FACCalibur Users Guide

Size: px
Start display at page:

Download "FACCalibur Users Guide"

Transcription

1 FACCalibur Users Guide FACSCalibur Start Up Procedure If the Instrument is OFF: 1. Check the sheath and waste tanks. Open the fluidics drawer (front of instrument) and check the levels of sheath fluid and waste in reservoirs. Sheath fluid should be filled to 3/4 capacity (marked line) ). Waste should be empty except for ~10ml of bleach. Fill and empty as necessary. 2. Pressurize the sheath tank. Flip the switch between the two fluidic reservoirs towards you (Pressurize). 3. Turn on cytometer. The FACSCalibur must be turned ON before the computer. If you have a problem connecting to the cytometer, then the cytometer was not turned on in time to be recognized by the computer. 4. Make sure Cytometer is in STANDBY mode. Make sure the tube of water is on the SIP, the STANDBY button is lit, and the speed is set to LO. If you are using the Red Laser (APC channel), then you must give the instrument 20 minutes to warm up

2 If the Instrument is ON already already: 1. Make sure the Cytometer is in STANDBY mode. Make sure the tube of water is on the SIP SIP, the STANDBY button is lit, and the speed is set to LO. 2. DePressurize the Sheath Tank Flip the switch between the two fluidic reservoirs away from you (depressurize-vent) (depressurize 3. Fill and Empty the Sheath and Waste Tanks as needed Sheath Fluid should be filled to ¾ capacity (marked line). Waste should be empty except for ~10ml of bleach. 4. Repressurize the Sheath Tank Flip the switch between the two fluidic reservoirs toward you (pressurize) CELLQuest Pro Software Start Up & Sample Acquisition 1. Turn on computer, log in and open CELLQuest Click on the CellQuest Pro Icon in your menu bar. The Following window will pop up:

3 2. Connect to the Cytometer. In the Menu Bar, Choose Acquire: Connect to Cytometer Cytometer. If you get an error message at this time, it means the computer was turned on BEFORE the instrument. You must restart the computer in order for the software to connect to the instrument. If the Following window (Acquisition Control) doesn t pop up when you connect to the cytometer: You can try disconnecting: and reconnecting to the software.

4 You can also control acquisition in the Parameter Description window. It can be found in the Menu Bar under: Acquire: Parameter Description. This is also where you will choose where you save your experiment, name your samples, and label your axes. 3. Open your Acquisition Template. In the Main Menu Bar, Choose File: Open Document... If you have a template saved, look in the DATA drive, under: Data: Your PI: Templates/Instrument Settings Folder **If you are making a new Template, use the toolbar on the left side of the screen to draw histograms and dot plots on the blank template (See New Template Protocol). Be sure to SAVE the Template in the Template/Instrument Settings folder for your lab!!! **If you try to run and don t see any events, or your axes are labeled P1, P2, etc. make sure your PLOT TYPE is ACQUISITION (or ACQUISITION>ANALYSIS). To check this, open: Plots: Format Histogram (or Dot Plot, depending on the type of plot you ve constructed)

5 The following window will pop up. This allows you to change Plot Type from Analysis to Acquistion.. You can also change things like your X parameter (histogram) or X and Y parameters (dot plot), how the plot is gated, etc. 4. Set up your Acquisition and Storage (the number of events to collect etc). Go to Acquire: Acquisition and Storage. (see explanation, next page)

6 a. Make sure your Acquisition Gate is set to Accept All events. b. Make sure your Collection Criteria is set properly. For example, if you would like to save 10,000 events in your live gate, you want to set collection criteria to event count and acquisition will stop when 10,000 of G1=R1 (your primary population gate) events are counted. c. Make sure your Storage Gate data file is set to contain All events and the file Resolution is d. Click on Parameters Saved and the window below on the left will pop up: Be sure you have ALL the proper parameters checked!!! e. Click on OK to exit Acquisition and Storage. 5. Create a folder for your data. a. Open the Parameter Description Window (Acquire: Parameter Description) b. In the Parameter Description window click on the Change button next to the Directory and choose a location in which to save your data files. c. Create a new folder for each experiment by clicking on Data then choosing your lab s folder and your specific folder, once you are there, in the bottom left corner of the window choose Create New Folder. d. Name the folder and Click on Select [your chosen folder name] to select that folder and exit Folder Information. (see next page)

7 e. Double check in the Parameter Description Window, that your Directory (or folder where your samples are being saved) is the proper folder. 6. Change the filename. a. In the Parameter Description window click on the Change button next to the File and choose a name for your data files b. Set the File Name Prefix refix to Sample ID. c. Set the File Name Suffix uffix to File Count and start file count from 001. d. Click on OK to exit File Information.

