Nikon A1-B Confocal Operating Manual. Start-up. Microscope
|
|
- Maud Evans
- 6 years ago
- Views:
Transcription
1 Nikon A1-B Confocal Operating Manual Start-up 1. Turn on Excite 120 LED power supply. a. No need for cool down as for mercury bulb b. Open shutter by pushing down contol knob. c. Adjust intensity by turning control knob. 2. Turn on power strip to turn on a. Transmitted light b. Stage controls c. Microscope-wait for microscope to initialize. 3. Turn on Nikon A1 controller. This is the control box for the confocal. Wait for the top light to stop blinking and turn green. 4. Turn on the laser control module by turning the Laser Power key to ON. 5. Push the button for each laser line that you would like to use. Buttons illiminate. Deselect any laser lines that you do not want to use. 6. Turn on the computer and monitor. 7. Login using your uniquename and level 1 Kerberos password. 8. Activate the NIS-Elements-confocal software. (Note-microscope must be initialized before activating NIS-Elements confocal.) a. Select desired screen lighting dark or light. b. Arrange controls panels as desired. 1. Ti Pad for microscope controls 2. A1plus Compact GUI (graphic user interface) 3. A1plus Scan Area c. If any for the control panels are missing, you can find them by right mouse clicking on the screen>acquisitions Control>A1plus Compact GUI or A1plus Scan Area or TiPad. Microscope 1. Tilt back tower for easy access to the stage. 2. Objective lenses a. Three ways of turning 1) By hand 2) With toggle switch on left of scope base 3) With software Ti Pad. b. 10X dry c. 20X dry d. 40X oil immersion e. 60X oil immersion
2 f. 60X water immersion g. (40X water immersion available by special request.) 3. Mount slide. (Move only left-hand slider on stage insert.) 4. Filter cubes a. Rotate with toggle switch on right of scope base or use TiPad. 1) DAPI 2) FITC/GFP 3) TxRed 4) Analyzer (for DIC) 5) A1 (for confocal) b. See readout on front of scope. 1) Display-arrow up or down to select desired readout. Tells which filter cube is in. 2) Brightness button for display c. Open fluorescent light source using push knob on control module. Adjust intensity by turning knob. d. Shutter button on right side of microscope base is non-functional. 5. Focus three ways to focus a. Knob on microscope base--toggle to coarse, fine, or extra-fine b. Knob on stage control-- Toggle to coarse, fine, or extra-fine c. In software using mouse wheel 6. Find region of interest using stage control joystick. Twist joystick to select movement as coarse, fine, or extra-fine. 7. Focus 8. Push Reset button on front of microscope. This sets the Z reading to 0, the working position. 9. To change to load position, push Escape on right side of scope. Push up button to return to the working position. 10. Transmitted light a. Turn on halogen light with software-tipad. 1) Click on DIA rectangle. Rectangle turns from green to red. 2) Click on yellow dot. 3) Use slider to adjust light level.
3
4
5 d. Click on range indicator dropdown list and select color for offset or black pixels. For example, you can choose blue to indicate black pixels. 13. Optimize light levels in each channel. a. Turn off all but one channel by unchecking them in GUI. b. Press Scan. c. Focus to find the channel s brightest focal plane. To focus, hover over the image and turn the mouse wheel. d. Hover mouse over HV control and turn wheel to adjust intensity up or down. e. Hover mouse over offset control and turn wheel to bring in black levels. f. Turn on next channel and turn off previous channel. g. Repeat light level adjustments for each channel. 14. Turn off range indicator. a. In top toolbar, turn off range indicator. b. Change back from grayscale to color by clicking below the images, bottom left, right mouse click on one of the channels>all color. 15. Select number of integrations. Click on integration icon to remove scan from Normal and change to Integrated mode. Click on the dropdown list to select number of scans, e.g., 4 or Capture image. Press on this icon to capture final image. Saving a single plane image 1. First, save as an ND2 Image File Format. This saves the header information needed for analysis software. a. File>Save As b. Establish path User folder>new folder titled with your unique name>new subfolder titled with today s date. c. Select file type ND2 Image File Format( *nd2) d. Type in file name e. Save all channels. (Or save selected channels if desired.) f. Save 2. Saving individual channels in Tif format. ND2 documents cannot be opened in Adobe Photoshop. This procedure automatically converts the images from 12 bit to 8 bit. a. File>Import Export>Export ND Document to tiff. b. Make sure correct path is chosen (Browse). Make Tif folder if desired. c. Select file type Tif. d. Click on Color Image for Each Channel. e. Check LUT box. f. Check convert 12 bit to 8 g. Click on Export. h. Software automatically adds a channel suffix to each image.
