Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

Similar documents
Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validation

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

Splitting-Based Statistical X-Ray CT Image Reconstruction with Blind Gain Correction

DUE to beam polychromacity in CT and the energy dependence

Superiorized polyenergetic reconstruction algorithm for reduction of metal artifacts in CT images

Reconstruction of CT Images from Sparse-View Polyenergetic Data Using Total Variation Minimization

Statistical iterative reconstruction using fast optimization transfer algorithm with successively increasing factor in Digital Breast Tomosynthesis

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

ATTENUATION correction is required for quantitatively

Accelerating CT Iterative Reconstruction Using ADMM and Nesterov s Methods

Introduction to Biomedical Imaging

ML reconstruction for CT

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Segmentation-Free Quasi-Newton Method for Polyenergetic CT Reconstruction

STATISTICAL image reconstruction methods have shown

Simultaneous Estimation of Attenuation and Activity Images Using Optimization Transfer

Augmented Lagrangian with Variable Splitting for Faster Non-Cartesian L 1 -SPIRiT MR Image Reconstruction: Supplementary Material

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Axial block coordinate descent (ABCD) algorithm for X-ray CT image reconstruction

CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

(RMSE). Reconstructions showed that modeling the incremental blur improved the resolution of the attenuation map and quantitative accuracy.

Iterative SPECT reconstruction with 3D detector response

Union of Learned Sparsifying Transforms Based Low-Dose 3D CT Image Reconstruction

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications

Ch. 4 Physical Principles of CT

Noise weighting with an exponent for transmission CT

STATISTICAL image reconstruction methods for X-ray

A Comparative Study of Two Methods for the Correction of Beam Hardening Artifacts in X-ray Computed Tomography

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Estimating 3D Respiratory Motion from Orbiting Views

Radon Transform and Filtered Backprojection

Simulation of Beam Hardening in Industrial CT with x-ray and Monoenergetic Source by Monte Carlo Code

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Convergent Incremental Optimization Transfer Algorithms: Application to Tomography

A new calibration-free beam hardening reduction method for industrial CT

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.

Scatter Correction Methods in Dimensional CT

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

BME I5000: Biomedical Imaging

Statistical Reconstruction Algorithms for Polyenergetic X-ray Computed Tomography

Central Slice Theorem

Consistency in Tomographic Reconstruction by Iterative Methods

Spectral analysis of non-stationary CT noise

1688 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 28, NO. 11, NOVEMBER We call the conventional CTACs single-energy computed tomography

Moscow-Bavarian Joint Advanced Student School 2006 / Medical Imaging Principles of Computerized Tomographic Imaging and Cone-Beam Reconstruction

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

MOTION of a patient induces artifacts such as streaks

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

DUAL energy X-ray radiography [1] can be used to separate

Algebraic Iterative Methods for Computed Tomography

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Optimal threshold selection for tomogram segmentation by reprojection of the reconstructed image

Bias-Variance Tradeos Analysis Using Uniform CR Bound. Mohammad Usman, Alfred O. Hero, Jerey A. Fessler and W. L. Rogers. University of Michigan

Statistical Image Reconstruction and Motion Estimation for Image-Guided Radiotherapy

Attenuation map reconstruction from TOF PET data

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 32, NO. 11, NOVEMBER

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

A Recursive Filter for Noise Reduction in Statistical Iterative Tomographic Imaging

Adaptive Reconstruction Methods for Low-Dose Computed Tomography

Carbondale, IL USA; University of North Carolina Chapel Hill, NC USA; USA; ABSTRACT

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Corso di laurea in Fisica A.A Fisica Medica 4 TC

arxiv: v1 [physics.med-ph] 18 Apr 2016

Enhanced material contrast by dual-energy microct imaging

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

Non-Stationary CT Image Noise Spectrum Analysis

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Image Acquisition Systems

Limited View Angle Iterative CT Reconstruction

Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm

Quantitative image-based spectral reconstruction for computed tomography

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

A prototype table-top inverse-geometry volumetric CT system

THE methods of maximum likelihood (ML) and maximum

Computer-Tomography I: Principles, History, Technology

A Curvelet based Sinogram Correction Method for Metal Artifact Reduction

Advanced Image Reconstruction Methods for Photoacoustic Tomography

MEDICAL IMAGING 2nd Part Computed Tomography

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

TOMOGRAPHIC reconstruction problems are found in

Research Article An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation

Feldkamp-type image reconstruction from equiangular data

Interior Reconstruction Using the Truncated Hilbert Transform via Singular Value Decomposition

