the data display in the load window. For operations in hold mode triggered by the BCD card, see Section G.5, "Output operations in hold mode". 1. Press once to display. 2. Press. The Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. A three-digit numerical value showing the current hold enable/disable status appears. Digit Hold target Third digit Output of load via option Second digit Result of comparison with the comparator First digit Data display in the load window Whether hold is executed is indicated by [0] (Not hold) or [1] (Hold). Default 001 For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set whether hold is executed. 5. Press. 6. Press twice to return to Measurement mode. 102

105 7 Settings Related to Operation 7.10 Initialization of setting [F-95], [F-99], [CF-99], [PF-99] In the CSD-709, you can initialize settings in functions, C functions, and P functions. When a memory card is mounted, you can erase the contents. To initialize the setting, press for the desired function. Function to be initialized Initialization functionn Function F-99 C function CF-99 P function PF-99 Memory card F-95 CAUTION For C functions, the contents of CF-50 to CF-69 to which the calibration data is stored will be retained. POINT To terminate the initialization, press. 103

106 8 Settings Related to Output Control 8 Settings Related to Output Control 8.1 External control input signal [F-60] to [F-65], [F-69] The CSD-709 can respond to up to six external contact inputs. You can assign the following functions to these contact inputs. Function Designate zero set. Designate zero set clear. Designate hold (level). Designate hold (pulse). Designate PEAK/TRACK. Designate RESET. Designate SEL1. Designate SEL2. Designate CHECK value. Designate comprehensive key lock. Designate print command (only when the serial interface card is mounted). Designate forced termination of multimode. Input type Pulse Pulse Level Pulse Level Level Level Level Level Level Pulse Pulse These functions can be set to functions F-60 to F-65 as operations of INPUT1 to INPUT6, respectively. You can also set response times against these operating signals to function F-69. Described below are the external control input setting procedures. POINT To select designation of SEL1 or SEL2, combine the two statuses as shown below and specify the change in calibration number. Status of SEL1 Status of SEL2 Calibration number to be changed Open Open 0 Short Open 1 Open Short 2 Short Short 3 For the wiring for the external control input, see Section 2.2.3, "Plugging in the external control input/output connector". For more information on functions available in the external control input, see the descriptions in the sections listed below. Function Referent Designate zero set. 7.8 Zero set and effective range Designate zero set clear. 7.8 Zero set and effective range Designate hold (level). D.1 Detection operation Designate hold (pulse). D.1 Detection operation Designate PEAK/TRACK. D.1 Detection operation Designate RESET. D.1 Detection operation Designate SEL Changing the calibration number Designate SEL Changing the calibration number Designate CHECK value. 7.6 Output of CHECK value Designate comprehensive key lock. 7.3 Lock Designate print command (only when serial interface card is mounted) K.3 Setting Designate forced termination of multimode. 5.2 Detection with multimode 104

107 8 Settings Related to Output Control 1. Press once to display. 2. Press. Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to any of,,,,, and, then press. A two-digit number showing the currently set external control input function appears. Number Function 00 Does not use the external control input function. 01 Designate zero set. 02 Designate zero set clear. 03 Designate hold (level). 04 Designate hold (pulse). 05 Designate PEAK/TRACK. 06 Designate RESET. 07 Designate SEL1. 08 Designate SEL2. 09 Designate CHECK value. 10 Designate comprehensive key lock. 11 Designate print command (only when serial interface card is mounted). 12 Designate forced termination of multimode. Default 00 For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set the external control input function, then press. 5. To set the external control input function to other functions, repeat Steps 3 and To set the response time against the external control input, display, then press. The number showing the currently set response time appears. Number Response time 0 50 ms 1 20 ms 2 10 ms 3 5 ms 4 2 ms 5 1 ms Default 0 (50 ms) The external control input function will be enabled after a signal longer than the set response time is input. 105

108 8 Settings Related to Output Control 7. Using, set the response time. Whenever is pressed, the number is cyclically changed from 0 to 5 in ascending order. 8. Press. Press twice to return to Measurement mode. 106

109 8 Settings Related to Output Control 8.2 External control output signal [F-70] to [F-75], [F-82], [F-83] The CSD-709 can provide up to six contact outputs to external devices. You can assign the following functions to these contact outputs: Show that the hold mode is active. Show that the [PEAK] LED is turned on. Show that the [MEAS.] LED is turned on. Show that the [END] LED is turned on. Show that multimode is active. Show that measurement is underway in multimode. Show that the results of waveform comparison and detection result judgment are acceptable. Show that the results of waveform comparison and detection result judgment are unacceptable. Show that the result of the detection result judgment exceeds the upper limit. Show that the open collector of S0 is turned on in the comparator. Show that the open collector of S1 is turned on in the comparator. Show that the open collector of S2 is turned on in the comparator. Show that an error occurred. Show that an overload occurred. Show that the zero set was executed. Use the level output to show that measurement is underway. Use the toggle output having a period of 500 ms to show that measurement is underway. These functions can be set to functions F-70 to F-75 as operations of OUTPUT1 to OUTPUT6, respectively. You can also set the output logic of these operations to functions F-82 and F-83. Described below are the external control output setting procedures. For the wiring for the external control output, see Section 2.2.3, "Plugging in the external control input/output connector". For more information on functions available in external control output, see the descriptions in the sections listed below. Function Referent Show that the hold mode is active. 7.9 Setting hold targets Show that the [PEAK] LED is turned on. D.1 Detection operation Show that the [MEAS.] LED is turned on. D.1 Detection operation Show that the [END] LED is turned on. D.1 Detection operation Show that multimode is active. 5.2 Detection with multimode Show that measurement is underway in multimode. 5.2 Detection with multimode Show that the results of waveform comparison and detection result judgment are acceptable. Show that the results of waveform comparison and detection result judgment are unacceptable. Show that the result of the detection result judgment exceeds the upper limit. Show that the open collector of S0 is turned on in the comparator. Show that the open collector of S1 is turned on in the comparator. Show that the open collector of S2 is turned on in the comparator. Show that an error occurred. Show that an overload occurred. Show that the zero set was executed. Use the level output to show that measurement is underway. Use the toggle output having a period of 500 ms to show that measurement is underway. 5.3 Detection result judgment and waveform comparison 5.3 Detection result judgment and waveform comparison 5.3 Detection result judgment and waveform comparison 6.1 Overview 6.1 Overview 6.1 Overview 9.1 Error code list 7.4 Setting the overload display 7.8 Zero set and effective range

110 8 Settings Related to Output Control 1. Press once to display. 2. Press. Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to any of,,,,, and, then press. A two-digit number showing the currently set external control output function appears. Number Function 00 Does not use the external control output function. 01 Show that the hold mode is active. 02 Show that the [PEAK] LED is turned on. 03 Show that the [MEAS.] LED is turned on. 04 Show that the [END] LED is turned on. 05 Show that multimode is active. 06 Show that measurement is underway in multimode Show that the results of waveform comparison and detection result judgment are acceptable. Show that the results of waveform comparison and detection result judgment are unacceptable. 09 Show that the result of the detection result judgment exceeds the upper limit. 10 Show that the open collector of S0 is turned on in the comparator. 11 Show that the open collector of S1 is turned on in the comparator. 12 Show that the open collector of S2 is turned on in the comparator. 13 Show that an error occurred. 14 Show that an overload occurred. 15 Show that the zero set was executed. 16 Use the level output to show that measurement is underway. 17 Use the toggle output having a period of 500 ms to show that measurement is underway. Default 00 For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using and, set the external control output function, then press. 5. To set the external control output function to other functions, repeat Steps 3 and

111 8 Settings Related to Output Control 6. To set the output logic for the external control output, display, then press. Five-digit numerical value showing the currently set output logic appears. Digit Operation Fifth digit OUTPUT5 set to F-74 Fourth digit OUTPUT4 set to F-73 Third digit OUTPUT3 set to F-72 Second digit OUTPUT2 set to F-71 First digit OUTPUT1 set to F-70 The output logic is indicated by [0] (Negative logic) or [1] (Positive logic). Default Using and, set the output logic for the external control output. 8. Press. 9. After confirming is displayed, press. The number showing the output logic of OUTPUT6 set to F-75 appears. Default Using, set the output logic. 11. Press. 12. Press twice to return to Measurement mode. 109

112 9 Troubleshooting 9 Troubleshooting 9.1 Error code list When an error occurs in the CSD-709, the code showing the type of error appears in the load window. The table below shows error codes, meanings, the behavior of the CSD-709 in response to errors, and remedial actions. Error code Meaning Behavior of CSD-709 Remedial action Zero set is executed outside the effective The error code flashes. After the error code flashes for two seconds, it is reset automatically. range. Invalid function is selected. The error code flashes. After the error code flashes for two seconds, it is reset automatically. Incorrect setting is made in calibration. AD conversion is terminated. Incorrect operation is executed during confirmation with CHECK value. Zero fine adjustment or span fine adjustment is executed although zero set, or peak is turned on. Value set to function is out of range. An attempt is made to change the settings of CF-30 to CF-41 although the model is not identified by [-07] or [-29]. Write error occurs on EEPROM. Option recognition error occurs. Read error occurs on EEPROM. Power-on sequence is executed with the hold signal turned on. Displayed value exceeds the upper limit of the allowable range. The error code flashes. The error code appears, and the measurement is terminated. The error code flashes. The error code flashes. The error code flashes. The error code flashes. The error code appears, and the measurement is terminated. The error code appears, and the measurement is terminated. The error code appears, and the measurement is terminated. The error code flashes. The error code is displayed. After the error code flashes for two seconds, it is reset automatically. Turn off power once and restart the equipment. After the error code flashes for two seconds, it is reset automatically. After the error code flashes for two seconds, it is reset automatically. After the error code flashes for two seconds, it is reset automatically. After the error code flashes for two seconds, it is reset automatically. Turn off power once and restart the equipment. Turn off power once and restart the equipment. Turn off power once and restart the equipment. Turn off the hold signal. Enter a value that stays within the allowable range. 110

113 9 Troubleshooting Error code Meaning Behavior of CSD-709 Remedial action Displayed value is smaller than the lower limit of the allowable range. The error code is displayed. Enter a value that stays within the allowable range. Result of AD conversion is overflown. Result of AD conversion is overflown. Actual load used in the calibration exceeds the specified value. Actual load used in calibration is smaller than the specified value. Zero point used in calibration exceeds the specified value. Zero point used in calibration is smaller than the specified value. The error code is displayed. The error code is displayed. The error code flashes for two seconds. The error code flashes. The error code flashes. The error code flashes. Enter a value that stays within the allowable range. Enter a value that stays within the allowable range. After the error code flashes for two seconds, it is reset automatically. However, when the simplified zero calibration is executed and the span therefore deviates from the specified range, even returning to Measurement mode will not resume measurement, and the error code will continue to flash. In this case, execute span calibration to make the span fall within the specified range. After the error code flashes for two seconds, it is reset automatically. After the error code flashes for two seconds, it is reset automatically. After the error code flashes for two seconds, it is reset automatically. 111

114 9 Troubleshooting 9.2 Troubleshooting When unintended operations are observed on the CSD-709, verify the phenomena and check-outs listed in the table below to take proper action. Phenomenon No value is displayed. Check-out Check whether the power cable is plugged in correctly. Check whether the power supply voltage is within the allowable range. Check whether the fuse is blown. The displayed load is faulty, and OL or -OL is observed. Check whether the cable coming from the strain-gage-based transducer is correctly plugged into the equipment. Check whether the calibration is correctly executed. Check whether the excitation voltage is at normal levels. The displayed value is fluctuating. Check whether the input currently displayed in Check mode agrees with the actual load. Check whether the measuring unit is stable. Check whether the excitation voltage is at normal levels. Check whether the input currently displayed in Check mode agrees with the actual load. Check whether the display becomes stable when strain-gage-based transducer connectors B (-SIG) and D (+SIG) are short-circuited. Check whether analog and digital filters are properly set. The displayed value fluctuates temporarily. The key operation is rejected. Values have been set to functions, or the data in the memory has been erased. Individual operation key lock or comprehensive key lock has been set. Comprehensive key lock has been set. The simplified calibration is rejected. The external control input does not operate. Simplified calibration lock has been enabled. Check whether signal cables are plugged in correctly. Check whether the operation is correctly set for the input signal. Check whether the pulse input is a 50-ms or longer signal. Check whether the level input is a 50-ms or longer continuous signal. The external control output does not operate. Check whether signal cables are plugged in correctly. Check whether the output is correctly set for the operation. Check whether the rated voltage or current is within the allowable range. The output logic in the external control output is different from the intended one. Communications based on the USB interface cannot be established. Change the output logic in the external control output. Check whether the driver has been installed on the host PC. Check whether the USB port is correctly recognized by the host PC. Check whether the USB port is correctly set on the host PC. The intended value cannot be output from the USB interface. Check whether the length of the USB cable is not more than 5 m. Check settings relative to the communication interface. Check settings for the output target. 112

115 9 Troubleshooting Phenomenon Communications based on the EzCTS cannot be established. Check-out Check whether the equipment is correctly connected to the host PC. Check whether the equipment is in EzCTS mode. 113

116 9 Troubleshooting 9.3 Replacing the fuse The fuse is mounted on the printed board. Described below are replacement procedures for the fuse. 1. Turn off the CSD Remove the two screws (A) from the rear panel and detach the top plate (B). 3. Replace the fuse (A) mounted on the printed board with a new one. CAUTION Problems will occur if the fuse is incorrectly mounted or its capacity is insufficient. 4. Reinstall the top plate and fix it with the two screws. 114

117 10 Communications 10 Communications via USB via USB 10.1 Communication specifications The CSD-709 is equipped with a USB interface conforming to the USB2.0 standard. The table below shows the communication specifications of the USB interface. Item Details Output connector B type mini-usb connector Transmission Half-duplex Sync mode Asynchronous Baud rate Select bps, bps, bps, bps, bps, bps, bps, or bps. Dataa bit length Parity bit Stop bit Terminator Communication Address Communication Select 7 bit or 8 bit. Select No parity, Even parity, or Odd parity. Select 1 bit or 2 bit. Select CR or CR+LF. ASCII code 00 (Fixed) Command mode or Stream mode CAUTION To use the USB interface, you must install the specialized driver to the PC used as host. POINT The USB interface of the CSD-709 accepts the device designed under the USB1.1 standard as well. 115