6 3. Saving merged image in Tif format. a. File>Import Export>ND Export to Tiff. b. Make sure correct path is chosen (Browse). Make Tif folder if desired. c. Select file type Tif. d. Click on Merged color image. e. Check LUT box. f. Check convert 12 bit to 8 bit. g. Click on Export. 4. Another way to save image as a Tif with slightly less resolution - a. Press x on the computer keyboard. This takes a snapshot of the display and automatically converts it to 8 bit. b. File>Save As c. Make sure correct path is selected (Browse). d. Select Tif format. e. Type in File name. f. Save 5. Saving display of all channels plus merged together in one panel. a. Set up display as desired with all channels plus merged image (e.g.four squares). b. Take a snapshot and save as for items 4 above. Adding a Scale Bar 1. On right side of image select scale bar icon. 2. With dropdown arrow, select properties. 3. Select desired line width, color, and font. 4. Select Burn into image or not. Setting up for a Z-Stack 1. Find ND acquisition panel. If it not already on desktop, right mouse click on desktop>acquisition Controls>ND Acquisition. 2. Deselect Save to File. (Otherwise, this would be a redundant copy of your Z-stack.) 3. Deselect time (or other) acquisition modes, select Z. 4. Define upper and then lower boundaries of the Z-stack. a. Change from Kalman to Normal scan mode. b. Press Scan.
7
8
9 7. Raise objective turret to working position. 8. Check MIL scheduling system to see if someone will be coming onto the scope after you. 9. Log off of the computer. 10. If you are the last user of the day, shut down in reverse order of the start-up. 6) and 5) Shut down computer and turn off monitor. 4) Leave laser line buttons illuminated. Turn off Laser Power key 3) Turn off Nikon A1 Controller. 2) Turn off Power Strip #2. 1) Turn off LED 120 power supply.
10
Multi-Photon Training
Multi-Photon Training Overview This training will take approximately 4 hours. The first 2 hours will be spent learning one-on-one with your trainer, while the following 2 hours will be your opportunity
More informationOlympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup F Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for
More informationStart-up system. Nikon Wide Field Microscope Ee1078. Complete manual
Start-up system Switch on the power socket located on the floor Switch devices on in the order of: o Mercury lamp control-unit switch on and ignite o Halogen lamp o Incubator control-box o Microscope stand
More informationOlympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused
More informationBasic Users Guide for the LSM 510 Confocal Microscope
Basic Users Guide for the LSM 510 Confocal Microscope Ian Jones and Adam Westmacott Department of Biology & Biochemistry, University of Bath, Bath, UK Updated 01 October 2003 This guide describes the basic
More informationCell Imaging Unit UIC Bárbara Fekete ZEISS AXIOIMAGER MICROSCOPE MANUAL
Cell Imaging Unit UIC Bárbara Fekete ZEISS AXIOIMAGER MICROSCOPE MANUAL 1 TABLE OF CONTENTS ANATOMY OF THE ZEISS AXIOIMAGE MICROSCOPE 3 INITIALISATION PROCEDURE 4 LEFT SIDE OF MICROSCOPE (DETAILED VIEW)
More informationZeiss AxioImager.Z2 Fluorescence Protocol
Zeiss AxioImager.Z2 Fluorescence Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationIndiana Center for Biological Microscopy. BioRad MRC 1024 MP Confocal & Multi-Photon Microscope
Indiana Center for Biological Microscopy BioRad MRC 1024 MP Confocal & Multi-Photon Microscope Microscope and the Attached Accessories A: B: C: D: E: F: G: H: Mercury Lamp Transmission Light Kr/Ar Laser
More informationManual ANGSTROM grid confocal microscope
Manual ANGSTROM grid confocal microscope Switching on the system 1) Start by switching on the different components of the microscope system. When using live cell imaging also start the temperature and
More informationOlympus DeltaVision Microscope Start-Up and Shut-Down Instructions.