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Transcription:

implified statistical image reconstruction algorithm for polyenergetic X-ray C omesh rivastava, tudent member, IEEE, and Jeffrey A. Fessler, enior member, IEEE Abstract In X-ray computed tomography (C), bony structures cause beam-hardening artifacts that appear on the reconstructed image as streaks and shadows. Currently, there are two classes of methods for correcting for bone-related beam hardening. he standard approach used with filtered backprojection (FBP) reconstruction is the Joseph and pital (J) method [1]. In the current simulation study (which is inspired by a clinical head scan), the J method requires a simple table or polynomial model for correcting water-related beam hardening, and two additional tuning parameters to compensate for bone. Like all FBP methods, it is sensitive to data noise. tatistical methods have also been proposed recently for image reconstruction from noisy polyenergetic X-ray data, [], [3]. However, these methods have required more knowledge of the X-ray spectrum than is needed in the J method, hampering their use in practice. his paper proposes a simplified statistical image reconstruction approach for polyenergetic X-ray C that uses the same calibration data and tuning parameters used in the J method, thereby facilitating its practical use. imulation results indicate that the proposed method provides improved image quality (reduced beam hardening artifacts and noise) compared to the J method, at the price of increased computation. he results also indicate that the image quality of the proposed method is comparable to a method requiring more beam-hardening information [3]. Index erms X-ray computed tomography, beam-hardening, Joseph-pital method, statistical reconstruction techniques. I. INRODUCION BEAM hardening causes artifacts in X-ray C images. hese artifacts manifest themselves as cupping due to water-related beam-hardening and as streaks or shadows due to bone-related beam-hardening. Current X-ray C systems are well-calibrated to compensate for water-related beam-hardening. Additional calibration for bone-related beamhardening is not required as use of a few tuning parameters in addition to the above calibration during the reconstruction phase (using the J method) is sufficient to compensate for bonerelated beam-hardening. As we shall see in a later section, the number of additional parameters required depends on the image viewing range and amount of soft-tissue and bone in the subject being scanned. Earlier statistical methods of [], [3] require more information about the X-ray spectrum and/or materials being imaged than is available to the standard J method. Consequently, these algorithms are less easily incorporated into the current X-ray C systems. his paper proposes a statistical upported in part by GE Healthcare echnologies and NIH grant CA87634. omesh rivastava and Jeffrey A. Fessler are with the Electrical Engineering: ystems Deptt., University of Michigan, Ann Arbor, MI 4819, UA. (E-mail: someshs@umich.edu, fessler@umich.edu). reconstruction algorithm that uses the same beam-hardening information as the J method. II. MODEL Polyenergetic X-ray C measurements for the ith ray through the object are assumed to be generated according to the following statistical model : { y i Poisson I i (E)e R } L µ(x,y,e)dl i de + r i, where, I i (E) is the energy spectrum of the incident X-rays, µ(x, y, E) is the X-ray attenuation coefficient for energy E at location (x, y) in the object, L i denotes the ith ray through the object and r i denotes scatter. We assume a two-material model (soft-tissue and bone 1 ) and express the X-ray attenuation coefficient as the product of mass attenuation coefficient, m(e) and material density, ρ(x, y). hus, we have, } y i Poisson {b i e fi(,i,b,i) + r i, (1) f i (,i, B,i ) Ii (E) log e ( m(e),i mb(e)b,i) de, b i,i L i ρ (x, y)dl [GI ρ] i, B,i L i ρ B (x, y)dl [GI B ρ] i, where, b i Ii (E)dE, ρ is a vector consisting of densities for each pixel, I and I B are diagonal matrices representing the image-domain masks for soft-tissue and bone respectively and G is the forward projection operator. Note that f i (, B ) is equal to the logarithm of the incident intensity of the X-ray spectrum to the exit intensity. In principle, this table could be measured using step phantoms. We call f(, B ) the water and bone correction table, and f(, ) f ( ) the water correction table. he method of [3] uses the entire water and bone correction table. his method will be called the Exact method from now on. he J method utilizes the water correction table and a few tuning parameters. he purpose of the additional parameters is to suitably extrapolate the water correction table to approximate the water and bone correction table within the tolerable error 1 he terms soft-tissue and water are used interchangably because the two materials have similar X-ray attenuation properties. he subscript i is dropped as each ray has the same incident energy spectrum.