8 7. Enter the names of your parameters (e.g. Antibody conjugates, fluorescent dyes). 1. Type the name directly into the box for each parameter. P1: FSC (Forward Scatter) P2: SSC (Side Scatter) P3: FL1 (FITC, GFP, Alexa 488) P4: FL2 (PE, Cy3) P5: FL3 (PI, PE-Cy5, PE-Cy7, 7AAD) P6: FL4 (APC) 8. Open the Counters window.. Go to Acquire: Counters. Please note: The Counters gives you a TOTAL event count ONLY (not cell count through gates). 9. Load Instrument Settings. If you would like to use the same Detectors/Amps, Compensation and Threshold settings as in a previous experiment, you can open them by doing the following: Go to Cytometer: Instrument settings. (picture, next page)

9 10. Open AND SET your Instrument Settings Click the OPEN button. You can choose a specific Instrument Settings file or a previously recorded data file from your folder (choosing to use a previously recorded data file will copy the instrument settings from when that file was collected). Remember that when you SET the instrument settings the color wheel will spin, and it might take a few seconds to set! Click on DONE to exit Instrument Settings. 11. Check sample settingss before recording a file. Tick the setup box in the Acquisition Control (or Parameter Description) window. Press Acquire to start viewing sample data. **you cannot record sample data file in setup mode**. Generally, the following steps (12-14) are done in SETUP mode. This way, you can change settings to get optimal signalss before running your experiment. Steps 12 and 13 are usually set using your Unstained Sample, and Step 14 is done using your single stain compensation samples.

10 12. Check Threshold settings. Go to Cytometer: Threshold and the window below will appear. Threshold is used to exclude data below a certain channel (or cells/debris below a certain size). size) Most of the time, threshold hreshold will remain the same. Howev However, er, if you have very small cells, cells or a lot of debris you don t want counted, it helps to lower or rraise the threshold hreshold respectively. Normal settings: Primary Parameter: Secondary Parameter: FSC-H value 52 None 13. Check Detectors/Amps settings settings. Go to Cytometer: Detectors/Amps the window on the next page will appear. Adjust each parameter as required. (see next page for guided instructions)

11 In set up mode (see step 11) run a sample of UNSTAINED cells and adjust the following: MODE: VOLTAGE: FL-4 (P7): Log amplification is often used to analyze samples with a large dynamic range of fluorescence signals. The log scale has four decades of range. Set all fluorescent channels to LOG (unless doing cell cycle analysis) Linear amplification is usually used for all light scatter parameters. It is also useful for fluorescent parameters used in DNA quantitation experiments. Set FSC and SSC to LIN Linear amplifier gain can be adjusted from The chosen value multiplies the signal by a factor of As the voltage is increased, the detector sensitivity increases, resulting in increased signal. As the voltage is decreased, the detector sensitivity decreases, resulting in decreased signal. PMT voltages are adjusted using the Voltage sliders In order to Acquire the FL FL-4 4 channel (most commonly APC) you must check the Four Color Box at the bottom of the Detector/Amps Window. 14. Check Compensation settings settings. Go to Cytometer: Compensation and the window on the next page will appear. appear (see Compensation tutorial for setting compensation on the FACSCalibur). In set up mode (see step 11) run each of your single stained samples, and adjust compensation so that there is minimal spillover of each signal into neighboring channels. (i.e. a FITC only sample doesn t appear to be FITC and PE positive)

12 **If you would like to save the changes your instrument settings in a separate file (instrument settings are the Threshold, Detectors/Amps and Compensation settings) go to Cytometer: Instrument Settings and in the window, click SAVE. Save the file in your Template/Instrument settings folder, and use it in future experiments for STEP 9 (Load Instrumentt Settings.) 15. Record a data file. a. Enter your Sample ID in the Parameter Description window. b. Uncheck the Setup box on the Acquisition Control window. Your first file (name.001) is ready for recording. c. Vortex your sample, double check that there are no aggregates in your sample, and place on SIP. d. Run cytometer on Low and click Acquire on the Acquisition Control window. e. The number of events per second (Counter panel) should not exceed If your cells/events are running faster then remove from the SIP and dilute. If they are running slower then increase you sample speed to Medium or High (buttons above the RUN/STANDBY buttons). f. If you do not reach the required number of events before the sample runs out, click Pause and then Save. g. Repeat steps a-f for each sample. 16. Backup your data at the end of each run. Either Bring a USB drive, or see the Data Transfer Guide for transferring your data to another computer The CIC facility is not equipped for data storage. Data is erased from all CIC computers (including workstations) on a monthly basis. 17. Please see the Data Analysis Protocols for analysis options.

