Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Similar documents
Mathematical methods and simulations tools useful in medical radiation physics

Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction

3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Simulations in emission tomography using GATE

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY

Scatter Correction Methods in Dimensional CT

Planar and SPECT Monte Carlo acceleration using a variance reduction technique in I 131 imaging

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

GPU implementation for rapid iterative image reconstruction algorithm

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING

James R Halama, PhD Loyola University Medical Center

Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology

Introduction to Positron Emission Tomography

Advanced Scatter Correction for Quantitative Cardiac SPECT

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Medical Imaging BMEN Spring 2016

Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code

SINGLE-PHOTON emission computed tomography

OBJECT ORIENTED DESIGN OF ANTHROPOMORPHIC PHANTOMS AND GEANT4-BASED IMPLEMENTATIONS ABSTRACT

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Master of Science Thesis. Activity quantification of planar gamma camera images

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

Quantitative capabilities of four state-of-the-art SPECT-CT cameras

Introduction to Emission Tomography

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Image-based Monte Carlo calculations for dosimetry

Concepts, Applications, and Requirements for Quantitative SPECT/CT. Conflict of Interest Disclosure

The new version of the GATE simulation platform

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Assesing multileaf collimator effect on the build-up region using Monte Carlo method

ATTENUATION CORRECTION IN SPECT DURING IMAGE RECONSTRUCTION USING INVERSE MONTE CARLO METHOD A SIMULATION STUDY *

Performance Evaluation of radionuclide imaging systems

A dedicated tool for PET scanner simulations using FLUKA

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Extremely Fast Detector for 511 kev Gamma

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

DUAL energy X-ray radiography [1] can be used to separate

Emission Computed Tomography Notes

Assessment of OSEM & FBP Reconstruction Techniques in Single Photon Emission Computed Tomography Using SPECT Phantom as Applied on Bone Scintigraphy

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

BME I5000: Biomedical Imaging

Modeling the ORTEC EX-100 Detector using MCNP

A publicly accessible Monte Carlo database for validation purposes in emission tomography

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

Ch. 4 Physical Principles of CT

3-D PET Scatter Correction

Quantitative imaging for clinical dosimetry

Application of MCNP Code in Shielding Design for Radioactive Sources

SPECT (single photon emission computed tomography)

USING cone-beam geometry with pinhole collimation,

CT vs. VolumeScope: image quality and dose comparison

Monte-Carlo simulation for scatter correction compensation studies in SPECT imaging using GATE software package

Non-Stationary CT Image Noise Spectrum Analysis

Scatter correction improvement based on the Convolution Subtraction Technique in SPECT imaging

Nad?ge Anizan 1, Hao Wang 2, Xian C Zhou 2, Robert F Hobbs 1,3, Richard L Wahl 1,2 and Eric C Frey 1*

Customizable and Advanced Software for Tomographic Reconstruction

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

UNIVERSITY OF SOUTHAMPTON

Continuation Format Page

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

Slab-by-Slab Blurring Model for Geometric Point Response Correction and Attenuation Correction Using Iterative Reconstruction Algorithms

ATLAS Tracking Detector Upgrade studies using the Fast Simulation Engine

Attenuation Coefficients for Layered Ceiling and Floor Shields in PET/CT Clinics

Performance Evaluation of the Philips Gemini PET/CT System

Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Deep Scatter Estimation (DSE): Feasibility of using a Deep Convolutional Neural Network for Real-Time X-Ray Scatter Prediction in Cone-Beam CT

Preface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

Characterizing the parallax error in multi-pinhole micro-spect reconstruction

Constructing System Matrices for SPECT Simulations and Reconstructions

Proton dose calculation algorithms and configuration data

Fits you like no other

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Scatter in an uncollimated x-ray CT machine based on a Geant4 Monte Carlo simulation. Wadeson, Nicola and Morton, Edward and Lionheart, William

Validation of the Small Animal Biospace Gamma Imager Model Using GATE Monte Carlo Simulations on the Grid