118 10 Communications via USB 10.2 Setting the USB interface [F-40] to [F-44], [CF-05] To use the USB interface, you should select any of three operating modes described below. Mode Command mode Stream mode Print command synchronization mode Operation Returns the data in response to the command and data received from the host PC. Continues outputting the target output data already determined. Outputs the target output data, when is pressed, if the print command communicated through the serial interface has been assigned as an F key operation. For the data output in any of these modes, the decimal point can be added based on the decimal point display position settings stored to C functions CF-01, CF-42, CF-43, and CF-44. For setting the F key operation, see Section 7.7, "Setting the F key operation". For setting the decimal point display position, see Section 3.3, "Setting the decimal point display position". When the Stream mode or Print command synchronization mode is selected as an operating mode, you should set any of the following items as the output target. Interlock with display in the load window Track Peak Combination of load display and status When [combination of load display and status] is selected, any of the following characters showing the unit set to C function CF-05 is output together with the data. The character showing the unit is determined by such a setting that the number of digits set to function F-44 is two or three. Unit Gram Kilogram Pound Ton Newton Kilonewton Pascal Kilopascal Megapascal Output character G Kg Lb T N kn P or Pa kp or kpa MP or MPa 116

119 10 Communications via USB The contents relative to the USB interface should be set to the following functions, respectively. Function Setting details F-40 Operating modes F-41 Output target in the Stream mode or Print command synchronization mode F-42 Communication method based on the USB interface F-43 Addition/Removal of decimal point to/from output data F-44 Number of digits of the unit to be output under combination of load display and status CF-05 Unit character to be output Described below are the procedures for setting the contents related to the USB interface. 1. Press once to display. 2. Press. Function mode is activated, and the display changes to a two-digit numerical value prefixed by [F-]. 3. Using and, change the numerical value to, then press. The number showing the current operating mode appears. Number Operating modes 0 Command mode 1 Stream mode 2 Print command synchronization mode Default 0 (Command mode) For setting procedures with and, see Section 1.6, "Basic key operation". 4. Using, set the operating mode, then press. Whenever is pressed, the number times through the values from 0 to 2 in ascending order. 5. After confirming is displayed, press. The number showing the currently set output target appears. Number Output target 0 Interlock with display in the load window 1 Track 2 Peak 3 Combination of load display and status Default 0 (Interlock with display in the load window) 6. Using, set the output target, then press. Whenever is pressed, the number times through the values from 0 to 3 in ascending order. The output target will be enabled after Stream mode or Print command synchronization mode is set as the operating mode. 117

120 10 Communications via USB 7. After confirming is displayed, press. A five-digit numerical value showing the current USB interface based communication method appears. Digit Details Fifth digit Terminator: [0] (CR) or [1] (CF+LF) Fourth digit Baud rate: [0] (1 200 bps), [1] (2 400 bps), [2] (4 800 bps), [3] (9 600 bps), [4] ( bps), [5] ( bps), [6] ( bps), or [7] ( bps) Third digit Stop bit: [0] (1 bit) or [1] (2 bit) Second digit Parity bit: [0] (No parity), [1] (Even parity), or [2] (Odd parity) First digit Data bit length: [0] (7 bit) or [1] (8 bit) Default (Terminator: CR+LF, Baud rate: bps, Stop bit: 1 bit, Parity bit: Odd parity, Data bit: 7 bit) 8. Using and, set the contents of the USB interface based communications, then press. 9. After confirming is displayed, press. 0 (Hide) or 1 (Add) showing the currently set decimal point on/off status appears. Default 0 (Hide) 10. Using, set the addition or removal of a decimal point, then press. Whenever is pressed, the number times between 0 and After confirming is displayed, press. 0 (two digits) or 1 (three digits) showing the currently set number of digits of the unit used in the combination of load display and status appears. Default 0 (Two digits) 12. Using, set the number of digits of the unit, then press. Whenever is pressed, the number times between 0 and Press once to display. 14. Press once to change the display to. 15. Press. The C function mode is activated, and the display changes to a two-digit numerical value prefixed by [CF-]. 118

121 10 Communications via USB 16. Using and, change the numerical value to, then press. The number showing the currently set output unit character appears. Number Unit Output character 0 - (None) - (None) 1 Gram G 2 Kilogram Kg 3 Pound Lb 4 Ton T 5 Newton N 6 Kilonewton kn 7 Pascal P or Pa 8 Kilopasca kp or kpa 9 Megapascal MP or MPa Default 0 (None) 17. Press to set the number showing the output unit character, then press. Whenever is pressed, the number times through the values from 0 to 9 in ascending order. 18. Press twice to return to Measurement mode. 119

122 10 Communications via USB 10.3 Data format This section describes the format of the data transferred through the USB interface per the operating mode Command mode When at sign address, command number, and terminator, all of which are described according to the following format, are sent from the host PC with the Command mode defined as current operating mode, the CSD-709 returns the response data. The address is fixed to [00]. Any of the following command numbers should be entered according to the data received. Command Command number Reference the displayed load 20 Reference the track 21 Reference the peak 22 Reference the displayed load with status 26 Reference the comparison value (S0) 30 Reference the comparison value (S1) 31 Reference the comparison value (S2) 32 Reference the current calibration number 39 Reference the status of the status window (LED) 40 Reference the status of the external control output 41 Reference the version 45 Set PEAK/TRACK 50 Set RESET 51 Execute the zero set 52 Execute the zero set clear 53 Execute the forced termination of multimode 54 Set the comparison value (S0) 60 Set the comparison value (S1) 61 Set the comparison value (S2) 62 Set the calibration number 69 Reference the setting (function) 70 Reference the setting (C function) 71 Reference the setting (P function) 73 Establish the setting (function) 80 Establish the setting (C function) 81 Establish the setting (P function)

123 10 Communications via USB Command Execute the calibration Execute the stream transferr Command number CAUTION If an error occurs during the data transfer in the Command mode, the data will be returned in the following format. This time, the Error code field has the numerical value describing the error generated, among those listed below. Numerical value Error description 01 The operation cannot be executed because the command is received in a mode other than Measurement mode. 02 Zero set error 10 Parity error 11 Stop bit detection error 12 Read error 13 An error occurs in the data after the command number. 14 The specified command number does not exist. Referencing the load data via command numbers 20 to 26 When you specify command number 20 (Reference the displayed load), 21 (Reference the track), or 22 (Reference the peak), the following data format is returned: 121

124 10 Communications via USB When you specify command number 26 (Reference the displayed load with status), the following data format is returned: This time, the Status 1 field has a two-letter alphabetic character describing the condition of the CSD-709, among those listed below. Alphabetic character Meaning OL OL TR TRACK PE PEAK ME MEAS ED END Also, Status 2 field has a two-letter alphabetic character showing the meaning of the numerical value entered into the Load field (eight-digit numerical value), among those listed below. Alphabetic character PE TR Meaning Peak Track POINT Whether the number of digits of unit should be two or three is determined by the setting of function F-44. Referencing the comparison value via command numbers 30 to 39 When you specify command number 30 (Reference the comparison value (S0)), 31 (Reference the comparison value (S1)), 32 (Reference the comparison value (S2)), or 39 (Reference the current calibration number), the following data format is returned: 122

125 10 Communications via USB Referencing via command numbers 40 to 45 When you specify command number 40 (Reference the status of the status window (LED)) or 41 (Reference the status of the external control output), the following data format is returned: This time, the nine-digit numerical value showing the status takes [0] (Off) or [1] (On). Individual digits have the following meanings. Digit Number in illustration 40 (Reference the status of the status window (LED)) 41 (Reference the status of the external control output) Ninth digit 6 SEL1 OUTPUT1 Eighth digit 7 SEL2 OUTPUT2 Seventh digit 8 CHECK OUTOUT3 Sixth digit 9 HOLD OUTOUT4 Fifth digit 10 PEAK OUTPUT5 Fourth digit 11 MEAS OUTPUT6 Third digit 12 END -(Not used) Second digit 13 LOCK -(Not used) First digit 14 - (Not used) -(Not used) When you specify command number 45 (Reference the version), the following data format is returned: Setting via command numbers 50 to 69 When you specify command number 50 (Set PEAK/TRACK), 51 (Set RESET), 52 (Execute the zero set), 53 (Execute the zero set clear), or 54 (Forcibly terminate multimode), the following command format is sent: 123

126 10 Communications via USB When you specify command number 60 (Set the comparison value (S0)), 61 (Set the comparison value (S1)), 62 (Set the comparison value (S2)), or 69 (Set the calibration number), the following command format is sent: Setting via command numbers 70 to 83 When you specify command number 70 (Reference the setting (function)), 71 ( Reference the setting (C function)), or 73 (Reference the setting (P function)), the following data format is returned: When you specify command number 80 (Establish the setting (function)), 81 (Establish the setting (C function)) ), or 83 (Establish the setting (P function)), the command number, function number, and setting are sent in the following data format: CAUTION For functions F-40 to F-52, you can check current settingss but cannot enter new settings. Calibration via command number 90 When you specify command number 90 (Execute the calibration), the command number and 01 (Transition to the Calibration mode) are sent in the following format: For more information on the Calibration mode, see Section 10.4, "Remote calibration via USB". 124

127 10 Communications via USB Referencing via command number 99 When you specify command number 99 (Execute the stream transfer), the data of the output target set with function F-41 are returned in the following format, just as in Stream mode: When you specify [3] (Combination of load display and status) with F-41, the Status 1 field has the two-letter alphabetic character showing the status of the CSD-709, among those listed below. Alphabetic character OL TR PE ME ED Meaning OL TRACK PEAK MEAS END Also, the Status 2 field has the two-letter alphabetic character showing the meaning of the numerical value entered into the Load field (eight-digit numerical value), among those listed below. Alphabetic character PE TR Meaning Peak Track POINT Whether the number of digits of unit should be two or three is determined by the setting of function F

128 10 Communications via USB Stream mode When Stream mode is defined as the current operating mode, the data of the output target set with function F-41 is output in the following format: When you specify [3] (Combination of load display and status) with F-41, the Status 1 field has the two-letter alphabetic character showing the status of the CSD-709, among those listed below. Alphabetic character OL TR PE ME ED Meaning OL TRACK PEAK MEAS END Also, the Status 2 field has the two-letter alphabetic character showing the meaning of the numerical value entered into the Load field (eight-digit numerical value), among those listed below. Alphabetic character PE TR Meaning Peak Track POINT Whether the number of digits of unit should be two or three is determined by the setting of function F

129 10 Communications via USB Print command synchronization mode When the Print command synchronization mode is defined as the current operating mode, the CSD-709 synchronizes with the print command, which has been assigned to the F key through the serial interface, and outputs the data of the output target set with function F-41 in the following format: When you specify [3] (Combination of load display and status) with F-41, the Status 1 field has the two-letter alphabetic character showing the status of the CSD-709, among those listed below. Alphabetic character OL TR PE ME ED Meaning OL TRACK PEAK MEAS END Also, the Status 2 field has the two-letter alphabetic character showing the meaning of the numerical value entered into the Load field (eight-digit numerical value), among those listed below. Alphabetic character PE TR Meaning Peak Track POINT Whether the number of digits of unit should be two or three is determined by the setting of function F-44. For setting the F key operation, see Section 7.7, "Setting the F key operation". 127

130 10 Communications via USB 10.4 Remote calibration via USB In Command mode, you can remotely calibrate the CSD-709 by specifying any of the following numbers in addition to command number 90 (Execute the calibration). No. Operation relative to calibration 01 Transition to calibration mode 02 Setting unit weight 03 Setting measuring weight 04 Setting weight mass 07 Zero calibration (actual load based) 08 Zero calibration (numerical value input based) 10 Span calibration (actual load based) 11 Span calibration (numerical value input based) 21 Setting decimal point display position 40 Confirmation of calibration status 99 Registration 00 Forced termination 128

131 10 Communications via USB Execute the remote calibration using the following steps. During remote calibration, observed in the load window. Common steps between actual load based and numerical value input based calibrations Transition to calibration mode using calibration number 01 is Setting decimal point display position using calibration number 21 Setting unit weight using calibration number 02 Setting measuring weight using calibration number 03 Setting weight mass using calibration number 04 Actual load based calibration Zero calibration (actual load based) using calibration number 07 Numerical value input based calibration Zero calibration (numerical value input based) using calibration number 08 Confirmation of calibration status using calibration number 40 Span calibration (actual load based) using calibration number 10 Span calibration (numerical value input based) using calibration number 11 Confirmation of calibration status using calibration number 40 Registration using calibration number 99 CAUTION Usually, you should execute the zero calibration (actual load based) using calibration number 07. To execute span calibration (actual load based) using calibration number 08, you should prepare a weight of at least two-thirds of the measuring weight. For more information on actual load based and numerical value input based calibrations, see Section 3.1, "Overview". 129

132 10 Communications via USB Setting with calibration numbers 01, 07, 10, 99, and 00 When you use calibration number 01 (Transition to calibration mode), 07 (Zero calibration (actual load based)) ), 10 (Span calibration (actual load based)), 99 (Registration), or 00 (Forced termination), the command number and calibration number are sent in the following format: CAUTION When the calibration number 00 (Forced termination) is sent during the calibration, the calibration data established up to that timing will not be reflected. Setting with calibration number 02 When you use calibration number 02 (Setting unit weight), the command number, calibration number, and any of numerical values (1, 2, 5, or 10) showing the unit weight, are sent in the following format: Setting with calibration numbers 03 and 04 When you use calibration number 03 (Setting measuring weight) or 04 (Setting weight mass) ), the command number, calibration number, and any of numerical values (ranging from 1 to ) showing the measuring weight or weight mass are sent in the following format: Setting with calibration numbers 08 and 11 When you use calibration number 08 (Zero calibration (numerical value input based)) or 11 (Span calibration (numerical value input based)), the command number, calibration number, and any of numerical values (ranging from to 3100) showing the zero calibration value or span calibration value are sent in the following format: 130

133 10 Communications via USB Setting with calibration number 21 When you use calibration number 21 (Setting decimal point display position), the command number, calibration number, and any of numerical values (ranging from 0 to 4) showing the decimal point display position are sent in the following format: Confirmation with calibration number 40 When you use calibration number 40 (Confirmation of calibration status), specifying both command number and calibration number returns the data in the following format: This time, the Status field has the numerical value describing the calibration status, among those listed below. Numerical value Calibration status 00 Complete 01 Under execution 20 Error code TE-L 21 Error code TE-H 22 Error code SP-L 23 Error code SP-I For error codes, see Section 9.1, "Error code list". 131