13 FACSCalibur Shut Down Procedure 1. Disconnect the software from the cytometer. After running the last sample, set the cytometer to STANDBY, make all necessary saves in CELLQuest Pro and choose Disconnect from Cytometer in the Acquire menu. 2. Clean flow cell with 10% Contrad for 3 minutes. Place a tube with 3 ml of 10% Contrad on the SIP with the arm in the open (back) position and aspirate for 30 seconds. Then close the arm (forward) and RUN the tube on HIGH for 3 minutes. 3. Rinse flow cell with water for 3 minutes. Place a tube with 3 ml of water on the SIP and aspirate for 30 seconds with the arm in the open position. Then close the arm and RUN on HIGH for another 3 minutes. 4. Put the instrument in STANDBY (turn the instrument off if necessary, see guidelines below) and depressurize the Sheath Tank by flipping the switch in the Fluidics Drawer to Vent- Change Tank. If someone is not on the instrument within 2 hours after your appointment, shut down the instrument If someone is on the instrument within 2 hours after your appointment, keep the instrument turned on (and in STANDBY mode) You can check the reservation schedule by double clicking on the reservation schedule shortcut on your desktop or going to

BD CellQuest Pro Acquisition Tutorial

BD CellQuest Pro Acquisition Tutorial BD CellQuest Pro Acquisition Tutorial Introduction This tutorial guides you through a CellQuest Pro Acquisition run like the one demonstrated in the CellQuest Pro Acquisition Movie on the BD FACStation

More information

BD CellQuest Pro Software Acquisition Tutorial

BD CellQuest Pro Software Acquisition Tutorial BD CellQuest Pro Software Acquisition Tutorial This tutorial guides you through a typical acquisition using BD CellQuest Pro software. If you are already familiar with previous versions of BD CellQuest

More information

CELLQuest Acquisition Tutorial

CELLQuest Acquisition Tutorial CELLQuest Acquisition Tutorial Introduction This tutorial guides you through a CELLQuest Acquisition run like the one demonstrated in the CELLQuest Acquisition Movie on the FACStation Overview CD-ROM.

More information

CyAn ADP Guide. Starting Up

CyAn ADP Guide. Starting Up CyAn ADP Guide Starting Up 1. Check the sheath and waste fluids, if needed, fill and empty: 1. To fill the Sheath Tank: a. Unscrew the top of the tank labeled Sheath Tank, and set it on top of the tank.

More information

FACSLSRFortessa SORP QUICK REFERENCE GUIDE

FACSLSRFortessa SORP QUICK REFERENCE GUIDE FACSLSRFortessa SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username

More information

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324 LSRFortessa Guide Cell Evaluation and Therapy Core HO324 Before Use: Prime the instrument twice with no tube and arm to the side Ensure that sheath is full and waste is empty To Begin: Log into ilabs tracker

More information

Unscrew and remove the cap from the waste tank

Unscrew and remove the cap from the waste tank Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish

More information

USER GUIDE TO BECKMAN COULTER FC500

USER GUIDE TO BECKMAN COULTER FC500 USER GUIDE TO BECKMAN COULTER FC500 Create a New Protocol Purpose: This procedure describes how to define the parameters and create rough analysis plots for a new experimental set-up. Some protocol templates

More information

Science is hard. Flow cytometry should be easy.

Science is hard. Flow cytometry should be easy. Science is hard. Flow cytometry should be easy. CFlow User Guide TABLE OF CONTENTS 1 INTRODUCTION TO CFLOW... 1 1.1 Installing CFlow... 1 1.2 Starting CFlow... 1 1.3 CFlow Workspace... 2 1.4 Opening a

More information

FACSLyric TM General User instructions

FACSLyric TM General User instructions flowkon Flow Cytometry User Instructions at the University of Konstanz FACSLyric TM General User instructions In BD FACSuite software, you can measure and analyze samples using either assays or experiments.

More information

Canto II Operation Protocol

Canto II Operation Protocol Canto II Operation Protocol Make sure the following actions have been taken before running your samples. Make sure the waste tank is empty and refilled with 500 ml clean (see page 15 for details) Make

More information

FACSAria Fusion SORP QUICK REFERENCE GUIDE

FACSAria Fusion SORP QUICK REFERENCE GUIDE FACSAria Fusion SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Admin Password: BDIS 2. Ensure the flow cell access door is open. You will want to

More information

UNC Flow Cytometry Core Facility LSR II Operation 2018

UNC Flow Cytometry Core Facility LSR II Operation 2018 Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration).

