COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

Similar documents
Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Lucy Phantom MR Grid Evaluation

RADIOMICS: potential role in the clinics and challenges

Conference Biomedical Engineering

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

TUMOR DETECTION IN MRI IMAGES

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction

Whole Body MRI Intensity Standardization

연구용유방초음파질관리 원광대학병원김혜원

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM

Image Acquisition Systems

Breast MRI Accreditation Program Clinical Image Quality Guide

Real-time self-calibration of a tracked augmented reality display

A Study of Medical Image Analysis System

The VesselGlyph: Focus & Context Visualization in CT-Angiography

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Interpolation of 3D magnetic resonance data

3-D Compounding of B-Scan Ultrasound Images

Tomographic Reconstruction

Fmri Spatial Processing

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

Prostate Detection Using Principal Component Analysis

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Magnetic Resonance Elastography (MRE) of Liver Disease

Advanced Image Reconstruction Methods for Photoacoustic Tomography

BME I5000: Biomedical Imaging

Catalog Numbers: CMB022, CMB022-AA, CMB022-UP, CMB033, CMB033-AA, CMB033-UP, CMB-CW, CMB-LK

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR

Use of MRI in Radiotherapy: Technical Consideration

Inter-slice Reconstruction of MRI Image Using One Dimensional Signal Interpolation

Motion artifact detection in four-dimensional computed tomography images

Role of Parallel Imaging in High Field Functional MRI

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

MEDICAL IMAGE COMPUTING (CAP 5937) LECTURE 4: Pre-Processing Medical Images (II)

Visualisation : Lecture 1. So what is visualisation? Visualisation

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

A study on image quality provided by a kilovoltage cone-beam computed tomography

Volume visualization. Volume visualization. Volume visualization methods. Sources of volume visualization. Sources of volume visualization

Quality control phantoms and protocol for a tomography system

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

arxiv: v1 [cs.cv] 6 Jun 2017

CT Protocol Clinical Graphics Move Forward 3D motion analysis service

Medical Imaging and Beyond

Global Thresholding Techniques to Classify Dead Cells in Diffusion Weighted Magnetic Resonant Images

Medical Image Analysis

3D Surface Reconstruction of the Brain based on Level Set Method

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

MR IMAGE SEGMENTATION

Institute of Cardiovascular Science, UCL Centre for Cardiovascular Imaging, London, United Kingdom, 2

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

An Improved Tracking Technique for Assessment of High Resolution Dynamic Radiography Kinematics

Computational Medical Imaging Analysis Chapter 4: Image Visualization

MEDICAL IMAGE ANALYSIS

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

Micro-CT Methodology Hasan Alsaid, PhD

Computer-Tomography I: Principles, History, Technology

Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

Computational Medical Imaging Analysis

Copyright 2017 Medical IP - Tutorial Medip v /2018, Revision

Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Correction of Partial Volume Effects in Arterial Spin Labeling MRI

The organization of the human cerebral cortex estimated by intrinsic functional connectivity

Generation of Hulls Encompassing Neuronal Pathways Based on Tetrahedralization and 3D Alpha Shapes

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

ACQUIRING AND PROCESSING SUSCEPTIBILITY WEIGHTED IMAGING (SWI) DATA ON GE 3.0T

Machine Learning for Medical Image Analysis. A. Criminisi

MEG Laboratory Reference Manual for research use

Computer-Aided Detection system for Hemorrhage contained region

Ch. 4 Physical Principles of CT

TITLE: Regularized Reconstruction of Dynamic Contrast-Enhanced MR Images for Evaluation of Breast Lesions

Automatic Cerebral Aneurysm Detection in Multimodal Angiographic Images

Performance Evaluation of the TINA Medical Image Segmentation Algorithm on Brainweb Simulated Images

Head Holder for MR-Guided Drug Delivery

MIDAS Signal Calibration

Mirrored LH Histograms for the Visualization of Material Boundaries

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI

Radon Transform and Filtered Backprojection

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Medical Images Analysis and Processing

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Fiber Selection from Diffusion Tensor Data based on Boolean Operators

A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties

Segmentation and Modeling of the Spinal Cord for Reality-based Surgical Simulator

Transcription:

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM Mažena MACIUSOVIČ *, Marius BURKANAS *, Jonas VENIUS *, ** * Medical Physics Department, National Cancer Institute, Vilnius, Lithuania ** Laboratory of Biomedical Physics, National Cancer Institute, Vilnius, Lithuania 12TH INTERNATIONAL CONFERENCE MEDICAL PHYSICS IN THE BALTIC STATES 2015 5-7 th of November Kaunas

What is MRI? A non-invasive imaging technique for studying soft tissue structures and characteristics. The MRI technique has been used for imaging the brain, heart, muscles and joints, for early diagnosing and staging of the diseases, monitoring the effects of therapies. 2

In short The patient is placed into a magnetic field. A radio frequency wave is sent in. The radio frequency wave is turned off. The patient emits a signal. The signal is received and used for reconstruction of the picture. 3 Bayer, 2010.

