FACCalibur Users Guide
|
|
- Jerome Nicholas Johnson
- 5 years ago
- Views:
Transcription
1 FACCalibur Users Guide FACSCalibur Start Up Procedure If the Instrument is OFF: 1. Check the sheath and waste tanks. Open the fluidics drawer (front of instrument) and check the levels of sheath fluid and waste in reservoirs. Sheath fluid should be filled to 3/4 capacity (marked line) ). Waste should be empty except for ~10ml of bleach. Fill and empty as necessary. 2. Pressurize the sheath tank. Flip the switch between the two fluidic reservoirs towards you (Pressurize). 3. Turn on cytometer. The FACSCalibur must be turned ON before the computer. If you have a problem connecting to the cytometer, then the cytometer was not turned on in time to be recognized by the computer. 4. Make sure Cytometer is in STANDBY mode. Make sure the tube of water is on the SIP, the STANDBY button is lit, and the speed is set to LO. If you are using the Red Laser (APC channel), then you must give the instrument 20 minutes to warm up
2 If the Instrument is ON already already: 1. Make sure the Cytometer is in STANDBY mode. Make sure the tube of water is on the SIP SIP, the STANDBY button is lit, and the speed is set to LO. 2. DePressurize the Sheath Tank Flip the switch between the two fluidic reservoirs away from you (depressurize-vent) (depressurize 3. Fill and Empty the Sheath and Waste Tanks as needed Sheath Fluid should be filled to ¾ capacity (marked line). Waste should be empty except for ~10ml of bleach. 4. Repressurize the Sheath Tank Flip the switch between the two fluidic reservoirs toward you (pressurize) CELLQuest Pro Software Start Up & Sample Acquisition 1. Turn on computer, log in and open CELLQuest Click on the CellQuest Pro Icon in your menu bar. The Following window will pop up:
3 2. Connect to the Cytometer. In the Menu Bar, Choose Acquire: Connect to Cytometer Cytometer. If you get an error message at this time, it means the computer was turned on BEFORE the instrument. You must restart the computer in order for the software to connect to the instrument. If the Following window (Acquisition Control) doesn t pop up when you connect to the cytometer: You can try disconnecting: and reconnecting to the software.
4 You can also control acquisition in the Parameter Description window. It can be found in the Menu Bar under: Acquire: Parameter Description. This is also where you will choose where you save your experiment, name your samples, and label your axes. 3. Open your Acquisition Template. In the Main Menu Bar, Choose File: Open Document... If you have a template saved, look in the DATA drive, under: Data: Your PI: Templates/Instrument Settings Folder **If you are making a new Template, use the toolbar on the left side of the screen to draw histograms and dot plots on the blank template (See New Template Protocol). Be sure to SAVE the Template in the Template/Instrument Settings folder for your lab!!! **If you try to run and don t see any events, or your axes are labeled P1, P2, etc. make sure your PLOT TYPE is ACQUISITION (or ACQUISITION>ANALYSIS). To check this, open: Plots: Format Histogram (or Dot Plot, depending on the type of plot you ve constructed)
5 The following window will pop up. This allows you to change Plot Type from Analysis to Acquistion.. You can also change things like your X parameter (histogram) or X and Y parameters (dot plot), how the plot is gated, etc. 4. Set up your Acquisition and Storage (the number of events to collect etc). Go to Acquire: Acquisition and Storage. (see explanation, next page)
6 a. Make sure your Acquisition Gate is set to Accept All events. b. Make sure your Collection Criteria is set properly. For example, if you would like to save 10,000 events in your live gate, you want to set collection criteria to event count and acquisition will stop when 10,000 of G1=R1 (your primary population gate) events are counted. c. Make sure your Storage Gate data file is set to contain All events and the file Resolution is d. Click on Parameters Saved and the window below on the left will pop up: Be sure you have ALL the proper parameters checked!!! e. Click on OK to exit Acquisition and Storage. 5. Create a folder for your data. a. Open the Parameter Description Window (Acquire: Parameter Description) b. In the Parameter Description window click on the Change button next to the Directory and choose a location in which to save your data files. c. Create a new folder for each experiment by clicking on Data then choosing your lab s folder and your specific folder, once you are there, in the bottom left corner of the window choose Create New Folder. d. Name the folder and Click on Select [your chosen folder name] to select that folder and exit Folder Information. (see next page)
7 e. Double check in the Parameter Description Window, that your Directory (or folder where your samples are being saved) is the proper folder. 6. Change the filename. a. In the Parameter Description window click on the Change button next to the File and choose a name for your data files b. Set the File Name Prefix refix to Sample ID. c. Set the File Name Suffix uffix to File Count and start file count from 001. d. Click on OK to exit File Information.