More information

FACScan 4 and 5 Color Upgrade Rainbow Users Guide

FACScan 4 and 5 Color Upgrade Rainbow Users Guide FACScan 4 and 5 Color Upgrade Rainbow Users Guide For Research Use Only 1 Table of Contents Overview... 3 Hardware Specifications... 3 Rainbow Software... 4 Rainbow Preferences Window... 5 Setting up the

More information

Setup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function

Setup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function Setup protocol for Firefly particles On BD LSR II and LSRFortessa cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction

More information

Gallios TM Quick Reference

Gallios TM Quick Reference Gallios TM Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic

More information

BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option

BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option This guide contains instructions for using BD FACSDiva software version 6 with BD FACSCanto and BD FACSCanto II systems

More information

BD FACSMelody Cell Sorter Quick Reference Guide

BD FACSMelody Cell Sorter Quick Reference Guide BD FACSMelody Cell Sorter Quick Reference Guide This guide contains instructions for using the BD FACSMelody cell sorter with BD FACSChorus software version. and later. Workflow Overview The following

More information

Setup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function

Setup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function Setup protocol for Firefly particles On BD FACSCanto cytometers with single tube loader function Version 1 Last Updated May 2016 Setup protocol for Firefly particles Contents Introduction Cytometer requirements

More information

Flow Cytometry & FACS Core Facility

Flow Cytometry & FACS Core Facility ZMBH - Zentrum für Molekulare Biologie der Universität Heidelberg Flow Cytometry & FACS Core Facility BD FACSCanto TM with HTS User's Guide For authorization to use the FACSCanto II cytometer, you first

More information

BD Influx Standard Operation Protocol - Basic Operation

BD Influx Standard Operation Protocol - Basic Operation Optimizing the drop phase BD Influx Standard Operation Protocol - Basic Operation 1. Verify the 2-tube holder with dummy 15ml tube is in place on the sort stage 2. Select 2 tube holder 2 way sort as the

More information

BD Multiwell AutoSampler Additional Features Tutorial

BD Multiwell AutoSampler Additional Features Tutorial BD Multiwell AutoSampler Additional Features Tutorial Introduction This tutorial provides step-by-step instructions on how to use the additional features available in BD Multiwell Plate Manager (MPM) software

More information

Release Notes Life Technologies Attune NxT Software v2.2

Release Notes Life Technologies Attune NxT Software v2.2 Release Notes Life Technologies Attune NxT Software v2.2 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software

More information

Spherotech, Inc Irma Lee Circle, Lake Forest, IL

Spherotech, Inc Irma Lee Circle, Lake Forest, IL New flow cytometers, with fluorescent channels from the UV to Far Red, and corresponding fluorescent conjugates are now available. As a result, we have developed the Ultra Rainbow Fluorescent Particles

More information

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: Shared Resources: Flow Cytometry and Cell Sorting Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: DHVI BD LSRII (L01/L02) and BD LSRFortessa

More information

Flow Cytometry/FACS Bio-Rad S3e Cell Sorter

Flow Cytometry/FACS Bio-Rad S3e Cell Sorter Flow Cytometry/FACS Bio-Rad S3e Cell Sorter Contents Brief Description... 3 Reference... 3 Location... 3 Suggestion for Materials and Methods... 4 Suggestion for Acknowledgements... 4 People and Contacts...

More information

The Danish Stem Cell Center (DanStem) Danish Stem Cell Center Flow Cytometry Core Facility. LSR Fortessa Guide

The Danish Stem Cell Center (DanStem) Danish Stem Cell Center Flow Cytometry Core Facility. LSR Fortessa Guide Danish Stem Cell Center Flow Cytometry Core Facility LSR Fortessa Guide LSR Fortessa Overview The Danish Stem Cell Center (DanStem) BD FACSFlow Supply System Fluidics Cart Sheath container Waste container

More information

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: Shared Resources: Flow Cytometry and Cell Sorting Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance: DHVI BD LSRII (L01/L02) and BD LSRFortessa

More information

Release Notes Life Technologies Attune NxT Software v2.3

Release Notes Life Technologies Attune NxT Software v2.3 Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software

More information

FloMax Software for Cytometry. Operating Manual Instrument Control and Acquisition. FloMax Operating Manual Instrument Control and Acquisition

FloMax Software for Cytometry. Operating Manual Instrument Control and Acquisition. FloMax Operating Manual Instrument Control and Acquisition FloMax Software for Cytometry Operating Manual Instrument Control and Acquisition 1 Contents Introduction to FloMax Software Acquisition and Instrument Control...3 Operating Basics - Preparing FloMax for