DeltaVision Instructions: 1. Start-up (Olympus DV) 2. Basic operation (Olympus DV) 2.1 set-up 2.2 designing and running an experiment 3. Transferring files via FTP 4. Deconvolving files 5. File conversion
More informationZeiss Elyra Microscope Standard Operation Protocol Basic Operation
Machine Startup Zeiss Elyra Microscope Standard Operation Protocol Basic Operation 1. Turn on the Incubator power to warm up the incubator for a couple of hours prior to sample acquisition as image could
More informationConfocal 6 Guide. Switching On Components of the system should be switched on in the following order: 1. Lasers Argon laser (488nm and 514nm)
Confocal 6 Guide The system is kept at 37⁰C. If you require the system at a lower temperature book at least one hour before the beginning of your session to allow the system to cool down. It is your responsibility
More informationConfocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by
Confocal Microscope - Radiance Basic Instructions (Stefanie Reichelt) The Radiance Confocal Scanning Microscope designed and developed by Brad Amos (MRC-LMB) Commercialised by Bio-Rad Inc. (1998 2004)
More informationOlympus BX51 on Aura. Introduction to the NRI-MCDB Microscopy Facility BX51 Upright Microscope
Olympus BX51 on Aura Introduction to the NRI-MCDB Microscopy Facility BX51 Upright Microscope Contents Start-up Preparing for Imaging Part I General Part II Transmitted Brightfield Part III Transmitted
More informationSpinning Disk Protocol
Spinning Disk Protocol 1) System Startup F Please note our sign-up policy You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time
More information1. Getting Started: Brief Step-By-Step Guide (PDF File)
Page 1 of 5 1. Getting Started: Brief Step-By-Step Guide (PDF File) Leica SP2 confocal microscope is controlled via software LCS (Leica Confocal Software). The latest version is V2.5. Bellowing is a brief
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated
More informationPCIC Plant Cell Imaging Center (BTI, room 104B) CONFOCAL LEICA SP5 user manual
PCIC Plant Cell Imaging Center (BTI, room 104B) CONFOCAL LEICA SP5 user manual -Turn on the mercury lamp if you need to see the fluorescence through the eye pieces. Check that its shutter is on the open
More informationDSU Start-Up instructions
DSU Start-Up instructions Always: - start with the 10x objective - properly center the stage around the current objective before changing to another objective - when done, leave the 10x objective in standby
More informationOIC. User Guide to Confocal Microscope: Zeiss LSM Pascal
User Guide to Confocal Microscope: Zeiss LSM Pascal This guide does not cover all the details in documentation that comes with the Zeiss micro-scope. Please do find time to go through the original documentation
More informationLEXT 3D Measuring LASER Microscope
LEXT 3D Measuring LASER Microscope Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. A) STARTUP 1. Computer
More informationWide Guy: Inverted Widefield Microscope
Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3
More informationUse of the Binocular Microscope
Use of the Binocular Microscope Before you begin this learning module be sure that you have the following materials in front of you: A microscope A packet of lens paper and Kimwipes A glass slide of a
More informationLSM510 Confocal Microscope Standard Operation Protocol Basic Operation