8 7. Enter the names of your parameters (e.g. Antibody conjugates, fluorescent dyes). 1. Type the name directly into the box for each parameter. P1: FSC (Forward Scatter) P2: SSC (Side Scatter) P3: FL1 (FITC, GFP, Alexa 488) P4: FL2 (PE, Cy3) P5: FL3 (PI, PE-Cy5, PE-Cy7, 7AAD) P6: FL4 (APC) 8. Open the Counters window.. Go to Acquire: Counters. Please note: The Counters gives you a TOTAL event count ONLY (not cell count through gates). 9. Load Instrument Settings. If you would like to use the same Detectors/Amps, Compensation and Threshold settings as in a previous experiment, you can open them by doing the following: Go to Cytometer: Instrument settings. (picture, next page)
9 10. Open AND SET your Instrument Settings Click the OPEN button. You can choose a specific Instrument Settings file or a previously recorded data file from your folder (choosing to use a previously recorded data file will copy the instrument settings from when that file was collected). Remember that when you SET the instrument settings the color wheel will spin, and it might take a few seconds to set! Click on DONE to exit Instrument Settings. 11. Check sample settingss before recording a file. Tick the setup box in the Acquisition Control (or Parameter Description) window. Press Acquire to start viewing sample data. **you cannot record sample data file in setup mode**. Generally, the following steps (12-14) are done in SETUP mode. This way, you can change settings to get optimal signalss before running your experiment. Steps 12 and 13 are usually set using your Unstained Sample, and Step 14 is done using your single stain compensation samples.
10 12. Check Threshold settings. Go to Cytometer: Threshold and the window below will appear. Threshold is used to exclude data below a certain channel (or cells/debris below a certain size). size) Most of the time, threshold hreshold will remain the same. Howev However, er, if you have very small cells, cells or a lot of debris you don t want counted, it helps to lower or rraise the threshold hreshold respectively. Normal settings: Primary Parameter: Secondary Parameter: FSC-H value 52 None 13. Check Detectors/Amps settings settings. Go to Cytometer: Detectors/Amps the window on the next page will appear. Adjust each parameter as required. (see next page for guided instructions)
11 In set up mode (see step 11) run a sample of UNSTAINED cells and adjust the following: MODE: VOLTAGE: FL-4 (P7): Log amplification is often used to analyze samples with a large dynamic range of fluorescence signals. The log scale has four decades of range. Set all fluorescent channels to LOG (unless doing cell cycle analysis) Linear amplification is usually used for all light scatter parameters. It is also useful for fluorescent parameters used in DNA quantitation experiments. Set FSC and SSC to LIN Linear amplifier gain can be adjusted from The chosen value multiplies the signal by a factor of As the voltage is increased, the detector sensitivity increases, resulting in increased signal. As the voltage is decreased, the detector sensitivity decreases, resulting in decreased signal. PMT voltages are adjusted using the Voltage sliders In order to Acquire the FL FL-4 4 channel (most commonly APC) you must check the Four Color Box at the bottom of the Detector/Amps Window. 14. Check Compensation settings settings. Go to Cytometer: Compensation and the window on the next page will appear. appear (see Compensation tutorial for setting compensation on the FACSCalibur). In set up mode (see step 11) run each of your single stained samples, and adjust compensation so that there is minimal spillover of each signal into neighboring channels. (i.e. a FITC only sample doesn t appear to be FITC and PE positive)
12 **If you would like to save the changes your instrument settings in a separate file (instrument settings are the Threshold, Detectors/Amps and Compensation settings) go to Cytometer: Instrument Settings and in the window, click SAVE. Save the file in your Template/Instrument settings folder, and use it in future experiments for STEP 9 (Load Instrumentt Settings.) 15. Record a data file. a. Enter your Sample ID in the Parameter Description window. b. Uncheck the Setup box on the Acquisition Control window. Your first file (name.001) is ready for recording. c. Vortex your sample, double check that there are no aggregates in your sample, and place on SIP. d. Run cytometer on Low and click Acquire on the Acquisition Control window. e. The number of events per second (Counter panel) should not exceed If your cells/events are running faster then remove from the SIP and dilute. If they are running slower then increase you sample speed to Medium or High (buttons above the RUN/STANDBY buttons). f. If you do not reach the required number of events before the sample runs out, click Pause and then Save. g. Repeat steps a-f for each sample. 16. Backup your data at the end of each run. Either Bring a USB drive, or see the Data Transfer Guide for transferring your data to another computer The CIC facility is not equipped for data storage. Data is erased from all CIC computers (including workstations) on a monthly basis. 17. Please see the Data Analysis Protocols for analysis options.