More information

Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0

Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0 Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0 In the following pages you will find instructions describing: 1. Software installation 2. New software features and fixes 3. Known

More information

Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer

Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer After completing this tutorial you will be able to: Create an Experiment with the Experiment Wizard in FCAP Array software. Export the Experiment

More information

NovoExpress Software Guide. Software Version 1.2.4

NovoExpress Software Guide. Software Version 1.2.4 NovoExpress Software Guide Software Version 1.2.4 Table of Content 1. Introduction 6 1.1 Revision History...6 1.2 About the NovoExpress Software...6 1.3 Conventions...6 2. Installation 8 2.1 Installation

More information

Applied Biosystems Attune Cytometric Software version 1.2 Release Notes

Applied Biosystems Attune Cytometric Software version 1.2 Release Notes Applied Biosystems Attune Cytometric Software version 1.2 Release Notes Customer ( you ) must review these notes prior to installing or operating the Attune Cytometric Software version 1.2 ( Attune Software

More information

A Bright Cytometer Easy to use

A Bright Cytometer Easy to use BriCyte E6 Flow Cytometer A Bright Cytometer Easy to use BriCyte E6 Flow Cytometer Flow cytometer is an essential tool to analyze cells, so it s used for diagnosis as well as research in laboratories worldwide.

More information

BD CellQuest Pro Analysis Tutorial

BD CellQuest Pro Analysis Tutorial BD CellQuest Pro Analysis Tutorial Introduction This tutorial guides you through a CellQuest Pro Analysis run like the one demonstrated in the CellQuest Pro Analysis Movie on the BD FACStation Software

More information

MERGE FILES TUTORIAL. Version 1.4

MERGE FILES TUTORIAL. Version 1.4 MERGE FILES TUTORIAL Version 1.4 1. INTRODUCTION The merge process consists on adding two or more sample data files from the same sample or from different samples. The file resulting from the merging will

More information

LEGENDplex Data Analysis Software Version 8 User Guide

LEGENDplex Data Analysis Software Version 8 User Guide LEGENDplex Data Analysis Software Version 8 User Guide Introduction Welcome to the user s guide for Version 8 of the LEGENDplex data analysis software for Windows based computers 1. This tutorial will

More information

Log into the computer using your CU Denver username and password. Log into the DIVA software using your CU and Username and the password flow

Log into the computer using your CU Denver username and password. Log into the DIVA software using your CU and Username and the password flow Log into the computer using your CU Denver username and password Log into the DIVA software using your CU and Username and the password flow When DIVA opens, IF a CST Mismatch pop up box appears, then

More information

STANDARD OPERATING PROCEDURE FOR THE LSR II

STANDARD OPERATING PROCEDURE FOR THE LSR II The Flow Cytometry Facility STANDARD OPERATING PROCEDURE FOR THE LSR II All users must follow the SOP & fill in the logbook. If you are unsure about anything please do not hesitate to ask for help. Before

More information

Release Notes Thermo Fisher Scientific Attune NxT Software v2.4

Release Notes Thermo Fisher Scientific Attune NxT Software v2.4 Release Notes Thermo Fisher Scientific Attune NxT Software v2.4 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance

More information

Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ).

Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ). Life Technologies Attune NxT Software version 1.1.0 Release Notes Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version 1.1.0 ( Attune Software v1.1.0

More information

Attune NxT Acoustic Focusing Cytometer

Attune NxT Acoustic Focusing Cytometer USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Numbers A24858, A24859, A24860, A24861, A24862, A24863, A24864, A28993 Publication Number 100024235 Revision C.0 For Research Use Only. Not for

More information

Quick Start Guide: COLLECT

Quick Start Guide: COLLECT Quick Start Guide: COLLECT Ready, Set, Run! The Collect tab is used to collect or import a sample, configure the run settings and set up plots and gates. Collect a Sample 1. Start the C6 Flow Cytometer,

More information

JMJ-Group Microbiology BMB SDU

JMJ-Group Microbiology BMB SDU JMJ-Group Microbiology BMB SDU Title: Daily use of FACS Aria II SOP number: SOPFACS0042_v01 Version number: 01 Date issued: 2013.01.16 Review date: 2014.12.01 Written by: Tina Kronborg and Michelle Madelung

More information

Hacking FlowJo VX. 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From

Hacking FlowJo VX. 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From Hacking FlowJo VX 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From Contents 1. Change the default name of your files. 2. Edit your workspace