134 A Specifications of CSD-709 A Specifications of CSD-709 A.1 Analog Specifications Bridge power supply Applicable transducers 10 ± 0.5 VDC, 60 ma, max (can be changed to 5 VDC or 2.5 VDC) Up to two strain-gage-based transducers (350 ) can be connected to the equipment in parallel at 10 VDC. Up to four strain-gage-based transducers (350 ) can be connected to the equipment in parallel at 5 VDC. Up to eight strain-gage-based transducers (350 ) can be connected to the equipment in parallel at 2.5 VDC. Input sensitivity 0.4 V/d (d = minimum scale) Input range -3.1 mv/v ~ 3.1 mv/v Monitor output Approx. 2 V per 1 mv/v of sensor input (@ 10 VDC of bridge power supply voltage) Load resistance: 2 k, min Non-linearity 0.02 %F.S.±1 digit Influence of temperature Zero point ±0.5 V/ C (for the calibration at input sensitivity of 0.4 V/d or higher) Sensitivity ± %F.S./ C (for the calibration at input sensitivity of 0.4 V/d or higher) Analog filter 100 Hz (can be changed to 10 Hz, 40 Hz, 400 Hz, 1 khz, 4 khz, 10 khz, or 30 khz) Digital low-pass filter 40 times/s: 0 (Off) ~ 4 Hz 400 times/s: 0 (Off) ~ 40 Hz times/s: 0 (Off) ~ 400Hz AD sampling times/s (can be changed to 40 or 400 times/s) AD internal resolution 24-bit CHECK value Approx. 0.3 mv/v (The value can be set at an increment of 0.1 mv/v within a range of about 0.1 mv/v ~ 2.0 mv/v.) * Minebea s standard extension cable CAB-502 (4-core cable) having a length of not more than 30 m shall be used. When a Zener barrier is used, this requirement is not applied. Analog peak hold Response speed According to the characteristics of the analog filter Accuracy 0.1 %F.S., max 132

135 A Specifications of CSD-709 A.2 Digital Specifications Main display (load window) Display range ~ Display increment 1 (can be changed to 2, 5, or 10) Display unit Green 7-segment LED, Character height: 17 mm Out-of-range display [-OL] appears in case of minus overflow. [OL] appears in case of plus overflow. AD overflow display [-OVF] appears in case of minus overflow. [OVF] appears in case of plus overflow. Status window SEL.1, SEL.2, CHECK, HOLD, PEAK, MEAS., END, LOCK Judgment window OUT1, OUT2, OUT3, OUT4 Display times 20 times/s (can be changed to 4, 50, or 100 times/s) Decimal point display Can be changed to no display, 101, 102, 103, and 104 A.3 Functions of Front Panel Sheet Key Switches Selects any mode other than Measurement mode, such as Function mode. Transitions to the simplified numerical value input calibration (by holding down this key for at least two seconds). Allows you to move to the next upper digit in the setting target. Transitions to the simplified zero calibration (by holding down this key for at least two seconds). Transitions to the simplified calibration lock mode (by simultaneously holding down and for at least two seconds). Increases the numerical value of the setting target by one. Transitions to the simplified span calibration (by holding down this key for at least two seconds). Selects the track value and various peak detection modes (detection of peak, bottom, peak-bottom, peak-to-peak, maximum, minimum, maximum-minimum difference, average, or inflection point) (entire section designation mode). Otherwise, controls the hold section in the section designation mode or time and section designation mode. Resets the peak. The display is fixed to [0] while this key is being turned on. Turns on/off the CHECK value. Selects [Off], [Hold], [Zero set], [Zero set clear], [Print command], [COMP setting mode], [CC-Link number], [Confirmation of RS-422/485 ID], or [Forced termination of multimode]. Registers the setting. 133

136 A Specifications of CSD-709 A.4 External Control Function A.4.1 External control input signal You can select any of six signals among those listed below. Off Zero set Zero set clear Hold (pulse) Print command (serial interface) Forced termination of multimode * These functions are enabled once when the pulse input width is 50 ms or longer. (The pulse width can be changed to 1, 2, 5, 10, or 20 ms.) Hold (level) PEAK/TRACK RESET SEL.1 SEL.2 CHECK Comprehensive key lock * These functions are enabled while the level that is 50 ms or longer is being input (at the time of the short-circuit.) (The level width can be changed to 1, 2, 5, 10, or 20 ms.) A.4.2 External control output signal You can select any of six signals among those listed below. Off Hold PEAK (turned on interlocking with the status display) MEAS. (turned on interlocking with the status display) END (turned on interlocking with the status display) MULTI (turned on in multimode) EXCUTE (turned on during measurement in multimode) OK (turned on when [lower limit < current load < upper limit] is met) NG (turned on under any condition, except for OK condition) Upper/lower limit (turned on under [upper limit current load] and [current load lower limit]) S0 S1 S2 ERROR Over error (±OL, ±OVF) Zero set response (turned off after elapse of 500 ms) RUN (level) RUN (toggle) 134

137 A Specifications of CSD-709 A.5 Comparator Function Setting ~ Number of settings 3 (S0, S1, and S2) Hysteresis data setting 0 ~ 99 digits Hysteresis time width setting 0 ~ 9.9 s Direction of hysteresis You can select [ON delay] or [OFF delay]. Comparator conversion frequency Synchronized with AD sampling Comparator operation target You can set peak and track independently. 135

138 A Specifications of CSD-709 A.6 Function Analog filter A low-pass filter that utilizes resistors and capacitors to attenuate those frequency components higher than the set frequency. This filter is enabled only when the analog peak hold is used. Digital Filter A filter that stabilizes the fluctuation during the moving-average process with the CPU. Digital low-pass filter A low-pass filter used in the operation by the CPU. This filter attenuates those frequency components higher than the set frequency. (Bessel characteristic secondary low-pass filter) Simplified calibration Holding down,, and respectively for at least two seconds allows you to enter the desired parameter setting screen or calibration screen. Communication calibration You can execute the calibration based on communications through the interface (optional RS-232C, RS-422/485, or CC-Link) without key operation. CHECK value An additional load equivalent to the set value. Using this value, for example, in the pre-work check, allows you to confirm the original calibration status is maintained. Hold target change You can combine [Display], [Comparison output], [Current output (option)], [Voltage output (option)], [BCD output (option)], [CC-Link (option)], and [Serial interface (option)] to define the resulting combination as the hold target. Analog output target change You can select [Track] or [Peak] as the analog output target. Brand code change You can make the CSD-709 memorize up to four different types of calibration data or comparator settings, then select the desired item using functions or external control inputs (SEL.1 and SEL.2). Peak mode selection You can select the peak detection mode from 36 modes configured through a combination of nine hold modes (peak hold, bottom hold, peak-bottom hold, peak-to-peak hold, maximum hold, minimum hold, maximum-minimum difference hold, average hold, and four inflection point holds) and four section modes (entire section, specified section, time-specified section, and automatic start time-specified section). Digital peak hold You can acquire the peak load through high-speed sampling of up to times/s. Analog peak hold You can acquire the peak load more speedily independent of the sampling rate. However, the analog circuit equipped with capacitors is used to execute the peak hold, and thus there are some restrictions (e.g., the current load cannot be determined. Monitor output This function outputs approximately 2 V per 1 mv/v (when bridge power supply is 10 VDC.) before A/D conversion of the input signal from strain gage type transducer. Monitor mode This mode converts the load currently applied to the sensor into a value expressed in mv/v and displays the result. It is effective in finding the cause when the displayed load is abnormal or unstable. Waveform comparison After an additional memory (option) is mounted, you can judge the integrity of the measurements using upper and lower limits determined with the specialized application and comparison waveforms. Measurement result log After an additional memory (option) is mounted, you can record comparison judgment results as well as time stamps. 136

139 A Specifications of CSD-709 A.7 USB Interface (for connection with EzCTS and specialized application) Specifications Output connector Compatible with USB2.0 (can also be used with USB1.1 compatible equipment) B type mini-usb connector CAUTION To use the USB interface, you must install the specialized driver to the PC used as host. A.8 General Specifications Operating Temperature -10 C ~ 50 C temperature/humidity Humidity 85 %RH, max (No condensation is permitted.) range Stored -20 C ~ 60 C temperature Power supply Power supply voltage 100 VAC ~ 240 VAC (Allowable fluctuation range: 85 VAC ~ 264 VAC) Power supply 50/60 Hz frequency Power Approx. 10 VA (@ 100 VAC; no option) consumption Approx. 18 VA, max (@ 100 VAC; equipped with optional CC-Link) Inrush current AC 5 A (1 ms) in 100 VAC average load condition (ordinary configuration temperaturee and cold start) 10 A (1 ms) in 200 VAC average load condition (ordinary temperaturee and cold start) External dimensions 96 mm 48 mmm 110 mm (excluding projections) (W x H x D) Weight Approx. 630 g (excluding options) A.9 Standard Shipment Specifications Bridge power supply 10 VDC Span adjustment The value [30 000] is displayed at 3.0 mv/v input Minimum scale 1 137

140 A Specifications of CSD-709 A.10 Accessories Startup Guide (in Japanese and English) 1 set each Midget fuse 1 (2.5 A) Unit seal 1 External control input/output Connector 1 Strain gage-based-transducer connector 1 BCD output plug 1 (available only when the optional BCD card is mounted in the CSD-709) Voltage output plug 1 (available only when the optional voltage output card is mounted in the CSD-709) Current output plug 1 (available only when the optional current output card is mounted in the CSD-709) RS-232C plug 1 (available only when the optional RS-232C card is mounted in the CSD-709) RS-422/485 plug 1 (available only when the optional RS-422/485 card is mounted in the CSD-709) Serial interface plug 1 (available only when the optional serial interface card is mounted in the CSD-709) CC-Link plug 1 (available only when the optional CC-Link card is mounted in the CSD-709) CD-ROM * (CSD-709D-01, EzCTS) USB cable 1 (available only when the optional additional memory is mounted in the CSD-709) 1 (available only when the optional additional memory is mounted in the CSD-709) * The CD-ROM attached when the optional additional memory is mounted in the CSD-709 includes both of the dedicated application CSD-709D-01 and the EzCTS. 138

141 A Specifications of CSD-709 A.11 Option A.11.1 Current output Model CSD Output Load resistance Resolution Non-linearity Specifications Over-range Output times Influence of temperature DC4 ma ~ 20 ma 510, max 1/12 000, min %F.S. Approx. DC 2.4 ma under display of [-OL], Approx. DC 21.6 ma under display of [OL]. Synchronized with AD sampling Zero point: 0.5 A/ C Sensitivity: %F.S./ C A.11.2 Voltage output Model CSD Output Load resistance Resolution Non-linearity Specifications Over-range Output times Influence of temperature DC ± 10 V 2 k, min 1/12 000, min %F.S. Approx. -11 V under display of [-OL], Approx. 11 V under display of [OL]. Synchronized with AD sampling Zero point: 0.6 mv/ C Sensitivity: %F.S./ C A.11.3 BCD output (sink type) Model CSD Specifications Output BCD five-digit parallel output Accompanied by polarity (POL.) (Output is turned on at negative polarity and turned off at positive polarity.) P.C. (print command): Turned on for a given period after completion of BCD output conversion ERROR: Turned on whenever any error occurs. OVER Output times Can be changed to 4, 20, 50, 100, 400, 1 000, 2 000, or times/s Input SEL.1, SEL2: You can select the desired brand code. HOLD: Display or BCD output is held. BCD-ENABLE: BCD-related outputs are forcibly turned off (high impedance). 139

142 A Specifications of CSD-709 A.11.4 BCD output (source type) Model CSD Specifications Output BCD five-digit parallel output Output times Input Accompanied by polarity (POL.) (Output is turned on at negative polarity and turned off at positive polarity.) P.C. (print command): Turned on for a given period after completion of BCD output conversion ERROR: Turned on whenever any error occurs. OVER Can be changed to 4, 20, 50, 100, 400, 1 000, 2 000, or times/s SEL.1, SEL2: You can select the desired brand code. HOLD: Display or BCD output is held. BCD-ENABLE: BCD-related outputs are forcibly turned off (high impedance). A.11.5 External control voltage input type circuit Model CSD Specifications Signal is output by applying a voltage between input and COM1 terminals. A.11.6 RS-232C interface Model CSD Specifications Baud rate Select bps, bps, bps, bps, bps, bps, bps, or bps. Data bit length Select [7 bit] or [8 bit]. Parity bit Select [No parity], [Even parity], or [Odd parity]. Stop bit Select [1 bit] or [2 bit]. Terminator Select [CR] or [CR+LF]. Transmission mode Half-duplex Sync mode Asynchronous Transmission data ASCII code Cable length 15 m, max; equipped with input/output monitor LED 140

143 A Specifications of CSD-709 A.11.7 RS-422/RS-485 interface Model CSD Specifications Baud rate Select bps, bps, bps, bps, bps, bps, bps, or bps. Data bit length Select [7 bit] or [8 bit]. Parity bit Select [No parity], [Even parity], or [Odd parity]. Stop bit Select [1 bit] or [2 bit]. Terminator Select [CR] or [CR+LF]. Transmission mode Half-duplex Sync mode Asynchronous Address Select one address within the range from 0 to 31. Transmission data ASCII code Cable length Approx. 1 km No. of connections 32 units, max (RS-422: 10 units) Termination Change between RS-422 and RS-485 Built in (Existence or nonexistence is selected by plugging in the connector board); equipped with input/output monitor LED Function is used. A.11.8 Serial interface (SI/F) Model CSD Specifications Baud rate 600 bps Data bit length Parity bit Stop bit Start bit Transmission mode 8 bit Odd 1 bit 1 bit Binary code or BCD A.11.9 Power supply voltage Model CSD Specifications Power supply 12 VDC ~ 24 VDC (10.6 VDC ~ 27.6 VDC) Power consumption Approx. 3.6 W (@ 24 VDC; no option) Approx. 4.9 W, max (@ 24 VDC; equipped with optional CC-Link) Inrush current 600 ma (20 ms) in 12 VDC average load condition (ordinary temperature and cold start) 250 ma (50 ms) in 24 VDC average load condition (ordinary temperature and cold start) 141