13 FACSCalibur Shut Down Procedure 1. Disconnect the software from the cytometer. After running the last sample, set the cytometer to STANDBY, make all necessary saves in CELLQuest Pro and choose Disconnect from Cytometer in the Acquire menu. 2. Clean flow cell with 10% Contrad for 3 minutes. Place a tube with 3 ml of 10% Contrad on the SIP with the arm in the open (back) position and aspirate for 30 seconds. Then close the arm (forward) and RUN the tube on HIGH for 3 minutes. 3. Rinse flow cell with water for 3 minutes. Place a tube with 3 ml of water on the SIP and aspirate for 30 seconds with the arm in the open position. Then close the arm and RUN on HIGH for another 3 minutes. 4. Put the instrument in STANDBY (turn the instrument off if necessary, see guidelines below) and depressurize the Sheath Tank by flipping the switch in the Fluidics Drawer to Vent- Change Tank. If someone is not on the instrument within 2 hours after your appointment, shut down the instrument If someone is on the instrument within 2 hours after your appointment, keep the instrument turned on (and in STANDBY mode) You can check the reservation schedule by double clicking on the reservation schedule shortcut on your desktop or going to
BD CellQuest Pro Acquisition Tutorial
BD CellQuest Pro Acquisition Tutorial Introduction This tutorial guides you through a CellQuest Pro Acquisition run like the one demonstrated in the CellQuest Pro Acquisition Movie on the BD FACStation
More informationBD CellQuest Pro Software Acquisition Tutorial
BD CellQuest Pro Software Acquisition Tutorial This tutorial guides you through a typical acquisition using BD CellQuest Pro software. If you are already familiar with previous versions of BD CellQuest
More informationCELLQuest Acquisition Tutorial
CELLQuest Acquisition Tutorial Introduction This tutorial guides you through a CELLQuest Acquisition run like the one demonstrated in the CELLQuest Acquisition Movie on the FACStation Overview CD-ROM.
More informationCyAn ADP Guide. Starting Up
CyAn ADP Guide Starting Up 1. Check the sheath and waste fluids, if needed, fill and empty: 1. To fill the Sheath Tank: a. Unscrew the top of the tank labeled Sheath Tank, and set it on top of the tank.
More informationFACSLSRFortessa SORP QUICK REFERENCE GUIDE
FACSLSRFortessa SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username
More informationLSRFortessa Guide. Cell Evaluation and Therapy Core HO324
LSRFortessa Guide Cell Evaluation and Therapy Core HO324 Before Use: Prime the instrument twice with no tube and arm to the side Ensure that sheath is full and waste is empty To Begin: Log into ilabs tracker
More informationUnscrew and remove the cap from the waste tank
Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish
More informationUSER GUIDE TO BECKMAN COULTER FC500
USER GUIDE TO BECKMAN COULTER FC500 Create a New Protocol Purpose: This procedure describes how to define the parameters and create rough analysis plots for a new experimental set-up. Some protocol templates
More informationScience is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. CFlow User Guide TABLE OF CONTENTS 1 INTRODUCTION TO CFLOW... 1 1.1 Installing CFlow... 1 1.2 Starting CFlow... 1 1.3 CFlow Workspace... 2 1.4 Opening a
More informationFACSLyric TM General User instructions
flowkon Flow Cytometry User Instructions at the University of Konstanz FACSLyric TM General User instructions In BD FACSuite software, you can measure and analyze samples using either assays or experiments.