LSM510 Confocal Microscope Standard Operation Protocol Basic Operation Please make sure that the COMPRESSED AIR has been TURNED ON prior to the use of the equipment. Kindly inform the administrator if
More informationLSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes
LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4.2 January 2007 Contents Page Starting the System...3 Setting the microscope...6 Configuring the beam path
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationOlympus cellsens TIRF Guide
Startup Power up the system as described in the "Startup & Environmental Control" guide, including power strip #2 ("TIRF"). ON For each laser line needed, turn the key on the control unit right and press
More informationIMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1
IMAGE STUDIO LITE Tutorial Guide Featuring Image Studio Analysis Software Version 3.1 Notice The information contained in this document is subject to change without notice. LI-COR MAKES NO WARRANTY OF
More informationZeiss LSM 510 Meta inverted confocal manual
Power Up Zeiss LSM 510 Meta inverted confocal manual *adapted from the Duke University Light Microscopy Core Facility manual 1. Turn on the mercury lamp (if required) and sign in the log book, recording
More informationANDOR/OLYMPUS SPINNING DISK USER GUIDE
ANDOR/OLYMPUS SPINNING DISK USER GUIDE CONTENTS Overview... 2 Turn off sequence:... 2 Turn on sequence:... 2 Getting started... 2 Multi Dimensional Acquisition (MDA)... 3 FRAP... 3 Focus motors and long-term
More informationTable of Contents. Imaging Nucleic Acid Gels on the LI-COR Odyssey FC Imager User Guide University of Puget Sound Updated June 2014 by Amy Replogle
Imaging Nucleic Acid Gels on the LI-COR Odyssey FC Imager User Guide University of Puget Sound Updated June 2014 by Amy Replogle Table of Contents I. Disclaimer!... 2 II. Introduction... 2 III. Safety
More informationEOS MOVIE Utility Ver. 1.5
ENGLISH Software for Playing, Merging and Movies EOS MOVIE Utility Ver. 1.5 Content of this Instruction Manual 1D C 1D X Mk II 1D X 5DS/5DS R 5D Mk III 6D 7D Mk II 70D EMU is used for EOS MOVIE Utility.
More informationECLIPSE LV Series Support Tools
M374E 07.1.NF.2 (3/3) ECLIPSE LV Series Support Tools (Setup software for ECLIPSE LV series microscopes) Software Manual Introduction Thank you for purchasing the Nikon products. This manual describes
More informationQuick & Simple Imaging. User Guide
Quick & Simple Imaging User Guide The Quick & Simple Imaging software package provides the user with a quick and simple way to search and find their documents, then view, print, add notes, or even e- mail
More informationApplication for Inverted Research Microscope ECLIPSE Ti2 Series Ti2 Control Ver Instruction Manual (for Android)
M705E 17.6.Nx.5 Application for Inverted Research Microscope ECLIPSE Ti2 Series Ti2 Control Ver.1.1.0 Instruction Manual (for Android) Introduction Introduction Thank you for purchasing a Nikon product.
More informationTable of Contents. Important Information... 4 Product Description... 4 Computer Requirements Windows Based PCs Mac OS X Based PCs...
Table of Contents Important Information... 4 Product Description... 4 Computer Requirements... 5 Windows Based PCs... 5 Mac OS X Based PCs... 5 Package Contents... 6 Product Overview... 7 Product Specifications...