144 A Specifications of CSD-709 A CC-Link interface Model CSD Specifications CC-Link version Ver.1.10 Number of shared stations Select 1, 2, or 4. Baud rate Communication system Sync mode Transmission path Transmission format Remote station number No. of connections Connecting cable Termination Status LEDs Select 156 bps, 625 bps, 2.5 Mbps, 5 Mbps, or 10 Mbps. Polling Bit synchronous RS-485 bus Conforming to HDLC 01 ~ 64 (one station is shared), 01 ~ 63 (two stations are shared), or 01 ~ 61 (four stations are shared) Up to 64 (one station is shared), up to 32 (two stations are shared), or up to 16 (four stations are shared) Shielded twisted-pair cable. The cable used shall be specialized for the CC-Link. External resistor Communication status is indicated with the associated LED (RUN, ERR, SD, or RD). A Additional memory Model CSD Specifications Calendar function is available. Comparator judgment result, peak, and error log can be stored. The stored data can be read using the specialized software. 142

145 B Warranty B Warranty B.1 Warranty The warranty for this product is valid for a period of one year from the date of delivery. For repair and after-sale service during the warranty period, contact your sales representative or dealer. B.2 Repairs Before requesting repair of your CSD-709, see Section 9.2, "Troubleshooting" to confirm you have properly conducted connections, settings, and adjustments. Particularly, check wires of the strain-gage-based transducer for disconnections or breaks. When it can be judged from the result of check-outs that the CSD-709 is in faulty condition, request a repair from your sales representative or dealer. When returning the equipment to us, always put it into an antistatic bag and make sure that the equipment will not be subject to impact during transportation. B.3 Component Service Lives Components used in the module have finite service lives. Service life will vary depending on operating methods and environmental conditions. Guidelines on service life are as shown below. Component Purpose Approximate life EEPROM Setting data storage Number of EEPROM write times: Approx Electrolytic capacitor Removal of power supply noise Approx. 10 years (@ operating temperature of 20 C) 143

146 C Character Display Pattern C Character Display Pattern Shown below are display patterns in the 7-segment LED of the CSD

147 D Detection Operation and Section D Detection Operation and Section D.1 Detection operation The CSD-709 uses the following detection operations to hold the loads. Peak hold Bottom hold Peak-bottom hold Peak-to-peak hold Maximum hold Minimum hold Maximum-minimum difference hold Inflection point hold Average D.1.1 Peak hold In the peak hold mode, the load measured immediately before is compared to the current load and, when [load measured immediately before current load] is established, the current load is held as a peak. A (held as peak) 145

148 D Detection Operation and Section In actual measurements, the peak shown below is held. When the RESET input is generated by, for example, pressing during the detection process, the displayed load will be [0]. 146

149 D Detection Operation and Section D.1.2 Bottom hold In bottom hold mode, the load measured immediately before is compared to the current load and, when [load measured immediately before current load] is established, the current load is held as a bottom. A (held as bottom) In actual measurements, the bottom shown below is held. When the RESET input is generated during the detection process, the displayed load will be [0]. 147

150 D Detection Operation and Section D.1.3 Peak-bottom hold In the peak-bottom hold mode, the absolute value of the difference between the load measured immediately before and the reference point set to P function PF-10 is compared to the absolute value of the difference between the current load and reference point and, when [ load measured immediately before - reference point current load - reference point ] is established, the current load is held. When X Y, the load, B, is held instead of B the load measured immediately before, A. Reference point Y X A In actual measurements, maximum and minimum loads against the reference point are held as shown below. 148

151 D Detection Operation and Section D.1.4 Peak-to-peak hold In the peak-to-peak hold mode, the difference between the peak and bottom is held. B (Peak) X (Peak-to-peak value) A (Bottom) D.1.5 Maximum, minimum, and maximum-minimum difference When the peak-to-peak value deviates from the detection data width set to P function PF-11, the current peak is held as a maximum. A (Maximum) Detection data width Peak-to-peak value B 149

152 D Detection Operation and Section In actual measurements, the maximum is held as shown below. When the peak-to-peak value deviates from the detection data width and then the difference between the bottom and the track exceeds the detection magnification set to P function PF-12, the bottom is held as a minimum. C A Y Detection data width X Detection magnification B (Minimum) 150

153 D Detection Operation and Section In actual measurements, the minimum is held as shown below. The difference between the maximum and minimum is held as the maximum-minimum difference. 151

154 D Detection Operation and Section D.1.6 Inflection point The CSD-709 can detect sharp changes in load using four inflection points. Inflection point B Inflection point C Inflection point A Inflection point D Inflection point A B C D Detection point The point at which application of load is initiated from the no-load condition The point at which the load reaches the peak in the load application process The point at which the load is reduced from the peak The point at which the no-load condition is established after continuation of load reduction. For example, the load change X between the start and end points of detection time A set to P function PF-21 is subtracted from the load change Y between the start and end points of detection time B set to P function PF-22, and the result is defined as value Z. When Z exceeds the detection data width set to PF-20, point [a] is detected as inflection point A. Inflection point 152

155 D Detection Operation and Section In actual held. measurements, the inflection point A is detected as shown below and the load at point A is D.1.7 Average The average is the value determined by the average update time set to P function PF-42. CAUTION The average can be used when the time-specified section or automatic start time-specified section is defined as the detection section. 153

156 D Detection Operation and Section D.2 Detection section The CSD-709 can execute the detection operation in the following four sections. Entire section Specified section Time-specified section Automatic start time-specified section D.2.1 Entire section In the entire section, the detection operation is initiated or completed with the PEAK/TRACK input enabled by, for example, pressing. Detection operation section PEAK/TRACK input 154

157 D Detection Operation and Section D.2.2 Specified section In the specified section, the detection operation and hold operation are toggled alternately with the PEAK/TRACK input enabled by, for example, pressing. Detection operation section Hold operation section Detection operation section PEAK/TRACK input D.2.3 Time-specified section In the time-specified section, the detection operation is executed with the PEAK/TRACK input enabled by, for example, pressing within the time period already set to P function PF-40. Detection operation section Set time period PEAK/TRACK input 155

158 D Detection Operation and Section D.2.4 Automatic start time-specified section In the automatic start time-specified section, the detection operation is initiated when the track crosses the threshold (automatic start level) set to P function PF-41. This time, the section determined by the time period set to P function PF-40 will be the detection section. Detection operation section Threshold Set time period In the automatic start time-specified section, the detection operation is initiated when the track crosses the threshold (automatic start level) set to P function PF-41 as shown below. Threshold When the track reaches the threshold but does not cross it as shown below, the detection operation is not initiated. Threshold 156

159 E Voltage Output Card CSD E Voltage Output Card CSD E.1 Wiring WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the voltage output card can output ±10 V in agreement with the displayedd value. Shown below are terminals arranged on the voltage output card. No Terminal Voltage output (+ +) Voltage output (-) Frame ground (F.G.) 157

160 E Voltage Output Card CSD E.2 Setting The voltage output card outputs the voltage according to settings of C functions CF-30 to CF-41 connected to the calibration numbers, as shown below. Calibration number C function used 0 CF-30 ~ 32 1 CF-33 ~ 35 2 CF-36 ~ 38 3 CF-39 ~ 41 In each of C functions connected to these calibration numbers, you should give the following settings. What parameter, display interlock, track, or peak, should be defined as the output target. Minimum display value Maximum display value Described below are setting procedures used to output the voltage according to the calibration numbers. 1. Press once to display. 2. Press once to change the display to. 3. Press. C Function mode is activated, and the display changes to a two-digit numerical value prefixed by [CF-]. 4. Using and, change the numerical value to any of,,, and, then press. Any of the following numbers showing the currently set output target appears. Number Output target 0 Display interlock 1 Track 2 Peak Default 0 (Display interlock) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the output target. Whenever is pressed, the number times through the values from 0 to 2 in ascending order. 6. Press. 7. After confirming,,, or is displayed, press. The currently set minimum display value appears. Default 0 158

161 E Voltage Output Card CSD Using and, set the maximum display value. You can set any value between and While the maximum display value is being displayed, the output from the voltage output card is 0 V. 9. Press. 10. After confirming,,, or is displayed, press. The currently set maximum display value appears. Default Using and, set the maximum display value. You can set any value between and While the maximum display value is being displayed, the output from the voltage output card is 10 V. 12. Press. 13. Press twice to return to Measurement mode. 159

162 E Voltage Output Card CSD E.3 Calibration Once you set the display range of ±10 V output from the voltage output card, you should calibrate the voltage actually output from the card using the following procedures. Transition to the CAL selection mode To execute calibration, enter CAL selection mode. Transition to the VCAL mode To calibrate the output voltage, enter VCAL mode. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the minimum display value. Adjustment of voltage Adjust the voltage of the minimum display value. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the maximum display value. Adjustment of voltage Adjust the voltage of the maximum display value. Registration Register the calibration data. Described below are output voltage calibration procedures. 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press twice to change the display to. 5. Press. The VCAL mode is activated and appears. 6. Press. The current load appears. Adjust the input from the measuring unit so that it will be minimum display value. 7. Using and, adjust the voltage of the minimum display value, then press. appears. 160

163 E Voltage Output Card CSD Press. The current load appears. Adjust the input from the measuring unit so that it will be the maximum display value. 9. Using and, adjust the voltage of the maximum display value, then press. appears. 10. Press. The calibration data are registered and the VCAL mode is completed. 11. Press twice to return to Measurement mode. 161

164 E Voltage Output Card CSD E.4 Adjustment After the voltage output from the voltage output card is calibrated, an error may be generated between displayed and actual voltages due to the environment in which the CSD-709 is used. In such a case, fine adjustment should be implemented for the output from the voltage output card using the following procedures. Transition to the CAL selection mode To execute calibration, enter CAL selection mode. Transition to the VADJ mode To fine-adjust the output voltage, enter VADJ mode. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the minimum display value. Adjustment of voltage Adjust the voltage of the minimum display value. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the maximum display value. Adjustment of voltage Adjust the voltage of the maximum display value. Registration Register the calibration data. Described below are fine adjustment procedures. 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press three times to change the display to. 5. Press. The VADJ mode is activated and appears. 6. Press. The current load appears. Adjust the input from the measuring unit so that it will be the minimum display value. 7. Using and, adjust the voltage of the minimum display value, then press. appears. 162

165 E Voltage Output Card CSD Press. The current load appears. Adjust the input from the measuring unit so that it will be the maximum display value. 9. Using and, adjust the voltage of the maximum display value, then press. appears. 10. Press. The adjustment data is registered and VADJ mode is completed. 11. Press twice to return to Measurement mode. 163

166 E Voltage Output Card CSD E.5 Output Operation in Hold Mode You can set whether any of the following three hold targets should be held using function F-18. Data display in the load window Result of comparison with the comparator Output of load via optional device (voltage output card, current output card, BCD card, serial interface card, or CC-Link card) In case a voltage output card is used, the output takes the following styles in the hold condition depending on the settings: execution/no-execution of hold determined by function F-18 and output targets established by C functions CF-30, CF-33, CF-36, and CF-39. Hold target established by F-18 Output target established by CF-30, CF-33, CF-36, and CF-39 Display interlock Track Peak Data display in the load The value held is output. The value is not held. The value is not held. window Option The value held is output. The value held is output. The value held is output. Whether the comparator s comparison result set to the second digit of F-18 is held will not affect the output targets established by CF-30, CF-33, CF-36, and CF-39. For setting of hold targets, see Section 7.9, "Setting hold targets". E.6 Checking on Problems If an unexpected operation occurs during the use of a voltage output card, verify phenomena and check-outs listed in the table below to take proper action. Phenomenon Voltage is not output correctly. Check-out Check whether the analog output cable is plugged in correctly. Check whether the load resistance is within the allowable range. Check whether the C function for the voltage output is correctly set. 164

167 F Current Output Card CSD F F.1 Current Outpu Card CSD Wiring WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the current output card can output current rangingg from 4 to 20 ma in agreement with the displayedd value. Shown below are terminals arranged on the current output card. No Terminal Current output (+ +) Current output (-) Frame ground (F.G.) 165

168 F Current Output Card CSD F.2 Setting The current output card outputs current according to the settings for C functions CF-30 to CF-41 connected to the calibration numbers, as shown below. Calibration number C function used 0 CF-30 ~ 32 1 CF-33 ~ 35 2 CF-36 ~ 38 3 CF-39 ~ 41 For each C function connected to these calibration numbers, make the following settings: What parameter, display interlock, track, or peak, should be defined as the output target. Minimum display value Maximum display value Described below are the setting procedures used to output current according to the calibration numbers. 1. Press once to display. 2. Press once to change the display to. 3. Press. C Function mode is activated, and the display changes to a two-digit numerical value prefixed with [CF-]. 4. Using and, change the numerical value to any of,,, and, and press. Any of the following numbers showing the currently set output target appears: Number Output target 0 Display interlock 1 Track 2 Peak Default 0 (Display interlock) For setting procedures with and, see Section 1.6, "Basic key operation". 5. Press to set the output target. Whenever is pressed, the number times through the values from 0 to 2 in ascending order. 6. Press. 7. After confirming,,, or is displayed, press. The currently set minimum display value appears. Default 0 166

169 F Current Output Card CSD Using and, set the minimum display value. You can set any value between and While the minimum display value is being displayed, the output from the current output card is 4 ma. 9. Press. 10. After confirming,,, or is displayed, press. The currently set maximum display value appears. Default Using and, set the maximum display value. You can set any value between and When the maximum display value is being displayed, the output from the current output card is 20 ma. 12. Press. 13. Press twice to return to Measurement mode. 167

170 F Current Output Card CSD F.3 Calibration Once you set the current output from the current output card within the display range from 4 to 20 ma, you should calibrate the current actually output from the card using the following procedures. Transition to the CAL selection mode To execute calibration, enter CAL selection mode. Transition to the VCAL mode To calibrate the output current, enter VCAL mode. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the minimum display value. Adjustment of current Adjust the current of the minimum display value. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the maximum display value. Adjustment of current Adjust the current of the maximum display value. Registration Register the calibration data. Described below are the output current calibration procedures. 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press twice to change the display to. 5. Press. The VCAL mode is activated, and appears. 6. Press. The current load appears. Adjust the input from the measuring unit so that it will be a minimum display value. 7. Using and, adjust the current of the minimum display value, then press. appears. 8. Press. The current load appears. Adjust the input from the measuring unit so that it will be a maximum display value. 168