More informationCanto II Operation Protocol
Canto II Operation Protocol Make sure the following actions have been taken before running your samples. Make sure the waste tank is empty and refilled with 500 ml clean (see page 15 for details) Make
More informationFACSAria Fusion SORP QUICK REFERENCE GUIDE
FACSAria Fusion SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Admin Password: BDIS 2. Ensure the flow cell access door is open. You will want to
More informationUNC Flow Cytometry Core Facility LSR II Operation 2018
Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration).
More informationFACScan 4 and 5 Color Upgrade Rainbow Users Guide
FACScan 4 and 5 Color Upgrade Rainbow Users Guide For Research Use Only 1 Table of Contents Overview... 3 Hardware Specifications... 3 Rainbow Software... 4 Rainbow Preferences Window... 5 Setting up the
More informationSetup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function
Setup protocol for Firefly particles On BD LSR II and LSRFortessa cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction
More informationGallios TM Quick Reference
Gallios TM Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic
More informationBD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option
BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option This guide contains instructions for using BD FACSDiva software version 6 with BD FACSCanto and BD FACSCanto II systems
More informationBD FACSMelody Cell Sorter Quick Reference Guide
BD FACSMelody Cell Sorter Quick Reference Guide This guide contains instructions for using the BD FACSMelody cell sorter with BD FACSChorus software version. and later. Workflow Overview The following
More informationSetup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function
Setup protocol for Firefly particles On BD FACSCanto cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction Cytometer requirements
More informationFlow Cytometry & FACS Core Facility
ZMBH - Zentrum für Molekulare Biologie der Universität Heidelberg Flow Cytometry & FACS Core Facility BD FACSCanto TM with HTS User's Guide For authorization to use the FACSCanto II cytometer, you first
More informationBD Influx Standard Operation Protocol - Basic Operation
Optimizing the drop phase BD Influx Standard Operation Protocol - Basic Operation 1. Verify the 2-tube holder with dummy 15ml tube is in place on the sort stage 2. Select 2 tube holder 2 way sort as the
More informationBD Multiwell AutoSampler Additional Features Tutorial
BD Multiwell AutoSampler Additional Features Tutorial Introduction This tutorial provides step-by-step instructions on how to use the additional features available in BD Multiwell Plate Manager (MPM) software
More informationRelease Notes Life Technologies Attune NxT Software v2.2
Release Notes Life Technologies Attune NxT Software v2.2 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationSpherotech, Inc Irma Lee Circle, Lake Forest, IL
New flow cytometers, with fluorescent channels from the UV to Far Red, and corresponding fluorescent conjugates are now available. As a result, we have developed the Ultra Rainbow Fluorescent Particles
More informationStandard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:
Shared Resources: Flow Cytometry and Cell Sorting Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: DHVI BD LSRII (L01/L02) and BD LSRFortessa
More informationFlow Cytometry/FACS Bio-Rad S3e Cell Sorter
Flow Cytometry/FACS Bio-Rad S3e Cell Sorter Contents Brief Description... 3 Reference... 3 Location... 3 Suggestion for Materials and Methods... 4 Suggestion for Acknowledgements... 4 People and Contacts...