More informationSetting up Micro-Magellan for Device control
Setting up Micro-Magellan for Device control From the plug-in menu, select Micro-Magellan and the unpopulated window opens Select Configure device control The window below will open, select devices to
More informationWork with Shapes. Concepts CHAPTER. Concepts, page 3-1 Procedures, page 3-5
3 CHAPTER Revised: November 15, 2011 Concepts, page 3-1, page 3-5 Concepts The Shapes Tool is Versatile, page 3-2 Guidelines for Shapes, page 3-2 Visual Density Transparent, Translucent, or Opaque?, page
More informationPrimary Use. Operating Principle
Primary Use The Leica DVM6 is an optical microscope that has the ability observe samples at a high magnification at a high resolution. The microscope allows users to view their sample with up to a 2350x
More informationPolarization Microscope. Omano OM349P Series. User Guide
Polarization Microscope Omano OM349P Series User Guide Table of Contents Getting Started... 3 Components... 3 Technical Specifications... 5 Setup Instructions... 6 Basic Operation... 7 Changing the Light
More informationDigital Recorder End User Guide. Official UK distribution partner
Digital Recorder End User Guide Official UK distribution partner 2 Installer Information Contents General Operation Mouse Operation Logging into the System Quick Menu PTZ Control 4 5 5 6 Playback Operation
More informationPrism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis
Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in
More informationZeiss Z.1 Lightsheet. Quick-Start Guide
Zeiss Z.1 Lightsheet Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 Chamber and Sample preparation Part 2 Image Acquisition Shut-down Start-up: Turn on the Microscope / Computers Turn
More informationOffice365 at Triton College
Office365 at Triton College Logging in to Office365 The new log in page for Triton email is http://outlook.com/triton.edu At that page, enter your username in this format: firstnamelastname@triton.edu
More informationAll-In-One. digital inverted microscope. High-quality Imaging Has Never Been Easier. evos-ca.com
All-In-One digital inverted microscope High-quality Imaging Has Never Been Easier evos-ca.com Fluorescence Imaging Is Now Easier Than Ever ALL-IN-ONE, digital inverted fluorescence microscope The easiest
More informationVolocity ver (2013) Standard Operation Protocol
Faculty Core Facility Volocity 6.3.0 (2013) SOP A-1 Volocity ver. 6.3.0 (2013) Standard Operation Protocol Faculty Core Facility Volocity 6.3.0 (2013) SOP A-2 A. Content Overview. 3 Start up. 3 Change
More information14. Using Illustrator CC with Other Adobe Applications
14. Using Illustrator CC with Other Adobe Applications Lesson overview In this lesson, you ll learn how to do the following: Place linked and embedded graphics in an Illustrator file. Place multiple images
More informationDigital Recorder End User Guide
Digital Recorder End User Guide metcalfeallen www.metcalfeallen.co.uk - 01884 38222 - admin@metcalfeallen.co.uk Installer Information Metcalfe Allen Ltd Unit 7 Saunders Way Kingsmill Industrial Estate
More informationWorking With Images: Intermediate Photoshop
Working With Images: Intermediate Photoshop Viewing Information in the Layers Palette 1. Choose File > Open and open the Start.psd file in the Lesson01 folder located in the PS_Workshop folder on the desktop.
More informationCreating Interactive PDF Forms
Creating Interactive PDF Forms Using Adobe Acrobat X Pro for the Mac University Information Technology Services Training, Outreach, Learning Technologies and Video Production Copyright 2012 KSU Department
More informationUser s Guide to the LMD Laser Micro-dissection. System
User s Guide to the LMD-6000 Laser Micro-dissection System Page 1 Glen MacDonald October 31, 2018 Start-up Procedure for Leica LMD-6000. 1. Turn on mercury lamp by pressing the rocker switch; a. beige
More information3-D. Here red spheres show the location of gold nanoparticles inside/around a cell nucleus.
3-D The CytoViva 3-D System allows the user can locate objects of interest in a 3-D space. It does this by acquiring multiple Z planes and performing our custom software routines to locate and observe
More informationThe walkthrough is available at /
The walkthrough is available at https://downloads.openmicroscopy.org/presentations/2018/gbi-sydney / Description We will demonstrate a number of features of the OMERO platform using an OMERO server based
More informationM20 Metallurgical Microscope User s Manual. Fein.
M20 Metallurgical Microscope User s Manual info@feinoptic.com M20 Microscope Components Lamp Housing C-Mount Adapter Eyepieces Analyzer Objective Lens Coarse Focus Mechanical Stage Stage X-Y Control Fine
More informationMi50 Inverted Microscope User s Manual. Fein.