1 6 4 3 B 1 Fig. 1. Plot of f(, B ). Units of and B are g/cm. Fig. 3. Image of the true phantom (14 14pixels). Window 4HU. 1.8 1.6 1.4 B 1 Fig.. Plot of γ(, B ). Units of and B are g/cm. range. he proposed method uses the same approximation as the J method. his paper is organized as follows. ection III describes the conditions under which f(, B ) can be approximated using two tuning parameters. ection IV describes the proposed method (the -parameter method) and other choices of statistical methods that can be used to eliminate noise and beam-hardening artifacts. he simulation setup is described in ection V and the results and a discussion of the results is presented in ection VI. III. APPROXIMAING HE WAER AND BONE CORRECION ABLE he water and bone correction table, f(, B ), can be computed exactly in simulations and is shown in Fig. 1. Note that the non-linearity of f(, B ) is the source of the beamhardening artifacts. In the current study, f(, B ) is computed for a 8kVp spectrum and two materials, soft-tissue and bone. We use the one-to-one nature of the water correction table 3 6 4 f in order to rewrite f(, B ) as : where, f(, B ) f( + γ(, B ) B, ) γ(, B ) f ( + γ(, B ) B ), { f 1 (f(,b)) B B, f lim 1 (f(, )) B. his formulation takes full advantage of the water correction table f. Fig. shows the plot of γ(, B ) corresponding to the f(, B ) shown in Fig. 1. he function γ(, B ) is equivalent to λ L defined in Eq.(13) of [1]. An accurate approximation of γ(, B ) by a function of two variables ( and B ) involving the tuning parameters is required for a good reconstruction. he number of tuning parameters needed for the approximation to γ(, B ) depends on the amount of tolerable error. he error tolerance depends on the image viewing window. We focus here on a 4 HU window (or.4g/cc in density units). he phantom also plays a role in the choice of number of tuning parameters. his is because the approximation to γ(, B ) has to be good over the range of and B determined by the amounts of soft-tissue and bone in the phantom. he phantom in this simulation study is designed based on a clinical case. In the current study, we use the following approximation for γ(, B ) : γ(, B ) A B B, () which is identical to Eq.() of [1]. he plot of γ(, B ) in Fig. suggests that the above approximation could be reasonable. It will be shown through simulations that this approximation works well for the head phantom. he thickness of bone mineral in the head phantom is around 1.mm.

Increasing the bone mineral thickness to about.mm causes this approximation to fail, necessitating the use of more tuning parameters. We consider only the 1.mm case in this paper. IV. AIICAL RECONRUCION MEHOD We compare the following statistical reconstruction methods. 1) 1-parameter method We use a constant to approximate γ(, B ) instead of using (), i.e., set B. his is equivalent to Eq.() of [1]. his method is used to verify that the amount of bone in the phantom is significant from the point of view of beam-hardening, and more than one tuning parameter is required for the approximation to be good. ) -parameter method his is the proposed method and uses () to approximate γ(, B ). 3) Exact method he exact value of γ(, B ), hence the exact value of f(, B ), is known in simulations. his method is almost identical to [3]. 4) Ad hoc method In this method, we perform the statistical reconstruction assuming a water-correction model and process the obtained image by the J method. We assume that the noise can be removed by including the water correction model (i.e. set A 1 and B ) only in the measurement model. Now, the J method is used to get rid of the beamhardening artifacts. his is not a systematically derived method but is attractive due to its simplicity. It is included here for the purpose of comparison with the -parameter method. A. 1-parameter and Ad hoc methods he negative log-likelihood for these methods are written as L(ρ) h i (f ([G 1 ρ] i )), i1 h i (t) (b i e t + r i ) y i log(b i e t + r i ), { G Ad hoc method G 1 G(I + AI B ) 1-parameter method. he function h i appears in Poisson transmission tomography problems and a quadratic surrogate for it was computed in [4]. he problem is regularized by a adding an edge-preserving (Huber) penalty function to the negative log-likelihood to create the final cost function. he cost function can be reduced monotonically by the use of quadratic surrogates [4]. Using the optimal curvatures computed in [4], the surrogate is expressed as Q 1 (ρ; ρ (n) ) h i (f ([G 1 ρ (n) ] i ))+ i1 h i(f ([G 1 ρ (n) ] i )) (f ([G 1 ρ] i ) f ([G 1 ρ (n) ] i ))+ 1 c i (f ([G 1 ρ] i ) f ([G 1 ρ (n) ] i )). Observations of f ( ) suggest that it is a concave function. Using Lemmas 3 and 4 of [4] we can prove the following : f (n) ( )( (n) ) f ( ) f ( (n) ), f ( (n) ) (n) (n) f ( ) f ( (n) ). Observations of f ( ) also suggest that f (n) f( () ). (n) hus, we have the final form for the surrogate : Q (ρ; ρ (n) ) h i (f ([G 1 ρ (n) ] i ))+ i1 h i (f ([G 1 ρ (n) ] i )) f ([G 1 ρ (n) ] i )[G 1 (ρ ρ (n) )] i + 1 c i (f ()) [G 1 (ρ ρ (n) )] i. A conjugate gradient method is used to reduce the value of Q (ρ; ρ (n) ) at every iteration. hese methods will not have nice convergence properties like monotonicity due to the beamhardening approximations made to the exact measurement model and the use of masks I and I B. A certain number of iterations are run and the iterate with the best image quality is taken as the final reconstruction. he 1-parameter method is initialized with a 1-parameter J reconstruction and the Ad hoc method is initialized with a water-corrected FBP reconstruction. B. -parameter and Exact methods he negative log-likelihood for these methods are written as L(ρ) h i (f([gi ρ] i, [GI B ρ] i )). i1 It might be possible to derive an exact -D surrogate for this negative log-likelihood along the lines of [4] but the procedure appears to be very complicated. A simpler way is to approximate f(, B ) by a local aylor series expansion for every pair of ( n, B n ) at the current iterate and hope that this approximation holds for the step sizes generated. f(, B ) f( n, n B)+ [ f 1, ( n, n B ) f,1 ( n, n B ) ] [ n B n B ].