More informationThe Danish Stem Cell Center (DanStem) Danish Stem Cell Center Flow Cytometry Core Facility. LSR Fortessa Guide
Danish Stem Cell Center Flow Cytometry Core Facility LSR Fortessa Guide LSR Fortessa Overview The Danish Stem Cell Center (DanStem) BD FACSFlow Supply System Fluidics Cart Sheath container Waste container
More informationStandard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:
Shared Resources: Flow Cytometry and Cell Sorting Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: DHVI BD LSRII (L01/L02) and BD LSRFortessa
More informationRelease Notes Life Technologies Attune NxT Software v2.3
Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationFloMax Software for Cytometry. Operating Manual Instrument Control and Acquisition. FloMax Operating Manual Instrument Control and Acquisition
FloMax Software for Cytometry Operating Manual Instrument Control and Acquisition 1 Contents Introduction to FloMax Software Acquisition and Instrument Control...3 Operating Basics - Preparing FloMax for
More informationRelease Notes Life Technologies Attune NxT Cytometric Software v 2.0.0
Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0 In the following pages you will find instructions describing: 1. Software installation 2. New software features and fixes 3. Known
More informationTutorial: FCAP Array Software with BD FACSArray Bioanalyzer
Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer After completing this tutorial you will be able to: Create an Experiment with the Experiment Wizard in FCAP Array software. Export the Experiment
More informationNovoExpress Software Guide. Software Version 1.2.4
NovoExpress Software Guide Software Version 1.2.4 Table of Content 1. Introduction 6 1.1 Revision History...6 1.2 About the NovoExpress Software...6 1.3 Conventions...6 2. Installation 8 2.1 Installation
More informationApplied Biosystems Attune Cytometric Software version 1.2 Release Notes
Applied Biosystems Attune Cytometric Software version 1.2 Release Notes Customer ( you ) must review these notes prior to installing or operating the Attune Cytometric Software version 1.2 ( Attune Software
More informationA Bright Cytometer Easy to use
BriCyte E6 Flow Cytometer A Bright Cytometer Easy to use BriCyte E6 Flow Cytometer Flow cytometer is an essential tool to analyze cells, so it s used for diagnosis as well as research in laboratories worldwide.
More informationBD CellQuest Pro Analysis Tutorial
BD CellQuest Pro Analysis Tutorial Introduction This tutorial guides you through a CellQuest Pro Analysis run like the one demonstrated in the CellQuest Pro Analysis Movie on the BD FACStation Software
More informationMERGE FILES TUTORIAL. Version 1.4
MERGE FILES TUTORIAL Version 1.4 1. INTRODUCTION The merge process consists on adding two or more sample data files from the same sample or from different samples. The file resulting from the merging will
More informationLEGENDplex Data Analysis Software Version 8 User Guide
LEGENDplex Data Analysis Software Version 8 User Guide Introduction Welcome to the user s guide for Version 8 of the LEGENDplex data analysis software for Windows based computers 1. This tutorial will
More informationLog into the computer using your CU Denver username and password. Log into the DIVA software using your CU and Username and the password flow
Log into the computer using your CU Denver username and password Log into the DIVA software using your CU and Username and the password flow When DIVA opens, IF a CST Mismatch pop up box appears, then
More informationSTANDARD OPERATING PROCEDURE FOR THE LSR II
The Flow Cytometry Facility STANDARD OPERATING PROCEDURE FOR THE LSR II All users must follow the SOP & fill in the logbook. If you are unsure about anything please do not hesitate to ask for help. Before
More informationRelease Notes Thermo Fisher Scientific Attune NxT Software v2.4
Release Notes Thermo Fisher Scientific Attune NxT Software v2.4 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance
More informationCustomer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ).