171 F Current Output Card CSD Using and, adjust the current of the maximum display value, then press. appears. 10. Press. The calibration data is registered, and VCAL mode is completed. 11. Press twice to return to Measurement mode. 169

172 F Current Output Card CSD F.4 Adjustment After the current output from the current output card is calibrated, an error may be generated between the displayed and actual current due to the environment in which the CSD-709 is used. In such cases, make fine adjustments for the output from the current output card using the following procedures. Transition to the CAL selection mode To execute calibration, enter CAL selection mode. Transition to the VADJ mode To fine-adjust the output current, enter VADJ mode. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the minimum display value. Adjustment of current Adjust the current of the minimum display value. Adjustment of input from measuring unit Adjust the input from the measuring unit so that it will be the maximum display value. Adjustment of current Adjust the current of the maximum display value. Registration Register the calibration data. Described below are fine adjustment procedures. 1. Press once to display. 2. Press three times to change the display to. 3. Press to change the display to. 4. Press three times to change the display to. 5. Press. The VADJ mode is activated, and appears. 6. Press. The current load appears. Adjust the input from the measuring unit so that it will be the minimum display value. 7. Using and, adjust the current of the minimum display value, then press. appears. 8. Press. The current load appears. Adjust the input from the measuring unit so that it will be the maximum display value. 170

173 F Current Output Card CSD Using and, adjust the current of the maximum display value, then press. appears. 10. Press. The adjustment data is registered, and VADJ mode is completed. 11. Press twice to return to Measurement mode. 171

174 F Current Output Card CSD F.5 Output Operation in Hold Mode You can set whether any of the following three hold targets should be held using function F-18. Data display in the load window Result of comparison with the comparator Output of load via optional device (voltage output card, current output card, BCD card, serial interface card, or CC-Link card) In case a current output card is used, the output takes the following styles in the hold condition depending on the settings: execution/non-execution of hold determined by function F-18 and output targets established by C functions CF-30, CF-33, CF-36, and CF-39. Hold target Output target established by CF-30, CF-33, CF-36, and CF-39 established by F-18 Display interlock Track Peak Data display in the load The value held is The value is not held. The value is not held. window output. Option The value held is output. The value held is output. The value held is output. Whether the comparator s comparison result set to the second digit of F-18 is held will not affect the output targets established by CF-30, CF-33, CF-36, and CF-39. For settings for hold targets, see Section 7.9, "Setting hold targets". F.6 Checking on Problems If an unexpected operation occurs during the use of a current output card, verify phenomena and check-outs listed in the table below to take proper action. Phenomenon Current is not output correctly. Check-out Check whether the analog output cable is plugged in correctly. Check whether the load resistance is within the allowable range. Check whether the C function for the current output is correctly set. 172

175 G G.1 BCD Wiring G BCD Card CSD /CSD Card CSD /CSD WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the BCD card can provide BCD output. Shown below are terminals arranged on the BCD card. 173

176 G BCD Card CSD /CSD BCD output (sink type) Pin No. Output/Input A1 - A2 Output A3 Output A4 Output A5 Output A6 Output A7 Output A8 Output A9 Output A10 Output A11 Output A12 Input A13 Output A14 Output A15 Input A16 - BCD output (source type) Pin No. Output/Input A1 - A2 Output A3 Output A4 Output A5 Output A6 Output A7 Output A8 Output A9 Output A10 Output A11 Output A12 Input A13 Output A14 Output A15 Input A16 - Signal COM Pin No. B1 B2 B3 Output/Input - Output Output SEL.1 POL. ERRORR HOLD N.C. Signal +24V B B B B B B11 SEL.1 B12 POL. B13 ERRORR B14 HOLD B15 0 V B16 B4 Output B5 Output B6 Output B7 Output B8 Output B9 Output B10 Output B11 Output B12 Input B13 Output B14 Output B15 Input B16 - Pin No. Output/Input B1 - B2 Output B3 Output B4 Output B5 Output CAUTION Output Output Output Output Output Output Input Output Output Input - Signal COM SEL.2 OVER P.C. BCD-ENABLE N.C. Signal +24V SEL.2 OVER P.C. BCD-ENABLE 0 V The card is insulated from the internal circuit of the CSD-709 by means of a photocoupler. Only BCD-ENABLE (pin number B15) of the source type becomes effective at the [OPEN]. Use this function after short-circuiting B15 and A1 or B1 when you do not use it. 174

177 G BCD Card CSD /CSD G.2 Outputt specifications Specifications BCD data POL. (polarity) P.C. (print command) ERROR OVER Decimal point display position (10 1, 10 2, 10 3, 10 4 ) STABLE GROSS HOLD BCD-ENABLE Input/Output Output Output Output Output Output Output Output Output Input Input Details Five-digit parallel output Output is turned on for the negative polarity while turned off for the positive polarity. Turned on for a given period after completion of BCD output conversion Turned on whenever any error occurs Turned on in synchronization with the OL display (overload) Turned on in synchronization with the decimal point display position setting Turned on upon detection of stable condition Turned on upon output of total value data Holds the BCD output. Forces the BCD output into high impedance condition. G.3 Setting Output operation mode, output target, outpu logic, PC width, and output frequency for the BCD are set with the following functions. Item Function Setting BCD output operation mode BCD output target BCD output logic F-30 F-31 F-32 0 (Stream mode), 1 (Print signal synchronization) 0 (Display interlock), 1 (Track), 2 (Peak) 0 (Negative logic) or 1 (Positive logic) is set using four digits. Digit Setting details First digit Dataa output logic (load data and decimal point display position) Second digit Polarity output logic Third digit Plug output logic (error, over, stable, total value) Fourth digit P.C.. (print command) output t logic BCD PC width F-33 0 (125 ms), 1 (25 ms), 2 (10 ms), 3 (5 ms), 4 (1.25 ms), 5 (0.5 ms), 6 (0.25 ms), 7 (0.125 ms) F-34 0 (4 times/s), 1 (20 times/s), 2 (50 times/s), 3 (100 times/s), BCD output 4 (400 times/s), 5 (1 000 times/s) ), 6 (2 000 times/s), frequency 7 (4 000 times/s) CAUTION During the rising stage of the print command, the BCD data is being converted, and the values are therefore unstable. For this reason, read the BCD data during the falling stage of the print command. 175

178 G BCD Card CSD /CSD G.4 Timing chart Measurement mode P.C. Output transistor turns ON (Negative logic in electrical theory) when all of the P.C., DATA and POL output the data. When data is over (±OL display) DATA POL ON P.C. ON OVR Output transistor of OVR signal turns ON (Negative logic in electrical theory) while OVR is output. Moreover, for all DATA when OVR is output, the output transistor becomes OFF status (Positive logic in electrical theory) When the error is occurred. ON DATA POL ON P.C. ON ERROR Output transistor of ERROR signal turns on (Negative logic in electrical theory) while ERROR is output. Moreover, for all DATA and POL when ERROR is output, the output transistor becomes OFF status (Positive logic in electrical theory). When HOLD signal is input. HOLD DATA POL P.C. Output transistor of P.C. turns off (Positive logic in electrical theory) while HOLD signal is input. The response time showing above will take for HOLD or reset of DATA and POL., after HOLD signal is input. (The input response time is set by function.) 176 DATA POL times/s: Approx ms at max.+input response time times/s: Approx. 0.5 ms at max.+input response time times/s: Approx. 1 ms at max.+input response time 400 times/s: Approx. 2.5 ms at max.+input response time 100 times/s: Approx. 10 ms at max.+input response time 50 times/s: Approx. 20 ms at max.+input response time 20 times/s: Approx. 50 ms at max.+input response time 4 times/s: Approx.250 ms at max +Input response time ON ON times/s: Approx.0.25 ms times/s: Approx.1 ms 100 times/s: Approx.10 ms 20 times/s: Approx.50 ms ON ON times/s: Approx. 0.25ms at max.+input response time, times/s: Approx. 1 ms at max.+input response time, 100 times/s: Approx. 10 ms at max.+input response time, 20 times/s: Approx. 50 ms at max.+input response time, times/s: Approx.0.5 ms 400 times/s: Approx.2.5 ms 50 times/s: Approx.20 ms 4 times/s: Approx.250 ms Changeable to approx ms, approx.0.25 ms, approx.0.5 ms, approx.1.25 ms, approx.5 ms, approx.10 ms, approx.25 ms or approx.125 ms times/s: Approx ms at max. +Input response time times/s: Approx. 0.5 ms at max.+input response time times/s: Approx. 1 ms at max.+input response time 400 times/s: Approx. 2.5 ms at max.+input response time 100 times/s: Approx. 10 ms at max. +Input response time 50 times/s: Approx. 20 ms at max. +Input response time 20 times/s: Approx. 50 ms at max. +Input response time 4 times/s: Approx.250 ms at max +Input response time times/s: Approx ms at max. +Input response time 400 times/s: Approx. 2.5 ms at max.+input response time 50 times/s: Approx. 20 ms at max.+input response time 4 times/s: Approx. 250 ms at max +Input response time

179 G BCD Card CSD /CSD G.5 Output Operation in Hold Mode You can set whether any of the following three hold targets should be held using function F-18. Data display in the load window Result of comparison with comparator Output of load via optional device (voltage output card, current output card, BCD card, serial interface card, or CC-Link card) In case a BCD card is used, the output takes the following styles in the hold condition depending on the settings: execution/non-execution of hold determined by function F-18 and output targets established by functions F-31. Hold target Output target established by F-31 established by F-18 Display interlock Track Peak Data display in the load The value held is The value is not held. The value is not held. window output. Option The value held is output. The value held is output. The value held is output. Whether the comparator s comparison result set to the second digit of F-18 is held will not affect the output targets established by F-31. For settings for hold targets, see Section 7.9, "Setting hold targets". G.6 Checking on Problems If an unexpected operation occurs during the use of a BCD card, verify phenomena and check-outs listed in the table below to take proper action. Phenomenon Communications are not established. Intended value is not output. BCD output is not changed. Check-out Check whether the BCD cable is plugged in correctly. Check the output logic setting. Check the output target setting. The output has been set as the hold target. 177

180 H RS-232C Card CSD H H.1 RS-232C Card Wiring CSD WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the RS-232C card can establish communications with the CSD-709 via the RS-232C interface. Shown below are the terminals arranged on the RS-232C card. RXD TXD Pin No Signal DTR TXD RXD S.G. Description Dataa terminal ready Transmission data Reception data Signal ground 178

181 H RS-232C Card CSD H.2 Communication specifications Item Transmission Sync mode Baud rate Data bit length Parity bit Stop bit Terminator Communication data Address Communication mode Half-duplex Asynchronous Details Select bps, bps, bps, bps, bps, bps, bps, or bps. Select [7 bit] or [8 bit]. Select [No parity], [Even parity], or [Odd parity]. Select [1 bit] or [2 bit]. Select [CR] or [CR+LF]. ASCII code 00 (Fixed) [Command mode] or [Stream mode] H.3 Setting Operation mode, output target, communication settings, addition of decimal point, and number of digits of unit for the RS-232C are set with the following functions. Item Function Setting Operating modes F-45 0 (Command mode), 1 (Stream mode), 2 (Print command synchronization), 3 (Modbus) Output target F-46 0 (Display interlock), 1 (Track), 2 (Peak), 3 (Combination of load display and status) Communication F-47 Set with a five-digit numerical value. settings Digit Details First digit Set the data bit length using 0 (7 bit) or 1 (8 bit). Second digit Set the parity bit using 0 (No parity), 1 (Even parity), or 2 (Odd parity). Third digit Set the stop bit using 0 (1 bit) or 1 (2 bit). Fourth digit Set the baud rate using 0 (1 200 bps), 1 (2 400 bps), 2 (4 800 bps), 3 (9 600 bps), 4 ( bps), 5 ( bps), 6 ( bps), or 7 ( bps). Fifth digit Set the terminator using 0 (CR) or 1 (CR+LF). Decimal point F-48 0 (Hide), 1 (Add) Number of digits F-49 0 (Two digits), 1 (Three digits) of unit Unit output character CF-05 0 (None), 1 (Gram), 2 (Kilogram), 3 (Pound), 4 (Ton), 5 (Newton), 6 (Kilonewton), 7 (Pascal), 8 (Kilopascal), 9 (Megapascal) For more information about settings, see Section 10.2, "Setting the USB interface". For information on Command mode, Stream mode, and Print command synchronization mode, see Section 10.3, "Data format". For information on Modbus, see Attachment J, "Communications Based on Modbus Protocol". 179

182 H RS-232C Card CSD H.4 Checking on Problems If an unexpected operation occurs during the use of an RS-232C card, verify the phenomena and check-outs listed in the table below to take proper action. Phenomenon Communications are not established. Intended value is not output. Check-out Check whether the RS-232C cable is plugged in correctly. Check the settings for the communication interface. Check whether the command number and data are correctly The flow control is enabled on the host PC. Check the output target setting. 180

183 I RS-422/485 Card CSD I RS-422/485 Card CSD I.1 Connection WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the RS-422/485 card can establish communications with the CSD-709 via the RS-422/485 interface. Shown below are the terminals arranged on the RS-422/485 card. RXD TXD Pin No. Signal Description 1 SDA Differential output (+) 2 SDB Differential output (-) 3 RDA Differential output (+) 4 RDB Differential output (-) 5 TRM Termination resistor 6 S.G. Signal ground To enablee the built-in terminationn resistor, short-circuit TRM and RDB. 181

184 I RS-422/485 Card CSD I.2 Communication specifications Item Transmission Sync mode Baud rate Data bit length Parity bit Stop bit Terminator Communication data Half-duplex Details Asynchronous Select bps, bps, bps, bps, bps, bps, bps, or bps. Select [7 bit] or [8 bit]. Select [No parity], [Even parity], or [Odd parity]. Select [1 bit] or [2 bit]. Select [CR] or [CR+LF]. ASCII code Address Select any value between 00 and 31. Communication mode [Command mode] or [Stream mode] 182