Fast Timing and TOF in PET Medical Imaging

Muon imaging for innovative tomography of large volume and heterogeneous cemented waste packages

SNIC Symposium, Stanford, California April The Hybrid Parallel Plates Gas Counter for Medical Imaging

Geant4 in Brachytherapy

ACOMPTON-SCATTERING gamma camera records two

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Transcription:

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract. Quantitative SPECT imaging of 111 In is of growing importance in nuclear medicine applications like patient-specific radionuclide dosimetry. In this study, the quality of a Monte-Carlo based model of the imaging chain from nuclear decay to detection in a gamma camera is investigated. Important characteristics like the camera point spread function (PSF), the phantom attenuation and scatter effects are compared between simulation and experiment. The model shows a good agreement to the actual measurements and allows for further use in the development of quantitative SPECT reconstruction which will be mandatory for future oncology. 1 Introduction Single Photon Emission Computed Tomography (SPECT) is an important nuclear imaging modality. In this study, imaging of 111 In is investigated, a radioisotope mainly used for applications in oncology. Here, the absolute assessment of local tracer activity provides access to both tumor viability staging and treatment prognosis. Due to its special characteristics including two equally dominant particle energy lines, 111 In imaging can become rather challenging in terms of image quality degradation e.g. due to down-scatter contamination. We present qualitative and quantitative comparisons between experimental and simulated gamma camera data acquisitions in order to prove the quality of a Monte-Carlo based imaging model to be used in the development of a next-generation quantitative SPECT reconstruction. 2 Materials and Methods 2.1 Phantom Acquisitions Gamma camera acquisitions from two different phantom types have been acquired on a Philips Precedence SPECT/CT scanner at Rigs Hospital in Copenhagen. For PSF assessment, a planar measurement with a point source realized as an activity of 7.4MBq 111 In located in a syringe tip was used. For the evaluation of attenuation and scatter impact, a SPECT scan from a NEMA PET

346 Schweizer & Goedicke phantom was acquired (Fig.1(a)). Two cylinders have been filled with different concentrations of 111In diluted in water: the outer cylinder (9.3kBq/cm 3, warm background ) and one insert (88.2kBq/cm 3, hot insert ). In addition to that, two cold cylinders (air and Teflon) have been installed. Applied 111 In activities were determined with a Capintec CRC-15R dose calibrator. Volume measures were derived by a segmentation procedure based on CT images of the phantom. For the point source measurement, a source-to-camera distance of 15 cm was chosen and 5 10 5 counts were detected in a planar configuration. For the NEMA PET phantom, 128 projections with (128 pixels) 2 resolution on a circular trajectory have been measured (Fig.1(a)). The acquisition time per projection was 30 s. In all acquisitions, a standard MEGP collimator was used and a standard dual 111 In energy window setting has been chosen for energy discrimination (1: 154 188keV, 2: 221 270keV). 2.2 Monte-Carlo Simulations For Monte-Carlo simulations of the imaging chain, an advanced SPECT simulation framework based on GEANT4 [1] has been used, which was accelerated via the Forced-Detection-based variance reduction technique [2][3]. As an input for the simulation, voxelized activity and attenuation maps for the different phantom parts as well as the patient table models have been generated from the acquired CT-scan data (Fig.1(b)). The camera model included a standard MEGP collimator as well as a 9.5 mm scintillator crystal plate of NaI. Post-processing of the simulated detector events was performed to emulate binning impact, signal blurring (light spread) and effects of the detector electronics. According to the activity in the experiment, the number of simulated photon histories per projection was 2.4 10 9 for the point source and 4.4 10 9 for the NEMA PET phantom. Fig. 1. Experimental setup for the NEMA PET phantom acquisition in the Philips Precedence SPECT/CT scanner (left). Modeling of the measurement in the GEANT4 based SPECT simulation framework (right).