Mi50 Inverted Microscope User s Manual Fein info@feinoptic.com Mi50 Microscope Components Siedentopf Trinocular Head Stage Plate Mechanical Stage Locking Screw Hex Wrench Storage Beam Splitter Dust Protector
More informationSamples Carolina sample slides (pollen, algae, ). Clean off oil with lens paper then OpticPad around lens (metal not glass) when done.
Bi/BE 227 Winter 2018 Assignment #1 Widefield and confocal laser scanning microscopy Schedule: Jan 3: Lecture Jan 3-12: Students get trained on how to use scopes, start on assignment Jan 3-17: Carrying
More informationApplication for Inverted Research Microscope ECLIPSE Ti2 Series Ti2Control Ver Instruction Manual (for Android)
M705E 17.3.Nx.4 Application for Inverted Research Microscope ECLIPSE Ti2 Series Ti2Control Ver.1.0.2 Instruction Manual (for Android) Introduction Introduction Thank you for purchasing a Nikon product.
More information2. The flight across the globe will begin and your final destination will be your home.
Google Earth 5.0 The Google Earth interface has a toolbar across the top and a panel down the left side. On the left panel, there are three sections: Search, Places and Layers. 1. In the Search section,
More informationStitchGraph User Guide V1.8
StitchGraph User Guide V1.8 Thanks for buying StitchGraph: the easy way to create stitch layouts for hardanger and other complex embroidery stitch types. StitchGraph is intended to allow you to create
More informationZeiss Efficient Navigation (ZEN) Blue Edition Standard Operation Protocol
Faculty Core Facility ZEN BLUE 2.3 SOP A-1 Zeiss Efficient Navigation (ZEN) Blue Edition Standard Operation Protocol Faculty Core Facility ZEN BLUE 2.3 SOP A-2 A. Content Overview. 3 Start up. 4 Display
More informationGetting Started with ShowcaseChapter1:
Chapter 1 Getting Started with ShowcaseChapter1: In this chapter, you learn the purpose of Autodesk Showcase, about its interface, and how to import geometry and adjust imported geometry. Objectives After
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More informationMulti Time Series Rev
Multi Time Series Rev. 4.0.12 The Multi Time Series program is designed to provide flexible programming of automated time dependent experiments. The basic programming unit is a single Time Series Block
More informationIntelligent Security & Fire Ltd
Copyright 2007-2009 Cortex Global Table of contents: Introduction... 3 1. Login... 3 1.1 Advanced login options... 4 2. Preview... 5 2.1 Main menu... 6 2.2 Camera list... 6 2.3 Main display... 6 2.4 Maps...
More informationVersion 4.0. Quick Start Guide For the Odyssey CLx Infrared Imaging System
Version 4.0 Quick Start Guide For the Odyssey CLx Infrared Imaging System 1 Install the Image Studio Software Important: You must be logged in to the Windows Operating System using an Administrator type
More informationInfinova 2217 Security Management System Client User Guide
Infinova 2217 Security Management System Client User Guide April 2015 Page 1 of 90 Contents 1. Quick Start... 4 1.1 Installing application... 4 1.2 Launching application... 4 1.3 Login... 4 2. Main screen...
More informationGet comfortable using computers
Mouse A computer mouse lets us click buttons, pick options, highlight sections, access files and folders, move around your computer, and more. Think of it as your digital hand for operating a computer.
More informationWork with RSS Feeds. Procedures. Add an RSS Text Object CHAPTER. Procedures, page 7-1
CHAPTER 7 Revised: November 15, 2011 s, page 7-1 s Add an RSS Text Object, page 7-1 Rename an RSS Text Object, page 7-2 Delete or Restore an RSS Text Object, page 7-4 Manipulate an RSS Text Object, page
More informationCopyright Notice. Trademarks
Copyright Notice All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording,
More informationWATER MAZE VIDEO TRACKING
instrumentation and software for research SOF-845 USER S MANUAL DOC-214 Rev. 2.2 Copyright 2016 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095
More informationExpression Design Lab Exercises
Expression Design Lab Exercises Creating Images with Expression Design 2 Beaches Around the World (Part 1: Beaches Around the World Series) Information in this document, including URL and other Internet
More informationFor detailed instructions, click the links below. To ask questions, request features, or report problems, visit feedback.photoshop.com.