185 I RS-422/485 Card CSD I.3 Setting Operation mode, output target, communication settings, addition of decimal point, number of digits of unit, address, change between RS-422 and RS-485 for the RS-422/485, return data delay for the RS-485 are set with the following functions. Item Function Setting Operating modes F-45 0 (Command mode), 1 (Stream mode), 2 (Print command synchronization), 3 (Modbus) Output target F-46 0 (Display interlock), 1 (Track), 2 (Peak), 3 (Combination of load display and status) Communication F-47 Set with a five-digit numerical value. settings Digit Details First digit Set the data bit length using 0 (7 bit) or 1 (8 bit). Second digit Set the parity bit using 0 (No parity), 1 (Even parity), or 2 (Odd parity). Third digit Set the stop bit using 0 (1 bit) or 1 (2 bit). Fourth digit Set the baud rate using 0 (1 200 bps), 1 (2 400 bps), 2 (4 800 bps), 3 (9 600 bps), 4 ( bps), 5 ( bps), 6 ( bps), or 7 ( bps). Fifth digit Set the terminator using 0 (CR) or 1 (CR+LF). Decimal point F-48 0 (Hide), 1 (Add) Number of digits F-49 0 (Two digits), 1 (Three digits) of unit Address F ~ 31 Change between F (RS-422), 1 (RS-485) RS-422 and RS-485 Return data delay F-52 0 ~ 99 (unit: 10 ms) in RS-485 Unit output character CF-05 0 (None), 1 (Gram), 2 (Kilogram), 3 (Pound), 4 (Ton), 5 (Newton), 6 (Kilonewton), 7 (Pascal), 8 (Kilopascal), 9 (Megapascal) For more information about settings, see Section 10.2, "Setting the USB interface". For information on Command mode, Stream mode, and Print command synchronization mode, see Section 10.3, "Data format". For information on Modbus, see Attachment J, "Communications Based on Modbus Protocol". 183

186 I RS-422/485 Card CSD I.4 Checking on Problems If an unexpected operation occurs during the use of an RS-422/485 card, verify phenomena and check-outs listed in the table below to take proper action. Phenomenon Communications are not established. Check-out Check whether the RS-422/485 cable is plugged in correctly. Check whether the address number is correctly set. Check the settings for the communication interface. Check whether the command number and data are correctly established. The flow control is enabled on the host PC. Check whether the change between RS-422 and RS-485 is correctly set. The return data delay in RS-485 is set to less than 1 ms. Intended value is not output. Check the output target setting. 184

187 J Communications Based on Modbus Protocol J Communications Based on Modbus Protocol J.1 Communication specifications Item Transmission Sync mode Baud rate Data bit length Parity bit Stop bit Half-duplex Asynchronous Details Select bps, bps, bps, bps, bps, bps, bps, or bps. Select [7 bit] or [8 bit]. Select [No parity], [Even parity], or [Odd parity]. Select [1 bit] or [2 bit]. Address Select any value between 00 and 31. Serial transmission mode Modbus RTU mode 185

188 J Communications Based on Modbus Protocol J.2 Setting After the CSD-709 is connected to the RS-232C or RS-422/485 card, you can establish communications using the Modbus protocol. The table below shows the functions used for communications based on the Modbus protocol. Item Function Setting Operating modes F-45 3 (Modbus) Output target F-46 0 (Display interlock), 1 (Track), 2 (Peak), 3 (Combination of load display and status) Communication F-47 Set with a five-digit numerical value. settings Digit Details First digit Set the data bit length using 0 (7 bit) or 1 (8 bit). Second digit Set the parity bit using 0 (No parity), 1 (Even parity), or 2 (Odd parity). Third digit Set the stop bit using 0 (1 bit) or 1 (2 bit). Fourth digit Set the baud rate using 0 (1 200 bps), 1 (2 400 bps), 2 (4 800 bps), 3 (9 600 bps), 4 ( bps), 5 ( bps), 6 ( bps), or 7 ( bps). Fifth digit Set the terminator using 0 (CR) or 1 (CR+LF). Decimal point F-48 0 (Hide), 1 (Add) Number of digits F-49 0 (Two digits), 1 (Three digits) of unit Address F ~ 31 Change between F (RS-422), 1 (RS-485) RS-422 and RS-485 Return data delay F-52 0 ~ 99 (unit: 10 ms) in RS-485 Unit output character CF-05 0 (None), 1 (Gram), 2 (Kilogram), 3 (Pound), 4 (Ton), 5 (Newton), 6 (Kilonewton), 7 (Pascal), 8 (Kilopascal), 9 (Megapascal) The address [00] shows the broadcast address. Any of the following values is determined as the silent interval of the Modbus protocol, depending on the baud rate set to function F-47 (communication settings). Baud rate (bps) Silent interval (ms)

189 J Communications Based on Modbus Protocol J.3 Modbus function forma J.3.1 Coil The coil offers reference and change functions using a binary-coded bit. Shown below is the data that can be referenced and changed by the coil. Data Address Zero set 1 Zero set clear 2 PEAK/TRACK 3 RESET 4 Forced termination of multimode 5 Change between hold and hold cancellation 20 CHECK (key operation)

190 J Communications Based on Modbus Protocol J.3.2 Input status The input status offers reference functions using a binary-coded bit. Shown below is the data that can be referenced by the input status. Data Address SEL1 17 SEL2 18 CHECK 19 HOLD 20 PEAK 21 MEAS. 22 END 23 LOCK 24 OUTPUT1 25 OUTPUT2 26 OUTPUT3 27 OUTPUT4 28 OUTPUT5 29 OUTPUT6 30 MULTI (multimode is used) 33 EXCUTE (measurement in the multimode) 34 OK 35 NG 36 Upper limit / lower limit 37 S0 38 S1 39 S2 40 Error 41 Faulty weight

191 J Communications Based on Modbus Protocol J.3.3 Input register The input register offers reference functions using a 16-bit length word. Shown below is the data that can be referenced by the input register. Data Address Decimal place 1 Unit 2 Track 5 Peak 7 Faulty weight 13 Error 15 Status 1 51 Status 2 52 Status error code 53 Track 54 ~ 55 Status 1 56 Status 2 57 Status error code 58 Peak 59 ~ 60 Status 1 61 Status 2 62 Status error code 63 Software version 99 In Status 1, the following statuses are represented by bits. Bit position Details 15 SEL1 14 SEL2 13 CHECK 12 HOLD 11 PEAK 10 MEAS. 9 END 8 LOCK 7 OUTPUT1 6 OUTPUT2 5 OUTPUT3 4 OUTPUT4 3 OUTPUT5 2 OUTPUT

192 J Communications Based on Modbus Protocol In Status 2, the following statuses are represented by bits. Bit position Details 15 MULTI (multimode is used) 14 EXCUTE (measurement in the multimode) 13 OK 12 NG 11 Upper limit / lower limit 10 S0 9 S1 8 S2 7 Error 6 Faulty weight When errors occur, error codes are represented by the status error codes shown below. Error code Cause 1 Zero set is executed outside the effective zero set range. 2 Invalid function is specified. 10 Load exceeds the display range. 11 Load falls below the display range. 12 Result of AD conversion is overflown. 13 Result of AD conversion is overflown. 20 Zero point in the calibration is below the specified value. 21 Zero point in the calibration is higher than the specified value. 22 Actual load in the calibration is lower than the specified value. 23 Actual load in the calibration is higher than the specified value. 190

193 J Communications Based on Modbus Protocol J.3.4 Hold register The hold register offers reference and change functions using a 16-bit length word. Shown below is the data that can be referenced and changed by the hold register. Data S0 (Calibration number 0) 11 S1 (Calibration number 0) 13 S2 (Calibration number 0) 15 S0 (Calibration number 1) 267 ~ 268 S1 (Calibration number 1) 269 ~ 270 S2 (Calibration number 1) 271 ~ 272 S0 (Calibration number 2) 523 ~ 524 S1 (Calibration number 2) 525 ~ 526 S2 (Calibration number 2) 527 ~ 528 S0 (Calibration number 3) 779 ~ 780 S1 (Calibration number 3) 781 ~ 782 S2 (Calibration number 3) 783 ~ 784 Calibration number Operating modes Address Unit weight ~ Setting of measuring weight ~ Setting of weight mass ~ Registration of zero point load cell output ~ Registration of span point load cell output ~ Actual load based calibration Under AD reading Actual load based calibration Registration of zero point Actual load based calibration Registration of span point Calibration complete Zero point fine adjustment Span point fine adjustment Fine adjustment mode, load display ~ Function Function Function Function Function Function Function 4 ~ ~ Function Function C function C function C function

194 J Communications Based on Modbus Protocol Data Address C function C function C function C function 4 ~ ~ C function C function P fuction P fuction P fuction P fuction P fuction 0 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 1 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 2 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 3 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 4 3 ~ ~ P fuction P fuction

195 J Communications Based on Modbus Protocol Data Address P fuction P fuction P fuction P fuction P fuction 5 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 6 3 ~ ~ P fuction P fuction P fuction P fuction P fuction P fuction P fuction 7 3 ~ ~ P fuction P fuction

196 J Communications Based on Modbus Protocol J.4 Calibration based on Modbus Using the following addresses in the hold register, you can execute remote calibration through the Modbus protocol. Data Calibration number Operating modes ~ Unit weight ~ Setting of measuring weight ~ Setting of weight mass Address ~ Registration of zero point load cell output ~ Registration of span point load cell output Actual load based calibration Under AD reading Actual load based calibration Registration of zero point Actual load based calibration Registration of span point Calibration complete Zero point fine adjustment Span point fine adjustment ~ Fine adjustment mode, load display Described below are the calibration procedures when the Modbus protocol is used. J.4.1 Calibration based on actual load 1. Set [9] to address (Operating modes) to enter Calibration mode. POINT When [0] is set to address 56833, you can terminate calibration. In this case, all settings established beforehand will be abandoned. 2. Set the calibration number to address Set the unit weight to address ~ Set the measuring weight to address ~ Set the weight mass to address ~ Set [1] to address (Actual load based calibration - Registration of zero point). 7. Refer to address (Actual load based calibration - Under AD reading) to check the setting is [0]. 8. Set [1] to address (Actual load based calibration - Registration of span point). 9. Refer to address (Actual load based calibration - Under AD reading) to check the setting is [0]. 10. Set [1] to address (Calibration complete). 194

197 J Communications Based on Modbus Protocol J.4.2 Calibration based on numerical value input 1. Set [9] to address (Operating modes) to enter Calibration mode. POINT When [0] is set to address 56833, you can terminate the calibration. In this case, all settings established beforehand will be abandoned. 2. Set the calibration number to address Set the unit weight to address ~ Set the measuring weight to address ~ Set the weight mass to address ~ Set the mv/v value of the zero point to address ~ in mv/v increment. 7. Set the mv/v value of the span to address ~ in mv/v increment. 8. Set [1] to address (Calibration complete). J.4.3 Fine adjustment 1. Set [10] to address (Operating modes) to enter Fine adjustment mode. POINT When [0] is set to address 56833, you can terminate the calibration. In this case, all settings established beforehand will be abandoned. 2. Set the calibration number to address Using address ~ (Fine adjustment mode, load display), refer to the currently set load. 4. Enter any of the following values to address to fine-adjust the zero point. Setting Fine adjustment 1 Slightly increases the adjustment amount. 10 Increases the adjustment amount. -1 Slightly decreases the adjustment amount. -10 Decreases the adjustment amount. 195

198 J Communications Based on Modbus Protocol 5. Enter any of the following values to address to fine-adjust the span. Setting Fine adjustment 1 Slightly increases the adjustment amount. 10 Increases the adjustment amount. -1 Slightly decreases the adjustment amount. -10 Decreases the adjustment amount. 6. Set [1] to address (Calibration complete). 196

199 K Serial Interface Card CSD K K.1 Serial Interfacee Card CSD Connection WARNING Turn off power before connecting or disconnecting cables. The CSD-709 is not equipped with a power switch. When you conduct work with the CSD-709 turned on, electric shock or equipment damage may result. CAUTION The connector board on the rear panel of the CSD-709 is a resin component. Do not drop it or subject it to strong impact. Always use shielded cables, except for power cables. Separate cables from noise sources, including power cables of peripheral devices and control I/O. In the conduit, arrange cables only for the CSD-709. Avoid joint use with other lines in the conduit. Ground the CSD-709 based on Type D single grounding using a protective ground connector. Also, do not use the grounding line of the CSD-709 to ground the power supply system. Using the serial interface card can output the load displayed on the CSD-709 to the printer. Shown below are terminals arranged on the serial interface card. Pin No. Signal 1 SI/ F+ Connect this pin plugged in. 2 SI/ /F- Connect this pin plugged in. 3 F.G. Frame ground Description to the serial interface terminal (+) of the device to be to the serial interface terminal (-) of the device to be 197

DIGITAL INDICATOR CSD-701B. Instruction Manual EN O

DIGITAL INDICATOR CSD-701B. Instruction Manual EN O DIGITAL INDICATOR CSD-701B Instruction Manual EN294-1127-O FOREWORD Thank you very much for your purchasing our Digital Indicator CSD 701B. This manual explains installation procedures and connecting

More information

Digital Indicator EN G

Digital Indicator EN G Digital Indicator EN294-1434-G Forwards Thank you very much for your purchasing Minebea s Digital Indicator CSD815B. This manual explains installation procedures and connecting method and also operating

More information

DIGITAL INDICATOR CSD-814B. Instruction Manual EN J

DIGITAL INDICATOR CSD-814B. Instruction Manual EN J DIGITAL INDICATOR CSD-814B Instruction Manual EN294-1213-J FOREWORD Thank you very much for your purchasing our Digital Indicator CSD 814B. This manual explains installation procedures and connecting

More information

V E1B Snap-in I/O Module

V E1B Snap-in I/O Module V200-18-E1B Snap-in I/O Module The V200-18-E1B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 16 isolated digital

More information

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions HITACHI EH-150 series PLC Resistance Temperature Detective input module Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction

More information

Molding Pressure Conversion Module MPC Instruction Manual

Molding Pressure Conversion Module MPC Instruction Manual Molding Pressure Conversion Module MPC-201-25 Instruction Manual EN294-1697F Introduction Thank you for purchasing the MPC-201-25 Molding Pressure Conversion Module. This instruction manual describes

More information

Model HU-224/225. Technical Information TI.224/ Humidity Transduce r FOR ADDITIONAL INFORMATION SEE HU-224/225 DATA SHEET

Model HU-224/225. Technical Information TI.224/ Humidity Transduce r FOR ADDITIONAL INFORMATION SEE HU-224/225 DATA SHEET Model /225 FOR ADDITIONAL INFORMATION SEE /225 DATA SHEET SPECIFICATIONS Accuracy*: ± 2% / ± 3% RH Range: 0-100% RH Hysteresis: ± 1% Supply Voltage:12-40 VDC 12-35 VAC (VDC output units only) Supply Current:

More information

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION

H Series PLC. ! : Indicates Compulsion. EH-150 Analog input module EH-AXH8M Instruction manual. Safety precautions DANGER CAUTION COMPULSION H Series PLC EH-150 Analog input module EH-AXH8M Instruction manual Thank you for purchasing a Hitachi Programmable Logic Controller. To operate it safely, please read this instruction manual and all the

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

V E2B Snap-in I/O Module

V E2B Snap-in I/O Module V200-18-E2B Snap-in I/O Module The V200-18-E2B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 16 isolated digital

More information

AD-4532B. Digital Indicator INSTRUCTION MANUAL 1WMPD A

AD-4532B. Digital Indicator INSTRUCTION MANUAL 1WMPD A AD-4532B Digital Indicator INSTRUCTI MANUAL 1WMPD4001683A WARNING DEFINITIS The warnings described in this manual have the following meanings: A potentially hazardous situation which, if not avoided, could

More information

TOUCH SCREEN CONTROLLER for Air Conditioning Control System

TOUCH SCREEN CONTROLLER for Air Conditioning Control System INSTALLATION MANUAL TOUCH SCREEN CONTROLLER for Air Conditioning Control System Model: BMS-TP064ACE BMS-TP064PWE BMS-TP52ACE BMS-TP52PWE Thank you very much for purchasing the TOSHIBA Touch Screen Controller.