More information

North America/International Japan Europe

North America/International Japan Europe North America/International Japan Europe 1730 North First Street San Jose, CA 95112 U.S.A. Voice: +1 800-275-5963 FAX: +1 408-352-4130 sales@sonybiotechnology.com http://www.sonybiotechnology.com 1-7-1,

More information

CyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System

CyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System CyFlow ML Healthcare Immunology Pathology Microbiology Cell Biology Multilaser 16 Parameter Desktop FCM System 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements

More information

Technical Note. Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6. Introduction

Technical Note. Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6. Introduction Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6 Contents 1 Introduction 2 Before You Begin 3 Using Global FlowJo Preferences 5 Usinig the Transform Button 5 Saving

More information

ENCENDIDO Y LAVADOS DEL CITÓMETRO

ENCENDIDO Y LAVADOS DEL CITÓMETRO ENCENDIDO Y LAVADOS DEL CITÓMETRO ANTES DE INICIAR EL PARATO SI EL ORDENADOR ESTÁ ENCENDIDO Y EL CITÓMETRO APAGADO: 1. APAGAR EL ORDENADOR 2. ENCENDER EL CITÓMETRO 3. ENCENDER EL ORDENADOR AL INICIAR EL

More information

MACSQuantify Software 2.11

MACSQuantify Software 2.11 MACSQuantify Software 2.11 User manual Version 1 Original instructions Copyright 2017 Miltenyi Biotec GmbH and/or its affiliates. All rights reserved. No part of this user manual may be reproduced, stored

More information

Sysmex CyFlow Cube Series

Sysmex CyFlow Cube Series Sysmex CyFlow Cube Series Reliable, Compact and Cost-effective Flow Cytometers For Research Use Only. Not for use in diagnostic or therapeutic procedures. Sysmex CyFlow Cube Series Compact, Economic Flow

More information

LSM510 Confocal Microscope Standard Operation Protocol Basic Operation

LSM510 Confocal Microscope Standard Operation Protocol Basic Operation LSM510 Confocal Microscope Standard Operation Protocol Basic Operation Please make sure that the COMPRESSED AIR has been TURNED ON prior to the use of the equipment. Kindly inform the administrator if

More information

FX500 Exchangeable Fluidics Cell Sorter

FX500 Exchangeable Fluidics Cell Sorter FX5 Exchangeable Fluidics Cell Sorter Sony Biotechnology Inc. FX5 Cell Sorter Sorting Made Simple The FX5 is a microfluidics chip based cell sorter featuring a fully replaceable fluidics system. This lets

More information

MA900 Multi-Application Cell Sorter

MA900 Multi-Application Cell Sorter MA9 Multi-Application Cell Sorter Sony Biotechnology Inc. MA9 Cell Sorter Sorting Made Simple The MA9 from Sony meets the needs of most sorting applications, supporting 12 fluorescence parameters and 4-way

More information

Gallios Flow Cytometer

Gallios Flow Cytometer Gallios Flow Cytometer For Research use only, not for use in diagnostic procedures. Quick Reference PN A75199AB.1 (May 2011) Beckman Coulter Inc. PN A75199AB.1 (May, 2011) 1 of 24 WARNINGS AND PRECAUTIONS

More information

INSTALLING THE PS3 XBOX READY SOFTWARE:

INSTALLING THE PS3 XBOX READY SOFTWARE: INSTALLING THE PS3 XBOX READY SOFTWARE: 1. Insert the Installation CD to CD-ROM drive and execute Ready_Setup.exe NOTE: If it is the first time for the target USB disk using under this software, the software

More information

for the Analysis of Flow Cytometry Listmode Data Files

for the Analysis of Flow Cytometry Listmode Data Files Purdue University Cytometry Laboratories Introduction to WinMDI 2.8 (J. Trotter 1993-1998) for the Analysis of Flow Cytometry Listmode Data Files Objectives of this tutorial: - Open a listmode data file

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2. Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and instrument

More information

Cytomics FC 500 Quick Reference

Cytomics FC 500 Quick Reference Cytomics FC 500 Quick Reference PN 722238 RevA (April 2005) Beckman Coulter Inc. Miami Education Center Miami Lakes, WARNINGS AND PRECAUTIONS READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED

More information

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2. Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.6 Contents 1. Description 1.1 Background information 1.2 Reagent and

More information

Instructions for Use. COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software. PN CA (September 2010)

Instructions for Use. COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software. PN CA (September 2010) READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software Instructions for Use CYTOMETER LASER SAMPLE FLOW LOW MED HIGH (September 2010)