Workspace basics For detailed instructions, click the links below. To ask questions, request features, or report problems, visit feedback.photoshop.com. You create and manipulate your documents and files
More informationQuick Start Guide - Contents. Opening Word Locating Big Lottery Fund Templates The Word 2013 Screen... 3
Quick Start Guide - Contents Opening Word... 1 Locating Big Lottery Fund Templates... 2 The Word 2013 Screen... 3 Things You Might Be Looking For... 4 What s New On The Ribbon... 5 The Quick Access Toolbar...
More informationTable of Contents WINDOWS 95
Table of Contents Accessories Active program button Active window Application Back-up Browse Cascade windows Check box Click Clipboard Close button Context menu Control Panel Copy Cursor Cut Default Desktop
More informationCCD Acquisition Function
File Edit View Props Region Open Flatfield Corr. Optics Image Stats Setup Open Background Corr. Reset Camera Image Profile Temperature Close Autoexpose Autofocus Correction Output Dual Shutter From the
More informationINTRODUCTION - VetScan HDmicroscope
INTRODUCTION - VetScan HDmicroscope Usage The microscope is an addition to our HM5 Hematology system in order to perform: Blood smears Cytological or Histological work Parasite studies Urine or fecal samples
More informationezimagex2 User s Guide Version 1.0
ezimagex2 User s Guide Version 1.0 Copyright and Trademark Information The products described in this document are copyrighted works of AVEN, Inc. 2015 AVEN, Inc. 4595 Platt Rd Ann Arbor, MI 48108 All
More informationCapstone Appendix. A guide to your lab computer software
Capstone Appendix A guide to your lab computer software Important Notes Many of the Images will look slightly different from what you will see in lab. This is because each lab setup is different and so
More informationLab 5 Microscopy. Introduction. Carrying the Microscope. Depth of Focus. Parts of the Compound Microscope. Magnification
Lab 5 Microscopy Introduction The microscope is an instrument that contains one or more lenses and is used to view objects that are too small be seen with the unaided eye. A magnifying glass is a simple
More informationCoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationGuidelines for Legible and Readable Text, page 2-1 Visual Density Transparent, Translucent, or Opaque?, page 2-3
CHAPTER 2 Revised: November 15, 2011 Concepts, page 2-1 s, page 2-4 Reference, page 2-25 Concepts Guidelines for Legible and Readable Text, page 2-1 Visual Density Transparent, Translucent, or Opaque?,
More informationThe viewer makes it easy to view and collaborate on virtually any file, including Microsoft Office documents, PDFs, CAD drawings, and image files.
Parts of this functionality will only be available in INTERAXO Pro. Introduction The viewer provides users with the capability to load a wide variety of document types online using a web browser. Documents
More informationSolidWorks 2½D Parts
SolidWorks 2½D Parts IDeATe Laser Micro Part 1b Dave Touretzky and Susan Finger 1. Create a new part In this lab, you ll create a CAD model of the 2 ½ D key fob below to make on the laser cutter. Select
More informationG500 SERIES MICROSCOPE OPERATION MANUAL For use with MFR# G502, G502T, G504, G504T, G505, G505T.
G500 SERIES MICROSCOPE OPERATION MANUAL For use with MFR# G502, G502T, G504, G504T, G505, G505T. United Products & Instruments, Inc. 182 Ridge Road, Suite E Dayton, NJ 08810 USA T. 732 274 1155; F. 732
More informationOnPoint s Guide to MimioStudio 9
1 OnPoint s Guide to MimioStudio 9 Getting started with MimioStudio 9 Mimio Studio 9 Notebook Overview.... 2 MimioStudio 9 Notebook...... 3 MimioStudio 9 ActivityWizard.. 4 MimioStudio 9 Tools Overview......