Life Technologies Attune NxT Software version 1.1.0 Release Notes Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version 1.1.0 ( Attune Software v1.1.0
More informationAttune NxT Acoustic Focusing Cytometer
USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Numbers A24858, A24859, A24860, A24861, A24862, A24863, A24864, A28993 Publication Number 100024235 Revision C.0 For Research Use Only. Not for
More informationQuick Start Guide: COLLECT
Quick Start Guide: COLLECT Ready, Set, Run! The Collect tab is used to collect or import a sample, configure the run settings and set up plots and gates. Collect a Sample 1. Start the C6 Flow Cytometer,
More informationJMJ-Group Microbiology BMB SDU
JMJ-Group Microbiology BMB SDU Title: Daily use of FACS Aria II SOP number: SOPFACS0042_v01 Version number: 01 Date issued: 2013.01.16 Review date: 2014.12.01 Written by: Tina Kronborg and Michelle Madelung
More informationHacking FlowJo VX. 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From
Hacking FlowJo VX 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From Contents 1. Change the default name of your files. 2. Edit your workspace
More informationNorth America/International Japan Europe
North America/International Japan Europe 1730 North First Street San Jose, CA 95112 U.S.A. Voice: +1 800-275-5963 FAX: +1 408-352-4130 sales@sonybiotechnology.com http://www.sonybiotechnology.com 1-7-1,
More informationCyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System
CyFlow ML Healthcare Immunology Pathology Microbiology Cell Biology Multilaser 16 Parameter Desktop FCM System 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements
More informationTechnical Note. Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6. Introduction
Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6 Contents 1 Introduction 2 Before You Begin 3 Using Global FlowJo Preferences 5 Usinig the Transform Button 5 Saving
More informationENCENDIDO Y LAVADOS DEL CITÓMETRO
ENCENDIDO Y LAVADOS DEL CITÓMETRO ANTES DE INICIAR EL PARATO SI EL ORDENADOR ESTÁ ENCENDIDO Y EL CITÓMETRO APAGADO: 1. APAGAR EL ORDENADOR 2. ENCENDER EL CITÓMETRO 3. ENCENDER EL ORDENADOR AL INICIAR EL
More informationMACSQuantify Software 2.11
MACSQuantify Software 2.11 User manual Version 1 Original instructions Copyright 2017 Miltenyi Biotec GmbH and/or its affiliates. All rights reserved. No part of this user manual may be reproduced, stored
More informationSysmex CyFlow Cube Series
Sysmex CyFlow Cube Series Reliable, Compact and Cost-effective Flow Cytometers For Research Use Only. Not for use in diagnostic or therapeutic procedures. Sysmex CyFlow Cube Series Compact, Economic Flow
More informationLSM510 Confocal Microscope Standard Operation Protocol Basic Operation
LSM510 Confocal Microscope Standard Operation Protocol Basic Operation Please make sure that the COMPRESSED AIR has been TURNED ON prior to the use of the equipment. Kindly inform the administrator if
More informationFX500 Exchangeable Fluidics Cell Sorter
FX5 Exchangeable Fluidics Cell Sorter Sony Biotechnology Inc. FX5 Cell Sorter Sorting Made Simple The FX5 is a microfluidics chip based cell sorter featuring a fully replaceable fluidics system. This lets
More informationMA900 Multi-Application Cell Sorter
MA9 Multi-Application Cell Sorter Sony Biotechnology Inc. MA9 Cell Sorter Sorting Made Simple The MA9 from Sony meets the needs of most sorting applications, supporting 12 fluorescence parameters and 4-way
More informationGallios Flow Cytometer
Gallios Flow Cytometer For Research use only, not for use in diagnostic procedures. Quick Reference PN A75199AB.1 (May 2011) Beckman Coulter Inc. PN A75199AB.1 (May, 2011) 1 of 24 WARNINGS AND PRECAUTIONS
More informationINSTALLING THE PS3 XBOX READY SOFTWARE:
INSTALLING THE PS3 XBOX READY SOFTWARE: 1. Insert the Installation CD to CD-ROM drive and execute Ready_Setup.exe NOTE: If it is the first time for the target USB disk using under this software, the software
More informationfor the Analysis of Flow Cytometry Listmode Data Files
Purdue University Cytometry Laboratories Introduction to WinMDI 2.8 (J. Trotter 1993-1998) for the Analysis of Flow Cytometry Listmode Data Files Objectives of this tutorial: - Open a listmode data file
More informationEquipment Overview: Safety:
PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system
More informationData acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and instrument