More information

2-2 Testing space Tension and Compression test can be available at the lower section of movable crosshead. Upper tension type : Sold separately.

2-2 Testing space Tension and Compression test can be available at the lower section of movable crosshead. Upper tension type : Sold separately. 1.General Specifications TECHNO GRAPH This TECHNO GRAPH, universal testing machine has developed with the concept of achieving Fulfillment of functions and simple operations. Adopted with AC servo motor,

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

Tempco Instruction Manual

Tempco Instruction Manual Tempco Instruction Manual 1/16 DIN Solid State Temperature Controller Relay Output Solid State Output For Heating Model Numbers: TEC-901, TEC-902, TEC-905 Temperature controls in this series are designed

More information

E2Q4 Square Proximity Sensor

E2Q4 Square Proximity Sensor Inductive Square Slim, compact size M2 Plug-in connection Integrated short circuit and reverse polarity protection Active face positioning: Y-axis 5, X-axis 90 incremets Ordering Information DC type Connection

More information

Programmable Relay ZEN V2 Units

Programmable Relay ZEN V2 Units Programmable Relay ZEN V2 Units Please read and understand this catalog before purchasing the products. Please consult your OMRON representative if you have any questions or comments. Refer to Warranty

More information

Thread length. 27 (40) mm. 27 (44) mm. 27 (40) mm. 34 (50) mm. 34 (49) mm. 39 (60) mm. 39 (54) mm. 44 (65) mm. 44 (59) mm

Thread length. 27 (40) mm. 27 (44) mm. 27 (40) mm. 34 (50) mm. 34 (49) mm. 39 (60) mm. 39 (54) mm. 44 (65) mm. 44 (59) mm Long Distance Cylindrical. Extra long for increased protection and sensing performance triple proximity sensors for flush mounting requirements. designed and tested for extra long life. Ordering Information

More information

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions

RTT2 ROOM TEMPERATURE SWITCH. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA STANDARDS OPERATIONAL DIAGRAMS WIRING AND CONNECTIONS

More information

Operation Manual. Fieldbus system EX510-GDN1. DeviceNet Compatible GW unit

Operation Manual. Fieldbus system EX510-GDN1. DeviceNet Compatible GW unit Fieldbus system DeviceNet Compatible GW unit Operation Manual EX50-GDN URL http://www.smcworld.com Akihabara UDX 5F, --, Sotokanda, Chiyoda-ku, Tokyo 0-00, JAPAN Phone: +8-507-89 Fax: +8-598-56 Note: Specifications

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

E2A. Ordering Information. Cylindrical Proximity Sensor. Safe Mounting with Greater Sensing Distance. E2A Cylindrical Proximity Sensor 1

E2A. Ordering Information. Cylindrical Proximity Sensor. Safe Mounting with Greater Sensing Distance. E2A Cylindrical Proximity Sensor 1 Cylindrical Proximity Sensor E2A Safe Mounting with Greater Sensing Distance Ensures a sensing distance approximately 1. to 2 times larger than that of any conventional OMRON Sensor. Problems such as the

More information

RXTP ROOM TEMPERATURE

RXTP ROOM TEMPERATURE ROOM TEMPERATURE CONTROLLER WITH PI CONTROL Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4

More information

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11 Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11 2011 Ashcroft Inc. 250 East Main Street, Stratford, CT 06614 USA Tel: 203-378-8281, Fax: 203-385-0402

More information

Thread length (overall length) material

Thread length (overall length) material Cylindrical Proximity Sensor for Mobile Usage Designed and tested to keep your machines moving IP69k tested and certified for highest water resistance e1 type approval (according to automotive directive

More information

DIGITAL INPUT/OUTPUT RELAY INTERFACE BMS-IFDD01E

DIGITAL INPUT/OUTPUT RELAY INTERFACE BMS-IFDD01E INSTALLATION MANUAL DIGITAL INPUT/OUTPUT RELAY INTERFACE BMS-IFDD01E Thank you very much for purchasing this TOSHIBA Digital Input/Output Relay Interface. Please read this manual carefully beforehand for

More information

DIO-1616LN-FIT. Features. Specifications. Ethernet Remote I/O F&eIT N Series Isolated Digital Input/Output Unit DIO-1616LN-FIT 1. Ver.1.

DIO-1616LN-FIT. Features. Specifications. Ethernet Remote I/O F&eIT N Series Isolated Digital Input/Output Unit DIO-1616LN-FIT 1. Ver.1. Ethernet Remote I/O F&eIT N Series Isolated Digital / DIO-1616LN-FIT * Specifications, color and design of the products are subject to change without notice. This product is an isolated digital input and

More information

SiO Controller. SiO Controller. SiO-C Ver onwards SiO2/SiO2PNP Ver onwards SiO3 Ver onwards INSTRUCTION MANUAL. Version 3.

SiO Controller. SiO Controller. SiO-C Ver onwards SiO2/SiO2PNP Ver onwards SiO3 Ver onwards INSTRUCTION MANUAL. Version 3. SiO Controller SiO-C Ver. 2.00 onwards SiO2/SiO2PNP Ver. 1.00 onwards SiO3 Ver. 1.00 onwards INSTRUCTION MANUAL Version 3.2 WARRANTY RANGE Warranty Period 1 year period after date of purchase 1. This product

More information

RТTH DUAL ROOM SWITCH

RТTH DUAL ROOM SWITCH DUAL ROOM SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS PRODUCT DESCRIPTION ARTICLE CODES INTENDED AREA OF USE TECHNICAL DATA

More information

7032 Digital-Analog Multimeter

7032 Digital-Analog Multimeter 7032 Digital-Analog Multimeter OPERATOR S MANUAL CONTENTS: 1. Safety precautions and procedures 1 1.1. Preliminary 1 1.2. During Use 2 1.3. After Use.. 2 2. General Description. 3 3. Preparation for Use..

More information

ELV12 Series. Instruction Manual

ELV12 Series. Instruction Manual Instruction Manual TDK-Lambda BEFORE USING POWER SUPPLY UNIT Be sure to read this instruction manual thoroughly before using this product. Pay attention to all warnings and cautions before using the unit.

More information

RXTH DUAL ROOM SENSOR / SWITCH

RXTH DUAL ROOM SENSOR / SWITCH DUAL ROOM SENSOR / SWITCH FOR TEMPERATURE AND RELATIVE HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE

More information

VM-11 SERIES MODEL VM-11B ACCELERATION INPUT VIBRATION SIGNAL CONDITIONER INSTRUCTION MANUAL. MANUAL No E1.8. Revised in Mar.

VM-11 SERIES MODEL VM-11B ACCELERATION INPUT VIBRATION SIGNAL CONDITIONER INSTRUCTION MANUAL. MANUAL No E1.8. Revised in Mar. MANUAL No. 29351E1.8 Published in Nov.1991 Revised in Mar.2006 VM-11 SERIES MODEL VM-11B ACCELERATION INPUT VIBRATION SIGNAL CONDITIONER INSTRUCTION MANUAL FOREWORD VM-11B For Use in Safety... Thank you

More information

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230 USER MANUAL VIBRATION CONTROL Version 2.2 1 IMPORTANT NOTES Electrical danger within the meaning of this documentation or the warning labels on the product itself respectively means that death, serious

More information

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions DUAL DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL

More information

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAM 5 WIRING AND

More information

MODEL SW6000 & SM6100 CENELEC INSTRUCTIONS

MODEL SW6000 & SM6100 CENELEC INSTRUCTIONS MODEL SW6000 & SM6100 CENELEC INSTRUCTIONS Installation Manual 1180 METRIX Experience Value 8824 Fallbrook Dr. Houston, TX 77064, USA Tel: 1-281-940-1802 After Hours Technical Assistance: 1-713-702-8805

More information

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA Model 206 Operating Instructions Setra Systems, Inc. 159 Swanson Road, Boxborough, MA 01719 800.257.3872 www.setra.com 1.0 General Information Every Model 206 has been tested and calibrated before shipment.

More information

Communication & Networking

Communication & Networking MicroSmart Master Module Capable of Connecting 62 Slaves Compliance with Ver. 2.1 specifi cations Digital and analog slaves can be connected. Configuration and slave monitoring can be done using LED indicators

More information

RS232C/RS485 Data can be easily monitored by LAN

RS232C/RS485 Data can be easily monitored by LAN KS SIGNAL CVERTER (AKS) RSC/RS8 Data can be easily monitored by LAN KS SIGNAL CVERTER FEATURES The connectors are located on the front panel. Easy to connect Easy to operate Can be connected to the LAN

More information

S8VE (60/90/120/180/240-W Models)

S8VE (60/90/120/180/240-W Models) Switch Mode Power Supply S8VE (60/90/120/180/240-W Models) CSM_S8VE_DS_E_1_1 60/90/120/180/240-W Models Improved Versions of Standard-type Power Supplies without Indication Monitor. Safety standards: UL508/60950-1,

More information

User safety and equipment protection guidelines

User safety and equipment protection guidelines Snap-in I/O Module The V200-18-E1 plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. The module offers: 3 analog inputs 16 digital

More information

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-920, IA001723 Nov. 2005 OPERATION MANUAL High Voltage Digitalmeter 149-30A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description Uni-I/O Modules Installation Guide UID-0808R, UID-0808T, UID-1600,UID-0016R, UID-0016T Uni-I/O is a family of Input/Output modules that are compatible with the UniStream control platform. This guide provides

More information

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description Uni-I/O Modules Installation Guide UID-0808THS Uni-I/O is a family of Input/Output modules that are compatible with the UniStream control platform. This guide provides basic installation information for

More information

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-820, IA001704 Mar. 2011 OPERATION MANUAL High Voltage Digitalmeter 149-10A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

Installation Sensing distance Connection Output configuration Operation mode NO Operation mode NC

Installation Sensing distance Connection Output configuration Operation mode NO Operation mode NC Standard (thin shape) Inductive Proximity Sensor Thin shape for space saving surface mounting Direct side wall mounting for bracket-less installation Ordering Information DC 3-wire Models Installation

More information

Conductive Level Controller

Conductive Level Controller Conductive Level Controller 61F-D21T-V1 Ideal for level control for industrial facilities and equipment. Outputs can be set to self-hold at ON or OFF using self-holding circuits. Sensitivity adjustment

More information

PX Series Balances. Quick Start Guide. Please download the user manual from

PX Series Balances. Quick Start Guide. Please download the user manual from Scan QR Code with your WebCam for downloading the user instruction manual. PX Series Balances Quick Start Guide Please download the user manual from www.ohaus.com. EN-1 1. INSTALLATION 1.1 Select the

More information

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION The DFI 7000 Signal Conditioner-Indicator is a complete 4-1/2-digit signal conditioner and indicator housed in a 1/8 DIN case. This unit provides

More information

TPE 1464 Series Pressure Transducer

TPE 1464 Series Pressure Transducer TPE 1464 Series Pressure Transducer Description The KMC TPE 1464 series of pressure transducers incorporate a gauge pressure transmitter featuring low hysteresis, excellent repeatability, and longterm

More information

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description Uni-I/O Wide Modules Installation Guide UID-W1616R, UID-W1616T Uni-I/O Wide is a family of Input/Output modules that are compatible with the UniStream control platform. Wide Modules are 1.5 times as wide

More information

Checker is the sensor. CheckMate is the software that runs on your PC. You use CheckMate to set up Checker mm mm 4.

Checker is the sensor. CheckMate is the software that runs on your PC. You use CheckMate to set up Checker mm mm 4. Quick Reference About Checker Checker is the sensor. CheckMate is the software that runs on your PC. You use CheckMate to set up Checker. Minimum PC Requirements Microsoft Windows 2000 or Windows XP 400MHz

More information

Operation Manual. Fieldbus system EX510-GPR1. PROFIBUS DP Compatible GW unit

Operation Manual. Fieldbus system EX510-GPR1. PROFIBUS DP Compatible GW unit Fieldbus system PROFIBUS DP Compatible GW unit Operation Manual EX50-GPR URL http://www.smcworld.com Akihabara UDX 5F, --, Sotokanda, Chiyoda-ku, Tokyo 0-00, JAPAN Phone: +8 3-507-89 Fax: +8 3-598-536

More information

OPLC Installation Guide

OPLC Installation Guide Samba OPLC SM35-J-R20/SM43-J-R20 SM70-J-R20 SM35-J-T20/SM43-J-T20 SM70-J-T20 OPLC Installation Guide 12 Digital Inputs, include 1 HSC/Shaft-encoder Input, 2 Analog inputs (only when the digital inputs

More information

Weighing Indicator USER S GUIDE. globalindustrial.com. globalindustrial.com. 11 Harbor Park Drive Port Washington, NY 11050

Weighing Indicator USER S GUIDE. globalindustrial.com. globalindustrial.com. 11 Harbor Park Drive Port Washington, NY 11050 Weighing Indicator USER S GUIDE 0713 11 Harbor Park Drive Port Washington, NY 11050 Table of Contents I Main Specifications.............................3 II Display and Keyboard..........................4

More information

Appearance Size (mm) Power supply voltage Model (W H D) 100 to 240 VAC K3SC to 240 VAC

Appearance Size (mm) Power supply voltage Model (W H D) 100 to 240 VAC K3SC to 240 VAC Interface Converter A compact converter that allows communications between RS-C/USB and RS-/8 devices. Ideal for industrial applications. Allows communications between RS-C/USB (Universal Serial Bus) and

More information

MT8050iE series. Installation Instruction (1) (2)

MT8050iE series. Installation Instruction (1) (2) MT8050iE series 3 Installation Instructions Installation Instruction Secure the operator panel in position, using all the fastening holes and the provided brackets and screws: (A) 1 Installation and Startup

More information

*Approved by CSA for non-hazardous locations (Group Safety Publication IEC Third Edition).