GEANT4 for 111 In SPECT Acquisition 347 2.3 Analysis Planar gamma camera images of the point source were projected along pixel rows to obtain 2D profile plots of the PSF. The absolute camera sensitivity S was calculated from point-source images as S = N/(A T), with N the number of detected counts, A the phantom activity and T the acquisition time. For the NEMA PET phantom, the detected counts in each projection were plotted against the projection number. The scatter fraction for each detection window was determined by applying a triple energy window (TEW) technique on the simulated projection data. 3 Results The analysis of the PSF-profile plots (Fig.2) shows a good agreement between the experiment (red) and simulation of an idealized point source in air (green) with respect to signal intensity and shape in the central part of the curve. A further significant improvement, especially for the non-central areas was observed after including the patient tray and the camera lead shielding and plastic cover in the simulation (blue curve). The achieved improvement indicates that these signal components are due to additional scatter events from these parts of the simulation geometry. The remaining differences in the wing area of the curve can be explained by the basic strategy of the forced detection method to exclude those parts of the geometry from the radiation transport which generally only yield negligible contributions to the main result. Fig. 2. Comparison between measured and simulated In-111 PSF on a logarithmic scale.

348 Schweizer & Goedicke However, it should be noted that the signal level here is more than two orders ofmagnitudebelowthepeaksignal(andevenlower, ifthedataisanalyzedinthe 2D detector area as opposed to a 1D-projection like in Fig.2). The agreement in the PSF amplitude is also reflected in the camera sensitivity figures, calculated Fig.3. DetectedcountsasafunctionofthecameraanglefortheNEMAPETphantom. Both curves are normalized to the same integral number of counts (top). Scatter fraction in lower and upper 111 In energy window as a function of the projection number for the NEMA PET phantom simulated in GEANT4 (bottom).

GEANT4 for 111 In SPECT Acquisition 349 as S = 163 cps/mbq for the measured and 157 cps/mbq for the simulated point source acquisitions. In Fig.3(a), the detected counts per projection in the experiment and the simulation are compared with each other. Note that the focus of this analysis is on an analysis of the attenuation and scatter impact. Therefore the two curves were normalized to the same integral number of counts in order to facilitate the shape comparison. Obviously, the curves show an extremely good agreement. The determined integral difference between both plots before normalization was 6.9%, which is clearly within the boundaries of the experimental activity calibration error. In addition to that, Fig.3(b), displays the scatter fraction in the simulated signal for the NEMA PET phantom. The mean scatter fractions of 39.5% for the lower energy window and 21.1% for the upper energy window are in good agreement with other studies from literature [4], where for a similar approach values of 43% and 22% respectively are reported. 4 Discussion The results achieved for both phantom types demonstrate a good agreement between experiment and simulation. In the PSF studies, the accurate modeling of important camera properties like PSF shape and absolute camera sensitivity could be demonstrated, especially when a realistic scanner model is included in the simulation geometry. This indicates that more complex phantom geometries can also be modeled accurately since such a case can be regarded as a superposition of point-sources from a mathematical view-point. In the NEMA PET phantom studies, detailed modeling of the complete imaging chain from nuclear decay to detection in the gamma camera could be proven. The quality of the Monte-Carlo simulation data warrants its use in quantitative SPECT. Future work based on these results will include the generation of highly realistic test datasets for algorithm development as well as use of accurate physical models in advanced SPECT reconstruction. References 1. Agostinelli S, et al. GEANT4: a simulation toolkit. Nucl Instrum Methods Phys Res A. 2003;506:250 303. 2. de Jong HWAM, Beekman FJ, Slijpen TP. Acceleration of Monte Carlo SPECT simulation using convolution-based forced detection. IEEE Nucl Sci Symp Conf Rec. 1999 24-30 Oct;3:1532 6. 3. Goedicke A, Schweizer B, Staelens S, et al. Fast simulation of realistic SPECT projections using forced detection in GEANT4. In: Proc Med Biol Eng Conf; 2005. 4. Holstensson M. Optimisation of Window Settings for Quantitative 111In Imaging: A Comparison of Measurements to Monte Carlo. Sweden: Lund University; 2006.