More information

BD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007

BD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007 BD Biosciences Technical Resources Page 1 For use with the BD Lyoplate Human Cell Surface Marker Screening Panel (Cat. No. 560747). Please check that your catalog numbers for the FCS Express Excel templates

More information

CyFlow Cube 6 Instrument Operating Manual

CyFlow Cube 6 Instrument Operating Manual CyFlow Cube 6 Instrument Operating Manual CyFlow Cube 6 CyFlow Cube 8 - Instrument Operating Manual For in vitro diagnostic use with Partec recommended IVD reagents. IVD The PartecCyFlow Cube Flow Cytometer

More information

Summit Software WorkList Builder Quick Start Guide

Summit Software WorkList Builder Quick Start Guide Summit Software WorkList Builder Quick Start Guide Document Number 0003185 Revision July 2006 2 Summit WorkList Builder Quick Start Guide Disclaimer Summit Worklist Builder is a Research Use Only software

More information

Nanosight NS300 NBTC User Instructions

Nanosight NS300 NBTC User Instructions Nanosight NS300 NBTC User Instructions Preliminary notes: -The laser module is delicate and expensive. Handle it with care and do not leave it on the bench unattended. Only leave it in place in the instrument

More information

FlowJo Software Lecture Outline:

FlowJo Software Lecture Outline: FlowJo Software Lecture Outline: Workspace Basics: 3 major components 1) The Ribbons (toolbar) The availability of buttons here can be customized. *One of the best assets of FlowJo is the help feature*

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Luminex Prima Theory of Operation Luminex 200 technology is based on flow cell

More information

ibidi PumpControl Instruction Manual Version 1.5.0

ibidi PumpControl Instruction Manual Version 1.5.0 ibidi PumpControl Instruction Manual Version 1.5.0 II ibidi GmbH 2011, version 1.5.0, revision 2, 2011-10-26, Andreas Rainer Table of contents 1 General Information... 1 1.1 System Requirements... 1 1.2

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

NovoSampler Pro Operator s Guide

NovoSampler Pro Operator s Guide NovoSampler Pro Operator s Guide Table of Content Prologue 5 I. Revision History...5 II. Contact Information...5 III. Trademark...6 IV. Intended Use...6 V. Using this Guide...6 VI. Conventions...6 VII.

More information

PANALYTICAL X PERT PRO XRD

PANALYTICAL X PERT PRO XRD STANDARD OPERATING PROCEDURE: PANALYTICAL X PERT PRO XRD Purpose of this Instrument: Identification of signal-phase materials and multi-phase mixture; Quantitative determination of amounts of different

More information

Homework Assignment 9 LabVIEW tutorial

Homework Assignment 9 LabVIEW tutorial Homework Assignment 9 LabVIEW tutorial Due date: Wednesday, December 8 (midnight) For this homework assignment, you will complete a tutorial on the LabVIEW data acquistion software. This can be done on

More information

SPA III Quick Reference Guide for Custom Panels

SPA III Quick Reference Guide for Custom Panels SPA III Quick Reference Guide for Custom Panels This guide contains instructions for using the BD FACS Sample Prep Assistant III (SPA III) with BD FACS SPA software version.0 and later. Initial Setup Custom

More information

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2. Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.8 Contents 1. Description 1.1 Background information 1.2 Reagent and

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

BD FACSDiVa Option. BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen

BD FACSDiVa Option. BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen BD FACSDiVa Option The Digital Advantage for the BD FACSVantage SE Flow Cytometry System BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen Proven Performance in Flow Cytometry

More information

Getting Started. COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software. PN BA (October 2010)

Getting Started. COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software. PN BA (October 2010) READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software Getting Started CYTOMETER LASER SAMPLE FLOW LOW MED HIGH (October 2010)

More information

Quick Guide NC-200. NucleoCounter. to the easiest cell count ever!

Quick Guide NC-200. NucleoCounter. to the easiest cell count ever! Quick Guide to the easiest cell count ever! NucleoCounter NC-00 How to get started Dear NucleoCounter Customer! Thank you for purchasing the NucleoCounter NC-00 offering not only the easiest one-step viability

More information

BD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual

BD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual 643274 BD CellQuest Pro Software Reference Manual BD CellQuest Pro Software Reference Manual bdbiosciences.com Part No. 643274 Rev. A November 2007 BD Biosciences

More information

Confocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by

Confocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by Confocal Microscope - Radiance Basic Instructions (Stefanie Reichelt) The Radiance Confocal Scanning Microscope designed and developed by Brad Amos (MRC-LMB) Commercialised by Bio-Rad Inc. (1998 2004)