Purpose To achieve the full potential of MRI and optimal images in a hospital there is a need for testing instrument performance and evaluating the system stability. Protocols for quality assurance (QA) and quality control (QC) of MRI system are required. A simple 3D printer compatible all in one phantom design was presented to fulfil the basic requirements for the assessment of stable MRI scanner operation. Daily/weekly/monthly measurements were performed and the suitability of the proposes system for the stability evaluation was confirmed. 4

Methods and materials Spherical phantom (20 cm Ø) has been made using a SOLIDWORKS software program. The phantom has been printed using a Dimension 1200es 3D Printer. Image and the view of the modelled phantom design rendered on SOLIDWORKS software. 5

Methods and materials The phantom has been printed in four different parts two hemispheres, the grid plate and the rest inner structures printed in one object (low contrast plate, spatial resolution plate, four side structures with crossing grooves). The structures were fixed together and filled with aqueous nickel chloride and sodium chloride solution. 6

Methods and materials Experiments were performed on a clinical 1.5 T MR-system Achieva XR at National Cancer Institute using Philips SENSE Head coil. All measurements were performed in one procedure no removing or repositioning of the phantom has been done. Measurements for every parameter have been performed five times, to acquire reliable data. 7

Results: Image uniformity The variation of percentage image uniformity over measurement period. PIU obtained for the properly functioning scanner should be more than 90%. Calculated PIU is 94.93% ± 1.54. 8

Results: SNR The variation of SNR values in the MRI images over measurement period. The observed SNR stability is 73.99 ± 6.05 (the allowed limits over measurement time ± 10 %). 9

Results: Geometrical distortion The variation of calculated maximum percent geometric distortion (PGD) over measurement period. The allowed value of maximum PGDs should not exceed 2%. The average maximum PGDs over the whole measurement period was 1.04 % ± 0.20. 10

Results: High contrast spatial resolution The plate consists of a line shape holes that form a high resolution test pattern of 1 16 line pair/cm. Spatial resolution is evaluated by visually estimating the amount of line pairs in centimetre that are still possible to distinguish as different lines. During the whole measurement period 6 lp/cm has been clearly visible. 11

Results: Low contrast resolution The variation of different contrast levels over measurement period. The average values of different contrast levels (from 1 to 6) are: 0.5 ± 0.01, 0.39 ± 0.01, 0.26 ± 0.01, 0.17 ± 0.01, 0.14 ± 0.01, 0.07 ± 0.004. The observed different contrast levels vary less than 10 %. 12

Results: Slice thickness Slice thickness is evaluated by measuring the width of the groove G seen in the images and multiplied by the factor of 0.5. The width of the groove is calculated as FWHM of pixel intensity profile (Line P) of the groove. 13

Results: Slice thickness The calculated slice thicknesses over measurement period. The calculated average slice thickness is 5.3 mm ± 0.15. The values are randomly distributed and no tendency can be identified. 14

Results: Slice position Verification is performed by measuring distances between crossing grooves seen in Slice 1 (distance d1) and Slice 6 (distance d2). The distance between slices is 25 mm. 15

Results: Slice position Deviations of the measured slice position from the set position over measurement period. The average difference from the set position is 0.72 mm ± 0.39 when the positioning distance is 25 mm. The deviations are distributed randomly and there is no tendency to lower or increase. 16

Conclusion The 3D printed phantom allowed to measure and calculate all required parameters to evaluate image uniformity, geometric accuracy, SNR, high contrast and low contrast spatial resolution, slice thickness and slice position. The deviation of the measured/calculated parameters has been insignificant therefore the observed constancy of these parameters is sufficient to declare the stable operation of the MRI scanner. To expand the abilities of the printable system further developments and investigations are planned to be performed. 17

18