he quadratic approximation can be now written as : Q 3 (ρ; ρ (n) ) K + (G ḣ) (ρ ρ (n) ) [ḣ] i + 1 (ρ ρ(n) ) G diag( c i )G (ρ ρ (n) ), ḣi(f([gi ρ (n) ] i, [GI B ρ (n) ] i )), G [ diag( f 1, ) diag( f,1 ) ] [ G G ] [ ] I. I B he problem is regularized by adding an edge-preserving (Huber) penalty function to the negative log-likelihood and a conjugate gradient method is used to reduce the value of the surrogate at every iteration. Due to the aylor series expansion and use of masks, these algorithms are not guaranteed to be monotonic. However, they have been found to perform satisfactorily in practice. he -parameter method is initialized by a -parameter J reconstruction and the Exact method is initialized by a J method using the water and bone correction table. 1-parameter and Ad hoc methods require one forward projection and one back-projection every iteration. he -parameter and Exact methods require two forward and back-projections every iteration. V. IMULAION EUP A -D fan-beam X-ray C scanner was simulated. he geometry of the scanner is similar to the central slice of a GE Lightspeed Pro scanner. inogram data was collected over a 36 rotation over a cm field-of-view (FOV). he sinogram dimensions were 984 angles by 888 bins per angle. An 8kVp spectrum was used and the blank scan count summed over the entire X-ray spectrum was 1.1 1 6 per bin. r i were set to zero for this study. he phantom size was 14 14 pixels and the reconstruction was done on a 6 6 grid. A 14 size grid permits the phantom to have an average bone mineral width of about 1.mm. Fig. 4. Fig.. Water-corrected FBP reconstruction. Water-corrected FBP reconstruction (zoomed in). VI. REUL AND DICUION he reconstruction from the water-corrected FBP method (Fig. 4 and Fig. ) is noisy and contains a streak artifact in the center of the image. he 1-parameter method (Fig. 6) provides a reconstruction which is free from noise but it still contains the streak. he reconstruction from the -parameter method (Fig. 7) is rid of the noise as well as the streak artifact. he image quality is comparable to that of the Exact method (Fig. 8). he Ad hoc method (Fig. 9) provides an image quality slightly worse than the -parameter method. he -parameter method should be the method of choice when compared to the Ad hoc method as it follows the measurement model more closely and uses the same beam-hardening information (water-correction table and tuning parameters). REFERENCE [1] P. M. Joseph and R. D. pital, A method for correcting bone induced artifacts in computed tomography scanners, J. Comp. Assisted omo., vol., pp. 1 8, Jan. 1978. [] B. De Man, J. Nuyts, P. Dupont, G. Marchal, and P. uetens, An iterative maximum-likelihood polychromatic algorithm for C, IEEE r. Med. Im., vol., no. 1, pp. 999 18, Oct. 1. [3] I. A. Elbakri and J. A. Fessler, tatistical image reconstruction for polyenergetic X-ray computed tomography, IEEE r. Med. Im., vol. 1, no., pp. 89 99, Feb.. [4] H. Erdoğan and J. A. Fessler, Monotonic algorithms for transmission tomography, IEEE r. Med. Im., vol. 18, no. 9, pp. 81 14, ept. 1999. ACKNOWLEDGMEN he authors would like to thank GE Healthcare echnologies for providing images and information about a clinical C scan of a human head and information about the GE Lightspeed Pro scanner.

Fig. 6. Reconstruction using the 1-parameter method. Window 4HU. Left, full image. Right, central part of the image. Fig. 7. Reconstruction using the -parameter method. Window 4HU. Left, full image. Right, central part of the image. Fig. 8. Reconstruction using the Exact method. Window 4HU. Left, full image. Right, central part of the image. Fig. 9. Reconstruction using the Ad hoc method. Window 4HU. Left, full image. Right, central part of the image.