More informationCytomics FC 500 Quick Reference
Cytomics FC 500 Quick Reference PN 722238 RevA (April 2005) Beckman Coulter Inc. Miami Education Center Miami Lakes, WARNINGS AND PRECAUTIONS READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED
More informationData acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.6 Contents 1. Description 1.1 Background information 1.2 Reagent and
More informationInstructions for Use. COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software. PN CA (September 2010)
READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software Instructions for Use CYTOMETER LASER SAMPLE FLOW LOW MED HIGH (September 2010)
More informationBD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007
BD Biosciences Technical Resources Page 1 For use with the BD Lyoplate Human Cell Surface Marker Screening Panel (Cat. No. 560747). Please check that your catalog numbers for the FCS Express Excel templates
More informationCyFlow Cube 6 Instrument Operating Manual
CyFlow Cube 6 Instrument Operating Manual CyFlow Cube 6 CyFlow Cube 8 - Instrument Operating Manual For in vitro diagnostic use with Partec recommended IVD reagents. IVD The PartecCyFlow Cube Flow Cytometer
More informationSummit Software WorkList Builder Quick Start Guide
Summit Software WorkList Builder Quick Start Guide Document Number 0003185 Revision July 2006 2 Summit WorkList Builder Quick Start Guide Disclaimer Summit Worklist Builder is a Research Use Only software
More informationNanosight NS300 NBTC User Instructions
Nanosight NS300 NBTC User Instructions Preliminary notes: -The laser module is delicate and expensive. Handle it with care and do not leave it on the bench unattended. Only leave it in place in the instrument
More informationFlowJo Software Lecture Outline:
FlowJo Software Lecture Outline: Workspace Basics: 3 major components 1) The Ribbons (toolbar) The availability of buttons here can be customized. *One of the best assets of FlowJo is the help feature*
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Luminex Prima Theory of Operation Luminex 200 technology is based on flow cell
More informationibidi PumpControl Instruction Manual Version 1.5.0
ibidi PumpControl Instruction Manual Version 1.5.0 II ibidi GmbH 2011, version 1.5.0, revision 2, 2011-10-26, Andreas Rainer Table of contents 1 General Information... 1 1.1 System Requirements... 1 1.2
More informationStandard Operating Procedure for the Horiba FluroMax-4
Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.
More informationNovoSampler Pro Operator s Guide
NovoSampler Pro Operator s Guide Table of Content Prologue 5 I. Revision History...5 II. Contact Information...5 III. Trademark...6 IV. Intended Use...6 V. Using this Guide...6 VI. Conventions...6 VII.
More informationPANALYTICAL X PERT PRO XRD
STANDARD OPERATING PROCEDURE: PANALYTICAL X PERT PRO XRD Purpose of this Instrument: Identification of signal-phase materials and multi-phase mixture; Quantitative determination of amounts of different
More informationHomework Assignment 9 LabVIEW tutorial
Homework Assignment 9 LabVIEW tutorial Due date: Wednesday, December 8 (midnight) For this homework assignment, you will complete a tutorial on the LabVIEW data acquistion software. This can be done on
More informationSPA III Quick Reference Guide for Custom Panels
SPA III Quick Reference Guide for Custom Panels This guide contains instructions for using the BD FACS Sample Prep Assistant III (SPA III) with BD FACS SPA software version.0 and later. Initial Setup Custom
More informationData acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.
Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and
More informationChromQuest 5.0 Quick Reference Guide
ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User
More informationBD FACSDiVa Option. BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen
BD FACSDiVa Option The Digital Advantage for the BD FACSVantage SE Flow Cytometry System BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen Proven Performance in Flow Cytometry
More informationGetting Started. COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software. PN BA (October 2010)
READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software Getting Started CYTOMETER LASER SAMPLE FLOW LOW MED HIGH (October 2010)
More informationQuick Guide NC-200. NucleoCounter. to the easiest cell count ever!