More information1. Move your mouse to the location you wish text to appear in the document. 2. Click the mouse. The insertion point appears.
Word 2010 Text Basics Introduction Page 1 It is important to know how to perform basic tasks with text when working in a word processing application. In this lesson you will learn the basics of working
More informationVisitron Spinning Disk Unit 1
Startup Procedure and Notes (Last update: Sept. 2014) 6 4 2 5 1 3 7 8 9 1. Switch on the fluorescence illumination for widefield viewing (HXP lamp) 2. Switch on power-bar on the left side of the shelf
More informationTutorials by All Creative Designs. Picasa 5 (3.9) Photo Editing Tutorial. How to download, install and use the Picasa Photo Editor
Tutorials by All Creative Designs www.allcreativedesigns.com.au Picasa 5 (3.9) Photo Editing Tutorial How to download, install and use the Picasa Photo Editor How to web optimize your photos for use on
More informationVision Pointer Tools
Vision Pointer Tools Pointer Tools - Uses Pointer Tools can be used in a variety of ways: during a Vision Demo to annotate on the master station s screen during a Remote Control session to annotate on
More informationWindow Designer. Opening Screen: When you start Window Designer, you will see the Opening Screen. Here you will be choosing from 4 options:
Window Designer Opening Screen: When you start Window Designer, you will see the Opening Screen. Here you will be choosing from 4 options: New Design: Use this option when no pre-built templates are available
More informationJEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM
Page 1 of 7 JEOL CarryScope SEM The JEOL CarryScope is a compact and portable SEM that utilizes a standard tungsten filament. It is to be used for inspecting and measuring samples processed in the NRF
More informationFirst Animated Model Yellow Submarine
Course: 3D Design Title: First Animated Model Yellow Submarine Blender: Version 2.6X Level: Beginning Author; Neal Hirsig (May 2012) First Animated Model Yellow Submarine Most of you are perhaps too young
More informationSolidWorks Intro Part 1b
SolidWorks Intro Part 1b Dave Touretzky and Susan Finger 1. Create a new part We ll create a CAD model of the 2 ½ D key fob below to make on the laser cutter. Select File New Templates IPSpart If the SolidWorks
More informationCoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationWhat is the Deal with Color?
What is the Deal with Color? What is the Deal with Color? Beginning from the beginning Our First Moves Diffuse Object Colors Specular Lighting Transparency Paint on Image Those sliders and things Diffuse
More informationUser Guide. DrawAnywhere.com: User Guide
DrawAnywhere.com: User Guide DrawAnywhere.com is an online diagramming & flow charting application with the look & feel of a desktop application! User Guide http://www.drawanywhere.com August, 2007 Table
More informationQuick Set-up Guide for the Biomedx Configured Olympus CX43
Quick Set-up Guide for the Biomedx Configured Olympus CX43 Page 1 of 15 Your microscope is supplied with an Olympus manual, please refer to that for more complete information on microscope operation. This
More informationVizit Pro User Manual
Vizit Pro User Manual 1 Table of Contents Vizit Pro User Manual... 1 Using Vizit Pro... 3 The Vizit Pro User Interface... 3 Toolbars... 4 File Tab Toolbar... 4 Edit Tab Toolbar... 5 Annotations Tab Toolbar...
More informationPress-Ready Cookbook Page Guidelines
Press-Ready Cookbook Page Guidelines table of contents These instructions are for all pages of your cookbook: Title Page, Special Pages, Table of Contents, Dividers, Recipe Pages, etc. WHAT IS PRESS-READY?
More informationBINOCULAR MICROSCOPES
BINOCULAR MICROSCOPES BOECO LABORATORY BINOCULAR MICROSCOPE MODEL BM-800 Specification BM-800: Optical System: Infinity corrected system, field diameter 22 mm Vieving Head: Bin- or Trinocular compensation
More information