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Special Procedures and Troubleshooting

Special Procedures and Troubleshooting READY ON FS SS AUX FL1 FL2 FL3 FL4 TM COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software Special Procedures and Troubleshooting CYTOMETER LASER SAMPLE FLOW LOW MED HIGH

More information

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b User Guide Biomedicum Imaging Unit (BIU) University of Helsinki www.biu.helsinki.fi 11.11.2016 1 GENERAL...1

More information

BD Accuri C6 with single tube loader protocol. For multiplex assays

BD Accuri C6 with single tube loader protocol. For multiplex assays BD Accuri C6 with single tube loader protocol For multiplex assays BD Accuri C6 with single tube loader protocol This protocol contains instructions for setting up an Accuri C6 with Tube Loader. If you

More information

Near UV Laser System Performance on the CytoFLEX S

Near UV Laser System Performance on the CytoFLEX S Near UV Laser System Performance on the CytoFLEX S APPLICATION NOTE Authors: Jeffrey Cobb James Tung Affiliation: Life Science, Beckman Coulter, Inc., Miami, FL, United States Introduction Material and

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

Cytometry Data Analysis in FlowJo V10. Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC

Cytometry Data Analysis in FlowJo V10. Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC Cytometry Data Analysis in FlowJo V10 Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC timc@flowjo.com Outline Part I Intro to FlowJo Navigating the V10 Workspace Customizing Ribbons Demo

More information

DIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only

DIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only DIAplex DIAplex Pro 1.0 Software Instructions for use For research use only Fast Track Your Research Table of Contents 1. Introduction... 2 2. DIAplex Pro 1.0 installation... 2 2.1. System Requirements...

More information

Windows 2000 Professional

Windows 2000 Professional The American University in Cairo Academic Computing Services Windows 2000 Professional prepared by Soumaia Ahmed Al Ayyat 4 August 2003 Table of Contents Starting Up the Computer Windows Environment Start

More information

Multi-Photon Training

Multi-Photon Training Multi-Photon Training Overview This training will take approximately 4 hours. The first 2 hours will be spent learning one-on-one with your trainer, while the following 2 hours will be your opportunity

More information

STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1)

STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) Aim To provide detailed description of all steps, material and methods used for fluorescent surface staining and preparation of whole blood cells for analysis

More information

TS MANAGER OPERATIONS MANUAL

TS MANAGER OPERATIONS MANUAL TS MANAGER OPERATIONS MANUAL Super Systems Inc. 7205 Edington Drive Cincinnati, OH 45249 513-772-0060 800-666-4330 Fax: 513-772-9466 www.supersystems.com Table of Contents Introduction... 3 Prerequisites...

More information

Basic Tutorial. Data Analysis Software for Flow Cytometry Windows/Mac user documentation

Basic Tutorial. Data Analysis Software for Flow Cytometry Windows/Mac user documentation Basic Tutorial Data Analysis Software for Flow Cytometry Windows/Mac user documentation FlowJo, its tutorials, documentation and web site are copyright FlowJo, LLC 2018. All Rights Reserved. Version 2.0

More information

User Manual. For Windows operating systems Excelsior Blvd. Suite Pécs St. Louis Park, MN Kedves street 20

User Manual. For Windows operating systems Excelsior Blvd. Suite Pécs St. Louis Park, MN Kedves street 20 User Manual For Windows operating systems Soft Flow, Inc. Soft Flow Hungary Ltd. 5009 Excelsior Blvd. Suite 154 7628 Pécs St. Louis Park, MN 55416 Kedves street 20 USA Hungary Tel: +1 (952) 681-7960 Tel:

More information

Introduction to IBM Rational HATS For IBM System i (5250)

Introduction to IBM Rational HATS For IBM System i (5250) Introduction to IBM Rational HATS For IBM System i (5250) Introduction to IBM Rational HATS 1 Lab instructions This lab teaches you how to use IBM Rational HATS to create a Web application capable of transforming

More information

DataPro Quick Start Guide

DataPro Quick Start Guide DataPro Quick Start Guide Introduction The DataPro application provides the user with the ability to download and analyze data acquired using the ULTRA-LITE PRO range of Auto Meter products. Please see

More information

FluoroSELECT Fluorometer User s Manual

FluoroSELECT Fluorometer User s Manual FluoroSELECT Fluorometer User s Manual Version 4.4.A Sigma-Aldrich Chemie GmbH Industriestrasse 25 CH-9470 Buchs Switzerland EurTechServ@sial.com techservice@sial.com (US and Canada) sigma-aldrich.com

More information