*Approved by CSA for non-hazardous locations (Group Safety Publication IEC Third Edition). PowerCore Model MPC-20 Installation Manual *Approved by CSA for non-hazardous locations (Group Safety Publication IEC 61010-1 Third Edition). Products covered in this document comply with European Council

More information

INSTRUCTION MANUAL TRIP CIRCUIT SUPERVISION RELAY GKAD1

INSTRUCTION MANUAL TRIP CIRCUIT SUPERVISION RELAY GKAD1 INSTRUCTION MANUAL TRIP CIRCUIT SUPERVISION RELAY GKAD1 TOSHIBA Corporation 2004 All Rights Reserved. ( Ver. 1.6 ) Safety Precautions Before using this product, please read this chapter carefully. This

More information

EDX-10 Series. Compact Recording System

EDX-10 Series. Compact Recording System 3-49 EDX-10 Series Compact Recording System Compact, lightweight, with a simple configuration, all channels synchronous 20 khz high-speed sampling (For 4 channels) Control Unit EDX-10B A unit controls

More information

OPERATOR S MANUAL 9000U ISOLATOR

OPERATOR S MANUAL 9000U ISOLATOR OPERATOR S MANUAL 9000U ISOLATOR Masibus Automation And Instrumentation Pvt. Ltd. B/30, GIDC Electronics Estate, Sector-25, Gandhinagar-382044, Gujarat, India Ph: 91 79 23287275-79 Fax: 91 79 23287281,

More information

TOXALERT MODEL AIR 2000

TOXALERT MODEL AIR 2000 TOXALERT MODEL AIR 2000 NOTE: Toxalert s Model GVU-CO 2 Sensor is the same as the Air2000R. Microprocessor-based, Infrared Environmental CO 2 Sensor OPERATOR S MANUAL TOXALERT TM INTERNATIONAL INC. P.O.

More information

CUS150M1 Instruction Manual

CUS150M1 Instruction Manual BEFORE USING THE POWER SUPPLY UNIT Be sure to read this instruction manual thoroughly before using this product. Pay attention to all cautions and warnings before using this product. Incorrect usage could

More information

V E5B Snap-in I/O Module

V E5B Snap-in I/O Module V200-18-E5B Snap-in I/O Module The V200-18-E5B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 18 isolated digital

More information

V E1B Snap-in I/O Module

V E1B Snap-in I/O Module V200-18-E1B Snap-in I/O Module The V200-18-E1B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 16 isolated digital

More information

INDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1

INDICATOR FOR PROCESS CONTROL. Love Controls. LCI108J-0x INSTRUCTION MANUAL. EDITION: January Valid for models with software version P-1 Love Controls INDICATOR FOR PROCESS CONTROL LCI108-0x LCI108J-0x INSTRUCTION MANUAL EDITION: January 2001 Valid for models with software version P-1 INTRODUCTION TO THE LOVE CONTROLS LCI108 SERIES This

More information

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS GB Contents Page 1 Introduction 4 2 Safety 5 3 Use 6 4 Checking on Delivered Equipment 6 5 General Specification and Dimensions 7 6 Positioning 10 7 Operating

More information

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description Uni-I/O Modules Installation Guide UIA-0402N Uni-I/O is a family of Input/Output modules that are compatible with the UniStream control platform. This guide provides basic installation information for

More information

EPWR Series. RoHS. Product Description. Available Products. Installation Instructions 01

EPWR Series. RoHS. Product Description. Available Products. Installation Instructions 01 01 Series RoHS Compliant Product Description The Series remote pressure transducers are designed for differential pressure applications. The sensors are remotely installed on existing plumbing runs. The

More information

DTY-EPS DTY-EPU. w e DTY-EPP- - w e. t y. q DTY-EPU PSU-BR DTY-EPP-CC- L DTY-EPS. ZD1 ZD4 Tr1 Tr4

DTY-EPS DTY-EPU. w e DTY-EPP- - w e. t y. q DTY-EPU PSU-BR DTY-EPP-CC- L DTY-EPS. ZD1 ZD4 Tr1 Tr4 DTY-EPU....... DTY-EPS. q DTY-EPU.... DTY-EPS DTY-EPP-- PSU-BR q w ew r DTY-EPP-CC-L D ZDZD TrTr q r w e t y q w e r t y.. w e r 0-00 -- - -- 00--00 t URL http://www.koganei.co.jp y DTY-EPU Thank you very

More information

G3PC. Model Number Structure. Solid State Relays with Failure Detection Function. Model Number Legend

G3PC. Model Number Structure. Solid State Relays with Failure Detection Function. Model Number Legend Solid State Relays with Failure Detection Function G3PC Refer to Warranty and Application Considerations (page 1), Safety Precautions (page 4), and Technical and Safety Information (page 6). Detects failures

More information

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions DIFFERENTIAL PRESSURE Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 5 OPERATIONAL

More information

Technical data. Your advantages

Technical data. Your advantages Technical data Nominal torque: 10.000 Nm to 25.000 Nm, bidirectional Customer-specific calibration upon request Rotational speed: 8.000 rpm Accuracy: ±0,5 % Temperature range: -30 C to +85 C Protection

More information

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES Load Cell and Weighing Indicators A Precision Measurement Instrument with Outstanding Features TRACKER 240 SERIES INDICATORS Ratiometric Measurement Tare and Auto Transducer Excitation

More information

RAD-DO8-IFS. I/O extension module, eight digital transistor outputs. Data sheet. 1 Description

RAD-DO8-IFS. I/O extension module, eight digital transistor outputs. Data sheet. 1 Description I/O extension module, eight digital transistor outputs Data sheet 105364_en_00 PHOENIX CONTACT 2013-03-26 1 Description The I/O extension module can be used in conjunction with Radioline wireless modules

More information

NTP-5521/5531/5561 SWITCHING MODE POWER SUPPLY

NTP-5521/5531/5561 SWITCHING MODE POWER SUPPLY NTP-5521/5531/5561 SWITCHING MODE POWER SUPPLY USER MANUAL Keep this manual in a safe place for quick reference at all times. This manual contains important safety and operation instructions for correct

More information

PCD-400A/430A. Sensor Interface. Carrier Wave Type Compact and Moderate Prices. Data Recorders/Analyzers MEASURING INSTRUMENTS

PCD-400A/430A. Sensor Interface. Carrier Wave Type Compact and Moderate Prices. Data Recorders/Analyzers MEASURING INSTRUMENTS 3-77 Sensor Interface 3-77 PCD-400A Carrier Wave Type Compact and Moderate Prices Easy sensor connection using various types of input adapters Connects to PC using USB interface Measurement using 1 unit

More information

F1000 User's Manual. (Version: V1.01)

F1000 User's Manual. (Version: V1.01) (Version: V1.01) Contents Chapter 1 Overview... 2 Chapter 2 Installation... 3 2.1 Installation guide... 3 2.1.1 Installation position... 3 2.1.2 NEMA4 standard installation... 3 2.1.3 Environment precautions...

More information

NPN output PNP output Cord connection Cord length: 2 m Through-beam 7 m Light-ON. 2 m. (selectable by wiring) Diffuse reflective 0.

NPN output PNP output Cord connection Cord length: 2 m Through-beam 7 m Light-ON. 2 m. (selectable by wiring) Diffuse reflective 0. Built-in Amplifier Photoelectric Sensor E3V3 Easy-to-use, Low-cost Photoelectric Sensor Incorporating s that can be clearly seen from a distance. A series of models with an M8 metal junction connector

More information

R1M-GH BEFORE USE... POINTS OF CAUTION INSTRUCTION MANUAL PC RECORDER MODEL. (thermocouple or DC input, 16 points)

R1M-GH BEFORE USE... POINTS OF CAUTION INSTRUCTION MANUAL PC RECORDER MODEL. (thermocouple or DC input, 16 points) INSTRUCTION MANUAL PC RECORDER (thermocouple or DC input, points) MODEL BEFORE USE... Thank you for choosing M-System. Before use, please check contents of the package you received as outlined below. If

More information

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description Uni-I/O Wide Modules Installation Guide UIS-WCB1 Uni-I/O Wide is a family of Input/Output modules that are compatible with the UniStream control platform. Wide Modules are 1.5 times as wide as Uni-I/O

More information

IO-DI8-TO8, IO-DI8-TO8-L I/O Expansion Modules 8 Inputs, 8 Outputs

IO-DI8-TO8, IO-DI8-TO8-L I/O Expansion Modules 8 Inputs, 8 Outputs IO-DI8-TO8, IO-DI8-TO8-L I/O Expansion Modules 8 Inputs, 8 Outputs The IO-DI8-TO8 and IO-DI8-TO8-L are I/O expansion modules that can be used in conjunction with specific Unitronics OPLC controllers. The

More information

Instruction book IQAN-LSL. Publ no HY /UK Edition 0301

Instruction book IQAN-LSL. Publ no HY /UK Edition 0301 Instruction book IQAN-LSL Publ no HY17-8367/UK Edition 0301 Contents 1 Introduction......................................................2 2 Precautions.......................................................3

More information

DC I/O Conversion Module FA-TB32XY

DC I/O Conversion Module FA-TB32XY DC I/O Conversion Module FA-TB32XY User's Manual Thank you for purchasing FA Goods product. Before using, please read this User s Manual and the relevant manuals carefully to ensure correct use. SAFETY

More information

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature

More information

EDX-10 Series. Compact Recording System

EDX-10 Series. Compact Recording System 3-51 EDX-10 Series Compact Recording System Compact & lightweight, with a simple configuration, all channels synchronous 20 khz high-speed sampling (For 4 channels) Control Unit EDX-10B A unit controls

More information

Zenith DS9000 Dispensing System

Zenith DS9000 Dispensing System Page Date: 04/2009 Zenith DS9000 Dispensing System Installation & Operation Manual Page 2 Table of Content Introduction... 3 Control Specification... 4 Wiring Diagram... 5 Wiring Instructions... 6 Field

More information

SORDS ELECTRIC ~ MA7200. Sensorless Vector AC Inverter.

SORDS ELECTRIC ~ MA7200. Sensorless Vector AC Inverter. www.sordselectric.com MA7200 Sensorless Vector AC Inverter Features and Benefits Sensorless Vector The MA7200 has precise speed and torque control for the most demanding system performance and simple set-up

More information

Model Number Structure

Model Number Structure Solid State Relays with Failure Detection Function G3PC CSM_G3PC_DS_E_1_1 Refer to Safety Precautions for All Solid State Relays. Detects failures in SSR used for heater temperature control and simultaneously

More information

Installation, Operation and Maintenance Manual

Installation, Operation and Maintenance Manual Document 481200 VGD-100 Vari-Green Drive Installation, Operation and Maintenance Manual Please read and save these instructions for future reference. Read carefully before attempting to assemble, install,

More information

GENERAL PRECAUTIONS GENERAL DESCRIPTION... 6

GENERAL PRECAUTIONS GENERAL DESCRIPTION... 6 Cat.No I173E-EN-01 RX Inverter Expansion I/O Board 3G3AX-EIO21-ROE USER S MANUAL GENERAL PRECAUTIONS... 3 1 GENERAL DESCRIPTION... 6 2 INSTALLATION PROCEDURE... 7 2.1 INSTALLING THE EXPANSION I/O BOARD...

More information

Description. Green. Red. Orange

Description. Green. Red. Orange GP3000 Series DIO Interface Installation Guide Caution Be sure to read the Warning/Caution Information on the attached sheet before using the product. CAUTION This manual describes the part names and general

More information

ACT-1B Series Panel Tachometer

ACT-1B Series Panel Tachometer MONARCH INSTRUMENT Instruction Manual ACT-1B Series Panel Tachometer Printed in the U.S.A. Copyright 2009 Monarch Instrument, all rights reserved 1071-4843-111R 0909 15 Columbia Drive Amherst, NH 03031

More information

Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V C30B and V T40B.

Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V C30B and V T40B. Vision OPLC Installation Guide V290 (Color) This guide provides basic information for Unitronics LCD color touchscreen models V290-19-C30B and V290-19-T40B. General Description Vision OPLCs are programmable

More information

V E62B Snap-in I/O Module

V E62B Snap-in I/O Module V200-18-E62B Snap-in I/O Module The V200-18-E62B plugs directly into the back of compatible Unitronics OPLCs, creating a selfcontained PLC unit with a local I/O configuration. Features 30 isolated digital

More information

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series TEMPSEN DEVICES Plot No : 2&3, Balaji Nagar, 4 th Street, Mettukuppam, Thoraipakkam, Chennai-600097 Tele fax : +91-44-24581758,Mobil

More information

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3

INSTALLATION MANUAL. LC 200 Electronic Overload Guard. Software versione PW0501 R 0.3 INSTALLATION MANUAL LC 200 Electronic Overload Guard Software versione PW0501 R 0.3 CONTENTS MAIN FEATURES LC 200 TECHNICAL FEATURES Page 2 SYMBOLS Page 3 WARNINGS Page 3 IDENTIFICATION DATA PLATE Page

More information

Analog Input Installation Manual

Analog Input Installation Manual Analog Input Installation Manual August 2011 Part Number: 144-23917 Copyright 2011 Magnetek 1. Preface and Safety Magnetek manufactures products used as components in a wide variety of industrial systems

More information