Quick Guide to the easiest cell count ever! NucleoCounter NC-00 How to get started Dear NucleoCounter Customer! Thank you for purchasing the NucleoCounter NC-00 offering not only the easiest one-step viability
More informationBD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual
BD CellQuest Pro Software Reference Manual 643274 BD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual bdbiosciences.com Part No. 643274 Rev. A November 2007 BD Biosciences
More informationConfocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by
Confocal Microscope - Radiance Basic Instructions (Stefanie Reichelt) The Radiance Confocal Scanning Microscope designed and developed by Brad Amos (MRC-LMB) Commercialised by Bio-Rad Inc. (1998 2004)
More informationPerkinElmer Series 200 HPLC: Operating Procedure
PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering
More informationSpecial Procedures and Troubleshooting
READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software Special Procedures and Troubleshooting CYTOMETER LASER SAMPLE FLOW LOW MED HIGH
More informationLeica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b
Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b User Guide Biomedicum Imaging Unit (BIU) University of Helsinki www.biu.helsinki.fi 11.11.2016 1 GENERAL...1
More informationBD Accuri C6 with single tube loader protocol. For multiplex assays
BD Accuri C6 with single tube loader protocol For multiplex assays BD Accuri C6 with single tube loader protocol This protocol contains instructions for setting up an Accuri C6 with Tube Loader. If you
More informationNear UV Laser System Performance on the CytoFLEX S
Near UV Laser System Performance on the CytoFLEX S APPLICATION NOTE Authors: Jeffrey Cobb James Tung Affiliation: Life Science, Beckman Coulter, Inc., Miami, FL, United States Introduction Material and
More informationCH142 Spring Spectrophotometers with Vernier Data Acquisition Software
Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample
More informationChromatography Software Training Materials. Contents
Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will
More informationCytometry Data Analysis in FlowJo V10. Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC
Cytometry Data Analysis in FlowJo V10 Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC timc@flowjo.com Outline Part I Intro to FlowJo Navigating the V10 Workspace Customizing Ribbons Demo
More informationDIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only
DIAplex DIAplex Pro 1.0 Software Instructions for use For research use only Fast Track Your Research Table of Contents 1. Introduction... 2 2. DIAplex Pro 1.0 installation... 2 2.1. System Requirements...
More informationWindows 2000 Professional
The American University in Cairo Academic Computing Services Windows 2000 Professional prepared by Soumaia Ahmed Al Ayyat 4 August 2003 Table of Contents Starting Up the Computer Windows Environment Start
More informationMulti-Photon Training
Multi-Photon Training Overview This training will take approximately 4 hours. The first 2 hours will be spent learning one-on-one with your trainer, while the following 2 hours will be your opportunity
More informationSTAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1)
STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) Aim To provide detailed description of all steps, material and methods used for fluorescent surface staining and preparation of whole blood cells for analysis
More informationTS MANAGER OPERATIONS MANUAL
TS MANAGER OPERATIONS MANUAL Super Systems Inc. 7205 Edington Drive Cincinnati, OH 45249 513-772-0060 800-666-4330 Fax: 513-772-9466 www.supersystems.com Table of Contents Introduction... 3 Prerequisites...
More informationBasic Tutorial. Data Analysis Software for Flow Cytometry Windows/Mac user documentation
Basic Tutorial Data Analysis Software for Flow Cytometry Windows/Mac user documentation FlowJo, its tutorials, documentation and web site are copyright FlowJo, LLC 2018. All Rights Reserved. Version 2.0
More informationUser Manual. For Windows operating systems Excelsior Blvd. Suite Pécs St. Louis Park, MN Kedves street 20
User Manual For Windows operating systems Soft Flow, Inc. Soft Flow Hungary Ltd. 5009 Excelsior Blvd. Suite 154 7628 Pécs St. Louis Park, MN 55416 Kedves street 20 USA Hungary Tel: +1 (952) 681-7960 Tel:
More informationIntroduction to IBM Rational HATS For IBM System i (5250)
Introduction to IBM Rational HATS For IBM System i (5250) Introduction to IBM Rational HATS 1 Lab instructions This lab teaches you how to use IBM Rational HATS to create a Web application capable of transforming
More informationDataPro Quick Start Guide
DataPro Quick Start Guide Introduction The DataPro application provides the user with the ability to download and analyze data acquired using the ULTRA-LITE PRO range of Auto Meter products. Please see
More informationFluoroSELECT Fluorometer User s Manual
FluoroSELECT Fluorometer User s Manual Version 4.4.A Sigma-Aldrich Chemie GmbH Industriestrasse 25 CH-9470 Buchs Switzerland EurTechServ@sial.com techservice@sial.com (US and Canada) sigma-aldrich.com
More information